TECHNICAL BULLETIN >>

TSH Immunoassay

OVERVIEW Measurement of TSH (Thyroid Stimulating (MES) buffer, pH 5.0. Detection anti-TSH
Luminex incorporates a proprietary process
® Hormone or Thyrotropin) is useful in screening monoclonal antibodies were biotinylated
to internally dye polystyrene microspheres for both hyperthyroidism, where TSH levels using EZ-Link-Sulfo-NHS-LC-Biotin (Pierce).
with two spectrally distinct fluorochromes. may be very low, and hypothyroidism, where The TSH calibration material was prepared in
Using precise ratios of these fluorochromes, TSH levels may be high. To date, no single horse serum to mimic human samples.
an array is created consisting of 100 different commercial assay is available that exhibits >> Reagent Selection: Antibodies were
microsphere sets – each with its own charac- both the exquisite sensitivity and broad screened for sensitivity by coupling thirteen
teristic spectral address. Each microsphere dynamic range required for screening for both different monoclonal antibodies to xMAP
can possess a different reactant on the surface. conditions. Using the xMAP® system, we microspheres and testing with 10 biotinylated
Since microsphere sets can be distinguished by developed a TSH immunoassay that combines detection monoclonal antibodies. From
their spectral addresses, they can be combined the third-generation sensitivity and the these results, OEM Concepts antibody,
to allow up to 100 different analytes to be dynamic range necessary for practical use as clone # 057-11003 was selected as the capture
measured simultaneously in a single reaction an analytical tool. antibody and Medix Biochemica antibody,
vessel. A third fluorochrome coupled to a clone # 5403, and a second antibody were
reporter molecule quantifies the biomolecular MATERIALS AND METHODS selected as detection antibodies.
interaction that has occurred at the microsphere >> Preparation of Assay Reagents: Anti-TSH Utilizing this antibody combination, we devel-
surface. Microspheres are interrogated indi- monoclonal antibodies were coupled to oped a 20 minute TSH assay with a sensitivity
vidually in a rapidly flowing fluid stream as xMAP microspheres using the recommended of <0.01 µIU/mL (Table 1).
they pass by two separate lasers in the Luminex procedure. The microspheres were In addition to sensitivity and speed, we
Luminex 100 TM
analyzer. High speed digital activated using 1-ethyl-3-(3-dimethylamino- sought to develop a TSH assay exhibiting
signal processing classifies the microsphere propyl) carbodiimide (EDC) and N-hydroxy- broad dynamic range. Using additional bead
based on its spectral address and quantifies sulfosuccinimide (Sulfo-NHS) in 0.1 M sodium sets, we included two more anti-TSH antibodies
the reaction on the surface in a few seconds phosphate buffer, pH 6.3 and the antibody (OEM Concepts antibody, clone # 204-12252,
per sample. We utilized this platform to develop was coupled to activated microspheres in and Medix Biochemica anitbody, clone # 5401)
a third-generation immunoassay for TSH. 0.05M 2-(N-morpholino) ethanesulfonic acid that had different affinities than the original
capture antibody.
Figure 1 shows that by using 3 different
antibody-bead sets, the assay had a dynamic
Table 1. Low-end Calibration Curve for TSH Assay
range greater than 5 orders of magnitude
(<0.01 to > 1000 µIU/mL).
TSH Calibrators Average Median Fluorescent SD
(µIU/mL) Intensity* (MFI) 4 reps
Assay Components:
0.08 524 112
>> Capture antibodies
0.04 275 50
0.02 148 38 2.5 µg antibody coupled per 106 beads
0.01 93 8 1000 beads for each of the 3 capture anti-
0.005 77 28 body bead sets
0.0025 39 9 PBS-BSA (1% BSA) buffer containing 2%
0.00 27 14 polyethylene glycol (MW 15,000)
>> Detection antibodies
*Uncoated bead fluorescence subtracted 20 µg/mL each
PBS-BSA (1% BSA)
Figure 1 Figure 2

Table 2. Intra-Instrument Assay Precision Table 3. Results for Samples below Limit of Detection for ACS 180

ACS180 xMAP
Antibody Bio-Rad Expected Observed Samples TSH Conc (µIU/mL) TSH Co
Bead Affinity Controls µIU/mL Std Dev % CV (µIU/mL)
040 High Affinity Control 1 0.48 0.64 0.03 5.39 CPL 26 <0.1 0.05
021 Med Affinity Control 2 5.00 4.20 0.22 5.24 CPL 27 <0.1 0.06
052 Low Affinity Control 3 28.00 21.00 1.39 6.64 CPL 28 <0.1 2.15
CPL 32 <0.1 0.02
CPL 36 <0.1 0.02
CPL 37 <0.1 0.05
CPL 38 <0.1 0.03
CPL 39 <0.1 0.02
CPL 41 <0.1 0.02

>> Wash buffer – PBS-BSA (1% BSA) assess the limit of detection for this xMAP ACS 180 and is significantly more sensitive.
>> Streptavidin-R-Phycoerythrin (100 µg/mL) TSH Assay. Defining the LDD as the mean
- in PBS-BSA (1% BSA) of the zero plus 3 SD, the LDD for this TSH CONCLUSIONS
>> Assay Protocol: Add 25 µL of patient sam- assay was 0.006 µIU/mL. These results demonstrate the use of xMAP
ple (undiluted serum) to a well. Add 10 µL >> Assay Correlation: Clinical samples were technology to develop a rapid, simple, third
of capture antibody-coupled beads and obtained from Clinical Pathology generation TSH assay with the combined sen-
incubate 10 minutes at 40°C. Add 10 µL of Laboratories (Austin, TX) with assigned sitivity and sufficiently large dynamic range
biotinylated detection antibodies and incu- values for TSH determined on the ACS to be useful as a clinical assay for TSH.
bate 5 minutes at 40°C. Add 10 µL of strep- 180. A total of 96 samples were tested
tavidin-R-phycoerythrin (SA-RPE) and incu- with this TSH Assay. Reportable results The xMAP TSH Assay:
bate 5 minutes at 40°C. Transfer to a filter were obtained for 89 samples of which 80 >> Sensitive – 0.006 µIU/mL
plate and wash 2 times with 200mL wash were within the dynamic range of the ref- >> Simple – requires only one wash step
buffer. Resuspend in 110 µL wash buffer erence ACS 180 TSH Assay (Figure 2). >> Rapid – complete in 20 minutes
and read on the Luminex 100 with the PMT Four samples were excluded from this corre- >> Accurate – correlates with the ACS 180
set to 800 mV. lation because there was a large difference in TSH Assay (r2 = 0.976)
the values of the two assays. The cause of the >> Practical – broad dynamic range ( >1000
RESULTS discrepancies is unknown and is under inves- µIU/mL)
Bio-Rad Controls were used to assess intra- tigation. Nine samples were below the limit of
assay precision with this xMAP TSH Assay. Ten detection for the reference assay, but were Call toll-free 1-888-219-8020 in North America,
replicates of each of the three Bio-Rad controls reportable with our xMAP TSH Assay (Table call 1-512-219-8020 outside of North America,
were measured on one plate (Table 2). 3). The result for one sample was at variance visit www.luminexcorp.com, or email
>> Assay Limit of Detection: Fifteen repli- with the reference value. These data demon- info@luminexcorp.com for more information.
cates of the zero calibrator were used to strate that our TSH assay correlates with the

United States Europe www.luminexcorp.com


Luminex Corporation Luminex B.V. info@luminexcorp.com
12212 Technology Blvd. Krombraak 15
Austin, Texas 78727 4906 CR Oosterhout
United States The Netherlands
Tel: 512.219.8020 Tel: +31.162.408333
Fax: 512.219.5195 Fax: +31.162.408337

© 2004 Luminex Corporation. xMAP, Luminex 100, Luminex HTS are trademarks of Luminex Corporation.
Specifications and features subject to change without notice.
070104-01-003-02