Installation Manual
Version 1.1.2
Roche
cobas 4800 system
Table of Contents
1. Preface................................................................................................................................................................. 8
1.1. Revision History .......................................................................................................................................................................... 8
1.2. Editor notice............................................................................................................................................................................... 10
1.3. Copyrights .................................................................................................................................................................................. 10
1.4. Trademarks ................................................................................................................................................................................ 10
1.5. Contact addresses................................................................................................................................................................... 10
1.6. Intended Use ............................................................................................................................................................................. 11
1.7. Configuration............................................................................................................................................................................. 11
1.8. About this Manual................................................................................................................................................................... 11
1.9. Order Information .................................................................................................................................................................... 12
1.10.
Document Reference....................................................................................................................................................... 12
1.11.
1.12.
Abbreviations...................................................................................................................................................................... 13
1.13.
Safety notes......................................................................................................................................................................... 14
1.14.
Data Security....................................................................................................................................................................... 16
1.15.
1.16.
Version 1.1.2
Page 3 of 175
Roche
cobas 4800 system
4.7.1.
4.7.2.
Reporting ................................................................................................................................................................... 42
4.7.3.
Test parameters....................................................................................................................................................... 43
4.7.4.
4.7.5.
4.7.6.
4.7.7.
4.12.
4.13.
4.15.
4.18.
4.19.
4.20.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
5.2. Restore Database..................................................................................................................................................................... 74
6. cobas x 480 instrument Installation ............................................................................................................ 76
6.1. Stationary Carrier and Heater/Shaker Unit..................................................................................................................... 76
6.2. Liquid Waste Installation Procedure ................................................................................................................................. 78
6.3. Firmware Verification.............................................................................................................................................................. 80
6.3.1.
Update Firmware..................................................................................................................................................... 81
9.2.2.
9.2.3.
9.3.2.
9.3.3.
9.3.4.
9.3.5.
Sending Results.....................................................................................................................................................100
9.4.2.
Physical Connection.......................................................................................................................................................104
10.2.
10.4.
11.2.
Installation Manual
Version 1.1.2
Page 5 of 175
Roche
cobas 4800 system
11.3.
Additional information...................................................................................................................................................122
12.2.
12.3.
12.4.
Goal ......................................................................................................................................................................................131
13.2.
Consumables ....................................................................................................................................................................132
Goal ......................................................................................................................................................................................142
14.2.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14.4.4. Load Reagent Reservoir Carrier 2-2 ..............................................................................................................147
14.4.5. Load Reagent Reservoir Carrier 2-3 ..............................................................................................................147
14.4.6. Load Reagent Reservoir Carrier 2-4 ..............................................................................................................147
14.4.7. Load Reagent Reservoir Carrier 2-5 ..............................................................................................................147
14.4.8. Load Reagent Vials Carrier 3-15,18,19, 23, 24 ...........................................................................................148
14.4.9. End of Sample Preparation................................................................................................................................149
14.4.10. Amplification and Detection .............................................................................................................................149
14.4.11. Control Curve Analysis........................................................................................................................................150
14.4.12. Finalize HPV Run...................................................................................................................................................150
14.4.13. Clean up...................................................................................................................................................................150
15. Completion of Installation............................................................................................................................151
16. Appendix..........................................................................................................................................................152
16.1.
16.2.
16.3.
Network Overview...........................................................................................................................................................160
16.4.
16.5.
16.6.
16.7.
Installation Manual
Version 1.1.2
Page 7 of 175
1
1.1
Preface
Revision History
Roche
cobas 4800 system
1. Preface
1.1. Revision History
Version
Revision Date
Revision Info
1.0
05-NOV-2009
First Version
1.0.1
05-FEB-2010
1.0.2
07-APR-2010
1.1.0
18-Nov-2010
New chapter
2 Upgrade from version 1.0.x to 1.1.0
Backup and Restore Manual is completely integrated into the installation manual
4 SW Installation control unit
Verification of installed BIOS
All subchapter revised with new cobas 4800 software 1.1.0.1042
4.17 Workaround exchanging of Workorder Config. File
4.18 Exchanging of Workorder Config. File
5 Upgrade Installation Restore Database5
9 LIS Interface Installation
10 LIS Interface Configuration with cobas IT firewall
11 Change LIS Interface Configuration
The Network Configuration Guide is completely implemented into the Installation
manual
12.4 Printer Installation in Laboratory Network (red Network)
16.3 Network Overview
16.7 Additional HPV Barcodes for WaterRun
Page 8 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Preface
Revision History
1
1.1
Moved to Appendix
Installation of OSXP Recovery Image
Removed Chapter
3.2 Verification RAID System
3.3. RAID Array configuration
6.4 Instrument supported barcodes
3.16.1. Firmware cobas z 480 analyzer
1.1.1
8-Dez-2010
1.16.
Material Required/Software
Software added: Host Interface Test Tool 7.1.5 or higher
3.4
9.3.2.
Transformer Configuration
Chapter added: Result Upload Configuration
9.3.3.
13.5.
2-Mar-2011
Installation Manual
Version 1.1.2
Page 9 of 175
1
1.2
Preface
Editor notice
Roche
cobas 4800 system
1.3. Copyrights
Copyright 2010, Roche Molecular Systems, Inc. All rights reserved.
2010, F. Hoffmann-La Roche Ltd. All rights reserved.
Microlab is a registered trademark of Hamilton Bonaduz Ltd, Switzerland
2010, Hamilton Company. All rights reserved.
Microsoft, Windows and Windows XP are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries.
1.4. Trademarks
The following trademarks are acknowledged:
COBAS, LightCycler and LIFE NEEDS ANSWERS is a registered trademark of F. Hoffmann-La Roche Ltd.
All other trademarks are the property of their respective owners.
Page 10 of 175
6343 Rotkreuz
Web: www.roche-rotkreuz.com
Switzerland
E-mail: rotkreuz.gssmdi@roche.com
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Preface
Intended Use
1
1.6
1.7. Configuration
This document is intended to install a complete cobas 4800 system including the following components:
Installation Manual
Version 1.1.2
Page 11 of 175
1
1.9
Preface
Order Information
Roche
cobas 4800 system
Symbol
Numbered lists
Used for
This are sequential procedures and have to be followed step by step
>
Used to indicate the navigation path on WinXP, application software as the cobas 4800
software, cobas 4800 Work Order Editor or GRIPS > iSDoc
BOLD
BOLD
Italic
Italic
Designating another section of this manual or another manual that should be consulted.
It also can refer to iSDoc.
Tip
Indicates additional information on correct use of the system or useful tips
Safety alert. The safety alert symbol by itself without a signal word is used to promote
awareness to hazards which are generic or to direct the reader to safety information
provided elsewhere in the document.
Page 12 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Preface
Abbreviations
1
1.12
1.12. Abbreviations
Acronym
Definition
cLLD
CO-RE
CT
Chlamydia Trachomatis
CU
Control Unit
DB
Database
DCS
DWP
EDC
e.g.
HPV
Human Papillomavirus
HDD
HW
Hardware
i.e.
IQ
Installation Qualification
iSDoc
iSWAP
IVD
In Vitro Diagnostic
GRIPS
LAN
LIS
MWP
n/a
Not applicable
NG
Neissera Gonorrhoeae
OQ
Operational Qualification
OS
Operation System
PC
PreservCyt
PCB
PCR
QC
Quality Control
Installation Manual
Version 1.1.2
Page 13 of 175
1
1.13
Preface
Safety notes
Roche
cobas 4800 system
Acronym
Definition
SP
Sure Path
STAR
TADM
USB
UT
UUT
WEEE
XML
Acronym
Definition
mL
uL
Used for
Warning indicates a hazardous situation which, if not avoided, could result in death or
serious injury.
Caution indicates a hazardous situation which, if not avoided, could result in minor or
moderate injury.
The safety alert symbol by itself without a signal word is used to promote awareness to
hazards which are generic or to direct the reader to safety information provided elsewhere in
the document
Notice indicates a hazardous situation which, if not avoided, may result in property damage.
Tip indicates additional information on correct use of the system or useful tips.
Page 14 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Symbol
Preface
Safety notes
1
1.13
Used for
Warning Moving Parts indicates a possible hazardous situation. Keep hands away from
moving parts.
Warning laser transmitter. Designates the presence of a laser transmitter. Do not stare
into the laser transmitter.
Electrical Warning. This label indicates that there is a danger of coming into contact with
electrical components when gaining access to parts of the analyzer marked with this label.
Refer to Operator's Manual cobas 4800 system for instructions on safe operation.
Electrostatic sensitive devices. Observe precautions for handling electrostatic sensitive
devices.
Components of your control unit (such as the computer, monitor, keyboard) which are
marked with this symbol are covered by the European Directive on Waste Electrical and
Electronic Equipment (WEEE, 2002/96/EC).
These items must be disposed of via designated collection facilities.For more information
about disposal of your old product, please contact your city office, waste disposal service or
your Roche representative.
Constraint: It is left to the responsible laboratory organization to determine whether control
unit components are contaminated or not. If contaminated, treat them in the same way as
the instrument and analyzer.
Installation Manual
Version 1.1.2
Page 15 of 175
1
1.14
Preface
Data Security
Roche
cobas 4800 system
Used for
Data Security
Unauthorized access and data loss due to malicious software and hacker attacks
Portable storage media can be infected with and transmit computer malware, which may be
used to gain unauthorized access to data or cause unwanted changes to software.
The cobas 4800 system is not protected against malicious software and hacker attacks.
The customers are responsible for IT security of their IT infrastructure and for protecting it
against malicious software and hacker attacks. Failure to do so may result in data loss or
render the cobas 4800 system unusable.
Roche recommends the following precautions:
Page 16 of 175
Ensure that all external devices are protected by appropriate security software.
Do not copy or install any software on the cobas 4800 software data station unless
it is part of the system software or you are instructed to do so by this installation
manual or Roche Service representative.
Do not use the USB ports to connect other storage devices unless you are instructed
to do so by official user documentation or contact GSS (Global System Support)
Ensure the storage devices is free of any threats (trojan, virus, worm) before
connecting or inserting the memory media to the cobas 4800 control unit.
Run an anti-virus program with latest virus definition file on another PC then the
control unit and scan any storage devices such as USB memory stick, USB HDD, CD,
DVD, etc. against harmful software before connecting it to the control unit .
Used dedicated storage devices for the cobas 4800 control unit. Do not those
storage device on public or home computers
Keep all external storage devices in a secure place and ensure that they can be
accessed by authorized persons only.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Preface
Safety precautions
1
1.15
When handling reagents, exercise the precautions required for handling reagents.
Be sure to wear protective equipment (such as goggles, gloves).
Observe the cautions given in the package insert and observe the information the
Material Safety Data Sheets available for Roche Diagnostics reagents solutions.
If a reagent or detergent comes into contact with your skin, wash it off immediately
with water and apply a disinfectant. Consult a physician immediately.
Installation Manual
Version 1.1.2
Page 17 of 175
1
1.16
Preface
Material Required
Roche
cobas 4800 system
Available on
Operators Manual cobas 4800 system (Language: EN, DE, ES, FR,...)
1.16.2. Software
Article
Mat.No
Available on
05831563001
DCS Mannheim
06240402001
DCS Mannheim
DCS Mannheim
DCS Mannheim
File: Workaround_Start_Menu_and_Shortcuts_Vx.x.zip
N/A
File: Cobas4800WorkorderEditor.exe.config.zip
N/A
N/A
N/A
Page 18 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Preface
Material Required
1
1.16
1.16.5. Consumables
Please refer to Ordering List cobas 4800 system available on GRIPS
Installation Manual
Version 1.1.2
Page 19 of 175
2
2.1
Roche
cobas 4800 system
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
6.
7.
Select folder 2_ORACLE and then press CTRL+C keys to copy the folder
8.
Browse to C:\ and then press CRTL+V to paste the folder 2_ORACLE
9.
Select folder 2_ORACLE and then press F2 key to rename the folder 2_ORACLE into ORACLE
Browse to C:\ORACLE\Dump
Make sure that cobas 4800 software is not running
2.
Page 20 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
3.
2
2.2
5.
6.
Press Exit
Creating the backup may take a few minutes, depending on the size of the database.
Installation Manual
Version 1.1.2
Page 21 of 175
2
2.3
Roche
cobas 4800 system
7.
8.
Press Exit
Browse to D:\
2.
Select the folder Archive and then press CTRL+C to copy the folder
3.
Connect or insert a storage device to cobas 4800 control unit (e.g. NTFS formatted USB 2.0 External
Hard Drive)
Do not use any USB flash drive for large databases (e.g USB Memory Stick). Flash drives have a
very low data transfer rate.
Use only NTFS files format. FAT32 has a files size limitation of 4GB
4.
5.
After successfully storing all back up and archive files, go ahead with installing cobas 4800 software
on the new cobas 4800 control unit hp rp5700
Please see next chapter 3.4 Installation cobas 4800 control unit, page 24
followed by chapter 4 SW Installation control unit, page 25
Page 22 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
New Installation
Delivery/ Unpacking
3
3.1
3. New Installation
3.1. Delivery/ Unpacking
The cobas 4800 system is transported in palletized boxes. The boxes should be carefully inspected for any
visual damage. In case of damages, report to your local Roche office and the shipping agent before
accepting the unit. The original shipping container must be transferred unopened to the installation site.
The cobas 4800 system should be unpacked and installed only by authorized Roche Service personnel. No
attempt should be made by the user to remove, set up or operate the system without Roche service
personnel being present.
Each instrument of the cobas 4800 system is delivered with a packing list. All instrument, components,
options, accessories and consumables are listed and should be verified against the delivered items.
Installation Manual
Version 1.1.2
Page 23 of 175
3
3.3
New Installation
Installation cobas z 480 analyzer
Roche
cobas 4800 system
Page 24 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.1
2.
During start-up process (blue hp invent screen) check for the version vE1.12 in the left bottom corner
If the installed BIOS version is vE1.12 continue with the next chapter.
4.3 Verification of installed image version
If the installed BIOS version is not vE1.12 please go to chapter
16.1 Installation of BIOS settings, page 152
Installation Manual
Version 1.1.2
Page 25 of 175
4
4.3
Roche
cobas 4800 system
2.
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
3.
4.
5.
6.
7.
If the installed Image version is OSXP Recovery Image 1.0.0 and No error messages occurred
during system startup, please continue with next chapter
4.4 Windows Time and Date Adjustments
If the installed Image version is not OSXP Recovery Image 1.0.0 or an error messages occurred
during system startup, please go to chapter
16.2 Installation of OSXP Recovery Image, page 153.
Page 26 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.4
Navigate to Start > Settings > Control Panel > Date and Time Properties
2.
Set the appropriate Time Zone & Daylight Saving as well as the correct Time
Press OK to apply and close the window.
Installation Manual
Version 1.1.2
Page 27 of 175
4
4.5
Roche
cobas 4800 system
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
6.
7.
8.
Click I Agree
Page 28 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
9.
4
4.5
Installation Manual
Version 1.1.2
Page 29 of 175
4
4.5
Roche
cobas 4800 system
11. Following installation screens belong to MTS SamplePrep Method Installer and its HW drivers.
Assistance is needed to accept and confirm settings.
Click OK
Page 30 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14. cobas 4800 1.1.0.1004 SamplePrep
Click on Install
4
4.5
15. Click OK
16. Following installation screen belong to Solid-Liquid Sensors and its HW driver
Click Next
Installation Manual
Version 1.1.2
Page 31 of 175
4
4.5
Roche
cobas 4800 system
18. Following installation screens belong to CAT SH10 IVD Model Heater/Shaker and its HW driver
Click Next
Page 32 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.5
20. Click Finish to complete cobas 4800 Sample Prep Method installation
21. Following installation screen belong to cobas 4800 Transformer and its HW driver
Click Next
Installation Manual
Version 1.1.2
Page 33 of 175
4
4.5
Roche
cobas 4800 system
23. Verify that only the following components are selected to install:
;
cobas 4800 Transformer
Start Menu Shortcuts
;
Additional Debug Information
Click Next
Page 34 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.5
Installation Manual
Version 1.1.2
Page 35 of 175
4
4.6
Roche
cobas 4800 system
2.
Select folder 2_ORACLE and then press CTRL+C keys to copy the folder
3.
Browse to C:\ and then press CRTL+V to paste the folder 2_ORACLE
4.
Select folder 2_ORACLE and then press F2 key to rename the folder 2_ORACLE into ORACLE
5.
6.
7.
8.
9.
Page 36 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.6
Installation Manual
Version 1.1.2
Page 37 of 175
4
4.6
Roche
cobas 4800 system
Page 38 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.6
16. Wait for this prompt, then press any key to continue
17. Wait for this prompt, then press any key to continue
18. Wait approx. for 3 minutes for this prompt, then press any key to continue
Installation Manual
Version 1.1.2
Page 39 of 175
4
4.6
Roche
cobas 4800 system
19. Wait approx. for 1.5 minutes for this prompt, then press any key to continue
Page 40 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.7
1.
2.
3.
For more passwords please see User Access Information for cobas 4800 system, available on
GRIPS
Installation Manual
Version 1.1.2
Page 41 of 175
4
4.7
4.7.2.
Roche
cobas 4800 system
Reporting
Navigate in the tab navigation bar to Utilities > Configuration > Reporting > Parameter Value
Enter Lab Name:
xxxxxx
Enter Lab Location:
xxxxxx
Activate Show Lab Header
;
Click Apply to take over the changes
Following example shows the location of Lab Name and Lab Location on the report if
Show Lab Header ;
is activated
Page 42 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4.7.3.
1.
4
4.7
Test parameters
Navigate in the tab navigation bar to Utilities > Configuration > Test
Click in column Value to activate customer required parameters
CT/NG workflow
CT/NG PCR only work
HPV Workflow
HPV PCR only workflow
CT/NG Cytology workflow
;
;
;
;
2.
Click Exit on the global action bar (right) followed by a Yes click to confirm
3.
4.
Installation Manual
Version 1.1.2
Page 43 of 175
4
4.7
4.7.4.
Roche
cobas 4800 system
1.
2.
3.
Wait approx one more minute then start cobas_4800_V1.1.0.1036 software by double clicking on the
desktop icon
After ever start up of the cobas 4800 control unit it is required to wait approx. 1 minute before
starting cobas 4800 software to avoid the following oracle error:
ORA-01033: ORACLE initialization or shutdown in progress
If such en error occurs close all windows and start cobas 4800 software again.
If any other oracle database error message than ORA-01033 occurs during cobas 4800
software start up. Shut down and start up control unit again (no restart). Start cobas 4800
software, if error still occurs reinstall OSXP Recovery Image.
Please go to chapter 16 Appendix \ 16.2 Installation of OSXP Recovery Image, page 153
4.
If no error occurs log on as user FSE, PW: 00rocheSRVC and close software.
4.7.5.
1.
Press CTRL+ALT+DEL keys, click on Log Off and click Log Off to confirm
Press & hold Shift key to interrupt auto log on
Log on as user Install, PW: Instcobas48000709
2.
Start cobas 4800 software and log on as user FSE, PW: 00rocheSRVC
For more passwords please see User Access Information for cobas 4800 system
GRIPS > Molecular Diagnostics > Systems > cobas 4800 > cobas z 480 > Installation Manual
3.
Page 44 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4.
4
4.7
Press Exit on the cobas 4800 software global action bar (right) followed by Yes click to confirm
4.7.6.
1.
2.
3.
4.
5.
Installation Manual
Version 1.1.2
Page 45 of 175
4
4.7
Roche
cobas 4800 system
6.
7.
8.
9.
4.7.7.
1.
2.
3.
Press SHIFT+DEL keys to permanently remove the folder Oracle with all its contents followed by a
Yes click to confirm.
Page 46 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.8
2.
3.
Click Next.
4.
Click Next
Installation Manual
Version 1.1.2
Page 47 of 175
4
4.8
5.
Click Install
6.
Click Finish
Page 48 of 175
Version 1.1.2
Roche
cobas 4800 system
Installation Manual
Roche
cobas 4800 system
4
4.9
Connect or insert the storage device (i.e. usb flash drive) to cobas 4800 control unit
8.
Browse to storage device F:\ and select the Operators Manual cobas 4800 system.
Press CTRL+C keys (to copy the file)
9.
17. Start cobas 4800 Work Order Editor (double click on desktop icon)
18. Click Info button at the right bottom corner and then click the button User Manual.
The Operators Manual (Users Manual.pdf) must open now.
Verify for the correct version, if all is correct close the Manual and click Exit to close cobas 4800
Work Order Editor
If the Operators manual does not open, check for correct file name UserManual.pdf and
Users Manual.pdf
19. Press Eject CD button and remove the CD
Installation Manual
Version 1.1.2
Page 49 of 175
4
4.10
Roche
cobas 4800 system
2.
Browse to E: \Service_Software
Run file (double-click) Microlab STAR Service 4_3_0_450.exe
3.
Click Next
4.
Click Finish
6.
Page 50 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.11
2.
3.
Remove CD
Connect the cobas x 480 instrument to cobas 4800 control unit via USB cable
A 3 meter USB cable for the connection between cobas 4800 control unit and cobas x 480
instrument is provided in the cobas x 480 accessory box. Cable length more then 5 meters may
lead to signal interferences.
Cable USB 2.0 high speed, gold plated connector Type A (m)Type B (m) with 2 ferrite cores.
2.
3.
Installation Manual
Version 1.1.2
Page 51 of 175
4
4.12
4.
5.
Click Finish
Roche
cobas 4800 system
2.
3.
4.
The LightCycler 480/1536 - cobas z 480 analyzer Service Software Install Shield Wizard appears and
click Next
5.
Click Next
Page 52 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
6.
7.
The wizard will install the service software and click Finish on the following screen
8.
9.
Select the shortcut named LightCycler 480 cobas z 480 Service Software and then press
SHIFT+DEL keys to permanently delete
4
4.12
Installation Manual
Version 1.1.2
Page 53 of 175
4
4.13
Roche
cobas 4800 system
Press CTRL+ALT+DEL keys to verify Logon Information. User Install is required, click Cancel
2.
3.
4.
Right Click on folder Log and then select in the context menu Properties
5.
a) Navigate to tab Security (a), Select Group Users, in Permissions for Users in column Allow click to
select Full Control check box, and then click Apply followed by a click on Advanced
b) Under tab Permissions (b), click to select check box Replace permission entries on all child
objects with entries shown here that apply to child objects, and then click OK
6.
Page 54 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.13
2.
Right click on folder Roche_x480_ConfigFiles and then select in the context menu Properties
3.
Navigate to tab Security (a), Select Group Administrators and then Click Advanced
4.
5.
Click to deselect check box Inherit from parent permission entries that apply to child objects. Include
these with entries explicitly defined here
6.
Click COPY in the security pop up window, click OK to close the window
Installation Manual
Version 1.1.2
Page 55 of 175
4
4.13
7.
Roche
cobas 4800 system
a) Navigate to tab Security (a), Select Group Lab Method Programmer, Under Allow click to
deselect Modify and Write check box, and then click Apply
b) Navigate to tab Security (b), Select Group Lab Operator, Under Allow click to deselect Modify
and Write check box, and then click Apply
c) Navigate to tab Security (c), Select Group Lab Operator 2, Under Allow click to deselect Modify
and Write check box, and then click Apply
8.
9.
Page 56 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.14
2.
3.
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
4.
5.
6.
7.
8.
The first created problem report (FSE report) will force the system to create the folder
FSEReport for user Operator.
9.
Installation Manual
Version 1.1.2
Page 57 of 175
4
4.14
Roche
cobas 4800 system
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.15
Press CTRL+ALT+DEL keys, click on Log Off and click Log Off to confirm
2.
3.
4.
Select in the menu bar File > New > Folder and then rename it as Work Orders
5.
6.
7.
8.
Right-click the folder Work Orders, select form the context menu Create Shortcut
9.
Installation Manual
Version 1.1.2
Page 59 of 175
4
4.15
Roche
cobas 4800 system
18. Click Exit (a) on the global action bar (right) and then click Yes (b) to confirm
2.
3.
4.
5.
6.
7.
8.
9.
2.
3.
4.
5.
Page 60 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.15
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
Switch OFF the cobas x 480 instrument to force the system creating the folder FSEReport for user
Operator. The cobas 4800 system safes the created Error Report in this folder
6.
7.
8.
9.
10. Right-click the folder FSEReport, select form the context menu Create Shortcut
11. Press F2 key to rename Shortcut to FSEReport as FSEReport
12. Select and then press CTRL+X keys to cut the shortcut
13. Browse to C:\Documments and Settings\All Users\Desktop and then press CTRL+V keys to paste
the shortcut FSEReport
14. Select the shortcut FSEReport and then press CTRL+C keys to copy
15. Browse to C:\Documments and Settings\All Users\Start Menu and then press CTRL+V keys to paste
shortcut FSEReport
FSE Reports are stored in user specific folders according Windows Log In
Procedure above creates the entries in Start Menu. See chapter 4.16.1Start Menu
(OPERATOR)
If the path C:\Documents and Settings\OPERATOR\ is not visible, log on as Operator
Perform procedure according chapter 4.15.1 Create a folder to save Work Orders
If the folder FSEReport is not displayed start cobas 4800 software, switch off cobas x 480
instrument an wait for the following error messages:
Processing on cobas x 480 was aborted by the instrument
Error on cobas x 480
The first created problem report (FSE report) will force the system to create the folder
FSEReport for the user operator
Installation Manual
Version 1.1.2
Page 61 of 175
4
4.15
Roche
cobas 4800 system
2.
Press CTRL+ALT+DEL keys to verify Logon Information. User Install is required, click Cancel
3.
4.
5.
Switch OFF the cobas 4800 instrument and wait for the following error messages:
Processing on cobas x 480 was aborted by the instrument
Error on cobas x 480
The first created problem report (FSE report) will force the system to create the folder
FSEReport for the user Install
6.
7.
Right-click the folder FSEReport, select form the context menu Create Shortcut
8.
9.
10. Browse to C:\Documments and Settings\INSTALL\Start Menu and then press CTRL+V keys to paste
shortcut FSEReport (Install)
2.
3.
4.
5.
6.
Page 62 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.15
2.
3.
4.
5.
6.
2.
3.
4.
2.
3.
4.
Installation Manual
Version 1.1.2
Page 63 of 175
4
4.15
Roche
cobas 4800 system
2.
3.
4.
Browse to C:\WINDOWS\system32
2.
Right-click on the file cmd.exe, select form the context menu Create Shortcut
3.
4.
5.
6.
Page 64 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.16
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
Installation Manual
Version 1.1.2
Page 65 of 175
4
4.16
Roche
cobas 4800 system
2.
Page 66 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.16
2.
3.
4.
Installation Manual
Version 1.1.2
Page 67 of 175
4
4.16
Roche
cobas 4800 system
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
6.
In order to correct the entries in start menu, browse to C:\Documents and Settings\Install\Start
Menu Icons stored in this folder are accessible for user Install only
Page 68 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.16
2.
Installation Manual
Version 1.1.2
Page 69 of 175
4
4.17
Roche
cobas 4800 system
2.
3.
Press CTRL+ALT+DEL keys to verify logon Information. User Install is required, click Cancel.
4.
Connect or insert the storage device (i.e. usb flash drive) to the cobas 4800 control unit
5.
6.
7.
8.
9.
Page 70 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.18
2.
Double click the File, mark Select the Program from a list and choose Notepad to open the file
3.
4.
2.
3.
Press Next
Installation Manual
Version 1.1.2
Page 71 of 175
4
4.19
Roche
cobas 4800 system
4.
Check that only PreservCyt Medium can be selected in section Media Type
5.
6.
7.
Press Next
8.
Check that UT, UUT and PreservCyt Medium can be selected in section Media Type
9.
Page 72 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4
4.20
2.
Select Internet Protocol (TCP/IP) in list This connection uses the following item
Click on Properties
3.
Installation Manual
Version 1.1.2
Page 73 of 175
5
5.1
Roche
cobas 4800 system
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
2.
3.
4.
5.
6.
7.
Select folder 2_ORACLE and then press CTRL+C keys to copy the folder
8.
Browse to C:\ and then press CRTL+V to paste the folder 2_ORACLE
9.
Select folder 2_ORACLE and then press F2 key to rename the folder 2_ORACLE into ORACLE
Browse to D:\Archive
2.
Create a new folder named MTS-Backup (Menu: File > New > Folder)
3.
Connect or insert the storage device to the cobas 4800 control unit that contains the back up files.
4.
5.
Browse to D:\Archive\MTS-Backup
6.
Press CTRL+V keys to paste the folder backup files INT_ADMIN.dmp and QUARTZ.dmp
7.
Browse to C:\Oracle\Dump
8.
9.
10. Press the Run button to start the restore process of the database.
Page 74 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
5
5.2
Installation Manual
Version 1.1.2
Page 75 of 175
6
6.1
Roche
cobas 4800 system
1.
2.
Adjust springs of the DWP Holder (a), check the fixing lug (b) and knurled screw and then verify that
the cable leaves the unit a the rear side (c) and not sidewards.
3.
Carefully load it the stationary carrier on the platform at position 41-47 and move it to the very end
4.
Fix it at the rear side of the stationary carrier with knurled screw
5.
Lay the CAT H/S Module cable into rear instrument channel and bring it either at the left or right out
of the instrument. Preferably to the right, next to the cobas 4800 control unit.
Page 76 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
6.
Place the Controller Box of the CAT H/S Unit on the top of the cobas 4800 control unit or any other
suitable place. Keep in mind that the CAT H/S Unit can only be switch on and off at the Controller Box
of the CAT H/S Unit and it should be easy accessible for operator/ service engineers
7.
Fix the Controller Box of the CAT H/S Unit with a large Velcro strip and bind it together with the cobas
4800 control unit
8.
Connect the CAT H/S Unit cable to controller box (Port ST)
9.
Connect the Controller Box of the CAT H/S Unit (Port RS232) with
the cobas 4800 control unit port (COM1) by using the supported serial cable
6
6.1
13. Stow the cables either into the instrument cable channel or to the trolley frame.
Make that cables are neatly stowed, fixed and safe against unwanted disconnection.
14. Switch ON at the controller box of the CAT H/S Module
Watch init procedure for proper function. After switching on the controller box, the
heater/shaker starts turning, accelerates and stops with a noticeable click
15. The installation of the Stationary Carrier and heater/shaker unit is complete.
Installation Manual
Version 1.1.2
Page 77 of 175
6
6.2
Roche
cobas 4800 system
Clean the surface from any dust and grease at position (a) and (b)
Stick the self-adhesive cable clamp to the waste station at position (a) and (b).
2.
Mount the liquid waste tubing assay (c) to the waste station.
3.
Cut the waste tube at 1m. This length ensures that the liquid waste container can be placed on the
floor but prevents building a large siphon tube which can get clogged due to the fact of little amount
of liquid lowing through this tube.
4.
5.
Page 78 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
6.
7.
Lay the Sensor cable along the waste tube, through the to cable clamps, then through the back cable
channel to the instrument electronic rack at the left hand side of cobas x 480 instrument
8.
Plug the sensor cable into the DIV port on the left instrument panel:
9.
6
6.2
10. Neatly fix the cable to the tube by using Velcro strip or cable tie
11. Install the Tip Eject Plate according Operators Manual
Place it carefully in the notch, verify the Tip Eject Plate is in horizontal position
See picture above (step 7)
Misplacement leads to a run abort!
12. Install the Waste Chute and Solid Waste Box or alternatively waste bag
Follow the Operators Manual instruction to install the Waste Chute and Solid Waste Box
13. Switch ON cobas x 480
Installation Manual
Version 1.1.2
Page 79 of 175
6
6.3
Roche
cobas 4800 system
2.
Click Download
3.
Verify the Firmware Versions given in the screen (a) with the requested versions given in table No. 5.
4.
Addr.
Version
File Name
Master
C0
6.0SE
GRUC0S60E.ACH
Auto load
I0
3.4S
GRUALS34.ACH
iSWAP
R0
4.0S
GRUROS40.ACH
3.6Sc
GRUPXS36c.ACH
X-Drive
1.2S
GRUX0S12.ACH
X0
In case of version mismatch please update according next chapter 6.3.1 Update
Page 80 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
6.3.1.
6
6.4
Update Firmware
1.
Download and reinstall firmware if any other version is displayed as the required one
2.
3.
4.
5.
Click Start (d) to download and install the firmware on the instrument controller
6.
Verify the version with the requested version in the 6.3 Firmware Verification
7.
8.
Installation Manual
Version 1.1.2
Page 81 of 175
7
7.1
Roche
cobas 4800 system
2.
Start LightCycler 480 z 480 Service Software and log on as user: LCservice PW: LCservice
For more passwords please see User Access Information for cobas 4800 system (1.16.1
Documentation)
3.
Please compare with latest firmware versions published in iSDoc for cobas z 480 analyzer
Page 82 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
7
7.2
Start LightCycler - z 480 Service Software Log in as user: LCservice PW: LCservice
2.
Navigate to Diagnostics > Checks and perform test 1.1 Perform Run (a) according description given
in the software
3.
Press Reboot (b) to restart the cobas z 480 analyzer and then close the software
Whenever cobas z 480 analyzer is controlled by LightCycler - z 480 Service Software, reboot or
restart analyzer before the analyzer gets controlled by cobas_4800_V1.1.0.1036 software
Installation Manual
Version 1.1.2
Page 83 of 175
8
8.1
Roche
cobas 4800 system
Information
Program Barcode
Page 84 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
5
No Check Character 39
10
Installation Manual
Version 1.1.2
8
8.2
Page 85 of 175
8
8.3
Roche
cobas 4800 system
Test Barcode
Page 86 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
9
9.1
Host computer
(e.g. Note Book)
IP Address
172.20.1.100
172.20.1.140
Subnet Mask
255.255.255.0
255.255.255.0
172.20.1.1
172.20.1.1
Gateway
Connect the Computer that will host the host interface simulator (e.g. Notebook) to the LabNET
network card of the cobas 4800 control unit using CAT5e Ethernet cable.
Please see connectors overview of control unit in chapter
3.4 unit, page 24
2.
Move the cursor over the network icon to see its name.
Do not proceed to the next chapter before a physical connection could be established, indicated
by the network icon (no red x)
Installation Manual
Version 1.1.2
Page 87 of 175
9
9.2
Roche
cobas 4800 system
1.
2.
3.
4.
Click on Properties
5.
Page 88 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
6.
Select Use the following IP address, enter address listed below and click OK
IP address: 172.20.1.100
Subnet mask: 255.255.255.0
Default gateway: 172.20.1.1
7.
9.2.2.
9
9.2
The LabNET of the cobas 4800 control unit is already set to the correct Network ID 172.20.1.xxx by default.
9.2.3.
1.
Go to the cobas 4800 control unit, log on as user Install, PW: Instcobas48000709
2.
3.
Installation Manual
Version 1.1.2
Page 89 of 175
9
9.3
Roche
cobas 4800 system
1.
2.
3.
4.
Browse to E:\4_LIS_Interface
5.
6.
Click Next
7.
Click to select check box I accept the terms in the License Agreement and press Next
Page 90 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
8.
9.
Click Install
Installation Manual
Version 1.1.2
9
9.3
Page 91 of 175
9
9.3
Roche
cobas 4800 system
11. Click OK
12. Click OK
Page 92 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
9.3.2.
9
9.3
Transformer Configuration
Protocol Configuration
1.
Go to C:\Program Files\RocheDiagnosticsAG\cobas_4800_V1.1.0.1036\bin.
2.
3.
Change Transformer Type to HL7, Protocol to HL7_24 and Protocol Version to 2.4 and save
Installation Manual
Version 1.1.2
Page 93 of 175
9
9.3
Roche
cobas 4800 system
2.
9.3.3.
1.
Download the latest Host Interface Test Tool from GRIPS>cobas z 480>Host Interface Test Tool
Host Interface Test Tool 7.1.5 or higher is required
2.
3.
Execute the file setup.exe on computer that will host the host interface simulator (e.g. Notebook)
4.
Click OK
Page 94 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
5.
6.
Click Continue
7.
Click Ok.
9.3.4.
9
9.3
The communication is always established by the cobas 4800 control unit; the cobas 4800 system is always
the TCP/IP client and the LIS (Host Interface Simulator) is always the TCP/IP server.
1.
2.
3.
Installation Manual
Version 1.1.2
Page 95 of 175
9
9.3
Roche
cobas 4800 system
4.
5.
Click OK
6.
7.
Select TCP/IP Server, enter Local Port No: 5555 and click OK
8.
Page 96 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
9.
9
9.3
The server IP address is the address of the network card and cannot be changed for a server.
It comes from the network settings of the computer hosting the simulator.
Installation Manual
Version 1.1.2
Page 97 of 175
9
9.3
Roche
cobas 4800 system
Compared to the old communication server of cobas 4800 SW v1.0.x the connection happens
immediately. (before only after every 1 min a connection was attempted)
If still no connection can be established (winsock status bar remains yellow)
Check Interface Settings (Port ID) and restart Control Unit
Do not proceed to the next chapter before the winsock status bar turns green
Page 98 of 175
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
9.3.5.
9
9.3
1.
2.
Go to utilities tab (1), select configuration tab (2), select System Settings (3) and enable the checkbox
LIS present (4).
3.
Click apply.
The cobas 4800 software is now configured for LIS communication. The workoder file workflow
can be reconstituted by unchecking the LIS present check box. The change will be active after
restarting the cobas 4800 software.
Installation Manual
Version 1.1.2
Page 99 of 175
9
9.4
Roche
cobas 4800 system
Sending Results
In this communication scenario results are sent from the cobas 4800 system to the Host Interface Simulator.
2.
Open the cobas 4800 software and log on as user FSE, PW: 00rocheSRVC
3.
Go to Results tab (1), select performed run (2), select sample (3), click Export (4).
Using the result export, all results of the selected runs are exported to the LIS. It is not possible
to export individual results. No matter if just one of the results or all 94 results are selected. It is
transmitted always batch wise! Results do not have to be accepted, Export works in any case.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
4.
9
9.4
Results can be verified in the communication trace window of the HL7 Host Interface Simulator
Installation Manual
Version 1.1.2
9
9.4
9.4.2.
Roche
cobas 4800 system
Sending an order
The simulator is able to reply to an order request. For this the simulator must have certain information in its
database. The simulator will use its database records to respond to the instrument requests. If the
instrument sends a request related to a certain Sample ID, the simulator will search its database for a
matching record. If one is found, the simulator will attempt to generate an order upload, using the
information stored in the simulator database record.
1.
2.
These entries can be edited by double clicking on a data field or added, duplicated via the Records
tab.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
3.
4.
5.
6.
7.
8.
9.
Load a sample rack with one sample tube (tracks 17-34), having the same sampleID as entered in
the simulator database
9
9.4
Installation Manual
Version 1.1.2
10
10.1
Roche
cobas 4800 system
Host Server
(e.g. Note Book)
cobas IT firewall
External Network
cobas IT firewall
Internal Network
cobas 4800
control unit
(LabNet)
IP Address
10.0.0.20
10.0.0.10
172.20.1.10
172.20.1.140
Subnet Mask
255.0.0.0
255.0.0.0
255.255.255.0
255.255.255.0
Gateway
10.0.0.1
10.0.0.1
172.20.1.1
172.20.1.1
Connect the Computer that will host the host interface simulator (e.g. Notebook) to the red port
(external network) of the cobas IT firewall and connect the green port (internal network) to the LabNet
network card of the cobas 4800 control unit using 2 CAT5e Ethernet cables.
Please see compare with the connection drawing above.
Please see connectors overview of control unit in chapter
3.4 unit, page 24
2.
Verify that the red x has disappeared on the LabNET icon of the cobas 4800 control unit and on the
Local Area Connection icon of the host computer.
Move the cursor over the network icon to see its name.
Do not proceed to the next chapter before a physical connection could be established, indicated
by the network icons (no red x)
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.2
Insert a paper-clip end or a small round pin into the reset-button access port and hold it.
2.
3.
Wait until all three led simultaneously blink four times. (After approximately 30 seconds).
4.
Release the reset button and allow 1 minute for the firewall to reboot.
Connect COE client to the Orange network of the firewall with a crossover cable.
2.
Select Obtain an IP address automatically for the Local Area Connection Properties
Installation Manual
Version 1.1.2
10
10.2
3.
Roche
cobas 4800 system
Wait until COE client has obtained an IP address for the cobas IT firewall.
Check for the network icon change in the windows task bar
4.
Double Click on the network icon and check for the correctly assigned IP address 172.19.1.10
5.
Download the latest cobas IT firewall Configuration Wizard from GRIPS > cobas IT firewall >
service software
2.
Java Runtime Environment is needed to run this file which is already preinstalled on all COE
clients
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
3.
4.
Installation Manual
Version 1.1.2
10
10.2
10
10.2
5.
6.
Roche
cobas 4800 system
The external network setting entered in the text field above are for the so called red interface at
the cobas IT firewall
The green network is configured according the orange network with the next following number.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
7.
Click Next
8.
Installation Manual
Version 1.1.2
10
10.2
10
10.2
9.
Roche
cobas 4800 system
Click Next
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.2
Installation Manual
Version 1.1.2
10
10.2
Roche
cobas 4800 system
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.2
Installation Manual
Version 1.1.2
10
10.2
Roche
cobas 4800 system
17. Select save this configuration to disk and upload it to the cobas IT firewall
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.2
Check all cables and make sure the firewall was reset before uploading
Installation Manual
Version 1.1.2
10
10.2
Roche
cobas 4800 system
Change IP address of the computer that is hosting the host interface simulator.
2.
Go to the cobas 4800 control unit, log on as user Install, PW: Instcobas48000709
2.
3.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.3
2.
3.
4.
Browse to E:\4_LIS_Interface
5.
6.
Click Next
7.
Click Change
Installation Manual
Version 1.1.2
10
10.3
8.
Click Next
9.
Click Change
Roche
cobas 4800 system
11. Click OK
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
10
10.4
12. Click OK
Installation Manual
Version 1.1.2
11
11.1
Roche
cobas 4800 system
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
11
11.3
Installation Manual
Version 1.1.2
11
11.4
Roche
cobas 4800 system
HL7 2.4
Transformer Type value= ``HL7``
Protocol value= ``HL7_24``
Protocolversion value= ``2.4``
HL7 2.5.1
Transformer Type value= ``HL7``
Protocol value= ``HL7_251``
Protocolversion value= ``2.5.1``
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Printer Installation
Printer Network Configuration
12
12.1
2.
3.
Connect the printer with a CAT5e network cable to the cobas 4800 control unit LabNET Network
Adapter (PCI Network Adapter, Intel PRO/1000PT Desktop Adapter)
Please see connectors overview of control unit in chapter
3.4 Installation cobas 4800 control unit, page 24
4.
5.
Manual IP configuration can be done directly done at the printer according following sub procedure:
6.
7.
Use the arrow buttons to select Network Config, and then press OK.
Installation Manual
Version 1.1.2
12
12.2
Printer Installation
Printer Driver Installation
Roche
cobas 4800 system
8.
Use the arrow buttons to select CFG TCP/IP, and then press OK.
9.
Use the arrow buttons to select BOOTP, make sure the value is No, and
10. Use the arrow buttons to select DHCP, make sure the value is No, and
11. Use the arrow buttons to select IP BYTE 1, and then press OK.
12. Use the arrow buttons to change the value of IP BYTE 1, and then press OK.
13. Repeat steps (11) and (12) for IP BYTE 2, IP BYTE 3, and IP BYTE 4. (172.20.1.170)
Also, the Subnet Mask (SM BYTE 1, SM BYTE 2, SM BYTE 3, SM BYTE 4) (255.255.255.0)
and for default gateway (BYTE 1, GW BYTE 2, GW BYTE 3, GW BYTE 4).
14. Use the Back or Cancel button to leave the Network Config menu
To verify the printers IP address, print the network configuration page at the printer by pressing
the printer function keys, 2*OK, 6*d followed by the OK button
2.
3.
4.
5.
Select a Create a new port, in the drop down menu select Standard TCP/IP Port
and then press Next
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Printer Installation
Printer Driver Installation
6.
7.
In the next window Add Port, type in the printers IP address 172.20.1.170
Press Next
8.
Installation Manual
Version 1.1.2
12
12.2
12
12.2
9.
Printer Installation
Printer Driver Installation
Roche
cobas 4800 system
Click Finish
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Printer Installation
Printer Driver Installation
12
12.2
13. Enter any printer name for printer (i.e. HP LaserJet P2055 DN)
Select Yes to use this printer as default
Click Next
Installation Manual
Version 1.1.2
12
12.2
Printer Installation
Printer Driver Installation
Roche
cobas 4800 system
17. Click OK
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Printer Installation
Set default printer for user OPERATOR
12
12.3
Press CTRL+ALT+DEL keys, click on Log Off and click Log Off to confirm
2.
3.
4.
5.
Press CTRL+ALT+DEL keys, Click on Log Off and click Log Off to confirm
6.
7.
Installation Manual
Version 1.1.2
12
12.4
Printer Installation
Printer Installation in Laboratory Network (red Network)
Roche
cobas 4800 system
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
Goal
13
13.1
Installation Manual
Version 1.1.2
13
13.3
Water Run
Consumables
Roche
cobas 4800 system
13.3. Consumables
13.3.1. Requirement
Make following consumables available:
Item #
Article
Mat. No
Qty
11
05232716001
05232724001
04639642001
10
05232732001
05232759001
Specimen Rack:
Take a 24 pos. Sample Rack SMP_CAR_24_A00 and insert the empty tube in pos 1-6.
Fill all 6 tubes with 2mL-10ml tap water.
2.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
Get started
13
13.4
2.
Press Win key to open Start Menu and then press the key d 6 times
Press ENTER to start Microlab STAR Method Editor
3.
4.
5.
6.
In the pop up window Enter Change Description, type in short description (i.e. Installation, water run)
Click OK
7.
8.
Installation Manual
Version 1.1.2
13
13.4
Water Run
Get started
Roche
cobas 4800 system
2.
3.
4.
5.
6.
7.
Barcode Information
Test Barcode
S1WaterRun
S1WaterRunU
S2WaterRun
S2WaterRunW
S3WaterRun
S3WaterRunY
S4WaterRun
S4WaterRun[
S5WaterRun
S5WaterRun]
S6WaterRun
S6WaterRun_
8.
9.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
New Run (WaterRun)
13
13.5
2.
3.
4.
5.
Insert the rack with the tube on pos 1-6 in track 20 (tracks 17-34)
6.
Click Load Samples at right bottom corner (skip step 6-12 when using Sample tube with barcodes)
7.
8.
9.
Scan the S1WaterRun from the previous chapter 13.4.3 Creating a new work order
10. Press Enter (c) to confirm the message Double blind entry failed!
11. Press Tab key to set the cursor to Confirm barcode (d) and scan the same barcode again
12. The status changes to green. Barcode is entered manually
13. The curser is set in field (b) for pos 2. Scan-Enter-Tab. Repeat step 8-11 accordingly for sample pos
2-5
14. Click Execute to proceed. Please wait approx 1 minute
15. In section Work list click to the browse button
WaterRun
Installation Manual
Version 1.1.2
13
13.5
Water Run
New Run (WaterRun)
Roche
cobas 4800 system
2.
3.
4.
5.
2.
3.
4.
5.
6.
7.
Barcode Information
Test Barcode
Washbuffer (WB)
0S0+#"BZf
0S0110302BZ
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
New Run (WaterRun)
13
13.5
2.
3.
4.
5.
6.
Barcode Information
Test Barcode
SDS Reagent
0SA+#"BZ2
0SA110302BZ
2.
3.
4.
5.
6.
Barcode Information
Test Barcode
Lysis (LYS)
3
0S6+#"BZx
Installation Manual
Version 1.1.2
0S6110302BZ
13
13.5
Water Run
New Run (WaterRun)
Roche
cobas 4800 system
2.
3.
Scan the Reagent Barcode Magnetic Glass Particles (MGP) provided below
4.
5.
Barcode Information
Test Barcode
0S2+#"BZl
0S2110302BZ
2.
3.
4.
5.
6.
Barcode Information
Test Barcode
0S4+#"BZr
7.
0S4110302BZ
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
New Run (WaterRun)
13
13.5
Take the second prepared Sample Rack SMP_CAR_24_A00 with sample tube at pos 15, 18, 19, 23,
24
2.
Verify the Liquid Level in the Dummy Reagent tube is at 80 mm (equal to 10.5 mL for SARSTEDT
tube)
3.
4.
5.
6.
7.
Barcode Information
Test Barcode
Proteinase K (PK)
0S8+#"BZ~
Installation Manual
Version 1.1.2
0S8110302BZ
2H1+#"BZU
2H1110302BZ
0NC+#"BZ.
0NC110302BZ
2H4+#"BZ^
2H4110302BZ
2H3+#"BZ[
2H3110302BZ
13
13.5
Water Run
New Run (WaterRun)
Roche
cobas 4800 system
8.
9.
Click Start Run to start the sample preparation on cobas x 480 instrument
Click Unload and wait till all racks have been unloaded by the system
2.
Click on Abort to cancel the run. The MWP is not transferred to cobas z 480 analyzer
2.
3.
4.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Water Run
New Run (WaterRun)
13
13.5
2.
Press Win key to open Start Menu and then press the key d 6 times
Press ENTER to start Microlab STAR Method Editor
3.
4.
5.
6.
In the pop up window Enter Change Description, type in short description (i.e. Installation, water run)
Click OK
7.
8.
9.
Press CTRL+ALT+DEL keys, select Shutdown and select Restart to reboot the control unit.
Installation Manual
Version 1.1.2
14
14.1
Roche
cobas 4800 system
Article
Mat. No
Qty
05232716001
05232724001
04639642001
10
05232732001
05232759001
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14
14.3
Item
[Kit]
Item
[Kit]
Washbuffer (WB)
Proteinase K (PK)
Large Bottle
[A]
SDS Reagent
Small Bottle
Lysis (LYS)
Small Bottle
[D]
9
[E]
Small Bottle
10
[C]
[D]
[C]
[B]
2
[B]
[B]
[E]
Mat. No 05235863190
Mat. No 05235812190
Mat. No 05235782190
Mat. No 05235855190
Mat. No 05235880190
Installation Manual
Version 1.1.2
14
14.3
Roche
cobas 4800 system
2.
3.
4.
5.
6.
7.
8.
Pos
9.
Barcode Information
Test Barcode
S11TestRunHPV
S11TestRunHPVs
S22TestRunHPV
S22TestRunHPVx
S33estRunHPV
S33TestRunHPV}
S44TestRunHPV
S44TestRunHPV
S55TestRunHPV
S55TestRunHPV
S66TestRunHPV
S66TestRunHPV%
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14
14.4
2.
3.
4.
Insert the rack with the tube on pos 1-6 in track 20 (tracks 17-34)
5.
Click Load Samples at right bottom corner (skip step 6-12 when using Sample tube with barcodes)
6.
7.
8.
Scan the S1WaterRun from the previous chapter 13.4.3 Creating a new work order
9.
Press Enter (c) to confirm the message Double blind entry failed!
10. Press Tab key to set the cursor to Confirm barcode (d) and scan the same barcode again
11. The status changes to green. Barcode is entered manually
12. The curser is set in field (b) for pos 2. Scan-Enter-Tab. Repeat step 8-11 accordingly for sample pos
2- 5
13. Click Execute to proceed. Please wait approx 1 minute
14. In section Work list click to the browse button
WaterRun
Installation Manual
Version 1.1.2
14
14.4
Roche
cobas 4800 system
2.
3.
4.
5.
2.
3.
4.
5.
6.
7.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14
14.4
2.
3.
4.
5.
6.
2.
3.
4.
5.
6.
2.
3.
4.
5.
6.
2.
3.
4.
5.
6.
7.
Installation Manual
Version 1.1.2
14
14.4
Roche
cobas 4800 system
2.
3.
Take the reagent vial PK, remove the transparent cap, insert in rack pos 15
4.
Take the reagent vial PC+, remove the red cap, insert in rack pos 18
5.
Take the reagent vial NC-, remove the blue cap, insert in rack pos 19
6.
Take the reagent vial MMX, remove yellow cap, insert in rack pos 23
7.
Take the reagent vial Mg/Mn, remove green cap, insert in rack pos 24
8.
9.
10. Click Start Run to start the sample preparation on cobas x 480 instrument
Error messages related to the negative control (PC-) are correct and the cause corresponds to
the error message.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
14
14.4
Click Unload and wait till all racks have been unloaded by the system
2.
Take the MWP and apply the sealing film by using the applicator
Follow the instruction given in the Operators Manual cobas 4800 system to seal the MWP
Amplification/ Detection must be started within 90 min. After this time the cobas 4800 software
changes the MWP status to expired
Click Next
2.
Press the Load button at cobas z 480 analyzer using instrument load button
3.
Place the sealed MWP in the plate holder of cobas z 480 analyzer
4.
5.
6.
7.
Discard DWP and all reagent vials, reagent reservoirs of the cobas x 480 instrument
Installation Manual
Version 1.1.2
14
14.4
Roche
cobas 4800 system
1.
Curve HPV (+) for Channel 1-4 is expected to rise after cycle 35
2.
2.
The HPVTestRun can be stated as Valid if the following requirements are fulfilled
3.
Positive Control (PC+) and Negative Control (NC-) are stated as Valid
4.
Samples 1-6 are flagged as Invalid due to the reaction limitation of water
14.4.13. Clean up
Unload and discard the MWP of cobas z 480 analyzer
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Completion of Installation
New Run (TestRunHPV)
15
14.4
Installation Manual
Version 1.1.2
16
16.1
Appendix
Installation of BIOS settings
Roche
cobas 4800 system
16. Appendix
16.1. Installation of BIOS settings
1.
2.
Insert DVD OSXP Recovery for cobas 4800 Control Unit (IVD workflow) 1.0.0 into the DVD drive.
3.
4.
Copy Cpqsetup.txt to the root of a storage device (i.e. usb flash drive)
5.
6.
Press key [F10]during boot session to enter to enter the BIOS Screen
7.
8.
9.
10. Select Restore from Removable Media, press F10 to accept the selection.
11. Select a drive USB press F10 to accept the selection.
12. Wait for this prompt and press any key to continue.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Installation of OSXP Recovery Image
16
16.2
1.
2.
Press F9 key during start-up process (blue hp invent screen) to enter the boot menu.
3.
Insert DVD OSXP Recovery for cobas 4800 Control Unit (IVD workflow) 1.0.0 into the DVD drive.
Ensure the cobas 4800 control unit model is according the model labeled on the DVD.
i.e. HP rp 5700
4.
Installation Manual
Version 1.1.2
16
16.2
Appendix
Installation of OSXP Recovery Image
Roche
cobas 4800 system
5.
6.
Press Y key to confirm that System Disk will be overwritten and killed.
7.
Ghost Recovery starts automatically and indicates progress. It takes approx. 33 seconds.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Installation of OSXP Recovery Image
8.
After successful completion, the green Restore System Disk window is displayed again.
9.
16
16.2
Installation Manual
Version 1.1.2
16
16.2
Appendix
Installation of OSXP Recovery Image
Roche
cobas 4800 system
11. After successful completion, the green Restore System Disk window is displayed again.
12. Press key no 3 for Full Recovery Backup Disk <D:\>
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Installation of OSXP Recovery Image
16
16.2
14. After successful completion, the green Restore System Disk window is displayed again
15. Press key e to Exit
Installation Manual
Version 1.1.2
16
16.2
Appendix
Installation of OSXP Recovery Image
Roche
cobas 4800 system
20. As Computer Name, type in the hp s/n (9-digit). See nameplate on top of the cobas 4800 control unit,
press ENTER key to confirm.
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Installation of OSXP Recovery Image
16
16.2
22. Click OK and wait for automatic reboot of cobas 4800 control unit. It takes approx. 90 seconds
23. After rebooting, the system will continue with final configuration. That takes approx. 90 seconds
Just wait until the system reboots the cobas 4800 control unit one more time.
24. After this boot process the system will log on automatically as user Operator.
a) No error messages appeared. Image of DVD: OSXP Recovery Image 1.0.0 is successfully installed.
Go ahead with installing cobas 4800 software.
Please see next chapter 4.5 cobas 4800 software installation, page 28
b) Error message appeared. There is an error in the installed recovery image file.
Reinstall OSXP recovery image.
Installation Manual
Version 1.1.2
16
16.3
Appendix
Network Overview
Roche
cobas 4800 system
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Installation Check List cobas 4800 control unit
16
16.4
Page:
25
65
66
67
68
69
Verify correct placement of the Tip Eject Plate (plate is in the notch)
78
80
84
Printer installed
123
141
WaterRun
135
141
142
Page:
24
82
83
Installation Manual
Version 1.1.2
16
16.6
Appendix
Installation Check List cobas x 480 instrument
Roche
cobas 4800 system
No
No
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Barcode Information
Test Barcode
S111TestRunHPV
S111TestRunHPVi
S112TestRunHPV
S112TestRunHPVm
S113TestRunHPV
S113TestRunHPVq
S114TestRunHPV
S114TestRunHPVu
S115TestRunHPV
S115TestRunHPVy
S116TestRunHPV
S116TestRunHPV}
S117TestRunHPV
S117TestRunHPV
S118TestRunHPV
S118TestRunHPV
S119TestRunHPV
S119TestRunHPV"
10
S120TestRunHPV
S120TestRunHPVh
Installation Manual
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Barcode Information
Test Barcode
Washbuffer (WB)
Kit 24, Volume 55 mL
Storage: 15-25 C/ Load: 15-25C
0S0**5BZ*'
0S0101021BZ1000
Container 200 mL
CONS
5`&!BZ1Hf
5640601BZ10040
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Rack 2
Item #
2
Barcode Information
Test Barcode
SDS Reagent
0SA**5BZ*!g
REAG
Container 50mL
CONS
0SA101021BZ1001
5`&"BZ1"=
5640602BZ10200
Lysis (LYS)
REAG
Container 50mL
CONS
0S6**5BZ*"S
0S6101021BZ1002
5`&"BZ1"!I
5640602BZ10201
REAG
Container 50mL
CONS
5
REAG
0S2101021BZ1003
5`&"BZ1""U
5640602BZ10202
Container 50mL
CONS
Installation Manual
0S2**5BZ*#T
0S4**5BZ*$g
0S4101021BZ1004
5`&"BZ1"#a
5640602BZ10203
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Rack 3
Item #
10
Barcode Information
Test Barcode
Proteinase K (PK)
Kit 24, Volume 0.9mL
Storage:15-25 C / Load:15-25C
0S8**5BZ*%
0S8101021BZ1005
2H1**5BZ*&d
2H1101021BZ1006
0NC**5BZ*'J
0NC101021BZ1007
2H4**5BZ*(
2H4101021BZ1008
2H3**5BZ*)*
2H3101021BZ1009
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Barcode Information
Test Barcode
S121TestRunHPV
S121TestRunHPVl
S122TestRunHPV
S122TestRunHPVp
S123estRunHPV
S123TestRunHPVt
S124TestRunHPV
S124TestRunHPVx
S125TestRunHPV
S125TestRunHPV|
S126TestRunHPV
S126TestRunHPV
S127TestRunHPV
S127TestRunHPV
S128TestRunHPV
S128TestRunHPV!
S129TestRunHPV
S129TestRunHPV%
10
S130TestRunHPV
S130TestRunHPVk
Installation Manual
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Barcode Information
Test Barcode
Washbuffer (WB)
Kit 24, Volume 55mL
Storage: 15-25 C/ Load: 15-25C
0S0**5BZ**B
0S0101021BZ1010
Container 200mL
CONS
5`&!BZ1Ir
5640601BZ10041
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Rack 2
Item #
2
Barcode Information
Test Barcode
SDS Reagent
0SA**5BZ*+
REAG
Container 50mL
CONS
0SA101021BZ1011
5`&"BZ1"$m
5640602BZ10204
Lysis (LYS)
REAG
Container 50mL
CONS
0S6**5BZ*,n
0S6101021BZ1012
5`&"BZ1"%y
5640602BZ10205
REAG
Container 50mL
CONS
5
REAG
0S2101021BZ1013
5`&"BZ1"&
5640602BZ10206
Container 50mL
CONS
Installation Manual
0S2**5BZ*-o
0S4**5BZ*.
0S4101021BZ1014
5`&"BZ1"'*
5640602BZ10207
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Rack 3
Item #
10
Barcode Information
Test Barcode
Proteinase K (PK)
Kit 24, Volume 0.9mL
Storage:15-25 C / Load:15-25C
0S8**5BZ*/4
0S8101021BZ1015
2H1**5BZ*0
2H1101021BZ1016
0NC**5BZ*1e
0NC101021BZ1017
2H4**5BZ*2;
2H4101021BZ1018
2H3**5BZ*3E
2H3101021BZ1019
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Barcode Information
Test Barcode
S131TestRunHPV
S131TestRunHPVo
S132TestRunHPV
S132TestRunHPVs
S133estRunHPV
S133TestRunHPVw
S134TestRunHPV
S134TestRunHPV{
S135TestRunHPV
S135TestRunHPV
S136TestRunHPV
S136TestRunHPV
S137TestRunHPV
S137TestRunHPV
S138TestRunHPV
S138TestRunHPV$
S139TestRunHPV
S139TestRunHPV(
10
S140TestRunHPV
S140TestRunHPVn
Installation Manual
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Barcode Information
Test Barcode
Washbuffer (WB)
Kit 24, Volume 55mL
Storage: 15-25 C/ Load: 15-25C
0S0**5BZ*4]
0S0101021BZ1020
Container 200mL
CONS
5`&!BZ1J~
5640601BZ10042
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
Additional HPV Barcodes for WaterRun
16
16.7
Rack 2
Item #
2
Barcode Information
Test Barcode
SDS Reagent
0SA**5BZ*56
REAG
Container 50mL
CONS
0SA101021BZ1021
5`&"BZ1"(6
5640602BZ10208
Lysis (LYS)
REAG
Container 50mL
CONS
0S6**5BZ*6"
0S6101021BZ1022
5`&"BZ1")B
5640602BZ10209
REAG
Container 50mL
CONS
5
REAG
0S2101021BZ1023
5`&"BZ1"*N
5640602BZ10210
Container 50mL
CONS
Installation Manual
0S2**5BZ*7#
0S4**5BZ*86
0S4101021BZ1024
5`&"BZ1"+Z
5640602BZ10211
Version 1.1.2
16
16.7
Appendix
Additional HPV Barcodes for WaterRun
Roche
cobas 4800 system
Rack 3
Item #
10
Barcode Information
Test Barcode
Proteinase K (PK)
Kit 24, Volume 0.9mL
Storage:15-25 C / Load:15-25C
0S8**5BZ*9O
0S8101021BZ1025
2H1**5BZ*:3
2H1101021BZ1026
0NC**5BZ*;
0NC101021BZ1027
2H4**5BZ*<V
2H4101021BZ1028
2H3**5BZ*=`
2H3101021BZ1029
Version 1.1.2
Installation Manual
Roche
cobas 4800 system
Appendix
User Access Information
16
16.8
Log in Information
Oprcobas48000709
Limited account, auto logon
00labOPR
Log in Information
Oprcobas48000709
Limited account, auto logon
00labMGR
Log in Information
Oprcobas48000709
Limited account, auto logon
Installation Manual
00rocheADM
For IT to switch on LIS
Version 1.1.2