Académique Documents
Professionnel Documents
Culture Documents
PAPER
Introduction
Many important decisions are based on the results of chemical
quantitative analysis. The results are used, for example, to check
materials against specifications or statutory limits and to estimate yields and monetary value. Whenever decisions are based
on analytical results, it is important to give some indication of the
quality of the results.
In some sectors of analytical chemistry, introducing quality
assurance (QA) measures is required by law in order to prove
that laboratories are capable of and are providing data of the
required quality. Such measures include: the use of validated
methods of analysis, the use of defined internal quality control
procedures, participation in proficiency testing schemes,
accreditation based on ISO 170251 or establishing traceability
of the results of the measurements. Further information on
analytical QA procedures is provided by ref. 2.
As a consequence, chemists are, on their part, coming under
increasing pressure to demonstrate the quality of their results and
particularly to demonstrate their fitness for purpose by giving
a measurement of the confidence that can be placed on the result.
They have to include the degree to which a result would be expected
to agree with other results or with other own results, normally
irrespective of the analytical method used. One of the useful
methods for this purpose is the measurement of uncertainty.2
The definition of the term uncertainty (of measurement) used
in this work and adopted for the International Vocabulary of
Basic and General Terms in Metrology3 is:
Experimental
Instrumentation
A manual SPME device (Supelco, Bellefonte, PA, USA) equipped with a fused silica fiber coated with a 100 mm film of poly(dimethylsiloxane) (PDMS) was used for extraction in all
experiments. GC separation was carried out on a HP-5890 gas
chromatograph equipped with a 15 m long, 0.53 mm id, 1.5 mm
film thickness DB-1 capillary column. The end of the GC column
was connected to a 0.5 m long, 0.32 mm id deactivated fused
silica capillary. Both the DB-1 column and the fused silica
capillary are located in the GC oven and are subjected to the
same heating conditions. The fused silica capillary was driven
through a pyrolysis tube before being coupled to a PSA 10.750
(PS Analyticals, Orpington, U.K.) atomic fluorescence (AFS)
mercury detector. Argon was used as the carrier gas. The split/
splitless injector of the GC was fitted with a 0.75 mm id (SPME)
liner and was used in splitless mode.
The schematic diagram of the instrumental setup is presented
in Fig. 1. The operational parameters of the GC-pyrolysis-AFS
system are shown in Table 1.
For sample preparation, a Realsonic RS 16-F ultrasonic bath
(Realsonic, Budapest, Hungary) operated at 50 W (dm3)1
ultrasonic power was used. The prepared samples were centrifuged with a Hettich Mikro 22R centrifuge (Hettich, Tuttlingen,
Value
100 kPa
190 C
100 C, 0.6 min
30 C min1
250 C, 1 min
800 C
4.85 0.42
0.151 0.036
5.12 0.16
0.152 0.013
a
Each sample was digested in triplicate and standard addition was used for quantification. Sample preparation procedures differed depending on the
MeHg content (see text for details). The expanded uncertainties of the measured values are expressed as the 95% confidence interval (see Table 5, Table 6
and text for details).
Number of
measurements
Mean, xi
Standard
uncertainty, u(xi)
c/pg mL1
V1/mL
V2/mL
V10/mL
L10/mL
L50/mL
V3/mL
6
6
6
6
6
6
202.75
0.9989
0.9989
9.9589
9.9839
49.9672
0.9989
Not quantified
0.0061
0.0061
0.0948
0.0332
0.1457
0.0061
Type B estimation
Source of
uncertainty
xi a
Type of
distribution
Standard
uncertainty
m/g
0.2500 0.0001
Rectangular
5.77 105
Number of
measurements
Mean, xi
Standard
uncertainty, u(xi)
c/pg mL1
V1/mL
V2/mL
V10/mL
L10/mL
V5/mL
V3/mL
6
6
6
6
6
6
63.56
0.9989
0.9989
9.9589
9.9839
4.9712
0.9989
Not quantified
0.0061
0.0061
0.0948
0.0332
0.0487
0.0055
Type B estimation
,
cV1 V2 V10
,
V1
y
L10 V5 m 106
V3
(2)
Source of
uncertainty
xi a
Type of distribution
Standard
uncertainty
m/g
0.2500 0.0001
Rectangular
5.77 105
ci
vy y xi uxi yxi
vxi
uxi
(4)
Accordingly, the u(y,xi) is just a difference between the values of y calculated for [xi + u(xi)] and xi, respectively (eqn (5)),
and it denotes the uncertainty in y arising from the uncertainty
in xi.
u(y,xi) y(xi + u(xi)) y(xi)
(5)
Table 5 The spreadsheet method and the experimental steps in order to determine the uncertainty of SPME, derivatization, digestion and calibration
processes in the case of the BCR 464 sample (see text for details)
c
xi
u(xi)
V1
202.75
c
V1
V2
V3
V10
L10
L50
m
y/mg g1
u(y.xi)
u(y.xi)2
4.84800
4.82115
V2
V3
V10
L10
0.9989
0.0061
0.9989
0.0061
0.9989
0.0061
9.9589
0.0948
9.9839
0.0332
49.9672
0.1457
0.2500
0.00006
202.75
1.005
0.9989
0.9989
9.9589
9.9839
49.9672
0.2500
202.75
0.9989
1.005
0.9989
9.9589
9.9839
49.9672
0.2500
202.75
0.9989
0.9989
1.005
9.9589
9.9839
49.9672
0.2500
202.75
0.9989
0.9989
0.9989
10.0537
9.9839
49.9672
0.2500
202.75
0.9989
0.9989
0.9989
9.9589
10.0171
49.9672
0.2500
202.75
0.9989
0.9989
0.9989
9.9589
9.9839
50.1129
0.2500
202.75
0.9989
0.9989
0.9989
9.9589
9.9839
49.9672
0.25006
4.85059
0.02685
7.21104
4.81869
0.00259
6.73 106
4.88656
0.02931
8.59 104
4.86424
0.03856
1.49 103
4.86226
0.01624
2.64 104
4.84696
0.01426
2.03 104
0.00104
1.09 106
u y
L50
V1
63.56
c
V1
V2
V3
V10
L10
V5
m
y/mg g1
u(y.xi)
u(y.xi)2
0.15120
0.15037
V2
V3
V10
L10
V5
0.9989
0.0061
0.9989
0.0061
0.9989
0.0061
9.9589
0.0948
9.9839
0.0332
4.9712
0.0487
0.2500
0.00006
63.56
1.005
0.9989
0.9989
9.9589
9.9839
4.9712
0.2500
63.56
0.9989
1.005
0.9989
9.9589
9.9839
4.9712
0.2500
63.56
0.9989
0.9989
1.005
9.9589
9.9839
4.9712
0.2500
63.56
0.9989
0.9989
0.9989
10.0537
9.9839
4.9712
0.2500
63.56
0.9989
0.9989
0.9989
9.9589
10.0171
4.9712
0.2500
63.56
0.9989
0.9989
0.9989
9.9589
9.9839
5.0199
0.2500
63.56
0.9989
0.9989
0.9989
9.9589
9.9839
4.9712
0.25006
0.15128
8.34 104
6.95 107
0.15029
8.44 105
7.12 109
0.15241
9.11 104
8.29 107
0.15171
1.21 103
1.45 106
0.15269
5.11 104
2.60 107
0.15117
1.49 103
2.22 106
2.90 105
8.42 1010
u y
b
u y Su yXi 2 , where the xi are the parameters: SPME,
u(yderivatization) 4.41 103 mg g1
.
derivatization, V1, V2, and V10
3 step. Investigation to determine the uncertainty of the digestion process
The RSD%
of the peak areas was 10.0993, consequently the u(y) 1.53 102 mg g1 dry weight.
q
u y
In the first step, the reproducibility of the SPME-desorptiondetection processes was investigated. For this purpose, in all
cases one previously derivatized solution (10 mL of acetate buffer
+ 1 mL of sample solution + 1 mL of NaBPh4 reagent solution,
see Analytical procedure section) was extracted and measured six
times. The standard deviation of the peak areas is derived just
from the uncertainty of SPME and desorption process in the
injector part of the GC and the GC-AFS system in this case.
Based on the RSD% of peak areas, the u(y) arising only from
the uncertainty of SPME-desorption-detection (u(ySPME)) was
determined. This value can be seen in Table 5 and Table 6
marked with an a.
In the second step, the uncertainty of the derivatization
process was studied. Six solutions were prepared (10 mL of
acetate buffer + 1 mL of sample solution + 1 mL of NaBPh4
reagent solution) from all samples and following derivatization,
these solutions were extracted and measured. The standard
deviation of the peak areas is derived now from the uncertainty
of the derivatization process, the above mentioned SPMEdesorption-detection processes and the V1, V2 and V10 sources.
Using eqn (5), the u(y,V1), u(y,V2) and u(y,V10) could be calculated, the u(ySPME) was already known and the u(y) was also
known (the relative standard deviation of peak areas was
1234 | J. Anal. At. Spectrom., 2009, 24, 12291236
Fig. 2 Uncertainties in MeHg determination in the case of measurements made with the BCR 464 sample.
Fig. 3 Uncertainties in MeHg determination in the case of measurements made with the TORT-2 sample.
The u(y,xi) values shown in Fig. 3 are taken from Table 6 and
refer to the measurements made with the TORT-2 sample.
Based on our results, it can be concluded that irrespective of
the sample concentration, the alkaline digestion has the strongest
influence on the reproducibility of the measurement results while
the calibration has the second strongest influence on it.
As for digestion, there are a lot of important parameters in
connection with the use of an ultrasonic bath during the sample
preparation, for example the number of samples prepared
simultaneously, the volume of the samples, the volume of water
in the bath, the position of the samples in the bath, etc. The
effects of the treatment-time and the applied temperature were
investigated during method development, the influence of the
other factors were not studied. Consequently, the large uncertainty of the digestion process may come from this fact.
The uncertainty of the applied standard addition calibration
probably derives from the fact that separate standard addition
solutions were applied for each individual digested sample
although the matrix was in all cases the same. In this way, the
measurements were over-calibrated.
Since the uncertainty of the calibration process was established
on the basis of experimental results, all uncertainty sources
associated with the calibration process has been taken into
account by calculation. Apart from this fact, it may be interesting
to determine the uncertainty of the purity of the calibrant within
the uncertainty of the calibration process.
The purity of the MeHgCl standard is given on the
certificate as 0.9840 0.0050 with a level of confidence
95%, so the standard uncertainty of purity is given by
0:0050
uPurity
0:00255. Based on the model equations
1:96
(the purity is a multiplier factor in the equation) and using eqn
(5), the u(yPurity) is 1.24 102 mg g1 in the case of the BCR 464
sample, and 3.86 104 mg g1 in the case of the TORT-2 sample.
These values denote the uncertainty in y (MeHg as Hg content,
mg g1 dry weight) arising from the uncertainty in the purity of
the calibrant.
The u(ycalibration) was found to be 1.17 101 mg g1 in the case
of the BCR 464 and 9.92 103 mg g1 in the case of the TORT-2
(see values marked with a d in Tables 5 and 6), so it can be seen
how the u(yPurity) contributes to the u(ycalibration) in the two cases.
Moreover, it can be concluded that the SPME was a more
critical step regarding reproducibility in the case of the low
content TORT-2 sample compared to BCR 464, while the
uncertainty of the derivatization process proved larger in the case
of the high content BCR 464 sample. In the case of TORT-2,
the higher methanol and matrix concentration, while in the case
of BCR 464 the potential precipitation reaction between potassium and sodium tetraphenylborate can explain the above
mentioned symptom.18
According to our expectations, the volume and weight
measurements during sample preparation were less dominant
processes in this respect.
Finally, it should be emphasized that this investigation tended
to determine the uncertainty arising from the reproducibility, the
uncertainty associated with the determination of the bias was not
taken into account.
The recovery is always an important question, its study is
often a complicated problem.
J. Anal. At. Spectrom., 2009, 24, 12291236 | 1235
Conclusion
The uncertainty statement of the developed mercury speciation
analytical method was evaluated and is presented in this paper.
The presented study was based on an investigation executed in
4 steps. During the experiments, the overall uncertainty determined experimentally was considered as the combined standard
uncertainty and, with knowledge of this value, the uncertainty
of the individual analytical steps was calculated according to
GUM. By this means, it became clear how the uncertainty of the
individual analytical processes contributes to the experimentally
determined overall uncertainty. Consequently, this study
provides an opportunity to find the steps in a multi-step procedure that have the largest contribution to the overall uncertainty.
Measurements applied to one of the two CRM samples represented analysis of samples with a high MeHg content (a few mg
g1), measurements with the other CRM represented analysis of
samples with low (a few ng g1) MeHg content.
As a result of the completed experiments, it can be concluded
that the alkaline digestion and the calibration are the most critical steps regarding reproducibility irrespective of the measured
CRM sample. Phenylation derivatization, SPME and volume
and weight measurements during sample preparation are less
dominant processes in this respect.
Knowing the uncertainty of each step, critical analytical
processes can be planned more carefully in the future.
Safety
Organomercurials are very toxic, therefore, inhalation or skin
contact must be avoided. These compounds must be handled
with the use of appropriate personal protection equipment and in
an adequately ventilated environment.
References
1 ISO/IEC 17025: 1999. General Requirements for the Competence of
Calibration and Testing Laboratories., ISO, Geneva, 1999.
2 M. Thomson et al., Analyst, 1995, 120, 2934.
3 International Vocabulary of basic and general terms in Metrology.
(ISBN 92-67-10175-1), ISO, Geneva, 1993.
4 Eurachem/CITAC Guide: Quantifying Uncertainty in Analytical
Measurement, 2000, ed. S. L. R. Ellison, M. Roesslein and A.
Williams, http://www.eurachem.org/.
5 Guide To The Expression Of Uncertainty In Measurement. (ISBN 9267-10188-9), ISO, Geneva, 1993.
6 M. M. Storelli, A. Storelli, R. Giacominelli-Stuffler and
G. O. Marcotrigiano, Food Chemistry, 2005, 89, 295300.
7 Mercury update: Impact on Fish Advisories, EPA-823-F-0111,
U.S.EPA, Washington, DC, 2001, http://www.epa.gov/.
8 Comission Regulation (EC) No. 221/2002 of 6 February 2002
amending Regulation (EC) No 466/2001 setting maximum levels for
certain contaminants in foodstuffs, Official Journal of the European
Communities, L 37/4, 7.2.2002, Brussels, http://eur-lex.europa.eu/
LexUriServ/LexUriServ.do?uriOJ:L:2002:037:0004:0006:EN:PDF.
9 SCAN (Scientific Advisory Commitee on Nutrition), Report: Advice on
fish consumption: benefits&risks, 2004, http://www.sacn.gov.uk/
reports_position_statements/reports/advice_on_fish_consumption_
benefits__risks.html.
10 A. I. C. Ortiz, Y. M. Albarran and C. C. Rica, J. Anal. At. Spectrom.,
2002, 17, 15951601.
11 Y. Cai and J. M. Bayona, J. Chromatogr., A, 1995, 696, 113122.
12 C. M. Tseng, A. DeDiego, F. M. Martin, D. Amouroux and
O. F. X. Donard, J. Anal. At. Spectrom., 1997, 12, 743750.
13 I. R. Pereiro, A. Wasik and R. Lobinski, J. Anal .At. Spectrom., 1998,
13, 743747.
14 S. R. Segade and C. Bendicho, J. Anal. At. Spectrom., 1999, 14, 263
268.
15 Z. Mester, J. Lam, R. Sturgeon and J. Pawliszyn, J. Anal. At.
Spectrom., 2000, 15, 837842.
16 S. Dez and J. M. Bayona, Talanta, 2008, 77, 2127.
17 A. M. Caricchia, G. Minervini, P. Soldati, S. Chiavarini, C. Ubaldi
and R. Morabito, Microchem. J., 1997, 55, 4455.
18 Z. J
okai, L. Abranko and P. Fodor, J. Agric. Food Chem., 2005, 53,
54995505.
19 C. R. Quetel, J. P. Snell, Y. Aregbe, J
okai Zs., L. Abrank
o,
C. Brunori, R. Morabito, W. Hagan, S. Azemard, E. Wyse,
V. Fajon, M. Horvat, M. Logar, O. Donard, E. Krupp, J. Entwisle,
R. Hearn, M. Schantz, K. Inagaki, A. Takatsu, P. Grinberg,
S. Willie, H. Hintelmann, E. Bj
orn, E. Blanco Gonzalez,
G. Centieno, J. Ignacio Garcia Alonso and A. Sanz-Medel, J. Anal.
At. Spectrom., 2005, 20, 10581066.
20 Y. Aregbe and P. D. P. Taylor, Metrologia, 2006, 43(Tech. Suppl.),
08006.