Académique Documents
Professionnel Documents
Culture Documents
V-2015
ARTICLE 10
ARTICLE 10
LEAK TESTING
T-1000
15
T-1010
INTRODUCTION
SCOPE
T-1020
T-1021.3 Procedure Qualification. When procedure qualification is specified by the referencing Code
Section, a change of a requirement in the applicable
Appendix Tables I-1021 through X-1021 identified as an
essential variable shall require requalification of the written procedure by demonstration. A change of a requirement identified as a nonessential variable does not
require requalification of the written procedure. All
changes of essential and nonessential elements from
those specified within the written procedure shall require
revision of, or an addendum to, the written procedure.
T-1022
Referencing Code
For the leak testing method(s) or technique(s) specified by the referencing Code, the referencing Code Section
shall then be consulted for the following:
(a) personnel qualification/certification
(b) technique(s)/calibration standards
(c) extent of examination
(d) acceptable test sensitivity or leakage rate
(e) report requirements
(f) retention of records
T-1030
EQUIPMENT
T-1031 Gages.
(a) Gage Range. When dial indicating and recording
pressure gage(s) are used in leak testing, they should preferably have the dial(s) graduated over a range of approximately double the intended maximum pressure,
but in no case shall the range be less than 11/2 nor more
than four times that pressure. These range limits do not
apply to dial indicating and recording vacuum gages.
Range requirements for other types of gages given in an
applicable Mandatory Appendix shall be as required by
that Appendix.
(b) Gage Location. When components are to be pressure/vacuum leak tested, the dial indicating gage(s) shall
be connected to the component or to the component from
a remote location, with the gage(s) readily visible to the
operator controlling the pressure/vacuum throughout
the duration of pressurizing, evacuating, testing, and
GENERAL