Thank you for purchasing the Signature Series Model 210 Laser Chair.

Please carefully read these


instructions and place it in a convenient location for future reference.

PRECAUTIONS
1. Installation
(1) Never position the chair where it will be exposed to moisture, direct sunlight, dust, salty
air, chemical storage areas, excessive temperature or humidity.
(2) Install the chair in a stable place, free of uneven floors, vibration and shock.
2. Always check, prior to operation, that all the cords are properly connected.
3. Take the proper steps to stop the chair (leaving the patient in a safe condition) if anything is
found wrong with the chair.
4. Be sure to observe the following after use:
(1) Never hold and pull the connecting cord for disconnecting plugs. Excessive force may be
applied and cause damage to inside connections.
(2) Never attempt to modify the chair in anyway. Always contact your dealer or Topcon for
repairs and service. For replacing fuses, read this manual carefully.
(3)

Before use make sure that the chair operates properly and safely when it has not been
used for any long period.
Revision B, 7/01

CONTENTS
1.

NOMENCLATURE
1.1
Model 210 Components

1
1

2.

PERFORMANCE AND SPECIFICATIONS

3.

SIGNIFICANT COMPONENTS
3.1
Chair Frame and Upholstery
3.2
Electric Lift
3.3
Footswitch Control
3.4
Optional Items

2
2
2
2
2

4.

ASSEMBLY
4.1
Tools
4.2
Model 210 Chair / Stand Position
4.3
Electrical Connections
4.4
Alternate Connections

2
2
2
2
3

5.

OPERATION
5.1
Elevation Control
5.2
Seat Rotation

3
3
3

6.

MAINTENANCE
6.1
Paint and Nonmetallic Parts
6.2
Upholstery
6.3
Fuse Replacement

4
4
4
4

7.

TROUBLE SHOOTING

ii

1.

NOMENCLATURE
1.1

Signature Series Model 210 Components


Figure 1 shows the primary features and key components of the Model 210 Laser
Chair. Use the figure as a reference.

Chair Frame
Electric Lift
Footswitch

Figure 1: Model 210 Laser Chair Features

2.

PERFORMANCE & SPECIFICATIONS


2.1

2.2

2.3

2.4

Chair Dimension
Overall:
Seat Height (retracted):
Seat Height (full stroke):

33.5 (High) x 18 (Wide) x 20 (Deep)


19 (H)
27 (H)

Electrical
110 Volt System:
220 Volt System:

115 V, 60 Hz, 6 Amp


220 V, 50 Hz, 8 Amp

Weight
Chair Capacity:
Shipping Wt.:

350 lbs
96 lbs

Movement
Elevation Stroke:

3.

4.

SIGNIFICANT COMPONENTS
3.1

Fixed Frame with Adjustable Backrests


The contoured frame and the adjustable backrest allow the patient to sit comfortably.

3.2

Electric Lift
The lift provides a reliable, stable and quiet lifting capability.

3.3

Footswitch Control
The footswitch controls the elevation of the chair

3.4

Optional Items
The optional items that can be used with the Model 210 are as follows:
1)
4-way locking rotation
2)
360 degree rotation
3)
8-way locking rotation

ASSEMBLY
4.1
Tools
None required.
4.2

Model 210 Chair / Stand Position


1)
Place the chair next to the base of the ophthalmic stand.
2)
Install the rubber pads under the base floorplate to provide traction.
3)
Locate the footswitch the appropriate position.

4.3

Electrical Connections
1)
Connect the chair power cable into the appropriate wall outlet or into the
ophthalmic stand.
2)
Connect the chair control cable to the ophthalmic stand and check the
elevation movement of the Model 210 chair.
NOTE: Auto return of the chair is not available on this model.

4.3.1 Configure for IS-800 Stand


Remove the shroud by following the steps below.
1)
Raise the lift to the highest position.
2)
Pull both sides of the shroud to release from the Velcro.
3)
Lift the shroud up.
4)
Remove the jumper connector marked No safety located on the cable. Cut
the wire tie and install the jumper marked Safety. This allows the IS-800
Table safety switch to be operable.
5)
Reinstall the shroud.

4.4

Alternate Connections
When the Model 210 laser chair is not coordinated with a Topcon ophthalmic stand,
please make sure the electrical input is correctly arranged.
Stand Cable Wire Code
Pin
Number
1
2
3
4
5
6
7
8
9

5.

Wire Function

Wire Color

Ground
Neutral
Hot
Up
Down
NA
NA
Safety Switch
Safety Switch

Green
White
Black
Red
Violet
Orange
Brown
Yellow
Blue

OPERATION

5.1

Elevation Control
The elevation of the Model 210 laser chair is controlled by the footswitch. Press the
right side of the switch to make the lift go up, and step on the left side of the switch
to lower the chair.

5.2

Seat Rotation (Optional)


Pull the rotation handle and rotate the chair to the desired position. Release the handle
at one of the predefined lock angles to secure the chair in place.

6.

MAINTENANCE

6.1

Paint and Nonmetallic Parts


If oil or dirt adheres to painted or nonmetallic parts of the chair, it can be easily
removed by using a cloth moistened with a diluted neutral detergent soap. Afterwards,
use caution when wiping the components of the ophthalmic chair by using a clean dry
cloth.

6.2

Upholstery
Use the same procedure as described above.

CAUTION:

6.3

Do not wipe painted, nonmetallic parts or upholstery with volatile


solvents such as benzene or paint thinner owing to the risk of damaging
these parts of the chair.

Fuse Replacement

WARNING: Do not attempt to lift the shroud without unplugging all power to the
chair.
There are no externally removable fuses for the Model 210. The entire power module
must be replaced. First the shroud must be removed by separating the plastic shroud
from the velcro on each side and then lifting the shroud up to the chair frame. Hook
the shroud on the top of the lift, and disconnect all cabling. Loosen the four screws
securing the power module and remove it. Replace it with the same unit, securing it in
reverse order the same way that it was removed.
7.

TROUBLE SHOOTING
If the chair does not function, check the following points before asking for assistance.
1. Check to see that all cords and/or cables are securely connected.
If problems persist, call your dealer or Topcon for assistance.

Topcon Medical Systems, Inc.


37 West Century Road
Paramus, NJ 07652
800-223-1130

www.topconmedical.com