1. Features
1. Various Measurements Supported.
Not only the usual refractometry and keratometry, but also corneal
diameter and base curve of contact lens can be measured with this
one instrument. Thus, measurements of eye and prescriptions for
glasses and contact lenses can be made more efficiently.
The foggy method of the eye fixation target makes examinees eye
comfortable and enables to get more accurate measurement data.
3. To connect this with other relevant equipments, consult with the dealer
about the way to make it.
5. Keep the objective glass of the examinee side clean. If it was stained, it
may cause on ERROR or inaccurate measurements.
8. When moving this URK-700, fix the stage by using clamping bolt and
stage holding knob, always keep power off, and then lift the bottom of
the unit with both hands.
9. If you leave URK-700 without using for certain period, disconnect the
power supply and protect the unit with dust cover.
3. Description.
3.1 Main Unit.
LCD Monitor
Functional Key
Measurement
button
Stage Clamping
Bolt
Printer
Headrest
Measurement
Chinrest
Stage Holding
EXT Video
RS-232 Port
Power IN LET
Fuse
Power Switch
Title
Technical skill
MODE
CLEAR
MENU
IOL
To measure the
Manual measurement
A.
Adjust the position and focus on the Mode Eye like the procedure 6 .
B.
C.
Automatic measurement
A.
B.
menu will be
When the bright dot enters the inner alignment mark and Model Eye is in focus
properly measurement starts automatically.
5. Measurement
5.1 Continuous Keratometry and Refractometry ( K/R Mode)
In the K/R Mode refractometry is performed after keratometry.
Press MODE button until R/K is displayed on the top of the display.
Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B.
Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
headrest.
Slide the operation joystick to the left so that the right eye of the examinee is
displayed on the monitor.
B.
Ask the examinee to look at the red balloon in the center of the scene (eye
fixation target).
C.
Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keep his/her eye opened wide until measurement ends.
D.
Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centered with the inner alignment mark
and focus on the KERATO RING, at the time adjusted the bar would be
appeared above the inner alignment mark.
(4) Measurement
A.
B.
C.
According to the 6.1 SETUP MODE, you can select and change VD, CYL
points.
B.
C.
CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
Slide the stage to the right side and measure the left eye.
B.
After both eyes are measured, pupil distance (PD) will be displayed on the
monitor.
(7) Print
A.
B.
Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B.
Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
face rest.
Slide the operation joystick to the left so that the right eye of the examinee is
displayed on the monitor.
B.
Ask the examinee to look at the red balloon in the center of the scene (eye
fixation target).
C.
Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keep his/her eye opened widely until measurement ends.
D.
Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centered with the inner alignment mark
and focus on the KERATO RING, at the time adjusted the bar would be
appeared above the inner alignment mark.
(4) Measurement
A.
B.
C.
According to the 6.1 SETUP MODE, you can select and change VD, CYL
points.
B.
C.
CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
Slide the stage to the right side and measure the left eye.
B.
After both eyes are measured, pupil distance (PD) will be displayed on the
monitor.
(7) Print
A.
B.
Press the MODE button until KER is displayed in the top of the display
Have the examinee sit and place his/her chin and forehead against the chin rest
and forehead rest.
B.
Adjust the height of the chin by rotating the height adjustment knob so that the eye
of the examinee is aligned with the height adjustment mark on the face rest.
Slide the operation joystick to the left so that the right eye of the examinee is
displayed on the monitor.
B.
Ask the examinee to look at the red balloon in the center of the scene (eye fixation
target).
C.
Looking at the monitor, check that image of the KERATO RING is not obscured by
the upper eyelid, if it is covering the ring, instruct the examinee to keep his/her eye
opened widely until measurement ends.
D.
Looking at the monitor, incline the operation joystick to right or left, and turn the
operation joystick so the pupil is concentric with the inner alignment mark and
focus on the KERATO RING, at the time adjusted the bar would be appeared
above the inner alignment mark.
(4) Measurement
A.
B.
C.
According to the 6.1 SETUP Mode you can select and change VD, CYL, KERATO
points.
B.
C.
CAUTION: before starting the new measurement you have to press CLEAR button
to remove previous data.
Slide the stage to the right side and measure the left eye.
B.
After both eyes are measured, pupil distance (PD) will be displayed on the monitor.
(7) Print
A.
Press the MODE button until CLBC MODE is displayed in the top of the display
Put water in the concave section of contact lens holder at backside of the TEST
Model Eye and place the contact lens on the holder with the concave surface
facing upward. Take care that the contact lens not to drop.
(3)
Remove the chin rest paper. Fix the Model Eye attached with contact lens using
fixation pin.
(4)
Perform Focusing
A.
Looking at the monitor, incline the operation joystick to right or left, and turn the
operation joystick so the pupil is centered with the inner alignment red mark.
B.
Moving operation joystick straight and back fix up KERATO RING to be clear at
the time adjusted the bar would be appeared above the inner alignment mark.
(5) Measurement
A.
B.
C.
According to the 6.1 SETUP Mode you can select and change VD, CYL, KERATO
points.
B.
C.
CAUTION: before starting the new measurement you have to press CLEAR button
to remove previous data.
(7) Print
A.
Mode
Press MODE button until SIZE MODE is displayed in the top of the display.
Have the examinee sit and place his/her chin and forehead against the chin
rest and forehead rest.
B.
Adjust the height of the chin by rotating the height adjustment knob so that
the eye of the examinee is aligned with the height adjustment mark on the
headrest.
Slide the operation joystick to the left so that the right eye of the examinee is
displayed on the monitor.
B.
Ask the examinee to look at the red balloon in the center of the scene ( eye
fixation target ).
C.
Looking at the monitor, check that image of the KERATO RING is not
obscured by the upper eyelid, if it is covering the ring, instruct the examinee
to keep his/her eye opened widely until measurement ends.
D.
Looking at the monitor, incline the operation joystick to right or left, and turn
the operation joystick so the pupil is centered with the inner alignment mark
and focus on the KERATO RING.
(4) Measurement
A.
B.
C.
D.
According to the 6.1 SETUP Mode you can select and change VD, CYL,
KERATO points.
B.
C.
CAUTION: before starting the new measurement you have to press CLEAR
button to remove previous data.
Slide the stage to the right side and measure the left eye.
(7) Print
A.
B.
(1)
STEP
To change measurement data units.
Select 0.25/0.12 pressing button /
(2)
VD
To select Vertex Distance (VD)
Select 0/10/12/13.5/15 pressing button /
(3)
CYL
To select Cylinder value and form
Select -/+/+- pressing button /
(4)
START
To select Manual or Auto measurement
Select Manual/Auto pressing button /.
(5)
REPEAT
To select repeated measurements.
Select 1/3/5/7 pressing button /.
(6)
TARGET
Target LED brightness
MONITOR
Monitor brightness control
(1 is the brightest)
(8)
PRINT
Change printout forms.
select ALL/ECONO/OFF pressing button /.
(9)
PRINT NO.
Printout Numbering.
Select ON/OFF/RESET pressing button /.
(10) DATA
Press PRINT button after DATA ON to display the latest measurements of KER mode
or REF mode.
select ON/OFF pressing button /.
(11) SAVE
Anytime you dont operate the instrument it will turn to the power saving mode.
select OFF/3/5/10 pressing button /.
(12) MESSAGE
Enter a message to be printed out with the results of measurement using the internal
printer in this window..
To display the users message select the one and press EDIT button.
Using buttons // select message and press SET button. You can edit the
message using ,, DEL, BAK buttons . then press SET button again. Select END
and SET button to complete message.
(13) DATE FORM
To setup order of displaying the date.
select YMD/DMY/MDY pressing button /.
(14) 24H MODE
to change 12 hour/24 hour time
change 12H/24H pressing button /.
(15) DATA
To setup year/month/day
Select year/month/d ay pressing button /and use +/- button.
(16) TIME
To setup hour/minute/second
Select
(17) CODE
To initialize program on the occasion of occurring errors among used data.
After pressing EDIT 5 Reload the factory data? will appear. If you answer yes
program can be initialized.