Académique Documents
Professionnel Documents
Culture Documents
1. MODEL NO.
: AA-7000
a. Power Requirements
b. Servo Voltage Stabilizer
:
:
c. Power Points
3. GAS REQUIREMENTS
a. Air
b. Acetylene
c. Nitrous Oxide
6.
INSTALLATION SPACE :
a. Required bench space for spectrophotometer alone is approx. 1100W(additional 200mm if
autosampler purchased) x 750D mm(additional 300mm if autosampler purchased) x 600Hmm.
Height of the Duct from bench top should be approx. 1000 mm. Additional space will be
required to accommodate PC, Printer and any optional accessories if you have procured the
same.
b. Installation bench should be capable of supporting approximately 70kg. of gross
spectrophotometer weight (additional 24Kg if autosampler purchased). PC, Printer and any
other accessory will add to this weight.
c. Proper place should be provided for keeping Air compressor so that standard gas tubing given
with instrument can be used from compressor to AA-6300.
7. VENTILATION SYSTEM
.
Damper
Fan
1100mm to 1200mm
8.
ENVIRONMENTAL CONDITIONS :
a. Room should be dust free with an air-conditioner of sufficient capacity recommended to
maintain temperature between 10oC to 35oC and operating relative humidity between 20% to
80%.
9. PRECAUTIONS :
Installation should be in a non corrosive and Dust free atmosphere.
a) Room should be Sufficiently Ventilated.
b) The equipment should not be exposed to direct sunlight, direct blow from an air conditioner or
high temperatures.
c) The equipment should not be installed in the vicinity of strong and continuous vibrations or near
electrical units generating intense electric / magnetic fields and high frequency power
d) Ambient temperature fluctuations during operations should be minimum.
10. LAMP REQUIRMENTS:
Any Hollow Cathode Lamp required for your analysis.
11. Standard Stock solutions requirement:
Standard Stock solutions for your elements should be of MERCK or Sigma Aldrich make of
1000ppm and they are to be procured from your side.
All necessary accessories for the dilution of above standards and sample preparations are to be kept
ready from your side.
12. COMPUTER REQUIREMENTS