www.thermoscientific.com/labwarranty
04 / 2013
Thermo Scientific
04 / 2013
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or errorfree and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
ThermoFisher
SCIENTIFIC
C
DECLARATION OF CONFORMITY
KO N FO NU ITATSERKLA NU
ru
Beschreibunq /description
Modellbezeichnunq / model name
Modellnr/ model no.
Gltig ab Equipmentnr.
Valid from equipment no.
: Labor-Zentrifuqe
: RC3 BP+
: 75007553
:4126277'l
61010-l :2004
61010-2-020: 2006
61326-1: 2006
5501 1A:2007
6un
Dr. Thomas Reck
Werkleiter
Erstellt
:reiqeqeben
Lienemann
Laaboubi
Jatum
17.08.2011
17 0A 20,11
ok0ment
to1 31
3/
Revision
00
WEEE Conformity
This product is subject to the regulations of the EU Waste Electrical & Electronic Equipment (WEEE)
Directive 2002/96. It is marked by the following symbol:
Thermo Fisher Scientific has entered into agreements with recycling and disposal companies in all EU
Member States for the recycling and disposal of this device. For information on recycling and disposal
companies in Germany and on the products of Thermo Fisher Scientific, which fall under the RoHS
Directive (Restriction of the use of certain hazardous substances in electrical and electronic
equipment), please visit the website www.thermoscientific.com/rohswee
T
Table of Contents
Preface ....................................................................................................................................... iii
Scope of Supply ....................................................................................................................... iii
Intended Use ..............................................................................................................................iv
Accident Prevention .................................................................................................................iv
Precautions .................................................................................................................................v
Thermo Scientific
Chapter 1
Chapter 2
Chapter 3
Chapter 4
Contents
Chapter 5
Chapter 6
Chapter 7
Chapter 8
ii
Thermo Scientific
Preface
Before starting to use the centrifuge, read through this instruction manual carefully and follow the
instructions.
The information contained in this instruction manual is the property of; it is forbidden to copy or pass
on this information without explicit approval.
Failure to follow the instructions and safety information in this instruction manual will result in the
expiration of the sellers warranty.
3~
Scope of Supply
Article number
Quantity
Centrifuge Sorvall RC 3BP Plus
50127698
70907249
Installation kit
Check
If any parts are missing, please contact your nearest Thermo Fisher Scientific representative.
Thermo Scientific
iii
Intended Use
Intended Use
This centrifuge is a laboratory product designed to separate components by generation of
centrifugal force It separates human samples (e.g. blood, urine and other body fluids) collected in
appropriate containers, either alone or after addition of reagents or other additives.
The centrifuge is designed to also run other containers filled with chemicals, environmental
samples and other non-human body samples.
This centrifuge should be operated by trained specialists only.
The centrifuge is to be used for separating materials of different density or particle size suspended
in a liquid.
g
Maximum sample density at maximum speed: 1.2 ---------3
cm
Accident Prevention
Prerequisite for the safe operation of the centrifuge is a work environment in compliance
with standards, directives and trade association safety regulations and proper instruction of
the user..
The safety regulations contain the following basic recommendations:
Maintain a radius of at least 30 cm around the centrifuge.
Implementation of special measures which ensure that no one can approach the
centrifuge for longer than absolutely necessary while it is running.
The mains plug and main switch must be freely accessible at all times. Pull out the power
supply plug or disconnect the power supply in an emergency.
iv
Thermo Scientific
Precautions
Precautions
In order to ensure safe operation of the centrifuge, the following general safety regulations
must be followed: Ignoring the safety regulations can result in severe damage and inguries.
Do not manipulate the safety devices
The centrifuge should be operated by trained specialists only.
The centrifuge is to be used for its intended purpose only.
Plug the centrifuge only into sockets which have been properly grounded.
Do not move the centrifuge while it is running.
Use only rotors and accessories for this centrifuge which have been approved by Thermo
Fisher Scientific. Exceptions to this rule are commercially available glass or plastic
centrifuge tubes, provided they have been approved for the speed or the RCF value of
the rotor.
Do not use rotors which show any signs of corrosion and/or cracks.
Do not touch the mechanical components of the rotor and do not make any changes to
the mechanical components.
Use only with rotors which have been properly installed. Follow the instructions in
section Rotor Installation on page 4-3.
Use only with rotors which have been loaded properly. Follow the instructions given in
the rotor manual.
Never overload the rotor. Follow the instructions given in the rotor manual.
Never start the centrifuge when the door is open.
Never open the door until the rotor has come to a complete stop and this has been
confirmed in the display.
The door emergency release may be used in emergencies only to recover the samples
from the centrifuge, e.g. during a power failure (see section Failure message on
page 8-3).
Never use the centrifuge if parts of its cover panels are damaged or missing.
Do not touch the electronic components of the centrifuge or alter any electronic or
mechanical components.
Please observe the safety instructions.
Thermo Scientific
Precautions
vi
Thermo Scientific
Thermo Scientific
1-1
1-2
Thermo Scientific
Technical Data
The technical data of the Sorvall RC 3BP Plus is listed in the following table.
Table 1-1. Technical Data
Feature
Value
Environmental conditions
+2 C to +40 C
+35.6 F to +104 F
Overvoltage category
II
II
II
Pollution degree
Heat dissipation
230 V
208 V
400 V
~5.2 kW
~4.5 kW
~5.8 kW
17700 BTU/h
15300 BTU/h
19700 BTU/h
IP
20
20
20
Running time
unlimited
unlimited
unlimited
5000 rpm
5000 rpm
5000 rpm
300 rpm
300 rpm
300 rpm
7277 xg
7277 xg
7277 xg
< 153 kJ
< 153 kJ
< 153 kJ
< 58 dB (A)
< 58 dB (A)
< 58 dB(A)
+2 C to +40 C
+2 C to +40 C
+2 C to +40 C
1178 mm
800 mm
905 mm
3000 mm
46.3 inch
31.5 inch
35.6 inch
118.1 inch
475 kg
1047 lb
Dimensions
Hight
Width
Depth
Length power cord
Weight without rotor
1
Thermo Scientific
1-3
220-230 V, 50 Hz;
400 V, 50 Hz
EN 61326-1
EN 55011B
EN 61000-6-2
208-220 V, 60 Hz;
230 V, 60 Hz
Description / Features
Structure / Housing
Chamber
Stainless steel
Drive
Controls
Microprocessor-controlled
Internal memory
Functions
Imbalance detected
Door lock
Automatic door closing and locking starting from an initial hold position
1-4
Thermo Scientific
Mains Supply
The following table contains an overview of the electrical connection data for the der Sorvall RC 3BP
Plus. This data is to be taken into consideration when selecting the mains connection socket.
Table 1-4. Electrical connection data
Article
number
Mains
voltage
Frequency
Rated
current
Powerconsumption
Equipment fuse
Building fuse
Mains plug
75007530
208-220 V
60 Hz
17.5 A
3300 W
25 A & 6 A
30 A
NEMA 6-30P
SinglePhase
75007533
220-230 V
50 Hz
18 A
3500 W
25 A & 6 A
32 A
NEMA 6-30P
SinglePhase
75007565
400 V, 3~
50 Hz
16.5 A
3200 W
25 A & 6 A
3x32 A
IEC60309 CEE
32 A-6h
1P+N+PE
Rotor Selection
The Sorvall RC 3BP Plus centrifuge is supplied without a rotor.
Various Thermo Scientific rotors are available to choose from.
H-6000A
11250
HBB-6
11753
HLR-6
11350
The technical data of the rotors and the corresponding adapters and reduction sleeves for various
commercially available containers can be found in the corresponding rotor operating manuals.
For more information visit our website at: www.thermoscientific.com/centrifuge.
Thermo Scientific
1-5
Before use
Contents
Thermo Scientific
2-1
2 Before use
Before Setting up
Before Setting up
1. Check the centrifuge and the packaging for any shipping damage.
Inform the shipping company and Thermo Fisher Scientific immediately if any damage is
discovered.
2. Remove the packaging.
3. Check the order for completeness (see Scope of Supply on page P-iii).
If the order is incomplete, please contact Thermo Fisher Scientific.
Location
The centrifuge should only be operated indoors.
The set-up location must fullfil the following requirements:
A safety zone of at least 30 cm must be maintained around the centrifuge.
People and hazardous substances must be kept out of the safety zone while centrifuging.
The location must be stable and free of resonance.
The location must be suitable for horizontal setup of the centrifuge.
The centrifuge should not be exposed to heat and strong sunlight.
WARNING UV rays reduce the stability of plastics.
Do not subject the centrifuge, rotors and plastic accessories to direct sunlight.
The set-up location must be well-ventilated at all times.
Setting up
WARNING The centrifuge operates with high kinetic energies. Do not run the centrifuge
without a radius of at least 30 cm (12 inch) around the centrifuge.
Do not run the centrifuge while standing on its casters.
1. Place the pallet with the centrifuge so that you have at least 2 m room in front of the pallet.
2. Release the rails from the pallet.
3. Screw the rails onto the pallet in front of the centrifuge.
2-2
Thermo Scientific
2 Before use
Setting up
Rails
5. Place the locking stabilizers under the studs and screw them down.
6. Remove the 4 screws that fix the centrifuge on the pallet. The heads are upside down in the pallet.
7. Remove the bolt spacers.
8. Screw the studs with the safety brackets entirely up. Otherwise they might get stuck when rolling
the centrifuge off the pallet.
9. Remove the rubber feet.
Safety brackets
Studs
Thermo Scientific
2-3
2 Before use
Setting up
2-4
Thermo Scientific
2 Before use
Transporting the Centrifuge
Mains Connection
1. Turn off the power supply switch located on a small recessed panel in the upper right corner of the front
cabinet panel (press "0").
2. Check whether the cable complies with the safety standards of your country.
3. Make sure that the voltage and frequency correspond to the figures on the rating plate.
4. Plug the centrifuge into grounded electrical sockets only.
Other Connections
Thermo Scientific
2-5
2 Before use
Storage
Storage
Before storing the centrifuge and the accessories it must be cleaned and if necessary desinfected and
decontaminated.
Store the centrifuge in a clean, dust-free location.
Be sure to place the centrifuge on its feet.
Avoid direct sunlight.
2-6
Thermo Scientific
Control Panel
Contents
Thermo Scientific
3-1
3 Control Panel
Control Panel
Control Panel
The control panel consists of set display, run display, menu key with option indicators, primary
function keys, a numeric keypad, a keypad lock and
,
, and
key.
Keypad
Lock
Option indicators
Run Display
Menu Key
Primary Function Keys
Set Display
Numeric Keypad
Numeric Keypad
Press the
Press the
Press the
key to open the door. The door only opens when the rotor is not spinning. The message
OPEN appears in the run display.
Press the
Use the numeric keypad to enter values into the set display. The
removes input values or messages
from the set display. The
toggles between positive and negative temperature values.
Run Display
The run display shows the actual values during a run.
Speed field
The speed field indicates either:
3-2
Thermo Scientific
3 Control Panel
Set Display
Time field
The speed field indicates either:
the remaining time,
the time elapsed, or
the calculated ACE value (Accumulated Centrifugal Effect). ACE indicator is lit.
Temperature field
The temperature field indicates the current calculated sample temperature in C. The centrifuge
assumes that the temperature of sample, rotor, and centrifuge are equal at the start of the run.
Set Display
The set display is used:
to input and display basic run parameters,
to select options or other advanced features, and
to display advisory messages.
The set display shows the currently selected rotor, values for speed, temperature, time, and the
number of the program if a recalled program is used.
After pressing the
key to access advanced feature options, pressing the
options and returns to the set display.
Press the
Press the
Thermo Scientific
Press the
Press the
3-3
3 Control Panel
Primary Functions
Press the
Press the
or
Primary Functions
ROTOR
Press the
key to select a different rotor. The available rotors are listed one after the other in the
set display. This is necessary if the rotor or bucket/carrier system are changed. The correct
identification of the installed rotor is necessary for the set speed litmits, correctly calculate RCF,
and accurately control sample temperature. If a rotor is able to use buckets/carriers with different
maximum radii, each configuration will be listed for selection, to calculate the different RCF.
CAUTION Make sure that you always select the correct rotor and accessories.
The
SPEED
Press the
Speed is selectable from 300 to 4700 rpm depending on the specified rotor. The equivalent RCF
value is calculated at the maximum radius of the specified rotor. Adapters are not accounted for.
Press the
key repeatedly to toggle between rpm and RCF. Pressing once causes the value to flash
so a change can be made in that mode. Pressing again changes to the other mode and displays a
value equivalent to the previous. Pressing the
key again causes the value to flash so a change can
be made in that mode. When pressing again the set display toggles between rpm and RCF. An
entered value will continue to flash in the set dispaly when it is out of range.
The formula used to calculate RCF is:
U/min
RZB = 11,18 Radius in cm ---------------
1000
Note When changing mode, rpm/RCF values may appear a digit or two off from the original set
value. The centrifuge translates set RCF values to rpm whole numbers for speed control purposes.
The whole numbers are calculated to the closest RCF value. The speed control accuracy is not
compromised, and the slight value difference will not affect your run.
3-4
Thermo Scientific
3 Control Panel
Primary Functions
TEMP
Press the
The sample temperature is calculated by the centrifuge during a run with the messured actual
values. The calculation assumes that the sample temperature, rotor assembly temperature, and set
temperature are all equal at the start of the run (as they must be any time that temperature control
is critical).
RECALL
Press the
key to access program memory. Recall a program by entering a number from 1-9.
The set values can be modified without fear of damaging the saved program. In order to modify or
replace a program you need to access the option by pressing the
key.
TIME
Press the
key to specify a length of a run (minutes and seconds, to a maximum of 99:59). In the
run display, minutes and seconds begin to count down.
The run terminates when the timer reaches zero. Now the deceleration of the rotor begins.
HOLD
Press the
key to start a continuous run. The run terminates when you press the
key. During
a continuous run the set dispaly shows the message HOLD and ELAPSED. The time passed since the
key has been pressed is shown in the run display.
The timer will accumulate up to 99 minutes 59 seconds - if a run continues beyond that, the
displayed run time will remain fixed at 99:59. Whenever it is possible for such an extended run to
occur, an external timing method is recommended.
ACE
Press the
key to define the ACE (Accumulated Centrifugal Effect). The ACE can be
reproduced and does not depend on the load.
When ACE control is selected, the centrifuge calculates the effect of speed in relation to time (the
shaded area in the illustration below). The run duration accounts for acceleration variation.
Speed
Running time
Thermo Scientific
3-5
3 Control Panel
Keypad Lock
The actual ACE (expressed as 2 dt value up to 9.99 x 1030 [displayed as 9.99e30]) is given in the
run display.
The run is over when the set ACE value has been reached. Now the deceleration of the rotor
begins. In the run display will continue to accumulate the ACE. Taken care you do not confuse the
run display's final accumulation with the controlling input/timeout value, found in the set display.
Compared to control by time, ACE selection provides a more advanced and relevant form of run
duration control, thereby improving separation consistency, run reproducibility, and dependability
of results.
During a run you can press the
, and
Note If you change the duration control mode to TIME during a run, time will begin to count
down from the set value the moment the change is made. If you change mode to ACE control, the
set value that exists at the moment the change is made will control run termination. If you change
to ACE after the run has exceeded the set value, the run will terminate immediately.
Between runs, the run display shows the previous run duration values.
Press the
Press the
key to see the MIN:SEC value in the run display after the rotor is in standstill.
Press the
key to see the ACE value in the run display after the rotor is in standstill.
Keypad Lock
The keypad lock can be used to limit control panel function.
A key can be removed in 3 positions. Functionality is as follows:
The upper position, FULL (full function), allows use of all keys and features, without
restriction;
The lower position, SINGLE (Run Single Program), minimizes functionality so that set
parameters cannot be changed (the only active keys are
,
,
and
);
3-6
Thermo Scientific
3 Control Panel
Keypad Lock
The middle position, ANY (Run Any Program), makes it so that users may only perform runs
using run parameters recalled from program memory (only
, the numeric keypad
and
the keys
,
,
and
are active).
Use of the KEYPAD LOCK may offer a considerable advantage, improving process control by
reducing the possibility for operator error, as well as restricting unauthorized run parameter or
program memory changes.
Thermo Scientific
3-7
Operation
Contents
Thermo Scientific
4-1
4 Operation
Switch on Centrifuge
Switch on Centrifuge
Note Make sure the centrifuge is properly installed before turning it ON for the first time, (see
Setting up on page 2-2).
1. Turn on the power switch located on a small recessed panel in the upper right corner of the front
cabinet panel into position (select "1").
The device performs a self-check of its software.
2. Watch the control panel LEDs to be sure that all segments light.
Note If any LED segments do not light, note which ones and inform Customer Service. Continued
operation when the centrifuge is unable to display all run information correctly could mislead an
unaware observer.
During start-up the SET DISPLAY shows:
Open door
1. Press the
key.
WARNING Do not reach into the crack between the door and the housing. The door is
drawn shut automatically.
Use the emergency release only for malfunctions and power failures (see Failure message
on page 8-3).
Close door
Close the door by pressing down on it lightly in the middle or on both sides of it. Two locks close the
door completely. The door will close automatically.
WARNING Do not reach into the crack between the door and the housing. The door is
drawn shut automatically.
Note The door should audibly click into place.
4-2
Thermo Scientific
4 Operation
Rotor Installation
Rotor Installation
The approved rotors for the centrifuge are listed in section Rotor Selection on page 1-5. Use only the
rotors and accessories from this list.
CAUTION Unapproved or incorrectly combined accessories can cause serious damage to
the centrifuge.
Proceed as follows:
1. Open the centrifige door and if necessary remove any dust, foreign objects or residue from the
chamber.
2. Follow the instructions given in the rotor manual in order to install, load, and balance the rotor.
3. Close the centrifuge door.
4. Make sure the rotor matches the rotor in the SET DISPLAY.
5. Press the
Normal Operation
Preselecting Speed / RCF
1. Press the
key.
The display shows the RPM or the RCF-value depending on the display setting. Press the
to toggle between the two modes.
, key
Thermo Scientific
4-3
4 Operation
Normal Operation
n 2
RCF = 11, 18 ------------- r
1000
r = centrifuging radius in cm
n = Rotational speed in rpm
The maximum RCF value is related to the maximum radius of the tube opening.
Remember that this value is reduced depending on the tubes and adapters used.
This can be accounted for in the calculation above if required.
ACE
1. Press the
HOLD
Press the
key in order to select the HOLD run. During continuous operation, the centrifuge will
continue running until you stop it manually with the
key.
4-4
Thermo Scientific
4 Operation
Centrifugation
Centrifugation
Once the rotor has been properly installed, the main switch has been turned on and the door has been
closed, you can start centrifuging.
Imbalance indicator
If a load is imbalanced, this will be indicated by the message ROTOR IMBALANCE - BALANCE ROTOR.
The run will terminate.
Check the loading and start the centrifuge once again. See the information on proper loading in the
rotor instruction manual. For information on troubleshooting, see section Failure message on
page 8-3.
Usually the run time is preset and you only have to wait until the centrifuge stops automatically when
the preset time limit expires.
As soon as the speed drops to zero, the message END will appear in the SPEED field. By pressing the
key, you can open the door and remove the centrifuge material.
You can also stop the centrifugation program manually at any time by pressing the
key.
ACE
When you select an ACE control run, the run will determine when the centrifuge has reached the set
ACE value.
HOLD
If you selected continuous operation (see HOLD on page 4-4), you will have to stop the centrifuge
manually. Press the
key in the control panel. The centrifuge will be decelerated at the designated
rate. The message END will illuminate, and after pressing the
key, the door will open and you can
remove the samples.
Thermo Scientific
4-5
4 Operation
Programmed Operation
Programmed Operation
Storing Selections to Program Memory
1. Establish all desired run parameters as explained above, Normal Operation on page 4-3.
2. Press the
3. Press the
4. Press the
key.
The following screen will appear in the SET DISPLAY:
Note Single-digit program numbers 1-9 are for storing sets of standard run parameters.
Double-digit program numbers 10-15 are for storing step-runs.
5. Enter a program number.
6. Press the
key.
The SET DISPLAY will then return to the HOME screen, and that program number will appear
in the PROG # field as confirmation:
Note To prevent accidental erasure of established programs, if the entered program number has
been previously assigned, the following message will appear:
4-6
Thermo Scientific
4 Operation
Temperature Control
key.
The rotor will accelerate according to the recalled parameters. After the run the message END will
appear in the SPEED display and remain there until the door is opened or the
key will be pressed.
Note If programs are used within a step run, that have not been defined before, a warning message
will be displayed.
Temperature Control
The centrifuge controls calculate sample temperature during a run. The software assumes that the
sample and the rotor are at SET temperature at the start of the run. If temperatures are not
equilibrated, adjustments made by the centrifuge's control system may not be appropriate Over time,
temperature control should come into range.
Note When temperature control is critical, the sample, rotor wtih accessory amd rotor chamber
should all be at SET temperature when
key is presssed. Remove the sample soon after a run has
ended. The temperature in the rotor chamber drops slightly when the rotor is no longer generating
heat due to air friction.
Either store the rotor in a controlled temperature environment (such as refrigerator or cold room), or
precool/preheat the rotor in the centrifuge chamber, until the rotor is the same temperature as the
sample and the required SET temperature.
When the centrifuge is used to run large volumes of sample for short amounts of time the temperature
control is less important. During a very short run it is difficult to significantly change sample
temperature.
key).
4. Check to be sure the correct rotor name is in the ROTOR field of the SET DISPLAY. If not, press
the
key until it is.
5. Press the
key.
If cooling, input 1500 rpm.
Thermo Scientific
4-7
4 Operation
Rotor Temperature Equilibration
If heating, input a value that is approximately 70% of the maximum rated speed of the selected
rotor.
6. Press the
7. Press the
8. Press the
9. The centrifuge will alert you that a sample overtemperature condition exists, and that pre-cooling
is taking place. Press the
key.
10. Monitor progress of the value in the RUN TEMPERATURE display. Note the time when the
displayed RUN temperature reaches the SET temperature. From the time it does, allow the run to
continue for additional 30 minutes, then press the
key.
4-8
Thermo Scientific
Advanced Features
Contents
Thermo Scientific
5-1
5 Advanced Features
Advanced Features
Advanced Features
The advanced features of the centrifuge allow:
performing quality control runs (QC RUN),
changing acceleration/deceleration settings,
changing overtemperature alert settings,
linking together programmed run parameters, and
saving run parameters to program memory for simple recall.
Advanced feature selection is usually reflected by LED option status indicators. When a feature has
been selected, the corresponding indicator will be lit.
If changes to advanced feature selection are required, press the
key and then the
key until
selection screen of the particular feature appears. The
key accesses the advanced feature options and
the program memory.
Press the
Press the
key to enter the advanced features where you can check and change parameters as well
as select the advanced features.
Press the
Press the
key to step through options without changing selections or settings. In the secondary
screens you must enter a value to select the option.
Press the
key to exit the options and return to the set display without entering/selecting a
flashing value.
5-2
Thermo Scientific
5 Advanced Features
Advanced Features
Set Display
MENU
START
MENU
0 or ENTER
MENU
1
0 or ENTER
ENTER
MENU
1
0 or ENTER
ENTER
MENU
1
0 or ENTER
ENTER
MENU
1
0 or ENTER
ENTER
MENU
1
0 or ENTER
ENTER
MENU
1
0 or ENTER
ENTER
Thermo Scientific
5-3
5 Advanced Features
Advanced Features
QC RUN
QC RUN allows simple, automatic quality control run speed/temperature verification when used
with the optional computer interface package. The QC RUN feature will run and document the
data from each set of parameters that has been saved in program memory.
When QC RUN is selected, the set display will prompt you to prepare for the run and press the
key to begin. The centrifuge will start running program number 1, and systematically run each
program in order up to program number 9.
While the QC RUN is in progress, the parameters of the program that is running will be in the set
display.
During a QC RUN the centrifuge checks if the preset speed and temperature can be reached.. In
order to save time during the QC RUN the next program will be run once the desired speed has
been reached and the temperature is within 2 C of the preset value.
Note If you plan on using the QC RUN feature, consider the following as you save runs to
program memory:
QC programs should all apply using the same rotor.
QC RUN programs should not use HOLD.
To minimize the time it takes to perform a QC RUN, we recommend the following: Save runs to
program memory. Sort first by temperature (highest first), then, if there is more than one run at a
single temperature, by speed (lowest first). Start with program 1.
The programs below are listed in correct order for using QC RUN (program values shown are for
example purposes only, actual set parameters will vary).
ROTOR
SPEED
TIMED
DEG C
PROG #
H-6000
2900
04:00 AM
22
H-6000
4000
03:00 AM
22
H-6000
3800
03:45 AM
H-6000
4700
05:30 AM
key.
2. Press the
key.
SLOW START chooses gentle acceleration from 0 to 250 rpm. At 250 rpm the centrifuge
accelerates at normal, maximum rate. SLOW START offers ten different acceleration rates.
5-4
Thermo Scientific
5 Advanced Features
Advanced Features
The profiles are numbered for ease of selection, with number 1 being the slowest, most gradual
rate, and each successive rate being incrementally faster up to number 10.
After SLOW START selections are completed, SELECT SLOW STOP? will appear in the set display.
SLOW STOP
SLOW STOP chooses gentle deceleration from 500 to 0 rpm. SLOW STOP offers ten different
acceleration rates.
The profiles are numbered for ease of selection, with number 1 being the slowest, most gradual
rate, and each successive rate being incrementally faster up to number 10.
Selection of BRAKE OFF will have an affect on SLOW STOP (see the NOTE under BRAKE
OFF).
BRAKE OFF
BRAKE OFF deactivates normal deceleration braking for a coasting stop from any specified speed
(in rpm).
The time it takes for the rotor to stop, depends on the transition speed, windage and inertia of the
rotor.
BRAKE OFF transition speed is set independently of set run speed, and is not affected by change
to set run speed. If the transition speed is set higher than the set run speed, at run termination, the
centrifuge will coast to a stop from set speed.
Note If SLOW STOP and BRAKE OFF are both selected:
If the BRAKE OFF transition speed is set to 500 rpm or higher, the SLOW STOP selection
will be ignored.
If the BRAKE OFF transition speed is set below 500 rpm, the centrifuge will decelerate with
full braking to 500, transition to the specified SLOW STOP rate, then change to a coasting
stop when the specified BRAKE OFF transition speed is reached.
After BRAKE OFF selections are completed, CHANGE OVERTEMPERATURE LIMIT? will appear in
the set display.
CHANGE OVERTEMPERATURE LIMIT
Thermo Scientific
5-5
5 Advanced Features
Advanced Features
Recalling program runs may change the overtemperature limit. Specific overtemperature limit
settings can be saved in run programs.
EXAMPLE: If you have a set temperature of 22 C and set the maximum temperature to 24 C, an
overtemperature alert will occur if the calculated sample temperature reaches 25 C. If the set
temperature is then changed to 4 C, the maximum temperature setting will automatically change
to 6 C.
During a run, if the calculated sample temperature in the RUN display goes above the maximum
allowable sample temperature, the run will terminate. The message SAMPLE TEMPERATURE
OVER LIMIT will appear in the set display and an alarm will sound. This may indicate a condition
requiring simple corrective action, or it could indicate a refrigeration problem (SAMPLE
TEMPERATURE OVER LIMIT on page 8-3).
Overtemperature termination is deactivated when a temperature change between programs occurs
in a QC RUN. If, during a QC RUN, the observed RUN temperature is more than 2 C over SET
temperature, the rotor continues spinning until set temperature is reached.
After CHANGE OVERTEMPERATURE LIMIT selections are completed, SELECT STEP RUN?
will appear in the set display.
STEP RUN
5-6
Thermo Scientific
5 Advanced Features
SAVE RUN allows pre-programming of up to 15 different sets of run parameters (9 standard runs,
and 6 step runs) for future simple recall and error-free run reproducibility.
In addition to basic parameters, all option selections and settings (excluding rotor name and
overtemperature limit) will also be saved to program memory.
To eliminate inadvertent loss of existing programs, the save run sequence alerts users before
overwriting.
The
To save a run, all run parameter selections should be made before selecting this option.
After SAVE RUN selection and assignment of a program number, the set display will return to the
home screen.
Thermo Scientific
5-7
Control Inspection
Contents
Thermo Scientific
6-1
6 Control Inspection
QC RUN
QC RUN
The QC RUN feature allows simple, automatic quality control run speed/temperature verification The
QC RUN feature will run and document the data from each set of parameters that has been saved in
program memory. In a QC RUN the centrifuge will start running program number 1, and
systematically run each program in order up to program number 9.
To minimize the time it takes to perform a QC RUN, we recommend the following:
Save the programs to the program memory so that the program 1 has the highest temperature. The
program temperatures should be lower the higher the program number is. If several programs use
the same temperatures, sort them by speed. The lowest speed should come first.
All programs must use the same rotor.
No program should use HOLD.
EXAMPLE: The programs below are listed in correct order for using QC RUN (program values shown
are for example purposes only, actual set parameters will vary).
ROTOR
SPEED
TIMED
DEG C
PROG #
H-6000
2900
04:00 AM
22
H-6000
4000
03:00 AM
22
H-6000
3800
03:45 AM
H-6000
4700
05:30 AM
While the QC RUN is in progress, the parameters of the program that is running will be in the
SET DISPLAY.
During a QC RUN the centrifuge checks if the preset speed and temperature can be reached. In
order to save time during the QC RUN the next program will be run once the desired speed has
been reached and the temperature is within 2 C of the preset value.
To select the QC RUN feature:
1. Press the
key.
2. Press the
key.
6-2
Thermo Scientific
6 Control Inspection
Customer Control Inspection
Speed Controls
1. Prepare a sample of test fluid.
2. Install the rotor in the rotor chamber.
3. Load and balance the rotor according to the instructions in the rotor manual.
4. Close the centrifuge door.
5. Set the run parameters for a commonly used protocol, with speed controlled by rpm, and run
duration set to HOLD.
6. Deselect RCF or SLOW START if indicators are lit.
7. Press the
key.
8. Wait for the speed value in the run display to reach the specified set speed. Wait an additional 5
minutes for speed to stabilize.
9. Check rotor speed through the viewing port in the centrifuge door using a strobe tachometer (in
rpm, following instructions supplied with the tachometer).
The tachometer should indicate rpm equal to the SET value 20 rpm or 1 % (2000 rpm and
below 20 rpm, above 2000 rpm 1 %).
10. If necessary, repeat this procedure at other speeds that are commonly used in your protocols.
Timer Controls
1. Prepare a sample of test fluid.
2. Install the rotor in the rotor chamber.
3. Load and balance the rotor according to the instructions in the rotor manual.
4. Close the centrifuge door.
5. Set the run parameters for a commonly used protocol, but set the run duration by time
(MIN:SEC) to 10:00.
6. Press the
key.
7. Using a stopwatch, begin timing precisely as the run time display counts down to 10:00. Then stop
timing precisely as the RUN TIME display counts down to 00:00.
The stopwatch should read between 9:30 to 10:30, representing 10:00 5 % (29 seconds).
8. If necessary, repeat this procedure at other speeds that are commonly used in your protocols.
Thermo Scientific
6-3
6 Control Inspection
Customer Control Inspection
Temperature Controls
When verifying temperature control, all test run conditions (TEMP, SPEED, TIME, ROTOR and
option selections) should match your required protocol(s). Test fluid must be at the same volume and
temperature as the actual sample (use a calibrated thermometer to confirm). The sample temperature,
rotor assembly temperature, and SET temperature must match also. Ideally, all components would
have been stored in the same controlled-temperature area overnight. If that is not possible, you can
precool or preheat the rotor as required Either store the rotor in a controlled temperature environment
(such as refrigerator or cold room), or precool/preheat the rotor in the centrifuge chamber, until the
rotor is the same temperature as the sample and the required SET temperature.
Immediately after the test run is performed, check test fluid temperature using the same thermometer
that was used before the run.
1. Enter the program parameters. Choose the protocol with the longest run time.
2. Rotor chamber, rotor, and test fluid must have the desired sample temperature.
3. Prepare two equal loads of test fluid (similar in volume to normal run samples) equilibrated to the
desired sample temperature. Check the sample temperature using a calibrated thermometer.
Note Test fluid must be compatible with aluminum and have a freezing point below the desired
sample temperature.
4. Install the rotor in the rotor chamber.
5. Load and balance the rotor according to the instructions in the rotor manual.
6. Close the centrifuge door.
7. Press the
key.
During the run, the run display temperature should be within 2 C of the set display temperature.
8. Immediately after the test run is performed, check test fluid temperature using the same
thermometer that was used before the run.
The measured sample temperature should be within 2 C of set temperature.
9. If necessary, repeat this procedure at other speeds that are commonly used in your protocols.
10. If an out-of-range temperature is observed, repeat the test run. Re-equilibrate the chamber, rotor,
and test fluid temperatures to match desired sample temperature, and repeat the test run.
6-4
Thermo Scientific
Thermo Scientific
7-1
Cleaning Intervals
For the sake of personal, environmental, and material protection, it is your duty to clean and if necessary
disinfect the centrifuge on a regular basis.
Maintenance
Recommended Interval
Clean rotor
Accessories
Cabinet
Ventilation holes
CAUTION Refrain from using any other cleaning or decontamination procedure than those
recommended here, if you are not entirely sure that the intended procedure is safe for the
equipment.
Use only approved cleansers.
If in doubt, contact Thermo Fisher Scientific.
Cleaning
General notes on cleaning
Use warm water with a neutral solvent.
Never use caustic cleaning agents such as soap suds, phosphoric acid, bleaching solutions or
scrubbing powder.
Rinse the cavities out thoroughly.
Use a soft brush without metal bristles to remove stubborn residue.
Afterwards rinse with distilled water.
Place the rotors on a plastic grate with their cavities pointing down.
If drying boxes are used, the temperature must never exceed 50 C C, since higher temperatures
could damage the material and shorten the lifetime of the parts.
Use only disinfectants with a pH of 6-8.
Dry aluminum parts off with a soft cloth.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil.
Store the aluminum parts at room temperature or in a cold-storage room with the cavities pointing
down.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
7-2
Thermo Scientific
CAUTION Observe that no liquids, particularly no organic solvents, get into the openings
at the motor shaft outlet, as they could leach out the grease of the bearings.
Clean centrifuge and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Prepare the rotor according to the rotor manual.
CAUTION Lift the rotor with two persons due to its weight.
key.
Thermo Scientific
7-3
Disinfection
Disinfect the centrifuge immediately whenever infectious material has spilled during centrifugation.
WARNING Infectious material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Disinfect the affected parts immediately.
Take other precautions if need be.
The rotor chamber and the rotor should be treated preferably with a neutral disinfectant.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
Observe the safety precautions and handling instructions for the cleaning agents used.
Contact the Service Department of Thermo Fisher Scientific for questions regarding the use of other
disinfectants.
Disinfect the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Prepare the rotor according to the rotor manual.
CAUTION Lift the rotor with two persons due to its weight.
5. Remove the centrifuge tubes and adaptors and dispose of them or disinfect them.
6. Treat the rotor and accessories according to the instructions for the disinfectant. Adhere strictly to
the given application times.
7. Be sure the disinfectant can drain off the rotor.
8. Rinse the rotor and rotor lid thoroughly with water and then rub down.
9. Dispose of the disinfectant according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 C.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil.
Tread the bolt of the swing out rotor with bolt grease (75003786).
7-4
Thermo Scientific
Decontamination
Decontaminate the centrifuge immediately whenever radioactive material has spilled during
centrifugation.
WARNING Radioactive material can get into the centrifuge when a tube breaks or as a
result of spills. Keep in mind the risk of infection when touching the rotor and take all
necessary precautions.
In case of contamination, make sure that others are not put at risk.
Decontaminate the affected parts immediately.
Take other precautions if need be.
CAUTION Before using any cleaning or decontamination methods except those
recommended by the manufacturer, users should check with the manufacturer that the
proposed method will not damage the equipment.
For general radioactive decontamination use a solution of equal parts of 70% ethanol, 10% SDS and
water.
Decontaminate the rotor and accessories as follows:
1. Open the centrifuge.
2. Turn off the centrifuge.
3. Pull out the power supply plug.
4. Prepare the rotor according to the rotor manual.
CAUTION Lift the rotor with two persons due to its weight.
5. Remove the centrifuge tubes and adaptors and dispose of them or disinfect them.
6. Rinse the rotor first with ethanol and then with de-ionized water.
Adhere strictly to the given application times.
7. Be sure the decontamination solution can drain off the rotor.
8. Rinse the rotor and accessories thoroughly with water.
9. Dispose of the decontamination solution according to the applicable guidelines.
10. Dry all of the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a
maximum temperature of 50 C.
After cleaning, treat the entire surface of aluminum parts with corrosion protection oil
(70009824). Also treat the cavities with oil.
Tread the bolt of the swing out rotor with bolt grease (75003786).
Thermo Scientific
7-5
7-6
Thermo Scientific
Troubleshooting
Contents
Thermo Scientific
8-1
8 Troubleshooting
Mechanical Emergency Door Release
Levers
8-2
key
Thermo Scientific
8 Troubleshooting
Circuit Breakers
Circuit Breakers
The centrifuge has circuit breakers. The circuit breakers prevent the centrifuge from damages due to
residual current. In the event of a failure inform Customer Service.
Failure message
If problems occur other than those listed in this table, the authorized customer service
representative must be contacted.
Failure message
NO ROTOR
DETECTED
ROTOR IMBALANCE
- BALANCE ROTOR
Imbalance detected.
CRITICAL
OVERTEMP
AIR TEMP
MEASURE ERROR
Thermo Scientific
8-3
8 Troubleshooting
Failure message
Failure message
MISSING MEMORY
CONTAINING
CONFIGURATION
DOOR CLOSED
SWITCH FAILURE
DOOR LOCK
SWITCH FAILURE
DOOR LOCK
SWITCH SHORTED
ROTOR NOT
SPINNING
MAIN CONTACT
FAILURE
8-4
Thermo Scientific
8 Troubleshooting
Failure message
Failure message
TACH FAULT
DETECTED
INTERNAL SYSTEM
ERROR OCCURRED
MEMORY NOT
INITIALISED
Thermo Scientific
8-5
8 Troubleshooting
Failure message
Failure message
INTERNAL
PROGRAM ERROR
FREQUENCY
CONVERTER
VOLTAGE TO HIGH
8-6
Thermo Scientific
BUNA N
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
POLYALLOMER
POLYCARBONATE
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
2-mercaptoethanol
Acetaldehyde
CHEMICAL
MATERIAL
ALUMINUM
Acetone
Acetonitrile
Alconox
Allyl Alcohol
Aluminum Chloride
Ammonium Acetate
Ammonium Carbonate
Ammoniumhydroxid (28%)
Ammonium Phosphate
Ammonium Sulfate
Amyl Alcohol
Aniline
Barium Salts
Benzene
Benzyl Alcohol
Boric Acid
Cesium Acetate
Cesium Bromide
Cesium Chloride
Cesium Formate
Cesium Iodide
Thermo Scientific
A-1
BUNA N
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
POLYALLOMER
POLYCARBONATE
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
Cesium Sulfate
Chloroform
Cresol Mixture
Cyclohexane
Deoxycholate
Distilled Water
Dextran
Diethyl Ether
Diethyl Ketone
Diethylpyrocarbonate
Dimethylsulfoxide
Dioxane
Ferric Chloride
Ethyl Acetate
Ethylene Dichloride
Ethylene Glycol
Ficoll-Hypaque
Formaldehyde (40%)
Glutaraldehyde
Glycerol
Guanidine Hydrochloride
Haemo-Sol
Hexane
Isobutyl Alcohol
Isopropyl Alcohol
Iodoacetic Acid
CHEMICAL
A-2
MATERIAL
ALUMINUM
Thermo Scientific
BUNA N
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
POLYALLOMER
POLYCARBONATE
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
Potassium Bromide
Potassium Carbonate
Potassium Chloride
CHEMICAL
MATERIAL
ALUMINUM
Potassium Permanganate
Calcium Chloride
Calcium Hypochlorite
Kerosene
Iodoacetic Acid
Carbon Tetrachloride
Aqua Regia
Magnesium Chloride
Mercaptoacetic Acid
Methyl Alcohol
Methylene Chloride
Metrizamide
N-Butyl Alcohol
N-Butyl Phthalate
N, N-Dimethylformamide
Sodium Borate
Sodium Bromide
Sodium Iodide
Sodium Nitrate
Sodium Sulfate
Sodium Sulfide
Sodium Sulfite
Nickel Salts
Oils (Petroleum)
Thermo Scientific
A-3
BUNA N
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
POLYALLOMER
POLYCARBONATE
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
Oils (Other)
Oleic Acid
Oxalic Acid
Phenol (5%)
Phenol (50%)
CHEMICAL
MATERIAL
ALUMINUM
Picric Acid
Pyridine (50%)
Rubidium Bromide
Rubidium Chloride
Sucrose
Sucrose, Alkaline
Sulfosalicylic Acid
Schwefelsure (konz.)
Stearic Acid
Tetrahydrofuran
Toluene
Trichloroacetic Acid
Trichloroethane
Trichloroethylene
Trisodium Phosphate
Triton X-100
Urea
A-4
Thermo Scientific
BUNA N
DELRIN
ETHYLENE PROPYLENE
GLASS
NEOPRENE
NORYL
NYLON
POLYALLOMER
POLYCARBONATE
POLYTHERMIDE
POLYRTHYLENE
POLYPROPYLENE
POLYSULFONE
POLYVINYL CHLORIDE
RULON A, TEFLON
SILICONE RUBBER
STAINLESS STEEL
TITANIUM
TYGON
VITON
Xylene
Zinc Chloride
Zinc Sulfate
MATERIAL
ALUMINUM
CHEMICAL
Polyethyleneterephthalate
Key
S
Satisfactory
Moderate attack, may be satisfactory for use in centrifuge depending on length of exposure, speed involved, etc. Suggest testing under actual
conditions of use.
--
Performance unknown; suggest testing, using sample to avoid loss of valuable material.
Chemical resistance data is included only as a guide to product use. No organized chemical resistance
data exists for materials under the stress of centrifugation. When in doubt we recommend pretesting
sample lots.
Thermo Scientific
A-5
Warranty
Thermo Fisher Scientific makes no warranty of any kind, expressed or implied, except as stated in this
warranty policy.
The Thermo Scientific Sorvall RC 3BP Plus Centrifuge is warranted (subject to the conditions
specified below and in the warranty clause of the Thermo Fisher Scientific terms and conditions of sale
in effect at the time of sale) to be free from defects in material and workmanship for a period of one (1)
year from the date of delivery. Thermo Fisher Scientific will repair or replace and return free of charge
any part which is returned to its factory within said period, transportation prepaid by user, and which
is found upon inspection to have been defective in materials or workmanship. This warranty does not
include normal wear from use, it does not apply to any instrument or part which has been altered by
anyone other than an employee of Thermo Fisher Scientific, nor to any instrument which has been
damaged through accident, negligence, failure to follow operating instructions, the use of electric
currents or circuits other than those specified on the plate affixed to the instrument, misuse or abuse.
Thermo Fisher Scientific reserves the right to change, alter, modify or improve any of its instruments
without any obligation whatever to make corresponding changes to any instrument previously sold or
shipped.
The foregoing obligations are in lieu of all other obligations and liabilities including negligence and all
warranties, of merchantability or otherwise, expressed or implied in fact or by law, and state our entire
and exclusive liability and buyer's exclusive remedy for any claim or damages in connection with the
sale or furnishing of goods or parts, their design, suitability for use, installation or operation Thermo
Fisher Scientific will in no event be liable for any special or consequential damages whatsoever, and our
liability under no circumstances will exceed the contract price for the goods for which liability is
claimed.
Terms may vary by country. Please contact your local sales office for further information.
Thermo Scientific
B-1
Contact Information
Thermo Scientific
Country
Telephone number
Latin America
Austria
+43 1 801 40 0
Belgium
+32 53 73 42 41
France
Germany
Italy
Netherlands
+31 76 579 55 55
Russia
+7 812 703 42 15
Spain / Portugal
+34 93 223 09 18
Switzerland
+41 44 454 12 22
UK / Ireland
China
India
Japan
Australia
New Zealand
C-1
Precautions ......................................................... v
Preface ............................................................... iii
Preselecting Speed / RCF ................................4-3
Preselecting the temperature ............................4-4
D
Decontamination ............................................ 7-5
Desinfection .................................................... 7-4
E
Explanation of RCF Value ..............................4-4
F
Failure messages .............................................. 8-3
H
HOLD ............................................................ 4-5
R
Rotor Installation ............................................4-3
S
Select run time ................................................4-4
Service of Thermo Fisher Scientific .................7-6
Set run time ....................................................4-5
Starting centrifuge program .............................4-5
Stopping the centrifugation program ...............4-5
Storage ............................................................2-6
T
Transporting the Centrifuge ............................2-5
Troubleshooting ..............................................8-1
I
Imbalance indicator ......................................... 4-5
L
Location .......................................................... 2-2
M
Mains Connections ......................................... 2-5
Maintenace ..................................................... 7-1
Thermo Scientific
I-1