Académique Documents
Professionnel Documents
Culture Documents
,& -J? f
~s <-.~li
'.
* ROTATINGANODE STARTER
FILAMENT
HIGH TENSIN POWER SUPPLY
AUTOMATIC EXPOSURE METER
W' - %t^ * - - %***" m
ST^4
:DA TA
s*^t m
MI
:;:;;;
4 FILTRO MAMM-" :^tMf/~~*/i
.;
t. Calibration safety
g&; ff Generalinformaton
'---: - ^^,
m*m-- ";
" " " ' ' ' - "'"- " ' - > ' : - ' - - :" """^ V"'V.C " ", ' . - " " - - " - . ' : = ' v : --- ^ ..'--' '"
jjfi:-4'CaHbraion
2. Control LED
:'
W^Kf;K^'
i
W^fiction configuration
3.1 Enablng brakng
. "
'......
..
-.'
--"'..^T::-
/---
-. -" "
- '-'-.-.-
.;9?-
",--".
. -.mf
,-- :
':.,
FLATIIITECHNICAL MANUAL
In this mode, no X-rays come from the unit and measurement time is sufficiently tong.
If necessary, the anode can be started even by short-circuiting the two collector and sender pins of the iC
photo-combiner. in this case use an insulated tool
The two safety circuits of the rotating anode starter must be calibrated so that an anomaly in at least one of tN
two circuits that causes the current to drop below the reference valu does not atlow or tnerrupt the exposure
normal
ALARM
ALARM
During the calibration phase, undesired safety circuit trpping could be an obstacle. If this happens, it is advisa
to exclude it by reducing, even to zero, the tensin on TP3 and TP4 through RV1 and RV2.
In special cases, the safety circuit can be completely exduded by short-circuiting JP4.
During calibration make sure that the anode is rotating.
1.2 Calibration
During a sufficiently long exposure, measure the tensin on TP1, then adjust RV1 until the TP3 tensin is 65%
that measured on TP1.
V(TP3)*0.65x(TP1)
V.
FLATIIITECHNICAL MANUAL
The tensin vales of the above mentioned Test Points are shown on the calibration sheet in the documentation
accompanying the unit,
VPT1
VPT2
COMET
0.9 V
2.2 V
!AE
1.2 V
2.5 V
2-CONTROL LED
LED1
LED2
LED3
LED5
LED6
3 - FUNCTION CONFIGURATION
--
PCB 92-074-3
JP1=open
JP1=ciosed
JP3=closed
JP3=open
FLATIIITECHNICAL MANUAL
METALTRONICA
92-074-X
000
ob
UOO SALDATORE
230V
240V
oto
oo*
^B
22QV
WELDED SIDE
4 - TROUBLE SHOOTING
Sf the dlsp!ay wih echnica! Information conainng ibis rnessage appears:
ROTOR SUPPLY
FAIL
Check ha the fat cable has been correcty inserted or 92-074 ana 01-170 boards.
Check tuses F1 anu F3 on 92-074 board.
"
ssszss msv
f>3
an
rf/P.
Of
:
-'
"
S?
"
^V
V ' - - / .
' - - . ':'
s-= ^
FLATIIITECHNICAL MANUAL
TUBE
On power supply board an 1/f converter is connected to th CPU 01-170 for a very ac(
mAs integration.
fer
nt~msi insfrum^nt
Bstore ntern any casbraton moiin, 'K ?!ee?S3y lo setecf tfi0 manya' mods and ~ yafus
of mAs <?!ua to 40 for fhe O.,3mm focos and 13 fc-r ?h O.lftif?? tecifS,
are active:
(Mi
4,
The anode current related to each valu kV for each focus depends on the characteriscs of
tha tf iH<s ncfoIioH ""H i* H->r* * K?^ <<< **>*> ! *.lu^*,*;-,^ , ? ? * **'? -irs.' = ?**?*. *'* -
>- L ?
<
FLATIIITECHNICAL MANUAL
To access mAL current calibration, it's necessary to act on DIP2 SW1 on board 01-170.
mAL
4
|3
ON
mAH
90
if 4,710
la 90,1
A
mA
mAs
40
UW l%
. T
SJ
c>\ ssntji(ifr
U!
K?
-.-:=
<--.*>..,:-<-,^'.--..-J;-.-3
-\ij
^'/^V^?
| ? S ?-; i. ?'-:"
^"'ff
Uf
ir;
UtiLUnaJp LjOOS U|
nq gsuu AKJ-X uuiijsn jaye -yo IIOIIMS t?O3T PUB j.a3T oeqn
u| uiu f. -o joj //vy jo SHQO
tj 3DBIJOA 8SB8JOUI '
JO B ijOOS U|
?'vC>>
J : : "
..
"'
r-.^ s^
U U
ijdsip uo
FiJOCQOBANOE
FUOCOPBCCH.O
SMALtFOCUS
sss.
Catibration of over-current safety circuit is executed at the end of anode currante calibratlc
The mximum filament current shall be drawn from he calibration report, for each focal m
Jo carry out cabraiors:
a) conroct a voltmeter 10 Vdc f.s. on TP6GND) and TP5,
b) setect LARGR FOCAt SPOT
c) selec he kV valu mtevant to he hghest filament current If
t) measurs volage on TP5/TP6 dunng preparaicn
el seiect SMAtl PlC&I SPOT
HHH IH K V
^ i>*i
FLATIIITECHNICAL MANUAL
The passage of anode current s indicated by LED5, whch during exposure lights up
OttOOiriv to inri annrie current:
__,,
Wfjtfrriotor
i H I Sil 1J-"
'?.<? i -..
': >'>*;*
i i
ti
i A V^ ^ -.
t~
* *-** r% y ~ V * A
^Y"-^'-'-
iS ifiSSilSu lu 1S /i"
S
i ,..,,., H
..,..!.. ..,,,,,,..,
7 - TROUBLZ SHOOTING
7.1 fttaimnt powersupply
The appearanee of the teehnicat screen wfth message:
"
-..:
;- - -..:,
1S rtia h sottWBrf* fnr fahrafjnn nararnAter? nnntrnt nff>nt?d ?>,n ffror n thf?
t ws; n-(
si"? n-15-K/ "-^V :")i'M*i"-*i"4';'*- HV VM*-^ *K" t ?V *-*"5*"?i ?'w M \MM JMI TH"w"i <"V* 'i?"** iii.U'f's H'IW^.
"
1* Descripcin
frpybleshppting
1 ~ DESCRIPTION
The X-ray tube used n this unit is the grounded cathode type.
The high tensin generator has positive polarity with rsped, to the yround.
>*
wp
M
""' T
The valu of high tensin required is set by the CPU thiough an S^bit opto-insulated bus
IC6/IC7/IC8.
It is important to point out that the ground return of the opto-insulated commands is
physically nterlocked to the ray button. This means that the commands can be transferred
from the CPU to the high voltage feeder orily when this is activated.
The PWM controller reglales ttie tensin applied to the bridge that is necessary to obtain
the kV valu set by the CPU.
A loop band width of about 3 k.Hz and the feedforward of the regulator ensure perfect
stabilization of the high voltage valu and lolally cancel tho doubie frequency mains ripple
from the output wave form.
The characteristics described above allow for a perfect wave form without overshoot and with
a kV valu that is totally independent o the anodic curren! valu supplied to the tube.
CAP 1 PRS>P 18 of 88
NO or SI OW IPIGPfP
The oscillograrn shows the high tensin form or 20 kV and 35 kV measured on TP1.
The kV wave form does not present overshoot in any operating stale and the kV valu is
iridependent of the current fed to the tube.
The current lirnitation of the chopper ensures perfec protection from current overloads at
output and a constant conductance of the kV rise time typical of 1.5 ms.
TEKTRONIX
2236
AUiTHTai
1U
IELE._L_ 2j]?.5 [
FLATIIITEGHNICAL MANUAL
2 - CHECKINGINTERNAL Kv (TP1/TP2)
The check can be made with a normal digital tester with a range of 1000 Vdc full scale arj
lOMOhm input impedance without the assistance of an High Voltage, divider inserted in tff
anodic circuit.
To make the check, connect the voltmeter or memory oscilloscope on TP1/TP2 (GND) of thl
94-132board.
Exclude ray nterruption with the [RESET alarm]+[kV+] button, valid only for one exposure
the DIP4 SW1 PCB 01-170 button, permanently, select the manual technique and a mf
valu that s high enough to consent a stable reading.
KV
20 kV
21 kV
22 kV
23 kV
24 kV
25 kV
26 kV
27 kV
28 kV
29 kV
30 kV
31 kV
32 kV
33 kV
34 kV
35 kV
VTP1/2(1%)
279 Vdc
293 Vdc
307 Vdc
320 Vdc
334 Vdc
348 Vdc
362 Vdc
376 Vdc
390 Vdc
404 Vdc
41 8 Vdc
432 Vdc
446 Vdc
460 Vdc
474 Vdc
488 Vdc
The table shows the kV vales and the corresponding vales measured on TP1/TP2.
NOTE
1) The:values in the table are valid f the reference voltage of the kV DAC measured o
TP8is5.00Vdc (standard calibration).
2) In exceptional instances kV vales can be calibrated with the RV5 trimmer (PCB 94
132). In this ristance the reference vales on TP1/TP2 are no longer valid. RegulatngJ
the kV made with RV5 is the only one for the whole range. It is advisable to correct the
28kV reference valu.
3) The kV vales are not affected by the mains voltage within the specified limits
4) The kV vales are not affected by the anodic load within the specified limits
5) DIP4 SW1 on board 01-170 must be in the ON position to exclude ray nterruption
measurement. The same effect can be obtained by pressing the [RESET alarm]+[kV-j
button, a valid function only for the next exposure
6) Measurement of the kV valu MUST be done exclusively in MANUAL mode since
Automatic mode exposure s always preceded by the pre-exposure impulse.
7) The kV vales are not affected by the replacement of one of blocks composing the HV
feeder and no recalibration is necessary after repairs have been made, if the kV valuej
has not been corrected with RV5
!
METALTRONICA
3 - CHECKING EXTERN AL kV
1)
METALTRONICA
CAP3Page21of88
PCB94-132
TP1/TP2
TP3
TP4
TP5
TP6
TP7
TP8
TP9
notused
TP10
internal input voltage +15Vdc
TP11
output current limitation reference
TP12
output current 1 Vdc=1 OA
TP13
reference grounding
PCBOO-16&
IS CHOPPER and BRIDGE output control
T.A.
INVERTER output current control 1 Vac=1 OA
5 -EXPOSURE TIME CONTROL
The inverter has an output current limit through the RV2 trimmer. The voltage reference on;
TP11 for the various configurations is:
1.5 Vdc standard
5kW 2.0Vdc option
1.7 Vdc bimetallic tube option
If the filament current can allow the anodic current to exceed the set valu limit, the output
kV valu is reduced in order not to exceed the mximum allowed power.
7- CONTROL LED
The green LED1 ndicates the soundness of the F2 fuse and the functioning of the internal||
input at+15Vdc.
The yellow LED1 turns on during exposure and can be used to verify continuous or pulsated||
X-ray emission.
FLATIIITECHNICAL MANUAL
8 - TROUBLE SHOOTING
There are no on-site reparable parts inside the inverter and the HV transformer.
Since dmage to the inverter could have been caused by damage to the HV transformer, it s
advisble to replace both of them to avoid the risk of damaging the new Inverter.
On-site checking, to see if the HV transformer functions, is not possible with the normal
instruments provided with the unit.
The Mosfet inverter and the HV transformer can both be repaired at the factory.
If the following message appears:
... INVERTER
POWER = OK SERVICE = NOK
Check fuse F2 on board 94-132
If the following message appears:
...
. INVERTER
POWER = NOK SERVICE = OK
Check the input power of the inverter and associated fuses.
Check the tensions and the control circuit of the power input.
The trip threshold of the control circuit s regulated by RV1 on the 92-082 motherboard and
must be 1.05dc on TP1 (TP2 gnd)
92-082
R30
If&f
R29
Hfr
CN4
fcd
Q>
H.
J
94-132
The control circui can be excluded during testing by short-circuiting the JP1 bridge on the
CPU 92-082 motherboard.
If the following message appears:
CAP 3 Page 23 of 88
CAP3Page24of8fc
S i I g e i > ^ f i s 5 f t ^ w H H
w^<f.o'-.-^^^*Jl^
4gli|tSi^^^
^.^:^;^&i;:?&^-&j-$3^^
,-.:.:;s3:;-s;!feKK;ix*
|fi||||||^
l^iSIfilllillllli
1!^^^^^^^^^^fj^^[fB'y^^^^^&^^^^i^^^^r
ftiSSMts;^.* invmsffsiii,iijiii,i'ii^i;KffM-if!imt^.i,i^aigfVfiii/iiit!tsff!i'fl'mxf!
'~*35S,:^iK^(!>s:S":ii.yiv.ifrS^:*^^
t;!tyrf^S^^S''''y^l^p^:WKKf'Sy&Vi'SMWt
'!"<f^Sfs^si^'^'-&':!:"^:^i&^:^^^iM
I^P^^W^'^S^j^^^^M^^IW^WwSS^IhW^w^^i^^m^ r/^.^'--'^'-^^'^.;*:^^^
:v.;.' ':-.: S^M^S-iW^fs^ v-ir-^!->*H^'^s/|^Ar.--vy?5^':?r;
^S^&'^W'^VnflnW/^/T^nlrl/^rt
KiRl^^i^tfeSfelSsft^S^taiS^^S
^^:;^:'^i>-.fc/.J;|vfc*|iNC /UCSI^VlCTIljII-lrvI.^iyf.f /Crlrf'f lt/f^^cS^"^<^^:f^^'^y', ;^:Cv:'vV::.-,C-4;-:::
KSSt'tK''i4;S!*"fVft?:,5aty.;:fi;1jv:'iMavi^--vm
*."J'!",;>sJJ
::
\iPTAT TRONICA
CAP 3 Pape 25 of 88
:
FLAT IIITECHNICAL MANUAL
1- INTRODUCTION
The unit has an Automatic Exposure Control (AEC) that is among the most advanced
available on thejnarket today.
Operation of the" AEC is based on the actual density of the organ to be examined, measured
by a very brief X-ray impulse called "pre-exposure".
This is the only way of knowing the density of the organ that, as such, cannot be validly
shown by its thickness.
The radiation [r] read by the detector during pre-exposure shows the density of the organ and
after being converted to a number that is inversely proportional to it.it will be used on the
process of exposure control that will continu in a completely numeric form, except for the
Dose integration.
29 -
somple
<10ms
LARGE = low
exposure
The above figure shows a time schematization of the exposure concept in ZERO POINT
technique, with completely automatic kV and mAs.
The parameter [p] derived directly from the [r], the density of the breast being examined, s
used as a poiner on two look-up tables containing the kV and mAs necessary for obtaining
an image with the right blackening and contrast to fully satisfy any Quality Protocol in the
Mammography field today.
A high valu [p] corresponds to a small density organ and low vales of kV and mAs are
taken from the tables.
A low valu [p] corresponds to a large density organ and high vales of kV and mAs are
taken from the tables
With the kV and mAs vales taken from the tables, the exposure that was interrupted after
the pre-exposure impulse is resumed.
The kV vales in the table cannot be modified, but four different tables are made available to
the operator, with different kV profiles suited to different requirements that go beyond the
possibilities offered by the well-known High Contrast (HC) and Low Dose (LD) techniques.
From the control panel, the operator can not only select the technique that he feels most
suitable at the time, he can also make fine-tuning in steps of 0.5kV on the kV range selected.
iFTAl TRONIPA
CAP3Page26of88
FLTIIITECHNIGAL MANUAL
The aboye figure is a diagram of the operating possibilites of the DEATA PLUS software in
its maxmurn confguration with Bimeallic Tube and automatic Mo/Rh Filter.
MAMMOGRAPHY,UNIT
IVFTALTRONTrA
CAP3Page27of88
To obtain a good X-ray mage, each component of the SYSTEM plays a determining role anc
must be appropriately managed and biended with the others.
No system s able to provide results suited to the requirements of present Quality Contro
protocols without an appropriate calibration that takes, into account all the specif<
characteristics of the single components.
The scope of this procedure is to crate the right blend amongst the system components fo
obtaining the required level of blackening and ts constancy within the variability limits o
breast density with the mnimum dose possible.
Some key information will be provided below on how to obtain a good image with minimun
dose.
1.2 - Selecting film/screen combinaion
For mximum image quality, it is advisable to use only film/screen combination
recommended by he manufacturer.
The United States regulations have made this recommendation a requirement.
The X-ray cassettes for mammography use have only one intensifying screen that convertirle X-rays to light that mpresses the image on film.
Every screen has ts own ray/light conversin factor that must be suited to the film to bi
used.
Films for mammography use are single emulsin and have various sensitivities an<
contrasts.
An appropriate choice of these producs is reflected in image quality. We leave thi.'
responsibility to the user, dutifully pointing out the difference in dosage that this can entail.
MINRH + MINR
Cm
KV
mAs
kV
mAs
5.5
28.0
480
23
28.0
190
24
The tables shows the mAs necessary for obtaining the same blackening on 5.5 cm o
Plexigls, using two different films/screens.
The difference is considerable: 480/190=2.5 times!
1.3 Developer
Film for mammography use has treatment characteristics that are very specific and difieren
from those used for other film.
A dedicated developer is recommended for treating mammography film, which guarantee;
temperature and development time suited to the specific film.
Development temperature is generally between 34 and 35 C and has a marked influence o
the radiological dose necessary to obtain the required blackening.
METAL TRONICA
CAP 3 Page 28 of 8:
2- CONTROL TESTS
In this configuration, calibrating the detector sensitivity is of mximum mportance in order to
guarantee the correct correspondence of the exposure parameters taken from the kV/mAs
tables, with the organ density.
To faciltate control and calibration of parameter [r] each unit has a plain block of Plexigls
(dummy) for a rapid control and a simple re-calibration of the time. The dummy s a relative
reference that is characterized in the factory after system calibration with an absolute
reference.
The dummy is an integrating part of the unit, can be used also by the operator and must
always be available to the operator throughout time.
During the test the system must be configured as shown n the following figure.
MFTA! TRONICA
CAP 3 Page 29 of 88
After exposure, check the [r] valu and if it difers more than +/-10 from the valu shown on
the dummy, recalibrate the [r] parameer.
2.1 - OFFSET [o] CALIBRATION
The valu of the detector offset appears in the lower left of the display when one of the two
automatic techniques is selected. The offset valu depends on the temperature.
The valu must be checked after at least 5 minutes from turning the unit on and t should be
105 with PHTM 9000. The system loses its linearity if the valu reaches zero or exceeds
20. Great variations of the [r] GAIN regulation influence the offset valu and require
correction.
Check and correct if necessary the OFFSET valu for all three available fields.
2.2 - DETECTOR [r] GAIN CALIBRATION
At the basis of good system operation the valu of [r] s correct.
Valu [r] depends on:
a) Precisin of the kV valu
b) Precisin of the mA anodic current
c) Detector gain calibration.
Before proceeding with gain calibration, make sure that the vales referred to n points a)
and b) are correct.
The valu of [r] appears in the lower left of the display when one of the two automatic
techniques are selected.
The valu must be checked after at least 5 minutes from turning the unit on and t should be
the same as the valu shown on he test dummy +/-10.
offset CW = increases
gain CW = decreases
o
o
o
O3
02
01
o 9 o
G1
METALTRONICA
G2
CAP3Page30of88
3 - PROCEDURE PREPARATION
3.1 - Developer
The developer must be n perfect conditions for use with developing and fixing replaced a
day beforehand.
3.2 - X-ry cassette for tests
Select and mark a cassette that will be the only one used or all the tests.
If possible, select the cassette with optical density as cise as possible to the average
measured on all those available.
3.3 - Film pack for tests
Make sure that there is enough film n the pack being used to complete the procedure.
Otherwise open a new pack.
CAP3Page31 of88
CALIBRATION FILE
Start the program with option 1 in order to continu working on an existing file or to crate a
new one
FILM/SCREEN COMBINATION
Select the Film/Screen number to calbrate in the 1 to 16 range..
=< FILH/SCREEM selecton >seect FILM SCREEN identity code in tlte range 1 to 16
FILM/SCREEM ID number =
ANODE/FILTER COMBINATION
Select the Anode/Filter combination to calbrate.
-< ftNODE/FILTER selection >=
each FILM SCREEN conbination has four independent c a l i b r a t i o n
files, one for each one of the four NODE/FILTER conbination
1 = anode Ho * f i l t e r fo
Z = anode Ha + filter Rh
3 = anode U + f i l t e r Bh
TftRGET/FILTER SELECT10M =
CAP3Page32of88
W!th DEATA PLUS all Anode/Filtercombinations can be used indefinitely, in both ZERO
POINT mode and ONE POINT mode.
MFTAT,TRONIC A
CAP 3 Page 33 of 88
NON-EXISTENT FILE
If the number entered does not correspond to any file, the following message will appear:
To crate a new file use N (new), and enter again the number of the file to crate.
It is advisable to use the same serial number as the unit so it will be easy to find in the future.
EXISTING FILE
If the file selected exists, the file heading information will appear.
(figure)
FOISNOOI.HH
Paee34
FLATIIITECHNICAL MANUAL
MAIN MEN
AII cabration operations can be performed from the main MEN.
1)
2)
3)
4)
5)
6)
7)
8)
9)
A)
B)
O
D)
E)
Q)
operatipg technique
uorking file
<
ENVIRONMENTALCONDITION
Gathering general information regarding the system and all ts components is very important
and useful for fture reference and in order to dentify file conten.
-'
Grid
Grid
Grid
Grid
Grid
Grid
Grid
Grid
STD
STD
EXT
EXT
HC
HC
LD
LD
MFTAI TRONICA
CAP3Page35of88
SENSITQMETRIC CURVES
For each operating technique regarding the film/screen pair, the sensitometric curves
devloped during calibration can be saved separately.
Since the calibration process can probably be completed n different times, this could be very
useful for comparing sensitometry throughout time, and a special SHOW ALL function is
provided to display the 16 curves at the same time in different colours.
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
step
IHE 22:59:52 CUTE 01-19-19%
TEF NUHBEB 10 REDEF. O.B. / C=COHFftBE / Q=QUII
18
11
12
13
11
15
16
17
18
19
28
Zl
O.D.
0.0."
O.D.=
O.D.O.D.=
O.D.=
O.D.=
O.D.=
O.D.=
O.D.
O.B.
O.D.=
O.D.D.B.O.D.*
O.D.*
D.D.=
O.D.=
O.D.*
O.B.O.D.=
O.B.=
Base Fog
Speed Index
Contr Index
8.18
0.1B
8.13
6.19
8.19
8.21
9.22
8.26
6.31
e.51
8.89
1.11
2.00
2.55
3.61
3.38
3.51
3.63
3.71
3.7S
3.78
3.
* 0.18
= 11
= 3
To enter the vales, first enter the step number, then ihe O.D. n the following form: X.XX.
Detailed Information about the sensitometry can be obtained from specific literature.
It is important to clarify that all successive points depend on the sensitometric curves and
only their stability in short and long term can give optimum O.D. results.
The C COMPARE function s a special one for comparing the current curve with one
contained n another file.
After pressing key "C", enter the ame of the file (including the extensin) and the two curves
will be displayed on the screen.
stC[)
step
step
step
step
step
step
step
step
step
step 18
step 11
step 12
step 13
step 11
step 15
step 16
step 17
step IB
step 19
step 26
step 21
lME 22:59:52 DOTE 01-19-1996
TEF MU1BEH TO BEBEF. O.D. / C^COtlPdRE / Q=QUII
MFTAT
Base * Fog
Speed Index
Contr Index
CAP3Page36of88
The figure shows two curves with FUJI UM-HA-HC. The highest is obtained with KODAK
developer with 180 seconds of development time, while the lowest with FUJI with 90 seconds
of development time.
The figure shows the curves obtained with the same configurations of the preceding figure,
with the highest curve obtained under normal conditions, while the lowest one under
unacceptable developer conditions for normal use.
The pre-existing curves can be used as reference for calibrating other developers.
OPERA TING TECH IQUES
Every film/screen combination has 16 different operating techniques, with parameters that
can be totally and independently defined.
01 = 1 rge focus uithout Bucky Mo/Ho
oz
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Bucky
Ducky
STflMDflRD
STANDARD
STANDARD
STANDARD
EXTENDED
EXTENDED
EXTENDED
EXTENDED
H16H CONTRAS!
tto/Ho HIGH CONTRAST
ho/Ho HIGH CONTRAST
No/No HIGH CONTRAST
Moxlo LOU DOSE
tto/*MoLOU DOSE
Mo/Mo LOU DOSE
H/Mo LOU DOSE
flo/Ho
ho/tto
Mo/Mo
Mo/Mo
No/fio
Moxlo
Ho/Mo
Ho/Ho
=?
The four standard operating techniques are repeated four times n relationship to the four
different kV ranges available to the operator at the control panel during normal operations in
ZERO POINT mode with kV and mAs completely set on automatic.
The four ranges are dentified as:
STD Standard
EXT Extended
HC High Contrast
LD Low Dose
The current work technique is highlighted on all screens and stored in the configuraron file,
to be automatically reloaded each time the file s opened again.
MF.TAT,TRONICA
CAP3Page37of88
Use the 0.5 thickness of Plexigls or else remove the compressor to make a distance from
the kV margin.
Repeat the test for all thicknesses that are to be used.
5.2 - Test X-rays and calibration data collecton
The following figure shows an example with three thicknesses of Plexigls: 2.5 cm, 4.5 cm,
mAs, p, i
are sent from the mammography unit to the computer and can be received and displayed
using function 4 "RS232 COMMUNICATION SOFTWARE" from the main men.
MF.T.A!.TRONICA
CAP 3 Pae 38 bf 88
LJ .
Oin
1 .2
;. :
,
.
pAH14.ri
. O D H . = 1.47
|mAsH=62.6
O.D.H.=1,54
= 286
^7S
mASH=191.8
.O.D.H.-1.49
=407
p=42
+01
raorioTn
Jrirtl
OQ7
mAsL=45.4
--'-=1-9
=ZL.
msL=52.6 1
O.D.L.-1.17
p=41
..-J
n 1
1
0.
o
2.5
4.5
6.5
cm PLEXI
Measure the Optcal Density of the film with the densitometer, check that this is higher ODH
or lower ODL than what s required and record the mAs, p, i parameters in the experimental
datatable.
Experimental data tabie
cm
2.5
4.5
6.5
CAP3Page39of88
nAs
=
O.D.
pulseC i ) =
nfts
O.D .
roAs
=
pulseCi) =
highest nAs
<ESC to escpe>
ualue =
each point is def ined uith tuo shots , one uith O.D.> 1.2 one uith O.D.< 1.2
The nterpolator is very easy to use since the software asks n sequence for data to be
entered, as specified in the table.
The procedure s to be repeated for each thickness in the table.
At the end of interpolating the data of each thickness, the interpolator gives the vales
mAs, I (pulse)
to be nserted in the table of interpolated vales.
Table of interpolated data
cm
2.5
4.5
6.5
mAs
11.9
49.1
155.9
pulse(i)
140
224
331
P'og(p)
O.D.
120
78
42
1.2
1.2
1.2
The valu of p remains the same as the preceding table since it is the density of the
Plexigls dummy for that specific thickness.
5.4 - Buldng the ZERO POINT calibration curve
Select option 5 from the main men to define the calibration curve, using the parameters
taken from the TABLE OF INTERPOLATED POINTS.
CAP3Paee40of88
TRIP
TKIP
IRIP
O.D.
i
i,
i '. ,'.'.;
:
',.
;
SID-
PULSE=Z160
PULSE=1038
PULSE= 681
PULSE= 534
PULSE= 415
PULSE= 3Z7
PULSE= 267
PULSE= 240
FOIMT tto/Rh
POIMI U/Rh
POIMI Lou p
correctlon
= 55
=41
= 10
= 120 x
0 8 1 2 3 4 4 5 6 7 8 8 9 1 1 1 1 1 1 1 1
6420864208601223456
FILrVSCREEM Large Grld ST 4 Z O 8 6 4 2 O
;-:'.';
PLOG
(Q) ?
From the screen display shown above t is possible to define the calibration curve; the
vertical dimensin (PULSE(i)) is automatically adjusted according to the mximum valu
inserted.
The number of PLOG/PULSE data pairs can be decreased (remove) or increased (insert).
Inser the p and i vales contained in the Table of Interpolated Data in ncreasing PLOG
order.
Aftr nsertig the p and i pairs of vales obtained from the preceding tests, it s necessary to
manallv adiust the beginninq (p=0) and the end (p=160) of the curve.
The number ofT impulses is selected in order to continu the natural progress of the curve
without formina dscontinuitv and points of inflexin.
If the curv obtained with only three thicknesses is not considered satisfactory, other
thicknesses can be defined to complete the Job.
5.5 - Storing data in the unit
After the building of the Zero Point curve s completed, the Calibration File must be updated
to avoid data loss and proceed to transferring these data to the Mammography unit through
the RS232 serial connection.
During data transfer the parameters are displayed on the screen and all other functions are
disabled.
Two different messages are displayed n the case of correct or incorrect transfer.
In the case of error, the transfer must be repeated and the connection checked.
da-t.4
upload
From
PC *o
MftMMQUMIT
i s:
inAc?
ndcsin-t
of
oinlyt>yPCsoF-twir'e
yo
ore:
uploodig
ress
anij
kctiji
-to
-to
FILH^-SCRKEN
-t=cop
oont; inue
:The:above message is shown before transfer, with the main data; transfer takes place only
iafter confirmation.
MF.T A T TRONICA
CAP3Page41of88
Cm
2.5
4.5
6.5
MAs
12.2
kV
plog(p)
pulse(i)
D.O.
25.5
119
142
50.9
26.5
78
224
152.6
27.5
38
312
1.25
1.23
1.12
CAP3Pape42of88
<I7TI
\r the definition of the % correction vales relative to each step, it is advisable to use: :
p|
Technique of the two Films with Density higher and lower than the desired
density.
Intrpolator
The % valu is easily obtained by a calculation based on the mAs vales obtained.
METALTRONICA
CAP3Page43of88
FLATIIITECHNICAL MANUAL
'.
100X
100X
lOOx
lOOx
lOOx
lOOx
lOOx
lOOx
100X
lOOx
150 x
140 X
130 X
120 X
110 X
100 X
90 X
80 X
70 X
60 X
50 X
e 6oa
A ^A VA. A
^ ii vAv Av v
A v'
V
v %f v
0 0 0 0 0 0 0 0 0 1 0 0 0 0 0 0 0 0 0.1
1 2 3 4 5 6 7 8 9 0 123 45 6 7 8 9 0
CAP3Page44of88
20mm
30mm
40mm
50mm
D(kV)
O.D.
OPTION2
To calbrate the reply at 30kV fixed and variable thicknesses.
kMlP >= 30kU > kUOP
11) Wlf = kUOP +5.5
12) kUlP = kUOP +6.O
13) kUlP = kUOP +6.5
14) JUlP = kUOP +7.0
15) kVlP = kUOP +7.5
16) kUlP = kUOP +8.0
17) kUlP = kUOP +8.5
18) kUlP = kUOP +9.0
19) kUlP = kUOP +9.5
20) kUlP = kUOP +10
100X
100X
100X
100X
100X
100X
100X
100X
100X
100X
150 X
140 X
130 X
IZO X
110 X
100
90
80
r0
60
50
X
X
X
X
X
X
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
00000000010000000001
1 2 3 4 5 6 7 8 9 0 1 2 34 5 67 89 O
C 30kU
kUlp > 30kU
50mm 60mm
Mas
plogCp)
Pulse(i)
D(kV)
O.D.
MF,TAT,TRONTCA
CAP3Page45of88
:f
OPTION 3
To calbrate kV1 P<kVOP
190 'A
180 X
;::i]::i:j::fj:::p9:i:::
: : : : : : : i/:
150 X
10)
11)
1Z)
13)
100 X
kUlP
kUlP
kUlP
kUlP
=
=
=
=
kUOP
kUOP
kUOP
kUOP
-5.0
-5.5
-6.0
-6.5
180X
190X
Z00X
ZlOx
130 X
: : : ;
^.;...i..|.ij^4.|.|..L,|..|,i
;..|.i.^.L|..L|...;...L|..;
i ; : ; ; : ; ; ! ; ;
iI^f..L.i.i.i.i..L..L..i...L..L.;
30000000011111
1 2 3 4 S 6 7 B 9 0 1 Z 3 4
ualue = |
The following table s fi'lled in, as an exarnple, with vales from a calibration:
45mm phantom
Mas
plog(p)
pulse(i)
D(kV)
O.D.
MPT A T TRONICA
26kV
120
28kV
79.8
77
466
-1
1.23
78
45.9
+3
1.28
30kV
54.8
78
459
+7
1.37
CAP3Page46of88
7 - CALIBRARON FORMS
parameters for O.D.>
mAsH O.D.H Pulse(i) plog(p)
cm
'
EXPERIMENTAL DATA
Cm
mAs
pulse(i) Plog(p)
O.D.
INTERPOLATED DATA
RLEpAf
FILM/SCREEN
HOSPITAL
OPERATOR
MFTAT.TRONICA
CAP 3 Page 47 of 88
Cm
MAs
KV
O.D.
Plog(p)
pulse(i)
./vi- -,
26 kV
30 kV
28 kV
Mas
plog(p)
pulse(i)
d(kV)
O.D.
20 mm
30 mm
40 mm
50 mm
P'og(p)
pulse(i)
D(kV)
O.D.
ONE POINT FINAL DATA (28 kV fixed and variable thicknesses)
30Kv
mAs
Plog(p)
pulse(i)
d(kV)
O.D.
50mm
60mm
CAP3Page48of88
FLATIIITECHNICAL MANUAL
3 Film/Screens instead of 1 6
Only kV Standard range
Only Mo/Mo Anode Filter combination
The reference Optical Density valu can be corrected with the coefficient of % calibration.
Access to the percentage correction coefficient is obtained with DIP6 SW1 on board 01-170.
The variation of percentage correction valu is done with keys mAs+ and mAs- and the
relative valu appears on the display in the place of the Manual Density valu.
The variation of percentage correction valu is done with keys mAs+ and mAs- and the
relative valu appears on the display n the place of the Manual Density valu.
mAs(
Regulating the Optical Density must be done on a Mdium (4cm) dummy thickness,
repeating the sequence correction / exposure / Density check until the desired valu s
obtained.
In this configuration the unit has 3 preset curves in order to have different slants that allow a
limited and emprica! linearization of the Optical Density.
Density Linearization
After adjusting the Density to 4 cm, make a Film with 2 cm and one with 6 cm of Plexigls.
Compare the three Films and if the blackening is insufficiently constant, repeat the test with
another Film/Screen combination until the best compromise is found.
MFTAT TRONIPA
CAP "?
40 of RR
*'
t-
FLATIIITECHNICAL MANUAL
>5 27.5RV 151.1mAs .30 Mo/Mos G OP D+1 F01\6 r0041 p041 0246 d O c100 %115 o00
The memory capacity s suficient to contain the information regarding the last 1300
exposures.
For data collection use option 5 rom the start men of the DEATA PLUS calibration
software, after having connected the RS232 cable to the COM1 port.
The parameters in the memory will be stored in two files with a TEC and .MED extensin.
'The first one contains technical parameters to be used by the Service. The second contains
radiological parameters exclusively, for use by the user.
Every file starts with a comment that can be entered by the operator, date and hour of
identification are added automatically.
At the end of data collection a third file with an .STT extensin is generated automatically,
containing a statistical analysis of the data collected (see dedicated paragraph).
The .TEC, .MED and .STT files are simple ASCII files, compatible with all existing text
editors, from DOS Edit.com to the latest versin of Word. They are created in APPEND,
which means that, using a ame of a file already n use, new nformation s added at the end
of the existing file.
WARNING: perform this operation only from the hard disk.
After starting the program and defining the file ame and comment, actvate the transfer from
'the control panel of the mammography unit.
|
Data download
press [BELL]+[man DENSITY+]
A beep will signal the download of each string, then 3 beeps signal that the last piece of
information has passed to the PC.
Data download nterrupton
press [BELLHman DENSITY-] and wait for 3 beeps.
Olean memory
press [BLL]+[man DENSITY-]+ [man DENSITY+] at the same time.
A 4-second beep shows that the memory has been cleaned.
It s advisable to clean the memory after each valid data coliection.
MFTALTRONICA
CAP 3 Page 50 of 88
FLATIIITECHNICAL MANUAL
n
"p
fe
not important
27.5kV
kV
CAP3Page51 of88
11
i. *-:?-- !&
IP
U
MFTAT TRONICA
CAP3Page52of8'
FLATIIITECHNICAL MANUAL
10 - STATISTICAL DATA
The data of the last 1300 exposures stored in the memory contain all the nformation for
making the tables that follow, automatically, at the end of the data download from the unit's
rnerory.
Ther are many ways that the data collected can be used by Health Physicists and service
technicians. By being able to modify some of the unit configuration parameters, they have a
valid instrument for evaluating the effects of the variations made.
The table that follows contains the simple count of exposures made, divided nto the two
focal spots.
This data s useful for understanding the type of use of the unit. A very low number of
exposures with Small Focus will emphasise that the unit is used mainly for Screening without
successive in-depth specific studies made with the enlargement technique.
= 1308
(table)
The frequency that the three techniques, Manual, Zero Point and One Point, are used tells
the user's habits or operating preferences.
A high number of exposures n manual technique is normally due to performing biopsy
exams and exams on patients with silicone prostheses.
A significant number of exams in One Point technique should be studied with the user to see
f it is justfied by special exam requirements or f it conceals problems due to incorrect
calibration or poor development conditions, and therefore technically solvable.
ll
The data in the next table allows us to evalate the tripping percentage of the Rhodium filter,
that n this case is about 10% of the patients.
A survey made on approximately 300 patients can be considered statistically representative
and therefore valid for comparing the data obtained with official statistics.
1
ife
MFTAT,TRONICA
CAP3Page53of88
A significantly different percentage could be corrected by appropriately moving the trip point
of the Rhodium filter.
The effect of this change could be evaluated and if necessary corrected again after having
collected the data^f the next 1300 exposures.
Knowing the data regarding the Total Dose, Average Dose and Average mAs valu is
mportant for respecting the legislation on Radioprotection that provides for controlling the
dose released in radiographic exams.
The Mammographic exam is defined at Low Dose and the effective estmate of the Dose
necessary for obtaining a good image is fundamental for correct system management.
The Dose and mAs vales are directly correlated to the development and sensitivity
conditions of the Film/Screen combination used.
Comparison of the data collected on the same system in different periods and with other
similar systems allows an in-depth evaluation of performance.
The correct selection and administration of Development, Film and Screen can heavily
condition the Dose, in some cases being able to halve it.
xxxx
xx.x
xx.x
The last table highlights the frequency of use of each kV valu and can be useful for a
further evaluation of the system usage conditions.
KV .
21 .OkV
22 .OkV
23 .OkV
24 .OkV
25 .OkV
26 .OkV
27 .OkV
28 .OkV
29 .OkV
30 .OkV
31 .OkV
32 .OkV
33 .OkV
34 .OkV
35 .OkV
METALTRONICA
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
14
25
26
34
138
169
288
52
8
6
0
0
0
0
0
21 .5kV
22 .5kV
23 .5kV
24 .5kV
25 .5kV
26 .5kV
27 .5kV
28 .5kV
29 .5kV
30 .5kV
31 .5kV
32 .5kV
33 .5kV
34 .5kV
=
=
=
=
=
=
=
=
=
=
=
=
=
=
2
7
6
17
83
273
152
4
4
0
0
0
0
0
CAP3Page54of88
,
FLAT III TECHNICAL MANUAL
sil
l>"^^&'^^
fUeti&M&rmSiii
-.'-;..; >.:;. -
. : . o
- - - ; : ^ - .
fi"
ir
H
1
CAP 3 Paae 55 of 88
FLATIIITECHNICAL MANUAL
1- INTRODUCTION
This is a Software-type Dosemeter. It can measure the input dose at skin level knowing :
1.
2.
3.
4.
kV of exposure
mAs supplied
distance from Focus to skin (resolution 5mm)
measured dose with dosemeter at a distance of 55 cm from Focus for
each kV valu and for each combination of anode and filter.
2 - MANAGEMENT SOFTWARE
Configuraron and calibration of the dosemeter are based on the SKDOSEXX.EXE
application software for PC designed to work in a DOS environment.
Given that the PC COM1 serial port s used for transferring calibration data to the
Mammography unit, it is recommended to sart the DOS mode since the PC is switched on,
Le. before Windows takes the control of the session.
Starting DOS emulation from nside Windows could result n some problms due to the
special PC configuration.
For a correct program operation, it is also necessary that the Default Data file
SKDOOO.GY
appears in the work directory, as well as the .EXE file.
Edit SKDOSE01 to start the program
METALTRONIOA
CAP 3 Page 56 of 88
FLATIIITECHNICAL MANUAL
list
. . . . . . : ,
-'
|" !
''';:-.-.
. ' ; fj'y.
r "v:'.!"-: 1
'
'
- ' ; '":'.
After the loading of an existing file or the creation of a new file, you can move to the central
part of the program from where you can chose among 9 operations.
1)
2)
3)
4)
5)
6)
7)
8)
9)
Q)
Option 1 allows you to chose the table relevant to the Anode / Filter combination to be
calibrated.
Molybdenum Molybdenum s the basic one and it appears on all mammography units.
Molybdenum Rhodum is only present on Mo Rh automatic Filter units.
Tungsten Rhodium is only available on units with double anode material Tube.
TRONICA
CAP 3 Paee 57 of 88
1)
2)
3)
Q)-
Calibration carried out for each Anode / Filter combination s valid for both focuses.
Option 2 allows you to edit a comment, which wil! be recorded n this file together with
calibration data.
Option 3 allows you to chose the type o physical quantity, i.e. mR or mGy, according to
which calibration data, resulting from the reference Dosemeter, will be entered.
If you have a Dosemeter calibrated in mR, the program will do the conversin into mGy.
Option 4 allows you to enter reference doses measured at 550mm of distance from the
Focus for each kV valu and for the valu o 50mAs.
In the able, vales in I GY will be entered for 1mAs after appropriae calculations,
autornatically executed by the Software.
Option 5 allows you o define some configuration parameters such as:
1)
2)
3}
4)
5}
Q)
current settings
METALTRONICA
CAP 3 Page 58 of 88
FLATIIITECHNICAL MANUAL
M P T A T TPONTP A
PAP 1 P.ion SO nf 88
ONICA
CAP3Page60of88
FLATIIITECHNICAL MANUAL
PADDLE OFFSET -
mm
FOCAL SPOT
fe:
A T TP OM rr1 A
FOCAL SPOT
- .^
METALTRONICA
CAP 3 Pape 62 of 88
Or on installation of:
FILTROMAMM
When calibration of the dosemeter can not be executed during technical interventions
concerning the above parts, t is recommended to totally stop operating as described at tem
2.1
CAP 3 Page 63 of 88
i ij
1)
3)
5)
7)
9)
11)
13)
15)
17)
19)
21)
23)
25)
27)
29)
-21 .0
22.0
23.0
24.0
25.0
26.0
27.0
28.0
29.0
30.0
31.0
32.0
33.0
34.0
35.0
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV =
2)
4)
6)
8)
10)
12)
14)
16)
18)
20)
22)
24)
26)
28)
21.5
22.5
23.5
24.5
25.5
26.5
27 .5
28.5
29.5
30.5
31.5
32.5
33.5
34.5
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
....
-.
=
=
=
=
=
=
=
=
=
=
=
=
=
=
-.
'
IVI=/Rh CALIBRATION
'
1)
3)
5)
7)
9)
11)
13)
15)
17)
19)
21)
23)
25)
27)
29)
j*~
*-
^ rrn A f
O
Ulurto
2 ) 21.5
4) 22.5
6) 23.5
8) 2 4 . 5
10) 2 5 . 5
12) 2 6 . 5
14) 27 .5
16) 2 8 . 5
18) 2 9 . 5
21.0 kV
2 2 . 0 kV
23.0 kV
2 4 . 0 kV
25.0 kV
2 6 . 0 kV
2 7 . 0 kV
28.0 kV
2 9 . 0 kV
30.0 kV
31.0 kV
3 2 . 0 kV
33.0 kV
34.0 kV
35.0 kV
20)
22)
24)
26)
28)
30.5
31.5
32.5
33.5
34.5
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
2)
4)
6)
8)
10)
12)
14)
16)
18)
20)
22)
24)
26)
28)
21.5
22 .5
23 .5
24.5
25.5
26.5
27.5
28.5
29.5
30.5
31.5
32.5
33.5
34.5
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
=
,.
W/Rh CALIBRATION
1)
3)
5)
7)
9)
11)
, 13)
15)
17)
19)
21)
23)
25)
27)
29)
21.0
22.0
23.0
24 .0
25.0
26.0
27.0
28.0
29.0
30.0
31.0
32.0
33.0
34,. 0
35'. 0
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
kV
PADDLE OFFSET =
FOCAL SPOT
rom
mm
CAP 3 Page 64 of 88
CALIBRATION VERIFICATION
KV
mAs
Reference
DOSE
Mammo Dose
% Error
20,0
21 ,0
21,5
22,0
22,5
23,0
23,5
24,0
24,5
25,0
25,5
26,0
26,5
27,0
27,5
28,0
28,5
29,0
29,5
30,0
30,5
31,0
31,5
32,0
32,5
33,0
33,5
34,0
34,5
35,0
:..
MFTAI,TRONICA
CAP 3 Paee 65 of 88
MFTAT TRONTPA
CAP 3 Page 66 Of 88
^$$fj^
:<Mjjjf^^
'jn^^li^i^&i^^^^ojj^^^S^^^^_
";":;?vsK
:v'.; :';; fe;
,,S-'S*
MRTAT,TRONICA
CAP 3 Page 67 of 88
1 - INTRODUCTION
The compression device of FLAT consists of a motor with geared motor coupled to the worm
gear by an electromagnetic clutch also acting as compression power reducer.
This system, allowihg both a motorized and manual breast compression, is also equipped
with a series of interchangeable compressors of different sizes, with floor control pedis and
a luminous centering device.
3 - MANUAL COMPRESSION
Manual Compression is made possible by the clutch relase obtained by pushing
simultaneously both pedis.
When pedis are released, compression s maintained at the manually set valu if this is
equal or less than the power set on the control panel.
4 - MOTORIZED COMPRESSION
Using floor pedis, the operator can oprate Upward/Downward movements of the
compressor keeping his/her hands free.
The luminous centering device automatically switches on when the compressor lowering is
activated, In this case, the compressor starts moving with a slight delay to avoid additional
pressing n case you want to switch the centering device on again for a check, once the
breast is already compressed.
The lowering movement is mmediate when the centering device is already on.
5 - SAFETY RELASE
For safety reasons, the compression power is reduced to zero in the following cases:
1. Equipment in off position.
2. Vertical movernent of the arm
3. Rotation of the arm
METALTRONICA
CAP 3 Page 68 of 88
During calibration, verify the compression power using a bathroom scales or a dynamometer.
NOTE
Compression voltage is applied to the clutch only during lowering. During ascending, voltage
is reduced to a just compressor weight supporting valu.
CAP ^ Pase 69 of 88
11
13 -
SR-DW6024-00
CAP 3 Page 70 of 88
MHTAI
CAP 1 P n p e V I of 88
METALTRONTCA
CAP 3 Page 72 of 88
4 - LIGHT/ RA YS Coincidence
No adjustment is required after removal/re-installaton or replacement of the tube.
An adjustment of the light field is possible using the screw which adjusts the inclination of the
collimator mirror.
(i
MHTAI TRONICA
CAP 3 Pase 73 of 88
METALTRONTCA
CAP 3 Page 74 of 88
MFTAT
CAP3Page75of88
1-POWER SU PPL Y
A completely independent circuit, equipped with its own microprocessor, controls all on/off
phases of the unit.
The circuit is located on the 92082 board and is fed by the T1 transformer connected to the
mains supply.
The transformer is equipped with a 220/230/240 voltage change possibility.
2 - ON/OFF PHASES
The IC20 microprocessor controls a complex sequence of events at switching on and off.
SWITCHING ON
a) to avoid a high current inrush, RYS2 s switched on and after some seconds RYS1
excludes the damping resistance.
b) to avoid random characters on screen, LCD voltage s delayed.
SWITCHING OFF
To avoid stripes on screen, LCD voltage is removed before +5 voltage of CPU.
RYS2
-15LCD
[""
|
RYS1
OFF
ON
ACI
F1Y10
R31
-- %--' -----
=^-I
r
; R65
AC2
...
,..
je ____
RA2
w
T5"
I
1
In this picture you can see the circuit pars carrying out the functions described.
The sequence of operations can be followed observing LED8/LED9/LED1/LED2/LED5
located on the 92082 board.
MFTAJ, TRONIC A
CAP3Page76of88
3 - FUNCTIONAL VARIATIONS
The units without mobile panel have no motor or microswitch.
Various versions of the IC20 microprocessor are used in the two cases.
With fixed panel and relevant software, the JP1 selection bridge allows you to invert the
function of LED1.
JP1 = CLOSED LED 1 on with unit in off position
JP1 = OPEN LED 1 on with unit in on position
4 - TROUBLE SHOOTING
In case of'operation failure, check that conductors referring to connectors CN16 and CN28
are correctly fixed and directly check on IC20 signis relevant to parts connected to CN16
and CN28.
With unit in on position, voltage between terminis 3 and 4 of RYS1 and RYS2 shall be of
approximately 12 Vcc.
Accurately check micro-switches MSWIN and MSWOUT as their failure can result in
damages to the motor.
In case of failure of the panel output also check the geared motor and external gears.
Motor and micro-switches are obviously absent in those units which are not equipped with
mobile control panel.
In this case, a different versin of the IC20 microprocessor s used.
With a fixed panel and connected software, the JP1 jumper will allow the inversin of the
LED1 function :
JP1 CLOSED = LED 1 on when the unit s n OFF position
JP1 OPEN = LED1 on when the unit is in ON position.
IMPORTAN!
When the mobile control panel is in use, JP2 on the 92-082 board must be open and a
special IC20 different from the one concerning the configuration for fixed panel must be
used.
MPTAT TRONICA
CAP 3 Page 77 of 88
I
MFTAI TRONCA
CAP 3 Page 78 nf 88
METAT/TRONICA
CAP3Page79of88
1 -DESCRIPTION
The Mo/Rh automatic filter is an optional accessory and it can be easily installed further on.
Filtromamm s sed.'to have a double ANODE/FILTER combination (Mo/Mo Mo/Rh) for a
better optrnization of the X-ray beam spectrum, in function of the real density of the
examined breast.
the breast density s quickly determined by a short exposure in ZERO POINTS technique
and the more appropriate ANODE/FILTER combination will be automatically selected
according to the breast density.
The Mo/Mo combination is used for exams of normal density breasts, whereas the Mo/Rh
combination is used for high density breasts.
The intervention threshold beyond which the Mo/Rh combination is selected, corresponds to
approximately 6 cm of RMI Equivalent Tissue and [r] = 62 for large focus (default).
After the sampling impulse, the valu [r] is compared with the reference level of 62 for
FILM/SCREEN 1 and 2 or with the selected level with the calibration software for the
FILM/SCREEN 3. If [r] s higher than the reference valu, Mo/Mo remains positioned, if [r] is
lower, the RY1 / RY 5 circuit on the PCB 92.082, change the filter in the dead time between
the sampng impulse and exposure.
RY1
RY5
CN23
INVMOT
INVERTER MOTOR
01-170
The filter change is temporizad and it does not depend on the status of MW12.
At the end of this time, MW12 is checked and if the filter has not reached the final position,
an alarm is generated.
MFTAT.TPONTrA
CAP 3 Pase 80 of 88
FLATIIITECHNICAL MANUAL
2 ~ ELECTRICAL COMPONENTS
Motor MT4 24 Vdc
Micro-switch MW10/11 in/out detectors
MW12 FILTER IN POSITION (detected by the CPU)
!!Mo/Rh FILTER V ERSION ONLY
~VJ"
Mo
1
Rh
MT4
"\7
Rh
/j
MW1 ) I
'" i
D6
i
LI^J
L _
MW11
Mo
0
MW12
FILTER O
FILTRO O <
D8
F
3 - INSTALLATION
Filtromamm has the same dimensions as the previously existing aluminum tube support.
Beore installation, remove the existing support, the Lead ring and the fixed filter in
Molybdenum.
Then position the Tube again as described in relevant manual section.
Electncal connections are already prepared and to actvate it move DIP 8 SW1 PCB 01-170
to the ON position to enable the FILTROMAMM.
PCB 92-080
rTTTTTTT
M F T A T TRONTPA
CAP3Page81 of88
MRTALTRONICA
CAP3Page.82of88
MF.TAT TRONTPA
CAP3Page83of88
FLATIIITECHNICAL MANUAL
1-DESCRlPTION
The urtit, configured for a three-dimensional Biopsy with BYM-3D, can be equipped with an
nterface o the 50x50mrrismall field digital image acquisition systern with a digital box having
the same dimensions as those'of a 18x24 cassette.
The digital interface is an optional of the standard configuration of the unit and it can also be
added after installation.
3 - CONNECTIONS
Connections of the whole mammography system fully equipped with Digital Cassette, BYM3D and PC with Digital Imaging software can be seen in the picture in he following page.
Before connection, you need to know some important issues to maintain the electrical safety
level of the system.
This system consists of the below listed and classified elements:
Medical Devices complying with IEC 601-1 for electrical safety
FLAT mammography unit
BYM-3D Biopsy Device
Digital Cassette and relevant mains power supply
Other devices not complying with IEC 601-1 for electrical safety
Personal Computer
Monitor
As you can see in this picture, I, H, L signal connections, are all optically solated to avoid
any direct elecrical connection between Medical Devices complying with IEC 601-1 and the
others.
Connections shall be n complance with these nstructons not to modfy the electrical
safetv ievel of the svstem.
NOTE
Connections from PC to the Mammography unit and the Stereotaxis device are made by two
CV0131 (short) and CV0132 (long) cables between which the optic isolator is inserted.
The CV0131 cable SHALL be positioned on the PC side for both connections, otherwise it
will not oprate.
MPTAT TRONICA
CAP 3 Page 84 of 88
A
B
c
D
E
r-
G
H
A
B
C
0
E
E
G
H
1
I.
un/I)
AC-DWG651-00
CAP 3 Page 85 of 88
1 >
2 5
3 '
4 X
5 X
Kv tens
kV units
7 X
8 k
9 V
kVtenths
10
'-
- ' . . -
11 X
12X
13X
mAs hundredths
mAs tents
mAs units
14
15X
16m
17A
18s
19
20
21 X
22 X
mAs tenths
23
24 X
25 X
267
27 X
28 X
29 Y
30
31 X
32
33 X
34 X
35
36 D
37 X
38 X
39
40 F
41 X
42X
typeofANODE Mo, W
typeofFILTER Mo, Rh
tecnique kV ZERO POINTS s=Standard h-HIGH CONTRAST
I=LOW DOSE e=EXTENDED
G = with grid space = without grid
OP = ZERO POINTS 1 P = ONE POINT
manual density
+/number of step from -5 a +5
film/screen
film screen number from 1 to 15
METAITROMPA
L
f DOSE IF ENABLED
J
CAP 3 Pas 86 of 88
FLATIIITECHNICAL MANUAL
43$5C
44$OD
4800 baud speed
8 bits
1 stop bit
no parity
ASCII line of 44 characters
KY min 20.0 KV max 35.0 KV
MAs min 0.1 mAs max 640.0 mAs
Dimensin of focal spot
Manual density +/- 5
Film/Screen 1/16 combination
The string is transferred once at the end of the exposure.
A .TFT A l TPOMIPA
CAP 3 Page 87 of 88
METALTRONICA
CAP 3 Pape 88 of 88