Académique Documents
Professionnel Documents
Culture Documents
Content
1. Introduction
2. Test Procedure
2.1.
2.2.
2.3.
2.4.
2.5.
2.6.
Bubble test3
Jet impingement test5
Thermal Stability Test 6
Foaming Tendency Test
Emulsion Stabilization Test
FTIR Finger Printed Spectra8
6
7
1. Introduction
The following are the testing type to be carried out:
1. Bubble Testing
2. Jet impingement Testing
3. Thermal stability Testing
4. Emulsion stabilizing testing
5. Foaming tendency
6. IR Spectroscopy
Bubble Testing
The bubble test is a flexible laboratory procedure for monitoring corrosion rates and inhibitor
performance. In the bubble test cell, the composition and temperature of the fluids in the field
can be simulated. The bubbles of the purging gas and very slowly stirred of magnetic bar
generate the fluid movement, so that no hydrodynamics correlation exists between the flow
condition in the test and those in pipelines.
The objective of this test is to provide the efficiency of corrosion inhibitor under low shear
stress condition. The corrosion inhibitors will be tested at 50 ppm/brine volume. The test will
be carried out in a sweet environment saturated with CO2 and a temperature of 50C in the
absence and in the presence of 20% hydrocarbon ( cyclohexane or kerosene) phase.
The testing glass cell consist of three electrodes system (Working Electrode = mild carbon
steel, Auxiliary Electrode = Pt, Reference electrode = saturated calomel) is filled with 1600 mL
of synthetic brine solution (1.9% NaCl, water analysis composition as Tabel 1) is exposed to
continuous CO2 bubbling with constant magnetic stirring. Figure 1 represent the cell set up for
bubble testing. Pre-corrosion time is 4 hours followed by LPR measurement to obtain the
baseline of corrosion rate. The inhibitors will be tested is injected to the testing solution (50
ppm/brine volume) after four hours LPR measurement of the baseline corrosion rate relatively
constant. Duration of the LPR measurement on the presence of 50 ppm corrosion inhibitor
would be carried out along 16 hours. pH and dissolved oxygen of testing solution will be
recorded at the beginning and at the end of LPR measurement. The bubble test shall be done
in duplicate.
CationWater
in1litersolution:
19,459gNaCl
2,058gCaCl2.2H2O,
5,284gMgCl2.6H2O
0,082gBasCl2.2H2O,
AnionWater
in1litersolution:
19,459gNaCl
0,067gNa2SO4.10H2O;
SyntheticBrine
(2L)
38,919gNaCl
2,058gCaCl2.2H2O,
5,284gMgCl2.6H2O
0,082gBaCl2.2H2O,
.
0,067gNa2SO4 10H2O;
Figure1.Schematicdiagramoftheelectrochemicalcellusedforthebubbletest
Resultingdatafromthetestshallcover:
1) Plot of the corrosion rate and the polarization resistance versus time of each corrosion
inhibitorproduct.
2) Calculationofefficiencyofthecorrosioninhibitorbasedoncorrosionrateandpolarization
resistance.
Optionaltestingrequirementtothetestingprocedure:
1) The fluid composition in the bubble cell is 80% v/v brine: 20% v/v hydrocarbon. Brine
compositionis1.9%NaCl.
2) Hydrocarbon is saturated with CO2 (at least 1 hour continuous CO2 bubbling before it is
transferredtothebubblecell.Nocontaminationwithairduringtransferthehydrocarbonto
bubblecell).
3) BrineissaturatedwithCO2(atleast12hourscontinuousCO2 bubblinginthestoragetank,
beforethebrineistransferredtothebubblecell).
4) Durationofprecorrosionis4hours.
5) Durationofcorrosionmeasurementis16hours.
6) The bubble test shall be done in duplicate for each corrosion inhibitor product .and the
resultwillbetheaverageofthetwotestresults.
% CPE =
Rpi Rp0
x 100
(1)
Rpi
Rpi =polarization resistance in the presence of 50 ppm corrosion inhibitor
Rp0 =polarization resistance) of blank solution
or
% CPE =
CR0 CRi
x 100
( 2)
Rp0
CRi =corrosion rate in the presence of 50 ppm corrosion inhibitor
CR0 =corrosion rate of blank solution
The objective of this test is to evaluate the efficiency of corrosion inhibitor performance under a
known shear stress.The inhibitor concentration is 100 ppm/brinevolume. The testwill becarried
outinasweetenvironmentsaturatedwithCO2andatemperatureof40C.Thecalculatedwallshear
stressduringeachtypicalexperimentare150250Paforjetflowrateofapproximately500liters
/hour. The change in corrosion rate is followed with the progression of time as the adsorption
develop during the increase in inhibitor concentration of 100 ppm through a direct titration with
1275ppmofaninhibitorsolutionattheadditionrateofaround16.7mL/min.thepossibledecrease
in transfer resistance is followed with time of aging at constant shear stress level. Very persistent
monolayer will remain adsorbed and provide protection as shown by slow drop of transfer
resistance.Themeasurementisundertakenunderconstant150ml/minflowrateofcarbondioxide
gas at the temperature of around 400C in the absence of any hydrocarbon phase (kerosene).The
corrosionratesmeasurementwasconductedusingPotentiostat/GalvanostatVoltaLab30,PGZ301
RadiometerandcontrolledwithcorrosionanalysissoftwaremodelVoltaMaster4.Jet impingement
testing shall be done once only.
ThechangesofcorrosionratemonitoredbyLPRmeasurement
Thejetimpingementtestareundertakenonmildsteelelectrode(workingelectrode,surfacearea=
2.355cm2),auxiliaryplatinumelectrode(surfacearea=0.785cm2),andsaturatedcalomelreference
electrode.Thejetimpingementtestconditionwastemperatureof55oCof5000mLsyntheticbrine
(plusaddedvolumeofcorrosioninhibitorsolution)weresaturatedwithCO2(CO2pressureof1020
mbar above atmospheric pressures as a deoxygenated continuous purging gas inside the jet
impingementtestchamber).Onthistestcondition,Protectionefficiencyfactor(in%)forcorrosion
inhibitorcouldbecalculatedbyusingformulae:
%CPE=
Rp Rp0
x 100
Rp
Rp=polarizationresistancewithcorrosioninhibitor
Rp0=polarizationresistanceofblanksolution
Electricalwireof
workingelectrode
(a)
(b)
Figure2.(a)Schematicdiagramoftheelectrochemicalcellusedforjetimpingementtest,
(b)Schematicofringworkingelectrodeusedforjetimpingementtest
Data and data presentation resulting from the test shall cover at the minimum:
1) Corrosion inhibitor efficiency calculation.
2) Plot of corrosion inhibitor efficiency versus time of each corrosion inhibitor product.
2.6 IR Spectroscopy
The objective of this test is to obtain fingerprint of the tested corrosion inhibitor product. It is carried
out with common IR spectrometric methods.The testing shall be done in once only.
Data and data presentation resulting from the test shall cover at the minimum:
1) Plot of fingerprint obtained from common IR Spectroscopy, transmitans vs wave number plot.