Leakage Current
Leakage current flowing between line and ground in electrical equipment can present
a lethal shock hazard to users
The susceptibility of a patient to serious injury or death increases with patientconnected medical equipment
The low impedance path between the patient and equipment creates an
environment where the patient may complete the fault circuit
3
8/9/2013
8/9/2013
NOTES:
Hospital Grade power strips cannot be used in general or critical patient care areas
They have not been evaluated for use where Article 517 of the National Electrical
Code requires Hospital Grade components
8/9/2013
Medical grade strips are designed to be used within (and outside) patient care
areas, including the patient care vicinity
Hospital Grade strips are not designed to be used within general or critical
patient care areas, but can be used in other areas throughout the facility