Académique Documents
Professionnel Documents
Culture Documents
Manual
8.109.8042EN
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
info@metrohm.com
www.metrohm.com
Manual
8.109.8042EN
10.2013 ebe
Teachware
Metrohm AG
CH-9100 Herisau
teachware@metrohm.com
Table of contents
Table of contents
1 Introduction
1.1
General .................................................................................. 1
1.2
Design of ion-selective electrodes (ISE) .............................. 1
1.2.1
Electrodes without exchangeable tip ........................................ 1
1.2.2
Electrodes with separate electrode tip ...................................... 2
1.3
2 Measuring
2.1
Preparing the electrodes ...................................................... 4
2.1.1
Crystal membrane electrodes ................................................... 4
2.1.2
Polymer membrane electrodes ................................................. 5
2.1.3
Glass membrane electrodes ..................................................... 5
2.2
Titration ................................................................................. 5
2.3
2.4
2.5
2.6
3 Storing electrodes
4 Troubleshooting
9
10
4.1
4.2
4.3
4.4
4.5
5 Technical specifications
13
5.1
5.2
Glossary
15
Index
18
III
1 Introduction
1 Introduction
1.1
General
Immediately after receiving the electrode, check to make sure that it
works properly. Electrodes that do not work properly must be sent back
for warranty processing within two months (starting from the day of delivery). If the defect is proven to be due to a material or manufacturing
defect, the electrode will be replaced at no charge. The transport costs are
to the customer's account.
1.2
1.2.1
Protective cap
Screw off and connect electrode cable.
Sensor surface
Do not touch with your fingers.
Filler opening
For filling in electrolyte solution.
1.2.2
Protective cap
Screw off and connect electrode cable.
Sensor surface
Do not touch with your fingers.
1.3
Reference electrodes
The reference system is integrated into the electrode body for measuring
with combined ion-selective electrodes; for ion-selective electrodes without integrated reference system, you will need a separate reference electrode. We recommend the 6.0750.100 LL ISE Reference as external reference electrode. Owing to its fixed ground-joint diaphragm, this electrode
is not sensitive to contamination and offers low-noise measuring signals.
The LL ISE Reference contains KCl 3 mol/L as reference and bridge electrolyte.
If K+ or Cl are interfering (e.g. because it is the measuring ion or because
of cross-sensitivity), the LL ISE Reference has to be filled with a suitable
bridge electrolyte, e.g. KNO3, NaCl or NH4Cl. Please keep in mind that,
1 Introduction
after an exchange of electrolyte, the sensor requires some time until the
system is stable again.
Filler opening
For filling in the bridge electrolyte, e.g. KCl
3 mol/L. Refill weekly and replace from time
to time.
Ground-joint diaphragm
Reference system
Ag/AgCl with KCl gel.
2 Measuring
CAUTION
Do not use the ultrasonic bath for electrodes, as they may be damaged
by such a treatment.
CAUTION
2.1
2.1.1
2.1.2
2 Measuring
2.1.3
2.2
Titration
Ion-selective electrodes are well-suited for potentiometric titrations. The
resulting titration curves are usually S-shaped and can be evaluated well in
automatic titration systems.
For application advice on working with ion-selective electrodes, please go
to www.metrohm.com.
2.3
2.4
2 Measuring
the same conditions apply as for standard addition. However, this method
is rarely used.
2.5
2.6
ISA/TISAB solutions
The following table gives examples for possible ISA/TISAB solutions:
Measuring
ion
ISA/TISAB
Ag+
KNO3 1 mol/L
10.11 g
Br
KNO3 1 mol/L
10.11 g
Remarks
or
NaNO3 2 mol/L
17.00 g
Ca2+
KCl 1 mol/L
7.46 g
Cd2+
KNO3 1 mol/L
10.11 g
Cl
KNO3 1 mol/L
10.11 g
or
KNO3
5.06 g
Ammonium acetate
3.85 g
2.8 mL
or
NaNO3 2 mol/L
17.00 g
CN
0.40 g
Cu2+
KNO3 1 mol/L
10.11 g
NaCl
5.84 g
5.75 mL
0.45 g
S2 interferes.
AB 82
Measuring
ion
ISA/TISAB
KNO3 1 mol/L
10.11 g
Remarks
or
NaNO3 2 mol/L
17.00 g
K+
0.584 - 5.844 g
Na+ (G)
12.11 g
or
Triethanolamine
Na+ (P)
CaCl2 1 mol/L
Set to between pH
8 and pH 10 using
HNO3 (AB 83).
7.50 mL
14.70 g
AB 83
CaCl22 H2O
NO3
[(NH4)2SO4] 1 mol/L
13.21 g
or
[Al2(SO4)3] 0.1 mol/L
3.42 g
Pb2+
NaClO4H2O 1 mol/L
14.05 g
S2
NaOH 2 mol/L
8.00 g
pH 13
SCN
KNO3 1 mol/L
10.11 g
or
Acetate buffer pH = 6,
CH3COOtot 1 mol/L
Abbreviations used:
AB: Metrohm Application Bulletin
Na+ (G): Glass membrane electrode
Na+ (P): Polymer membrane electrode
3 Storing electrodes
3 Storing electrodes
Electrode
Type
Storage
For short periods
Ag+
dry
Br
Ca2+
Cd2+
Cl
CN
Cu2+
KI 0.1 mol/L
K+
Na+ (G)
Na+ (P)
dry
NO3
Pb2+
dry
S2
dry
SCN
dry
*) A small amount of residual moisture on the electrode helps keep the reference electrode
operational.
Abbreviations used:
G: Glass membrane electrode
C: Crystal membrane electrode
P: Polymer membrane electrode
4 Troubleshooting
The following malfunctions could occur:
These problems may arise with the reference electrode or the ion-selective
electrode.
4.1
4.2
10
4.3
4 Troubleshooting
4.4
Application-related problems
4.5
Interfering ions
The concentrations in mol/L of the interfering ions, which generate an
analysis error of approximately 10%, are specified in the following table.
Measuring ion
Interferences
Ag+
Br
Ca2+
c(Na+) < 0.26; c(Pb2+) < 5.3106; c(Fe2+) < 6.7105; c(Zn2+) < 2.6106;
c(Cu2+) < 6.7105; c(Mg2+) < 8.3103
Cd2+
Cl
11
Measuring ion
Interferences
CN
K+
c(Na+) < 5101; c(TRIS+) < 0.1; c(Li+) < 1; c(NH4+) < 3.3103; c(Cs+) < 104;
c(H+) < 102
Na+ (G)
pH > 4; c(Li+) < 5102; c(K+) < 0.1; c(Rb+) < 3; c(NH4+) < 3; c(Ag+) < 3107
Na+ (P)
NO3
c(Br) < 7104; c(NO2) < 7104; c(SO42) < 1; c(Cl) < 2.5102; c(F) < 1;
c(CH3COO) < 0.1
Pb2+
S2
SCN
c(OH) < 101; c(Cl) < 2102; c(Br) < 3106; c(I) < 109;
c(S2) < 109; c(CN) < 7106; c(S2O32) < 1.3
Abbreviations used:
Na+ (G): Glass membrane electrode
Na+ (P): Polymer membrane electrode
12
5 Technical specifications
5 Technical specifications
5.1
Measured data
Electrode
Type
Measuring
range [mol/L]
Reproducibility
pH range
Temperature
range [C]
Membrane
resistance
[M]
Ag+
1107 - 1
2%
2-8
0 - 80
Br
5106 - 1
2%
0 - 14
0 - 50
0.1
Ca2+
5107 - 1
4%
2 - 12
0 - 40
1-4
Cd2+
1107 - 101
4%
2 - 12 (recommended pH
7)
0 - 80
Cl
5105 - 1
2%
0 - 14
0 - 50
0.1
CN
8106 - 102
2%
10 - 14
(HCN is
present at
pH < 10)
0 - 80
30
Cu2+
1108 - 101
4%
2 - 12 (recommended pH
6)
0 - 80
1106 - sat.
2%
5-7
0 - 80
0.15 - 0.2
5108 - 1
2%
0 - 14
0 - 50
0.1
K+
1106 - 1
2%
2.5 - 11
0 - 40
10 - 20
Na+
1105 - 1
2%
5-9
0 - 80
300
Na+
1106 - 1
2%
3 - 12
0 - 40
< 50
NO3
7106 - 1
2%
2.5 - 11
0 - 40
1-5
Pb2+
1106 - 101
4%
4-7
0 - 80
S2
1107 - 1
2%
2 - 12
0 - 80
SCN
5106 - 1
2%
2 - 10 (recommended pH
4)
0 - 50
0.1
13
Abbreviations used:
G: Glass membrane electrode
C: Crystal membrane electrode
P: Polymer membrane electrode
5.2
14
Glossary
Glossary
Activity coefficient
The activity coefficient yi is a measure for the nonideal behavior of a solution. It can be calculated approximately and is valid up to about
c < 0.1 mol/L. The activity coefficient depends on the solvent, the temperature, the ionic strength and the effective size of the hydrated ion.
Ag/AgCl reference system
The reference system consists of silver, silver chloride and a KCl solution,
usually KCl 3 mol/L.
Calibration curve
The calibration curve shows the measured potential U as a function of the
logarithm of the ion activity. The following diagram shows the calibration
curve of a fluoride ISE (note the negative slope of the calibration curve,
which is due to the ion charge z of F = 1!):
U / mV
100 mV
log ai
-7
Figure 1
-6
-5
-4
-3
-2
-1
For low ion concentrations, the curve is nonlinear, which means that in
this range the electrode responds not only to the measuring ion but also
to interfering ions (e.g. to OH for the fluoride ISE).
Cross-sensitivity
See interfering ions.
15
Double-junction electrode
Electrode which contains another electrolyte solution (so-called bridge
electrolyte solution) between the sample solution and the reference system. For an example of a double-junction electrode, see figure page 3.
Electrode slope
The potential U of ion-selective electrodes depends on the activity ai of
free ions in the solution. This relation is described by the Nernst equation
as follows:
U = U0 +
2.303RT
log ai = U 0 + U N log ai
z i F
U0:
R:
T:
Absolute temperature
zi:
F:
ai:
UN:
Nernst slope
The electrode slope depends on the charge of the ion (including its sign)
as well as on the temperature.
Electrolyte solution
Solution with electric conductivity.
Interfering ions
Ion-selective electrodes often respond to other ions besides the measuring
ion (cross-sensitivity). The lower the concentration of the measuring ion,
the greater the interference of such ions. In cases of low measuring ion
concentrations, the calibration curve is no longer linear.
Ion activity, ion concentration
Many of the laws that apply to solutions lose their validity with rising ion
concentrations ci. These equations are again applicable to the ion activity
if the activity coefficient yi is taken into account. Activity ai and concentration ci of the ion i in solutions containing only free ions are related as follows:
ai = yi ci
16
Glossary
Ion-selective electrodes respond only to ion activities. For low concentrations, yi 1 applies, which means that concentration and activity are
equal.
Ionic strength
A measure for interionic interactions in a solution. The ionic strength I
depends on the concentration ci and the charge zi of the ions:
1
I=
2
ci z i2
i
TISAB
This abbreviation stands for Total Ionic Strength Adjustment Buffer and
designates a buffer solution with high ionic strength that maintains the
ionic strength of a sample solution as well as its pH value at a constant
level. TISAB solutions are mainly used for measurements using calibration
curves.
17
Index
Index
A
Activity coefficient .................... 15
Ag/AgCl reference system ......... 15
C
Calibration curve ...................... 15
Cross-sensitivity .................. 11, 15
I
Interfering ions ................... 11, 16
Ion activity ................................ 16
Ion concentration ..................... 16
D
Design of the electrodes ............. 1
Direct measurement ................... 6
Double-junction electrode ........ 16
E
Electrode
Clean .................................... 7
Prepare ................................. 4
Store .................................... 9
18
M
Measured data ......................... 13
N
Nernst slope ............................. 16
P
Polymer membrane electrode
Lifetime .............................. 14
R
Reference electrode .................... 2
S
Standard addition ....................... 6
T
Technical specifications ............ 13
TISAB ................................... 7, 17
Titration ..................................... 5
Troubleshooting
Applications ....................... 11
Crystal membrane electrode
........................................... 10
Polymer membrane electrode
........................................... 11
Reference electrode ............ 10