PA 349 - 06/1

Programming instructions
CADAS 100 LPG 210

Phosphate-Phosphorus

Cuvette-Test
LCK 349 Page 1

ortho and total

Edition 01/06

Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key
on the keyboard.

Programming of LCK 349 Phosphate-Phosphorus


 Select the program VAR.
Select function key below symbol 349 E.
Phosphate-Phosphorus (PO4-P) (Program VAR)
Display

Input via keyboard

Program number:

for example 1

Symbol:

349 E

Name:

PO4 - P LCK 349 E

Pro:

STA

FPRG

Press function key below 1

= 850 nm

W:
Factor:

1.595

Constant:

- 0.009

Unit:

mg/l

Measuring limit:

yes

Lower limit:

0.05 mg/l

Upper limit:

1.5 mg/l

no

Press function key below yes

Programming of LCK 349 Phosphate


 Select the program VAR.
Select function key below symbol 349.
Phosphate (PO4 ) (Program VAR)
Display

Input via keyboard

Program number:

for example 1

Symbol:

349

Name:

PO4 LCK 349

Pro:

STA

FPRG

Press function key below 1

= 850 nm

W:
Factor:

4.89

Constant:

- 0.016

Unit:

mg/l

Measuring limit:

yes

Lower limit:

0.15 mg/l

Upper limit:

4.5 mg/l

no

Press function key below yes

All data entries are documented on the printer; their transfer to storage is confirmed by the printer message stored.
You should check the printed data to ensure that all entries are correct.

PA 349 - 06/1

Programming instructions
CADAS 100 LPG 210

Phosphate-Phosphorus

Cuvette-Test
LCK 349 Page 2

ortho and total

Edition 01/06

Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key
on the keyboard.

Programming of LCK 349 Phosphorus-Pentoxide


 Select the program VAR.
Select function key below symbol 349 P.
Phosphorus-Pentoxide (P2 O5 ) (Program VAR)
Display

Input via keyboard

Program number:

for example 1

Symbol:

349 P

Name:

P2 O5 LCK 349 P

Pro:

STA

FPRG

Press function key below 1

= 850 nm

W:
Factor:

3.654

Constant:

- 0.01

Unit:

mg/l

Measuring limit:

yes

Lower limit:

0.15 mg/l

Upper limit:

3.50 mg/l

no

Press function key below yes

All data entries are documented on the printer; their transfer to storage is confirmed by the printer message stored.
You should check the printed data to ensure that all entries are correct.