Académique Documents
Professionnel Documents
Culture Documents
Home
About
News
Careers
Support us
Contact
Health Professionals
Patients and Families
Departments and Services
Research
Oxygen delivery
Introduction
Definition of terms
Normal values
Indications for oxygen delivery
Nurse initiated oxygen
Delivery mode
o
o
o
o
Considerations
o
o
Introduction
The goal of oxygen delivery is to maintain targeted SpO2 levels in children through the
provision of supplemental oxygen in a safe and effective way which is tolerated by
infants and children to:
o
o
o
o
o
Definition of terms
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
o
Minute ventilation: The total amount of gas moving into and out of the lungs per
minute. The minute ventilation (volume) is calculated by multiplying the tidal
volume by the respiration rate, measured in litres per minute.
Peak Inspiratory Flow Rate (PIFR): The fastest flow rate of air during
inspiration, measured in litres per second.
Tidal Volume: The amount of gas that moves in, and out, of the lungs with each
breath, measured in millilitres (6-10 ml/kg).
Ventilation - Perfusion (VQ) mismatch: An imbalance between alveolar
ventilation and pulmonary capillary blood flow.
Normal Values
o
o
o
o
NB: The above values are genearlised to the paediatric population, for age specific
ranges please consult CLARA and/or the medical team.
The above values are expected target ranges. Any deviation should be documented on the
observation chart as MET modifications.
If you require further information please click here for the Assessment of Severe
Respiratory Conditions guideline.
o
o
Oxygen is a drug and requires a medical order. Each episode of oxygen delivery
should be ordered on the medication chart either as a one-off order or on-going
treatment.
o
o
o
o
Note: Some flow meters may deliver greater than the maximum flow indicated on the
flow meter if the ball is set above the highest amount. Use caution when adjusting the
flow meter.
Note: Oxygen therapy should not be delayed in the treatment of life threatening hypoxia.
Note: In low flow systems the flow is usually titrated (on the flow meter) and recorded in
litres per minute (LPM).
Ventilators
CPAP/BiPaP drivers
Face mask or tracheostomy mask used in conjunction with an entrainment device
or AIRVO 2 Humidifier
High flow nasal prongs (HFNP)
Documentation:
o
Document the FiO2 as indicated on air entrainment device & total flow as per Air
Entrainer chart below and wall O2 flowmeter flow.
Humidification
Oxygen therapy can be delivered using a low flow or high flow system. All high flow
systems require humidification. The type of humidification device selected will depend
on the oxygen delivery system in use, and the patient's requirements. The humidifier
should always be placed at a level below the patient's head.
Rationale:
o
o
o
o
Indications:
o
o
o
o
o
RCH predominantly uses the Fisher & Paykel MR850 Humidifier & AIRVO 2
Humidifier. Please consult user manuals for any other models in use.
Invasive Mode - delivers saturated gas as close to body temperature (37 degrees,
44mg/L) as possible.
Suitable for patients with bypassed airways:
Invasive Ventilation
Tracheostomy attachment or mask
Nasal Prongs
Non-Invasive Mode delivers gas at a comfortable level of humidity (31-36
degrees, >10mg/L).
Suitable for patients receiving:
Face mask therapy:
Non-invasive ventilation (CPAP/BIPAP)
Nebuliser mask (with RT308 circuit)
AIRVO 2 Humidifier
Follow instructions in the AIRVO 2 User Manual in conjunction with this Guideline.
Has two modes:
o
Junior Mode
Suitable for patients using Optiflow Junior Infant and Paediatric Nasal
Prongs
Standard Mode
Suitable for patients using:
Optiflow adult nasal prongs
Nebuliser mask (via Mask Interface Adaptor)
Tracheostomy mask (via Mask Interface Adaptor)
Tracheostomy direct connection
For routine cleaning instructions please refer to the following link: RCH Equipment
Cleaning Table Prepared by Infection Prevention and Control Team
When commencing therapy on a new patient, ensure the disinfection cycle was
performed. On device start up, a green traffic light confirms the AIRVO 2 is safe for use
on a new patient. An orange traffic light confirms the AIRVO 2 has not been cleaned and
disinfected since last use, and is not safe for use on a new patient.
Delivery Mode
Click to view the delivery mode quick reference table
room air around the nasal prongs and a complete seal is not created the prong size should
be approximately half the diameter of the nares. Select the appropriate size nasal prong
for the patient's age and size.
Note: Do not use air entrainment device with simple nasal prongs.
With the above flow rates humidification is not usually required. However, if
humidification is clinically indicated - set up as per the recommended guidelines for the
specific equipment used. As with the other delivery systems the inspired FiO2 depends on
the flow rate of oxygen and varies according to the patient's minute ventilation.
Care and considerations of child with simple nasal prongs:
o
o
o
o
o
Position the nasal prongs along the patient's cheek and secure the nasal prongs on
the patient's face with adhesive tape.
Position the tubing over the ears and secure behind the patient's head. Ensure
straps and tubing are away from the patient's neck to prevent risk of airway
obstruction.
Check nasal prong and tubing for patency, kinks or twists at any point in the
tubing and clear or change prongs if necessary.
Check nares for patency - clear with suction as required.
Change the adhesive tape weekly or more frequently as required
Optiflow nasal prongs are compatible for use in humidified low or high flow oxygen
delivery.
Note: MR850 Humidifier should be placed in Invasive Mode for Nasal Prongs Therapy.
See guides below for recommended patient sizing and flow rates.
Premature
Neonate
Infant
Paediatric
See Appendix A for further information regarding appropriate junior range sizing: Fisher
and Paykel Optiflow junior range sizing guide
Fisher and Paykel Optiflow nasal cannula standard range
Three sizes of prongs
o
Small
o
o
Medium
Large
Paediatric Patients
(RT330 circuit - click here for instructions for use)
o
o
o
o
The main safety feature of the RT330 Oxygen Therapy System is the pressure relief
valve. The pressure relief valve has been set to a limit of < 40 cm H20. This valve has
been designed to minimize the risk of excessive pressure being delivered to the infant in
the event that the nasal prongs seal around the infant's nares while the mouth is closed.
3 sizes of prongs:
Small
Medium
Large
Low flow, a maximum of:
4 LPM for children over 2 years of age
FiO2 21-100% (direct from O2 wall source or via a blender)
High flow (in approved areas only, see relevant recommendations above)
Flow of 2L/kg/min up to 10kg, plus 0.5L/kg/min for each kg above 10kg
(to a maximum of 50LPM)
FiO2 above 50% requires PICU review
Junior Mode
Standard Mode
Junior Mode
o
Patients requiring FiO2 > 50% require a medical review and close
monitoring. Consider PICU transfer if after one hour no clinical
improvement has occurred
Fisher and Paykel Optiflow nasal cannula junior range for AIRVO 2
Below is an image of the Fisher and Paykel Optiflow nasal cannula junior range for
AIRVO 2
Standard Mode
o
Three sizes of Optiflow nasal prongs suitable for use with AIRVO 2 Humidifer
(click here for: Fisher and Paykel Optiflow (adult) nasal cannula standard range
guide)
Small
Medium
Large
High Flow (in approved areas only)
Flow of 2L/kg/min up to 10kg, plus 0.5L/kg/min for each kg above 10kg
(to a maximum of 50 LPM)
FiO2 21-50%
Patients requiring FiO2 > 50% require a medical review and close
monitoring. Consider PICU transfer if after one hour no clinical
improvement has occurred
Optiflow Nasal Prong junior and standard humidification and flow rate guide for Airvo.
Face Mask
Click to view the delivery mode quick reference table
The minimum flow rate through any face mask or tracheostomy mask is 4 LPM as
this prevents the possibility of CO2 accumulation, CO2 re-breathing and drowsiness.
Select a mask which best fits from the child's bridge of nose to the cleft of jaw, and adjust
the nose clip and head strap to secure in place.
Oxygen (via intact upper airway) via a simple face mask at flow rates of 4LPM does not
require routine humidification. However, as compressed gas is drying and may damage
the tracheal mucosa humidification might be indicated/appropriate for patients with
Nebuliser mask
Nebuliser mask or tracheostomy mask in conjunction with a humidification system
A nebuliser mask or tracheostomy mask with an air entrainment device is intended for
use with a MR850 Humidifier). By entraining room air into the delivery system, the total
gas flow to the patient can be increased up to approximately 45 LPM. When the
instructions on the air entrainment device are followed, it is possible to deliver a specific
FiO2. This system is useful in accurately delivering low concentrations of oxygen (2850%).
A nebuliser mask or tracheostomy mask with a mask interface adaptor is intended for use
with an AIRVO 2 Humidifier. A tracheostomy direct connection may be used with an
AIRVO 2 Humidifier. The AIRVO 2 Humidifier flow rate should be set to meet or
exceed the patient's entire ventilatory demand, to ensure the desired FiO2 is actually
inspired by the patient. This system is useful in accurately delivering concentrations of
oxygen (21 95%). Patients who require an FiO2 greater than 50% require medical
review.
With both systems, as the gas flow is > 4LPM re-breathing of expired gas is not a
potential problem. Therefore this system reduces the risk of carbon dioxide retention.
NOTE: While a specific FiO2 is delivered to the patient the FiO2 that is actually inspired
by the patient (ie what the patient actually receives) varies depending on:
o
o
o
flow rate delivered to the patient (see air entrainment device below)
mask size and fit
the patient's respiratory rate
o
o
Ensure the flow rate from the wall to the mask is adequate to maintain a fully
inflated reservoir bag during the whole respiratory cycle (i.e. inspiration and
expiration).
Do not use with humidification system as this can cause excessive 'rain out' in the
reservoir bag.
Not routinely used outside of ED and PICU and should only be used in
consultation with the medical team.
Tracheostomy
Click to view the delivery mode quick reference table
Tracheostomy HME - Heat Moisture Exchanger (Swedish nose filter) with oxygen
attachment
In spontaneously breathing tracheostomy patients who require oxygen flow rates of less
than 4 LPM there are two options available:
o
o
OXY-VENT with Tubing: The adaptor sits over the TRACH-VENT and the
tubing attaches to the oxygen source (flow meter).
TRACH-VENT+: Alternatively a Hudson RCI HME - TRACH-VENT+
(Swedish nose filter) has an integrated oxygen side port which connects directly
to oxygen tubing which is attached to the oxygen source (flow meter).
Note: HME are used without a heated humidifier circuit.
Considerations:
o
o
The Hudson Trach-Vent HME has a dead space of 10mL and is recommended
for use in patients who have tidal volumes of 50mL and above.
Trach-Vent's are changed daily or as required if contaminated or blocked by
secretions.
Isolette
At the RCH, oxygen therapy via an isolette is usually only for use in the Butterfly
neonatal intensive care unit. (See Isolette use in paediatric wards, RCH internal link
only.)
Considerations
o
o
o
o
o
o
o
CO2 Narcosis - This occurs in patients who have chronic respiratory obstruction
or respiratory insufficiency which results in them developing hypercapnea (i.e.
raised PaCO2). In these patients the respiratory centre relies on hypoxaemia to
maintain adequate ventilation. If these patients are given oxygen this can reduce
their respiratory drive, causing respiratory depression and a further rise in PaCO2
resulting in increased CO2 levels in the blood and CO2 narcosis.
Monitoring of SpO2 or SaO2 informs of oxygenation only. Therefore, beware of
the use of high FiO2 in the presence of reduced minute ventilation.
Pulmonary Atelectasis
Pulmonary oxygen toxicity - High concentrations of oxygen (>60%) may
damage the alveolar membrane when inhaled for more than 48 hours resulting in
pathological lung changes.
Retrolental fibroplasia (also known as retinopathy of prematurity) An alteration
of the normal retinal vascular development, mainly affecting premature neonates
(<32 weeks gestation or 1250g birthweight), which can lead to visual impairment
and blindness.
Substernal pain-due: characterised by difficulty in breathing and pain within
the chest, occurring when breathing elevated pressures of oxygen for
extended periods.
Oxygen safety
Oxygen is not a flammable gas but it does support combustion (rapid burning). Due to
this the following rules should be followed:
o
o
o
o
o
o
o
Evidence Table
... coming soon
References
Bateman, N.T. & Leach, R.M. (1998). ABC of Oxygen - Acute oxygen therapy. BMJ,
September 19; 317(7161): 798-801.
Dunn, L., & Chisholm, H. (1998). Oxygen Therapy. Nursing Standard, 13(7), 57 - 60.
Fell, H., & Boehm, M. (1998). Easing the discomfit of oxygen therapy. Nursing Times,
94 (38), 56 - 58.
Frey, B., & Shann, F. (2003). Oxygen administration in infants. Archives of Disease in
Childhood - Fetal and Neonatal Edition, 88, F84 - F88.
Oh, T.E. (1990). Intensive Care Manual 3rd Edition. Sydney: Butterworths.
Shann, F., Gatachalian, S., & Hutchinson, R. (1988). Nasopharyngeal oxygen in children.
The Lancet. 1238 -1240.
St. Clair, N., Touch, S. M., & Greenspan, S. (2001) Supplemental Oxygen Delivery to the
Nonventilated Neonate. Neonatal Network. 20 (6), 39-45.
Bersten, A. & Soni, N. (Eds). (2009). Oh's Intensive Care Manual 6th Edition. China:
Butterworth Heinemann Elsevier
Schibler, A., Pham, T.,Dunster, K., Foster, K., Barlow, A., Gibbons, K., and Hough, J.
(2011) Reduced intubation rates for infants after introduction of high-flow nasal prong
oxygen delivery. Intensive Care Medicine. May;37(5):847-52
McKieman, C., Chua, L.C., Visintainer, P. and Allen, P. (2010) High Flow Nasal
Cannulae Therapy in Infants with Bronchiolitis. Journal of Pediatrics 156:634-38
Spentzas, T., Minarik, M., Patters, AB., Vinson, B. and Stidham, G. (2009) Children with
respiratory distress treated with high-flow nasal cannula. Journal of Intensive Care
Medicine. 24(5): 323-8
Revision of the Oxygen delivery guideline, origionally published Oct 2012, was
coordinated by Sueellan Jones, Respiratory Nurse Consultant, Department of Respiratory
Medicine, and Brenda Savill, Nurse Educator, Nursing Education. Approved by the
Clinical Effectiveness Committee. Authorised by Bernadette Twomey, Executive
Director Nursing Services. Revised guidelines published November 2013.
Support us