RESEARCH ARTICLE
1
Department of Physics, University of Helsinki, Helsinki, Finland; 2HUS Medical Imaging Center, University of Helsinki and
Helsinki University Hospital, Helsinki, Finland; 3Oral Radiology, Department of Oral and Maxillofacial Diseases, University of
Helsinki, Helsinki, Finland; 4STUKRadiation and Nuclear Safety Authority, Helsinki, Finland
Objectives: Dental radiography may involve situations where the patient is known to be
pregnant or the pregnancy is noticed after the X-ray procedure. In such cases, the radiation
dose to the foetus, though low, needs to be estimated. Uniform and widely used guidance on
dental X-ray procedures during pregnancy are presently lacking, the usefulness of lead shields
is unclear and practices vary.
Methods: Upper estimates of radiation doses to the foetus and breasts of the pregnant
patient were estimated with an anthropomorphic female phantom in intraoral, panoramic,
cephalometric and CBCT dental modalities with and without lead shields.
Results: The upper estimates of foetal doses varied from 0.009 to 6.9 mGy, and doses at the
breast level varied from 0.602 to 75.4 mGy. With lead shields, the foetal doses varied from
0.005 to 2.1 mGy, and breast doses varied from 0.002 to 10.4 mGy.
Conclusions: The foetal dose levels without lead shielding were ,1% of the annual dose limit
of 1 mSv for a member of the public. Albeit the relative shielding effect, the exposure-induced
increase in the risk of breast cancer death for the pregnant patient (based on the breast dose
only) and the exposure-induced increase in the risk of childhood cancer death for the unborn
child are minimal, and therefore, need for foetal and breast lead shielding was considered
irrelevant. Most important is that pregnancy is never a reason to avoid or to postpone
a clinically justified dental radiographic examination.
Dentomaxillofacial Radiology (2016) 45, 20150095. doi: 10.1259/dmfr.20150095
Cite this article as: Kelaranta A, Ekholm M, Toroi P, Kortesniemi M. Radiation exposure to
foetus and breasts from dental X-ray examinations: effect of lead shields. Dentomaxillofac
Radiol 2016; 45: 20150095.
Keywords: dentistry; radiation protection; phantoms; imaging; radiography; panoramic;
tomography; CBCT
Introduction
The number of dental radiographs taken annually in
Finland is approximately 2.7 million. In 2014, the annual number of intraoral, panoramic, cephalometric
and CBCT examinations in Finland was 2.35 million,
300,000, 35,000 and 7500, respectively (T Helasvuo,
June 11, 2015, personal communication).1 In 2004, the
Correspondence to: Ms Anna Kelaranta. E-mail: anna.kelaranta@alumni.
helsinki.fi
This study was supported by the State Subsidy for University Hospitals in
Finland (AK).
Received 19 March 2015; revised 13 August 2015; accepted 26 August 2015
2 of 9
Table 1 Uterus and breast doses from studies with different dental modalities and head CT for comparison
Modality
Dental device
Bitewing (film, rect. coll.)
FM intraoral (film, rect. coll.)
FM intraoral (film, round cone coll.)
Intraoral (film)
Siemens Heliodent
Intraoral (film)
Siemens Heliodent
Intraoral (film)
Siemens Heliodent
Intraoral (digital)
Gendex
Panoramic (digital)
SCANORA 3D
Panoramic (digital)
ProMax 3D
Panoramic
Instrumentarium
Panoramic
PA cephalometric
Lateral cephalometric
CBCT
3D Accuitomo
CBCT
CB MercuRay
CBCT
3D Accuitomo
CBCT
CB MercuRay
CBCT (small FOV)
CBCT (medium FOV)
CBCT (large FOV)
CBCT
Alphard VEGA
CBCT
3DX multi-image micro CT
Head CT
Toshiba Xpress
2.66 (2.36)
2.4 (2.23)
31.839.0 (35.953.8) 2.327.4 (5.136.6)
19.875.6 (17.385.4) 7.184.1 (3.892.8)
7.97 (2.24)
1.77 (0.04)
0.56 (0.01)
0.62 (0.01)
0.010.03
0.16d
0.050.16
9.1032.25
1.46d
145.91d
8.153.7 (0.21.2)
7.661.6 (0.21.4)
7.562.6 (0.21.4)
0.1,e 0.20.4,f 1g,h
13,e 1621,f 34,g 27h
36i
320 (75)
Reference
Ludlow17,a
Ludlow17,a
Ludlow17,a
Lecomber and Faulkner18
Lecomber and Faulkner18
Buch et al19
Buch et al19
Rottke et al20
Rottke et al20
Buch et al19
Ludlow17,a
Ludlow17,a
Ludlow17,a
Okano et al21
Okano et al21
Okano et al21
Okano et al21
Ludlow17,a
Ludlow17,a
Ludlow17
Okano et al22
Okano et al22
Beaconsfield et al23
CBVI, cone-beam volumetric imaging; FM, full mouth; FOV, field of view; PA, posteroanterior; rect.coll., rectangular collimator.
Doses reduced with a lead shield, when available, appear in parenthesis
Values are presented with the accuracy of the original publications.
Dental device manufacturers are as follows: 3D Accuitomo (J Morita Mfg. Corp., Kyoto, Japan); 3DX multi-image micro CT (J Morita Mfg.
Corp.); Alphard VEGA (Asahi Roentgen Ind., Co., Ltd, Kyoto, Japan); CB MercuRay (Hitachi Medical Corp., Tokyo, Japan); Gendex (Gendex
Dental Systems, Hatfield, PA); Instrumentarium (Instrumentarium Dental, Tuusula, Finland); ProMax 3D (Planmeca Oy, Helsinki, Finland);
SCANORA 3D (Soredex, Tuusula, Finland); Siemens Heliodent (Siemens, Erlangen, Germany); Toshiba Xpress (Toshiba Electron Tubes &
Devices Co., Ltd, Tochigi, Japan).
a
Absorbed doses calculated from the effective doses given by Ludlow17 using percentages of effective dose contribution of indirectly irradiated
tissues. For the dental devices, see Ludlow.17
b
Bisecting angle technique.
c
Paralleling technique.
d
Implant mode (FOV 102 3 102 mm2).
e
Dental mode (FOV 51 3 51 mm2).
f
Special implant mode (FOV 102 3 60 mm2), implant mode (FOV 102 3 102 mm2).
g
Panoramic mode (FOV 154 3 154 mm2).
h
Cephalo mode (FOV 200 3 179 mm2).
i
FOV 40 3 30 mm2.
birpublications.org/dmfr
Dental modalities
3 of 9
birpublications.org/dmfr
4 of 9
Tube
voltage (kV)
Filtration
(mmAl)
Tube
current (mA)
Time
(s)
70
60
63
min. 2.5
min. 2.5
min. 2.5
6
7
6
0.2
0.1
0.2
66
66
min. 2.5
min. 2.5
8
10
15.8
6.4
FSD/
FID (cm)
FOV
(cm 3 cm)
DAP
(mGy cm2)
6 (diam.)
6 (diam.)
6 (diam.)
21
9
17
30
30
30
14 3 30
24 3 18
71
16
50
170
HVL
(mmAl)
2.78
2.37
2.50
DAP, dosearea product; diam., diameter; FID, focus-to-image distance; FOV, field of view; FSD, focus-to-skin distance; HVL, half value layer;
min., minimum; mmAL, millimetres of aluminium.
Figure 1 Measurement settings in three intraoral projections: upper occlusal (a), mandibular incisor (b) and right maxilla premolar (c) and in
panoramic (d), cephalometric (e) and CBCT (f) exposures. The dose measurements were performed inside the phantom at the slice 21 position (gap
by wooden spacers) and at the breast level outside the phantom. Both measurement positions were used for all modalities.
birpublications.org/dmfr
5 of 9
Tube
voltage (kV)
Filtration
(mmAl 1 mmCu)
90
90
90
90
min.
min.
min.
min.
2.5 1 0.5
2.5 1 0.5
2.5 1 0.5
2.5 1 0.5
Tube
current (mA)
Time
(s)
FOV
(cm 3 cm)
DAP
(mGy cm2)
FID
(cm)
CTDI
(mGy)
10
10
10
10
12
12
12
27
435
835
838
20 3 18
557
820
1093
2491
60
60
60
60
8.2
11.2
9.6
5.4
CTDI, CT dose index; DAP, dosearea product; FID, focus-to-image distance; min., minimum; mmAL, millimetres of aluminium; mmCu,
millimetres of copper.
Discussion
The aim of this study was to determine the upper estimate of radiation exposure to the foetus and the breasts
of a pregnant patient from different dental X-ray
examinations both with and without lead shielding.
The results will serve as the basis of guidance for good
practices in dental radiography during pregnancy. The
European guidelines on radiation protection in dental
radiology2 indicate that national practices may vary. It
is therefore essential to establish consistent practices to
improve patient safety and to put the need for radiation
shields into perspective with actual radiation doses in
dental radiography.
In this study, the upper estimates of foetal doses
varied from 0.009 to 6.9 mGy without lead shielding and
from 0.005 to 2.1 mGy with lead shielding. The foetal
doses with or without the lead shields were far below the
level associated with any practical radiation detriment
to the foetus. However, exposure situations can vary
considerably depending on the imaging techniques applied. The cephalometric projection, for example, was
directed laterally, which explains the minimal reduction
in the foetal dose with the frontal shielding, since the
dose accumulates almost completely as internal scatter.
In panoramic radiography and CBCT, the rotating scan
geometry causes more frontal exposure without the
shields, so these cases usually have relatively higher
shielding efficiency (Table 5). However, the focus of this
study was on pregnant patients, and cephalometric exami-
Figure 2 Lead shield for thyroid only (Shield 1, left); lead shield for thyroid, breasts and abdominopelvic region (Shield 2, middle); and lead apron
for breasts and upper abdomen (Shield 3, right).
birpublications.org/dmfr
6 of 9
Table 4 Intraoral foetal and breast doses with effect of the lead shield expressed as a relative dose reduction in percentages. Dose conversion
factors are calculated as the foetal or breast dose divided by the corresponding dosearea product value
Modality
Intraoral (mGy)
Upper occlusal
Dose reduction (%)
Conversion factor
(mGy Gy cm22)
Mandibular incisor
Dose reduction (%)
Conversion factor
(mGy Gy cm22)
Maxillary
premolar (right)
Dose reduction (%)
Conversion factor
(mGy Gy cm22)
Foetal dose
No
Lead thyroid
shield shield (Shield 1)
0.553
27.0
0.009
1.0
0.022
96
1.1
0.017
97
0.8
91.7
0.006
39
0.6
0.602
0.005
0.994
47
48.5
0.016
99
0.8
0.283
53
31.8
0.002
.99
0.3
0.512
0.004
22
30.5
99
0.2
0.659
57
0.3
39.3
nations are mostly performed for (non-pregnant) paediatric patients.31 The highest relative shielding occurs in
upper occlusal projection, as one would expect with the
downward frontal direction of the exposure geometry
(Figure 1a). It should be noted, however, that in clinical
practice, implementation of the shielding geometry and
coverage inevitably varies considerably owing to individual working methods and competence, patient morphology and other specific factors involved in physical
examinations. This also leads to considerable variance in
the effectiveness of the shields.
The doses at the breast level varied from 0.602 to 75.4 mGy
without lead shielding and from 0.002 to 10.4 mGy with
lead shielding. The breast dose reduction in the cephalometric projection was one of the highest, being at the same
level with the intraoral dose reductions with much more
covering shield. This can be explained by the fact that
breasts were more shielded by the frontal shield than the
estimated foetal position. For patients with frequent dental
examinations, especially in orthodontic radiography in
paediatric female patients, the accumulated dose to the
breasts can be a notable addition to the personal radiation
burden in relative terms.
The use of lead shields reduced the foetal dose by
3997% and reduced the breast dose by 22%99%.
However, the absolute foetal dose was negligible even
without shielding. According to the European Council
Directive (2013/59/Euratom),32 it is correct to enquire
whether the patient is pregnant or breastfeeding unless it
can be ruled out for obvious reasons or it is not relevant
for the radiological procedure. Furthermore, the directive also states that if pregnancy cannot be ruled out,
special attention shall be given to the justification and
optimization of the radiological procedure, in particular
if abdominal and pelvic regions are involved. The International Atomic Energy Agency Basic Safety Standards33
state that ascertaining the pregnancy status shall be ensured when significant foetal doses are expected. The oral
region is far from the abdominopelvic region, and the
Dentomaxillofac Radiol, 45, 20150095
1.882
67.7
0.012
0.7
Breast dose
No
Lead thyroid
shield shield (Shield 1)
birpublications.org/dmfr
Modality
Panoramic dose (mGy)
Dose reduction (%)
Conversion factor
(mGy Gy cm22)
Cephalometric
dose (mGy)
Dose reduction (%)
Conversion factor
(mGy Gy cm22)
CBCT dose (mGy)
Small FOV
Dose
reduction (%)
Conversion
factor
(mGy Gy cm22)
Medium FOV
Dose
reduction (%)
Conversion
factor
(mGy Gy cm22)
Large FOV
Dose
reduction (%)
Conversion
factor
(mGy Gy cm22)
Extralarge FOV
Dose
reduction (%)
Conversion
factor
(mGy Gy cm22)
FOV, field of view.
Foetal dose
No
Lead apron
shield (Shield 3)
0.04
0.11
61
1.5
0.6
Breast dose
No
Lead apron
shield (Shield 3)
0.61
3.57
83
50.3
8.6
0.69
0.71
44.4
0.08
4.33
3
43.1
270.8
98
5.2
2.64
0.80
70
43.20
5.74
87
4.7
1.4
77.6
10.3
3.75
1.10
71
61.9
9.09
85
4.6
1.3
75.5
11.1
4.52
1.28
72
75.26
10.36
86
4.1
1.2
68.9
9.5
6.93
2.11
69
75.38
8.24
89
2.8
0.8
30.3
3.3
7 of 9
8 of 9
Conclusions
Acknowledgments
The authors are grateful to sales manager Jari Outavaara of Planmeca for letting us use the dental devices
in Planmeca, and to product manager Erkki Hiltunen of
Planmeca for helping us to use them.
References
1. Helasvuo T, ed. Number of radiological examinations in Finland in
2011. STUK-B 161. Helsinki, Finland: STUK; 2013.
2. European Commission. Radiation protection no. 136. European
guidelines on radiation protection in dental radiology. The safe use
of radiographs in dental practice. Luxembourg: EC; 2004. Available from: ec.europa.eu/energy/sites/ener/files/documents/136.pdf
3. Brennan J. An introduction to digital radiography in dentistry.
J Orthod 2002; 29: 669. doi: 10.1093/ortho/29.1.66
4. Wenzel A, Mystad A. Work flow with digital intraoral radiography: a systematic review. Acta Odontol Scand 2010; 68: 10614.
doi: 10.3109/00016350903514426
5. Berkhout E, Sanderink G, van der Stelt P. Digital intra-oral radiography in dentistry. Diagnostic efficacy and dose considerations. Oral Radiol 2003; 19: 113. doi: 10.1007/BF02493286
6. Pauwels R, Beinsberger J, Collaert B, Theodorakou C, Rogers J,
Walker A, et al; SEDENTEXCT Project Consortium. Effective
dose range for dental cone beam computed tomography scanners.
Eur J Radiol 2012; 81: 26771. doi: 10.1016/j.ejrad.2010.11.028
7. Roberts JA, Drage NA, Davies J, Thomas DW. Effective dose
from cone beam CT examinations in dentistry. Br J Radiol 2009;
82: 3540. doi: 10.1259/bjr/31419627
8. Ludlow JB, Davies-Ludlow LE, Brooks SL. Dosimetry of two
extraoral direct digital imaging devices: NewTom cone beam CT
and Orthophos Plus DS panoramic unit. Dentomaxillofac Radiol
2003; 32: 22934. doi: 10.1259/dmfr/26310390
9. Ludlow JB, Davies-Ludlow LE, Brooks SL, Howerton WB.
Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT. Dentomaxillofac
Radiol 2006; 35: 21926. doi: 10.1259/dmfr/14340323
10. Li G. Patient radiation dose and protection from cone-beam
computed tomography. Imaging Sci Dent 2013; 43: 639. doi:
10.5624/isd.2013.43.2.63
11. American College of Radiology. ACR practice guideline for imaging pregnant or potentially pregnant adolescents and women with
ionizing radiation: resolution 26. Reston, VA: American College of
Radiology; 2008.
12. Weber J, Ewen K, Schubel F. Determining organ doses in the
uterus during dental x-ray examinations. [In German.] Dtsch
Zahnarztl Z 1989; 44: 3403.
13. Orsini S, Campoleoni M, Rozza M, Conti U, Landini A, Eulisse G, et al.
The doses absorbed by the patients and the exposure of the operators in
dental radiodiagnosis. [In Italian.] Radiol Med 1992; 83: 1015.
birpublications.org/dmfr
27.
28.
29.
30.
31.
32.
33.
low dose Cone Beam CT. Eur J Radiol 2005; 56: 41317. doi:
10.1016/j.ejrad.2005.05.011
Miller MC. The pregnant dental patient. Oral Health 1996;
86: 1725.
Sandley E, Emmerling H. Dental radiography. RDH 2011; 6973.
European Commission. Radiation protection no. 172. Cone beam CT
for dental and maxillofacial radiology. Evidence based guidelines.
Luxembourg: EC; 2012. ec.europa.eu/energy/nuclear_protection/
doc/publication/172.pdf
Pina PM, Douglass J. Practices and opinions of Connecticut
general dentists regarding dental treatment during pregnancy. Gen
Dent 2011; 59: e2531.
Esmaeili EP, Ekholm M, Haukka J, Waltimo-Siren J. Quality
assessment of orthodontic radiography in children. Eur J Orthodont 2015. Epub ahead of print.
Council Directive 2013/59/Euratom. Official J Eur Union 2014;
L 013. ec.europa.eu/energy/sites/ener/files/documents/CELEX32013L0059-EN-TXT.pdf
International Atomic Energy Agency. IAEA safety standards for
protecting people and the environment. Radiation protection and
safety of radiation sources: international basic safety standards.
General safety requirements part 3, 2014. Vienna, Austria: IAEA.
9 of 9