Programming instructions
CADAS 100 LPG 210
Phosphate-Phosphorus
Cuvette-Test
LCK 349 Page 1
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key
on the keyboard.
Program number:
for example 1
Symbol:
349 E
Name:
Pro:
STA
FPRG
= 850 nm
W:
Factor:
1.595
Constant:
- 0.009
Unit:
mg/l
Measuring limit:
yes
Lower limit:
0.05 mg/l
Upper limit:
1.5 mg/l
no
Program number:
for example 1
Symbol:
349
Name:
Pro:
STA
FPRG
= 850 nm
W:
Factor:
4.89
Constant:
- 0.016
Unit:
mg/l
Measuring limit:
yes
Lower limit:
0.15 mg/l
Upper limit:
4.5 mg/l
no
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message stored.
You should check the printed data to ensure that all entries are correct.
PA 349 - 06/1
Programming instructions
CADAS 100 LPG 210
Phosphate-Phosphorus
Cuvette-Test
LCK 349 Page 2
Edition 01/06
Attention! The edition date of this programming instruction must be identical to the edition date of the instruction
attached to the reagents. This is the only way to ensure that your instrument will be programmed with the correct data.
Please note!
If you make an error during programming, you can cancel the last entry by pressing the key
on the keyboard.
Program number:
for example 1
Symbol:
349 P
Name:
P2 O5 LCK 349 P
Pro:
STA
FPRG
= 850 nm
W:
Factor:
3.654
Constant:
- 0.01
Unit:
mg/l
Measuring limit:
yes
Lower limit:
0.15 mg/l
Upper limit:
3.50 mg/l
no
All data entries are documented on the printer; their transfer to storage is confirmed by the printer message stored.
You should check the printed data to ensure that all entries are correct.