Fabius GS

Technical
Documentation

Copyright by Drger Medical AG & Co. KGaA, Lbeck, Federal Republic of Germany
Copying is prohibited for commercial purposes.
Subject to change without notice. No replacement in case of changes.
Attention is to be paid to the Instructions for Use/Operating Instructions.
This technical documentation does not replace the Instructions for Use/Operating
Instructions.
The warranty and liability conditions of the general terms and conditions for business
transactions of Drger Medical AG & Co. KGaA are not extended by this technical
documentation.
Safety Regulations:
Reference is hereby made to the observance of the relevant safety provisions, such
as the Medical Equipment Ordinance (Medizingerteverordnung), the Pressure
Container Ordinance (Druckbehlterverordnung), the Technical Rules for Pressurised
Gases (Technische Regeln Druckgase) or the Occupational Health and Safety
Provisions (Unfallverhtungsvorschriften).
Insofar as reference is made to laws, regulations or standards, these are based on the
legal system of the Federal Republic of Germany. Follow your local laws and
regulations.

&RQWHQWV
*HQHUDO


6\PEROVDQG'HILQLWLRQV

)XQFWLRQDO'HVFULSWLRQ


*HQHUDO,QIRUPDWLRQDERXWWKH)DELXV*6

)DELXV*6)XQFWLRQ'LDJUDP



%DWWHU\%DFNXS



)DELXV*63LSLQJ'LDJUDP



)XQFWLRQ'HVFULSWLRQRI*DV%R[



625& 6HQVLWLYH2[\JHQ5DWLR&RQWUROOHU 



&RPSDFW%UHDWKLQJ6\VWHP&RV\,,



For internal use only. Copyright reserved.

K6020001IECIVZ.fm 01.02.02

7.1

Function Description: Manual Ventilation .............................................. 21

7.2

Function Description: Spontaneous Breathing ..................................... 23

7.3

Function Description: Volume Mode Ventilation ................................... 25

7.4

CosyII Absorber .................................................................................... 28

9HQWLODWRU



8.1

High Pressure Safety Valve .................................................................. 31

8.2

Negative Pressure Relief Valve ............................................................ 31

3QHXPDWLF6\VWHP



9.1

PEEP Pmax Valve Control .................................................................... 32

9.2

APL Bypass Valve Control .................................................................... 33

6020.010 Fabius GS

Contents

Page 3 of 5

&RQWHQWV


(OHFWULFDO%ORFN'LDJUDP





)XQFWLRQ'HVFULSWLRQ&RQWURO3&%



11.1 Flowmeter ............................................................................................. 35


11.2 Patient Interface .................................................................................... 35
11.3 Ventilator ............................................................................................... 35
11.4 Front Panel Functions ........................................................................... 35
11.5 Pneumatic Assembly Control ................................................................ 35
11.6 Serial Port Interface .............................................................................. 36
11.7 Battery .................................................................................................. 36


&RQWURO3DQHO$VVHPEO\





)L20HDVXUHPHQW





5HVSLUDWRU\)ORZ0HDVXUHPHQW





*DV)ORZ5DWH0HDVXUHPHQW





9DSRUL]HU



6SDUHDQG5HSODFHPHQW3DUWV

For internal use only. Copyright reserved.

K6020001IECIVZ.fm 01.02.02

6SDUHSDUWVOLVW



1.1

Trolley assembly 4116401 .................................................................... 44

1.2

Breathing system assembly w/APL bypass .......................................... 47

1.3

Mount, auxiliary assembly, fresh gas .................................................... 53

1.4

Fresh Gas Flow Meter [part of flow meter bezel assembly (3-gas) or (2gas)] ...................................................................................................... 54

1.5

Flow control valves and valve manifold


[part of flow meter bezel assembly (3-gas) or (2-gas)] ......................... 55

6020.010 Fabius GS

Contents

Page 4 of 5

&RQWHQWV
1.6

Fresh gas flow sensors and filter assembly


[part of flow meter bezel assembly (3-gas) or (2-gas)] ......................... 57

1.7

Pipeline pressure gauges


[part of flow meter bezel assembly (3-gas) or (2-gas)] ......................... 59

1.8

SORC
[part of flow meter bezel assembly (3-gas) or (2-gas)] ......................... 60

1.9

Fresh gas display PCB


[part of flow meter bezel assembly (3-gas) or (2-gas)] ......................... 61

1.10 Cylinder Yokes, Regulators, and Gauges: ............................................ 62


1.11 Controller PCB and related items (part of controller assembly ) ........... 66
1.12 Pneumatic (PEEP Control) Assembly ................................................... 68
1.13 Bezel Assembly, Monitor ...................................................................... 70
1.14 Ventilator ............................................................................................... 72

For internal use only. Copyright reserved.

K6020001IECIVZ.fm 01.02.02

1.15 Scavenger, AGS M33300 ..................................................................... 76

6020.010 Fabius GS

Contents

Page 5 of 5

*HQHUDO

This Technical Documentation/Service Manual conforms to the International Standard


IEC 60601-1.
Read each step in every procedure thoroughly before beginning any test. Always use
the proper tools and specified test equipment. If you deviate from the instructions
and/or recommendations in this Technical Documentation/Service Manual, the
equipment may operate improperly or unsafely, or the equipment could be damaged.
Use only original Drger parts and supplies.
The maintenance procedures described in this Technical Documentation may be
performed by qualified service personnel only. These maintenance procedures do not
replace inspections and servicing by Drger Medical AG & Co.KGaA.

6WULFWO\IROORZWKH,QVWUXFWLRQVIRU8VH2SHUDWLQJ,QVWUXFWLRQV
7KLV7HFKQLFDO'RFXPHQWDWLRQGRHVQRWUHSODFHWKH,QVWUXFWLRQV
IRU8VH2SHUDWLQJ,QVWUXFWLRQV$Q\XVHRIWKHSURGXFWUHTXLUHV
IXOOXQGHUVWDQGLQJDQGVWULFWREVHUYDWLRQRIWKHSURGXFWVSHFLILF
,QVWUXFWLRQVIRU8VH2SHUDWLQJ,QVWUXFWLRQV

For internal use only. Copyright reserved.

K6020001A01.fm 01.02.02

Unless otherwise stated, reference is made to laws, regulations or


standards (as amended) applicable in the Federal Republic of
Germany.

5664.900 Savina

Version 2.0

General

6\PEROVDQG'HILQLWLRQV
A dash () is used to identify items on a list.
A bullet () is used to identify a sentence that contains direct work instructions.

7KLVV\PEROLVXVHGWRSURYLGHLPSRUWDQWLQIRUPDWLRQWKDWLI
LJQRUHGFRXOGOHDGGLUHFWO\WRDSDWLHQWVRURSHUDWRUVLQMXU\,WLV
DOVRXVHGWRSURYLGHLPSRUWDQWLQIRUPDWLRQWKDWLILJQRUHGFRXOG
OHDGGLUHFWO\WRHTXLSPHQWGDPDJHDQGLQGLUHFWO\WRDSDWLHQWV
LQMXU\

This symbol is used to provide additional information, operating tips,


or maintenance suggestions.

Inspection

examination of actual condition

Servicing

measures to maintain specified condition

Repair

measures to restore specified condition

Maintenance

inspection, servicing, and repair

For internal use only. Copyright reserved.

K6020001A01.fm 01.02.02

Definitions according to German standard DIN 31051:

5664.900 Savina

Version 2.0

General

)XQFWLRQDO'HVFULSWLRQ

*HQHUDO,QIRUPDWLRQDERXWWKH)DELXV*6

The Fabius GS comprises the following assemblies:


Bezel assembly: Display and Control Panel
Flowmeter assembly
Gas Box: Gas Inlet Assembly and related items
Breathing system
Pneumatic Assembly
Ventilator
Anesthetic vaporizer(s)
Trolley

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

Monitoring, Electrical and Gas connections are illustrated on following pages.

6020.010 Fabius GS

Functional Description

Page 8 of 42

DISPLAY

AUXILIARY
OXYGEN
FLOWMETER

CONTROL PANEL
VENTILATOR
VAPORIZER

BREATHING
SYSTEM

CYLINDER
PRESSURE
GAUGES

O2 FLUSH

PIPELINE
PRESSURE
GAUGES

TROLLEY

FLOW
CONTROLS

TOTAL FRESH GAS


FLOWMETER

SV00543

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Front View of Fabius GS Anesthesia System

6020.010 Fabius GS

Functional Description

Page 9 of 42

SV00457

SERIAL COMMUNICATION
PORT
HOSE CONNECTION FOR
PEEP VALVE
HOSE CONNECTION FOR
APL BYPASS VALVE

O2 SENSOR
CORD CONNECTION

BREATHING PRESSURE
PILOT LINE CONNECTION
FLOW (VOLUME) SENSOR
CORD CONNECTION
AC POWER CONNECTION

ACCESS PANEL FOR MAINS


FUSES (2X 2.5A)

ON-OFF SWITCH

BATTERY FUSE

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Back View Showing Interface Panel and Power Entry

6020.010 Fabius GS

Functional Description

Page 10 of 42

SV00505

PIPELINE
HOSE
CONNECTIONS

O2 CYLINDER YOKE

N2O OR AIR
CYLINDER YOKE

O2 CYLINDER YOKE

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Back View Showing Gas Pipeline and Cylinder Yokes

6020.010 Fabius GS

Functional Description

Page 11 of 42

N2O CYLINDER CONNECTION


O2 CYLINDER CONNECTION

PIPELINE
HOSE
CONNECTIONS

SV00551

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Back View Showing Gas Pipeline and Cylinder Hose Connections (machines
without cylinder yokes)

6020.010 Fabius GS

Functional Description

Page 12 of 42

)DELXV*6)XQFWLRQ'LDJUDP

FRESH GAS

FRESH GAS
DECOUPLING
VALVE

INSPIRATORY
VALVE

AUXILIARY
AIR
VALVE

BREATHING
BAG

PAW

O2
LUNG

SAFETY
VALVE
VENTILATOR

P VENT
(AIR
ENTRAINMENT)

ABSORBER

VACUUM

EXPIRATORY
VALVE

PEEP
CONTROL
VALVE
*V

APL VALVE
MAN

SPONT/IPPV

ELECTRONIC
PEEP
VALVE

APL
BYPASS
VALVE
PISTON
PUMP
SV00499

OUTLET

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Function diagram of Fabius GS

6020.010 Fabius GS

Functional Description

Page 13 of 42

%DWWHU\%DFNXS

Fabius GS backup power is provided by two series-connected 12 V rechargeable


batteries. These batteries remain on charge as long as the machine is plugged into an
active AC outlet. Should power to the main controller PCB assembly fail while the
machine is in operation, the batteries will allow the machine to continue operating for
a minimum of 45 minutes, provided that the batteries are fully charged.
The batteries are located within the controller housing and are accessible by opening
the ventilator compartment. The 3.15A battery fuse is located on the power entry
panel at the back of the machine.

MAIN CONTROLLER PCB ASSEMBLY


+ 12 V

+ 12 V
J5

FUSE
3.15A

4.0A

BATTERY
MONITORING
AND CHARGING
POWER

J6

28 V POWER SUPPLY

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Battery Backup Arrangement

6020.010 Fabius GS

Functional Description

Page 14 of 42

6020.010 Fabius GS

N2O OR
AIR YOKE

O2 YOKE

O2 YOKE

O2 FLUSH

YOKE OPTION

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

VAPOR
MOUNT

N2O GAUGE

O2 GAUGE

FRESH GAS

GAS INLET ASM

TOTAL
FLOWMETER

O2 GAUGE

AIR GAUGE

N2O GAUGE

FLOW CONTROL
VALVE MANIFOLD

SORC

SORC BY-PASS ASM

FLOWMETER ASM

FLOW
SENSORS

SAFETY VALVE
(10 PSI)


)DELXV*63LSLQJ'LDJUDP

SV00454

)LJ  Fabius GS Piping Diagram

Functional Description

Page 15 of 42

)XQFWLRQ'HVFULSWLRQRI*DV%R[

The supply gases flow through the filters and non-return valves in the gas inlet
assembly. Pipeline supply pressures are indicated on gauges located on the
flowmeter assembly. Cylinder pressure gauges are located on the trolley assembly.
The pressures of O2 and N2O delivered to the flowmeter assembly are set by
regulators on the gas inlet assembly.
Should the O2 supply fail or if its pressure decrease below a certain limit, the O2
supply pressure alarm switch signals an alarm.

O2
FLUSH

O2 SUPPLY
PRESSURE
ALARM SWITCH

TO
FRESH GAS
OUTLET

P8

P7

O2

P6
P4

AIR

P5

TO
FLOW
METERS

N2O

PRESSURE
REGULATORS

O2

N2O
PRESSURE
GAUGES
P1

O2

P2

GAS INLET
ASSEMBLY

P3

P11
P10

AIR
P9
N2O
PIPELINE
SUPPLIES

FILTERS

NON-RETURN
VALVES

SV00489

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Gas Box Function Diagram, part 1

6020.010 Fabius GS

Functional Description

Page 16 of 42

If the O2 flush button is pressed, oxygen is delivered to the fresh gas outlet. The
fresh-gas ejector prevents the fresh gas from flowing back into the anesthetic
vaporizer. This avoids an increase in the anesthetic gas concentration.

O2 FLUSH
FRESH GAS

FLOWMETER
ASSEMBLY

TOTAL
FLOWMETER

(O2)
ANESTHETIC
VAPORIZER

FROM THE
GAS BOX (O2)

FROM THE
GAS BOX (AIR)

SAFETY VALVE
(10 PSI) 0.7 BAR

FROM THE
GAS BOX (N2O)

SV00490

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Gas Box Function Diagram, part 2

6020.010 Fabius GS

Functional Description

Page 17 of 42

625& 6HQVLWLYH2[\JHQ5DWLR&RQWUROOHU

The SORC is a control element that functions like an N2O shut-off device and ensures
a vital O2 concentration in the fresh gas. In the event of an O2 shortage, the SORC
limits the N2O flow such that the O2 concentration in the fresh gas does not decrease
below 21 vol.%. For the SORC to function properly O2 and N2O must have different
pressures.
If the O2 flow control valve is closed or if the O2 flow is lower than or equal to
200 mL/min, the SORC interrupts the N2O flow.
N2O can be added when the O2 flow is approx. 300 mL/min. In this case, the SORC
also prevents O2 concentrations below 21 vol.%.
The inlet valve vent (SORC bypass) allows O2 to bypass the restrictor in the SORC
when O2 flows above 10 L/min. are needed.

FRESH GAS
FROM
O2 FLUSH
FLOW
SENSOR

FLOWMETER
ASSEMBLY

TOTAL
FLOWMETER

FLOW
CONTROL
VALVE

ANESTHETIC
VAPORIZER

FROM THE
GAS BOX (O2)

SAFETY
VALVE
(10 PSI)0.7 BAR

FROM THE
GAS BOX (AIR)

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

INLET
VALVE,
VENT
S-ORC
FROM THE
GAS BOX (N2O)

SV00491

)LJ  SORC function diagram, part 1

6020.010 Fabius GS

Functional Description

Page 18 of 42

The flow control valves are used to adjust the O2 and N2O flows.
Restrictors located at the outlets of the SORC generate back-pressures. These backpressures exert a force on the control diaphragms of the SORC. The O2 backpressure opens the SORC. The N2O back-pressure closes the SORC. The pressure
ratio at the control diaphragm affects the N2O flow.
The restrictors and the spring tension are dimensioned such that a minimum
concentration of 21 vol.% O2 is always ensured. The maximum O2 flow is approx. 9
L/min.

O2

N2O

restrictors

control
diaphragms

N2O check valve

operating-point
adjusting screw
O2

N2O
flow control
valves

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  SORC function diagram, part 2

6020.010 Fabius GS

Functional Description

Page 19 of 42

&RPSDFW%UHDWKLQJ6\VWHP&RV\,,

The Cosy II compact breathing system allows three modes of patient ventilation:
manual and spontaneous breathing, volume controlled, and pressure controlled. The
APL valve (adjustable pressure limiting valve) has a selector which can be used to
toggle between "MAN" and "MAN/SPONT".
In the "MAN" position, the compact breathing system is closed to atmosphere. This
position is the default position for manual ventilation of the patient. The APL valve
opening pressure can be adjusted from 5 to 70 cmH2O.
In the "MAN/SPONT" position the APL valve is open to atmosphere. This position is
the default position for volume controlled ventilation and spontaneous breathing.
The pressure limit (Pmax) can also be adjusted during volume control from 15
cmH2O to 70 cmH2O using the control box and the PEEP Pmax valve.

10

1
3

9
A

7
11

6
4
5

12

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

16
15
13
14

)LJ  Compact breathing system, CosyII

6020.010 Fabius GS

Functional Description

Page 20 of 42

.H\
1

PEEP/Pmax valve

Expiratory valve

Flow sensor (Spirolog)

10

Expiratory port

MAN/SPONT-Volume Control selector

11

APL Bypass valve

(APL valve)

12

Fresh-gas port

Fresh-gas decoupling valve

13

Ventilator port

Breathing bag hook

14

Anesthetic gas scavenging port

Inspiratory port

15

Absorber

Inspiratory valve and O2 sensor

16

Pressure sensor connector

Breathing bag terminal and Y-piece plug



)XQFWLRQ'HVFULSWLRQ0DQXDO9HQWLODWLRQ



0DQXDO9HQWLODWLRQ*HQHUDO

During manual ventilation, the APL valve is set to the "MAN" position. The patient
system safety valve is activated. The piston of the ventilator is in the upper end
position in order to reduce the dead space volume of the ventilator.


0DQXDO9HQWLODWLRQ,QVSLUDWLRQ

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

During inspiration, expiratory valve remains closed. When the clinician compresses
the breathing bag the gas mixture (expiratory gas and fresh gas) flows through the
fresh-gas decoupling valve, the inspiratory valve, the O2 sensor, the inspiratory hose,
and the Y-piece into the patients lung. The pressure sensor measures the airway
pressure. The APL valve limits the ventilation pressure. Any excess amount of the
gas mixture flows through the APL valve and the non-return valve to the anesthetic
gas scavenging system.

6020.010 Fabius GS

Functional Description

Page 21 of 42

OXYGEN SENSOR
PRESSURE SENSOR
INSPIRATORY VALVE
OPEN

FLOW SENSOR
PEEP/PMAX
VALVE OPEN

VENTILATOR
INACTIVE

FRESH GAS
DECOUPLER
OPEN

FRESH GAS INLET

EXPIRATORY VALVE
CLOSED
APL BYPASS VALVE
CLOSED
APL VALVE
ADJUSTED/OPEN

BREATHING BAG
ABSORBER

EXHAUST VALVE
CLOSED

)LJ  Manual ventilation (inspiration)




0DQXDO9HQWLODWLRQ([SLUDWLRQ

During expiration, the inspiratory valve remains closed and thus prevents the
expiratory gas from flowing back into the inspiratory branch.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

After releasing the breathing bag, the expiratory gas from the lung flows through the
expiratory hose, the flow sensor, the PEEP Pmax valve, the expiratory valve and
through the absorberinto the breathing bag. At the same time, new fresh gas flows
into the breathing bag.

6020.010 Fabius GS

Functional Description

Page 22 of 42

OXYGEN SENSOR
PRESSURE SENSOR
INSPIRATORY VALVE
OPEN

FLOW SENSOR

PEEP/PMAX
VALVE VARIABLE
VENTILATOR
INACTIVE

FRESH GAS
DECOUPLER
OPEN

FRESH GAS INLET

EXPIRATORY VALVE
OPEN
APL BYPASS VALVE
CLOSED
APL VALVE
ADJUSTED/OPEN

BREATHING BAG

ABSORBER

EXHAUST
VALVE OPENS
AS REQUIRED

)LJ  Manual ventilation (expiration)



)XQFWLRQ'HVFULSWLRQ6SRQWDQHRXV%UHDWKLQJ



6SRQWDQHRXV%UHDWKLQJ*HQHUDO

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

A prerequisite for spontaneous breathing is that the patient is supplied with a sufficient
amount of fresh gas. The APL valve selector must be set to the "SPONT" position. No
pressure builds up in the compact breathing system.


6SRQWDQHRXV%UHDWKLQJ,QVSLUDWLRQ

During inspiration, the expiratory valve remains closed thus preventing rebreathing of
expiratory gas containing CO2.

6020.010 Fabius GS

Functional Description

Page 23 of 42

The patient inhales the gas mixture (expiratory gas and fresh gas) from the breathing
bag. The gas mixture flows through the fresh-gas decoupling valve, the inspiratory
valve, the O2 sensor, the inspiratory hose, and through the Y-piece into the lung. The
pressure sensor measures the airway pressure.

OXYGEN SENSOR

PRESSURE SENSOR
FLOW SENSOR

INSPIRATORY VALVE
OPEN
VENTILATOR
INACTIVE

FRESH GAS INLET

PEEP/PMAX
VALVE OPEN
FRESH GAS
DECOUPLER
OPEN

EXPIRATORY VALVE
CLOSED
APL BYPASS VALVE
CLOSED

ABSORBER
APL VALVE
ADJUSTED/OPEN

BREATHING BAG

EXHAUST
VALVE CLOSED

)LJ  Spontaneous breathing (inspiration)




6SRQWDQHRXV%UHDWKLQJ([SLUDWLRQ

During expiration, the inspiratory valve remains closed thus preventing the expiratory
gas from flowing back into the inspiratory branch.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

The APL valve is open, regardless of its pressure setting.


The expiratory gas flows from the lung through the expiratory hose, the flow sensor,
the PEEP control valve, the expiratory valve, and through the absorber into the
breathing bag. At the same time, new fresh gas flows into the breathing bag.
When the breathing bag is full, any excess gas mixture flows through the non-return
valve into the anesthetic gas scavenging system.

6020.010 Fabius GS

Functional Description

Page 24 of 42

OXYGEN SENSOR
PRESSURE SENSOR
FLOW SENSOR
INSPIRATORY VALVE
OPEN
PEEP/PMAX
VARIABLE
VENTILATOR
INACTIVE
FRESH GAS INLET

FRESH GAS
DECOUPLER
OPEN

EXPIRATORY VALVE
OPEN
APL BYPASS VALVE
CLOSED

ABSORBER
APL VALVE
ADJUSTED/OPEN

EXHAUST
VALVE OPENS
AS REQUIRED

BREATHING BAG

)LJ  Spontaneous breathing (expiration)

The CO2 is scrubbed from the expiratory gas by the soda lime contained in the
absorber. The fresh gas replaces the anesthetic and oxygen taken up by the patient.



)XQFWLRQ'HVFULSWLRQ9ROXPH0RGH9HQWLODWLRQ



9ROXPH&RQWURO0RGH*HQHUDO

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

A prerequisite for volume control is that the patient is supplied with a sufficient amount
of fresh gas.
The APL bypass valve opens in volume mode, allowing excess gas to be vented to
the scavenging system regardless of the MAN-SPONT valve setting.
The safety valve of the patient system makes sure that no pressures greater than 75
cmH2O build up in the system. The pressure limit (Pmax) can be adjusted on the
control box.

6020.010 Fabius GS

Functional Description

Page 25 of 42



9ROXPH&RQWURO0RGH,QVSLUDWLRQ

During inspiration, the PEEP Pmax valve remains closed. The control pressure
present at the PEEP Pmax valve varies with the set pressure limit (Pmax).
The pressure generated by the ventilators piston closes the fresh-gas decoupling
valve. The gas mixture (expiratory gas and fresh gas) flows through the inspiratory
valve, the O2 sensor, the inspiratory hose, and through the Y-piece into the lung. The
pressure sensor measures the airway pressure. The ventilation pressure cannot
exceed the pressure limit (Pmax) set on the control box because the PEEP Pmax
valve opens. The fresh gas then fills the breathing bag.
Any excess fresh gas flows through the open APL bypass valve, and through the nonreturn valve into the anesthetic gas scavenging system.

OXYGEN SENSOR

PRESSURE SENSOR
INSPIRATORY VALVE

FLOW SENSOR

PEEP/PMAX
VALVE
VENTILATOR

FRESH GAS
DECOUPLER
CLOSED
EXPIRATORY VALVE

FRESH GAS INLET

APL BYPASS VALVE

APL VALVE

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

BREATHING BAG

ABSORBER

EXHAUST
VALVE

)LJ  Volume control ventilation (inspiration)

6020.010 Fabius GS

Functional Description

Page 26 of 42



9ROXPH&RQWURO0RGH([SLUDWLRQ

During expiration, the inspiratory valve remains closed thus preventing rebreathing
into the inspiratory branch.
The expiratory gas from the lung flows through the expiratory hose, the flow sensor,
the PEEP/Pmax valve, the expiratory valve, and through the absorber back into the
breathing bag mixing with fresh gas also flowing into the breathing bag.
The ventilators piston moves back drawing the gas mixture needed for the next
inspiration into the piston space.
Any excess fresh-gas flows through the APL bypass valve, and through the nonreturn valveinto the anesthetic gas scavenging system

OXYGEN SENSOR
PRESSURE SENSOR
INSPIRATORY VALVE
CLOSED
VENTILATOR
PISTON
RETURNS TO
START
POSITION

FLOW SENSOR

PEEP/PMAX
VALVE OPEN
FRESH GAS
DECOUPLER
OPEN
EXPIRATORY VALVE
OPEN

FRESH GAS INLET

APL BYPASS VALVE


OPEN

APL VALVE

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

BREATHING BAG
FILLS

ABSORBER

EXHAUST
VALVE 2MBAR
RESISTANCE
CLOSED

)LJ  Volume control ventilation (expiration)

6020.010 Fabius GS

Functional Description

Page 27 of 42



&RV\,,$EVRUEHU

The absorber canister is filled with fresh soda lime. The soda lime scrubs CO2 from
the respiratory expired gas.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

Expired soda lime changes its color. The soda lime must be replaced when two thirds
of the soda lime in the absorber canister is discolored.

6020.010 Fabius GS

Functional Description

Page 28 of 42

9HQWLODWRU

The ventilator is located in a swing-out compartment at the left side of the Fabius GS.
A hose terminal is provided on the left side of the compartment for connection to the
breathing system. Fresh gas is delivered to the patient by a piston that is driven by a
motor and ball-screw arrangement. A sight window on the compartment allows the
operator to verify movement of the piston.
Two diaphragms (upper and lower) comprise a bag-type rolling seal that surrounds
the piston. Vacuum from the pneumatic assembly (described in a later paragraph) is
provided between the outside of the seal and the cylinder, to ensure proper operation
of the seal during piston movement.
During inspiration the ventilator delivers fresh gas at a given volume, pressure and
frequency. These parameters are set at the control panel. Refer to the Operators
Manual for details on ventilator settings, displays and controls. During expiration, the
bag-type rolling seal fills with expired gas from the patient and with fresh gas stored in
the breathing bag.
Power for the ventilator motor is distributed from the control PCB. A home sensor on
the ventilator signals the control PCB when the piston reaches its lower limit. An
incremental encoder on the motor shaft determines the number of revolutions and
provides piston travel information to the control PCB.
Ventilator pressure is monitored by a transducer on the control PCB. Should the
negative pressure relief valve on the patient assembly open, a Fresh Gas Low alarm
is displayed.
The ventilator pressure transducer is the same type as the one used for measuring
airway pressure. A hose connects the transducers positive pressure port to a hose
barb located on the top cover of the ventilator. The purpose of this transducer is to
allow the software to sense when a condition exists that would cause the ventilator
negative pressure relief valve to open. The threshold that is used by the software for
this condition is -3 mbar. In normal use the primary cause for this condition is an
insufficient amount of reserve gas in the breathing bag. The operator is alerted when
this condition exists, with a medium priority FRESH GAS LOW alarm. A future
software release after 1.10 may add an alarm based on transducer diagnostics.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

The ventilator assembly is illustrated on the next page.

6020.010 Fabius GS

Functional Description

Page 29 of 42

NEGATIVE PRESSURE
RELIEF VALVE
VENT CHAMBER
PRESSURE
SENSOR LINE

HIGH PRESSURE
VALVE

VACUUM LINE
TO PNEUMATIC
ASSEMBLY

UPPER
DIAPHRAGM
CYLINDER
PISTON
LOWER
DIAPHRAGM

MOTOR AND
BALLSCREW
ASSEMBLY

INCREMENTAL ENCODER

SV00513

)LJ  Ventilator (piston shown in down position)

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

The top of the ventilator assembly (patient system) contains two valves:

6020.010 Fabius GS

Functional Description

Page 30 of 42



+LJK3UHVVXUH6DIHW\9DOYH

If the pressure limit control fails, the patient system high pressure safety valve limits
the gas pressure. This valve is set to open at approximately 75 cmH2O.

screw

spring

valve disc
washer

)LJ  Sectional View of the Safety Valve



1HJDWLYH3UHVVXUH5HOLHI9DOYH

The negative pressure relief valve allows the patient to spontaneously breathe
ambient air should the medical gas supply and/or Fabius GS fail. The opening
pressure of this valve is set at -3 cmH2O.

threaded ring
valve seat

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

valve disc

valve cross with spring

)LJ  Sectional View of the Negative Pressure Relief Valve

6020.010 Fabius GS

Functional Description

Page 31 of 42

3QHXPDWLF6\VWHP

The pneumatic assembly provides pressure for the PEEP valve control, and also
provides vacuum for the ventilator bag-type rolling seals and the APL bypass valve
control.

TO UPPER PORT OF
LS1 ON PCB ASM
TO UPPER PORT OF
LS3 ON PCB ASM

PNEUMATIC ASSEMBLY
VARIABLE
RESTRICTOR

VENTILATOR CHAMBER
PRESSURE

FILTER

VENTILATOR

V1
V2

ELECTRONIC
PEEP
VALVE

PUMP
RESERVOIR

APL BYPASS
VALVE
PEEP VALVE

INTERFACE
PANEL

CONTROL SIGNAL
FROM J7 ON PCB ASM

BREATHING
SYSTEM

)LJ  Pneumatic Control System Schematic

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02



3((33PD[9DOYH&RQWURO

When the Fabius GS is operating in the automatic mode, the pump on the pneumatic
assembly is running, and the electronic PEEP valve receives a signal from the main
control PCB. The amount of current supplied to the coil of the electronic PEEP valve
is proportional to the PEEP value set by the operator, and controls the position of the
diaphragm within the electronic PEEP valve. This then determines the control
pressure applied to the proportional PEEP valve in the breathing system, which
maintains the desired amount of PEEP during patient expiration. The V1 reservoir
smooths out pressure variations caused by the pump. See Figure 21.

6020.010 Fabius GS

Functional Description

Page 32 of 42



$3/%\SDVV9DOYH&RQWURO

When the Fabius GS is operating in the automatic mode, the pneumatic assembly
provides a vacuum signal to hold open the APL bypass valve in the breathing system.
The V2 reservoir and filter provide noise damping, and the variable restrictor is used
to set the vacuum level in the range of -150 to -240 cmH2O.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

When the machine is operating in the Manual mode, the pump on the pneumatic
assembly (and the ventilator) is stopped, and the spring-loaded APL bypass valve in
the breathing system closes, directing exhaled gas through the APL valve.

6020.010 Fabius GS

Functional Description

Page 33 of 42



(OHFWULFDO%ORFN'LDJUDP

GAS INLET ASM

FLOW METER BEZEL ASM

P3

P2

FRESH J3
GAS
DISPLA
J1

O2

AIR

N2O

P4

PNEUMATIC ASM

J1

J16

J1

J7

PEEP
VALVE

LOW
O2

J1

SERIAL PORT
INTERFACE
O2 SENSOR

J8
J1

PUMP
MAIN CONTROLLER
PCB ASSEMBLY

FRONT
PANEL
DISPLAY

J2

J3
J4

J5

J21

J6

TABLE
LAMP

MAINS

TACH
SENSOR

J1

HOME
SENSOR

F4
F1

VENTILATOR

BOOS
T

ON/OF
ROTARY
ENCODER
BATT

MOTOR

J11
J9

J1
KEYPAD
INTERFACE
J2

J1

SPIROLOG
VOL SENSOR

J3
J1

POWER
SUPPLY

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

PWR ENTRY

)LJ  Electrical Block Diagram

6020.010 Fabius GS

Functional Description

Page 34 of 42



)XQFWLRQ'HVFULSWLRQ&RQWURO3&%

The control PCB provides the following functions in the Fabius GS:



)ORZPHWHU

Receives flow rate data from the N2O, Air, and O2 flow sensors and provides fresh
gas display information for these gases.



3DWLHQW,QWHUIDFH

Receives information from the oxygen sensor and processes it for display.
Receives flow information from the Spirolog sensor and processes it for display.
Converts airway breathing pressure to an electrical signal and processes it for display.



9HQWLODWRU

Provides ventilator motor drive, receives ventilator piston position and movement
information. Provides ventilator pressure information.



)URQW3DQHO)XQFWLRQV

Provides power for the display panel and LED table lamp.
Provides video signals to the display.
Receives information from the keypad and rotary encoder for making display and
operating selections, setting alarm limits, and service functions.



3QHXPDWLF$VVHPEO\&RQWURO

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

Provides power to the pump on the pneumatic assembly.


Provides control signal to the electronic PEEP valve on the pneumatic assembly in
response to operator setting.
Monitors pump vacuum to stabilize PEEP control.

6020.010 Fabius GS

Functional Description

Page 35 of 42



6HULDO3RUW,QWHUIDFH

Provides isolated port for connecting external monitors and downloading software.



%DWWHU\

Provides charging current for battery and monitors state of battery charge.

LATCH

BUFFER

FLASH
MEM

XCVR

FRESH GAS DISPLAY


FRESH GAS
FLOW SENSORS
DISPLAY
PANEL

VIDEO
CONTROL

RAM
O2 SENSOR
MICROCONTROLLER

SPKR
VOL
SENSOR

VACUUM

LS1

AIRWAY
PRESS

LS2

VENT
PRESS

LS3

PWR IN
BATT

DRV

LEDs

Q15

TABLE LAMP

KEYPAD & KNOB

PROG
LOGIC

VENTILATOR TACH

VENT
MOTOR
DRIVE

VENTILATOR MOTOR

PWR
MGMT

VENTILATOR HOME SENS

+28V
OPTO
ISO

DRIVER
RCVR

SERIAL
PORT

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Controller Functional Block Diagram

6020.010 Fabius GS

Functional Description

Page 36 of 42



&RQWURO3DQHO$VVHPEO\

The control panel assembly consists of a 320 x 240 pixel graphical display, a table
lamp with six LEDs, a membrane keypad, rotary encoder and speaker.
Data and power for the display comes from the main controller PCB via a 20conductor ribbon cable. The keypad interface is connected to the main controller PCB
by a 30-conductor ribbon cable. A block diagram of the control panel assembly is
shown in the following illustration.

TO J2 ON MAIN CONTROL PCB

TO J9 ON MAIN CONTROL PCB

GRAPHICAL
DISPLAY

KEYPAD & LEDs


(USER INTERFACE)
SPEAKER
J3

J2
TABLE LAMP LEDs

ROTARY
ENCODER

)LJ  Control Panel Block diagram

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

An illustration of the control panel is shown on the next page.

6020.010 Fabius GS

Functional Description

Page 37 of 42

Descriptions of the numbered items in the following illustration are given in the table
below.

25%10:50 AM

Service Log

Preventive
Maintenance

General

Entries
Real Time Values

Activate

Serial Number
Reset Hours Run
Reset Last Ser-

Calibration

Flowmeter

Serial
Port

Fresh Gas Flow

O2 Position

Parameters

vice Date

Service













For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Fabius GS Control Panel (Main Service Screen Illustrated)

Item

Function

Selects volume controlled ventilation mode. Refer to Operators Manual

Selects pressure controlled ventilation mode. Refer to Operators Manual

Reserved for future use

Controls table lamp: Off/On

Places ventilator in Man Spont mode. Refer to Operators Manual

Soft keys: activate the corresponding function that appears on screen above the
key

For setting alarm limits. Refer to Operators Manual

Setup key: activates sub-screens for monitoring functions. Refer to Operators


Manual

Home key: returns display to main screen shown before standby

10

Rotary control: moves the cursor on the screen; confirms selection when pressed

11

Alarm Status indicators:


Flashing Red: Warning; Flashing Yellow: Caution; Solid Yellow: Advisory

12

Alarm Silence key: silences all active alarms for two minutes

13

Power ON indicator: lighted when machine is plugged into an active AC outlet

14

Returns display to Standby screen

6020.010 Fabius GS

Functional Description

Page 38 of 42



)L20HDVXUHPHQW

The O2 sensor measures the fraction of inspired O2 (FiO2) in the respiratory gas.
The O2 sensor contains an alkaline electrolyte, a lead anode, two gold cathodes, and
a Teflon membrane. The spatial separation of the two gold cathodes allows a voltage
comparison to be made as explained below.
The O2 sensor is an electrochemical cell that generates a voltage from the ion
current.

TEMPERATURE
COMPENSATING
RESISTORS

LEAD ANODE

ALKALAI
ELECTROLYTE

GOLD CATHODE A
GOLD CATHODE B
TEFLON MEMBRANE

)LJ  O2 sensor

The O2 to be measured diffuses through the Teflon membrane, reacts at the gold
cathodes (negative polarity) and forms lead oxide and water at the lead anode
(positive polarity). During this chemical process, a voltage is generated which is
proportional to the O2 partial pressure.

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

The internal resistance of the cell is determined by the surface area of the gold
cathodes, the O2 diffusion velocity, the distance between the gold cathodes and the
lead anode. This resistance is approximately 700 ohms.
The chemical process is temperature-sensitive. Therefore temperature-sensitive
resistors are connected in parallel with the O2 sensor. These resistors and the internal
resistance of the O2 sensor correct the measuring voltage. Because two cathodes
used in the O2 sensor cell, two different voltages are generated. These voltages are
compared with each other. If their difference exceeds a certain value, the machine
prompts the operator to check the cell.
If the O2 sensor fails, the control box will indicate an error on the graphics display.

6020.010 Fabius GS

Functional Description

Page 39 of 42



5HVSLUDWRU\)ORZ0HDVXUHPHQW

The flow sensor functions according to the constant temperature hot-wire


anemometer principle. Respiratory gas flows past a thin platinum wire. This platinum
wire (A) is located in a measuring tube and is electrically heated. The platinum wire is
held at a constant temperature. Gas flow removes heat from the hot wire. The higher
the gas flow rate, the greater the heat removal. The amount of electrical current
needed to maintain a constant platinum wire temperature is thus proportional to the
gas flow rate.
A second platinum wire (B) in the measuring tube is used to compensate for
interferences from different gases present in the respiratory gas. The heat removed
from the second platinum wire is measured during inspiration when the gas flow is
zero.
The different gases present in the respiratory gas have a different thermal
conductivity. The amount of heat removed from the second platinum wire is thus an
indicator of respiratory gas composition.
Internal calibration tables for O2/N2O mixtures, Air and 100% O2 are used to linearize
the measured flow.

Pla tinum Wire A

Pla tinum Wire B

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

)LJ  Respiratory Flow Sensor

6020.010 Fabius GS

Functional Description

Page 40 of 42



*DV)ORZ5DWH0HDVXUHPHQW

The gas flow sensors operate on the principle of specific heat for individual gases. In
each sensor, as the gas flows through a heated chamber the gas molecules carry
away a certain amount of heat relative to the specific heat index for that gas.
A known amount of electrical current is required to maintain the temperature in the
heated chamber. The higher the gas flow rate, the more heat is removed from the
chamber and more current is required to maintain the temperature in the chamber.
This current is then scaled and displayed as liters per minute flow rate for each gas.

ELECTRICAL
CONNECTION

TUBING
CONNECTOR

ELECTRONIC
COMPONENTS

GAS OUTLET
PORT
(TO MANIFOLD)

GAS INLET
ASSEMBLY

HEATED
CHAMBER

MOUNTING POSTS

)LJ  Flow Sensor Details

FROM NITROUS
OXIDE FLOW
CONTROL VALVE

TOTAL FRESH GAS FLOWMETER

FROM AIR FLOW


CONTROL VALVE

FRESH GAS
MANIFOLD

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

FROM OXYGEN
FLOW CONTROL
VALVE

)LJ  Flow of Gases Through Sensors

6020.010 Fabius GS

Functional Description

Page 41 of 42



9DSRUL]HU

For internal use only. Copyright reserved.

F6020001T01.fm 01.02.02

Refer to separate technical documentation of the anesthetic vaporizer.

6020.010 Fabius GS

Functional Description

Page 42 of 42

6SDUHDQG5HSODFHPHQW3DUWV

6SDUHSDUWVOLVW

Part numbers for field-replaceable items on the Fabius GS anesthesia system are
listed on the following pages, along with part numbers for related hardware and
cables.

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

The item numbers are keyed to the accompanying illustrations to aid in identifying the
item and its location.

6020.010 Fabius GS

Spare and Replacement Parts

Page 43 of 76



7UROOH\DVVHPEO\
SV00522

19

20
22
21
13

18

16

12

11

15

17
5

6
7
8

14
13
3

10

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.

Fresh gas hose assembly


Connector cone, mix gas
O-ring
Pressure nozzle
O-ring, 6mm ID x 1mm W
Connector lump
Angle connector
Straight fitting, 1/4 tube x M12
Screw, M4 x 0.7 x 16 cap skt hd (2x)
Lock washer, M4 split (2x)
Valve, 2/2 way

6020.010 Fabius GS

Spare and Replacement Parts

4117046
E20278
4117167
2600503
M30953

1293680

Page 44 of 76

12.
13.
14.
15.

16.
17.
18.
19.
20.
21.
22.

Straight connector, M5
Washer, M4 white
Angle connector
Label, O2 flush:
Germany, Austria, France, Spain
USA
Cabinet/frame assembly
Caster, locking
Caster
Drawer lock
Drawer (top) includes clips
Drawer (middle & bottom) includes clips (2x)
Drawer Slide (6x)

M30952
M31602
M30953

M28934
4117033
8411945
8411946
4117035
AF00007
4117008

1RWVKRZQ
Auxiliary outlet strip, USA/CSA
Alternate part number
Auxiliary outlet strip, France
Alternate part number

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

Auxiliary outlet strip, Spain


Alternate part number

6020.010 Fabius GS

Spare and Replacement Parts

Page 45 of 76

1
4

3
7

5
6

8
2

SV00558

1.
2.

8403735
4199951
4199950
6850645

4106388

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

3.
4.
5.
6.
7.
8.
9.

Spirolog flow sensor, 5 pack


Spirolog cable assembly, COSY left
Spirolog cable assembly, COSY right
Oxygen Sensor, complete assembly
Sensor capsule
Housing assembly, less capsule
Adapter (cover)
O-ring, #025 (Viton)
O-ring, #016 (Viton) (2x)
Fresh gas hose assembly

6020.010 Fabius GS

Spare and Replacement Parts

Page 46 of 76



%UHDWKLQJV\VWHPDVVHPEO\Z$3/E\SDVV

6
1

FRESH GAS
DECOUPLING VALVE
PEEP/PMAX VALVE

10

APL BYPASSVALVE

4
11

3
12

13
5
14

SV00508

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

Breathing system assembly w/APL bypass, Service Exchange


Breathing system assembly w/APL bypass and sample gas return fitting
1.
2.
3.
4.
5.
6.
7.
8.
9.

Dome, insp valve


Plug assembly, O2 sensor
Plunger
Breathing system
Scavenger port assembly
APL valve assembly
Retaining nut
Union nut (2x)
Packing ring (2x)

6020.010 Fabius GS

M30448
M30453
M09231

Spare and Replacement Parts

Page 47 of 76

10. Valve disk (2x)


11. Exp. hose terminal (incl. O-ring)
12. Bag mount port
O-ring
13. Insp. hose terminal
O-ring
14. Absorber top & canister

M23225
8603275
8603276
M20622
M20622
See parts lists on subsequent pages

Fresh gas decoupling valve:


Cap
O-ring, cap
Spring Cross
Diaphragm, Viton
Crater

2600734
M17450
2600737

PEEP/Pmax valve:
Cap & hose barb assembly
O-ring
Spacer
Diaphragm asm
Crater
Sealing ring

M30384
2600731
2600479

4116929

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

APL bypass valve:


Cap & hose barb assembly
Spring
Spacer, spring holder
Crater assembly

6020.010 Fabius GS

Spare and Replacement Parts

Page 48 of 76

4
3

SV00418

6
7

Absorber top
Lip seal
O-ring
O-ring
O-ring (3x)
Washer (9x)
Screw (3x)
Washer (3x)

M30455
R50313
R18352
4302469
2600459
1266225
1329472

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.
8.

6020.010 Fabius GS

Spare and Replacement Parts

Page 49 of 76

6
4
5
3

3
SV00419

Screen insert
Canister
Packing ring
Bead chain
Screw, oval hd AM4 x 6 DIN 85 - A4
Acorn nut, M4 DIN 1587 - M A4/051
Screw, M4 x 8 DIN 921 - A2

M29999
M29994
M30563
M30562
1334751
1336029

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.

6020.010 Fabius GS

Spare and Replacement Parts

Page 50 of 76

E6020001T01.fm 01.02.02

For internal use only. Copyright reserved.

6020.010 Fabius GS
Spare and Replacement Parts

Page 51 of 76

1.

Breathing system mount assembly

4101872

1190520

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

Gauge assembly, patient airway pressure


2. Pressure gauge assembly, NMM
3. O-ring, #010 (neoprene)
4. Gauge mount adapter
5. Hose barb coupling, 90 deg
6. T-fitting, 5/32 ID tube
7. Hose, 4 x 1.5 SI NF clear
8. Screw, 8-32 x 1/2 in. btn hd skt
9. Lock Washer, #8 int-t
10. Side Panel Option, left

6020.010 Fabius GS

Spare and Replacement Parts

Page 52 of 76



0RXQWDX[LOLDU\DVVHPEO\IUHVKJDV
4

Outlet, aux fresh gas


Bracket, aux fresh gas
Slide plate, GCX black
Screw, M3 x 0.5 x 10 MM, btn hd skt (2x)
Screw, 1/4-20 x 7/8 in. btn hd skt (2x)
Nut, acorn, 1/4-20 (2x)
Lock washer, 1/4 split (2x)
Knob, thumb screw, GCX

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.
8.

SV00557

6020.010 Fabius GS

Spare and Replacement Parts

Page 53 of 76



)UHVK*DV)ORZ0HWHU>SDUWRIIORZPHWHUEH]HODVVHPEO\ 
JDV RU JDV @

SV00477

1
1

1
2

Flowmeter tube assembly (incl. brackets and mounting nuts)


Luer fitting
Collar, 6 mm white

M30960
M31603

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.

6020.010 Fabius GS

Spare and Replacement Parts

Page 54 of 76



)ORZFRQWUROYDOYHVDQGYDOYHPDQLIROG
>SDUWRIIORZPHWHUEH]HODVVHPEO\ JDV RU JDV @

8
5
3

9
7

13
13
16

14

14

17

16

15

14

15

12

18
15

10

19

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

SV00479

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.

17

11

Flow control valve (3x)


Spacer (3x)
Collet (3x)
Chuck (3x)
Knob, N2O & Air (2x)
Knob, O2
Washer, DIN 1751 (3x)
Spring washer (3x)
Nut (3x)
Manifold
Screw, M5 x 0.8 x 8 mm btn hd
Screw, M5 x 25 cap skt hd (2x)
Washer, 4 mm blue
Angle connector
Straight connector, M5
Ferrule, pilot to Divan
Washer, M4 white

6020.010 Fabius GS

M24452
M24900

M30937
M30953
M30952
M30963
M31602

Spare and Replacement Parts

Page 55 of 76

18. Lock washer, M5 split (2x)


19. Flat washer, M5 (2x)
.QREFDSVSDUWQXPEHUV
O2

Air

N2O

Color

P/N

Color

P/N

Color

P/N

Germany,
Austria

Black

M34305

Black

M34307

Black

M34306

USA

Green

M25147

Yellow

M25797

Blue

M24901

France,
Spain,
UK/Ireland

White cover
screw

Blk/wh

M26205

Blue

M24901

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

Country

6020.010 Fabius GS

Spare and Replacement Parts

Page 56 of 76



For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.

)UHVKJDVIORZVHQVRUVDQGILOWHUDVVHPEO\
>SDUWRIIORZPHWHUEH]HODVVHPEO\ JDV RU JDV @

Flow sensor bracket


Flow sensor assembly (3x)
Cable assembly, gas flow sensor
Safety valve
O-ring, #113 (silicone)
Filter assembly
Flow sensor manifold, 3-gas
Flow sensor manifold, 2-gas
O-ring, #106 (neoprene) (3x)
Collar, 6 mm white
Angle connector, M5
Kep nut, M5 (2x)
Angle adapter, 1/4 MPT to M5 (3x)
Straight connector, M5
Washer, M4 white
Y-plug
Ferrule, Pilot to Divan
Washer, 4 mm blue
Kep nut, M4 x 0.5 (2x)
Input valve, complete

6020.010 Fabius GS

Spare and Replacement Parts

M16201

4109322
M31603
M30961

M30952
M31602
M30962
M30963
M30937

Page 57 of 76

1340778

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

20. Screw, M4 x 8 (2x)


21. Screw, M5 x 10 pan hd (2x)

6020.010 Fabius GS

Spare and Replacement Parts

Page 58 of 76



3LSHOLQHSUHVVXUHJDXJHV
>SDUWRIIORZPHWHUEH]HODVVHPEO\ JDV RU JDV @

1
SV00480

2
4

Plate, pipeline gauge


Pipeline pressure gauge (3x)
Screw, M4 x 12 mm self tap (4x)
Washer, 4 mm blue
Straight connector, M5
Ferrule, pilot to Divan
Washer, M4 white

4117021
M30937
M30952
M30963
M31602

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.

6020.010 Fabius GS

Spare and Replacement Parts

Page 59 of 76



625&
>SDUWRIIORZPHWHUEH]HODVVHPEO\ JDV RU JDV @

SORC

3
7

4
2

5
6
4
5

SV00456

Kep nut, M5 (4x)


SORC (proportional valve)
Resistor, O2 red
O-ring
O-ring, DIN 3771
Resistor, N2O blue
Filter (inside unit)

M32940
M27543
E20274
M30308
M27539

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.

6020.010 Fabius GS

Spare and Replacement Parts

Page 60 of 76



)UHVKJDVGLVSOD\3&%
>SDUWRIIORZPHWHUEH]HODVVHPEO\ JDV RU JDV @

SV00482

1
3

4
5

Insulator, flowmeter display


PCB assembly, fresh gas display
Cable assembly
Cable assembly, O2 pressure
Screw, M3 x 5 (4x)
Lock washer, M3 int-t (4x)

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.

6020.010 Fabius GS

Spare and Replacement Parts

Page 61 of 76



&\OLQGHU<RNHV5HJXODWRUVDQG*DXJHV

21

20

17

22

18

O2

O2

19

12

O2

14
15

9
13

14

11

13

10

16

1
2

SV00470

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.

2.
3.

O2 yoke (2x)1101620
N2O yoke1101621
Air yoke1101625
Index pin - screw, sltd 0.157 OD 6-32 x 0.718 L (2x per yoke)
Label, 3/8 DOT O2
Label, 3/8 DOT N2O
Label, 3/8 DOT Air

6020.010 Fabius GS

Spare and Replacement Parts

Page 62 of 76

Screw, 10-32 x 3/8 rd hd


Plug assembly
Yoke handle
Screw, 5/16-24 x 3/4 cap skt hd (2x per yoke)
Lock washer, 5/16 split (2x per yoke)
Check valve assembly (2x)
Check valve assembly
O2 cylinder pressure regulator, 36 psi
N2O cylinder pressure regulator, 36 psi
Straight fitting, 1/4 MPT x 1/8 FPT (2x)
Check valve (2x)
Washer, 4 mm white
Washer, 4 mm blue
T-fitting, 3/16 tube x 3/16 tube x 1/8 MPT
Nut, 3/16 tube
L-fitting, 3/16 tube x 1/4 MPT
Cylinder pressure gauge, 3000 psi, NMM (2x)
Kep nut, 6-32 (2x per gauge)
Straight fitting, 3/16 tube x 1/8 FT (2x)

M30946
M31602
M30937

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.

6020.010 Fabius GS

Spare and Replacement Parts

Page 63 of 76

SV00462

10

11
12

9
13
17
15

20

19

18

1
14

16

2
8
5

6
23

24
3
21

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

22

6020.010 Fabius GS

Spare and Replacement Parts

Page 64 of 76

Gas Inlet Assembly, 3-connector DISS


Gas Inlet Assembly, 5-connector NIST
1.
2.
3.
4.
5.
6.
7.

4116537
4116946
M30935
M30953
M28816
M31348
M30963
4117143
M31602
M30937
4116942
4102532
4116980
4101882
HW01127

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.

Back panel (5 conn plate)


Spacer (plate, cylinder block)
Screw, M5 x 0.8 x 8 btn hd skt (4x)
Lock washer, #10 split (4x)
Connector, NIST N2O male
Connector, DISS N2O
Connector, NIST Air male
Connector, DISS Air
Connector, NIST O2 male
Connector, DISS O2
Inlet assembly, 3-gas/5-conn
Inlet regulator (2x)
90 deg connector G 1/8 (3x)
Angle connector (5x)
Fitting, pilot to Divan (2x)
Contact 04 mm (3x)
Ferrule, Pilot to Divan (2x)
Pressure switch, O2 supply
Washer, M4 white (3x)
Washer, M4 blue (3x)
Filter holder (5x)
Filter (5x)
O-ring, #008 (EPDM) (5x)
Check valve (5x)
O-ring, #022 (Viton) (3x)
Screw, M4 x 40, cap skt hd (6x)
Lock washer, #10 split (6x)

6020.010 Fabius GS

Spare and Replacement Parts

Page 65 of 76



&RQWUROOHU3&%DQGUHODWHGLWHPV SDUWRIFRQWUROOHU
DVVHPEO\

SV00521

2
1
3

17

8
9

7
10

11
5

14

13

15

16

12

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

16

6020.010 Fabius GS

Spare and Replacement Parts

Page 66 of 76

1.
2.

3.
4.
5.
6.
7.
8.
9.

Plate, control PCB


PCB assembly
PCB assembly, Service Exchange
Fuse, 5 x 20 mm time lag, 1.6 A
Fuse, 5 x 20 mm time lag, 2.5 A
Fuse, 5 x 20 mm time lag, 4.0 A
Battery, 3V lithium
Screw, M3 x 0.5 x 6 btn hd skt (6x)
Lock washer, #4 int-t (6x)
Fuse Holder
Fuse, 5 x 20 mm time lag, 3.15 A
Cable assembly, power switch
Power inlet & mains fuse holder
Fuse, 5 x 20 mm time lag, 2.5 A (2x)

4116632

4199948

1832662
4117053

Assembly mounting screws not shown:


10. Screw, M4 x 0.7 x 8 mm cap skt hd (6x)
11. Lock washer, #8 int-t (6x)
12. Power supply
4117060
13. Cable assembly, power supply AC input
4117050
14. Cable assembly, power supply output (incl. capacitors & inductor) 4117051
15. Clip, power cable (2x)
16. Tie strap
17. Cable assembly, charge capacitors (incl. capacitors)
Power cord assembly:

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

USA, 15 ft.
UK/Ireland
Alternate part number
Spain/France/Germany/Austria
Alternate part number

6020.010 Fabius GS

Spare and Replacement Parts

Page 67 of 76



3QHXPDWLF 3((3&RQWURO $VVHPEO\

SV00519

24

23

25
17

22

14

21

18

19

20

13
10

12

9
11
15

5
2
1
3
16

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.

Bracket
Top Gasket
Muffler
Kep nut, M5
PEEP valve assembly
Superseded by
Screw, M5 x 0.8 x 8 mm btn hd skt (2x)

6020.010 Fabius GS

Spare and Replacement Parts

2600573

Page 68 of 76

Lock washer, #10 int-t (2x)


Filter
Top plate
Straight fitting, 1/16 ID x 10-32 w/seal
Straight fitting, 3/32 ID hose x 10-32M (4x)
Adapter fitting for 1/16 ID hose x 1/8 MPT
Lock washer, #10 int-t
Screw, M5 x 80 cap skt hd
Pump assembly
L-fitting, 3/32 ID hose x 10-32M
Variable restrictor
Hose, 2.5 x 2 SI (0.025 M)
Tie strap, 0.09 x 4 1/8
Hose, 2.5 x 2 SI (0.127 M)
Hose, natural, 0.075 ID x 0.030 W (0.178 M)
Hose, 4 x 1.5 SI NF clear (0.012 M)
Hose, 2.5 x 2 SI (0.040 M)
Hose, 2.5 x 2 SI (0.127 M)

2600564
4107995

1190520

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.
18.
19.
20.
21.
22.
23.
24.

6020.010 Fabius GS

Spare and Replacement Parts

Page 69 of 76



%H]HO$VVHPEO\0RQLWRU

10
12

5
3

6
8
7

11
4

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

SV00461

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.

REF

Keypad (control panel)


Bezel
Flat panel display
Knob
LED lamp PCB assembly
Rotary switch (incl. mtg. hardware)
Flat washer, 0.168 ID x 0.624 OD
Screw, self tap, M4 x 12 mm
Spacer, 0.162 ID x 0.205 L (4x)
Kep nut, M3 x 0.5 (4x)
Kep nut, M4 x 0.5 (5x)
Lock washer, #8 ext-t

6020.010 Fabius GS

Spare and Replacement Parts

4116941
4113755
M29655
4117017
4117012
4117021

Page 70 of 76

Switch panel inserts:

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

Export
USA
France

6020.010 Fabius GS

Spare and Replacement Parts

Page 71 of 76



9HQWLODWRU

10

6
7

4
3

11
1

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

SV00512

1.
2.
3.
4.
5.
6.

Motor & case assembly


Motor/ball screw assembly (includes encoder)
Ground wire assembly
Diaphragm, patient side
Diaphragm, piston
O-ring
Band clasp

6020.010 Fabius GS

Spare and Replacement Parts

4117097
4199947
2600650
2600651
G60337

Page 72 of 76

Screw, M6 x 80 hex hd
Patient assembly
High Pressure valve2600680
9. Negative Pressure Relief valve
10. Cable assembly, light barrier

2600775
2600670

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

7.
8.

6020.010 Fabius GS

Spare and Replacement Parts

Page 73 of 76

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
13.
14.
15.
16.
17.

Auxiliary O2 Flow Meter


Set screw, cup pt 6-32 x in.
Knob, auxiliary oxygen
Screw, btn hd skt 6-32 x in. (2x)
Label, 5/8 DOT - green/wht rings
Set screw, cup pt 10-32 x 7/8 in. (2x)
Label AUXILIARY OXYGEN
Tie Strap 0.09 W x 4 1/8 L (3x)
Hose, 0.171 ID x 0.040 W 20 in.
Hose, 0.075 ID x 0.030 W 24 in.
Coupling, 1/8 NPT (F)
T-fitting, 1/8 BSPP x 4 mm OD
Adapter, 1/16 ID hose x 1/8 MPT
Label, O2'
Kep nut, 10-32 (2x)
Hose clamp, press-on
L-fitting, NY 1/16 ID hose x 10-32 M

6020.010 Fabius GS

Spare and Replacement Parts

Page 74 of 76

1.

Related parts:
3. L-fitting, 1/4 tube x 1/8 MPT
4. Angle connector, M5
5. Collar, 6 mm white
6. Hose, 4 mm x 1 mm, PAE NF Clear

M30961
M31603
1210173

6020.010 Fabius GS

Page 75 of 76

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

2.

Two-vapor mounting system (illustrated)


Alternate part number
O-rings, vaporizer mount (4x)
One-vapor mounting system
Alternate part number
O-rings, vaporizer mount (2x)

Spare and Replacement Parts



6FDYHQJHU$*60

AGS

3
1

SV00560

Container
Filter
Flow tube
Screw plug
Coupling

M33292
M33294
M33293
M33291
M33149

For internal use only. Copyright reserved.

E6020001T01.fm 01.02.02

1.
2.
3.
4.
5.

6020.010 Fabius GS

Spare and Replacement Parts

Page 76 of 76

Drger Medical AG & Co. KGaA


Moislinger Allee 53 55
D-23542 Lbeck
Federal Republic of Germany

Phone (++49) 451/882 - 3127


Fax
(++49) 451/882 - 2160

Fabius GS en