FormulaforCalculatingtheNumberofTheoreticalPlates
HPLCABasicKnowledgeofAnalysis
CONTACTSEARCH
ProductInformation
Quotation
TechnicalInquiry
OtherInquiry
Introduction
N,thenumberoftheoreticalplates,isoneindexusedtodeterminetheperformanceandeffectivenessofcolumns,andis
calculatedusingequation(1).
1)wheretr:retentiontime,andW:peakwidth
Thispeakwidth,W,isbasedonthebaselineinterceptsoftangentlinestoaGaussianpeak,whichisequivalenttothe
peakwidthat13.4%ofthepeakheight.
However,tosimplifythecalculationandaccommodatenonGaussianpeaks,thefollowingcalculationmethodsareusedin
actualpractice.
June2016
ShimadzuNewApplications
arenowavailable.
May2016
ShimadzuNewApplications
arenowavailable.
April2016
ThelatestissueofShimadzu
Journalfocusingonnew
energyisnowavailable.
April2016
ShimadzuNewApplications
arenowavailable.
March2016
Youcandownloadthe
postersthatShimadzu
presentedatPittcon2016.
Fig.1PeakWidth
1.TangentLineMethod
Peakwidthisthedistancebetweenpointswherelinestangenttothepeak'sleftandrightinflectionpointsintersectthe
baseline,andiscalculatedusingequation(1).TheUSP(UnitedStatesPharmacopeia)usesthismethod.Thisresultsin
smallNvalueswhenpeakoverlapislarge.
Thisalsopresentsaproblemifthepeakisdistorted,sothatithasmultipleinflectionpoints.
1)
2.HalfPeakHeightMethod
Widthiscalculatedfromthewidthathalfthepeakheight(W0.5).Sincewidthcanbecalculatedeasilybyhand,itisthe
mostwidelyusedmethod.ThisisthemethodusedbytheDAB(GermanPharmacopeia),BP(BritishPharmacopeia),and
EP(EuropeanPharmacopeia).
2)
TheJapanesePharmacopoeia15threvisionissuedinApril2006changedthecoefficientfrom5.55to5.54.
(LCsolutionallowsselectingthecoefficientviathe[ColumnPerformance]setting,wherethecalculation
methodfor5.54is"JP"andfor5.55is"JP2."
Forbroaderpeaks,thehalfpeakheightmethodresultsinlargerNvaluesthanothercalculationmethods.
2)
3.AreaHeightMethod
Widthiscalculatedfromthepeakareaandheightvalues.Thismethodprovidesrelativelyaccurateandreproducible
widths,evenfordistortedpeaks,butresultsinsomewhatlargerNvalueswhenpeakoverlapissignificant.
3)A:Area,H:Height
4.EMGMethod(ExponentiallyModifiedGaussian)
Thismethodintroducesparametersthataccommodatetheasymmetryofpeaks,andusesthepeakwidthat10%ofthe
peakheight(W0.1).Sinceitusesawidthnearthebaseline,itresultsinNvalueslargerthanothermethodsforbroad
peaks.Furthermore,itcannotcalculatethewidthunlessthepeakiscompletelyseparated.
4)a0.1:Widthoffirsthalfofpeakat10%heightb0.1:Widthofsecondhalfofpeakat
10%height
ComparisonofCalculationMethods
GivenaGaussianpeak,eachofthesecalculationmethodsresultsinthesameNvalue.However,normallypeakstendto
havesometailing,whichresultsindifferentNvaluesfordifferentcalculationmethods.
Therefore,thefourcalculationmethodswerecomparedusingchromatograms.ProfileAshowsatypicalchromatogram
(withsometailing),whereasprofileBshowsachromatogramwithsignificanttailing.Thetheoreticalnumberofplates
calculatedusingthefourmethodsareindicatedinthetablebelow.ResultsforNvariedevenforchromatogramA.Also,
peakswithmoresignificantdistortion,suchasatpeak1inprofileB,canresultinNvaluesthatdifferbymanytimes.
Akeyfactorforperformingreliablequantitativeanalysisiswhetherornotseparationispossible,sothereisacommon
opinionthatacalculationmethodthatjudgesbroaderpeaks,suchaswithtailing,moreseverelyismorepractical.
However,unfortunately,thereseemstobenoconsensusonopinionsregardingNandW.
Consequently,ifacertainmethodisalreadybeingusedforevaluation,thentoachievecorrelation,itisprobably
preferabletokeepusingthesamemethod.
Fig.2Chromatograms
ComparisonofTheoreticalNumberofPlates
A(roughlytypicalpeak)
B(significanttailing)
HalfPeakHeightMethod
15649
20444
20389
22245
5972
7917
9957
TangentLineMethod
14061
18516
20309
21447
5773
7692
5795
9707
AreaHeightMethod
13828
19207
17917
21020
4084
7845
6217
8641
EMGMethod
10171
15058
14766
17836
1356
4671
Ahyphenindicatescalculationwasnotpossible.Inthehalfpeakheightmethod,5.54wasusedasthecoefficient.
Shimadzu'sLCworkstationsoftwareisabletooutputperformancereportsusinganyofthemethodsindicatedabove1.tangent
line,2.halfpeakheight(5.54),2'.halfpeakheight(5.55),3.areaheight,orEMG.Werecommendrecordingthecorresponding
columnperformanceresultsalongwithanalyticalresults!
LiquidChromatography
LiquidChromatographMassSpectrometry
ForResearchUseOnly.Notforuseindiagnosticprocedures.
Thispagemaycontainreferencestoproductsthatarenotavailableinyourcountry.
Pleasecontactustochecktheavailabilityoftheseproductsinyourcountry.
PRIVACYPOLICY
TERMSOFUSE
SITEMAP