Vous êtes sur la page 1sur 8

Problems Using Flat Reference Blocks for

Calibrating Ultrasonic Systems for Testing


Cylindrical Material
by Kenneth H. Beck*
It is a common industrial practice to use flat reference blocks for
testing cylindrical materials. This article points out a number of
problems with this approach and provides an analytical explanation
for the inadequacy of this practice. This article also demonstrates
the importance of selecting proper transducers and reference
blocks. It should be of interest to anyone involved in the ultrasonic
testing of cylindrical materials.
G.P. Singh
Associate Technical Editor

Introduction

It has been the practice for many years in some industries to use the response from flat
reference blocks to standardize the sensitivity and focal conditions for the testing of
cylindrical material. Some specifications and procedures such as MIL-STD-2154 (US
Department of Defense, 1982) indicate that this is at least acceptable when applied to
material over 100 mm (4 in.) in radius. This practice may be justified where years of
essentially empirical experience have demonstrated that all important discontinuities are
discovered in a given product by its employment. However, there are a number of
problems to this approach which should preclude its incorporation in a document
intended to prescribe a general practice for such tests. Some have been discussed
elsewhere (Beck, 1999) and others are listed below.

Most specifications for testing materials for critical end use specify that
reference standards be made from material of the same composition,
surface finish and heat treatment as the parts to be tested. There is a very
small likelihood of these conditions being met by a flat reference standard
when test material is cylindrical.

This represents a compromise with good practice that should be used only in
special circumstances.

To ensure reliable test results, most specifications for other than manual
contact tests require that the rejection sensitivity be verified by initial and
periodic checking of the reference standard at the full scanning speed
used for the test. Depending upon the nature of the scanning system
employed for the cylindrical material, this requirement may be difficult or
impossible to meet with flat reference blocks
Most importantly, dimensions and focal conditions of a transducer selected
to optimize response from a target at a certain depth in a flat reference
block bear no relation to those needed by a transducer for testing of
cylindrical material at the same depth. This point is demonstrated by the
following analysis. Furthermore, transducer selection based on the use of
a flat reference block would obviate transducer dimensional selection
procedures such as are described in Beck (2003).

Analysis
An example of the inadequacy of transducer selection for cylindrical materials testing at
a given depth, based on optimization of its characteristics to maximize response from a
target at the same depth in a flat reference standard, is demonstrated by the following
analysis.
Refer to Figure 1, which shows, in simplified form, the beam from a transducer focused
for maximum response from a target at a depth d in a block with a flat surface.

Figure 1 - Normalized transducer focused at a given depth in a flat block.


The focus of the refracted beam as shown in this figure may be achieved by a lens, by a
phased array or may be due to the natural near field convergence of the beam from a
flat transducer. As shown, it is an approximation of actual focal conditions because the
wave front from transducers is never "flat" in the optical sense and because the ratio of
transducer element size to wavelength is so much smaller than is required for this
pseudooptical representation (Krautkramer and Krautkramer, 1990). However, the use
of short pulse lengths smoothes amplitude variations in the near field (Singh and Rose,
1982) and allows this representation to adequately demonstrate the conditions to be
described.
Equation 1 expresses the depth d from the top surface of a flat block to the approximate
location of the refracted beam focus point in terms of the focal length Lf of the
transducer (which may be the Y0+ [last near field maximum] distance of a flat
transducer), the water path length Lw and the ratio K of the sound wave velocity in the
material to the velocity in the couplant. Since greatest beam intensity is achieved at this
point, it is assumed that d represents the center of the depth range to be examined.

(1)

The water path must be selected to place the first water path multiple beyond the back
surface of the block. Since a cylindrical bar of radius R is to be tested, the minimum
water path length is given by Equation 2.

(2)

Substituting this in Equation 1 yields for the flat block:

(3)

Solving this for the proper transducer focal length for maximum intensity at depth d
results in Equation 4.

(4)

The expression equivalent to Equation 1 for the depth of focus in a cylinder (Beck,
1991) with the Equation 2 substitution is:

(5)

The equation for minimum focal length required to obtain normalized focus at any depth
d in a solid cylinder may be obtained from Beck (1991), with the substitution for water
path given in Equation 2 above as:

(6)

Example
If it is desired to have maximum beam intensity at the center of a cylinder of 127 mm (5
in.) radius (d = R = 127 mm [5 in.]) and the velocity ratio K is assumed to be 4, the
required minimum water path length is found from Equation 2 to be:

If the calibration is based on optimizing on a flat block, the focal length (or Y0+) of the
transducer is found from Equation 4 to be:

If this focal length is substituted in Equation 5, the actual depth for maximum beam
intensity is found to be:

The minus sign indicates that the focus is really a virtual one above the top surface.
This condition is shown in Figure 2. This obviously does not produce the desired
maximum intensity at the center of the bar.

Figure 2 - Dispersion of refracted rays in a round bar from a transducer focused in a flat
block.

The proper focal length for center focus in a bar of 127 mm (5 in.) radius is obtained
from Equation 6 by substituting d = R. This yields:

This is seen to equal the water path length plus the material radius (Lw = R), which is
obviously correct for a transducer focused at the center of a round since all incident rays
enter the surface perpendicular to it (that is, along radii) and so have zero incidence
angle and are therefore not refracted.

The disparity between focal length determined by optimization on a flat block and that
determined by a round one becomes even worse at smaller diameters and is not
negligible until very large diameters are considered. For instance, to optimize at a depth
of 254 mm (10 in.) in a flat bar Lf = 1143 mm (45 in.) while the Lf for focus at the center
of a 508 mm (20 in.) diameter bar is 381 mm (15 in.).
Large transducer elements are needed to produce the long focal lengths required for
optimization at long distances in a flat block. (For example, element diameter = 39 mm
[1.45 in.] for Y0+ = 1143 mm [45 in.] at 5 MHz and 55 mm [2.16 in.] for operation at 2.25
MHz.) If available, elements of these diameters would produce poor signal to noise
ratios for operation on round bars.
It might be argued that, after using a transducer with a large focal length to optimize the
response from a target such as a flat bottom hole at a given depth in a flat block, the
water path length could be changed to produce optimum focus at the same depth in
round material to be tested. There are two problems with this approach in addition to
those mentioned in the introduction. First, the extra long water path required during
actual testing would increase the opportunity for spurious signals from small bubbles
and other possible contaminants in the couplant. Second, and probably more important,
at long water path lengths the stability of the transducer position and angulation with
respect to the material surface becomes more difficult to maintain. This could lead to
degradation of the test reliability.

Conclusions
The practice of allowing the use of flat blocks for material over 100 mm (4 in.) should
not be permitted in general specifications or recommended practices except as a
special case sanctioned by written permission from the procuring agency. Such
permission might best be obtained in cases where years of empirical evidence has
shown satisfactory results based on a specifically defined system standardization
technique with flat blocks for cylindrical material testing. Even so, this represents a
compromise with good practice that should be used only in special circumstances.

References
Beck, K.H., "Ultrasonic Transducer Focusing for Inspection of Cylindrical Material,"
Materials Evaluation, Vol. 49, 1991, pp. 875-882.
Beck, K.H., "Limitations to the Use of Reference Blocks for Periodic and Preinspection
Calibration of Ultrasonic Inspection Instruments and Systems," Materials Evaluation,
Vol. 57, 1999, pp. 323-326.
Beck, K.H., "Ultrasonic Transducer Selection for Fixed Water Path Scanning of a Range
of Bar Sizes," Materials Evaluation, Vol. 61, 2003, pp. 517-522.

Krautkramer, J., and H. Krautkramer, Ultrasonic Testing of Materials, fourth edition, New
York, Springer-Verlag, 1990.
Singh, G.P. and J.L. Rose, "A Simple Model for Computing Ultrasonic Beam Behavior of
Broad Band Transducers," Materials Evaluation, Vol. 40, 1982, pp. 880-885.
US Department of Defense, Military Standard, MIL-STD-2154, Inspection, Ultrasonic,
Wrought Metals, Process For, 1982.

Vous aimerez peut-être aussi