Question 1
(a) A tungsten /5% rhenium-tungsten /26% rhenium thermocouple has an output e.m.f. as shown in
the following table when its hot (measuring) junction is at the temperature shown. Estimate the
sensitivity of measurement for the thermocouple in mV/0C.
0
C
mV
200
4.37
400
8.74
600
13.11
800
17.48
[8 marks]
(b) The following data are observed in one electrical laboratory work meter measurement
R = 2200 1%
I = 20 mA 2%
Compute the power and probable error.
[12 marks]
Question 2
An AC voltmeter has a 100A movement and a half wave rectifier. If it is set on its 10V range to
measure the voltage across the 15k resistor of Figure 1, estimate:
(i)
(ii)
[10 marks]
[10 marks]
Figure 1
Section B
CONFIDENTIAL
Question 1
A deflection bridge circuit, as shown in Figure 2, is used to measure the value of unknown
capacitance CX. The components in the bridge have the following values:
Vs=20Vrms, C1=100F, R2=60, R3=40
If CX=100F, calculate the output voltage VO.
[20 marks]
Figure 2
Question 2
CONFIDENTIAL
(a)
The following results are obtained for cooper-constant thermocouple after measured heat.
Temperature
(oC)
0
370
The resistance of a copper RTD at 25 oC and 0 oC is 900 and 100.0 respectively. Calculate the
RTD resistance at 40 oC when its temperature coefficient is 0.0038 oC-1.
[6 marks]
(c)
A semiconductor thermistor, =3510K, is used in order to monitor the temperature of the heater
house system. The resistance at 68oF is 1500. Calculate the final resistance if the temperature
is increased to 104oF.
[oK=oC+273.15; oK= (oF-32)*(9/5) +273.15; F=(C*(9/5))+32]
[8 marks]
Question 3
CONFIDENTIAL
Figure 3
Question 4
A PMMC meter movement having a full scale deflection of 500A and a coil resistance of 150 is
to be used as part of an AC voltmeter with full-wave rectification. If the voltmeters full scale
reading 15Vrms. Determine the meters:
CONFIDENTIAL
(a) The DC sensitivity
[4 marks]
(b) The AC sensitivity
[4 marks]
(c) The value of the multiplier resistor, Rs
[12 marks]