Vous êtes sur la page 1sur 28

Service Manual

MMED6000DP-M7

Version: VS1.0M7
Issue date: 22/March

2010

Copyright
Beijing Choice Electronic Technology Co., Ltd. (hereinafter called Beijing Choice) owns all rights to
this unpublished work and intends to maintain this work as confidential. Beijing Choice may also seek
to maintain this work as an unpublished copyright. This publication is to be used solely for the purpose
of reference, operation, maintenance, or repair of Choice equipment. No part of this can be
disseminated for other purposes.
In the event of inadvertent or deliberate publication, Beijing Choice intends to enforce its right to this
work under copyright laws as a published work. Those having access to this work may not copy, use,
or disclose the information in this work unless expressly authorized by Beijing Choice to do so.
All information contained in this publication is believed to be correct. Choice shall not be liable for
errors contained herein nor for incidental or consequential damages in connection with the furnishing,
performance, or use of this material. This publication may refer to information and protected by
copyrights or patents and does not convey any license under the patent rights of Beijing Choice, nor
the rights of others. Beijing Choice does not assume any liability arising out of any infringements of
patents or other rights of third parties.
Content of the manual is subject to changes without prior notice.
BEIJING CHOICE ELECTRONIC TECHNOLOGY CO., LTD.
ALL RIGHTS RESERVED

Precautions:

Maintenance of the monitor must be performed by technical personnel who have been trained on
professional technology and have mastered the maintenance knowledge.

To prevent electrical hazard, do not handle the monitor in high temperature, high voltage, gas
fumigation, or soaking in liquid, cut of the main power before cleaning or sterilizing the monitor.

If the monitor is wetted accidentally during the operation or maintenance, please stop it and clean
and dry the affected parts.

While replacing the fuse, please replace with the fuse of the same specification and the same
rated current.

Before practical operation of the system, check all the cable connections for damage, and replace
the damaged cable and joints immediately. Before using the system, the operation must check
the system for correction working situation and operation condition.

If the monitor is not stored and used according to the stipulated environment conditions,
performance and precision of the system may be affected.

If the icon of alarm bell or sound switch is marked with X, there will be no sound alarming under
any circumstance.

Do not rely on the sound alarm system of the patient monitoring only. During the monitoring, when
the alarm volume is turned down or closed, the medical care personnel may not be notified when
the patient is dangerous. Please remember, the most reliable monitoring method is to combine
close personnel monitoring with correct operation of the monitoring equipment.

After connecting the monitor to the central monitoring system, please check the alarm system for
normal function.

When the monitor is connected with external equipments, the equipment shall accord with the
electric leakage indicator of IEC 60601-1 standard to prevent possible dangerous electric
leakage.

When the monitor is used continuously for a long time, please connect it to network power supply.
Please connect the monitor to network power supply as soon as the power down alarm rings
when using battery, otherwise, the monitor will be closed automatically to cause monitoring
interruption.

Wireless LAN equipment contains built-in RF radiator which may interrupt other medical
equipments including patient implanting equipments. Before installation and connecting new
medical equipment to the coverage of wireless LAN, please perform electromagnetic
compatibility test according to the instructions of Installation and Maintenance Manual for
Wireless LAN System.

The electrodes of lead shall not connect with other conductors and ground.

When the monitor is used with defibrillator, use special defibrillating leads to protect discharging
efficiency of the cardiac defibrillator, and during the defibrillating, do not touch the patient, leads,
instrument, and the table of placing the instrument.

Please refer to the User Manual for precautions of other operations.

Contents
CHAPTER 1. Introduction...................................................................................................1
1.1 General ..................................................................................................................................................... 1
1.2 Brief Description .................................................................................................................................... 1
1.3 Appearance introduction ..................................................................................................................... 1

CHAPTER 2. Function principle ......................................................................................4


2.1 Principle of hardware............................................................................................................................ 4
2.1.1 Power supply Module.................................................................................................................. 4
2.1.2 Main control board ....................................................................................................................... 5
2.1.3 Main function board (CSN608 board)...................................................................................... 5
2.1.4 Keyboard ........................................................................................................................................ 7
2.2 Principle of software ............................................................................................................................. 8
2.2.1 Real-time operating system....................................................................................................... 8

CHAPTER 3. Verification of monitoring parameter ..........................................................10


3.1 System inspection ............................................................................................................................... 10
3.1.1 Appearance and assembly inspection of the whole machine ........................................ 10
3.1.2 Start-up inspection..................................................................................................................... 10
3.1.3 Normal operation inspection................................................................................................... 10
3.1.4 Record ........................................................................................................................................... 10
3.1.5 Net working function ................................................................................................................. 10
3.1.6 Clock inspection......................................................................................................................... 10
3.2 Engineer menu...................................................................................................................................... 11
3.3 Parameter verification ........................................................................................................................ 11
3.3.1 Inspection of ECG and RESP parameter .............................................................................. 11
3.3.2 Inspection of NIBP parameters ............................................................................................... 11
3.3.3 SpO2 parameter inspection ...................................................................................................... 13
3.3.4 TEMP parameter inspection..................................................................................................... 13

CHAPTER 4. Service.................................................................................................................15
4.1 Disassembly diagrams ....................................................................................................................... 15
4.2 Fault clearance guidance................................................................................................................... 15
4.2.1 Fault of whole machine............................................................................................................. 15
4.2.2 Display fault ................................................................................................................................. 16
4.2.3 Operation, record, and net working faults ........................................................................... 16
4.2.4 Power panel fault........................................................................................................................ 16
4.2.5 Monitoring parameter fault ...................................................................................................... 16

CHAPTER 5. Monitor Installation ..........................................................................................18


5.1 Core connection of the monitor ....................................................................................................... 18
5.2 Casing inspection of the monitor .................................................................................................... 19
5.3 Electrifying inspection of the monitor............................................................................................ 19
5.4 Installation of printer........................................................................................................................... 19

CHAPTER 6. Basic Operation ................................................................................................20

6.1 Key pressing ......................................................................................................................................... 20


6.2 Screen display area ............................................................................................................................. 20

CHAPTER 7. Cleaning and Maintenance ............................................................................21


7.1 Cleaning ................................................................................................................................................. 21
7.1.1 Cleaning ........................................................................................................................................ 21
7.1.2 Sterilizing...................................................................................................................................... 21
7.1.3 Disinfecting .................................................................................................................................. 21
7.2 Maintenance .......................................................................................................................................... 21

CHAPTER 8. Net working ........................................................................................................22


8.1 Networking performance ................................................................................................................... 22
8.2 Use ........................................................................................................................................................... 22

Appendix 1 Keys brief Description.......................................................................................22


Appendix 2 Alarm instructions..............................................................................................23

Service Manual

VS1.0M7

CHAPTER 1. Introduction
1.1 General
This service manual is used for maintaining the monitor by technicians authorized by our company. This manual introduces the repairing and
maintenance of MMED6000DP-M7 monitor.

1.2 Brief Description


The vital signals which the MMED6000DP-M7 monitors include: ECG, SpO2, NIBP, RESP, TEMP etc. MMED6000DP-M7 multi-parameter
monitor adopts fine resolution TFT display displaying the measured data, waveform and parameters. This monitor can display 7 leads
waveforms and all monitoring parameters. Moreover, the monitor can be connected to the network and communicate real-time data with the
central monitoring system.

1.3 Appearance introduction

Front panel

Fig. 1-1
(1)

Indicator light: Indicating the power supplying state of the monitor


Green: power supplying

(2)

Power switch: Press the button for about 3/4 seconds to power on/off.

(3)

Indicator light: Indicating battery charging status


Orange and flash: the battery is in the process of quick charging.
Orange: the battery is slowly being charged or the charge is finished.

(4)

FREEZE
Pressing the button will freeze the waveform of ECG. Pressing FREEZE button again will restore the normal monitor status.

(5)

SILENCE
ALARM function button. When pressing the button, the icon in menu status is shown as

which indicates all the audible alarms have

been shut off. Once a new alarm event occurs, the audible alarm function will be activated automatically. Or the audible alarm will be
restarted automatically if the alarm event still exists after 3 minutes.You can also restore the audible alarm by pressing the button again.
The icon shown as

which indicates the audible alarm function is restored.

NOTE When mark appears, the system can not give the audible alarm prompt. Therefore, the operator should use this
function carefully.
(6)

NIBP
Start/Cancel NIBP measuring button. Pressing the button will inflate the cuff to start a new NIBP measurement. During the measurement,
press this button to cancel the measurement and deflate the cuff, including the CYCLE and STAT (NOT STOP) measuring mode.

(7)

PRINT
PRINT button. Pressing this button will motivate the recorder to output the results if the monitor is equipped with them.

(8)

MENU
Pressing the button will exit the submenu and refresh the screen. If the CF card is inserted, you can store the changed parameters setting
as configuration files in a CF card only by pressing the REFRESH button. Any setting is not refreshed in a CF card before you press the
REFRESH button.

Service Manual
(9)

VS1.0M7

Rotary Knob
The operator uses the rotary knob to select the menu item and modify the setup. It can be rotated clockwise or
counter-clockwise and pressed like other buttons. The operator uses the knob to realize the operations.
The method of using the knob to execute the operation:
Rotary knob is just like the cursor of computer. When operator rotates the knob on the icon where the operation is wanted,
the icon will be automatically highlighted. Then press the knob, operator will open the setup menu of the corresponding
parameter so as to set up the menu.

Fig. 2-2

(10) ALARM Indicator


If there is physiological alarm, it demonstrates that some parameters measurement is beyond alarm limitation, in the meantime, the
indicator begins flashing. There are three alarm-levels in all for selection:
High level: when alarm occurs, the color of the indicator becomes red and starts flashing once every one second.
Mid level: when alarm occurs, the color of the indicator becomes orange and starts flashing once every one second.
Low level: when alarm occurs, the color of the indicator becomes green and starts flashing once every one second.

Side Panel

Fig. 1-2
(1). NIBP socketconnect NIBP cuff with extension tube.
(2). Temperature socket1: connecting the TEMP probe.
(3). Temperature socket2: connecting the TEMP probe.
(4). ECG socketConnect ECG 5-lead wire with 6 PIN connector.
(5). SpO2 socketConnect the SpO2 sensor with extension cable.
(6). Battery cover.

CF card and Recorder

Fig. 1-3
(1)

CF card interface: CF card is a memory. Plug a CF card in this interface. If the CF card has been inserted before power on, the
configuration files (including all the set parameters) and the measurement data stored in the CF card will be read after power on. During
the operation, the measured data will be automatically stored in the CF card. You can store the changed parameters setting as
configuration files in the CF card only by pressing the REFRESH button.
NOTE:
z

DO NOT plug and insert the CF card when the monitor is working!

Any setting is not refreshed in a CF card before you press the REFRESH button.

Service Manual
(2)

VS1.0M7

A thermal array recorder with standard 50mm (+1/-1) wide printout paper is used for the portable patient monitor. About the detailed
information, please refer to Section 6.3 Print Setup.
NOTE: The thermal array recorder should be installed by the adequate technician.

Rear Panel

Fig. 1-4
(1) Power socket AC Power supply: 100-240 (VAC), 47-63(Hz); Fuse: standard T 2.0A 250(VAC)
(2) DC power supply (reserved)
(3) Equivalent electric ground access for connecting with the hospitals grounding system.
(4) NET access point: connecting with the central monitoring system through the standard RJ-45.

Service Manual

VS1.0M7

CHAPTER 2. Function principle


2.1 Principle of hardware
This section introduces the functions and operating principle of every function board of the MMED6000DP-M7 multi-parameter monitor.

2.1.1 Power supply Module


Power supply module of MMED6000DP-M7 patient monitor composed of two parts: AC-DC Power Supply Module and CSN 512 power supply
board. The principle of power supply module is shown in the fig.2-1

Figure 2-1 Block diagram


1. Isolated Switching Power Supply Module
The AC-DC Power Supply Module supplied power supply board CSN512. Input voltage: AC90V~260VOutput voltage: DC14.85~15.15V
Output Continuous Current : 2.8A.
2.CSN512 power supply board
AC-DC Power Supply Module supplies CSN512 power supply board and charges the battery though the CSN512. The board CSN512
composed of switching circuit, fan control, audio control and charging circuit. And the Temperature sensor is positioned on the CSN512 board.
The voltages of the three DC currents which are output from the switching circuit are 12V, 6V and 5V. Those currents supply fan controlling
circuit and printer, audio controlling circuit and CSN608 board. The temperature sensor DS1820 is positioned on the CSN512 board and its
signal is acquired by keyboard. The method of cooling for the monitor is fanning.
3. Key point of the power supply board
The voltages of key points of the power supply board are discripted as bellow:
Key point

Reference Voltage

Pin 34 of P2

DC15V5%

Pin 34 of P9

DC6V5%

Pin 34 of P11

DC5V5%

Pin 2 of P4

Actual voltage of battery

4.Dimension of CSN 512:

Figure 2-2
5.Interfaces of CSN 512:

Figure 2-3
CSN512 power supply board supplies the following components including: CSN720 mainboardCSN608 functional boardkey board
rechargeable batterylaudspeakerfuninverterprinter etc.

Service Manual

VS1.0M7

The functions of the components above are descripted as the following table:NoteThe square pin of the interface is the first.
NO.

Function

NOTE

P1

Interface to keyboard

P2

Input of the power supply board

To the output of the AC-DC Power Supply Module

P3

Interface of fan and loudspeaker control

To the key board

P4

Interface of the battery.

P5

Interface of the laudspeaker

P6

Input of audio signal

P7

Interface of the fan

P8

Interface of the printers power supply

1pin: 5V2pin: ground3pin: 12V

P9

Power supply interface of CSN608

1&2pin:ground3-4pin:6V

P10

Power supply interface of inverter2Pin

Not connect

P11

Power supply interface of CSN720

1&2pin:ground3-4pin:5V

P12

Interface of the temperature sensor

To key board

To the mainboard CSN710

2.1.2 Main control board


The main board and functional boards and key board communicate with each other via serial port.
The layout of CSN720 is shown as below

Figure 2-4
The functions of the main board are described as follows: NoteThe square pin of the interface is the first.
NO.

Function

Note

J3

Interface of LCD

To key board

J4

Serial port for functional board CSN608

3-Line TTL serial port

J5

Audio interface

To power supply board CSN512

J6

Serial port for key board

3-Line TTL serial port

J7

Serial port for printer

3-Line TTL serial port

J10

Interface to power supply board CSN512

DC 5V

CON2

Signal interface, To CF card reader and networks card adapter

2.1.3 Main function board (CSN608 board)


CSN608 main function board is composed of ECG/ RESP/TEMP module, NIBP and SPO2 modules. This section introduces the working
principle of these functional modules in detail.
(1). ECG/RESP/TEMP module
ECG/RESP/TEMP module is composed of five parts, including ECG amplification circuit, respiration circuit, temperature circuit, control and
data acquisition circuit, isolation power supply and isolation photocoupler circuit. ECG amplification circuit is divided into three parts, including
ECG pre-amplification grade, channel I and II of ECG main amplification. Its functional block diagram is shown as figure 2-5.

Service Manual

VS1.0M7

Figure 2-5 Functional block diagram of ECG/RESP/TEMP module


ECG pre-amplification grade includes signal buffering circuit and lead-off detection circuit. The cardiac electricity signal from the body will flow
to the lead-off detection circuit through the signal buffering circuit to judge whether the lead off or not, and then flow to the main amplification
channel of cardiac electricity and be amplified and filtered under the control of the MCU so as to obtain ECG signal. The main amplification
channel II of ECG includes pacing pulse detection circuit which is to detect the pacing pulse signal superimposed in the ECG signal. Thoracic
Impedance signal from the body is regulated by the preliminary respiration circuit and then transferred the post respiration circuit for being
demodulated, amplified, and filtered, so as to obtain the respiration wave signal. The body temperature circuit can amplify and filter the input
signal by the body temperature probe and output a voltage signal corresponding to the temperature. This circuit is composed of sampling
switch, constant current source and filter. MCU of the control and data acquisition circuit converts the analog signals, including the ECG signal,
respiration wave signal and temperature signal, to digital signals. And it can carry out date acquisition, ECG circuit control (including lead
switching, gains selection, filtering selection and lead off status reading), body temperature circuit control(selection of probe type and channel),
as well as communication with the main system, acquisition the control signal and upload waveform data and status; the isolated power supply
and isolated photocoupler circuit convert the +6V power from the main system into 8V analog +3.3V, and digital +3.3V power supplies by the
ECG/RESP/TEMP circuits through the DC/DC convertor, and realize electric isolation. The data of ECG/RESP/TEMP is transmitted through
the isolated photocoupler circuit and the serial interface, and then realize isolated transmission of signals.
Besides, the ECG/RESP/TEMP module has three interfaces, among which P200 is the ECG sensor interface used for connecting ECG Lead
cable; P201 is the body temperature probe interface; P10 is the serial interface connecting with the main control board and P11 is the input
power supply from the power board.
(2) NIBP module
NIBP module is composed of blood pressure circuit, one charging motor, and two deflation valves. The functional block diagram is as shown in
figure 2-6.

Figure 2-6 Functional block diagram of blood pressure measuring


Among which, the analog circuit is divided into cuff pressure, pulse wave detection and cuff overpressure protection circuit. The digital circuit is
the control component composed of Z80180 and its peripheral devices. The pressure sensor converts the current air pressure into voltage
signal which will be amplified by the amplification then the voltage signal is transferred to A/D converter.
Core device of the circuit is Z80180 which communicates with the main control CPU through the serial interface and accomplishes functions
such as blood pressure measuring according to the order of CPU and transfers the measured results and working condition to the main control
CPU for display. Z80180 controls the motors start and switch of the air relief valve through the peripheral circuit and monitors the pressure
amplification circuit and motor operation circuit through A/D chip.
The input of the functional board supplies the charging motor and deflation valve, and then generates +5V power through the LDO on the
circuit board and acts as working power supply of the whole blood pressure system.

Service Manual

VS1.0M7

(3) SPO2 module


SPO2 module includes CPU, photoelectricity driver, signal processing circuit and power supply circuit. The block diagram of SpO2 module is
shown in the figure below. CPU of SpO2 module accomplishes internal circuit control, AD sampling and computation, as well as
communications with serial interfaces of 608 main control boards CPU. CPU uses the C8051F005 produced by XinHua Dragon, with a working
clock of 11.0592MHz. The single chip C8051F005 contains a 32K FALSH memory, a 2304 bites RAM, 12-bite AD converter and 2-way
12-position D/A converter. The photoelectricity driver, under the control of the single chip, produces driving current for driving the LED inside
the SpO2 sensor. The photoelectricity signal produced by the sensor will be amplified by the differential amplification circuit first, and then be
amplified again after filtering the direct component. Finally, the AC components are amplified by the biasing amplification circuit (U105A) to
obtain sampling signal which will be transferred to the AD input end of U100 for AD switching. The power supply circuit used the 3.3V digital
power supply on board 608. The SpO2 board has only one external interface, namely the SpO2 sensor interface (P11), and it communicates
with 608 main control boards CPU through the serial interface. Its working functional block diagram is shown in Figure 2-7.

Figure 2-7
(4) Interfaces and descriptions of CSN608
Positions and dimensions of the interfaces of the 608 function board are shown in the figure below:

Figure 2-8
The figure above shows the design of initial electric input and output of CSN608 module, and the table below describes all the pins and status
of the connections. (Note: the square pin of all interface sockets is no. 1):
No.

Function

Remark

P10

Serial port of functional board

pin1:TXD; pin2:RXD; pin3:GND

P11

Interface of power supply

Pin1,2: VDD; pin3,4:GND

P200

Interface of ECG cable

Connect with the side panel

P201

Interface of TEMP cable

Connect with the side panel

P105

Interface of SpO2 cable

Connect with the side panel

J1

Connect with deflating valve

Connect with the pump valve board

J6

Connect with deflating valve

Connect with the pump valve board

J2

Connect with inflating pump

Connect with the pump valve board

2.1.4 Keyboard
The keyboard of MMED6000DP-M7 monitor includes: keyboard identification circuit, function control circuit and driving circuit. This section
introduces the operating principle of the keyboard in detail.
The keyboard is fixed on the front shell and has six keys, S2, S4, S5, S7 and S8 and the input S1 of the knob coder. The keyboard adopts
Atmega48 single chip produced by ATMEL Company as the processing core to accomplish the function of keyboard scanning, charging control,
key code data sending and receiving, and indicator driving through related peripheral circuit.
Signals from the key-press and knob are transmitted to the main board after being identified by the single chip through the serial interface. The
sound and alarm are controlled by the single chip according to the serial interface data received from the main board. The keyboard sends
control signals to the power panel through P6 and P3. P5 is the temperature sensor wire which is connected to power panel P12. P2 is the
interface of function indicator, and P4 is the serial interface of keyboard.

Service Manual

VS1.0M7

Figure 2-9 Functional block diagram of the key board


Interfaces of the keyboard and their functions are described as below:
No.

Function

Remark

P1
P2
P3
P4
P5
P6
P7
J1J2

Program download port of the key board


Interface of double color alarm indicator
Interface of fan and loudspeaker control
Serial interface of keyboard
Interface of temperature sensor
Keyboard control interface
Interface of inverter
Interface of LCD display

Use to upgrade the monitors program


Connect with front panel
Connect with power board
Connect with main board
Connect with power board
Connect with power board
Connect with inverter
Connect with LCD display

2.2 Principle of software


2.2.1 Real-time operating system
This sytem is an embedded real-time single task system supported by real-time operating system, and characteristics of the application
procedure operated include:
(1) Real time
(2) Response capability of asynchronous event
(3) Interuption management capability
(4) switching time and time interuption defering shall be definite

2.2.2 General information of the software system


(1). Data flow diagram:

Figure 2-13 Block diagram of data flow

Service Manual

VS1.0M7

(2) Task division


Data devision task
Data processing task
Arrhythmia analysis task
Interface processing task
Task of sending control command
Output task of the recorder
Network sending task
Network reception task
Task of refreshing the new waveform regularly
(3)

Task design

Data division task: revert the data recieved by the serial interfaces to original real data and put the data to corresponding data region, such
as waveform data region, parameter data region, and status data region.
Data processing task: accomplish computation of the parameters and alarm for distinguishing, if there is any alarm, the system forms an
alarm data region, and in the meantime form a short tendency and long tendency database.
Arrhythmia analysis task: perform arrhythmia analysis store the waveform of arrhythmia and corresponding information.
Interface processing task: accomplish human-computer interactions, process system faults, bounce out menus, modify alarm limits and
alarm switch of various parameters, and processing of alarm disply and the system fucntion settings, setting of information of the patients,
display of breath oxygenation diagram, display of abnormal analysis waveform, switching of the display interfaces, history data of the EVENT,
and tendency data, and drive the task of sending control commands and the task of recorder.
Task of sending control command: send out various control commands, such as: version number reading command, lead switching of the
multi-functional board, gain changing, filtering means changing command, start and stop measuring command of NIBP, and send alarm
information and sound information to the panel.
Output task of the recorder: accomplish real-time waveform and parameter recording and parameter alarm recording.
Network sending task: send the waveform, parameter, maximum and minimum alarm value, and other related information of this bed to the
central station.
Network reception task: send the modified maximum and minium alarm values and related settings (such as: analog demonstration, NIBP
measuring switch, mode, channel, and lead selection, etc.) of the central station to the monitor.
Task of refreshing the waveform regularly: fresh the waveform in real time.

CHAPTER 3. Verification of monitoring parameter


3.1 System inspection
This section will introduce the system inspection of MMED6000DP series monitors, including: appearance and assembly inspection of the
whole machine, start-up inspection, configuration normal operation inspection, normal work inspection of the recorder, net working inspection,
clock inspection, ECG output inspection, long-time operation inspection of the monitor, and safety detection.
3.1.1 Appearance and assembly inspection of the whole machine
(1) There is no besmirchment, obvious scoring, breakage, and crack on casing of the machine, and assebly is firm, without loose part or
broken pieces, etc.
(2) Operaton of keys is fine, and the interfaces function are normal.
(3) Silk on surface of the instrument is complete and correct.
(4) Accessories are complete and perfect.
(5) Perform vibration test to the whole machine before other tests.
3.1.2 Start-up inspection
After powering on, the monitor works normally and there is no abnormality on the screen and no instruction of error message; check whether
waveform scanning of all channels are normal and whether parameters refreshing and clock are normal.
3.1.3 Normal operation inspection
(1) Operate every key, the monitor should act or respond normally according to the Operatiors Manual, and abnormal operation shall not
cause crash.
(2) Revolve and press the knob, the monitor should act or respond normally according to the User Manual, and abnormal operation shall
not cause crash.
(3) Operate the monitor according to Operators Manual, check whether the sound of key pressing and each level of alarms are normal,
whether the volume control is normal and whether the alarm silence and alarm inhibit function are effective.
(4) Inspection of the recorder.
3.1.4 Record
The recorder shall record according to the conditions and content specified in the User Manual, including:
(1) Record shall be correct and accord with the waveform or other records displayed on the screen.
(2) Recording shall be stopped immediately by pressing the stop key during recording.
(3) Faults of hardware of the recorder can be indicated on the screen.
(4) It works normally for a long time and no paper jamming happens.
3.1.5 Net working function
(1) No matter which one is turned on first, the centration station or the monitor, net working shall work and number of the bed shall be
correct.
(2) Data receiving and sending of the monitor and centration station are normal.
(3) Waveform parameter, alarm limit, related information about NIBP measuring, alarm switch, and ECG lead receiving by the central station
shall be correct.
(4) The alarm limit, related information about NIBP measuring, alarm switch, upper and lower time synchronization, and ECG lead returned
from the central station shall be correct.
(5) After the monitor is turned off, response of the centration station shall be correct.
(6) When any connection of the network is cut off, the centration station can indicate that this bed is off line, and network and operation will
be normal after recovering the connection.
3.1.6 Clock inspection
Correct indiction, even running, without sudden change, without stop.
3.1.7 If operate the monitor for a long time, check whether there are any circumstances such as crash or scanning suspension of the wave.
3.1.8 Factory menu setting of the monitor
Except for the factory setting menu, set the values as initially uploaded by the software.
3.1.9 Safety inspection
(1) Grounding leakage current detection
Normal: a voltage with the effective value of 250V is applied between the short circuit of fire wire and zero wire and grounding wire of the power
supply, with the current less than 0.5mA.
Single failure: a voltage with the effective value of 250 is applied between the fire wire and grounding wire of the power supply, with the current
less than 10mA.

Service Manual

VS1.0M7

(2) Leakage current detection: including three monitoring parameters, ECG, SpO2, and TEMP. Take ECG as an example.
Normally, connect all the leads of ECG and apply a voltage with the effective value of 250V between the leads and ground, with the current not
less than 10A.

3.2 Engineer menu


Under the monitoring status, revolve the knob to locate the cursor at the icon of system setting

, and then press the knob, the menu

bouncing out at bottom of the screen is as below: engineers with different rights enter with different codes, the codes are: 2222 and 7134.
3.3.1 Enter the main menu and select [EXTEND] [SYSTEM], and then select the code 2222, refers to figure 3-3click [PW-ENTER] to enter
the system menu. There are four options, including DEMO, ECG CAL, USERPW and SETUP.

TIME

4
VOLUME

PRINT

COLOR

AUTO
FAN EXTEND

RETURN

Figure 3-1

0
VER608

RESET

SYSTEM

BED

MESSAGE

OTHER

RETUN

Figure 3-2
Entering password: revolve the knob to select the item to be set, and then press the knob, characters of the selected item will become white
and change the value in this item by revolving the knob. After setting the code, press [PW-ENTER] to complete the setting and enter the
system menu, refers to figure 3-3 and 3-4.

1ST

2ND

3RD

4TH

ENTER

RETUN

Figure 3-3

OFF
DEMO

OFF
ECGCAL

SETUP

USERPW

RETURN

Figure 3-4
In the [SETUP] menu, the user can set DEF", LAN, HzCho, TEM-H and TEM-L.
DEFDefault settings.
LAN: EN or . Users can set the display language is English or Chinese.
HzCho50Hz or 60Hz Users can set the display language is 50Hz or 60Hz.
TEM-H and TEM-LWhen the switch of the fan is auto, if the temperature of the monitor is beyond the alarm limits, the fun will run and
the alarm will be heard until the temperature is normal.

3.3 Parameter verification


To ensure acuracy of the monitor, inspect and calibrate the parameters regularly, generally once in a year.
3.3.1 Inspection of ECG and RESP parameter
(1) Inspection Device
Physiological signal simulator
(2) Inspection procedures
. Connect the physiological signal simulator with the ECG interface of the monitor.
. Check the number of ECG waveforms on the screen of monitor, and ECG module menu shall accord with the selections in the factory
menu.
. Check whether the ECG waveform and RESP waveform are normal.
. Set parameters of the simulator: HR=30 (gains4), RR=120
. Check the waveforms of ECG and RESP and check whether the values of HR and RR are the same as the parameters set by the
simulator.
. Check whether the monitor will make report immediately when the ECG lead falls off.
3.3.2 Inspection of NIBP parameters
(1)

Inspection Device
NIBP stimulator

(2)

Inspection procedures

Use the NIBP stimulator with calibration function to calibrate and judge the acuracy of blood pressure pump. After finishing the calibration,
perform the inspection below:
A. Both the simulator and monitor are set to the adult mode.
B Select a group of blood pressure value within the measuring range on the NIBP stimulator, such as: S=90;

M=70;

C Check whether the acutual results measured by the monitor are consistent with the value set by the stimulator.
D Change the blood pressure value set by the stimulator, and measure again.
E Check whether the acutual results measured by the monitor are consistent with the value set.

11

D=60

Service Manual

VS1.0M7

(3) Software calibration


Under the monitoring status, revolve the knob to locate the cursor at the position of NIBP in the block of blood pressure parameters within
the digital display area.

Figure 3-5
And then click the knob, the menu will bounce out at the bottom of the screen as below.

mmHg MANUAL
0
ALARM UNIT CYCLE MODE

0
NON-S

RETURN

Figure 3-6

SYS
Click Alarm to enter the alarm setting menu, and then select SYS to adjust the maximum and minimum of the systolic pressure alarm to the
most.

SYS

MEAN

DIAS

OFF
ALARM

RETURN

Figure 3-7

255

220

HIGH

LOW

RETURN

Figure 3-8
Return to [ALARM] menu, there are the QUOTIETY and OFFSET whose ranges are -10~10 and used to calibrate the systolic pressure, the
principle is: SBP(after)= SBP (before)+ SBP(before)XQUOTIETY/100+ OFFSET.

mmHg
0
ALARM UNIT SQUTY

0
SOFFS

OFF
NON-S

RETURN

Figure 3-9

MEAN

Enter the NIBP menu and click ALARM to enter the alarm menu, and then select MEAN to adjust the maximum and minimum of the mean
pressure alarms to the most as Figure 3-11.

SYS

MEAN

OFF
DIAS

ALARM

RETURN

Figure 3-10

220

200

HIGH

LOW

Figure 3-11
Return to the [ALARM] menu, there are the OFFSET whose ranges are
CUFF(after)=CUFF(before)+OFFSET.

SYS

MEAN

DIAS

OFF
ALARM

RETURN

-50~50 and used to calibrate the CUFF. The principle is

OFF
IBRAT

0
OFFSET

RETURN

Figure 3-12
Return to the [NIBP] menu, there is the QUOTIETY and OFFSET whose ranges are -10~10 and used to calibrate CUFF, the principle is:
MBP(after)= MBP(before)+ MBP(before)XQUOTIETY/100+OFFSET.

ALARM

mmHg
0
UNIT MQUTY

0
MOFFS

OFF
NON-S

RETURN

Figure 3-13

DIAS

Enter the NIBP menu and click ALARM to enter the alarm menu, and then select DIAS to adjust the maximum and minimum of the diastolic
pressure alarm to the most as figure 3-15

SYS

MEAN

DIAS

OFF
ALARM

RETURN

Figure 3-14

220

180

HIGH

LOW

RETURN

Figure 3-15
Return to the previous menu again. And there are the QUOTIETY and OFFSET whose ranges are -10~10 and used to calibrate the diastolic
pressure, the principle is: DBP(after)= DBP (before)+ DBP(before)XQUOTIETY/100+ OFFSET.

ALARM

mmHg
0
UNIT DQUTY

0
DOFFS

Figure 3-16

12

OFF
NON-S

RETURN

Service Manual

VS1.0M7

3.3.3 SpO2 parameter inspection


(1) Inspection tool
SpO2 stimulator
(2) Inspection procedures
A.

Connect the SpO2 stimulator with the monitor SpO2.

B.

Set parameters of the SpO2 stimulator to be: SpO2=98

PR=70

Check whether the values of SpO2 and PR of the monitor are consistent with the stimulator.
(Note: for the value of PR, heart rate source must be SPO2 in the blood oxygen menu.)
D.

(3)

Change the values of SpO2 and PR of the SpO2 stimulator.

E.

Check whether the values of monitor are consistent with the values set.

F.

Pull out the SpO2 probe, the monitor will make report.

Software calibration

Under the monitoring status, revolve the knob to locate the cursor at the position of SPO2 in the block of SPO2 parameter within the digital
display area.

Figure 3-17
And then click the knob, the menu will bounce out at the bottom of the screen. Operations refer to Figure 3-18 to 3-19.

3
AMPL

ECG
HRFROM

ALARM

OFF
PR SOU

FILL
WAVE

RETURN

Figure 3-18

OFF
PROBE

PR AL

100
HIGH

80
LOW

OFF
ALARM

RETURN

Figure 3-19
Adjust SpO2 upper limit to the maximum 100, and the lower limit to the maximum 99. Then return to previous menu of SPO2, there is the
OFFSET whose range is -55 and used to calibrate the value of SpO2. The principle is: SpO2(after calibration)= SpO2(before calibration)+
deviation of SpO2.

3
AMPL

0
OFFSET

ALARM

OFF
PR SOU

FILL
WAVE

RETURN

Figure 3-20
3.3.4 TEMP parameter inspection
(1) Inspection Device
Physiological signal stimulator
(2) Inspection procedures

(3)

A.

Connect one end of the temperature sensor with the stimulator and the other with the TEMP port of monitor.

B.

Set the stimulator: TEMP=34.

C.

Check whether the value of TEMP displayed on the screen of monitor is 34.

D.

Change the value of stimulator: TEMP=40.

E.

Check whether the value of TEMP displayed on the screen of monitor is 40.

Software calibration

Under the monitoring status, revolve the knob to locate the cursor at the position of TEMP in the block of TEMP parameter within the digit
display area.

Figure 3-21
And then click the knob, the menu will bounce out at the bottom of the screen. Operate according to Figure 3-32 and 3-33.

Cel
UNIT

ALARM

RETURN
Figure 3-22

45.0
HIGH

44.5 MEDIUM
2.0
LOW ALARM DI_LMT

OFF
TEMPDI
Figure 3-23

13

RETURN

Service Manual

VS1.0M7

Adjust the upper limit and lower limit of body temperature to the most, 45.0 and 44.5 respectively. Return to current menu and enter T1 OFFS
and T2 OFFS whose ranges are -1.01.0 and used to calibrate the value of body temperature. The principle is:
T1 (after) = T1 (before)+ T1 Cal;
T2 (after) = T2 (before) + T1 Cal.

39.5
HIGH

36.0
LOW

MEDIUM
ALARM

0.0

0.0

T1 Cal

T2 Cal

Figure 3-24

14

RETURN

Service Manual

VS1.0M7

CHAPTER 4. Service
4.1 Disassembly diagrams

Figure 4-1

Figure 4-2
According to the different function boards used, the core can be installed with: blood oxygen board, invasive blood pressure board, and CPT
template, one or two of the three. Installing position is as shown in the figure. If only one function board is used, it shall be installed at the
position indicated on the left preferentially, and if two function boards are used, they shall be installed at the positions indicated on the left and
right.

4.2 Fault clearance guidance


Faults may be caused to the MMED6000DP-M7 monitor due to various factors (such as instable of grid voltage, vibration, impact, and aging,
etc.) during transportation, storage, and use and the monitor fails to work normally. In such cases, professional personnel with certain
experiences in electronic medical instrument maintenance shall perform component level overhaul with reference to the classification of faults
below. The component level overhaul is to position the fault to components of the instrument, such as power panel, master control board, TFT
components, measuring cable or the function boards, by analysis, replacement, and trial operation of the components. Maintenance of the
components, namely maintenance of the elements and devices, shall be carried out by experienced engineer of the manufacturer with special
instrument and equipment and under specific environment and conditions.

4.2.1 Fault of whole machine


Phenomenon

Possible Reasons

Solution

There is no display when the


machine is turned on, and the
power indicator lamp does not
brighten up and the fan does not
run.

The fuse is damaged

Replace the fuse

Power supply is damaged

Replace the power supply

Other components are shorted

Find out the shorted component and replace it

Failure of the power panel

Replace the power panel

There is no display when the


machine is turned on or the screen
is black while working, however
the power indicator lamp brightens
and the fan works normally

Failure of the master control board


or display.

Turn to the display fault confirmation and replace


corresponding component.

Backlight inverter of the screen is


damaged.

Replace the inverter

Error in data communication between


the master control board and function
board.

According to the error indication, replace the master control


board, patch panel or corresponding function board to
confirm the fault.

Character display is normal while


waveform is intermittent

15

Service Manual

VS1.0M7

An operation or measuring
function fails

The master control board or


corresponding component is
damaged.

Replace the master control board

The keyboard is damaged or fails.

Replace the keyboard

Instant strong interruption of the


electricity grid.
Performance of the power panel is
poor.
Performance of the master control
board is poor.
Connection of the power supply
and master control board is poor.

Accidental crash

Check the power supply and grounding system


Replace the power panel
Replace the master control board
Replace or repair the connector

4.2.2 Display fault


Phenomenon

Possible reasons

Power supply of start-up is normal, but


there is no display or the screen
becomes black during normal work

Backlight inverter is
damaged
Connecting wire of the
LCD is poor
Master control board is
damaged

Solution
Connect external VGA display to confirm the fault
Repair or replace the wire of LCD
Replace the master control board

4.2.3 Operation, record, and net working faults


Phenomenon
Keys or the rotary encoder
fails
The sound is hoarse or
fails

The recorder does not print

Paper feeding of the


recorder is deflective
Net working fails

Possible reasons

Solution

Keyboard or rotary encoder is


damaged

Replace the keyboard or rotary encoder

Wiring of the keyboard fails

Replace or repair the wire of keyboard

The keyboard fails

Replace the keyboard

The horn or wiring fails

Replace the horn or wiring

Audio amplifier of the power panel fails

Replace the power panel

There is no paper in the recorder or the


paper bail has not been pressed

Load paper for the recorder and press the paper bail

The recorder fails

Replace the recorder

Drive power supply of the recorder fails

Replace the power panel

Wiring of the recorder fails

Replace or repair wiring of the reorder

Installing direction of the printing paper


is incorrect
Installation and positioning of the
recorder is poor

Reinstall the printing paper


Adjust installation of the recorder

Network wiring fails

Check and repair the network wire

The master control board fails

Replace the master control board

4.2.4 Power panel fault


Phenomenon

Possible reasons

Solution

The fuse is burnt when turning on the machine.

The power supply or other parts


are shorted

Turn on the machine for further inspection

When all the loads are disconnected, the fuse is till


burnt

Power supply fails

Replace the power supply

The fuse is burnt when connecting a component

This component is shorted

Replace this component

+12V DC power
damaged

Replace the power panel

Power indicator lamp and indicator lamp of the


master control board brighten up, but the fan does
not run and indicator lamp of the patch panel does
not brighten up
Indicator lamp of the master control board does not
brighten up, but the fan runs normally and indicator
lamp of the patch panel brightens

supply is

+5V DC power supply is


damaged

Replace the power panel

4.2.5 Monitoring parameter fault


Phenomenon

There is no cardiac electricity wave

Possible cause

Remedy

Contact of the electrode pad of


cardiac electricity is poor

Use a new electrode pad and ensure good contact

RL electrode is hanging in the air

Connect RL electrode

CSN608 board is damaged

Replace CSN608 board

Connecting means of the electrode

Connect the measuring electrode pad correctly

16

Service Manual

VS1.0M7
is incorrect

Cardiac electricity wave is abnormal or


interrupted

There is no breathing or the breath wave


is abnormal

Measured value of body temperature is


incorrect

There is electrode pad hanging in


the air
The AC power supply has no
ground wire

Remove the unused electrode pad


Use three-wire power supply and ground it, or
connect the grounding terminal to the ground

Filtering means of ECG is incorrect

Use a suitable filtering means

The 608 function board is poor

Replace the CSN608 board

The electrode is in poor connection

Reinstall the electrode to correct the connecting


position

The patient moves frequently

Make the patient quiet

The CSN608 board is poor

Replace the CSN608 board

The measuring probe is in poor


connection

Connect and fix the body temperature probe

The CSN608 board is poor

Replace the CSN608 board

Heart rate is inaccurate and abnormal


and analysis of ST section is incorrect

Measuring wave of cardiac electricity


is poor

Adjust the connection to make the measuring wave


normal

NIBP can not be charged

The air pipe is folded or leaks

Adjust or repair the air pipe

Blood pressure can not be measured


once in a while

The cuff is loose or the patient moves

Make the patient stay still and tighten the cuff

Size of the cuff is not fit for the


patient

The CSN608 board fails

Replace the CSN608 board

The probe or blood oxygen module


is damaged

Replace the probe and check the fault

The blood oxygen wave is interrupted


greatly

The patient is moving

Make the patient quiet

Ambient light is too strong

Weaken ambient light

Measured value of SpO2 is incorrect

The patient is injected with stain

Clear the stain before measuring

Error of blood pressure measurement is


big
There is no blood oxygen wave

17

Use a suitable cuff

Service Manual

VS1.0M7

CHAPTER 5. Monitor Installation


5.1 Core connection of the monitor

Figure 5-1
NO.

From

TO

Type

NO.

From

TO

Type

P9 of power supply board

P11of CSN608

CS6.644.1236A

P201 of CSN608

Rear shell T1T2

CS6.644.1149B

P11 of power supply board

Lanyu mainboard

CS6.644.1374

P200 of CSN608

Rear shell ECG

CS6.644.1151C

P7 of keyboard

J10 of converter

CS6.644.1373

J1 of CSN608

Pump cable

CS6.644.1249

P8 of power supply board

J6 of printer

CS6.644.1369

P2 of keyboard

Alarm lamp board

CS6.644.1115

P1 of power supply board

P6 of keyboard

CS6.644.1237

P5 of keyboard

P12 of power supply board

CS6.644.1245

P4 of keyboard

J6 of Lanyu mainboard

CS6.644.1371

CON2 of Lanyu mainboard

Networks board

CS6.644.1375

P3 of power supply board

P3 of keyboard

CS6.644.1246

J1 of keyboard

Line of LCD

provide for oneself

P4 of power supply board

Battery

CS6.644.1372 21

J2 of keyboard

J3 of Lanyu mainboard

CS6.644.1363

P2 of power supply board

power adapter

CS6.644.1240 22

CON2 of Lanyu mainboard

CF card reader

provide for oneself

AC-DC Module

220V power socket

CS6.644.1260A 23

P7 of power supply board

Fan

CS6.644.021F

J7 of Lanyu mainboard

J4 of printer

CS6.644.1368 24

P5 of power supply board

Laud Speaker

CS6.644.021F

J2J6 ofCSN608

Valve cable

CS6.644.1248 25

J4 of Lanyu mainboard

P10 of CSN608

CS6.644.1370

P105 of CSN608

SpO2 on Rear shell

CS6.644.1365A 26

P6 of power supply board

J5 of Lanyu mainboard

CS6.644.1244

18

Service Manual

VS1.0M7

5.2 Casing inspection of the monitor


Check the name, quantity and specification of each item according to the packing list inside the packing case. If there is any discrepancy,
contact the service department of our company.

5.3 Electrifying inspection of the monitor


Perform necessary inspection before electrifying the monitor:
(1). Check whether voltage of the electricity grid accords with the operating voltage of the monitor.
(2). The monitor must use three-term power line and three-term socket.
(3). When the monitor is used with other medical equipments at the same time, connect the equal electric potential terminals of the
monitor with those of other equipments effectively and reliably.
(4). Prevent liquid from entering the monitor
Check whether display on the screen is correct after electrifying the monitor.

5.4 Installation of printer


1. Insert the power cable and serial port cable to the corresponding sockets on the printer. Refer to figure 5-2.
2. Put the printer in the monitor correctly and fix it with two M312 screws, as shown in figure 5-3

Fig. 5-2

Fig. 5-3

Fig. 5-4 Printer cable Connection

19

Service Manual

VS1.0M7

CHAPTER 6. Basic Operation


6.1 Key pressing
Look the monitor from the front side, there are several keys on the keyboard of the monitor:
(1). FREEZE (freeze/unfreeze key)
Under normal monitoring, press this key to stop all wave refreshing on the screen; under freezing status, press this key to recover refreshing of
all waves.
(2). SILENCE (alarm suspension/reset key, mute)
Press this key for a short moment: alarm suspended for 3 minutes, and counting down of 3 minutes' alarm will be displayed on upper right corner
of the screen.
(3). PRINT (record/stop)
Press this key to start a real-time recording; during recording, press this key to stop this recording.
Length of recording can be chosen through printing length in the menu of printing setting.
(4) NIBP (start measuring/stop measuring):
Under non-measuring status: press this key to charge the cuff and start a manual blood pressure measuring.
Under measuring status: if this measurement is to be cancelled due to some reason, press this key to stop the measurement and discharge.
Note: press this key during continuous measuring, not only this blood pressure measurement is given up, but the continuous blood pressure
measurement is cancelled;
(5) MENU
Refresh the screen manually, press this key in any menu to exit the menu and return to the display interface of main screen.

6.2 Screen display area


The screen is divided into three parts, indication area, waveform area, and parameter area.
(1). Indication area
This area is on tht top of the screen and for display bed No., name of patient, system time/date, regular alarm and indication information.
(2). Waveform area
The waveform area can display 7 waves at most. When the seven leads of ECG are displayed on the same screen, upper left corner of each
wave is identified with name.
(3). Parameter area
The parameter area is on the right of the screen and noted with names of the parameters. Use the rotary button to select these parameters to
enter corresponding menu and work on the items of the menu to achieve the purpose of the operator. Alarm information of corresponding
parameter will be displayed in corresponding parameter area or waveform area, NIBP will display such information as monitoring mode, unit,
measuring mode, and measuring time, and TEMP will display unit.

20

Service Manual

VS1.0M7

CHAPTER 7. Cleaning and Maintenance


7.1 Cleaning
Cleaning of MMED6000DP-M7 monitor shall be carried out according to the stipulated method. Recommended material and agent shall be
used to prevent the monitor from being polluted and damaged or even bringing bacteria that may endanger peoples health. In the meantime,
the user shall not use deteriorated or damaged lead, sensor and accessories to protect the patient and medical personnel.
7.1.1 Cleaning
Cleaning of MMED6000DP-M7 monitor shall:
(1). Use such cleaning instruments as soft cotton
(2). For the mainframe and power supply box, soap, amido or ethanol based cleaning material can be used.
(3). For reusable cardiac electricity lead, blood pressure sensor, blood oxygen probe, and body temperature probe, soap or amido based
cleaning materials can be used.
Requirements for detergent:
Ammonia: diluted ammonia<3%
Ethanol: Ethanol 70%, isopropyl alcohol 70%
Note: do not use cleaning stuff made of friction materials, such as wire brush or lapping agent, etc. Liquid shall not enter the instrument or
soak the instrument.
7.1.2 Sterilizing
Sterilize the monitor when necessary, and pay attention that:
(1). For the mainframe and power supply box, ethanol and acetaldehyde based detergent shall be used;
(2). Acetaldehyde based detergent shall be used for the cardiac electricity lead;
(3). Ethanol and acetaldehyde based detergent shall be used for reusable cuff, blood oxygen probe, and body temperature probe.
Requirements for bactericide:
Ethanol: ethanol 70%
Acetaldehyde: formaldehyde (35%-37%)
Bleacher: diluent of sodium hypochlorite (bleacher for washing), with the concentration of 500ppm-5000ppm.
Note: Prevent the sterilizing detergent from entering or soaking the instrument and accessories.
Prevent the sterilizing liquid from staying on surface of the instrument.

7.1.3 Disinfecting
Disinfect the instrument when necessary based on regular cleaning. Disinfection must be performed strictly according to the standards of the
hospital and pay special attention to safety. Generally, EtO gas can be used for disinfection.

7.2 Maintenance
In order to give play to the best performance of MMED6000DP-M7 monitor, ensure accuracy of long-term work of the instrument, and prevent
various measuring errors, the monitor shall be maintained properly. In the meantime, the monitor shall be cleaned, sterilized and disinfected
regularly to prevent cross infection.
Maintenance of the monitor includes:

Check the monitor, leads, sensor, and accessories regularly;

Clean the monitor as required;

Perform safety test once in year;

Calibrate NIBP and body temperature parameter one in a year;

Perform function test to the whole machine once in a year;

Perform safety test after the machine is opened for repairing;

21

CHAPTER 8. Net working


8.1 Networking performance
(1). Longest distance (through a hub) between the monitor and central station can be up to 100M;
(2). Longest time between start-up to successful connection is 40 seconds;
(3). Deferring of online data <5 seconds;
(4).The monitor has the Plug&Play function, namely the network joint connected with the MMED6000DP-M7 portable monitor can be
plugged and pulled out dynamically and it is unnecessary to turn off the power supply.

8.2 Use
(1). Technicians of the company shall be responsible for layout analysis, wiring, and installation of the network system;
(2). Turn on the machine for trial connection after connection of the net line is proved to be proper;
(3). If there is any fault during the use, refer to Section 4.2 for fault handling.

Appendix 1 Keys brief Description


1.1 REFRESH (return to the main screen)
Press this key to refresh the main screen or return to the main screen. Press this key under any menu, the system will stop current operation
automatically and return to the main interface. Press hits key in the main interface, and system will refresh the whole screen automatically.

1.2 FREEZE (freeze/off-frozen key)


Under normal monitoring, press this to stop all wave refreshing on the screen; under freezing status, press this key to recover refreshing of all
waves.

1.3 SILENCE (alarm suspension/reset key, mute)


Press this key for a short moment: alarm suspended for 3 minutes, and countdown of 3 minutes' alarm will be displayed on upper right corner of
the screen.
Alarm suspension can clear the alarm mark, namely: when alarm suspension is over, the cleared alarm will no longer generate alarm
indication. Only when new alarm condition happens, there will be sound-light alarm indication.
When alarm is suspended, the system will not perform any physiological alarm identification, namely not generating any physiological
alarm indication.
When alarm is suspended, the old technical alarm will still be displayed on the screen. However, the newly generated technical alarm
will stop the suspension of the system automatically and recover sound-light alarm. Volume of the alarm accords with the setting of menu.
When alarm is suspended, press this key again for a short while to stop the alarm suspension and recover normal alarm status.

1.4 PRINT (record/stop key)


Press this key to start a real-time recording; during recording, press this key to stop this recording.

1.5 NIBP (start measuring/stop measuring)


Under non-measuring status: press this key to charge the cuff and start a manual blood pressure measuring.
Under measuring status: if this measurement is to be cancelled due to some reason, press this key to stop the measurement and discharge.
Note: press this key during continuous measuring, not only this blood pressure measurement is given up, but the continuous blood pressure
measurement is cancelled;

Service Manual

VS1.0M7

Appendix 2 Alarm instructions


Alarm instructions mainly include: technical alarm information during normal use, physiological alarm instruction, function safety alarm, and
machine status information.
When power of the battery is low, icon of the battery power becomes red for warning. When the temperature is too high, the temperature
becomes red for warning, and when parameters are higher than the upper limit or lower than the lower limit, the parameters will become
different colors (red, blue, white) which are divided into three levels. Alarms by strings are as shown in the table below:
Type

Chinese string

English string

Remark

ECG

Error

ERROR

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Lead off

LDOFF

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Signal low

SINGAL LOW

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Ventricular

VFIB

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Atrial fibrillation

AFIB

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Pulse arrest

ASYST

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Tachycardia

TACHY

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Room speed

VENT TACHY

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Bradycardia

BRADY

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Beat lost

MISS BEAT

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Frequency PVC

FREQ PVC

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Running PVC

RUN PVC

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Trigeminy

TRIG

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Biogeminy

BIG

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

RonT

RonT

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Insert PVC

INTER PVC

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Paired PVC

COUPL

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

PVC

PVC

Displayed in the ECG alarm area, ECG parameter display area or ECG waveform area

Overpressure

OVERPRESSUR

Displayed at the left of top of the screen

fibrillation

NIBP

E
RESP

SPO2

RECORD

Printer busy

PRINTER BUSY

Displayed in the RESP alarm areaRESP parameter display area or RESP waveform area

Printing stop

PRINTER STOP

Displayed in the RESP alarm areaRESP parameter display area or RESP waveform area

Printer error

PRINTER ERROR Displayed in the RESP alarm areaRESP parameter display area or RESP waveform area

Alarm off

ALARM OFF

Displayed in the SpO2 alarm areaSpO2 parameter display area or SpO2 waveform area

Help call

HELP CALL

Displayed in the SpO2 alarm areaSpO2 parameter display area or SpO2 waveform area

Printing

PRINTING

Displayed at the right of top of the sreen

Printing finished PNT STOP

Displayed at the right of top of the sreen

23

Vous aimerez peut-être aussi