Model QS-550

Free-Standing
Tubestand

Service Manual

Manual Part No. DC30-009 Revision G


This manual is copyrighted and all rights are reserved. No portion of this document
may be copied, photocopied, reproduced, translated, or reduced to any electronic
medium or machine readable form without prior consent in writing from Quantum
Medical Imaging, LLC. (QMI)

This Service Manual is incomplete without Operator’s Manual DC30-034

Copyright© 2010 QMI

Quantum Medical Imaging, LLC Made in U.S.A.


2002-B Orville Drive North
Ronkonkoma, New York 11779 USA
Phone: (631) 567-5800
Fax: (631) 567-5074
E-mail: info@qmiteam.com
www.quantummedical.net
Revision History

REVISION DATE TYPE OF MODIFICATION


A 12/7/00 Initial Release.

B 2/28/01 Minor modification made to installation of tube-


stand on floor track procedure.

C 9/19/01 Incorporated ECO’s 0396, 0402, 0423.

D 6/13/02 Minor revisions to assembly and installation proce-


dure. Incorporated ECO 0788.

E 3/7/03 Incorporated ECO’s 0945, 1057.

F 3/15/04 Incorporated ECO’s 1359, 1372.

G 3/22/10 Incorporated ECO 1404, 1691, 1889, added Tech-


Vision option

Page Page Page


Number Rev Number Rev Number Rev

i - iv G
1-1 thru 1-6 G
2-1 thru 2-26 G
3-1 thru 3-6 G
4-1 thru 4-10 G
5-1 thru 5-6 G

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC i
Revision History

This product has been classified as Class I, Type B by Underwriters Laboratories, Inc.
Equipment not suitable for use in the presence of a flammable anesthetic mixture of
air with oxygen or with nitrous oxide.

MEDICAL ELECTRICAL EQUIPMENT


WITH RESPECT TO ELECTRIC SHOCK, FIRE,
MECHANICAL HAZARDS ONLY IN ACCORDANCE
WITH UL 60601-1 AND CAN/CSA C22.2 NO. 601.1
98UA

EU Authorized Representative:
Medizintechnik Berlin GmbH
Altentreptower, Strasse 59
12683 Berlin - Germany
Phone: +49-302-82 4726
Fax: +49-302-82 6382
E-mail: Medizintechnik.Berlin.GmbH@t-online.de

Revision G X-ray Tube Stand - Service Manual


ii Quantum Medical Imaging, LLC
Table of Contents

CHAPTER 1 - SPECIFICATIONS
PHYSICAL SPECIFICATIONS ................................................................. 1-3
SYSTEM SPECIFICATIONS .............................................................. 1-3
HANDGRIP SPECIFICATIONS .......................................................... 1-3
PERFORMANCE SPECIFICATIONS .................................................... 1-3
ELECTRICAL SPECIFICATIONS .............................................................. 1-4
SYSTEM OPTIONS ................................................................................ 1-4
SYSTEM OPERATING ENVIRONMENT .................................................... 1-4
NON-OPERATING ENVIRONMENT ......................................................... 1-5
COMPATIBILITY STATEMENT ............................................................... 1-5

CHAPTER 2 - ASSEMBLY & INSTALLATION


OVERVIEW .......................................................................................... 2-3
ROOM CONSTRUCTION .................................................................. 2-3
INSTALLATION PRECAUTIONS ........................................................ 2-3
CDRH CERTIFICATION ................................................................... 2-3
REQUIRED TOOLS AND MATERIALS ................................................ 2-3
UNPACKING ................................................................................... 2-4
INSTALLATION INSTRUCTIONS ...................................................... 2-4
Floor Track Installation ............................................................. 2-4
Vertical Column/Base Assembly Attachment ............................... 2-4
INSTALLING TUBESTAND ON FLOOR TRACK ................................... 2-6
INSTALLING THE X-RAY TUBE, COLLIMATOR
AND HANDGRIPS ........................................................................... 2-7
Trunnion Configuration ............................................................. 2-7
Tube Mounting Fork Configuration .......................................... 2-13
Leveling the Collimator ........................................................... 2-14
TRANSVERSE ARM COVERS INSTALLATION ............................. 2-14
VERTICAL COUNTERWEIGHT BALANCING INSTRUCTIONS ............. 2-16
VERTICAL COLUMN ROTATION ADJUSTMENT ................................ 2-17
ADJUSTING THE LONGITUDINAL BRAKE
(ELECTROMAGNETIC LOCK) ......................................................... 2-17
HORIZONTAL SID SWITCH ASSEMBLY INSTALLATION ................... 2-18
VERTICAL SID SWITCH ASSEMBLY INSTALLATION ........................ 2-20
TUBESTAND INPUT POWER .......................................................... 2-21
CONNECTION OF THE HIGH-VOLTAGE
CABLES/CABLE ROUTING ............................................................. 2-22
HORIZONTAL SID SWITCH ACTUATOR/INDICATOR
INSTALLATION ............................................................................ 2-24
VERTICAL SID SWITCH ACTUATOR INSTALLATION ....................... 2-24
VERTICAL DISTANCE SCALES ....................................................... 2-25
Bucky SID ............................................................................. 2-25
Table SID .............................................................................. 2-25
ALIGNMENT OF THE SYSTEM ....................................................... 2-26
Collimator Light Field Versus Actual X-Ray Field ........................ 2-26

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC iii
Table of Contents

CHAPTER 3 - THEORY OF OPERATION


OVERVIEW .......................................................................................... 3-3
THEORY OF OPERATION ...................................................................... 3-3
VERTICAL MOTION OPERATION ..................................................... 3-3
LONGITUDINAL MOTION OPERATION ............................................ 3-3
ANGULATION MOTION OPERATION ................................................ 3-3
TRANSVERSE MOTION TRAVEL ...................................................... 3-4
ALL LOCKS RELEASE OPERATION ................................................... 3-4
AUTO-STOP OPERATION ................................................................ 3-4
HAND GRIP DISPLAYS ................................................................... 3-5

CHAPTER 4 - SERVICE INSTRUCTIONS


OVERVIEW .......................................................................................... 4-3
SERVICE MAINTENANCE ................................................................ 4-3
Visual Inspection ..................................................................... 4-3
Perform Every Six Months: ....................................................... 4-3
Perform Annually: .................................................................... 4-4
TROUBLESHOOTING PROCEDURE .................................................. 4-4
REPLACEMENT PARTS AND ORDERING INFORMATION .................... 4-5
ORDERING INFORMATION ............................................................. 4-6

CHAPTER 5 - DIAGRAM
Model QS-550 Tubestand Sample Configuration Diagram ....................... 5-3
Model QS-550 Tubestand Wiring Diagram - Manual Collimator ................ 5-4
Model QS-550 Tubestand Wiring Diagram - Automatic Collimator
(Progeny Linear II) .............................................................................. 5-5
Model QS-550 Tubestand Wiring Diagram - Progeny Linear IV
Automatic Collimator ........................................................................... 5-6

Revision G X-ray Tube Stand - Service Manual


iv Quantum Medical Imaging, LLC
Chapter
1

SPECIFICATIONS

1-1
1-2
Chapter 1 Specifications

PHYSICAL SPECIFICATIONS
Refer to Figure 5-1 in Chapter 5, DIAGRAMS for Model QS-550 Tubestand dimensions.
SYSTEM SPECIFICATIONS
• Column Height: 92.0 inches (2336.8 mm)
• Tube Mount: Fork or Trunnion
• Weight: 600 lbs. (272.0 kg)
• Collimator
Compatibility: Manual, Selectable or Automatic, 10-20 lbs. (typi-
cal). Mounting to fork/tube housing requires four
1/4-20 or M6 x 16 bolts spaced on a 3.625" diame-
ter circle
• X-ray Tube
Compatibility: 3" or 4" rotating anode type, 30 - 40 lbs. (typical)
HANDGRIP SPECIFICATIONS
• Push Button
Switches: Vertical lock release, Longitudinal lock release,
Angulation lock release, Transverse lock release, All
Locks release, Auto-Stop
• LED Indicators: 40” (100 cm) SID, 72” (180 cm) SID

PERFORMANCE SPECIFICATIONS
• Tubestand
Longitudinal Travel: 96.0 inches (1752.6 mm)
• Vertical Travel (x-ray
tube): 60.0 inches (1524.0 mm)
• X-ray Tube Height: 72.5 inches (1848.2 mm) maximum to
13.75 inches (349.3 mm) minimum
• Transverse Tube
Travel: ±5 inches (127 mm) from centerline of table
• Transverse Travel
Detent: Center position
• X-ray Tube
Angulation: 240° (±120°)
• X-ray Tube Angulation
Detents: 0°, 90°(CW), 90°(CCW)
• Column Rotation: 180° (±90°)
• Column Rotation
Detents: 0°, 90°(L), 90°(R)

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 1-3
Chapter 1 Specifications

ELECTRICAL SPECIFICATIONS
Note: The following specifications are for the tubestand only and do not
apply to the x-ray tube or collimator. Refer to x-ray tube and collimator
manufacturer’s documentation (supplied with the tubestand) for
applicable specifications.
• Locking System
Power: +24 VDC, 1.0 Ampere
• Mode of Operation: Continuous
• Power Supply Type: Double (or Reinforced) Insulated (must meet con-
struction/performance criteria per UL 60601-1)
• Electrical
Interfacing: Tubestand-mounted terminal strip for interfacing
+24 VDC and, when equipped with automatic colli-
mator, tray sensing signals
SYSTEM OPTIONS
• Auto-Tracking System, with Motorized Vertical Motion (QS-MAT)
• Fork Mount (replaces Trunnion Mount)
• Automatic collimator interface (QS-ACT-550)
• Floor Grout Kit
• Model QS-550 tubestands equipped with TechVision Hand Grips option:
• TechVision replaces the standard equipment hand grips to provide tube
stand position controls and indicators and serves as a generator remote
control panel. Refer to the table below for the appropriate documentation for
the application:

OPERATOR SERVICE
APPLICATION TYPE
MANUAL MANUAL

Quantum(Non-integrated)
and DiRex(Integrated) DC30-048 DC30-049

Canon Integration DC30-102 DC30-049

NX-DR Integration DC30-113 DC30-049

Revision G X-ray Tube Stand - Service Manual


1-4 Quantum Medical Imaging, LLC
Chapter 1 Specifications

SYSTEM OPERATING ENVIRONMENT


Note: The tubestand is not designed for use in the presence of explosive
or flammable gases as might be found in operating rooms.
Ambient Temperature: +10°C to +40°C
Relative Humidity: 30 to 75%, non-condensing
Atmospheric Pressure: 700 hPa to 1060 hPa
NON-OPERATING ENVIRONMENT
Note: Operation or prolonged storage temperatures below 32 degrees
Fahrenheit should be avoided as it will cause lubricants to become more
resistive to manually-operated motions.
Ambient Temperature: -18°C to +70°C
Relative Humidity: 20 to 95%, non-condensing
Atmospheric Pressure: 500 hPa to 1060 hPa
COMPATIBILITY STATEMENT
The Quantum Medical Imaging, LLC tubestand is compatible with all Quantum Medi-
cal Imaging, LLC. manufactured tables, wall stands, and high-voltage x-ray genera-
tors, and with other manufacturer’s equipment having equivalent means for indication
of SID and perpendicularity.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 1-5
Chapter 1 Specifications

THIS PAGE INTENTIONALLY LEFT BLANK

Revision G X-ray Tube Stand - Service Manual


1-6 Quantum Medical Imaging, LLC
Chapter
2

ASSEMBLY &
INSTALLATION

2-1
2-2
Chapter 2 Assembly & Installation

OVERVIEW
This chapter describes the steps required to assemble and install the Model QS-550
tubestand, including making all required electrical connections, and performing all
required alignments and calibrations.
ROOM CONSTRUCTION
A conventional radiographic room construction should be used in the area where
the tubestand is to be installed. Consult with your State Health Department or
local Building Codes for specific radiation shielding requirements.
INSTALLATION PRECAUTIONS
Install the tubestand before installing the radiographic table.
Check that the space location, conduit arrangements and junction boxes are pre-
pared for the permanent installation of the tubestand. Verify that the location
and space allowed is sufficient for the full range of tubestand/Transverse Arm
travel without hitting any walls or structures in the room (see Figure 5-1 for
equipment dimensions). Note that the Transverse Arm in/out adjustment and the
Vertical Column position relative to the rear wall are critical. Column rotation in
the counterclockwise (right-hand) direction could result in the Tube Housing strik-
ing the wall before the 90 degree index position is reached.
CDRH CERTIFICATION
The tubestand is not a CDRH certifiable product. However, it must be compatible
and operate in conjunction with other components in the x-ray system so that the
x-ray system performs in compliance with CDRH performance standards.
REQUIRED TOOLS AND MATERIALS
The following tools and materials are required to complete the assembly and
installation procedures:
• Double-bubble level - two feet minimum length
• Tape Measure - eight feet minimum length
• Standard hand tools
• Alcohol - to clean the high-tension cable ends, tubehead, and high- tension
transformer sockets
• Dow-Corning DC4 vapor-proofing compound, silicon insulating grease or
equivalent - for packing x-ray tubehead and high-tension cable terminals
• Special film cassette (11" x 14" cassette with scribed centerline)
• Electric drill and drill bits
• Lubricant - white lithium grease

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-3
Chapter 2 Assembly & Installation

UNPACKING
Open crate or carton marked "packing list enclosed" first. Remove packing list
and use it as a guide to open the remaining cartons. Do not dispose of packing
material until packing list is matched with actual parts received. Should there be
a shortage or damage, notify Quantum Medical Imaging, LLC.’s customer service
department immediately. The manufacturer is relieved of any responsibility for
damage during shipment after unit is picked up by the carrier.
INSTALLATION INSTRUCTIONS
NOTE
When the tubestand is used with a vertical wall frame or wall stand, care should
Examine all
be taken during room planning and pre-installation to allow for the accurate align- cartons and
ment of these components. Alignment accuracy must be within 2% of the SID crates at time
used. After the tubestand Floor Track is aligned with the table receptor, the track of delivery. If
cannot be adjusted transversely. Therefore, transverse alignment of the x-ray damage is
field to the image receptor in the vertical wall frame or wall stand must be made apparent, have
delivery driver
by adjusting the position of the image receptor within the wall stand or relocation write a "Dam-
of the entire wall stand. aged Ship-
All mounting procedures must be approved before and checked after installation ment Note" on
copies of
by the hospital or facility engineering department and a registered, professional, freight bill, sign
structural engineer. It is recommended that a leakage current measurement be it, and file
taken following installation to verify system meets allowable levels. appropriate
carrier claim.
To assemble and install the tubestand, proceed as follows: Should you dis-
Floor Track Installation cover con-
cealed
The area of the floor on which the Floor Track is to be mounted must be damage,
level in all directions. Shims may be used, however, grouting is recom- immediately
mended. notify the
transporting
1. Position the Floor Track in the proper location for installation. agent and ask
for an "Inspec-
2. Using a bubble-level, check that the Floor Track is level both lon- tion of Dam-
gitudinally and transversely. If not, shims can be used to level age". Carrier
track, however, grout is recommended. will not accept
concealed
3. Fasten the Floor Track to the floor using appropriate fastening damage claim
hardware (capable of supporting 100 lb. pull-load) depending on if filed after 15
the installation site (see Figure 2-1). days from date
of receipt of
4. Install five (5) Hole Plugs in holes for floor mounting screws. merchandise
Vertical Column/Base Assembly Attachment
1. Remove the shipping bolt located in the end of the Vertical Col-
umn shaft.
2. Remove eight (8) screws securing the left and right side Base
Access Panels to the Base Assembly and remove panels.
3. Mount the Base Assembly to the Vertical Column and secure to
Vertical Column Shaft using the supplied retaining ring (see Fig-
ure 2-2).

Revision G X-ray Tube Stand - Service Manual


2-4 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

MOUNTING
HOLES (5X)

Figure 2-1. Floor Track Installation

RETAINING
RING

Figure 2-2. Installing the Vertical Column Retaining Ring

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-5
Chapter 2 Assembly & Installation

INSTALLING TUBESTAND ON FLOOR TRACK


1. Remove the End Cap from one end of the Floor Track.
2. At the end of Floor Track without End Cap, slide the Front and Rear Bearing
Bracket Assemblies into the Floor Track.
3. Raise the Base and Vertical Column onto the Floor Track.
4. Secure the Front and Rear Bearing Brackets to the Base Assembly (see Fig-
ures 2-3 and 2-4).

FRONT
BEARING
BRACKET

Figure 2-3. Bearing Bracket Locations

BEARING W/
ECCENTRIC
BUSHING (2X)
REAR
BEARING
BRACKET (2X)

Figure 2-4. Vertical Column Angle Adjustment

Revision G X-ray Tube Stand - Service Manual


2-6 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

CAUTION! When Bearing Brackets are removed,


extreme care must be taken not to tip over the
Base Assembly and Vertical Column. The Base
Assembly and Vertical Column should not be left
standing unattended at any time. Re-install Front
and Rear Bearing Brackets immediately once the
column is raised.

1. Re-install the End Cap to the end of the Floor Track using four (4) 5/16-18 x
1/2" hex head bolts, lockwashers and flat washers.
2. Using a two-foot (minimum) double-bubble level, placed on the front of the
Vertical Column, check that the Vertical Column is plumb. If not, adjustment
can be made by rotating the two (2) Vertical Angle Adjustment Bearings with
eccentric bushings (see Figure 2-4). Loosen the two (2) eccentric bearings
mounted to the rear of the Base Assembly; rotate the bearings to adjust. Re-
tighten the bearings after the column angle adjustment is made.
INSTALLING THE X-RAY TUBE, COLLIMATOR AND HANDGRIPS

CAUTION! The tube angulation lock is not a


mechanical lock; electrical power must be applied
to hold angulation. During installation, the x-ray
tube and collimator must be held to prevent free-
turning. Mount the collimator per instructions
supplied by the collimator manufacturer.

The system is supplied with either a Trunnion type x-ray tube mounting bracket or
a Tube Mounting Fork. Refer to the appropriate paragraph that follows corre-
sponding with the type of x-ray tube mounting bracket used on your system. On
systems equipped with optional TechVision Hand Grips, refer to TechVision Ser-
vice Manual for complete hand grips mounting instructions.
Trunnion Configuration
1. Install the Trunnion Assembly onto the Transverse Arm using four
(4) 1/4-20 x 2" hex head bolts and lockwashers (see Figures 2-5
and 2-6). There are two sets of four (4) Trunnion mounting holes
that can be used as shown in Figure 2-5. The upper set of holes
is used when mounting 3-inch x-ray tubes. The lower set of
holes is used for mounting 4-inch x-ray tubes.
2. Remove top half of Trunnion Rings.
3. Place x-ray tube onto lower half of Trunnion Rings and re-mount
upper half.
4. Install Trunnion Handle on Trunnion Assembly using two (2) 1/4-
20 x 1/2" button head socket screws (see Figure 2-7).

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-7
Chapter 2 Assembly & Installation

5. Check trunnion handle locking. If too loose, remove Trunnion


Handle and rotate each Trunnion Cam clockwise a few turns. If
too tight, remove Trunnion Handle and rotate each Trunnion Cam
counterclockwise a few turns. Re-install Trunnion Handle and re-
check locking. Repeat until properly adjusted.

Figure 2-5. Trunnion Mounting Holes

Revision G X-ray Tube Stand - Service Manual


2-8 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

TUBE SUPPORT
BRACKET

TRUNNION
ASSEMBLY

1/4-20 X 2" HEX COLLIMATOR


HEAD BOLT MOUNTING
AND LOCK- PLATE
WASHER (4X)

Figure 2-6. Mounting the Tube and Trunnion Assembly

TRUNNION
HANDLE
1/4-20 X 1/2"
BUTTON HEAD
TRUNNION SOCKET SCREW
CAM (2X)

Figure 2-7. Trunnion Handle Installed

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-9
Chapter 2 Assembly & Installation

6. Remove the Collimator Mounting Plate from the collimator.


7. Attach the Collimator Mounting Plate to the Tube Support Bracket
and x-ray tube. Use required spacer(s) supplied with collimator
between x-ray tube and Tube Support Bracket.
8. Mount Tube Support Bracket onto Trunnion using four (4) 1/4-20
hex head bolts.
9. Mount Handgrip Assembly on Tube Support Bracket (in front of
the x-ray tube) using four (4) 1/4-20 x 1" hex head bolts (see
Figure 2-8).

HANDGRIP
ASSEMBLY

1/4-20 X 1" HEX


HEAD BOLT (4X)

Figure 2-8. Installing the Handgrips


10. Rotate the Handgrips/tube 120° (i.e., nearly upside down).
Install collimator on Collimator Mounting Plate using Collimator
Mounting Ring (see Figure 2-9).
11. Affix supplied rubber bumper to Angulation Plate in location
where collimator contacts Angulation Plate when in fully rotated
position.

Revision G X-ray Tube Stand - Service Manual


2-10 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

12. Affix Trunnion angulation labels (scales and pointers) as follows:


a. Affix pointers on x-ray tube.
b. Place angulation scales on Trunnion Mounting Rings so that a
scale can be viewed from either side of the tubestand, and
with the pointers (on tube) pointing to 0° on angulation
scales.
13. Route Handgrip cable (from terminal block TB1) to D-connector
on rear side of the Handgrip Assembly and connect (see note).
Note: When installing Techvision Option, route the Ethernet Cable
from the switch or hospital LAN to J2 at rear of TechVision, the
Generator Logic Board cable to J3, and the cable from TB1 on the
Vertical Column to J5 (see Figure 2-9b). Refer to the TechVision
Service Manual (DC30-049) for additional information on install-
ing TechVision).
14. Connect collimator cables as required (refer to collimator manual
and to Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS).

COLLIMATOR
MOUNTING RING

Figure 2-9. Collimator Installed

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-11
Chapter 2 Assembly & Installation

Figure 2-9a. TechVision Option

Figure 2-9b. TechVision Electrical and Communications Connections

Revision G X-ray Tube Stand - Service Manual


2-12 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

Tube Mounting Fork Configuration


1. Position the Tube Support Bracket on the Tube Arm Extension,
with the top two slots centered on mounting holes, and secure
using four (4) 1/4-20 x 2" hex head bolts and lockwashers (see
Figure 2-10). Remove Collimator Mounting Plate from collimator
and install Collimator Mounting Plate (with appropriate spacers,
supplied with collimator) and x-ray tube on Tube Fork.
2. Mount the Handgrip Assembly (in front of the x-ray tube) using
four (4) 1/4-20 hex head bolts (see Figure 2-8).
3. Install collimator on Collimator Mounting Plate using Collimator
Mounting Ring (see Figure 2-9).
4. Route Handgrip cable (from terminal block TB1) to D-connector
on rear side of the Handgrip Assembly and connect.
5. Connect collimator cables as required (refer to collimator manual
and to Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS).

TUBE FORK

1/4-20 X 2" HEX


HEAD BOLTS (4X)

Figure 2-10. Mounting the Tube Fork

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-13
Chapter 2 Assembly & Installation

Leveling the Collimator


Place a level on top or bottom of the collimator (from front to back). If
not level, proceed as follows:
1. Loosen four (4) 1/4-20 x 2" hex head bolts securing Trunnion (or
Tube Mounting Fork) to Transverse Arm.
2. Adjust set screws as required to level collimator (see Figure
2-11).
3. Re-tighten four (4) 1/4-20 x 2" hex head bolts.

COLLIMATOR
LEVELING SET
SCREW (4X)

Figure 2-11. Collimator Leveling Adjustment Set Screws


TRANSVERSE ARM COVERS INSTALLATION
Refer to Figures 2-12 and 2-13 and proceed as follows:
1. If not already installed, position the Transverse Arm Inner Cover as shown in
Figure 2-12 and secure to Transverse Arm using six (6) 8-32 x 1/4" button
head socket screws.
2. Carefully position the Transverse Arm Outer Cover over the Transverse Arm
Inner Cover. The holes in the four (4) mounting tabs attached to the inside
bottom edge of the Transverse Arm Outer Cover should be aligned with the
holes in the four (4) mounting blocks on the underside of the Transverse Arm
Assembly.
3. Position the Transverse Arm Lower Cover as shown in Figure 2-13 and secure
it to the Transverse Arm using four (4) 6-32 x 3/4" flat head screws.

Revision G X-ray Tube Stand - Service Manual


2-14 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

TRANSVERSE ARM
INNER COVER

8-32 X 1/4" BUTTON


HEAD SOCKET
SCREW (6X)

Figure 2-12. Installing the Transverse Arm Inner Cover.

TRANSVERSE ARM
OUTER COVER

6-32 X 3/4" FLAT


HEAD SCREW (4X)
TRANSVERSE ARM
LOWER COVER

Figure 2-13. Installing the Lower Transverse Arm Cover

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-15
Chapter 2 Assembly & Installation

VERTICAL COUNTERWEIGHT BALANCING INSTRUCTIONS


1. Verify that the vertical counterweight is balanced properly, so that it provides
a smooth up and down motion of the tube head.
2. If additional weights are necessary, add trim weights as required. There are
two configurations of the Vertical Column; one has the trim weight access
opening located in the front side, the other has it located in both the front
and rear sides). To add trim weight to rear side opening, remove access
opening cover, add required trim weight(s), and re-install cover. To add trim
weight to front side opening, proceed as follows:
a. Bring the tube head down to minimum vertical height position.
b. Remove top two Column Center Cover mounting screws and carefully
tilt cover forward enough to gain access to trim weight opening (see
Figure 2-14).
c. Bring the tube head up until the top of the counterweight is in level
with bottom of opening.
d. Add trim weight as required.
e. Re-install Column Center Cover mounting screws.
NOTE
Be careful not
to drop trim
weights when
ACCESS adding through
OPENING opening in Ver-
tical Column.

Figure 2-14. Trim Weight Access Opening (Front Side of Vertical Column)

Revision G X-ray Tube Stand - Service Manual


2-16 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

VERTICAL COLUMN ROTATION ADJUSTMENT


Rotate the Vertical Column to feel the mechanical indexing. If too loose or too
tight, the tension of indexing is adjusted by loosening or tightening two (2) 1/4-
20 socket head screws, one on each side of the Vertical Column near the base
(see Figure 2-15).

1/-20 x 1 HEX BOLTS


FOR MOUNTING
LONGITUDINAL
BRAKE ASSEMBLY

LONGITUDINAL
BRAKE
ASSEMBLY

COLUMN ROTATION
ADJUSTMENT
SCREW (2X)

Figure 2-15. Vertical Column Rotation Adjustment Screws

ADJUSTING THE LONGITUDINAL BRAKE (ELECTROMAGNETIC LOCK)


The Longitudinal Brake Assembly is mounted inside the Base Assembly (on the
right side when facing the front of column). To adjust height, proceed as follows:
1. Loosen two (2) 1/4-20 x 1” hex head bolts securing the Longitudinal Brake
Assembly to the Base Assembly (see Figure 2-15).
2. The space between electromagnetic lock and contacting part of the Floor
Track should not exceed 20/1000" (ideally between 0.010” and 0.020”).
Adjust electromagnetic lock brake height as required.
3. When electromagnetic lock height is correct, re-tighten two (2) 1/4-20 x 1”
hex head bolts.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-17
Chapter 2 Assembly & Installation

HORIZONTAL SID SWITCH ASSEMBLY INSTALLATION

HORIZONTAL
SID SWITCH
ASSEMBLY

Figure 2-16. Horizontal SID Switch Installation

The Horizontal SID Switch Assembly mounts inside the Base Assembly (on the left
side when facing front of column) as shown in Figure 2-16. Proceed as follows:
1. Loosely install Horizontal SID Switch Assembly on rear wall of Base Assembly
using two (2) 10-32 x 1/2" phillips pan head screws, flat washers and kep
nuts.
2. Using a Switch Actuator placed on the Floor Track as a reference (do not peel
off adhesive backing at this time), adjust the height of the switch so that
when it contacts the actuator, the switch activates (clicks).
3. When the switch height is correct, tighten the two (2) 10-32 x 1/2" phillips
pan head screws.
4. Remove terminal strip cover from back of tubestand.

Revision G X-ray Tube Stand - Service Manual


2-18 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

5. Connect wires from Horizontal SID Switch Assembly and Longitudinal Brake
Assembly to terminal block TB1 as follows:
Table 2-1. Manual Collimator TB1 Wiring

WIRE CONNECTION
WIRE DESCRIPTION
COLOR ON TB1
LONGITUDINAL MAGNET RED TB1-6

LONGITUDINAL MAGNET BLACK TB1-2

40" SID WALL SWITCH YELLOW TB1-1

40" SID WALL SWITCH ORANGE TB1-4

72" SID WALL SWITCH GREEN TB1-1

72" SID WALL SWITCH BROWN TB1-5

40" SID TABLE SWITCH RED TB1-12

40" SID TABLE SWITCH BLACK TB1-4

6. If tubestand is being interfaced with automatic collimator, connect four (4)


wires from Horizontal SID Switch Assembly as follows:

Table 2-2. Automatic Collimator TB1 Wiring

WIRE CONNECTION
WIRE DESCRIPTION
COLOR ON TB1
LONGITUDINAL MAGNET RED TB1-6

LONGITUDINAL MAGNET BLACK TB1-2

40" SID WALL SWITCH BLUE TB1-18

40" SID WALL SWITCH PURPLE TB1-19

72" SID WALL SWITCH WHITE TB1-16

72" SID WALL SWITCH GRAY TB1-19

40" SID TABLE SWITCH BROWN TB1-17

40" SID TABLE SWITCH ORANGE TB1-20

7. Re-install terminal strip cover on back of tubestand.


8. Install left and right Base Access Panels on the Base Assembly.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-19
Chapter 2 Assembly & Installation

VERTICAL SID SWITCH ASSEMBLY INSTALLATION


The Vertical SID Switch Assembly mounts to the lower section of the Cable Clamp
(on the anode side) as shown in Figure 2-17. Proceed as follows:

VERTICAL SID
SWITCH VERTICAL
ASSEMBLY COLUMN

SID SWITCH
CABLE CLAMP CABLE
LOWER HALF

Figure 2-17. Vertical SID Switch Installation


1. Install two (2) Cable Clamp lower sections on each side of Tube Arm Exten-
sion using two (2) 1/4-20 x 2" socket head cap screws.
2. Loosely mount the Vertical SID Switch Assembly to the lower section of
anode-side Cable Clamp using two (2) 10-32 x 1/4" socket head cap screws.
3. Using a switch actuator placed on the Vertical Column as a reference (do not
peel off adhesive backing at this time), adjust the height of the switch so that
when it contacts the actuator, the switch activates (clicks).
4. When switch height is correct, tighten the two (2) 10-32 x 1/4" socket head
cap screws.
5. On SID switch cable (from terminal block TB1), connect red and black wires
to NO and COM terminals, respectively, on Vertical SID Switch Assembly
switch S7. On automatic collimator systems, connect brown and orange
wires to NO and COM terminals, respectively, on Vertical SID Switch Assembly
switch S10.

Revision G X-ray Tube Stand - Service Manual


2-20 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

TUBESTAND INPUT POWER


The tubestand requires a +24 VDC power input connection. Refer to Figure 5-2,
5-3 or 5-4 in Chapter 5, DIAGRAMS and proceed as follows:
1. Locate 25’ power input cable provided with system.
2. Connect power input cable as follows:
a. Connect end of 25’ power cable with fast-on terminals to tubestand
terminal block TB1-1 (+24 VDC) and TB1-2 (24 VDC RTN). Terminal
block TB1 is located on the rear side of the Vertical Column.
a. Connect opposite end of cable (end without fast-on terminals) to
power source (e.g., x-ray generator) +24 VDC and 24 VDC RTN ter-
minals. The input power source(s) must be double (or reinforced)
insulated and meet construction/performance criteria per UL 2601-1.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-21
Chapter 2 Assembly & Installation

CONNECTION OF THE HIGH-VOLTAGE CABLES/CABLE ROUTING


To make the required electrical connections, refer to the appropriate system inter-
connection diagram (see Figure 5-2, 5-3 or 5-4 in Chapter 5, DIAGRAMS) and
proceed as follows:
1. Connect the Anode and Cathode high-voltage cables to the x-ray tube hous-
ing. Allow sufficient cable drape (see Figure 2-18) in order to angulate the x-
ray tube 90° clockwise or counterclockwise.

CABLE
CLAMP (2X) TERMINAL
STRIP COVER

Figure 2-18. Cable Clamping and Management


2. Insert the high-voltage cables along with the cables from the collimator,
Handgrip Assembly, and Vertical SID Switch Assembly into corrugated (slit)
tubing provided with system (the tubing will need to be cut to required
lengths). Use tie-wraps to secure cable in slit tubing as required.

Revision G X-ray Tube Stand - Service Manual


2-22 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

3. Install a High Voltage Cable Strain Relief on each high voltage cable attach-
ment to x-ray tube horns. (see Figure 2-19).

HIGH VOLTAGE
CABLE
STRAIN RELIEFS
(2X)

Figure 2-19. Installing the High Voltage Cable Strain Reliefs


4. Fasten the corrugated tubing into the Cable Clamps (see Figure 18) on each
side of the Vertical Trolley by installing upper halves of Cable Clamps to lower
halves using two (2) 8-32 x 1” socket head cap screws.
5. From the Terminal Strip Assembly, route the tubestand cables along with the
+24 VDC power input cable, to the x-ray generator.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-23
Chapter 2 Assembly & Installation

HORIZONTAL SID SWITCH ACTUATOR/INDICATOR INSTALLATION


Install the Horizontal SID switches actuators and indicator decals as follows:
1. Position the tubestand at 40" and 72" SID and mark each position on Floor
Track.
2. Clean marked area with alcohol and secure actuator in marked location using
double-sided tape. Check switch height to ensure activation at 40" and 72"
SID.
3. Affix 40” and 72” (wall) SID indicators on Floor Track at each SID position
(see Figure 2-20).

Figure 2-20. Installing SID Indicator Labels


VERTICAL SID SWITCH ACTUATOR INSTALLATION
The Vertical SID Switch actuating arm faces towards the Vertical Column.
1. Attach the Vertical SID Indicator Bracket to the Vertical Trolley. This is done
by inserting the end of the bracket with the square cut-out and slot into the
opening between the trolley and Vertical Column. The square opening in the
bracket is then hung onto the inside of the trolley. Pull down on bracket to
ensure it is fully seated in slot.
2. Bring the Vertical Trolley down to the desired 40" SID mark on the Vertical
Column.
3. Mark the location for the Vertical SID Switch actuator.

Revision G X-ray Tube Stand - Service Manual


2-24 Quantum Medical Imaging, LLC
Chapter 2 Assembly & Installation

4. Clean marked area with alcohol and secure actuator in marked location using
double-sided tape.
5. Check switch height to ensure activation at 40" SID.
VERTICAL DISTANCE SCALES
The Vertical Distance Scales mount on the front side of the Vertical Column (see
Figure 2-21). The Vertical Distance Scales, calibrated in either centimeters (met-
ric system) or inches (SAE system), indicate the focal spot-to-bucky and focal
NOTE
spot-to-tabletop distances
Bucky and table Bucky SID
SID measure-
1. Set tube position for 40" (100 cm) bucky SID.
ments and
markings are 2. Affix label on Vertical Column so that the 40" (100 cm) mark lines
made sepa- up with the arrow in the bucky SID window of the SID Indicator
rately and inde- Bracket.
pendently of
each other. Table SID
1. Set tube position for 40" (100 cm) tabletop SID.
2. Affix label on Vertical Column so that the 40" (100 cm) mark lines
up with the arrow in the table SID window of the SID Indicator
Bracket.

"BUCKY"
VERTICAL SID SID INDICATOR
INDICATOR BRACKET

"TABLE"
VERTICAL SID
INDICATOR

Figure 2-21. Vertical SID Scales Mounted on Vertical Column

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 2-25
Chapter 2 Assembly & Installation

ALIGNMENT OF THE SYSTEM


Once installed, the tubestand must be aligned with other components of the x-ray
system, such as a radiographic table and/or vertical wall stand. If both a radio-
graphic table and vertical wall stand are used, align the table first, then the verti-
cal wall stand. Alignment instructions are provided in the installation and
operation manual provided with each component. NOTE
For this test, it
Refer to the collimator manual for additional alignments and performance verifi-
is very impor-
cations associated with beam limiting device. tant to use the
collimator light
Collimator Light Field Versus Actual X-Ray Field
for aligning the
1. Using any size film with cassette, set the proper technique in the table and verti-
cal wall stand.
x-ray generator.
2. Make an exposure and develop the film to verify actual x-ray field
versus collimator light field alignment is within required specifica-
tions.
3. Use collimator horizontal cross light to align the table. Rotate the
x-ray tube to check the pitch adjustment as required.
4. Verify vertical movement and alignment with bucky stand or cas-
sette holder using collimator light.

Revision G X-ray Tube Stand - Service Manual


2-26 Quantum Medical Imaging, LLC
Chapter
3

THEORY
OF OPERATION

3-1
3-2
Chapter 3 Theory of Operation

OVERVIEW
The Model QS-550 tubestand is a free-standing x-ray tube support device (i.e., does
not require wall and/or ceiling mounting) that provides a wide range of tube positions
through longitudinal, vertical, rotational, and transverse tubestand motions. The
Handgrip Assembly is designed for operator convenience with ergonomically posi-
tioned push button switches that enable tubestand movement when pressed.
The following paragraphs provide detailed theory of operation of the various tube-
stand functions. See Figure 5-2, 5-3, 5-4 in Chapter 5, DIAGRAMS for system sche-
matics.

THEORY OF OPERATION
VERTICAL MOTION OPERATION
The vertical motion of the x-ray tube is controlled by a fail-safe type "vertical
brake"; when power is removed from the system, the brake remains mechanically
locked. A +24 VDC input, typically provided from the radiographic system’s x-ray
generator, is applied through connector TB1-1 and switch S1-2 to Vertical Lock
Switch S2-1. When S2 is pressed, the +24 VDC is applied to electromagnetic Ver-
tical Lock Coil L1, releasing the brake (i.e., magnetic lock) and enabling vertical
positioning of the x-ray tube. When Vertical Lock Switch S2 is released, +24 VDC
is immediately removed from L1 and the vertical brake (magnetic lock) is acti-
vated inhibiting x-ray tube vertical motion.
LONGITUDINAL MOTION OPERATION
The longitudinal motion of the x-ray tube (i.e., entire Vertical Column) is con-
trolled by a "longitudinal brake" located in the tubestand’s Base Assembly. The
+24 VDC input is applied through connector TB1-1 and switch S1 to Longitudinal
Lock Switch S3-1. When S3 is pressed, the +24 VDC is removed from electro-
magnetic Longitudinal Lock Coil L3, releasing the brake (i.e., magnetic lock) and
enabling longitudinal positioning of the x-ray tube. When push button S3 is
released, +24 VDC is immediately applied to L3 and the longitudinal brake (mag-
netic lock) is activated inhibiting x-ray tube longitudinal motion.
ANGULATION MOTION OPERATION
The angulation motion of the x-ray tube (i.e., clockwise and counter-clockwise
rotation of the x-ray tube) is controlled by an "angulation brake" located in the
tubestand’s Transverse Arm Assembly. The +24 VDC input is applied through
connector TB1-1 and switch S1 to Angulation Lock Switch S4-1. When S4 is
pressed, the +24 VDC is removed from electromagnetic Angulation Lock Coil L2,
releasing the brake (i.e., magnetic lock) and enabling rotation of the x-ray tube.
When push button S4 is released, +24 VDC is immediately applied to L2 and the
angulation brake (magnetic lock) is activated inhibiting x-ray tube angulation
motion.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 3-3
Chapter 3 Theory of Operation

TRANSVERSE MOTION TRAVEL


The transverse motion of the x-ray tube (i.e., cross-table movement of the x-ray
tube) is controlled by a "transverse brake" located in the tubestand’s Transverse
Arm Assembly. The +24 VDC input is applied through connector TB1-1 and
switch S1 to Transverse Lock Switch S5-1. When S5 is pressed, the +24 VDC is
removed from electromagnetic Transverse Lock Coil L4, releasing the brake (i.e.,
magnetic lock) and enabling transverse motion of the x-ray tube. When push
button S5 is released, +24 VDC is immediately applied to L4 and the transverse
brake (magnetic lock) is activated inhibiting x-ray tube transverse motion.
ALL LOCKS RELEASE OPERATION
The All Locks Release button provides the operator with a single button that,
when pressed, enables x-ray tube positioning in the vertical, longitudinal, rota-
tional, and transverse directions simultaneously. The +24 VDC input is applied
through connector TB1-1 to normally-closed switch S1-2 and S1-5. With the
switch closed, the +24 VDC power input is routed through contact S1-1 to the
Transverse Lock Switch S5-1, Longitudinal Lock Switch S3-1, and Angulation Lock
Switch S4-1 and to contact S1-4, which is not connected. The +24 VDC is then
routed through switches S5, S3 and S4 to the Transverse, Longitudinal, and
Angulation Lock Coils L4, L3, and L2, respectively, which activate the locks for
each of these motions. The fail-safe Vertical lock remains locked since the +24
VDC is not connected in this switch state.
When switch S1 is pressed, the +24 VDC power input is routed to switch contacts
S1-3 and S1-4. In this state, +24 VDC input power is removed from the Trans-
verse Lock Switch S5-1, Longitudinal Lock Switch S3-1, and Angulation Lock
Switch S4-1 disabling the locks (thus enabling transverse, longitudinal, and angu-
lation x-ray tube motions). The +24 VDC power input is routed through switch
contact S1-6 to the fail-safe Vertical Lock, releasing the brake, thereby enabling
vertical motion of the x-ray tube.
AUTO-STOP OPERATION
The Auto-Stop function is provided to enable quick longitudinal positioning of the
x-ray tube at the 40" (or 100 cm) and 72 "(or 180 cm) SID positions. This func-
tion requires that the operator simultaneously press the Longitudinal Lock
Release button (switch S3) and the Auto-Stop button (switch S6). The circuit
operation is as follows:
The +24 VDC input is applied through connector TB1-1 to 40" Table SID switch
S7, 40" Wall SID switch S8, and 72" Wall SID switch S9. If the tubestand’s Verti-
cal Column is moved over a 40" or 72" SID actuator on the floor track, it closes
the corresponding SID switch, which routes the +24 VDC to the Longitudinal Lock
Coil L3. The longitudinal brake (magnetic lock) is activated, locking the x-ray
tube in the current SID position. To re-activate the Auto-Stop function, both the
Longitudinal Lock Release and Auto-Stop buttons must be pressed.

Revision G X-ray Tube Stand - Service Manual


3-4 Quantum Medical Imaging, LLC
Chapter 3 Theory of Operation

HAND GRIP DISPLAYS


The analog-type Hand Grip display includes an analog angle indicator for display-
ing the rotation angle (i.e., "angulation") of the x-ray tube, and two SID light-
emitting diodes (LEDs). When the Vertical Column is positioned over the 72" SID
actuator, located on the floor track, +24 VDC is applied from connector TB1-1
through 72" Wall SID switch S9, to the 72 Inch LED, lighting the LED. Similarly,
when the Vertical Column or the Vertical Trolley is positioned over the 40" SID
actuator, +24 VDC is applied from connector TB1-1 through 40" Wall SID switch
S8 or 40" Table SID switch S7, respectively, to the 40 Inch LED, lighting the LED.
Optionally, TechVision Hand Grips replace the standard analog-type hand grips on
systems equipped with this feature. Refer to the TechVision Service Manual for
theory of operation.

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 3-5
Chapter 3 Theory of Operation

THIS PAGE INTENTIONALLY LEFT BLANK

Revision G X-ray Tube Stand - Service Manual


3-6 Quantum Medical Imaging, LLC
Chapter
4

SERVICE
INSTRUCTIONS

4-1
4-2
Chapter 4 Service Instructions

OVERVIEW
This chapter is designed to assist the service technician in maintaining the smooth
operation of the tubestand. The tubestand will function reliably when maintained
according to the instructions provided in this chapter. Safe equipment performance
requires the attention of service personnel who are specifically trained and experi-
enced with medical x-ray apparatus. Applicable preventive maintenance or any repair
service should be performed by these skilled individuals. Failure to follow manufac-
turer’s or service personnel’s recommendations may result in serious injury.

WARNING! Personnel engaged in maintenance


activities should exercise normal caution and care
while working with electro-mechanical equip-
ment. Before removing or opening any electrical
power panels or covers, verify that the incoming
power supply is turned OFF. In the event mainte-
nance procedures require power to be supplied to
the unit, extreme care MUST be exercised to
insure the safety of service or other personnel in
the area. Verify equipment is properly grounded
before attempting any electrical operation.

SERVICE MAINTENANCE

WARNING! Always disconnect the equipment


from the main power supply prior to any cleaning.

Visual Inspection
The following inspection and adjustment procedures are recommended to
maintain the system in its original operating condition.
Perform Every Six Months:
• Check for evidence of loose hardware or loose wires
• Check all bearings and bearing surfaces for cleanliness and
corrosion
• Clean Bearing Tracks, Floor Track
• Conduct a general inspection for worn or damaged parts
• Check for proper counterbalancing and smoothness of
vertical (up/down) tubestand motion
• Verify that all ground conductors are properly and securely
installed and free of corrosion or damage
• Check all electrical cabling and wiring for wear and fraying

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 4-3
Chapter 4 Service Instructions

Perform Annually:
• Lubricate tube rotation detents, angulation plate, transverse
assembly bearing wheels, column rotation detents using
lithium grease
TROUBLESHOOTING PROCEDURE
The following table provides instructions for troubleshooting problems that could
be encountered on Model QS-550 tubestand.

Table 4-1. Troubleshooting Information

Problem/ Possible Cause(s) Remedial Action


Indication
System will not oper- No Power to the tubestand 1. Verify +24 VDC present
ate across terminal block TB1-1
and TB1-2.

40" SID or 72" SID • Loss of +24 VDC power. 1. Check for +24 VDC on hand-
indicators not function- grips cable connector A3P1-
• Horizontal Switch (40"
ing 24.
WALL SID or 72" WALL
2. Check Horizontal Switch (40"
SID) failure
WALL SID or 72" WALL SID).
• SID lamp failure 3. Replace SID lamp.

Revision G X-ray Tube Stand - Service Manual


4-4 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

REPLACEMENT PARTS AND ORDERING INFORMATION


Table 4-2 provides a list of replaceable parts for the tubestand. Figures 4-1
through 4-10 show the locations of the replaceable parts in the system.
Table 4-2. Replaceable Parts

Part Qty Used


Item Description Number on System
1 Magnet, Small (Transverse) EL58-007 1

2 Magnet, Large (Longitudinal, Angula- EL58-006 2


tion)

3 Fail-safe Brake EL80-002 1

4 SID Switch Actuator ME10-218 3 (manual col-


limator)
4 (automatic
collimator)

5 SID Switch EL50-006 2

6 SID Indicator Light, Blue EL31-003 2

7 Push-button Switch, Transverse EL50-002-1 1

8 Push-button Switch, Longitudinal EL50-002-2 1

9 Push-button Switch, All Locks Release EL50-002-3 1

10 Push-button Switch, Auto-Stop EL50-002-4 1

11 Push-button Switch, Angulation EL50-002-5 1

12 Push-button Switch, Vertical EL50-002-6 1

13 Bearing, R6ZZ ME40-018 4

14 Bearing, 1628-ZZ ME40-017 8

15 +24 VDC Power Cable EL75-005 35 ft.


(not shown)

16 Horiz. SID Switch, DPDT (automatic EL50-019 2


(not shown) collimator systems only)

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 4-5
Chapter 4 Service Instructions

ORDERING INFORMATION
To order replacement parts for the tubestand, contact the Service Department at;
Quantum Medical Imaging, LLC.
2002-B Orville Drive North
Ronkonkoma, New York 11779-7661 USA
Phone: (631) 567-5800 (x2)
Fax: (631) 567-5074
e-mail: service@qmiteam.com
When ordering replacement parts, supply the following information:
• Model and serial number of equipment
• Part number
• Part description
• Quantity required
• P.O. Number
• Shipping Instructions
When ordering components or parts not listed in Table 4-2, a complete descrip-
tion of the part, including its function and location should be provided with the
model number and serial number of the unit. Use only original replacement
parts, as supplied or recommended by Quantum Medical Imaging, LLC

1
2
(ANGULATION
MAGNET)

Figure 4-1. Model QS-550 Tubestand Parts Location Diagram

Revision G X-ray Tube Stand - Service Manual


4-6 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

2
(LONGITUDINAL
MAGNET)

Figure 4-2. Model QS-550 Tubestand Parts Location Diagram

Figure 4-3. Model QS-550 Tubestand Parts Location Diagram

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 4-7
Chapter 4 Service Instructions

5
NOTE: Vertical SID
switch and actuator
are shown in this
figure; the
Horizontal SID
switch, located in
the Base Assembly,
and longitudinal
actuators have the
same respective
part numbers.

Figure 4-4. Model QS-550 Tubestand Parts Location Diagram

6
7 10

8 11

9 12

Figure 4-5. Model QS-550 Tubestand Parts Location Diagram

Revision G X-ray Tube Stand - Service Manual


4-8 Quantum Medical Imaging, LLC
Chapter 4 Service Instructions

14

13

Figure 4-6. Model QS-550 Tubestand Parts Location Diagram

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 4-9
Chapter 4 Service Instructions

THIS PAGE INTENTIONALLY LEFT BLANK

Revision G X-ray Tube Stand - Service Manual


4-10 Quantum Medical Imaging, LLC
Chapter
5

DIAGRAMS

5-1
5-2
Chapter 5 Diagrams

(5) .750 [19] DIAMETER MOUNTING


4" [102] HOLES 28" [711] APART

61/2" [165]
1/2
18 " [470]

52.34 [1329] ±5" [127]


NOTE 1
9. DIMENSION WITH AUTO-TRACK (Q-TRACK) OPTION. 55.84±5" NOTE 9
10. WHEN ACTIVATED, RECEPTOR AND TUBE WILL
AUTOMATICALLY TRACK RECEPTOR TRAVEL. 45.50 [1156] ±5" [127]
(Q-TRACK OPTION ONLY). NOTE 1
49.00±5" NOTE 9
OF TUBESTAND 40" F.F.D.
ROTATION [1016] REF
FILM
PLANE
DELUXE 72" F.F.D.
FLOOR MOUNTED [1829] REF
TUBESTAND
QS-550
VERTI-Q
WALL STAND
REFERENCE ONLY
QUIET-LIFT TABLE, FOCAL
REFERENCE SPOT
ONLY

AUTO-TRACK
ENABLE SWITCH
72.75 MAX NOTE 10
92.00 [1848]
[2337] 40"
NOTE 8 [1016] C.G.
93.38 7.92 NOTE 7
NOTE 9
72.75 MAX.
[1848]
13.00 MIN.
32.5 [330] AUTO-TRACK
[825] ENABLE SWITCH
NOTE 10

13.00 MIN
[330]

5.00 [127] C.G. NOTE 7


NOTE 4 NOTE 2 DC80-017 Rev. D
8.50 MIN. NOTE 9 10' [3048]
17.7 [450]
C.G. LOCATION

Figure 5-1. Model QS-550 Tubestand


Sample Configuration Diagram

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 5-3
Chapter 5 Diagrams

NO 1 RED +24 VDC


2 COM BLACK BLACK +24 VDC RTN FROM
NC 3 GENERATOR

40" TABLE SID SW


S7 GREEN/YELLOW

ORANGE
D7 ANGLE LONGITUDINAL TRANSVERSE VERTICAL

BROWN
NO 1 GROUND STUD LOCK LOCK LOCK LOCK
2 COM

RED
1N4005 GENERATOR GROUND
NC 3 L2 L3 L4 L1
D6
40" WALL SID SW 1N4005 D2 D3 D4 D1
S8 NO 1
2 COM

RED

ORANGE
YELLOW

RED
TUBESTAND NC 3 1N4005 1N4005 1N4005 1N4005

RED
GREEN
ASSEMBLY 72" WALL SID SW
S9

BROWN
BLACK

BLACK

BLACK
24 VAC
FROM 24 VAC
GENERATOR GROUND

9
8
7
6
5
4
3
2
1
20
19
18
17
16
15
14
13
12

10
11
TB1

20
19
18
17
16
15
14
13
12

10
11

9
8
7
6
5
4
3
2
1
AY30-023
WHITE
COLLIMATOR BLACK
GREEN

GRY

ORN

WHT
GRN
RED

BRN
PUR

BLU

YEL

BLK
10
15

14

13

12
P1

11

5
10
15

14

13

12

11

5
J1
40 INCHES 72 INCHES
LED 1 (LEFT) LED 2 (RIGHT)

GREEN/YELLOW

BROWN
WHITE
ORANGE

YELLOW
GRAY

BLUE

BLACK
RED
PURPLE

RED
S6
AUTOSTOP SW GROUND
D5 STUD
HANDGRIP
RED
ASSEMBLY

NC
NC
1N4005 BLACK
(AY20-020)

NC

2
NO

2
S3 S4
2

S5 LONG SW ANGLE SW
S2 TRANS SW

1
VERT SW

COM
COM
1
1

COM
COM

+24V
+24V
ORN

S1
NC 1 COM
2
NO 3 SC50-002T REV. E

NC 4 COM
5
NO 6

ALL LOCKS REL

Figure 5-2. Model QS-550 Tubestand Wiring Diagram


- Manual Collimator

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 5-4
Chapter 5 Diagrams

S10
NC 1 40" TABLE SID SWITCH
2 COM
NO 3 RED +24 VDC
S7 BLACK +24 VDC RTN FROM
40" TABLE SID SW
NC 3 GENERATOR
2 COM BLACK
NO 1

BROWN

ORANGE
GREEN/YELLOW

BROWN
ORANGE
S8 D7 ANGLE LONGITUDINAL TRANSVERSE VERTICAL
NC 6 COM S9 GROUND STUD LOCK LOCK LOCK LOCK

RED
5 NC 6 COM 1N4005 GENERATOR GROUND
YELLOW NO 4 5 L2 L3 L4 L1
NO 4 D6
NC 3 COM 1N4005 D2 D3 D4 D1
2 NC 3
NO 1 2 COM

RED

ORANGE
GREEN

RED
PURPLE
TUBESTAND NO 1 1N4005 1N4005 1N4005 1N4005

RED
BLUE
ASSEMBLY

GRAY
40" WALL SID SW DPDT

WHITE
72" WALL SID SW DPDT

BROWN
BLACK

BLACK

BLACK
9
8
7
6
5
4
3
2
1
20 20
19 19
18 18
17 17
16 16
15 15
14 14
13 13
12 12

10 10
11 11
TB1

9
8
7
6
5
4
3
2
1
VERTICAL COM
Automatic LONG. COMMON AY30-023
Collimator 40" SID WALL
Interface 40" SID TABLE
Box 72" SID WALL

GRY

ORN

WHT
GRN
RED

BRN
PUR

BLU

BLK
YEL
10
15

14

13

12
P1

11

5
15

14

13

12

11

10

5
J1
40 INCHES 72 INCHES
LED 1 (LEFT) LED 2 (RIGHT)

GREEN/YELLOW

BROWN
WHITE
GRAY
ORANGE

YELLOW
BLUE

BLACK
RED
PURPLE

RED
GROUND
S6
STUD
D5
HANDGRIP
RED
ASSEMBLY

2 NC

BLACK
2 NC

BLACK
(AY20-020) 1N4005
NC

AUTOSTOP SW
NO

S3 S4
2

S5 LONG SW ANGLE SW
S2 TRANS SW

1
VERT SW
1
COM

COM

COM
1
COM

+24V
+24V
ORN

S1
NC 1 COM
2
NO 3

NC 4 COM
5
SC50-010T REV. E
NO 6

ALL LOCKS REL

Figure 5-3. Model QS-550 Tubestand Wiring Diagram


- Automatic Collimator (Progeny Linear II)

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 5-5
Chapter 5 Diagrams

40" TABLE SID MONITOR


RED +24 VDC
S7 40" TABLE SID SWITCH BLACK +24 VDC RTN FROM
R1
BLACK NC 3 GENERATOR

3
2 COM COM BLACK
POT
NO 1
GREEN/YELLOW
1K

ORANGE

BROWN
2 BROWN
S8 D7 ANGLE LONGITUDINAL TRANSVERSE VERTICAL
S9
NC 6 GROUND STUD LOCK LOCK LOCK LOCK
COM NC 6

RED
RED 5 1N4005 GENERATOR GROUND

1
YELLOW NO 4 5 COM L2 L3 L4 L1
NO 4 D6
NC 3 1N4005 D2 D3 D4 D1
P/N: EL20-022 COM
2 NC 3
NO 1 2 COM

RED

ORANGE
PURPLE

GREEN

RED
TUBESTAND NO 1 1N4005 1N4005 1N4005 1N4005

RED
BLUE
ASSEMBLY

GRAY
40" WALL SID SW DPDT
72" WALL SID SW DPDT

WHITE

BROWN
BLACK

BLACK

BLACK
QT-750 TABLE
PBL CIRCUIT

TB2
12

10
11

9
8
7
6
5
4
3
2
1
20
19
18
17
16
15
14
13
12

10
11
9
8
7
6
5
4
3
2
1

TB1

20
19
18
17
16
15
14
13
12

10
11

9
8
7
6
5
4
3
2
1
AY30-023

Collimator Interface Box (LINEAR IV)


BLACK/BROWN VERTICAL COMMON

TS4 LONG. COMMON


GREEN 40" SID WALL
15
RED RED 40" SID TABLE
14
BLACK 72" SID WALL
13
12

GRAY

ORN

WHT
GRN
PUR

RED

BRN
BLU

YEL

BLK
11
10
9

15

14

13

12

10
P1

11

5
TS5
8

15

14

13

12

10
11

5
7 1 J1
6 2 HANDGRIP 40 INCHES 72 INCHES
LED 1 (LEFT) LED 2 (RIGHT)

GREEN/YELLOW
5 3 ASSEMBLY
4 4

BROWN
AY20-020

WHITE
3 5 ORANGE

GRAY

YELLOW
BLUE

BLACK
2 6
PURPLE

RED
1 7
8
9 RED

BLACK
10

BLACK
S6 GROUND
11
D5 STUD
12
RED
13

NC
2 NC
14 AUTOSTOP SW
1N4005
2 NC
15

2
TS6
2 NO

16
S3 S4
5
4
1

17
S5 LONG SW ANGLE SW
18
S2 TRANS SW
19

COM 1

COM 1
VERT SW
20
COM 1
COLL RDY

21
COM 1

22
GND

+24V
+24V
ORN

GENERATOR S1
COM
NC 1
2
NO 3
SC50-043T REV. D

NC 4 COM
1
2
3
4
5
6
7

TB2 5
INTERFACE BOARD NO 6

ALL LOCKS REL

Figure 5-4. Model QS-550 Tubestand Wiring Diagram


- Progeny Linear IV Automatic Collimator

X-ray Tube Stand - Service Manual Revision G


Quantum Medical Imaging, LLC 5-6