Académique Documents
Professionnel Documents
Culture Documents
doi: 10.1111/pce.12315
Original Article
ABSTRACT
Since 2005, an unresolved debate has questioned whether
R-shaped vulnerability curves (VCs) might be an artefact of
the centrifuge method of measuring VCs. VCs with R-shape
show loss of stem conductivity from approximately zero
tension, and if true, this suggests that some plants either refill
embolized vessels every night or function well with a high
percentage of vessels permanently embolized. The R-shaped
curves occur more in species with vessels greater than half
the length of the segments spun in a centrifuge. Many have
hypothesized that the embolism is seeded by agents (bubbles
or particles) entering the stem end and travelling towards
the axis of rotation in long vessels, causing premature cavitation. VCs were measured on Robinia pseudoacacia L. by
three different techniques to yield three different VCs;
R-shaped: Cavitron P50 = 0.30 MPa and S-shaped: air injection P50 = 1.48 MPa and bench top dehydration P50 =
3.57 MPa. Stem conductivity measured in the Cavitron was
unstable and is a function of vessel length when measured
repeatedly with constant tension, and this observation is discussed in terms of stability of air bubbles drawn into cut-open
vessels during repeated Cavitron measurement of conductivity; hence, R-shaped curves measured in a Cavitron are probably invalid.
Key-words: Cavitron; centrifuge method; exponential curves;
vulnerability curve.
INTRODUCTION
Plants transport metastable water (tensile water) under negative pressure (Tyree & Zimmermann 2002). The transport
system is vulnerable to embolisms because if an air bubble
enters a conduit filled with tensile water, the bubble (or
vacuum void) will expand to fill the entire conduit, rendering
it incapable of further transport of tensile water until it can be
refilled by some mechanism. Transport of tensile water is
possible because of the unique structure of xylem conduits,
which tend to be numerous for pathway redundancy and
Correspondence: J. Cai. Fax: +86 029 87080363; e-mail: cjcaijing
@163.com; and M. T. Tyree. Fax: +86 029 87080363; e-mail: mel.tyree
@cantab.net
2014 John Wiley & Sons Ltd
2668
R. Wang et al.
Length of vessels
Vessel length was measured by the air method as described
by Cohen et al. (2003), with some modification. Briefly, long
shoots were cut with a sharp razor blade and injected with
compressed air (0.15 MPa) from the distal side of 6-mmdiameter stems while the basal end was immersed in water.
Stem segments (2 cm long) were sequentially excised until
bubbling was observed, and the length of remaining stem was
recorded as the maximum length of the vessel.
The bubbles flowing out from the basal end were collected
in a volumetric cylinder by the water displacement method,
and the volume of air was determined when the water surface
inside and outside the cylinder were at the same level in
order to be sure that the air volume was measured at atmospheric pressure. The gravimetric water displacement method
was used to measure air volume since this allowed more
precision than reading the scale on the volumetric cylinder.
Firstly, the cylinder was filled to the top with water, and then
the water was delivered to a balance to measure the
maximum water content (Wi). After measuring the air displacement for a measured time interval (T), the cylinder
was held at a shallow angle with the open end below the
water and the air/water interface at the water level. The end
of the cylinder was covered to avoid spilling of water and the
cylinder was lifted out of the water. The water remaining in
the cylinder was again delivered to a balance and measured
(Wf). The air flow rate [Q (mL/min)] was computed from
(Wi Wf)/(T), where is the density of water displaced by
the air.
The air flow rate [Q (mL/min)] was measured at different
stem lengths with the air pressure at the inlet and outlet ends
being held constant. Under such conditions of constant air
pressure drop, the hydraulic conductivity of the cut-open
vessels to the air (C) should be proportional to Qx (see eqn
4 in Cohen et al. 2003), where x is the stem length at which the
air flow rate (Q) was measured. Using the theory of Cohen
et al. (2003), we assume that
C = C0 exp ( kx ) or Qx = Q0 x0 exp ( kx )
Plant material
Experiments were conducted from May to December in 2013
on R. pseudoacacia L. trees growing about 5 m above the
water level on the bank of the Weihe River in Yangling,
Shaanxi, China (3416N 1084 E, 457 m a.s.l.). Most samples
were collected from four trees of 57 m tall. All of the measurements were carried out on current-year stems harvested
from the southern crown of the trees. For laboratory experiments, shoots of 1.42 m long were excised because preliminary air injection experiments indicated that maximum
vessel length was <1 m. After spraying leaves with water,
(1)
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
Vulnerability curves
VCs were measured by the Cochard Cavitron technique
(Cochard 2002; Cochard et al. 2005) using a custom-modified
centrifuge. Details about the construction and use of the
Cochard Cavitron are described in Cai & Tyree (2010).
Briefly, the Cavitron allows measurement of stem-segment
hydraulic conductivity while the stems are spinning in a centrifuge. The maximum conductivity (Kmax) was measured at a
slow spin rate, which causes little or no embolism. Increasing
the spin rate increased the tension (T) and the stem conductivity Kh declined. Percentage loss conductivity was computed from
(2)
(3a)
(3b)
2669
P50 = B [ ln ( 2 )]
1C
(4)
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2670
R. Wang et al.
100
(100 NPLC )
(5)
Stomatal conductance
Stomatal conductance (gs) was measured using a Li-6400XT
portable photosynthesis system (Li-Cor Inc., Lincoln, NE,
USA) in a sunny day in August. Ten leaves from the four
sample trees were tagged and measured from 0700 to 2000 h
at 1 h intervals. Leaves were measured with natural sunlight,
air temperature and ambient CO2 while keeping the leaves in
their native growing position during the measurement.
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2671
RESULTS
Diameter and length of vessels
The frequency distribution of vessel diameter percentage of
vessels versus bin diameter size class (D) is shown in Fig. 1a
together with the hydraulic weight of the size classes. The
hydraulic weight of each size classes is defined as %
vessels D4 in each diameter size class. The fourth power is
used to weigh the hydraulic conductivity of each bin size class
according to the HagenPoiseuille law. It can be seen that
large-diameter vessels have a disproportionate hydraulic
weight relative to the percentage of each bin diameter size
class. Staining was used to determine which vessels were
embolized (unstained) in the native state; the number of
embolized vessels was 269 compared to the total vessels
measured, 1239 (stained and unstained). In terms of hydraulic weight, the embolized vessels tended to be a little smaller,
where the mode diameter of embolized vessels was about
60 m versus 75 m for all vessels.
We chose to measure vessel length by the air injection
technique (Cohen et al. 2003) because it also averages
(weighs) the vessel length towards the more conductive conduits. In the air injection technique, one can plot ln(C/C0)
versus distance from the air injection surface, where C is the
pneumatic conductivity of the stem at distance x and C0 is C
at x = 0. The use of C for pneumatic flow is similar to water
flow, which, like the HagenPoisseuille law, weighs air flow by
the fourth power of the diameter of the conduit through
which the air flows. According to Cohen et al. (2003), the
slope of ln(C/C0) versus distance x from the injection surface
can be used to calculate vessel length. Three representative
plots out of 12 experiments in which air was injected from the
apical or basal end of six stems each are shown in Fig. 1b.
Other advantages of the air injection technique are that it is
faster than other standard techniques (Cai et al. 2010) and
has a high R2 (0.99) for the regression of the plot of ln(C/C0)
versus x used to determine mean vessel length. A final advantage of normalizing the plot of C to C0, that is, ln(C/C0), is that
the anti-log directly yields the fraction of vessels still open at
any given distance x from the injection surface (see right axis
values in % in Fig. 1b).
There was no significant difference between vessel length
computed whether air was injected from the apex or base of
stems, so the values were pooled to yield a mean vessel length
of 20 1.9 cm (n = 12); all errors in this article are standard
errors of the mean, unless otherwise stated. However, there
was a large variation in the mean vessel length between stems
between a maximum and minimum of 30.0 and 12.7 cm,
respectively. The length of stem segments used in our
Cavitron is 27.4 cm, the mean vessel length and percentage of
vessels still open at the length and half-length of these segments are given in Table 1 for each segment as well as the
overall means.
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2672
R. Wang et al.
Table 1 Shown are vessel lengths of n = 12 stems injected with air from the apex or base
Method
Vessel length
(cm)
% > 27.4 cm
long
% > 13.7 cm
long
Longest
vessel (cm)
Apex injection
Apex injection
Apex injection
Apex injection
Apex injection
Apex injection
Base injection
Base injection
Base injection
Base injection
Base injection
Base injection
Mean
SD
SE
17.2
32.0 (max)
19.2
16.8
15.9
27.9
14.9
29.0
22.1
12.7 (min)
16.3
16.1
20.0
6.3
1.9
4.1%
18.0%
5.8%
3.8%
3.2%
14.1%
2.6%
15.1%
8.3%
1.3%
3.4%
3.3%
6.9%
5.7%
1.7%
20.3%
42.5%
24.0%
19.5%
17.8%
37.5%
16.0%
38.9%
28.9%
11.6%
18.5%
18.2%
24.5%
10.1%
3.0%
61
59
57
54
56
58
32
53
46
44
51
42
Maximum and minimum values are indicated in the table; however, in Fig. 4b, we report a mean vessel length of 5.2 cm in one stem segment. The
longest vessel length has a resolution of 2 cm (=length of segments excised).
slowed down as the tension grew higher than 0.599 MPa and
reached 95% at 2.947 MPa. The shapes of VCs were not
dramatically changed regardless of whether there is flushing
(Fig. 2a) or not (Fig. 2b) prior to the VC measurements, and
the P50 values were not significantly different (P = 0.941),
with the values of 0.319 and 0.323 MPa for flushed and
unflushed stems, respectively.
The VC constructed by the bench top dehydration technique is shown in Fig. 2b, which displayed an S-shaped
curves. The PLC stayed at low levels (<10%) until tension
exceeded 2.9 MPa. PLC dramatically increased from 6.39 to
97.27% as xylem tension rose from 2.9 to 4.6 MPa. The P50
equalled 3.57 MPa for bench top dehydration, which was
about 11 times the P50 value calculated from Cavitron VCs.
The VCs obtained by the air injection technique were also of
Table 2 Shown are the parameters from the single and dual
Weibull fits (WF) in Eqns 3a and 3b, respectively
Dual WF
B1
C1
B2
C2
RMSerror
Mean
SD
SE
P1/2
0.280
1.595
1.695
1.240
68.2
2.393
0.081
0.717
1.037
0.731
10.389
0.854
0.022
0.199
0.288
0.203
2.881
0.237
0.223
Mean
SD
SE
P50
0.436
0.939
5.516
0.124
0.233
1.027
0.034
0.065
0.285
0.295
1.261
Single WF
B
C
RMSerror
The RMSerror of the fits are significantly different (P < 108), n = 14.
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2673
There was no significant difference in Ki, Kmax or PLC measured by the LPFM or Cavitron (data not shown).
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2674
R. Wang et al.
the segment. This experiment was possible because of the
high variability in mean vessel length between Robinia segments. It can be seen that the percentage of the initial Kh is a
sigmoid function of the mean vessel length. These results are
consistent with the PSE hypothesis and are consistent with
the existence of very vulnerable particles that cause embolism at low tension.
DISCUSSION
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2675
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2676
R. Wang et al.
2
= ( Pb PW )
r
(6a)
Tcrit =
2
r
(6b)
Figure 5 shows a plot of Tcrit versus r. The open circles represent the T values used in Fig. 4. In Fig. 4a, the RPM was 500,
corresponding to T = 32 kPa, and in Fig. 4b, the RPM was 900,
corresponding to T = 72 kPa. From the critical tension curve,
we can conclude that the bubbles needed to be bigger than
4.5 m (T = 32 kPa) and 2.0 m (T = 72 kPa) to expand and
cause a decline in Kh as the bubbles passed through the centre
of the stem segments. From Fig. 1a (squares), it can be seen
that the (most common) vessel radius that accounts for Kh is
around 35 m (70 m diameter on the x-axis); hence, bubbles
as large as predicted in Fig. 5 could easily pass through the
vessels of Robinia because vessel elements have simple perforation plates capable of passing air bubbles >5 m (Sherwin
Carlquist, personal communication) and this is confirmed
using a maceration technique.
More work needs to be done to confirm the suggestions
advanced in the previous paragraph, that is, that bubbles
cause the time-dependent embolism in Fig. 4 at constant
tension. However, the results in Fig. 4 stand even if the effervescent water hypothesis turns out to be wrong.
Figure 5. (a) Shown is a plot of Eqn 6b. The y-axis is the critical
tension for bubble stability versus bubble radius on the x-axis.
When a bubble experiences water at a tension more than Tcrit, it
should expand to seed a cavitation and the consequent loss of Kh,
and when the water tension is <Tcrit, it should slowly collapse as
the air dissolves in the surrounding water. (b) This diagram shows
the regions of instability for a bubble as it passes through a stem
segment in a Cavitron spinning with a central tension of 72 kPa.
The bell-shaped curve is a plot of tension versus distance from the
axis of rotation. A bubble of 9 m radius will have a critical
tension of 25 kPa. In the grey areas, the bubble will shrink, and in
the clear areas, it will expand. The bubble is stable only at the
interface between the grey and clear areas.
(7)
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
CONCLUSIONS
The R-shaped curves are invalid because they do not fit with
the water relations of Robinia (Fig. 3), because the R-shaped
curve predicted a very high midday NPLC not observed in our
data. The preponderance of evidence presented in this article
supports the notion that the air bubbles cause premature
2677
ACKNOWLEDGMENTS
The authors wish to acknowledge the following grants that
made this research possible: Natural Science Foundation of
China (Grant No. 31270646) to J.C. and the thousand talent
program grants to M.T.T.
REFERENCES
An F., Cai J., Jiang Z., Zhang Y., Lan G. & Zhang S. (2006) Refilling of
embolism in the xylem of eight tree species and its relationship with
pressure-volume parameters. Journal of Northwest A & F University
(Natural Science Edition) 34, 3844.
Arango-Velez A., Zwiazek J.J., Thomas B.R. & Tyree M.T. (2011) Stomatal
factors and vulnerability of stem xylem to cavitation in poplars. Physiologia
Plantarum 143, 154165.
Brodersen C.R., McElrone A.J., Choat B., Matthews M.A. & Shackel K.A.
(2010) The dynamics of embolism repair in xylem: in vivo visualizations
using high-resolution computed tomography. Plant Physiology 154, 1088
1095.
Cai J. & Tyree M.T. (2010) The impact of vessel size on vulnerability curves:
data and models for within-species variability in saplings of aspen, Populus
tremuloides Michx. Plant, Cell & Environment 33, 10591069.
Cai J., Zhang S. & Tyree M.T. (2010) A computational algorithm addressing
how vessel length might depend on vessel diameter. Plant, Cell & Environment 33, 12341238.
Cai J., Li S., Zhang H., Zhang S. & Tyree M.T. (2014) Recalcitrant vulnerability
curves: methods of analysis and the concept of fibre bridges for enhanced
cavitation resistance. Plant, Cell & Environment 37, 3544.
Cochard H. (2002) A technique for measuring xylem hydraulic conductance
under high negative pressures. Plant, Cell & Environment 25, 815819.
Cochard H., Cruiziat P. & Tyree M.T. (1992) Use of positive pressures to
establish vulnerability curves: further support for the air-seeding hypothesis
and implications for pressure-volume analysis. Plant Physiology 100, 205
209.
Cochard H., Breda N. & Granier A. (1996) Whole tree hydraulic conductance
and water loss regulation in Quercus during drought: evidence for stomatal
control of embolism? Annales des Sciences Forestieres. Sciences 53, 197
206.
Cochard H., Damour G., Bodet C., Tharwat I., Poirier M. & Amglio T. (2005)
Evaluation of a new centrifuge technique for rapid generation of xylem
vulnerability curves. Physiologia Plantarum 124, 410418.
Cochard H., Herbette S., Barigah T., Badel E., Ennajeh M. & Vilagrosa A.
(2010) Does sample length influence the shape of xylem embolism vulnerability curves? A test with the Cavitron spinning technique. Plant, Cell &
Environment 33, 15431552.
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678
2678
R. Wang et al.
Cochard H., Badel E., Herbette S., Delzon S., Choat B. & Jansen S. (2013)
Methods for measuring plant vulnerability to cavitation: a critical review.
Journal of Experimental Botany 64, 47794791.
Cohen S., Bennink J. & Tyree M.T. (2003) Air method measurements of apple
vessel length distributions with improved apparatus and theory. Journal of
Experimental Botany 54, 18891897.
Crombie D., Milburn J. & Hipkins M. (1985) Maximum sustainable xylem sap
tensions in Rhododendron and other species. Planta 163, 2733.
Ennajeh M., Nouiri M., Khemira H. & Cochard H. (2011) Improvement to the
air-injection technique to estimate xylem vulnerability to cavitation. Trees
25, 705710.
Jones H. & Sutherland R. (1991) Stomatal control of xylem embolism. Plant,
Cell & Environment 14, 607612.
Salleo S., Lo Gullo M.A., De Paoli D. & Zippo M. (1996) Xylem recovery from
cavitation-induced embolism in young plants of Laurus nobilis: a possible
mechanism. New Phytologist 132, 4756.
Sperry J. (2013) Cutting-edge research or cutting-edge artifact? An overdue
control experiment complicates the xylem refilling story. Plant, Cell & Environment 36, 19161918.
Sperry J., Donnelly J. & Tyree M. (1988) A method for measuring hydraulic
conductivity and embolism in xylem. Plant, Cell & Environment 11, 3540.
Sperry J.S., Christman M.A., Torres-Ruiz J.M., Taneda H. & Smith D.D. (2012)
Vulnerability curves by centrifugation: is there an open vessel artefact, and
are r shaped curves necessarily invalid? Plant, Cell & Environment 35,
601610.
Taneda H. & Sperry J.S. (2008) A case-study of water transport in co-occurring
ring-versus diffuse-porous trees: contrasts in water-status, conducting capacity, cavitation and vessel refilling. Tree Physiology 28, 16411651.
Tyree M.T. & Dixon M.A. (1986) Water stress induced cavitation and embolism in some woody plants. Physiologia Plantarum 66, 397405.
Tyree M.T. & Sperry J.S. (1988) Do woody plants operate near the point of
catastrophic xylem dysfunction caused by dynamic water stress? Answers
from a model. Plant Physiology 88, 574580.
Tyree M.T. & Yang S. (1992) Hydraulic conductivity recovery versus water
pressure in xylem of Acer saccharum. Plant Physiology 100, 669676.
Tyree M.T. & Zimmermann M.H. (2002) Xylem Structure and the Ascent of
Sap, 2nd edn, Springer-Verlag, Berlin.
Tyree M.T., Cochard H., Cruiziat P., Sinclair B. & Ameglio T. (1993) Droughtinduced leaf shedding in walnut: evidence for vulnerability segmentation.
Plant, Cell & Environment 16, 879882.
Wang M., Tyree M.T. & Wasylishen R.E. (2013) Magnetic resonance imaging
of water ascent in embolized xylem vessels of grapevine stem segments.
Canadian Journal of Plant Science 93, 879893.
Wheeler J.K., Huggett B.A., Tofte A.N., Rockwell F.E. & Holbrook N.M.
(2013) Cutting xylem under tension or supersaturated with gas can generate
PLC and the appearance of rapid recovery from embolism. Plant, Cell &
Environment 36, 19381949.
Yang S. & Tyree M.T. (1992) A theoretical model of hydraulic conductivity
recovery from embolism with comparison to experimental data on Acer
saccharum. Plant, Cell & Environment 15, 633643.
Zhang S. & Richter H. (1996) The refilling of embolized xylem in Taxus
baccata L. Journal of Northwest Forestry College 11, 58.
Zwieniecki M. & Holbrook N. (1998) Diurnal variation in xylem hydraulic
conductivity in white ash (Fraxinus americana L.), red maple (Acer rubrum
L.) and red spruce (Picea rubens Sarg.). Plant, Cell & Environment 21,
11731180.
2014 John Wiley & Sons Ltd, Plant, Cell and Environment, 37, 26672678