Conten
CKIC
Overview
Testing Procedure
General element
testing
Related standards
Instrument
procedure
Application &
Stud
&Instrum
Element In Coal
CKIC
S:0.3~4.
0%
0.8~5.0
%
45~85
%
0.20~1.5
%
CKIC
CKIC
Why TestC,H,N
?
Carbon makes
H2O evaporation
will take away
calorific value.
To calculate the
Net C.V,
H value is
calorific
H2O evaporation
Why Tes S ?
CKIC
Safe
ty
Producti
Soci
al
Responsib
Eficie
nt
Energy
Sulf
ur
Testin Method
CKIC
C&
H
Liebig
method
Electric
QuantityWeight
Infrared
Absorpti
on
method
N
Semimicro
Kjeldahl
Semimicro
Gasificati
on
Thermal
Conductiv
ity
method
Eschka
method
Calcul
High
at ing
temperature
combustion
metho
method
CoulometricTitrat
d
ion metho
d
Infrared
metho
Absorption
d
Summary
of Element
CKIC
Principleof InfraredAbsorption
CKIC
Infrared Cell
Principleof Thermal
CKIC
Thermal
Conductivity
Gas
Molecular formula
Molecular weight
Coefficient of
thermal conductivity
Hydrogen
H2
39
Helium
He
33
Neon
Ne
20
10.4
Oxygen
O2
32
5.7
Nitrogen
N2
28
5.6
Air (dry)
Air
29
5.4
Carbon monoxide
CO
28
5.4
Water vapor
H2O
18
4.0
Argon
Ar
40
3.8
Carbon dioxide
CO2
44
3.3
Sulfur dioxide
SO2
64
1.6
Befor
e
Analys
Reference
Cell
He
He
Measuring
Cell
Stable
Durin
g
Analys
Reference
Cell
N2
He
Measuring
Cell
Balance
CKIC
Sulfur Testing
Standard
CKIC
GB/T
25214
STANDAR
DS
Determination of Total Sulfur in CoalInfrared
Absorpti Metho
on
d
Standard
CKIC
Esch
ka
Meth
Coulo
mb
M
d
e
I
R
Absorpt
Chemical Method
Physical Method
Physical Method
GB
Approved
ISO
GB
Approved
Apply for
ISO
GB
Approved
ISO
Approved
Standards-General
Procedure
CKIC
Instrument Preparation
Perform the instrument set up according to
the instruction manual:
Temperature setting, leakage checking.
Standards-General Testing
CKIC
Calibration
Calibration should be processed before any
normal determinations.
Verification of the calibration.
Standards-General Testing
CKIC
Determinations
Carry out the determination by weighing out a
test portion of each analysis sample within a
mass range recommended by the manufacturer.
CKIC
Sulfur Testing
--Coulomb Titration Method
Standard
CKIC
Standard
CKIC
Maintenance
CKIC
Dust cleaning
Surface cleaning
Internal cleaning: filters, electrolytic cell
Leakage checking
O-ring, combustion/reagent tube
Reagent
KI solution, silica gel
Crucible replacement
Application
CKIC
Application
CKIC
Material
Heavy Oil
Plant, Feed,
Grain, and
Coal, Coke,
and
Fertilizers
Pretreatment
0.1 gram, add quartz
sand
to sampl
cove entir
r
e
e
0.1 gram, uniform
powder
0.05 gram, uniform
powder
0.05-0.1 gram, uniform
powde
r
CKIC
Sulfur Testing
--Infrared Absorption Method
Standard
CKIC
Standard
CKIC
Maintenance
CKIC
Dust cleaning
Surface cleaning
Internal cleaning: reagent tube, filters, IR cell chamber
Leakage checking
Gas supply, o-ring, combustion/reagent tube
Reagent
Magnesium perchlorate
Crucible replacement
Application
CKIC
Application
CKIC
Material
Cement,
Limestone, Lime,
andPlant,
Similar-Type
Feed,
Grain, and
Flo and
Coal, Coke,
Graphite
Ore
s
Fertilizers
Plant tissue
Pretreatment
0.3 gram, uniform powder
0.35 gram, uniform powder
0.3 gram, uniform powder
0.1 gram, uniform powder
0.05-0.1 gram, uniform
powder
Weigh 0.25 gram sample
which should be of uniform
consistency
and dried at 85 for two
hours
prior to analysis.
Instrumen
CKIC
Medium Model
IR Absorption
Economic Model
Coulomb Method
5EIRS3000
5EIRSII
5ES3200
Advanced Model
IR Absorption
CKIC
CHN Testing
Standard
CKIC
ISO
29541
ASTM
D5373
GB
?
STANDAR
DS
Solid mineral
fuels Determination
of total
carbon, hydrogen and nitrogen
Standard Test Methods for
Instrumental Determination of
Carbon, Hydrogen, and Nitrogen
in Laboratory Samples of Coal
Determination of total carbon,
hydrogen and nitrogen content in
coal by Instrumental method
Standards-General Testing
CKIC
Instrument Set-Up
Verify the specifications according to the
instrument operating manual.
Verify the condition and quantities of all
chemicals.
Leakage checking.
Standards-General Testing
CKIC
Blank Analyses
Perform blank analyses to establish carbon,
hydrogen and nitrogen levels in the
combustion and carrier gases.
Standards-General Testing
CKIC
Conditioning
Select a proper coal to check the stability of
the instrument, whether a corrective action
needed before proceeding with calibration.
Standards-General Testing
CKIC
Calibration
Calibration should be processed before any
normal determinations.
Verification of the calibration.
Standards-General Testing
CKIC
Determinations
Carry out the determination by weighing out a
test portion of each analysis sample within a
mass range recommended by the manufacturer.
Standards-General Testin
Procedure
CKIC
Determinations
Preparing foil
cup
Weighing
sample
Making foil
capsule
Loading
sample
InstrumentCKIC
Analysis Process
Each sample analysis is divided into
three
Purg
e
Clean sample
Clean gas blending tank
Clean gas path tube
Read blank voltage
stages
.
Combusti
on
Sample fully
combusts
Remove gas
impurities and dust
particle.
Analys
is
Instrument18
99.995%
O2
Sample
C Result
reagents
H Result
H2O IR Cell
CO2 IR Cell
Thermal
Conductivity Cell
CO2
H2O
SOx
NOx
Purify
N2
CO2
H2O
NOx
O2
CO2
H2O
NOx
O2
CO2 Reduced
H2O
N2
N Result
InstrumentCKIC
Control
Circuit
Infrare cell
d
Double layer
sample
automatic
loade
r
InstrumentCKIC
MODEL
Analysis
time(min.)
Auto-Sampler
Analysis Range
5E-CNH 2200
Approx. 5-6
35 Samples, stackable to 70
samples
C:0.005%~100%50ppm;H:0.02%~50%200ppm;N:0.0
08~50%80ppm
Sample Mass
(mg)
80-100recommended
Analysis
Method
C&H:
by IR Absorption Method N:Thermal
detection
method
Carrier Gas: Helium gas, purity 99.995%,
(0.250.01)Mpa
Combustion Gas: Oxygen,for measuring C&H:
purity99.50%;
for measuring nitrogen, the purity of higher
99.995%
0.250.01Mpa, power gas: Compressed air or
Gas
Required
Furnace
Type
C0.5%;H0.15%;N0.08%
Maintenance
CKIC
Dust cleaning
Surface cleaning
Internal cleaning: reagent tube, filters, ballast tank, IR cell
chamber
Leakage checking
Gas supply, o-ring, combustion/reduction tube
Reagent
Magnesium perchlorate, CO2 absorb reagent, high purity
copper, N-catalyst
Crucible replacement
Application
CKIC
Power plants
Coal mines
Commercial Labs
Steel Plant
Organic Chemicals
Milling
Applications
(Feeds, Oilseeds)
Agriculture
Plant Food
industry
Fertilizer Plant
Application
CKIC
Need do pre
Application
CKIC
Application
CKIC
Application
CKIC
Nitrogen in fertilizer
Weigh 0.05 gram uniform powder into tin foil cup,
calibration should be processed with EDTA before
any analysis.
Application
CKIC
Nitrogen in Plastics
Weigh 0.15-0.25 gram uniform powder into bigger
size of tin foil cup, 24-hole carousel instead of the
normal 35-hole carousel of the instrument.
C,H in Biomass
Weigh 0.15 gram uniform powder into bigger size
of tin foil cup, 24-hole carousel instead of the
normal 35-hole carousel of the instrument.
Case Stud
27
Characteristic
Solution
Note
Case
28
N content
<300ppm
Case
29
Increase
weight
80mg
800mg
Case
30
5ml
10ml
Case Stud
31
Questio
CKIC