MANUAL

Large Capacity
Refrigerated Centrifuge

V
VS
S- 11 00 00 00

Before using this unit , check instruction


manual enough.

Vision Scientific Co., Ltd.


W W W . V I S I O N S C I . C O . K R

Tel. 82-32-672-7779 FAX:82-32-676-3977


- 1 -

Pre-Caution

Prohibition

VS-1000

Caution

Instruction

Apart 30Cm from


the wall

- 2 -

PRE-CAUTION

VS-1000

Caution in installation
1. Do not install in high temperature & humidity. Install on well ventilated place.
2. Install apart from wall by 30Cm. Please set on place which is plain & no vibration.
3. Do not install on flammable or explosive place.
4. Please clarify proper electronic source before using. (Voltage, Power)
5. Please use in place where the voltage variable rate is within 10 %.
6. Please use the ground outlet.

Caution in operation

1. Please keep the circumstantial conditions as temperature from 2 to 40, and


humidity lower than 80%. Unit may not operate correctly.

2. Do not put the plug with wet to prevent the electric shock.
3. Please keep the circumstantial temperature not to changed rapidly.
Humidity may caused on unit inside. And it can cause trouble on unit.
(If dew is foamed on chamber, clean with soft dry fabric.)
4. Please put the contents into ROTOR symmetrically with same weight of contents.
If not, unit is operated in UNBALANCE and causes the trouble.
5. If you use under room temperature, please operate the refrigerated before 30 minutes
from input contents.
If you put the contents into CHAMBER before getting the aimed temperature, contents
can be degenerated.
6. Once refrigerator is operated, do not touch the chamber in wet. Hand can be frozen or
stuck on chamber by cold.
7. Door is closed during rotation. Do not open by forced for safety.
After end of operation, DOOR can be opened but chamber is still rotating for a while.
Therefore, please put out the contents after rotor is fully stopped.
8. When move the unit or do not use for a long time, please separate the rotor from unit.
If do not used for a long time, rotor may cannot be separated form shaft.

Additional cautions.
1. Use mild detergent for cleaning.
2. Please use proper tubes or Blood Bag.
- Plastic tubes of Nalgene, Falcon, Corning, Nunc, Sastedit.
- It it is possible, don't use glass tubes. In case of using glass tubes, take Corex
tubes or Firex tubes and don't set speed higher than 2,500RPM.
3. In case of smoke, strange smell or Error occurs, cut off main power switch and inform
to contact our service center.
4. Do not alternate or repair by an unauthorized person.

-- 3 --

VS-1000

Contents

Pre Caution................................................
Composition

2-3

................................................

- Composition

......................................

- Installation .............................................

Instruction ................................................ 6
Specification .............................................. 7
- Rotor Kinds

- Relative Tube

Operation

..................................................

- Control Panel

..........................................

- Operation

9
9

................................... 10

- General use

11

- Program use

12

- Error

12

Message

- Reference

13

Service Manual.............................................

14

Warranty.................................................

30

-- 4 --

Composition

VS-1000

Composition
After unit is arrived, check about the composition.

Content

Main Body

1Set

Swing Rotor +

1Set

Shield Chamber included Lid


(LID Open)

Qty

(LID Close)

Bucket

1Set

Tube Rack
4

(One of tuberack when you order Tube Rack x 6

1Set

this unit)
5

Instruction Manual(EN)

1EA

Emergency Door Wrench

1EA

Installation
1. Install on place where is plane & no vibration. And apart from the wall by 30Cm.
(By using Balance device)
2. Check the outside and all contents
3. Connect the AC Power Cord and

on rear part and connect to the grounded outlet. (220v)

And then turn on the main power switch on rear part.

Please use proper voltage to prevent fire, electrical accident or damage on unit.
4. Put the Swing rotor on the chamber.

- When unit is delivery, swing rotor and Shield chamber is fixed.


5. Put the Buckets on Swing rotor, Put property tube rack on Buckets.
- After BUCKET is put on the swing rotor, check these buckets is moved smoothly .
** Use sample material with exact balance and same volume .
Also , put them on tube racks symmetrically. If not, big vibration and noise while
unit is running. Also, it can be caused for unit broken.
- Close LID of the shield chamber, press the door. Door is closed automatically.

-- 5 --

Instruction

VS-1000

Meaning of word
1. RPM : Rotation Per Minute. It means speed of centrifuge.
2. RCF : Relative Centrifugal Force.

3.

dt : Integral Sigma Square dt. Total centrifugal value.

Feature
1. VS-1000 is Large Capacity Refrigerated centrifuge by Inverter Motor Control.
Test tube or Blood bag can be put on this unit.
2. Material of inside Chamber is made of STAINLESS STEEL (SUS 304), Shield chamber and
Buckets are made of Aluminium. Exterior material is made of Steel(CR41) with powder
coating.
3. Deceleration and Acceleration mode can be controlled by 9 steps.
4. RPM and RCF can be exchanged by Microprocessor controller.
20 Memories can be available for easy test.
5. Self diagnostic function detects an error and shows on display.
6. Hermetically sealed compressor makes low noise.
7. Available check test progress by Touch Screen Pad easily.
8. Strong chamber and rotor ,Self-Diagnostic Function and Safety device for Unbalance
, Temperature and Speed.
9. Automatic Door is built for safety and convenience.

-- 6 --

Specification

VS-1000

Specification

Max. Speed

5,000rpm (Swing rotor)

Max. Force

7,406 g (Swing rotor)

Max. Capacity

6000ml, 1000ml 6Tubes(Swing rotor)

Standard Rotor

Swing rotor (VR-1000S) with Shield Chamber

Door

Automatic Door Lid System

Main Controller

Microprocessor

Display

Touch Screen Display

Program

20 Memories

Accel/Decel Rates

Accel : 9 Files , Decel : 9 Files

Timer

9 Hrs. 59 Min, 59Sec & Hold Long Run

Temp. Range

-20 ~ + 40

Dimensions

840 x 900 x 880(H)mm

Drive System

Inverter Motor

Safety Devices

Over/Under speed protector, Overload protector


Unbalance detector, Temp. protector

Power Supply

220V, 60Hz, 3.5KW,16A

-- 7 --

Operation

VS-1000

Rotor Kinds
Swing rotor and Tube Rack

Rotor No.

Max.5,000RPM, 7,406 x G,

Swing Rotor

Radius 265mm

Shield Chamber included


LID

Sealing Cap

Bucket

(Option)
Aluminium Alloy
Tube Rack

Relative Tube( Option)


No

Volume

10ml

15ml

30ml

50ml

85ml(100ml)

250ml

500ml

1000ml

P.No
3114- 0010
3115- 0010
3118- 0010
3119- 0010
3110- 0150
3117- 0150
3114- 0030
3115- 0030
3118- 0030
3119- 0030
3114- 0050
3115- 0050
3118- 0050
3119- 0050
3118- 0085
3120- 0250
3122- 0250
3120- 9500
3122- 0500
3120- 1000
3122- 1000

Description
Oak Ridge Tube,
Screw Top
Round Tube, Open Top
Round Tube, Open Top

Oak Ridge Tube, Screw Top

Oak Ridge Tube, Screw Top

Oak Ridge Tube,


Screw Top
Bottle, Screw Top
Bottle, Screw Top
Bottle, Screw Top

Dia x Height
Dia x mm
16.0 x 81.5
16.0 x 82.0
16.1 x 81.7
16.0 x 81.4
15.9 x 113.9
16.1 x 114.3
25.7 x 93.7
25.7 x 94.5
25.7 x 94.5
25.5 x 94.3
28.8 x 107.7
29.0 x 107.7
28.8 x 107.0
28.8 x 106.7
38.2 x 105.7
61.8
61.8
69.5
69.5
97.5
97.6

x
x
x
x
x
x

127.7
127.6
170.2
169.6
184.5
188.4

Material
FEP
PSF
PC
PPCO
PPCO
PC
FEP
PSF
PC
PPCO
FEP
PSF
PC
PPCO
PC

50,000

50,000

PPCO
PC
PPCO
PC
PPCO
PC

27,500
27,500
4,800
13,700
7,100
7,100

27,500
27,500
4,800
13,700
7,100
7,100

* The tubes can be changed for promotion and better quality without any notice

-- 8 --

Min RCF
4' C
22'C
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
20,000
50,000
50,000
50,000
50,000
50,000
50,000

Operation

VS-1000

CONTROL PANEL

No.

Name

Function
RPM Setting / Display for Present Speed

SPEED Display

(If you press


, RPM & RCF can be changed)
Setting Value is shown on bottom of Speed Button(Blue)
Present value is shown on right space.
TIME Setting

TIME Display

TEMPERATURE Display
ROTOR No. Display
Acceleration Display
Deceleration Display
Program Display

(If you press


, Time & dt can be changed
Setting value is shown on bottom of TIME Button(Blue)
Present value is shown on right space.
Temperature is shown
Rotor No. is shown
Acceleration profile is shown
Deceleration profile is shown
Program No. is shown

Key

Pad

to

set

the

Speed,Time,

Temperature,

rotor

Acceleration/Deceleration and etc..


* ENTER Button must be pressed after setting to save the
setting value.

Key Pad

Program button
START Button
STOP Button
Working Lamp
Error Lamp
Status of the unit

Button to set the SPEED


Button to set the TIME
Button to set the TEMPERATURE
Button to set the ROTOR No.
Button to set the Accel
Button to set the Decel
ENTER Button. Input value is saved after setting.
CLEAR Button. Input value can be cancelled.
Buttons to set the number
Button to select the program

START Button
STOP Button
If there is No problem, Green is shown
If there is some problem, Red is shown. Unit is stopped.
Error message or Setting status are shown.

-- 9 --

No.,

Operation

VS-1000

Operation
1. Turn on the Main switch on the rear side.
2. If there is no problem on the unit, current status is displayed on each display.

If any error message is displayed, please cut off the main power and contact
your agent or our A/S team directly.

3. Open the door and set the rotor correctly.


4. Make a balance of unit and rotor by using balance. (** Balance is option)

** Make a balance by using Balance on top, front and side of the unit.
** Install this unit on horizontal place. Make a balance by adjustable device on wheel.
4. Close the LID of shield chamber after Putting the test tube or blood bag on the
Bucket, Press the door a little. Automatic Door Lid System is working to close the
door automatically. Press the door more. Door is closed fully.
* Refer to below picture when tube or blood bag is put on the buckets.

Same Volume tube or blood bag must be put on the buckets or tube rack symmetrically
(** If there are no balance, some accident can be occurred while unit is running.)
EX) Swing Rotor

a. Put same volume tube or blood bag on bucket or tube


rack symmetrically.
b. If you want only 1 test tube, 6 tubes (Same Volume)
must put on each tube rack.
c. If many tube should be put, tube must be put
symmetrically
d. If you put blood bag on buckets, put same volume
blood bag on each buckets symmetrically

* If you want to put exact volume on the tube, you should use Pippet or precision balance.
-- 10 --

Operation

VS-1000

5. Check below order to set the value on parameter.


6. If you want to use without PROGRAM Mode......
( Go to No.7 if you want to use PROGRAM MODE.)
6-1. Check Page.9. Input the value by Page.9(Control Panel)
If you press -

(SPEED Button), Available setting the SPEED.


Setting SPEED is shown on bottom of SPEED BUTTON (BLUE)
After setting the value, press the ENTER Button. Value is saved.

If you press -

(TIME Button), Available setting the TIME.


Setting TIME is shown on bottom of TIME BUTTON(BLUE)
After setting the value, press the ENTER Button. Value is saved.

If you press -

(TEMP Button), available setting the TEMPERATURE


Setting TEMP is shown on bottom of TEMP BUTTON(BLUE)
After setting the value, press the ENTER Button. Value is saved.

If you press -

(Rotor Button), available setting the Rotor No.

After setting the value, press the ENTER Button. Value is saved.
If you press -

(Accel Button),available setting the ACCEL.

After setting the value, press the ENTER Button. Value is saved.
If you press -

(Decel Button),available setting the DECEL

After setting the value, press the ENTER Button. Value is saved.
If you press -

(Program Button), available setting PROGRAM.

After setting the value, press the ENTER Button. Value is saved.
** Check Page.12(No.7)
If you press -
If you press -

(Clear Button) while you set the value, input value is cancelled.
after setting the value, all value is saved.

-- 11 --

VS-1000

7. PROGRAM MODE. ( AVAILABLE 20 Program)


(PROGRAM Button).

7-1. Press

7-2. Press program number you want by using number button..


7-2. Go to No.6 on Page.11).
7-3. After setting the value, Press the ENTER button. Value is saved.
You can make various PROGRAM.
8. Press

(START button), unit is starting.

9. If working is finished, centrifuge is stopped automatically


Or , If you press

(STOP button) while unit is running, working is stopped immediately.

10. After unit is stopped fully, press

(STOP button) one more.

Door is opened automatically. Open the door fully by using hand. Open the LID of Shield
chamber without shake. Take out the TEST tube or blood bag.

If main power is failed suddenly or door is not opened by


STOP button, use emergency key in front of unit.

On emergency situation, take off the main power immediately.

Emergency
Key

After unit is stopped completely, check the front of unit like


photo. There is a hole. Use wrench we offered.
Door is opened manually .

** Do not use Emergency key while unit is running.


After unit is stopped completely, use this ,please.

11. After Rotor is stopped completely, press STOP button. Door is opened automatically.
12. After using Centrifuge, turn off the main power and Power circuit breaker.

Error Message (If there are some problem, Error message is shown on the display)
Error
sErr_Motor
sErr_StartThe
sErr_DoorDoor
sErr_RpmSetOver
sErr_RpmSetLow

DISPLAY
" Motor overheat "
" Motor does not rotate "
" Locking State is Invalid "
" Setting Rpm is above Rotor "
" Limit RPM is above "

Contents
Motor TEMP is raised over than 110'C
Motor is not started
Door is not closed.
Setting Max RPM wrong
Setting Min RPM wrong

sErr_TempLimit

" Temperature escaped a setting limit " TEMP on Chamber is over 20'C than setting TEMP.

sErr_SetAccel
sErr_SetDecel
sErr_TempSetLow
sErr_TempSetOver
sErr_Unbalance
sErr_OverRpmOver
sErr_Inverter

"
"
"
"
"
"
"

Invalid Accel number "


Invalid Decel number "
Setting Up Chamber Temperature "
Setting Up Chamber Temperature "
Unbalance signal occurrence "
Under speed Error "
Inverter Control Error "

Setting ACCEL wrong


Setting DECEL wrong
Setting TEMP less -20'C
Setting TEMP is over 60'C
Balance of Rotor or unit is not fit.
RPM is raised over 5% than setting RPM
There is no signal from the Invertor Control

* Do not check the unit freely when Error message. Contact to us immediately.
-- 12 --

Operation

VS-1000

Reference
1. In case of low temperature, please set the temp. value before 30 minutes to secure
cooling status. If the setting temp. is lower than ambient, refrigerator starts
automatically.
2. Once you select program number, you can review all setting value of that program.
3. To check TOTAL RPM, please turn on the centrifuge with pressing of Speed mode key.
4. In case of failure of power supply, door is not opened by STOP Key. In that case, you
can open the lid forcefully by Emergency hole.
You can find it in front of unit.

This guarantee is valid only if the unit has been used in accordance
with the instructions in this manual and for its intended purpose.
Vision Scientific Co., Ltd. shall not be responsible for consequential
damages or damages resulting from the abuse or misuse or this
equipment.
* * This information will be required in the event that service becomes necessary.
(You can get these information from bar- code on rear part.)
- Specification (Description, Model No. , Serial No.)
- Date of Purchase.

-- 13 --

SERVICE

MANUAL

VS- 1000

SERVICE MANUAL

VS-1000

V I S I O N
S C I E N T I F I C
C O . , L T D
823 Toprip-Dong Yuseong-Gu Daejeon-Si Korea
Office TEL.032-672-7779 FAX.032-676-3977
HQ TEL.042-714-7700 FAX.042-714-7720

-- 14 --

SERVICE

MANUAL

VS- 1000

Contents

General information..........................................

16

Safety Information

...................................

17

Composition

.....................................
.

18

3- 1 Name and Function

....................... 18

3- 2 Control Panel

....................... 19

3- 3 Dimension

........................ 20

3- 4 Specification

........................ 20

3- 5 Composition

..........................21

Installation

Operation

Disassembly/ Assembly

..............................

........................................

22

23

.......................................24

6- 1 Main

........................................24

6- 2 Block Diagram

....................................... 26

6- 3 Circuit Diagram

.......................................27

Error Message

.................

Cleaning and maintenance

..................

29

Warranty

..................

30

-- 15 --

28

SERVICE

MANUAL

VS- 1000

1. General Information
This service manual is designed to help implement the exact and prompt service for
Centrifuge (Model : VS- 1000) of VISION SCIENTIFIC CO.,LTD.
Therefore, a service representative should fully understand the manual and use it to
be the proper reference materials.
Please follow the manual because mechanical damages can happen and a user can
be hurt during installation, operation and A/ S, if you do not follow the manual.
The operation should be conducted by the man who can handle with the equipment
that has electric current. If the manual is lost, please ask the head office or an
agency.
If you don't understand the contents of the manual, please feel free to contact the
head office or an agency.
Please understand that there could be some differences in some of the contents,
pictures and images if the parts change for safety reason or quality and performance
improvement reason.

We are not responsible for troubles or injury resulting from disassembly without any
proper tools or maintenance facilities during working.
If you contact our marketing team about any improvement or illogicality during
working, we will improve after closely reviewing it.

VISION SCIENTIFIC CO., LTD.


823 Toprip- Dong Yuseong- Gu Daejeon- Si Korea
Office TEL.032- 672- 7779 FAX.032- 676- 3977
HQ TEL.042- 714- 7700 FAX.042- 714- 7720
www.visionsci.co.kr

-- 16 --

SERVICE

MANUAL

VS- 1000

2. Safety information
SIGN

DANGER

CAUTION

SAFETY
CLASSIFICATION
The serious injury or death
of a user or loss of
property could happen, if
you do not follow safety
precaution.

The serious injury of a


user or loss of property
could happen, if you do
not follow safety
precaution.

DESCRIPTION
Do not operate the unit in places to be directly exposed to
moisture or water.
Do not operate the unit in places with flames.
Do not touch the power cord with wet hands.
Do not use any damaged or loose plugs.
Do no operate the unit except people who are well informed of
the operating way.
Do not install the unit in places to be exposed to moisture or
water.
Do not unplug by pulling on the cord.
(There could be electric shock or fire)
Do not plug several electric products in to an electrical outlet.
(There could be electric shock or fire)
Keep the plug and electrical outlet clean.
(There could be electric shock or fire)
Connect the power cord to a ground outlet and ground the

unit.

It is essential for safety.

Maintain level during installation or operation.


Do not wear loose clothing or a scarf during working.
Do not bend the power cord or press it with a heavy stuff.
Operate only the unit with a rated current.
Use Only the prescribed fuse to avoid fire.

CAUTION

The serious injury of a


user or loss of property
could happen, if you do
not follow safety
precaution.

Do not install the unit in a place with moisture or dust.


Use only the recommended parts and accessories by Vision
Scientific CO., Ltd.
We are not responsible for the damage of the unit or any
accidents resulting from the use of the unrecommended parts
and accessories.
When using the hazardous materials (pathogenicity, toxicity,
radiation), a user certainly should understand the materials for
safety.
If the hazardous materials are contaminated inside or outside
the unit, take safety measures(ventilation, isolation) after
removing the materials.
Do not use combustible and explosive materials.
Before use, keep the unit dry.
Be careful of the corrosion by a high level acidity, base, salts.
Use the proper tools.

-- 17 --

SERVICE

MANUAL

VS- 1000

3. Elements of the Product


3- 1. Name and Function

No.

Name

Function

Door /Viewer

Door / Viewer to see inside chamber while unit is running.

Main Power S/W

Power Circuit Breaker for Main S/W

Motor

Motor is built in the unit.

Door Lock

Door Locker(Automatic Door Lid System)

Chamber

Space that rotor can be running .

Shield Chamber

Inside chamber to be controlled wind of the swing rotor.

Control Panel

Control panel to set or check the value

Refrigerator

Refrigerator is built in the unit for cooling

Wheel

Wheel with Locker for making balance or movement

10

Emergency Door S/W

Emergency key to open the door

-- 18 --

SERVICE

MANUAL

VS- 1000

Control Panel

No.

Name

SPEED Display

Function

RPM Setting / Display for Present Speed


(If you press
Setting Value

is

, RPM & RCF can be changed)


shown on bottom of Speed

Button(Blue)
Present value is shown on right space.
TIME Setting

TIME Display

TEMPERATURE Display
ROTOR No. Display
Acceleration Display
Deceleration Display
Program Display

(If you press


, Time & dt can be changed
Setting value is shown on bottom of TIME Button(Blue)
Present value is shown on right space.
Temperature is shown
Rotor No. is shown
Acceleration profile is shown
Deceleration profile is shown
Program No. is shown
Key Pad to set the Speed,Time, Temperature, rotor No.,
Acceleration/Deceleration and etc..
* ENTER Button must be pressed after setting to save the

Program button
START Button
STOP Button
Working Lamp
Error Lamp

setting value.
Button to set the SPEED
Button to set the TIME
Button to set the TEMPERATURE
Button to set the ROTOR No.
Button to set the Accel
Button to set the Decel
ENTER Button. Input value is saved after setting.
CLEAR Button. Input value can be cancelled.
Buttons to set the number
Button to select the program
START Button
STOP Button
If there is No problem, Green is shown
If there is some problem, Red is shown. Unit is stopped.

Status of the unit

Error message or Setting status are shown.

Key Pad

-- 19 --

SERVICE

MANUAL

VS- 1000

3- 3 Dimension

3- 4 Specification
Max. Speed

5,000rpm (Swing rotor)

Max. Force

7,406 g (Swing rotor)

Max. Capacity

6000ml, 1000ml 6Tubes(Swing rotor)

Standard Rotor

Swing rotor (VR-1000S) with Shield Chamber

Door

Automatic Door Lid System

Main Controller

Microprocessor

Display

Touch Screen Display

Program

20 Memories

Accel/Decel Rates

Accel : 9 Files , Decel : 9 Files

Timer

9 Hrs. 59 Min, 59Sec & Hold Long Run

Temp. Range

-20 ~ + 40

Dimensions

840 x 900 x 880(H)mm

Drive System

Inverter Motor

Safety Devices

Over/Under speed protector, Overload protector


Unbalance detector, Temp. protector

Power Supply

220V, 60Hz, 3.5KW,16A

-- 20 --

SERVICE

MANUAL

VS- 1000

3- 5 Composition
Rotor Kinds
Rotor No.
Swing

rotor

+
1

Shield Chamber included


LID
5,000RPM, 7,406 x G, RADIUS 265mm

Sealing Cap

Bucket

(Option)
Aluminium Alloy
Tube Rack

Relative Tube
No

Volume

10ml

15ml

30ml

50ml

85ml(100ml)

250ml

500ml

1000ml

P.No
3114- 0010
3115- 0010
3118- 0010
3119- 0010
3110- 0150
3117- 0150
3114- 0030
3115- 0030
3118- 0030
3119- 0030
3114- 0050
3115- 0050
3118- 0050
3119- 0050
3118- 0085
3120- 0250
3122- 0250
3120- 9500
3122- 0500
3120- 1000
3122- 1000

Description
Oak Ridge Tube,
Screw Top
Round Tube, Open Top
Round Tube, Open Top
Oak Ridge Tube, Screw
Top

Oak Ridge Tube, Screw


Top
Oak Ridge Tube,
Screw Top
Bottle, Screw Top
Bottle, Screw Top
Bottle, Screw Top

Dia x Height
Dia x mm
16.0 x 81.5
16.0 x 82.0
16.1 x 81.7
16.0 x 81.4
15.9 x 113.9
16.1 x 114.3
25.7 x 93.7
25.7 x 94.5
25.7 x 94.5
25.5 x 94.3
28.8 x 107.7
29.0 x 107.7
28.8 x 107.0
28.8 x 106.7
38.2 x 105.7
61.8
61.8
69.5
69.5
97.5
97.6

x
x
x
x
x
x

127.7
127.6
170.2
169.6
184.5
188.4

Material
FEP
PSF
PC
PPCO
PPCO
PC
FEP
PSF
PC
PPCO
FEP
PSF
PC
PPCO

Min RCF
4'C
22' C
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
20,000
50,000
50,000
50,000
50,000
50,000
50,000

PC

50,000

50,000

PPCO
PC
PPCO
PC
PPCO
PC

27,500
27,500
4,800
13,700
7,100
7,100

27,500
27,500
4,800
13,700
7,100
7,100

* The tubes can be changed for promotion and better quality without any notice

-- 21 --

SERVICE

MANUAL

VS- 1000

4. Installation
4- 1. Instruction.
1) When a centrifuge arrives, check closely the damages for packing.
2) When the packing do not have any damages, open and check the damages for the unit.
3) Check if the unit has all accessories.
4) Check standard power inquiry. Use the Power socket with Ground.
In case of bent plug, stop to use it. Contact to us.
If you use Power socket or plug for ground, Fire or electric accident can be occurred.
Voltage and hertz are mentioned on serial barcode of unit.
Before using the unit, check it in advance.
* If you find some problem or broken for unit, contact to us.
Warranty is attached on last page of this manual .

4- 2. Location
1) Put this unit on wide space enough. Unit should be put apart 30cm from the wall as
Standard IEC1010- 2- 020.
2) To turn the power immediately when error is occurred on unit, unit should be put around
Emergency S/ W for power.
3) Locate this unit on horizontal place.
4) While centrifuge is working, do not put anything on top of the centrifuge.
It can be blocked for good working and it can be gotten some damage.
5) Be careful Test sample do not be dropped on centrifuge.
6) All condition for mechanical, electric and location are fine, plug in . Then, check 4- 3.
(4- 3. Refer to Electric inquiry)

4- 3. Electric inquiry
1) Use a proper wire because the centrifuge requires special electrical requirements.

205 ~ 250Vac, 50Hz/ 60H, SINGLE PHASE, 16A


2) The large outfit during acceleration is the rated power to reduce acceleration time.
3) The wire is 30A wire(3C x 2.5) including a ground wire.
4) Power cord is connected to main switch board directly.
5) The centrifuge is equip with Power Breaker S/ W(50A) just in case of a
short circuit or over- current by the regional difference of electric power.
6) Check the voltage by a electrical measuring instrument and then check to meet the
requirement of the fifth clause.

-- 22 --

SERVICE

MANUAL

VS- 1000

7) If the measured voltage is 10% over 230V, the malfunction or errors of the unit could
happen.
8) Check if the ground wire of a plug is grounded or the ground connection part is
disconnected.

5. Operation
5- 1. Function Test Table
No.

Description

Check unit is installed on horizontal place by using balnce device.

Run the centrifuge on Minimum RPM. Is it run well?

Run the centrifuge on Maximum RPM. Is it run well?

Check Same volume tube are put on bucket symmetrically or not.

Unit can be started while door is opened or not.

While unit is running, door is opened or not.

Property value is set on the unit or not.

* * This is the Test method to check unit is running well or not.


If there are some problem, contact to us, please.

-- 23 --

Check Point

SERVICE

MANUAL

VS- 1000

6. Disassembly
6- 1 Main Body (Before disassembly, check with engineer enough)
You should unplug when you disassemble or change parts to prevent electrical accidents.

No.

Description

Pictures

Front lower Open


There are 3 screws around lower part of unit.
Lift up the lower panel like left photo after release them.
Front lower case would be opened.

Front Upper Open


After remove the lower case,
Check around middle part of unit like left photo.
Release 3 screws. Then, lift up the upper case.
Upper case would be opened.

2-1

Side Open
Release 3 screws around left or right side of unit.
If you push to downward, case is opened.
Emergency Switch
There is a hole for emergency to open the door.

If some emergency is occurred, turn off the main power


first of all.
Check unit is stopped completely.
Spin this hole by using offered wrench, Door is opened
manually.
** Do not use this switch while unit is running.
After unit is stopped completely, use this,please.
Inverter motor controller

**
Do not change the saving value on Inverter motor control
without our approval. If you change the factory initial
value, most program can be twisted.
-- 24 --

SERVICE

MANUAL

VS- 1000

No.

Description

Pictures
Front lower case
A: Condenser
B : Motor
C : Door Device
D : Refrigerator(2HP)
E : Inverter motor Controller

Motor
RPM Sensor is located in bottom of motor.
Unbalance Sensor S/W is located around motor.
6

** Do not disassembly motor without engineer approval .


If motor is put wrong, some problem or accident can be
occurred..

Main Control PCB


PCB is located on upper paner.
This PCB can be controlled for SPEED, TEMP, TIME.
(Refer to Power Circuit diagram.).

Disassembly of Swing rotor.


8

Swing rotor and Shield chamber is attached when we


delivery the unit. If you want to disassembly swing rotor
and Shield chamber, contact to us, please.

-- 25 --

SERVICE

MANUAL

VS- 1000

6- 2

Block Diagram

-- 26 --

SERVICE MANUAL

VS- 1000

6- 3 Circuit Diagram

-- 27 --

SERVICE

MANUAL

VS- 1000

7. Error Messages
If Error messages are shown while unit is running, contact to us immediately.
Do not disassembly it without our approval.
Error Message
Error

DISPLAY

Contents

sErr_Motor

" Motor overheat "

Motor TEMP is raised over than 110'C

sErr_StartThe

" Motor does not rotate "

Motor is not started

sErr_DoorDoor

" Locking State is Invalid "

Door is not closed.

sErr_RpmSetOver

" Setting Rpm is above Rotor "

Setting Max RPM wrong

sErr_RpmSetLow

" Limit RPM is above "

Setting Min RPM wrong

sErr_TempLimit

" Temperature escaped a setting limit " TEMP on Chamber is over 20'C than setting TEMP.

sErr_SetAccel

" Invalid Accel number "

Setting ACCEL wrong

sErr_SetDecel

" Invalid Decel number "

Setting DECEL wrong

sErr_TempSetLow

" Setting Up Chamber Temperature "

Setting TEMP less -20'C

sErr_TempSetOver

" Setting Up Chamber Temperature "

Setting TEMP is over 60'C

sErr_Unbalance

" Unbalance signal occurrence "

Balance of Rotor or unit is not fit.

sErr_OverRpmOver

" Under speed Error "

RPM is raised over 5% than setting RPM

sErr_Inverter

" Inverter Control Error "

There is no signal from the Invertor Control

Reference
1. In case of low temperature, please set the temp. value before 30 minutes to secure
cooling status. If the setting temp. is lower than ambient, refrigerator starts
automatically.
2. Once you select program number, you can review all setting value of that program.
3. To check TOTAL RPM, please turn on the centrifuge with pressing of Speed mode key.
4. In case of failure of power supply, door is not opened by STOP Key. In that case, you
can open the lid forcefully by Emergency hole.
You can find it in front of unit.

This guarantee is valid only if the unit has been used in accordance
with the instructions in this manual and for its intended purpose.

Vision Scientific Co., Ltd. shall not be responsible for consequential


damages or damages resulting from the abuse or misuse or this
equipment.
* * This information will be required in the event that service becomes necessary.
(You can get these information from bar- code on rear part.)
- Specification (Description, Model No. Serial No.)
- Date of Purchase.

-- 28 --

SERVICE

MANUAL

VS- 1000

8. Cleaning and Maintenance


8- 1. Cleaning
1) Unit

The outside of a unit should be cleaned by soft and dry cloth.


If the outside needs to be cleaned, keep the unit dry by wiping with dry cloth after
cleaning with soapy water.

Do not use caustic detergents such as alcohol, benzene, thinner.


Use a damp sponge with a mild liquid detergent.
Be careful of a scratch during cleaning.
Do not use steel wool. It could create corrosion.

-- 29 --

SERVICE

MANUAL

VS- 1000

WARRANTY & SERVICE

VS-1000 are

warranted by VISION SCIE NT IF IC CO., LT D for a period of one

year after the date of purchase.


The warranty applies only to defects due to the fault of product workmanship or materials,
and is not valid for the defects resulting from accidental damage, misuse or neglect and in
case of alternations or repair carried out by an unauthorized person.
Service under the warranty is provided only upon the presentation of a declaration form.
Service is available in all countries where the product is officially distributed by VISION
Scientific Co., Ltd. If you need further information or have any difficulties, please contact
your supplier or directly VISION SCIENTIFIC CO., LTD.
Address

823 Toprip-Dong Yuseong-Gu Daejeon-Si Korea


Office TEL.032-672-7779 FAX.032-676-3977
HQ TEL.042-714-7700 FAX.042-714-7720 E-mail:

-- 30 --