VS-1000
Caution
Instruction
- 2 -
PRE-CAUTION
VS-1000
Caution in installation
1. Do not install in high temperature & humidity. Install on well ventilated place.
2. Install apart from wall by 30Cm. Please set on place which is plain & no vibration.
3. Do not install on flammable or explosive place.
4. Please clarify proper electronic source before using. (Voltage, Power)
5. Please use in place where the voltage variable rate is within 10 %.
6. Please use the ground outlet.
Caution in operation
2. Do not put the plug with wet to prevent the electric shock.
3. Please keep the circumstantial temperature not to changed rapidly.
Humidity may caused on unit inside. And it can cause trouble on unit.
(If dew is foamed on chamber, clean with soft dry fabric.)
4. Please put the contents into ROTOR symmetrically with same weight of contents.
If not, unit is operated in UNBALANCE and causes the trouble.
5. If you use under room temperature, please operate the refrigerated before 30 minutes
from input contents.
If you put the contents into CHAMBER before getting the aimed temperature, contents
can be degenerated.
6. Once refrigerator is operated, do not touch the chamber in wet. Hand can be frozen or
stuck on chamber by cold.
7. Door is closed during rotation. Do not open by forced for safety.
After end of operation, DOOR can be opened but chamber is still rotating for a while.
Therefore, please put out the contents after rotor is fully stopped.
8. When move the unit or do not use for a long time, please separate the rotor from unit.
If do not used for a long time, rotor may cannot be separated form shaft.
Additional cautions.
1. Use mild detergent for cleaning.
2. Please use proper tubes or Blood Bag.
- Plastic tubes of Nalgene, Falcon, Corning, Nunc, Sastedit.
- It it is possible, don't use glass tubes. In case of using glass tubes, take Corex
tubes or Firex tubes and don't set speed higher than 2,500RPM.
3. In case of smoke, strange smell or Error occurs, cut off main power switch and inform
to contact our service center.
4. Do not alternate or repair by an unauthorized person.
-- 3 --
VS-1000
Contents
Pre Caution................................................
Composition
2-3
................................................
- Composition
......................................
- Installation .............................................
Instruction ................................................ 6
Specification .............................................. 7
- Rotor Kinds
- Relative Tube
Operation
..................................................
- Control Panel
..........................................
- Operation
9
9
................................... 10
- General use
11
- Program use
12
- Error
12
Message
- Reference
13
Service Manual.............................................
14
Warranty.................................................
30
-- 4 --
Composition
VS-1000
Composition
After unit is arrived, check about the composition.
Content
Main Body
1Set
Swing Rotor +
1Set
Qty
(LID Close)
Bucket
1Set
Tube Rack
4
1Set
this unit)
5
Instruction Manual(EN)
1EA
1EA
Installation
1. Install on place where is plane & no vibration. And apart from the wall by 30Cm.
(By using Balance device)
2. Check the outside and all contents
3. Connect the AC Power Cord and
Please use proper voltage to prevent fire, electrical accident or damage on unit.
4. Put the Swing rotor on the chamber.
-- 5 --
Instruction
VS-1000
Meaning of word
1. RPM : Rotation Per Minute. It means speed of centrifuge.
2. RCF : Relative Centrifugal Force.
3.
Feature
1. VS-1000 is Large Capacity Refrigerated centrifuge by Inverter Motor Control.
Test tube or Blood bag can be put on this unit.
2. Material of inside Chamber is made of STAINLESS STEEL (SUS 304), Shield chamber and
Buckets are made of Aluminium. Exterior material is made of Steel(CR41) with powder
coating.
3. Deceleration and Acceleration mode can be controlled by 9 steps.
4. RPM and RCF can be exchanged by Microprocessor controller.
20 Memories can be available for easy test.
5. Self diagnostic function detects an error and shows on display.
6. Hermetically sealed compressor makes low noise.
7. Available check test progress by Touch Screen Pad easily.
8. Strong chamber and rotor ,Self-Diagnostic Function and Safety device for Unbalance
, Temperature and Speed.
9. Automatic Door is built for safety and convenience.
-- 6 --
Specification
VS-1000
Specification
Max. Speed
Max. Force
Max. Capacity
Standard Rotor
Door
Main Controller
Microprocessor
Display
Program
20 Memories
Accel/Decel Rates
Timer
Temp. Range
-20 ~ + 40
Dimensions
Drive System
Inverter Motor
Safety Devices
Power Supply
-- 7 --
Operation
VS-1000
Rotor Kinds
Swing rotor and Tube Rack
Rotor No.
Max.5,000RPM, 7,406 x G,
Swing Rotor
Radius 265mm
Sealing Cap
Bucket
(Option)
Aluminium Alloy
Tube Rack
Volume
10ml
15ml
30ml
50ml
85ml(100ml)
250ml
500ml
1000ml
P.No
3114- 0010
3115- 0010
3118- 0010
3119- 0010
3110- 0150
3117- 0150
3114- 0030
3115- 0030
3118- 0030
3119- 0030
3114- 0050
3115- 0050
3118- 0050
3119- 0050
3118- 0085
3120- 0250
3122- 0250
3120- 9500
3122- 0500
3120- 1000
3122- 1000
Description
Oak Ridge Tube,
Screw Top
Round Tube, Open Top
Round Tube, Open Top
Dia x Height
Dia x mm
16.0 x 81.5
16.0 x 82.0
16.1 x 81.7
16.0 x 81.4
15.9 x 113.9
16.1 x 114.3
25.7 x 93.7
25.7 x 94.5
25.7 x 94.5
25.5 x 94.3
28.8 x 107.7
29.0 x 107.7
28.8 x 107.0
28.8 x 106.7
38.2 x 105.7
61.8
61.8
69.5
69.5
97.5
97.6
x
x
x
x
x
x
127.7
127.6
170.2
169.6
184.5
188.4
Material
FEP
PSF
PC
PPCO
PPCO
PC
FEP
PSF
PC
PPCO
FEP
PSF
PC
PPCO
PC
50,000
50,000
PPCO
PC
PPCO
PC
PPCO
PC
27,500
27,500
4,800
13,700
7,100
7,100
27,500
27,500
4,800
13,700
7,100
7,100
* The tubes can be changed for promotion and better quality without any notice
-- 8 --
Min RCF
4' C
22'C
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
20,000
50,000
50,000
50,000
50,000
50,000
50,000
Operation
VS-1000
CONTROL PANEL
No.
Name
Function
RPM Setting / Display for Present Speed
SPEED Display
TIME Display
TEMPERATURE Display
ROTOR No. Display
Acceleration Display
Deceleration Display
Program Display
Key
Pad
to
set
the
Speed,Time,
Temperature,
rotor
Key Pad
Program button
START Button
STOP Button
Working Lamp
Error Lamp
Status of the unit
START Button
STOP Button
If there is No problem, Green is shown
If there is some problem, Red is shown. Unit is stopped.
Error message or Setting status are shown.
-- 9 --
No.,
Operation
VS-1000
Operation
1. Turn on the Main switch on the rear side.
2. If there is no problem on the unit, current status is displayed on each display.
If any error message is displayed, please cut off the main power and contact
your agent or our A/S team directly.
** Make a balance by using Balance on top, front and side of the unit.
** Install this unit on horizontal place. Make a balance by adjustable device on wheel.
4. Close the LID of shield chamber after Putting the test tube or blood bag on the
Bucket, Press the door a little. Automatic Door Lid System is working to close the
door automatically. Press the door more. Door is closed fully.
* Refer to below picture when tube or blood bag is put on the buckets.
Same Volume tube or blood bag must be put on the buckets or tube rack symmetrically
(** If there are no balance, some accident can be occurred while unit is running.)
EX) Swing Rotor
* If you want to put exact volume on the tube, you should use Pippet or precision balance.
-- 10 --
Operation
VS-1000
If you press -
If you press -
If you press -
After setting the value, press the ENTER Button. Value is saved.
If you press -
After setting the value, press the ENTER Button. Value is saved.
If you press -
After setting the value, press the ENTER Button. Value is saved.
If you press -
After setting the value, press the ENTER Button. Value is saved.
** Check Page.12(No.7)
If you press -
If you press -
(Clear Button) while you set the value, input value is cancelled.
after setting the value, all value is saved.
-- 11 --
VS-1000
7-1. Press
Door is opened automatically. Open the door fully by using hand. Open the LID of Shield
chamber without shake. Take out the TEST tube or blood bag.
Emergency
Key
11. After Rotor is stopped completely, press STOP button. Door is opened automatically.
12. After using Centrifuge, turn off the main power and Power circuit breaker.
Error Message (If there are some problem, Error message is shown on the display)
Error
sErr_Motor
sErr_StartThe
sErr_DoorDoor
sErr_RpmSetOver
sErr_RpmSetLow
DISPLAY
" Motor overheat "
" Motor does not rotate "
" Locking State is Invalid "
" Setting Rpm is above Rotor "
" Limit RPM is above "
Contents
Motor TEMP is raised over than 110'C
Motor is not started
Door is not closed.
Setting Max RPM wrong
Setting Min RPM wrong
sErr_TempLimit
" Temperature escaped a setting limit " TEMP on Chamber is over 20'C than setting TEMP.
sErr_SetAccel
sErr_SetDecel
sErr_TempSetLow
sErr_TempSetOver
sErr_Unbalance
sErr_OverRpmOver
sErr_Inverter
"
"
"
"
"
"
"
* Do not check the unit freely when Error message. Contact to us immediately.
-- 12 --
Operation
VS-1000
Reference
1. In case of low temperature, please set the temp. value before 30 minutes to secure
cooling status. If the setting temp. is lower than ambient, refrigerator starts
automatically.
2. Once you select program number, you can review all setting value of that program.
3. To check TOTAL RPM, please turn on the centrifuge with pressing of Speed mode key.
4. In case of failure of power supply, door is not opened by STOP Key. In that case, you
can open the lid forcefully by Emergency hole.
You can find it in front of unit.
This guarantee is valid only if the unit has been used in accordance
with the instructions in this manual and for its intended purpose.
Vision Scientific Co., Ltd. shall not be responsible for consequential
damages or damages resulting from the abuse or misuse or this
equipment.
* * This information will be required in the event that service becomes necessary.
(You can get these information from bar- code on rear part.)
- Specification (Description, Model No. , Serial No.)
- Date of Purchase.
-- 13 --
SERVICE
MANUAL
VS- 1000
SERVICE MANUAL
VS-1000
V I S I O N
S C I E N T I F I C
C O . , L T D
823 Toprip-Dong Yuseong-Gu Daejeon-Si Korea
Office TEL.032-672-7779 FAX.032-676-3977
HQ TEL.042-714-7700 FAX.042-714-7720
-- 14 --
SERVICE
MANUAL
VS- 1000
Contents
General information..........................................
16
Safety Information
...................................
17
Composition
.....................................
.
18
....................... 18
3- 2 Control Panel
....................... 19
3- 3 Dimension
........................ 20
3- 4 Specification
........................ 20
3- 5 Composition
..........................21
Installation
Operation
Disassembly/ Assembly
..............................
........................................
22
23
.......................................24
6- 1 Main
........................................24
6- 2 Block Diagram
....................................... 26
6- 3 Circuit Diagram
.......................................27
Error Message
.................
..................
29
Warranty
..................
30
-- 15 --
28
SERVICE
MANUAL
VS- 1000
1. General Information
This service manual is designed to help implement the exact and prompt service for
Centrifuge (Model : VS- 1000) of VISION SCIENTIFIC CO.,LTD.
Therefore, a service representative should fully understand the manual and use it to
be the proper reference materials.
Please follow the manual because mechanical damages can happen and a user can
be hurt during installation, operation and A/ S, if you do not follow the manual.
The operation should be conducted by the man who can handle with the equipment
that has electric current. If the manual is lost, please ask the head office or an
agency.
If you don't understand the contents of the manual, please feel free to contact the
head office or an agency.
Please understand that there could be some differences in some of the contents,
pictures and images if the parts change for safety reason or quality and performance
improvement reason.
We are not responsible for troubles or injury resulting from disassembly without any
proper tools or maintenance facilities during working.
If you contact our marketing team about any improvement or illogicality during
working, we will improve after closely reviewing it.
-- 16 --
SERVICE
MANUAL
VS- 1000
2. Safety information
SIGN
DANGER
CAUTION
SAFETY
CLASSIFICATION
The serious injury or death
of a user or loss of
property could happen, if
you do not follow safety
precaution.
DESCRIPTION
Do not operate the unit in places to be directly exposed to
moisture or water.
Do not operate the unit in places with flames.
Do not touch the power cord with wet hands.
Do not use any damaged or loose plugs.
Do no operate the unit except people who are well informed of
the operating way.
Do not install the unit in places to be exposed to moisture or
water.
Do not unplug by pulling on the cord.
(There could be electric shock or fire)
Do not plug several electric products in to an electrical outlet.
(There could be electric shock or fire)
Keep the plug and electrical outlet clean.
(There could be electric shock or fire)
Connect the power cord to a ground outlet and ground the
unit.
CAUTION
-- 17 --
SERVICE
MANUAL
VS- 1000
No.
Name
Function
Door /Viewer
Motor
Door Lock
Chamber
Shield Chamber
Control Panel
Refrigerator
Wheel
10
-- 18 --
SERVICE
MANUAL
VS- 1000
Control Panel
No.
Name
SPEED Display
Function
is
Button(Blue)
Present value is shown on right space.
TIME Setting
TIME Display
TEMPERATURE Display
ROTOR No. Display
Acceleration Display
Deceleration Display
Program Display
Program button
START Button
STOP Button
Working Lamp
Error Lamp
setting value.
Button to set the SPEED
Button to set the TIME
Button to set the TEMPERATURE
Button to set the ROTOR No.
Button to set the Accel
Button to set the Decel
ENTER Button. Input value is saved after setting.
CLEAR Button. Input value can be cancelled.
Buttons to set the number
Button to select the program
START Button
STOP Button
If there is No problem, Green is shown
If there is some problem, Red is shown. Unit is stopped.
Key Pad
-- 19 --
SERVICE
MANUAL
VS- 1000
3- 3 Dimension
3- 4 Specification
Max. Speed
Max. Force
Max. Capacity
Standard Rotor
Door
Main Controller
Microprocessor
Display
Program
20 Memories
Accel/Decel Rates
Timer
Temp. Range
-20 ~ + 40
Dimensions
Drive System
Inverter Motor
Safety Devices
Power Supply
-- 20 --
SERVICE
MANUAL
VS- 1000
3- 5 Composition
Rotor Kinds
Rotor No.
Swing
rotor
+
1
Sealing Cap
Bucket
(Option)
Aluminium Alloy
Tube Rack
Relative Tube
No
Volume
10ml
15ml
30ml
50ml
85ml(100ml)
250ml
500ml
1000ml
P.No
3114- 0010
3115- 0010
3118- 0010
3119- 0010
3110- 0150
3117- 0150
3114- 0030
3115- 0030
3118- 0030
3119- 0030
3114- 0050
3115- 0050
3118- 0050
3119- 0050
3118- 0085
3120- 0250
3122- 0250
3120- 9500
3122- 0500
3120- 1000
3122- 1000
Description
Oak Ridge Tube,
Screw Top
Round Tube, Open Top
Round Tube, Open Top
Oak Ridge Tube, Screw
Top
Dia x Height
Dia x mm
16.0 x 81.5
16.0 x 82.0
16.1 x 81.7
16.0 x 81.4
15.9 x 113.9
16.1 x 114.3
25.7 x 93.7
25.7 x 94.5
25.7 x 94.5
25.5 x 94.3
28.8 x 107.7
29.0 x 107.7
28.8 x 107.0
28.8 x 106.7
38.2 x 105.7
61.8
61.8
69.5
69.5
97.5
97.6
x
x
x
x
x
x
127.7
127.6
170.2
169.6
184.5
188.4
Material
FEP
PSF
PC
PPCO
PPCO
PC
FEP
PSF
PC
PPCO
FEP
PSF
PC
PPCO
Min RCF
4'C
22' C
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
50,000
20,000
50,000
50,000
50,000
50,000
50,000
50,000
PC
50,000
50,000
PPCO
PC
PPCO
PC
PPCO
PC
27,500
27,500
4,800
13,700
7,100
7,100
27,500
27,500
4,800
13,700
7,100
7,100
* The tubes can be changed for promotion and better quality without any notice
-- 21 --
SERVICE
MANUAL
VS- 1000
4. Installation
4- 1. Instruction.
1) When a centrifuge arrives, check closely the damages for packing.
2) When the packing do not have any damages, open and check the damages for the unit.
3) Check if the unit has all accessories.
4) Check standard power inquiry. Use the Power socket with Ground.
In case of bent plug, stop to use it. Contact to us.
If you use Power socket or plug for ground, Fire or electric accident can be occurred.
Voltage and hertz are mentioned on serial barcode of unit.
Before using the unit, check it in advance.
* If you find some problem or broken for unit, contact to us.
Warranty is attached on last page of this manual .
4- 2. Location
1) Put this unit on wide space enough. Unit should be put apart 30cm from the wall as
Standard IEC1010- 2- 020.
2) To turn the power immediately when error is occurred on unit, unit should be put around
Emergency S/ W for power.
3) Locate this unit on horizontal place.
4) While centrifuge is working, do not put anything on top of the centrifuge.
It can be blocked for good working and it can be gotten some damage.
5) Be careful Test sample do not be dropped on centrifuge.
6) All condition for mechanical, electric and location are fine, plug in . Then, check 4- 3.
(4- 3. Refer to Electric inquiry)
4- 3. Electric inquiry
1) Use a proper wire because the centrifuge requires special electrical requirements.
-- 22 --
SERVICE
MANUAL
VS- 1000
7) If the measured voltage is 10% over 230V, the malfunction or errors of the unit could
happen.
8) Check if the ground wire of a plug is grounded or the ground connection part is
disconnected.
5. Operation
5- 1. Function Test Table
No.
Description
-- 23 --
Check Point
SERVICE
MANUAL
VS- 1000
6. Disassembly
6- 1 Main Body (Before disassembly, check with engineer enough)
You should unplug when you disassemble or change parts to prevent electrical accidents.
No.
Description
Pictures
2-1
Side Open
Release 3 screws around left or right side of unit.
If you push to downward, case is opened.
Emergency Switch
There is a hole for emergency to open the door.
**
Do not change the saving value on Inverter motor control
without our approval. If you change the factory initial
value, most program can be twisted.
-- 24 --
SERVICE
MANUAL
VS- 1000
No.
Description
Pictures
Front lower case
A: Condenser
B : Motor
C : Door Device
D : Refrigerator(2HP)
E : Inverter motor Controller
Motor
RPM Sensor is located in bottom of motor.
Unbalance Sensor S/W is located around motor.
6
-- 25 --
SERVICE
MANUAL
VS- 1000
6- 2
Block Diagram
-- 26 --
SERVICE MANUAL
VS- 1000
6- 3 Circuit Diagram
-- 27 --
SERVICE
MANUAL
VS- 1000
7. Error Messages
If Error messages are shown while unit is running, contact to us immediately.
Do not disassembly it without our approval.
Error Message
Error
DISPLAY
Contents
sErr_Motor
sErr_StartThe
sErr_DoorDoor
sErr_RpmSetOver
sErr_RpmSetLow
sErr_TempLimit
" Temperature escaped a setting limit " TEMP on Chamber is over 20'C than setting TEMP.
sErr_SetAccel
sErr_SetDecel
sErr_TempSetLow
sErr_TempSetOver
sErr_Unbalance
sErr_OverRpmOver
sErr_Inverter
Reference
1. In case of low temperature, please set the temp. value before 30 minutes to secure
cooling status. If the setting temp. is lower than ambient, refrigerator starts
automatically.
2. Once you select program number, you can review all setting value of that program.
3. To check TOTAL RPM, please turn on the centrifuge with pressing of Speed mode key.
4. In case of failure of power supply, door is not opened by STOP Key. In that case, you
can open the lid forcefully by Emergency hole.
You can find it in front of unit.
This guarantee is valid only if the unit has been used in accordance
with the instructions in this manual and for its intended purpose.
-- 28 --
SERVICE
MANUAL
VS- 1000
-- 29 --
SERVICE
MANUAL
VS- 1000
VS-1000 are
-- 30 --