Vous êtes sur la page 1sur 63

MMED6000DP/S6-CSN801

Service Manual

Copyright @ 2005 Choice Electronic Technology Co. Ltd.


Edition: MMED6000DP-CSN801
Issued date: Aug.2005

MMED60000DP (S6) Service Manual

COPYRIGHT
Beijing Choice Electronic Technology Co., Ltd. (hereinafter called Beijing Choice) owns all
right to this unpublished work and intends to maintain this work as confidential. Beijing Choice
may also seek to maintain this work as an unpublished copyright. This publication is to be
used solely for the purpose of reference, operation, maintenance, or repair of Choice
equipment. No part of this can be disseminated for other purposes.
In the event of inadvertent or deliberate publication, Beijing Choice intends to enforce its right
to this work under copyright laws as a published work. Those having access to this work may
not copy, use, or disclose the information in this work unless expressly authorized by Beijing
Choice to do so.
All information contained in this publication is believed to be correct. Choice shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. This publication may refer to information and
protected by copyrights or patents and does not convey any license under the patent rights of
Beijing Choice, nor the rights of others. Beijing Choice does not assume any liability arising
out of any infringements of patents or other rights of third parties.
Content of the manual is subject to changes without prior notice.
PROPERTY OF BEIJING CHOICE ELECTRONIC TECHNOLOGY CO., LTD.
ALL RIGHTS RESERVED

MMED60000DP (S6) Service Manual

CONTENT
Chapter 1 Introduction

General

Description of abbreviation

Explanation of symbols

Special feature

Appearance of monitor

Specification

13

Chapter 2 Special operation in menu

17

Chapter 3 Monitor function and principle

19

Power supply unit

19

CSN801 board

19

Chapter 4 Component and structure


Component

21
21

Structure
Chapter 5 Troubleshooting
Diagram of solutions to various troubles
Troubleshooting
Error Message

21
30
30
40
43

Chapter 6 Calibration of monitor

44

Required testing tool

44

Power on the monitor

44

Front panel test

44

Audio system test

45

Test of automatic storage

45

Alarm test

45

ECG test

46

Respiration test

47

Temperature test

47

NIBP test

47

SPO2 test

49

Chapter 7 Adjustment procedure

50

MMED60000DP (S6) Service Manual

ECG adjustment
NIBP adjustment
SpO2 adjustment
IBP Calibration
ET-CO2 calibration
Chapter 8 Disassembly procedure
Monitor Disassembly
Battery Removal
Chapter 9 Cleaning and Disinfection
Maintenance check
General cleaning
Sterilization
Precondition and cleaning
TEMP sensor cleaning and disinfection
SpO2 sensor cleaning and disinfection
Chapter 10 Warranty
Warranty and repair
Exemption and restriction
Customer guarantee
Non-warranty and non-replacement policy
Customer special warranty period
Repackaging

50
50
50
52
54
55
55
55
57
57
57
58
59
59
60
61
61
62
62
62
62
62

MMED60000DP (S6) Service Manual

CHAPTER 1 INTRODUCTION
1.1 General
MMED6000DP is a Portable Patient Monitor that has abundant monitoring functions and is
used for the clinical monitoring of adult, pediatric and neonate. Besides, the user may select
the different parameter configuration according to different requirements.
MMED6000DP can be connected to the central working station to form a network monitoring
system.
MMED6000DP can monitor vital signals as ECG, RR, SpO2, PR, NIBP, TEMP, Dual-channel
IBP (optional), CO2 (optional). It integrates parameter measuring modules, display and
recorder in one device, featuring in compactness, lightweight and portability. Large true color
display provides clear view of 6 waveforms and full monitoring parameters.

1.2 Description of Abbreviation


HR:
SPO2:
NIBP:
IBP:
S:
M:
D:
PR:
RR:
TEMP1:
TEMP2:
CO2:

heart rate
2-channel ECG waveform
arterial oxygen saturation
non-invasive blood pressure
invasive blood pressure
systolic
mean blood pressure
diastolic
pulse rate
respiration rate
temperature channel 1
temperature channel 2
end-tidal CO2 concentration

1.3 Explanation of Symbols


Attention Refer to the relevant
the prompt.
Resistant defibrillator
BF type equipment

Class II Equipment
Heart Beat Detected

Power on/off

Rotated knob
Counter-clockwise
and clockwise

Alarm on

Alarm off

Sound on

Silence

MMED60000DP (S6) Service Manual

System setup menu

Trend menu

Equipotential grounding
terminal

AC
50/60 HZ

Fuse

This symbol indicates that the instrument is IEC 60601-1 Type CF equipment. The
unit displaying this symbol contains an F-type isolated (floating) patient applied part
providing a high degree of degree of protection against shock, and is suitable for
use during defibrillation.
This item is compliant with Medical Device Directive 93/42/EEC of 14 June
1993, a directive of the European Economic Community.

1.4 Special Features


l
l
l
l
l
l
l

Portable, compact, AC power and internal rechargeable battery;


Automatically and rapid charge, 80-90% power will be recharged within 3 hours;
Resistant high-frequency electrotome design, reliable and special module is used in
operation room;
Optional inner printer with 3 channels wave or external printer;
Adjust volume more accurately by digital system;
Menu design adopts Huffman decode, operating more effectively;
Support Ethernet, wireless LAN and could connect with Central monitoring system.

MMED60000DP (S6) Service Manual

1.5

Appearance of Monitor

1.5.1. Screen display


u 4 channels
(1)

(2)

(4)

(3)
(6)

(5)

Fig.1-1
The display of MMED6000DP patient monitor is TFT LCD, which displays the patient
parameters, waveforms, historical data and monitor status.
The screen is divided into five areas: (1) monitor status, (2) waveforms, (3) the table of
historical data, (4) parameters (5) menu as well as (6) frame of menu status.

Fig. 1-2

MMED60000DP (S6) Service Manual

Monitoring status frame


Bed No.: indicates the bed number of the patient being monitored when the patient
monitor is connected with the workstation.
DEMO: indicates the device is in DEMO monitoring mode now. In clinical application,
this function is not recommended because the DEMO will mislead the hospital
workers to treat the waveform and parameter as actual data of the patient, which may
result in delay of treatment or mistreatment.
Start time: indicates the time powered on the monitor, it shows in 16:25.
Current time: indicates the current date and time, it shows in 05/16/2004, 10:10

Waveforms displaying area


The waveform area can maximally display 6 waveforms. The colour of waveforms
matches with the parameters on the right of the screen. For the maximum
configuration, the waveforms provided by the system for selection are (from up to
bottom):
2-lead ECG waveforms
SpO2 waveform
RESP waveform
There are two means to gain RESP waveform, trans-thoracic impedance and Et-CO2
(optional).
If operator selects Et-CO2 (side stream) to monitor the patients respiration, the CO2
waveform will instead of the RESP waveform acquired by Trans-thoracic impedance.
2 IBP waveforms(option)
If operator selects IBP operation, the table of historical data will be instead of 2 IBP
waveforms.

l Numeric displaying area


The parameter area lies to the right side of the graphic area, which includes:
ECG: heart rate (unit: beats/minute)
NIBP: from left to right, there are Systolic pressure, Mean pressure and Diastolic
pressure (unit: mmHg or kPa)
IBP: the BP of channel 1 and 2. From left to right, there are Systolic pressure, Mean
pressure and Diastolic pressure (unit: mmHg or kPa)
SpO2: SpO2(unit: %)
pulse rate (unit: beats/minute)
RESP: respiration rate (unit: breaths/minute)
TEMP: temperature of channel 1 and 2 (unit:
or )
Et CO2: unit: mmHg or kPa
l

Menu
The menu always occupies the fixed position on the bottom of the screen, user can
set the system and perform operation.

Frame of menu status

MMED60000DP (S6) Service Manual

Silence on/off: showing the status of silence on or off

Alarm on/off: shows the status of alarm on

or off

Trend: the function is for operator observing the patients latest 72 hours change,
including historical data table, HR, NIBP, SpO2, ST segment.
System setup: you can configure various aspects of the monitor, including system
time, simulation, print setup, color, display wave, language, color, etc..
Menu bar: it shows the different menus of every operation.

Other modes:
IBP EXT:

ECG EXT:

MMED60000DP (S6) Service Manual

1.5.2 Function buttons

(1)
(2)
(3)
(4)
(5)
(6)

(7)

(8)

Fig.1-3
1

Alarm
ALARM button, pressing the button will prohibit all the technically audio alarm and
physically audio alarm for 3 minutes. The audio alarm function will be restored
automatically after 3 minutes or be activated when new alarm occurs. As pressing the

button, the icon in menu status shows

which indicates all the audio alarms

have been shut off.


NOTE
When

mark appears, the system can not give out the audio alarm

prompt. Therefore, the operator should use this function carefully.


2 Silence
SILENCE button, when pressing the button, this
mark appears in the menu
status, indicating that all kinds of sound including the audio alarm and heart beep have
been manually muted until the button has been pressed again, the system will
immediately restore the normal status,
and the icon shows as

NOTE
When pressing SILENCE, the system cannot give out the audio alarm and
heart beep; therefore, the operator should be considerate in using the
function.
3 Freeze
FREEZE button, pressing the button can freeze the waveform of ECG. The screen
displays two traces of ECG waveform. When one trace is active, pressing the FREEZE

MMED60000DP (S6) Service Manual

button will freeze another trace. Pressing FREEZE button again will restore the normal
monitor status.
4 Start/Cancel
START/CANCEL BP measuring button, pressing the button will inflate the cuff to start a
new NIBP measurement. When measuring, press it to cancel the measurement and
deflate the cuff, including the CYCLE and STAT (NOT STOP) measuring mode.
5 Print
PRINT button, pressing this button will motivate the recorder or desktop printer to output
the result if the monitor is equipped with them.
(6) Menu
Main MENU button, pressing the button will exit the submenu and refresh the screen.
7 Rotary Knob
The operator uses the rotary knob to select the menu
item and modify the setup. It can be rotated clockwise
or counter-clockwise and pressed like other buttons.
The operator uses the knob to realize the operations.

The method of using the knob to execute the operation:

Fig. 1-4

Rotary knob is just like the cursor of computer. When operator rotates the knob on
the icon where the operation is wanted, the icon will be automatic highlight. Then
pressing the knob, operator will open the setup menu of the corresponding
parameter so as to set up the menu.
(8) Power Supply Indicator
Indicating the power supplying state of the monitor
Orange: 220Vac main power supplying
Orange and flash: 220Vac main power supplying and battery charging
Green and flash: internal battery power supplying and 50% power has been
consumed.
1.5.3. Side Panel

MMED60000DP (S6) Service Manual

(2)
(1)

(3)

(4)
(7)

(5)

(6)

Fig.1-5
(1) ECG socket connect ECG 5-lead wire with 6 PIN connector
(2) NIBP socket connect NIBP cuff with extension tube
(3)(4) IBP (IBP1, IBP2) socket: Single, Dual channel (optional)
(5)(6) Temperature socket: connecting the TEMP probe
(7)SPO2 socket connect the SpO2 sensor with extension cable
(8) Et CO2 socket connect the nasal cannula with connector.

(8)
Fig.1-6

Fig.1-7

1.5.4 Recorder

Fig. 1-8

MMED60000DP (S6) Service Manual

A thermal dot matrices recorder with standard 50mm (+1/-1) wide printout paper is used for
MMED6000DP portable patient monitor.
NOTE
The thermal array recorder should be installed by the adequate technician.

1.5.5 Rear panel

(1)

(2)
(3)
(5)

(6)
(4)
Fig.1-9

(1) Power switch


(2) NET: Internet access point: connecting with the central monitoring system through the
standard RJ-45.
(3) CRT: exteriorly connecting screen and standard VGA.
(4) Printer: exteriorly connecting printer, which type is designated by the manufacturer.
(5) AC Power supply: 100-230 (VAC), 50/60(Hz); Fuse: standard T 1.6A
(6) Equivalent electric ground access for connecting with the hospitals grounding system.

1.6 Specification
Environment
The operation environment should comply with the following conditions:
Operating Temperature:
5 to 45
Relative Humidity:
0 to 80%, non-condensing
Height of sea level:
-500m ~ 4600m
The transport and storage environment should comply with the following conditions:
Storage Temperature:
-20 to 70
Relative Humidity:
0 to 93%, non-condensing
Height of sea level:
-500 m~ 13100m

MMED60000DP (S6) Service Manual

Display
Type:
Screen Size:

640X480 pixel color TFT


10.4 inch diagonal

Displayed Parameters
Time:
Alarms:
ECG:
Heart Rate:
NIBP:
Pulse Oximeter:
Respiration Rate:
Trends:
Temperatures:
Trace Freeze

Battery-backed quartz crystal clock


High and low limits selectable on patient parameters
ECG Waveform Scale, displayed lead
Derived from ECG OR Spo2
Pressure (systolic, mean and diastolic)
Pulse Rate, Pulse waveform, and percent saturation.
Respiration rate derived from ECG.
Heart rate, respiration rate, NIBP (systolic, mean, diastolic),
Temperatures and SpO2
Two channels
Traces A or B

ECG
Protected against defibrillator and electro surgery potentials
Standard Lead
I II III aVL aVR aVF V1 V2 V3 V4 V5 V6
Display Gain Scales
5mm/mV 7mm/mV 10mm/mV 15mm/mV 20mm/mV
25mm/mV
Sweep Speed
12mm/s 25mm/s 50mm/s
Input Resistance
> 5M Ohm (at 10 Hz, not including patient cable)
Frequency Response
0.05Hz-100Hz (3dB)
Common
Mode >60dB
Rejection Ratio
Electrode
Offset Maximum
0.3V
Potential
Baseline Recovery
<5s after 5KV defibrillation
Heart Rate range
15 to 300bpm
Resolution
1bpm
Accuracy
100bpm
1%
100bpm
2%
Alarm
Heart rate high and low limits alarm delay<12s
Lead Off condition
Detected and displayed
ST segment
ST segment range
-0.8mV to +0.8mV
Accuracy
0.05mV

Spo2
Display range
Accuracy
LED Specifications

Alarm delay
Display Update
Resolution

0 to 100%
80-100%: 2digits;
70-80%: 3digits;0-69% Unspecified
Wavelength
Radiant Power
RED
1.8mW
660 2nm
IR
2.0mW
905 10nm
Spo2 high and low limits alarm delay<7s
<5s
1%

MMED60000DP (S6) Service Manual

Pulse Rate
Measurement Range
Resolution
Accuracy
Alarm delay

20 to 250bpm
1bpm
1bpm
Pulse rate high and low limits alarm delay<7s

NIBP
Technique
Patient Types
Cuff Inflation Time
Cuff Inflation Pressure

Measurement Modes

Measurement
Interval Time
Measurement Range
Systolic
Diastolic
Mean Arterial

Oscillometric method (with inflatable cuff)


Determines systolic, diastolic and mean arterial pressures.
Adult, Pediatric and Neonate
3-15 seconds depending on cuff size.
Initially 180mmHg for Adult/Pediatric (100 or 70 mmHg for
Neonate). Subsequent inflation pressures determined by last
measured systolic pressure.
Manual: Immediate upon operator command
AUTO: Determinations automatically made with selectable
intervals
STAT: Determinations continues in 5 minutes
1-240min
Step:1min(1-10min)
5min(10-30min)
10min(30-90min)
30min(90-240min)

Adult
30-255mmHg
15-220mmHg
20-235mmHg

Infant
Systolic
Diastolic
Mean Arterial
Pressure Resolution
Accuracy

Determination Time
Overpressure Valve
Adult/Neonate
Overtime Protection
Adult/Neonate
Alarm delay

25-220mmHg
10-150mmHg
15-160mmHg
1mmHg
Cuff Pressure Range: 0 to 275mmHg
Pressure Span Accuracy: 3mmHg
Mean difference: 5mmHg
Standard deviation: 8mmHg
Typically 25seconds.Varies with patients pulse rate, pulse
pressure and amount of artifact present.
Automatically releases cuff pressure if inflation pressure
exceeds 280mmHg/150mmHg
Stop determinates if the measurement time exceeds 120s/90s.
Pressure high and low limits alarm delay<7s

RR
Technique
Range
Resolution
Accuracy
Alarm delay

Resistance method (RA-LL)


0-100 rpm
1 rpm
2 rpm
0-60rpm
Respiration rate high and low limits alarm delay<7s

MMED60000DP (S6) Service Manual

TEMP
Technique
Channel
Scales
Probes
Range
Revolution
Accuracy
Alarm delay

Resistance
2 (T1 and T2)
F. Or
Resistive; recta and skin (reusable and disposable)
YSI 400 Series types
0-50
0.1
0.1
Body temperature high and low limits alarm delay<7s

CO2 (Optional)
Technique
Side Stream, non-dispersive infrared (NDIR) absorption technique. Including multiple
water trapped/filtration system and microprocessor logic control of sample handling and
calibration.
0-99mmHg
Measurement Range
1-99 breaths per minute
Respiration rate
+/- 3 mmHg 0 - 40 mmHg
Accuracy
+/- 8% of reading 41 - 76 mmHg
+/- 10% of reading 77 - 99 mmHg
Flow rate range

User-selectable, variable from 90 to 200 ml/min (defaults


to 175 ml/min)

IBP
Range
Sensitivity
Gain Accuracy
Bandwidth

-50 to +350mmHg
5uV/V/mmHg
0.5%
0 to 12Hz

Power Requirements
Power Supply:
Internal Battery:
Operating Time:
(fully charged battery)

100~240VAC 50/60Hz 1.6A max


12V 7AH; Type- sealed lead-acid
2 hours typical at 25 ,no printing, one NIBP measurement
per 15 min.

MMED60000DP (S6) Service Manual

CHAPTER 2 SPECIAL OPERATION IN MENU


Some operations of monitor aim at the dealer but not the doctor, such as Language setup or
Hz selection, the dealer should note the following points before monitor getting end user.
NOTE: you must keep the correct setup according customers condition before using
the equipment.
In monitoring status, rotate the knob to highlight the
MENU icon, and then press the
knob, the menu bar of system setup appears on the bottom of screen.

Fig.2-1
Rotate the knob and select EXTEND item to access the sub-menu shown as below Fig.1-2

Fig.2-2
The functions of the items in the submenu please refer to section 6.5 of chapter 6 system
menu in Operators manual.
Select SYS-ACC item to access the sub-menu shown as Fig.2-3,

Fig.2-3
Then Input 2 in each of the items, press PW-ENTER to enter another sub-menu shown as
Fig.2-5:

Fig.2-4

Fig.2-5
USE TIME ,RUN TIME, relate with the password of opening equipment. For instance, set
USE TIME to 10, as powered on equipment at 11th time, the monitor will require you to input
password, just pressing function button START twice, PRINT once you are able to enter
the normal monitoring status.

MMED60000DP (S6) Service Manual

Rotate the knob to select USEDATE menu to enter in the sub-menu shown as Fig.2-6

Fig.2-6
Set the USEYEAR,USEMON,USEDAY for expiring date,CLR LOG means clear the record.
Rotate the knob to select SETUP menu to enter in the sub-menu shown as Fig.2-7

Fig.2-7
Language Setup
The system offers two languages: English and Chinese. Technician is able to set the required
menu as per customers demand.
Screen Selection
If the screen of monitor is 12.1 inch, you should select 800*600 for the item, otherwise
please select 640*480 for 10.4 inch.
SPO2 CFG
There are two options offering to you: 201&801 , if your monitor use CSN801 module to
supply SPO2,please select 801 option, if your monitor use BCI sop2 module to supply
SPO2 ,please select 201 option.

MMED60000DP (S6) Service Manual

CHAPTER 3 MONITOR FUNCTION and PRINCIPLE


3.1 Power supply unit
Power supply unit consist of AC-DC adapter, battery and Power board. The AC-DC
adapter converts 220VAC into 15VDC, and the Power board provides 5VDC to mainboard
and CSN801 board, 12VDC to TFT displayer and fan, power Supply Indication signal to
LED, charge battery port to battery, and speaker signal to speaker.

3.2 CSN801 board


3.2.1 Power
Board input voltage is 5-6v, through the DC-DC, the voltage is changed into 9v, though
manostat voltage is changed into 5v and +3.3v.
3.2.2 CPU
The board includes ECG part, SpO2 part, respiration part, temperature part, NIBP part, and
there are 3 CPU in the board altogether. ATEMEGA128L (clock 8MHz) is used as major
CPU. 128K Bytes of In-System programmable Flash, it can endure 1,000 Write/Erase
Cycles.4K Bytes EEPROM, and endure 100,000 Write/Erase Cycles. It is responsible for
the whole control of the board. Using the timer interrupt, CPU is detecting the R wave and
ST segment from ECG waveform input by ADC, calculating heart rate, the respiration rate,
and receiving the data from NIBP part, temperature part and SpO2 part, and transferring
the data to the host computer.
Z80 is another CPU, and responsible for NIBP measurement .The third CPU is 8051F007;
it is responsible for SpO2 measuring.
3.2.3 Serial communication
There are two serial RS232 I/O ports, one for transferring data to the host computer, and
another is for receiving the NIBP data from the NIBP part.
3.2.4 ECG module
3.2.4.1 Input device
Signal through the ECG lead wires is inputted into three multiplexer. One multiplexer can
selects the I, II, III, avR, avL, avF, V1, V2, V3, V4, V5, V6 lead and CAL(1mV scale voltage),
the second multiplexer can select filter mode including monitor mode, diagnosis mode,
operation mode, the software filter mode, and the third select feedback and driver. After
through the low pass filter and high pass filter, the signal made 1000 times as large by the
amplifier, and then output to the A/D Conversion of CPU.
3.2.4.2 Leadoff detector
There are two ways to detect the leadoff. One is from the input circuits, and the other is
from the feedback circuits. When one of two conditions occurs, the leadoff is detected.
3.2.4.3 Filter

MMED60000DP (S6) Service Manual

Hardware filter consists of the low pass filter and the high pass filter. The bandwidths are
different according the various filter modes. In Diagnosis mode it is 0.05-100Hz, in Monitor
mode is 0.5-75Hz, in Operation mode is 1-25Hz.
After calculating and processing from the CPU, the ECG data are transferred to the host
board.
3.2.5 NIBP
Blood pressure measurement module can be used for monitoring the blood pressure from
neonates to adults. Measurement adopts the oscillimetric method, CPU controls the pump,
valve to inflate and deflate the cuff, and the sensor transforms the pressure signals into the
electrical signals. After that, the signals are amplified by the AD623, and sampled by
1293AD. At the same time, 1293A/D converter sampled the pluses wave. Based on the
pluses wave, the systolic, diastolic and mean pressures are confirmed. According to the
difference of Neonate mode and Adults mode, overpressure protection circuit resets the
pressure value of blood pressure module to ensure the patientssafety.
3.2.6 SpO2
Pulse oximetry works by applying a sensor to a pulsating arteriolar vascular bed. The
sensor contains a dual light source and photodetector. Bone, tissue, pigmentation, and
venous vessels normally absorb a constant amount of light over time. The arteriolar bed
normally pulsates and absorbs variable amounts of light during systole and diastole, as
blood volume increases and decreases. The ratio of light absorbed at systole and diastole
is translated into an oxygen saturation measurement. This measurement is referred to
SpO2. Based on this principle, the CPU (C8051F007) controls the MOS IRF 7303,
IRF7304 to light on the red and infrared. After the signals are received, filtered, and
amplified, they are transferred to inner A/D conversion of C8051F007 and calculated. So
Spo2 and pluses rate values are obtained.
3.2.7 TEMP
The resistances is sampled by ADS1100, and transferred into the CPU, through matching
the resistance values, the temperature values are calculated.
3.2.8 RESP
Respiration wave is inputted by ECG lead. Through amplifying, filtering, the signals are
sampled by ADS1100. The digital signals are transferred to CPU by IIC bus. CPU calculates
respiration rate, transfers the respiration wave value and respiration rate to the host computer,
the host computer reappearances the respiration wave.

MMED60000DP (S6) Service Manual

CHAPTER 4 COMPONENT and STRUCTURE


4.1 Component
4.1.1 Main components of monitor
Standard:

No.
1
2
3
4
5
6
7
8
9
10
12
13
14

Production Description
Main board
SSD (solid state disk)
CSN801 board
Power board
Power inverter
Build-in power adapter
Build-in battery
040 program chip
TFT screen
Deflation valve
Inflation pump
Keyboard
Fan

Standard Quantity
1 piece
1 piece
1 piece
1 piece
1 piece
1 piece
1 piece
2 piece
1 piece
2 piece
1 piece
1 piece
1 piece

Option:

No.
Production Description
1
IBP module
2
Et-CO2 module (side-stream)
3
Thermal array recorder

Standard Quantity
1 piece
1 piece
1 piece

4.1.2 Standard accessaries

No.

Production Description

Standard Quantity

MMED6000DP(S6) Portable patient monitor

1 set

Choice reusable adult SpO2 sensor with extension cable

1 piece

Choice adult NIBP cuff with connector(25~35cm)

1 piece

Choice NIBP extension hose with connector (1.5m)

1 piece

Choice ECG cable with 5 lead wire and 6PIN connector( AHA)

1 piece

Choice TEMP skin probe

1 piece

Choice ECG electrodes (10 pieces/ pack)

1 pack

Triphase power wire

1 piece

Ground wire

1 piece

10 Build-in battery

1 piece

11 Wired net card

1 piece

4.2 Structure
4.2.1 Front panel

MMED60000DP (S6) Service Manual

2
4

5
Fig.4-1 Front panel
1.
2.
3.
4.
5.

Power Supply Indicator(LED)


TFT display screen
Rotary Knob (Optical encoder)
Keyboard
Inverter

Fig. 4-2 Panel structure

MMED60000DP (S6) Service Manual

Fig.4-3 Circuit diagram of front panel

Fig.4-4 Circuit diagram of keyboard


4.2.3 Side panel

MMED60000DP (S6) Service Manual

Fig.4-5 Side panel

Fig.4-6 Side panel structure


4.2.4 inside parts

MMED60000DP (S6) Service Manual

1
7
2

Fig.4-7 Inside view


1.
2.
3.
4.
5.
6.
7.

Pump and valve fix board


Power board
Side Panel board
AC-DC adapter
Output: 15V, 3A
Battery:
Sealed lead acid battery, 12V, 7Ah
Recorder
Mainboard

Fig. 4-8 Connection of inside board

MMED60000DP (S6) Service Manual

Fig.4-9 Profile
1.
2.
3.
4.
5.

CSN801 board
Pump and valve fix board
Power board
Mainboard
SSD

Fig. 4-10 Structure of profile


4.2.5 Pump & Valve fix board connection

MMED60000DP (S6) Service Manual

Fig.4-11

MMED60000DP (S6) Service Manual

4.2.6 FLOW-CHART DIAGRAMS


(1) Signal Flow
General Block Diagram

Inverter

TXD
RXD

TFT
Displayer

CSN801 board
ECG, NIBP,
Respiration,
Temperature

SpO2,
PUMP
VALVE

DATA

SSD

TXD

ETCO2 module

COM1

Pc104

RXD

P2
COM4

DOC
DATA

Main board

TXD
COM3
RXD

PRN
Indication
LED
DATA
Rotary Knob

Keyboard

Recorder

IBP Module

MMED60000DP (S6) Service Manual

(2) Power Flow

AC/DC
Power Supply

AC 220V

DC 15V

Power
Board

DC 5V

Main board

DC 5V

CSN801 BOARD

DC 5V
DC 12V
LED
&Volume
Signal

DC 5V

DC 12V
Key
Board

LED

Battery

Battery
charge
circuit

DC 12V

Fan

SSD

Inverter

MMED60000DP (S6) Service Manual

CHAPTER 5 TROUBLESHOOTING
5.1 Diagram of solutions to various troubles
A.
NO DISPLAY WITH AC MAIN POWER ON

NO
Is fuse blown?

NO
Is Power Supply
Indicator
(LED)
lighted?

A1
2

YES
NO

YES

Replace fuse

NO
Are there any
loose connectors?

Is TFT screen
backlight
lighted?

YES

YES

Is fuse blown again?


YES
Reconnect the connectors
Change
AC/DC power supply

A2

NO
Replace
inverter.

the

MMED60000DP (S6) Service Manual

A1
2

Does
AC/DC
power
supply
output 15V?

NO

Change
AC/DC POWER SUPPLY

YES
A2
Change
the power board

Can you hear the voice


of mainboard power-on
self-checking?

NO

Is
mainboard
supply voltage
OK?

YES

Change the TFT


displayer

YES

Change
mainboard

the

NO

Change
the power board

MMED60000DP (S6) Service Manual

B.
NO BLOOD PRESSURE READING

Is
connector
connected
properly?

NO

Reconnect
connectors.

cuff/hose

YES
NO
Does cuff inflate?

NO
Does PUMP work?

B3

YES

YES

NO
B1

Is there air leak?


YES
B2

Change the CSN801


board

MMED60000DP (S6) Service Manual

B1

NO
Is pressure inflated to
pre-set valve?

NO
Is there air leak?

Change the CSN801


board

YES
YES
B2
Is deflation speed
OK?
YES
Change the CSN801
board

B2

NO

Are CUFF and


AIR HOSE OK?

NO

Change CUFF
AIR HOSE

and

YES

Is 5V supplied to
the valves?
YES

Change CUFF and AIR


HOSE

NO

Change the CSN801


board

MMED60000DP (S6) Service Manual

B3

Turn power OFF and


ON again.

NO
Does PUMP work?

NO
Is 5V supplied to
the pump?

YES
END
Change the pump

Change the CSN801


board

MMED60000DP (S6) Service Manual

C.
NO ECG MEASUREMENT

NO
Is there ECG
ERROR alarm?
YES

Is 5V supplied
voltage is OK?

NO

Change
Power board

YES
Change
CSN801 board

NO
Change ECG cable and
Lead wire

Are ECG cable and


lead wires OK?

YES
NO
Are
electrodes
placed OK?

YES

Change CSN801 board

Replace the electrodes


refer to operation manual

MMED60000DP (S6) Service Manual

D.
NO TEMPERATURE READING

NO
Is
temperature
PROBE OK?

Replace PROBE

YES
NO
Reconnect connector

Is connection OK?

YES
NO
Is supply voltage
OK?

YES

Change
CSN801 board

Check power source

MMED60000DP (S6) Service Manual

E.
NO RESPIRATION READING

NO
Are ECG cable and
lead wires OK?

Replace PROBE

YES
NO
Replace electrodes

Are electrodes be
placed OK?

YES

Is respiration type
same
as
the
internal setup?

YES

Change
CSN801 board

NO

Select the respiration


type correctly

MMED60000DP (S6) Service Manual

F.
NO SPO2 READING

NO
Is SPO2
OK?

sensor

Replace SPO2 sensor

YES

NO
Reconnect connector

Is connection OK?

YES

NO
Is supply voltage
OK?

YES

Change
CSN801 board

Change power source

MMED60000DP (S6) Service Manual

5.2 Troubleshooting
5.2.1 The main trouble of ECG/HR
Failure
Possible cause
Display
shows
ECG CSN801
board
has
ERROR when you turn on problem.
the equipment
No ECG waveform
Poor connection of ECG
electrode films
RL
electrode
is
suspended
CSN801
board
is
damaged.
The patients skin is too
dry.
ECG waveform is abnormal
or has interference.

Only one or some leads


could display waves, some
leads display straight lines.

ECG waveform
disorderly

displays

Electrodes are connected


incorrectly.
There is suspending
electrode.
AC power has no
grounding wire.
CSN801
board
is
damaged.
Some electrodes are
not contacted well.
The lead wires are
damaged.
The CSN801 board has
problem.
Patient
is
moving
constantly.
Position of electrode
The mode of monitor
Grounded receptacle

HR
data
fluctuate
remarkably
ECG base line is not
smooth

Patient
is
moving
constantly.
Power supply is inferior

5.2.2 The main trouble with RESP


Failure
Possible cause
No RESP waveform or
Electrodes
are
RESP
waveform
is
connected incorrectly.
abnormal.
Patient
is
moving
constantly.
CSN801
board
is
damaged.

Solution
Replace the CSN801 board

Use new electrode to


ensure good contact.
Connect RL electrode.
Replace it.
Note the precondition of
skin.

Correctly
connect
electrode.
Remove electrode that
are not used.
Use 3-wire power
Replace it.
Connect the lead with ECG
simulator, if the signal is
normal, the trouble lies in
the electrodes or leads
(identify which electrode
should be changed) if the
signal is abnormal still,
change the CSN801 board.
Keep patient quiet.
The electrodes should
not be too close.
Select the right mode of
monitor
Connected
with
grounded receptacle
Keep patient quiet.
Do not use the same power
panel with high power
equipment.

Solution
Use RL-LL electrode,
connect to the correct
position.
Keep patient quiet
Replace it.

MMED60000DP (S6) Service Manual

The
data
change
remarkably
and
have
deviation from the real data.

No RESP wave

The dryness of skin


Inferior
quality
electrode
The
position
electrodes

of
of

The mode of RESP is


wrong.

5.2.3 The main trouble with TEMP


Failure
Possible cause
TEMP value is incorrect.
Measuring sensor is poorly
connected.
The measuring data is low.
The measuring time is
short.
The position of probe

5.2.4 The main trouble of Blood pressure


Failure
Possible cause
NIBP cuff cannot be inflated 1,Air way is folded or has
leakage.
2,The keyboard is bad
Blood pressure cannot be Cuff becomes loose or
measured occasionally.
patient is moving.
Error of blood pressure Cuff size does not fit the
measurement is too great.
patient.
Cuff inflates repeatedly but Pressure board or the
no data
deflation valve have trouble,
cuff and the inner pipe of
monitor have leakage.

5.2.5 The main trouble of SpO2


Failure
Possible cause
No SpO2 waveform
Sensor is damaged.
SpO2 waveform has strong
interference.

Patient is moving.
Environment light
very intensive.

is

SpO2 value is inaccurate.

Coloring agent has been


injected into patient body.

SpO2 sensor is hot.


SpO2 wave displays as
infill, no SpO2 data.
SpO2 wave displays as
zigzag waveform.

SpO2 sensor is damaged.


CSN801
board
has
problem.
The problem of ground wire

Note the precondition of


skin
Replace
the
new
electrodes.
The electrodes should
not be too close.
Select Impedance type in
RESP mode.

Solution
Connect TEMP
sensor
stably.
The measuring time
should be long.
Move the position of
probe.

Solution
1,Adjust or repair the air
way.
2, check the cable or replace it
Keep the patient quiet; bind
the cuff correctly and safely.
Use
the
cuff
with
appropriate size.
Using other cuff check if the
cuff is leak. If the cuff is
excellent, you should check
if the gas pipe is leak.
Finally replace the pressure
board and deflation valve.

Solution
Replace the sensor and
confirm the failure.
Keep the patient quiet.
Weaken
the
light
intensity
in
the
environment.
Remove the coloring agent
before
perform
measurement.
Replace the SpO2 sensor.
Replace the board.
Reconnect the ground wire.

MMED60000DP (S6) Service Manual

5.2.6 Display failures


Failure
When powering on the
device, power supply is in
normal operation, however,
screen goes black during
normal operation.

Possible cause
Backlight board damage
Bad connecting wire of
display
Damage of main board
The
battery
is
undercharge.

5.2.7 Operation, recording, network linking failure


Failure
Possible cause
Key or rotary encoder is
Keyboard or rotary
disabled.
encoder is damaged.
Connecting wire of
keyboard is damaged.
Sound is raucous or there is
no sound.

Keyboard failure
Speaker or connecting
wire failure.

Recorder cannot execute


printing operation.

Recorder has no paper.


Recorder failure
Connecting wire of the
recorder is damaged.
The model of printer has
been set incorrectly.

Record paper goes out


skew.
Cannot be linked into
network.

Bad recorder installing or


positioning.
Network linking wire is
damaged.
Main board failure

5.2.8 Other troubles


Failure
Rechargeable battery does
not charge.

Some parameter areas are


blank.
Not enter the monitor status
Start the monitor, the
monitor
can
check
automatically, but it still
stays starting Rom-Dos
User
Limited.
Enter
password

Possible cause
The
battery
is
exhausted.
The power board is
damaged.
The color setup is incorrect.
Password is locked
Mainboards
damaged

BIOS

is

1,the expire date is earlier


than the current date
2, the runtime is over the
usetime
3. the mainboard is bad

Solution
Connect external VGA
display and confirm the
failure.
Repair
or
replace
connecting wire
Replace main board
Charge the battery.

Solution
Replace keyboard or
rotary encoder.
Replaced or repair
connecting wire of
keyboard.
Replace keyboard.
Replace speaker or
connecting wire.
Install paper.
Replace the recorder.
Replace or repair the
connecting wire of the
recorder.
Set the mode of printer
correctly.
Adjust the installation of
recorder.
Check
and
repair
network-linking wire.
Replace main board.

Solution
Charge for over 24
hours.
Replace the battery and
the power board.
Reset
the
color
of
parameter.
Pressing START twice,
PRINT once release lock.
rewrite BIOS programme
or replace the main board

1,change the setting


2,change the setting
3,replace it

MMED60000DP (S6) Service Manual

Boot failure
Insert disk in unit A

Mainboard is damaged

replace

5.3 Error Message


Message

Cause of Error

ECG ERROR

Cannot communicate with ECG part of


CSN801 board

LDOFF

ECG cable or electrodes are off

PLETH ERROR

Cannot communicate with SPO2 part of


CSN801 board

PROBE OFF

SPO2 sensor is off

MMED60000DP (S6) Service Manual

CHAPTER 6 CALIBRATION OF MONITOR


6.1 Required testing tools
The following tools are required to perform calibration and verification on the
MMED6000DP monitor.
l

Mercury Manometer

Patient Simulator(ECG: 30 to 300bpm


wave at 10Hz)

Temperature Simulator

NIBP Simulator (0 to 250 mmHg)

0.5-2.0mV amplitude, 1mV square

25 to 45

l Respiration testing tool (Respiration: 30bpm, 0.1 to 3.0 Ohm impedance,


baseline impedance 500 Ohm).
l

SPO2 signal Simulator

6.2 Power on the monitor


l

For AC operation, confirm that AC Power wire is plugged in the connector on


the rear panel, and LED Power Indicator is lit Orange.

For internal battery Operation, confirm that the LED Power Indicator is
illuminated Green.

Confirm that the Screen is displayed without any error messages when
powering monitor on.

6.3 Front panel test


6.3.1 Rotary knob test
Rotate the knob in the clockwise direction while verifying that the highlight moves
from icon to icon with a clockwise movement.
Rotate the knob in the counter-clockwise direction while verifying that the highlight
moves from icon to icon with a counterclockwise movement.
Highlight any parameter icon, rotate the knob clockwise in the parameter area, verify
that the movement of cursor is from up to down.
Highlight any parameter icon, rotate the knob counter-clockwise in the parameter
area, verify that the movement of cursor is from down to up.
Highlight the RETURN menu item, press the knob and verify that the menu is return
to the superior menu.
6.3.2 Verification of Keyboard control
Press ALARM key, observe that the alarm icon in frame of menu status will change.
Press SILENCE key, observe the silence icon in frame of menu status will change.
Press FREEZE key, one channel of the ECG waveform should be frozen,
Press START key will start NIBP measurement, and press the key again will cancel
this measurement.
Press PRINT key, it can print the relevant waveform and measured data if the printer

MMED60000DP (S6) Service Manual

is installed.
Press MENU key, the display is refreshed.

6.4 Audio system test


l

Set the simulation item ON, the detailed operation please refer to the Section
9.5 ECG Menu of Operators Manual

Select ON for SOUND in ECG menu, the detailed operation please refer to the
Section 9.5 ECG Menu of Operators Manual

You will hear the simulated heart beep

6.5 Test of automatic storage


l

Enter the ECG setup menu.

Change the lead setup, e.g. select aVF for LEAD.

Turn off the monitor, and power on it after 5 minutes.

The lead should be the value you set up just now.

6.6 Alarm test


6.6.1 ECG alarm test
Connect the ECG Patient Simulator to the ECG input on the monitor side panel.
Set the Patient Simulator to output a 1mV QRS pulse at 60bpm, set the lead of
ECG submenu as II
Set the heart rate high limit as 58bpm, low limit as 30bpm.
Set the heart rate alarm as on.
Verify the heart rate visual and auditory alarms, the heart rate data should flash
and dudu voice should be heard.
Disconnect the ECG Patient Simulator
6.6.2 NIBP alarm test
Connect the NIBP Patient Simulator to the NIBP input on the monitor side panel.
Set the Patient Simulator to output Systolic: 120, Diastolic: 80
Set the systolic high limit as 110, low limit as 60.
Set the NIBP alarm as on.
Verify the systolic visual and auditory alarms, the systolic data should flash and
dudu voice should be heard.
6.6.3 SPO2 alarm verification
Connect SPO2 Probe to the SPO2 connector on the monitor side panel.
Insert the operators finger into the finger sensor, the SPO2 measured value of
healthy person should be 96%.
Set the SPO2 high limit as 90, low limit as 80.
Set the SPO2 alarm as on.

MMED60000DP (S6) Service Manual

Verify the SPO2 visual and auditory alarms, the SPO2 data should flash and
dudu voice should be heard.
Disconnect the ECG Patient Simulator

6.7 ECG test


Attach the lead to the ECG patient simulator.
Set the patient simulator to output a 1mV 10Hz sine wave, set the wave of ECG
submenu as 10mm/Mv, the ECG waveform should show the sine wave that the
height is 10mm.
Set the patient simulator to output a 1mV QRS pulse at 60bpm, set the lead of
ECG submenu as II.
Verify the ECG waveform moving QRS pulses without jittering, the number
indicate an ECG count of 60bpm.
Set the patient simulator to output a 1mV QRS pulse at 120bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an ECG
count of 120bpm.
Set the patient simulator to output a 1mV QRS pulse at 240bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an ECG
count of 240bpm.
Set the patient simulator to output a 0.5mV QRS pulse at 30bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an ECG
count of 30bpm.
Set the patient simulator to output a 2mV QRS pulse at 240bpm, verify the ECG
waveform moving QRS pulses without jittering and the number indicate an ECG
count of 240bpm.
6.7.1 ECG LEAD OFF test
Perform the following procedure to verify if the function of ECG Lead Off works
orderly.
Attach the lead to the ECG patient simulator.
Set the patient Simulator to output a 1.0mV QRS pulse at 60BPM.
Select ECG Lead for
Remove one lead from simulator in turn (that pertains to the selected lead) until
all the leads have been tested. In this situation, verify that LEADOFF
message appears in ECG displaying frame of parameter area in 12S.
Disconnect the Patient Simulator from the monitor front panel, the LEADOFF
message appears.
6.7.2 ST segment test
Perform the following procedure to verify if ST segment program works orderly.
Rotate and press the knob to enter ECG menu.
Rotate knob to select the lead
item.
Set Patient Simulator to 1mV, 60bpm, POS at 0.8 setting.
Verify ST reading of +0.8 ( 0.05).
Set Patient Simulator to 1mV, 60bpm, POS at -0.8 setting.
Verify ST reading of -0.8 ( 0.05).

MMED60000DP (S6) Service Manual

6.8 Respiration test


Attach the lead to the ECG patient simulator.
Rotate and press the knob to enter RESP menu.
Select INDEP for RESPTYPE item in the RESP submenu.
Set the Respiration Simulator to 0.5 Ohms impedance with baseline impedance
of 1000 Ohms
Set the Respiration Simulator to 15bpm, verify the RESP waveform correct and
the number indicate an RESP count of 15bpm.
Set the Respiration Simulator to 30bpm, verify the RESP waveform correct and
the number indicate an RESP count of 30bpm.
Set the Respiration Simulator to 60bpm, verify the RESP waveform correct and
the number indicate an RESP count of 60bpm.

6.9 Temperature test


Connect a Temperature Simulator to the T1 input on the monitor side panel.
Verify the following readings:
1) 30
0.1
2) 35 0.1
3) 40 0.
Connect a Temperature Simulator to the T1 input on the monitor side panel.
Verify the following readings:
1) 30
0.1
2) 35 0.1
3) 40 0.

6.10 NIBP test


Connect the manual mercury manometer to the NIBP socket.
Set the CAL ON at the NIBP submenu of the monitor.
Inflate the cuff pressure manually to 50mmHg according to the reading of the
manual mercury manometer, then the cuff reading of the monitor should be 50
3mmHg.
Verify the following points
Reading of the manual mercury manometer: 100mmHg
Cuff reading of the monitor : 100 3mmHg
Reading of the manual mercury manometer: 150mmHg
Cuff reading of the monitor : 150 3mmHg
Reading of the manual mercury manometer: 250mmHg
Cuff reading of the monitor : 250 3mmHg
If an NIBP Simulator is available, test the NIBP range and accuracy as follows:
Adult range and accuracy verification:
Systolic
Diastolic
Tolerance
60
30
5mmHg
80
50
5mmHg

MMED60000DP (S6) Service Manual

120
80
5mmHg
150
100
10mmHg
200
150
10mmHg
NOTE: the uncertainty specifications of the NIBP Simulator must be added to
the monitor tolerances for proper accuracy verification.
Test of over pressure:
1) Connect a sphygmomanometer to the NIBP inflate port
as following figure.

2) If the patient selection is NEONATTE (located under the NIBP icon),


perform the following to select ADULT:
a) Turn the Rotary Knob to highlight NIBP icon.
b) Press the Rotary Knob to select NIBP menu.
c) Turn the Rotary Knob to highlight MODE selection
d) Press the Rotary Knob and select AUTO selection.
3) Perform the following to set CAL ON:
a) Turn the Rotary Knob to highlight NIBP icon.
b) Press the Rotary Knob to select NIBP menu.
c) Turn the Rotary Knob to highlight ALARM selection
d) Press the Rotary Knob to select ALARM selection.
e) Turn the Rotary Knob to highlight CAL selection
f) Press the Rotary Knob to select CAL selection.
g) Turn the Rotary Knob to select ON selection.
4) Inflating the cuff manually by pressing the rubber ballonet. With the
pressure rising, the reading of sphygmomanometer is higher. When the
pressure over 280mmHg, the valve should deflate.

MMED60000DP (S6) Service Manual

6.11 SPO2 test


Connect SPO2 Probe to the SPO2 connector on the monitor side panel.
Insert the operators finger into the finger sensor, the SPO2 measured value of
healthy person should be from 95% to 99%, and the pulse rate is same as heart
rate, the waveform of the SPO2 is smooth with no step.
If SPO2 Simulator is available, verify the accuracy of Oxygen Saturation Value
with BCI probes as follows:
Oxygen Saturation
Tolerance
96%
2%
86%
2%
76%
3%

MMED60000DP (S6) Service Manual

CHAPTER 7 ADJUSTMENT PROCEDURE


NOTE
The MMED6000DP Monitor contains a great number of Static Sensitive circuits. All
service procedures must be done by grounded personnel.

7.1 ECG adjustment


ECG gain is fixed. User cannot adjust it. If ECG gain dose not meet the standard
demand. Please contact the service person of Choice Co.

7.2 NIBP adjustment


7.2.1 Cuff pressure calibration
Perform cuff pressure calibration as follows:
Connect a manometer with the NIBP inflating port as Fig. 7-1

Fig. 7-1
Perform the following to set CAL ON:
a) Rotate and press the knob to enter NIBP menu.
b) Then highlight ALARM to select the submenu.
c) Select ON for CAL.
Inflate the cuff manually by pressing the rubber ballonet.
Inflate the cuff to 200mmHg and observe the NIBP value displayed on the
monitor, if which equals to the value displayed on the manometer.
If the error exceeds 3mmHg, adjust the VR1 on the CSN801 board shown as Fig.
7-2.

MMED60000DP (S6) Service Manual

VR1

Fig. 7-2
7.2.2 Setup of deflection and offset
Rotate the knob to call up ALARM submenu of NIBP, set the high limit and low limit to
maximum. For example, if you want to adjust systolic, set the high limit of systolic to 255,
low limit to 220, then press RETURN to return the superior menu, next press RETURN of
the superior menu again back the main menu, at that time, you will find NIBP menu has
changed as Fig.7-3:

Fig.7-3
SYSQUTY: quotiety of diastolic and systolic value
SYSOFFS: offset of diastolic and systolic value

7.3 SpO2 adjustment


Rotate the knob to highlight the SPO2 icon in the displaying area, pick the SpO2
icon to call up the SpO2 setup menu as Fig. 7-4

Fig.7-4
Then press ALARM icon to enter ALARM submenu, set the high limit to
100(maximum), and low limit to 99(maximum).

Fig.7-5
Press RETURN to enter the SpO2 main menu, which has changed as Fig.7-6,
OFFSET item displays on the screen. You can adjust the value in this item. For
Example, if the SpO2 value is 3 lower than standard, you can adjust OFFSET
value as 3 (Figure 4); If the value is higher than standard, you can adjust value
as 2.

Fig.7-6

MMED60000DP (S6) Service Manual

7.4 IBP Calibration


7.4.1 Zeroing the Transducer
WARNING
Close the one way of the transducers three-way connected to the patient before zeroing.
l Transducer should be connected with air before zeroing
l Put the transducer at the same level with the patients cardio, namely
approximately mid-axillary lines level.
l The zeroing must be done before the monitor used to monitor the patients and
at least one time everyday. (The zeroing must be done after inserting or taking
out the power supply wire.)
1. Connect the transducer with cable, install DOME. The both ends of DOME should be
connected with air (e.g.: IBP1)
2. Set the pressure revising value to zero.
3. The pressure value will be shown on the IBP channel1, S, D value also will be shown,
and please take the shown S value as criterion
4. The S value should be 0, 1, 2, or 3. The tolerance is 0.4 kPa (3mmHg).
5.If the error exceeds the tolerance, please enter into IBP1 pressure revising item to
adjust IBP1 revising value and let S value be zero.

Fig.7-7

7.4.2 Mercury manometer calibrations


Keys
Mercury manometer calibration should be processed when starting to use the new
transducer or regularly be processed according to hospitals rules.
l The purpose of the calibration is to insure that the system can give accurate
measure result.
l You must zero the transducer before the mercury manometer calibration.
l If you want to calibrate the transducer by yourself, required equipment should
be:
Standard sphygmomanometer
3-way switch
About 25 centimeter pipeline
Calibration steps

MMED60000DP (S6) Service Manual

Fig.7.8 IBP calibration


WARNING
Never perform this calibration while patient is being monitored.
1. Close the 3-way that was open to atmospheric pressure for the zero calibration
2. Connect the pipeline to the sphygmomanometer
3. Ensure the patient isnt connected with the transducer.
4. Connect another 3-way switch with the 3-way that disconnects with the patients
pipeline (when patient is being monitored). Connect injector to one port of the 3-way,
then the sphygmomanometer and pipeline be connected the other port, open one way of
the 3-way which connect with sphygmomanometer. Ensure pressure value is zero, press
ZERO item in the menu of pressure revise value.

Fig.7.9
When presses ZERO, the menu shows as follow:

Fig.7.10
5 Inflating the mercury to the 200mmHg(or other appointed value)
6. The pressure value will be shown on the IBP channel 1, S, D value also will be
shown. And take the S value as criterion.
7.The S value should be 200mmHg, The tolerance is 0.4Kpa(3mmHg).
8 If exceed the tolerance, setup TWO menu. Under TWO item, the pressure setup
value can be adjusted between 0-400mmHg(step distance: 20mmHg) .
WARNING:
The pressure value should be as same as mercury pressure value. For example: setup
this calibration to 200mmHg under TWO item, then return.
9 After one time calibration the mercury manometer can be deflated, then add pressure
renewably, checking the transducers veracity after calibration. If not correct, please

MMED60000DP (S6) Service Manual

repeat the above steps and redoing calibration operation.


10 Remove the blood-pressure pipeline and the additive 3-way.

7.5 Et-co2 adjustment

7.5 ET-CO2 calibration


7.5.1 Rotate the knob to highlight the RESP icon in the displaying area, pick the RESP
icon to call up the RESP setup menu as Fig. 7-11, then select CO2 for the type
item.

Fig.7-11

7.5.2 Then press ALARM icon to enter ALARM submenu, set the high limit to 75
(maximum), and low limit to 70(maximum), next please press RETURN to back
the former menu.

Fig.7-12

7.5.3 Enter into the CO2SETUP again, OFFSET item is appeared on the menu, then
you can adjust the offset from 20 to +20 to calibrate the CO2 value.

Fig.7-13

MMED60000DP (S6) Service Manual

CHAPTER 8 DISASSEMBLY PROCEDURE


8.1 Monitor disassembly
NOTE
The disassembly procedures should be performed by qualified person.
Perform the following procedure to open external cabinet:
Disconnect AC power wire from monitor rear panel.
Unscrew the 3 screws from the rear panel of monitor.
Screw 3
Screw 1

Screw 2

Fig.8-1
After unscrewing 3 screws, the front panel and rear panel are separated.

8.2 Battery disassembly


After opening the front panel, the inside components are exposed.

MMED60000DP (S6) Service Manual

Battery
AC-DC
adapter

Fig.8-2
Unscrew the 6 screws on the metal shell; the battery and adapter are disassembled
together from inner. Draw the battery out the shell.
Unplug the battery cable from the power board, the battery can be removed.

Fig.8-3

MMED60000DP (S6) Service Manual

CHAPTER 9 CLEANING AND DISINFECTION

NOTE
Before cleaning the monitor or the sensor, make sure to turn off the power and
disconnect the AC power.

9.1 Maintenance checks


Before using the monitor, do the following:
1. Check if there is any mechanical damage;
2. Check all the outer cables, inserted modules and accessories;
3. Check all the functions of the monitor to make sure that the monitor is in good
condition.
If finding any damage on the monitor, stop using the monitor on patient.
6. The overall check of the monitor, including the safety check, should be performed
only by qualified person once every 6 to 12 month and each time after fix up.

9.2 General cleaning


1. MMED6000DP Patient Monitor must be kept dust-free.
2. It is recommended to regularly cleaning the monitor shell and the screen. Use only
non-caustic detergents such as soap and water.
NOTE
Please pay special attention to the following items to avoid damaging
MMED6000DP:
1.

Avoid using ammonia-based or acetone-based cleaners such as acetone.

2.

Most cleaning agents must be diluted before use. Follow the manufacturer's
directions carefully for dilution.

3.

Do not use the grinding material, such as steel wool etc.

4.

Do not let the cleaning agent enter into the chassis of the system. Do not
emerge any part of the device into any liquid.

5.

Do not leave the cleaning agents on any part of the device surface.

6. Except for those cleaning agents listed in NOTE part, following disinfectants

MMED60000DP (S6) Service Manual

can be used on the instrument:


l

Diluted Ammonia Water

Diluted Sodium Hyoichlo (Bleaching agent).

NOTE
The diluted sodium hyoichlo from 500ppm(1:100 diluted bleaching agent) to
5000ppm (1:10 bleaching agents) is very effective. The concentration of the diluted
sodium hyocihlo depends on how many organisms (blood, mucus) on the surface
of the chassis to be cleaned.
l

Diluted Mindrayhylene Oxide 35% -- 37%

Hydrogen Peroxide 3%

Alcohol

Isopropanol

NOTE
l

MMED6000DP monitor and sensor surface can be cleaned with hospital-grade


ethanol and dried in air or with soft and clean cloth.

Choice Co. has no responsibility for the effectiveness of controlling infectious


disease using these chemical agents. Please contact infectious disease
experts in your hospital for details.

9.3 Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when
stipulated as necessary in the Hospital Maintenance Schedule. Sterilization facilities
should be cleaned first.
Recommended sterilization material: Ethylate, and Acetaldehyde.
CAUTION
1. Follow the manufacturers instruction to dilute the solution, or adopt the lowest
possible density.
2. Do not let liquid enter the monitor.
3. No part of this monitor can be subjected to immersion in liquid.
4. Do not pour liquid onto the monitor during sterilization.

MMED60000DP (S6) Service Manual

5. Use a moistened cloth to wipe off any agent remained on the monitor.
6. To avoid extended damage to the equipment, disinfecting is only recommended
when stipulated as necessary in the Hospital Maintenance Schedule.
Disinfecting facilities should be cleaned first.
7.Appropriate disinfecting materials for ECG lead, SpO2 sensor, blood pressure
cuff, TEMP probe, IBP sensor are introduced Operators Manual respectively.
8.

Do not use EtO gas or formaldehyde to disinfect the monitor.

9.4 Precondition and cleaning


NOTE
Before cleaning the monitor or the sensor, make sure to turn off the power and
disconnect the AC power.
If ECG cable is damaged or aged, replace with a new ECG cable.
1 Cleaning
MMED6000DP monitor and sensor surface can be cleaned with hospital-grade ethanol
and dried in air or with soft and clean cloth.
2 Sterilization
To avoid extended damage to the equipment, sterilization is only recommended when
stipulated as necessary in the Hospital Maintenance Schedule. Sterilization facilities
should be cleaned first.
3 Materials recommended for use in sterilization
Ethylate: 70%
ethanol: 70%
Acetaldehyde
4 Disinfection
To avoid extended damage to the equipment, disinfection is only recommended when
stipulated as necessary in the Hospital Maintenance Schedule. Disinfection facilities
should be cleaned first.

9.5 TEMP sensor cleaning and disinfection


1. The TEMP probe should not be heated above 100

(212). It should only be

MMED60000DP (S6) Service Manual

subjected briefly to temperatures between 80

(176) and 100

(212).

2. The probe must not be sterilized in steam.


3. Only detergents containing no alcohol can be used for disaffection.
4. The rectal probes should be used, if possible, in conjunction with a protective rubber
cover.
5. To clean the probe, hold the tip with one hand and with the other hand rubbing the
probe down in the direction of the connector using a moist lint-free cloth.
NOTE
l

Disposable TEMP probe must not be re-sterilized or reused.

For protecting environment, the disposable TEMP probe must be recycled or


disposed of properly.

9.6 SpO2 sensor cleaning and disinfection


NOTE
Do not subject the sensor to autoclaving.
Do not immerse the sensor into any liquid.
Do not use any sensor or cable that may be damaged or deteriorated.

1. Use a cotton ball or a soft mull moistened with hospital-grade ethanol to wipe the
surface of the sensor, and then dry it with a cloth. This cleaning method can also be
applied to the luminotron and receiving unit.
2. The cable can be cleaned with 3% hydrogen dioxide, 7% isopropanol, or other active
reagent. However, connector of the sensor shall not be subjected to such solution.

MMED60000DP (S6) Service Manual

CHAPTER 10 WARRANTY
10.1 Warranty and repair
Repair response time: AM9:00 to PM17: 30 on Monday to Friday except legal holiday.
Repair time: AM9:00 to PM17: 30 on Monday to Friday except legal holiday.
Repair service: Including telephone support, field inspecting, fittings replacement.
l

Telephone support: we can give guidance to customers engineer to inspecting


the instrument when you dial our service line (86-10-86863551). Professional

repair engineer online provides technical support.


Field inspecting: we will send engineers to repair the instrument if necessary.
Certified engineers of our company or local repair team trained by our company
provide this service.
Fittings replacement: if necessary, we will replace the damaged fittings according
to contract. The damaged fittings should be returned to us expect for special
reason.

Spare machine for repair: it is used to replace the damaged machine for customer
using, customer should send the damaged machine to us to repair.
Repair for sponsoring and contributing machine: customer should send the machine
to us to repair.
Updating software is free.

10.2 Exemption and restriction


Warranty does not apply to the damage or loss sustained due to well-known act
of god, such as fire, earthquake, flood, thunder, cyclone,
hail, electrical storm, blast, building collapse, commotion,
etc.
Non-service items:
The cost and insurance of dismantling and testing, overhauling, reinstall, transfer,
moving the instrument or parts.
Damage or loss sustained due to inspected or repaired by other institute that is not
certified
Damage or alteration by anyone other than our company authorized service
personnel.
The damage or lose sustained due to connection to peripheral equipment (such as
printer, computer etc.), that are not provided by our company are not covered by the
warranty.
Obligation restriction: In the duration of warranty, if the operators use other fittings that
are not provided by us, we reserve the right to cancel warranty.

MMED60000DP (S6) Service Manual

10.3 Customer guarantee


Read the user manual in details before operation.
Operation and maintenance according to the user manual, and guarantee the
requests of power and environment.

10.4 Non-warranty and non-replacement policy


The work environment is not eligible. For example, if the relative humidity
exceeds 70%, circuit boards of the instrument may be damaged due to
condensate.
If voltage of power supply is fluctuant and exceeds 240VAC, the power
adapter may be damaged.
There is smear or marks that are not belong to the instrument and cannot
be removed from the outside surface of the instrument.
The instrument or its fittings are mechanically damaged.
The circuit is short and damaged due to liquor or other stuff flow in the
instrument or its fittings.
All probe and its accessories are not free replacement.
Leakage of air cell of blood pressure sleeve due to improper storage or
operation is not free replacement.
The malfunction with result form improper repair by anyone other than our
company authorized service personnel.
The malfunction with result from improper use.

10.5 Customer special warranty period


Due to we stipulate the warranty period according to the relevant electronic regulation of
country, which we stipulate is on year, accessory is three months. When customer
requires to extending the warranty period, you should consider whether it is reasonable.
Because electronic product quickly replace, as to the warranty period over three years,
purchased accessories may be out of stock. In this case, we will adopt to entirely
upgrade or replace the old, you should pay the minimum acceptable cost of renewed
device.

10.6 Repackaging
Remove all the detectors, leads and accessories and put them into the plastic bag.
Try to use the original packaging case and materials. Any damage due to the improper
packaging during the transportation shall be responsible by the user.
If you are still within the period of warranty, please present the warranty card and one
copy of the invoice or receipt.
Please present a written note detailing all the troubles when repairing the
instrument.

MMED60000DP (S6) Service Manual

For further information or assistance with this product:


Beijing Choice Electronic Technology Co., Ltd.
Room 1126-1127,BuildingB, BAILANGYUAN No.A36, Fuxing Road, Beijing China
Zip Code: 100039
Tel: +86-10-88203551 +86-10-88203520
Fax: +86-10-88204632
E-mail:market@choicemmed.com;intlsupport@choicemmed.com
http://www.choicemmed.com

Vous aimerez peut-être aussi