Vous êtes sur la page 1sur 2

Manual Supplement

Manual Title:
Part Number:
Print Date:
Revision/Date:

7103 Users Guide


3729337
December 2013

Supplement Issue:
Issue Date:
Page Count:

1
1/15
1

This supplement contains information necessary to ensure the accuracy of the above manual.

2015 Fluke Corporation. All rights reserved.

7103 Users Guide

Manual Supplement

Change #1, 412


On page 49, under 12.2 Calibration Procedure, replace step 1 with:
1.

Choose three set-points to use in the calibration of the R0, ALPHA, DELTA and BETA
parameters. These set-points are generally -25 C, 0C, 50 C and 100 C but pther setpoints may be used if desired or necessary.

On page 50,
Change:

E = R4 - T3
F = R3 - T2
delta =

To:

E = R3 - R2
F = R2 - R1
delta =

Change:

=
=

To:

=
=

1/15

2 4
4 1 = 2 3

2 4
4 1 2 3