PSL Document PSL 61000-4-30 CLASS A Certificate - Final - A-Eberle - PQBox100-200 - Last update 8/7/2013

PSL
Power Standards Lab
2020 Challenger Drive #100
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
A-Eberle PQ Box 100 and PQ Box 200
when equipped with the following accessories or options:
GPS satellite receiver Art. 111.9024.45

IEC 61000-4-30 Ed. 2


230V, 50/60 Hz, L-N Udin for all parameters
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation
intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Yes

Yes

Yes

Yes
Yes
Yes

Yes
Yes
(N/A)

Yes
(N/A)
(N/A)

Remarks

(N/A) Not Applicable. There is no requirement in the Standard.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document #PSL 61000-4-30 Ed 2 Test Report A-Eberle
PQBox100-200, dated 7 August 2013. PSL tested two samples, S/N 1320-101 and 1317-105, at
230VAC, 50/60 Hz. Manufacturer states that these samples are representative of the PQ Box
series.

PQ Box 100 and PQ Box 200

________________________________
Alex McEachern
7 August 2013
Alex@PowerStandards.com

Statement of IEC 61000-4-30 Compliance

PSL Document PSL 61000-4-30 CLASS A Certificate - iSTATM355 - ALSTOM - Final- Last update 8/14/2014

PSL
Power Standards Lab
2020 Challenger Drive #100
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
ALSTOM iSTAT M355
with compatible GPS receiver or NTP server

IEC 61000-4-30 Ed. 2


230V, 50/60 Hz, L-N Udin for all parameters
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation
intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Yes

Yes

Yes

Yes
Yes
Yes

Yes
Yes
(N/A)

Yes
(N/A)
(N/A)

Remarks

Tested with NTP server

(N/A) Not Applicable. There is no requirement in the Standard.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document #PSL 61000-4-30 Ed 2 Test Report iSTAT M355
ALSTOM, dated 11 July 2014. PSL tested two samples, S/N MC015976 and MC015975, at
230VAC, 50/60 Hz. Manufacturer states that these samples are representative of the iSTAT
M355 series.

iSTAT M355 - ALSTOM

_______________________________
Alex McEachern
14 August 2014
Alex@PowerStandards.com

Statement of IEC 61000-4-30 Compliance

PSL Document PSL 61000-4-30 CLASS A Certificate - Nexus 1500.doc - Last update 1/11/2010

PSL
Power Standards Lab
1201 Marina Village Parkway #101
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
Electro Industries Nexus 1500
equipped with compatible IRIG-B GPS receiver
IEC 61000-4-30 Ed. 2
230V, 50/60 Hz, L-N Udin
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)

Remarks

See Note 1 below

Yes
Yes
Yes
Yes
Yes
(N/A)

Yes
Yes
Yes
(N/A)
(N/A)

See Note 2 below

(N/A) Not Applicable. There is no requirement in the Standard.


Note 1: EUT meets Class A requirements at 230V, 50Hz.
Note 2: Transients applied to EUT measuring terminals and power terminals.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, PSL document # ELIND-001, dated 11 January 2010. PSL tested
one sample, S/N 910-0035301315 at 230VAC, 50/60 Hz. Manufacturer states that this sample is
representative of the Nexus 1500 series.

________________________________
Alex McEachern
11 January 2010
Alex@PowerStandards.com
.
Electro Industries Nexus 1500

Statement of IEC 61000-4-30 Compliance

PSL Document PSL 61000-4-30 CLASS A Certificate - Final - ELSPEC 4400 and 4500-3500.doc - Last update 8/17/2010

PSL
Power Standards Lab
1201 Marina Village Parkway #101
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
ELSPEC 4400 series (443x/442x/441x) and 4500/3500 series
IEC 61000-4-30 Ed. 2
230V, 50/60 Hz, L-N Udin
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)
Yes
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)
Yes
Yes
Yes
Yes
Yes
(N/A)
Yes
Yes
(N/A)
(N/A)

Remarks

See Note 1

(N/A) Not Applicable. There is no requirement in the Standard.


Note 1: The Elspec G4K series are not mains signaling meters.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document # PSL ELSPEC-005-30, dated 17 August 2010. PSL
tested two samples, Model G4420 S/N 00-60-35-03-46-96 and Model G4500 S/N 00-60-35-0B96-D6 at 230VAC, 50/60 Hz. Manufacturer states that these samples are representative of the
ELSPEC 4400 (443x/442x/441x) and 4500/3500 series.

________________________________
Alex McEachern
17 August 2010
Alex@PowerStandards.com
Left: ELSPEC G4420
Right: ELSPEC G4500

Statement of IEC 61000-4-30 Compliance

PSL Document FLUKE-005-30 - Last update 10/29/2010

PSL
Power Standards Lab
1201 Marina Village Parkway #101
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Ed. 2 Class A
Fluke 1760
equipped with Voltage Probes 600V and
Garmin GPS18 LVC GPS Receiver
(or other GPS receiver with equivalent accuracy and functionality)

IEC 61000-4-30 Ed. 2


230V, 50/60 Hz, L-N Udin
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class B
Compliance
Yes
Yes

(N/A)

(N/A)

Yes
Yes
Yes

Yes
Yes

(N/A)

Remarks

(N/A)

Yes
Yes
Yes
Yes
Yes
Yes

(N/A)
(N/A)

(N/A) Not Applicable. There is no requirement in the Standard.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document # PSL FLUKE-003-30a, dated 17 May 2007 and
Addendum 1 for Ed. 2 Update, dated 29 October 2010. PSL tested one sample, S/N
V7-59941-IC at 230VAC, 50/60 Hz. Manufacturer states that this sample is representative of the
FLUKE 1760 series.

________________________________
FLUKE 1760

Alex McEachern
29 October 2010
Alex@PowerStandards.com

Statement of IEC 61000-4-30 Compliance

PSL Document PSL-001-30 - Last update 7/8/2011

PSL
Power Standards Lab
2020 Challenger Drive #100
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A, Class S, and Class B
PSL PQube
with Firmware 1.5 or later

IEC 61000-4-30 Ed. 2


230V 50Hz, L-N Udin
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation
intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
All except
150/180
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
All except
150/180
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)
Yes
Yes
Yes
Yes
Yes
-

Remarks
See Section 4.4
At 230V 50Hz

See Section 4.4


See Note 4
See 5.8
See Note 1
See Note 2

Yes
Yes
(N/A)
(N/A)

See Note 3
Pass
Pass

Note 1: PQube is not a mains signaling meter


Note 2: Underdeviation and overdeviation are not recorded
Note 3: SNTP time synchronization available, 2 seconds guaranteed. Meets Class A backup requirements after losing synchronization
Note 4: Measurements are correct and accurate, methods are correct but not gapless. See 4.4.
(N/A) Not Applicable. There is no requirement in the Standard.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document # PSL PSL-001-30, dated 8 February 2011. PSL tested
S/N P002887 at 230V 50Hz. Manufacturer states this sample is representative of the PQube.

________________________________
Alex McEachern
8 February 2011
Alex@PowerStandards.com
.
PSL PQube

Statement of IEC 61000-4-30 Compliance

PSL Document SIEMENS-009-30 - Last update 8/27/2009

PSL
Power Standards Lab
1201 Marina Village Parkway #101
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
Siemens SIMEAS Q80
equipped with Garmin GPS18x LVC
(or other GPS receiver with equivalent accuracy and functionality)
IEC 61000-4-30 Ed. 2
230V, 50/60 Hz, L-N Udin
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
-

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
-

Class B
Compliance
Yes
Yes

Yes

4.4

Measurement aggregation intervals

Yes

No

4.6
4.7
6.1

Time-clock uncertainty
Flagging
Transient influence quantities

Yes
Yes
Yes

Yes
Yes
(N/A)

(N/A)

Remarks

See Note 1 below

Yes
Yes
Yes
Yes
Yes
Yes

See Note 2 below


Class A and Class S are
mutually exclusive

Yes
(N/A)
(N/A)

See Note 3 below

(N/A) Not Applicable. There is no requirement in the Standard.


Note 1: Flicker is only defined at 230V, 50Hz and 120V, 60Hz. EUT meets Class A requirements at 230V, 50Hz.
Note 2: Overdeviation and underdeviation parameters are not measured by the Siemens SIMEAS Q80.
Note 3: Transients applied to EUT measuring terminals and power terminals.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document # PSL SIEMENS-009-30, dated 27 August 2009. PSL
tested two samples, S/N 140148 and 140149 at 230VAC, 50/60 Hz. Manufacturer states that
these samples are representative of the SIMEAS Q80 series.

________________________________
Alex McEachern
27 August 2009
Alex@PowerStandards.com
.
Siemens SIMEAS Q80

Statement of IEC 61000-4-30 Compliance

PSL Document PSL

61000-4-30 CLASS A Certificate - PQM-702, 703, 710 and 711 - Sonel_Final Last update 11/17/2014

PSL
Power Standards Lab
2020 Challenger Drive #100
Alameda, CA 94501 USA
TEL ++1-510-522-4400
FAX ++1-510-522-4455

www.PowerStandards.com

Certificate of Conformity
IEC 61000-4-30 Class A
SONEL PQM-702, PQM-703, PQM-710 & PQM-711 Series
with compatible GPS antenna

IEC 61000-4-30 Ed. 2


230V, 50/60 Hz, L-N Udin for all parameters
61000-4-30
Section
5.1
5.2
5.3
5.4
5.5
5.7
5.8
5.9
5.10
5.12
4.4
4.6
4.7
6.1

Power Quality Parameter


Power frequency
Magnitude of the supply voltage
Flicker
Supply voltage dips and swells
Voltage interruptions
Supply voltage unbalance
Voltage harmonics
Voltage interharmonics
Mains signaling voltage
Underdeviation and overdeviation
Measurement aggregation
intervals
Time-clock uncertainty
Flagging
Transient influence quantities

Class A
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes

Class S
Compliance
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Class B
Compliance
Yes
Yes
(N/A)
Yes
Yes
Yes
Yes
Yes
Yes
(N/A)

Yes

Yes

Yes

Yes
Yes
Yes

Yes
Yes
(N/A)

Yes
(N/A)
(N/A)

Remarks

Tested with GPS Antenna


Tested PQM 711

(N/A) Not Applicable. There is no requirement in the Standard.

This certificate summarizes the results of the PSL IEC 61000-4-30 Power Quality Measurement
Methods Compliance Report, document #PSL 61000-4-30 Ed 2 Test Report Sonel PQM dated
30 September 2014. PSL tested two samples, Model PQM-702 S/N AZ0191 & PQM-711 S/N
BS0009, at 230VAC, 50/60 Hz. Manufacturer states that these samples are representative of the
PQM-702, PQM-703, PQM-710 & PQM-711 series.

PQM-702, 703, 710 & 711 - Sonel

______________________________
Alex McEachern

30 September, 2014

Alex@PowerStandards.com
Statement of IEC 61000-4-30 Compliance

K
KALIB
BRERIINGSB
BEVIS
S
utffrdat av riksmtplats 01
CA
ALIBRATION CERTIF
FICATE issued by a Sw
wedish National Labooratory
Contact person

Date

Page

Sttefan Svenssson

2012
2-03-23

1 (2))

Measurement Teechnology
M
+446 10 516 54 155
stefan.svensson@
@sp.se

Referencee

MTeePX22199-K
K06 rev1
Unip
power AB
P. O.. Box 411
SE-4
441 28 Alingss
Swed
den

C
Calibratio
on of Freq
quency
R 1: Summ
Rev
mary includedd, editorial changes
c

I
Identifica
ation
O
Object
Object state
O
C
Calibration
loocation
C
Calibration
d
date

Power Qualitty meter, Uniipower UP-22210R, Seriaal number


P
2
22120402
U
Upon
arrival the object haad no visual damage.
B
Bors
2
2012-02-16

M
Measurem
ment meth
hods and procedurres
The instrumeent was calibbrated by com
T
mparison with
h the electriccal voltage annd frequency
y
s
standards
of SP.

M
Measurem
ment cond
ditions
Room temperature
R
W
Warm-up
tim
me

223.0 2.0 C
C
> 12 h (conneected to main
ns)

R
Results
The result is presented beelow as the error
T
e
expand
ded uncertainnty, or as sett value, meassured
v
value
and exppanded unceertainty. The reported exp
panded uncerrtainty of meeasurement iss stated
a the standarrd uncertaintty of measurrement multip
as
plied by the coverage facctor k = 2, wh
hich for
a normal disttribution corrresponds to a coverage probability off approximattely 95%. Th
he
s
standard
unceertainty has been
b
determined in accorrdance with EA
E Publicatiion EA-4/02. The
l
long
term staability of the calibrated obbject is not in
ncluded in thhe reported eexpanded unccertainty
o measurem
of
ment.

T
Tables
of thee result are foound on folloowing pages..

S Technical Research Ins


SP
stitute of Swe
eden
Poostal address

Office locationn

Phonne / Fax / E-mail

SP
P
Boox 857
SE
E-501 15 BORS
Sw
weden

Vstersen
Brinellgatan 4
B
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@
@sp.se

National Laborattories are designateed by the Swedish Government


according to the Act (1992:1119) annd the Decree (19993:1065) concerning
testing and metroology. SP operatess under ISO 17025, supervised by
SWEDAC. This document
d
may not be reproduced other than in full, exceept
with the prior writtten approval of SP
P.

KALIBRERIN
NGSBEVIS
CA
ALIBRATION CERTIFICA
ATE

Date

Page

2012
2-03-23

2 (2))

Referencee

MTeePX22199-K
K06 rev1

S
Supply
frequuency

Voltage
(V)

Set Supply
frequency
(Hz)

Readingg EUT
(Hzz)

Expanded
uncertainty
u
(Hz)

42.500
50.000
57.500

42.5001
50.0001
57.5000

0.001
0.001
0.001

63.5

S
Summary:
Inn the tested points
p
the resuults are in co
ompliance with IEC 610000-4-30 Classs A.

T
Traceabil
lity
SP is Nationaal Laboratoryy for electriccal quantitiess and time annd frequency by appointm
S
ment of
t Swedish government.. SP realizes fundamental units such as
the
a volt, ohm and second from
p
primary
standdards. Traceaability for otther units is established
e
fr
from
these realizations by
y means
o in-house calibrations
of
c
a scientificc analyses. To ensure inteernational eqquivalence an
and
nd
a
acceptance
of the establisshed traceabiility, interlab
boratory com
mparisons aree made betweeen
n
national
labooratories.

E
Equipmen
nt
Voltage amplifier Fluke 5215A
V
5
s/n 4555014
F
Frequency
coounter Fluke PM6680B SPno
S
502997
7
D
DAC
Iotech DAC488HR
R/2, SPno 5022132

S Techniccal Researcch Institute of Sweden


SP
n
M
Measuremen
nt Technoloogy - Electriccity
P
Performed
by
Signed by: Stefan Svensson
Reason: I am the author of this document
Date & Time: 2012-04-13 13:54:23 +02:00
___Signature_1

S
Stefan
Svenssson

S Technical Research Ins


SP
stitute of Swe
eden

__S
Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2012-03-23

1 (3)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX22199-K01 rev1
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Voltage and Unbalance


Rev 1 changes due to calculation error, summary included

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2012-02-16

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature
Warm-up time

23.0 2.0 C
> 12 h (connected to mains)

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.

Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

2 (3)

Reference

MTePX22199-K01 rev1

Supply voltage

Voltage

(Hz)
50

Set
Supply
Voltage
(V)

Reading
EUT-L1

Reading
EUT-L2

Reading
EUT-L3

Expanded
uncertainty

(V)

(V)

(V)

(V)

95.25
63.5
19.0
6.3

95.278
63.521
19.005
6.304

95.288
63.526
19.009
6.306

95.294
63.528
19.010
6.306

0.020
0.020
0.010
0.006

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.
Unbalance

Nom
Voltage
(V)

Set
Unbalance
(%)

Reading EUT
(%)

Expanded
uncertainty
(%)

63.5

5.05*
4.95**

5.02
4.98

0.01
0.01

* In accordance with IEC 61000-4-30 table 2,testing state 2


** In accordance with IEC 61000-4-30 table 2, testing state 3
Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

3 (3)

Reference

MTePX22199-K01 rev1

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
Voltage amplifier Fluke 5215A s/n 455014
Frequency counter Fluke PM6680B SPno 502997
DAC Iotech DAC488HR/2, SPno 502132
Reference ZERA COM303-3, s/n 503044
Three phase generator HEG System_4200, s/n 503048

SP Technical Research Institute of Sweden


Measurement Technology - Electricity
Performed by
Signed by: Stefan Svensson
Reason: I am the author of this document
Date & Time: 2012-04-13 13:52:54 +02:00
__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2014-07-01

1 (2)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTe4P04418-K01
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Voltage Unbalance


Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2014-06-11

Measurement methods and procedures


The instrument was calibrated by comparison with the standards of SP.

Measurement conditions
Room temperature

23.0 1.0C

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.

Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (2011:791) and the Decree (2011:811) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-07-01

2 (2)

Reference

MTe4P04418-K01

Calibration of unbalance
The calibration is an extension of earlier calibration with certificate number MTePX22199K01

Nom
Voltage
(V)

Set
Unbalance
(U2%)

Set
Unbalance
(U0%)

Reading EUT

Reading EUT

(U2%)

(U0%)

Expanded
uncertainty
(%)

60
60

0.01*
2.22**

0,01
2,24

0.06
2.33

0,05
2.34

0.01
0.01

* Corresponding to IEC 61000-4-30 table 2, testing state 1


** Approximately half the value of IEC 61000-4-30 table 2, testing state 2
Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

Equipment
Three phase comparator Zera Com303-3, SPno 503044

SP Technical Research Institute of Sweden


Measurement Technology - Electricity
Performed by

__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2012-03-23

1 (2)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX22199-K04 rev1
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Flicker
Revision 1: summary included.

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2012-02-16

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature
Warm-up time
Frequency

23.0 2.0 C
> 12 h (connected to mains)
50.0 0.1 Hz

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)/true value*100 %.
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

2 (2)

Reference

MTePX22199-K04 rev1

Flicker at 63.5 V, values in accordance with SS_EN 61000-4-15

Voltage

(V)

Modulation
frequency
(Changes per
minute)

Set Pst

Reading EUT

Expanded
uncertainty

(-)

(-)

(-)

4000
4000
1620
1620
110
7
1
1
1

1.000
10.00
1.000
0.200
1.000
1.000
10.00
1.000
0.200

1.010
9.887
0.992
0.200
1.013
1.027
10.383
1.022
0.200

0.010
0.030
0.010
0.010
0.010
0.010
0.030
0.010
0.010

63.5

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

Equipment
Voltage amplifier Fluke 5215A s/n 455014
DAC Iotech DAC488HR/2, SPno 502132
Software RectangularFlicker ver 1.2

SP Technical Research Institute of Sweden


Measurement Technology
Performed by
Signed by: Stefan Svensson
Reason: I am the author of this document
Date & Time: 2012-04-13 13:53:39 +02:00
__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2012-03-23

1 (2)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX22199-K02 rev1
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Dips, Swells


Rev 1: Changes due to calculation error, summary included

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2012-02-16

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature
Warm-up time
Frequency

23.0 2.0 C
> 12 h (connected to mains)
50.0 0.1 Hz

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)/true value*100 %.
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

2 (2)

Reference

MTePX22199-K02 rev1

Dips and swells, from 60 V, start of dip/swell at zero degrees.

Voltage
(V)

Set
Dip/Swell
level
(%)

Set
Dip/swell
length
(ms)

Read
Dip/swell
length
(ms)

85
50
115

100
100
100

100
100
120

63.5

Read level EUT


(%)

Expanded
uncertatinty
(%)

85.01
50.01
115.06

0.01
0.01
0.01

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
Voltage amplifier Fluke 5215A s/n 455014
DAC Iotech DAC488HR/2, SPno 502132

SP Technical Research Institute of Sweden


Measurement Technology - Electricity
Performed by
Signed by: Stefan Svensson
Reason: I am the author of this document
Date & Time: 2012-04-13 13:52:10 +02:00
__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

REPORT
Contact person

Date

Bengt Andersson

2014-10-09 4P06655-E1

Reference

Page

1 (3)

Electronics
+46 10 516 56 55
bengt.andersson@sp.se

Unipower AB
Box 411
441 28 Alingss

Verification of Power Quality meter Unipower UP-2210R


Test object
Power Quality meter type Unipower UP-2210R
Serial number: 22120402

Commission
Verification of the voltage interruption detection function and interruption time measurement
according to EN 61000-4-30:2009 / IEC 61000-4-30:2008, clause 5.5.2.
Interruption time to be verified:
Interruption level:
Interruption threshold set to:
Interruption hysteresis set to

2000 ms
5% and 15%
10 %
2%

Result summary
The requirements according to EN/IEC 61000-4-30 Class A were fulfilled in the tested points.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

This document may not be reproduced other than in full, except with the
prior written approval of SP.

REPORT

Date
2014-10-07

Date

Reference

2014-10-09 4P06655-E1

Temperature
23 C 3 C

Page

2 (3)

Humidity
31 % 5 %

Measurement equipment
Variable voltage supply HP 6813B
Voltage interruption generator T 3000
Multimeter Fluke 83
Variable transformer
Temperature and humidity meter Testo 615

503 091
901 284
503 418
503 518
503 418

Test set-up and procedure


The power supply HP 6813B was adjusted to 65.0 V 50 Hz.
The variable transformer was connected to the 65 V source and adjusted to
3.25 0.20 V respective 9.75 0.20 V (5 % resp 15% of 65 V).
The output of the power supply and the transformer was connected to the interruption
generator T 3000. The interruption generator was set to specified switch time, with switch
between the voltage sources at zero crossing of the voltage. The voltage was steady state
65.0 V 60 s before the voltage dip.
The output voltage from the interruption generator was connected to all phases of the test
object together, i.e. simultaneous voltage change on all phases of the voltage input of the test
object and no phase shift.
The read out from the test object was displayed by an external computer. The client was
responsible for the for the operation of the external computer. The test result was saved on the
log file SP2S15P5P UP2210 on the clients computer.
The test set-up can be seen in the picture below:

SP Technical Research Institute of Sweden

REPORT

Date

Reference

2014-10-09 4P06655-E1

Page

3 (3)

Result
Voltage dip level Voltage dip time
9.75 V (15 %)
2 000 ms
3.25 V (5 %)
2 000 ms

Result
Not detected as voltage interruption (> 10 %)
Detected interruption.
Registered interruption time 2.010s

The required dip duration uncertainty according to EN/IEC 61000-4-30


clause 5.4.5.2 is 20 ms.
The registered interruption time was within the requirement as specified in
EN/IEC 61000-4-30.

SP Technical Research Institute of Sweden


Electronics - EMC
Performed by

Examined by

__Signature_1

__Signature_2

Bengt Andersson

SP Technical Research Institute of Sweden

Tomas Bodeklint

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2012-03-23

1 (6)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX22199-K03 rev2
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Harmonics
Rev1: Editorial changes, summary included, rev2: editorial.

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2012-02-16

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature
Warm-up time

23.0 2.0 C
> 12 h (connected to mains)

Results
The result is presented below as the error expanded uncertainty, or as set value,
measured value and expanded uncertainty. The reported expanded uncertainty of
measurement is stated as the standard uncertainty of measurement multiplied by the
coverage factor k = 2, which for a normal distribution corresponds to a coverage
probability of approximately 95%. The standard uncertainty has been determined in
accordance with EA Publication EA-4/02. The long term stability of the calibrated object
is not included in the reported expanded uncertainty of measurement.
The error is calculated as (measured value true value) or for percentage values
(measured percentage value true percentage value).
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

2 (6)

Reference

MTePX22199-K03 rev2

Harmonics, levels 200 % of class 3 in accordance with of IEC 61000-2-4


Set voltage

L1
Error

L2
Error

L3
Error

Expanded
uncertainty

/Harm. nbr

/V

/mV

/mV

/mV

/mV

60

-16

-4

-4

/Harm. nbr

/%

/%

/%

/%

/%

2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
23
33
38

6.00
12.00
3.00
16.00
2.00
14.00
2.00
5.00
2.00
10.00
2.00
9.00
2.00
4.00
2.00
8.00
2.00
7.00
2.00
5.60
2.00
2.00

0.00
0.01
0.01
0.04
0.01
0.05
0.01
0.02
0.01
0.05
0.01
0.04
0.01
0.02
0.01
0.04
0.01
0.03
0.01
0.02
0.00
0.00

0.01
0.03
0.01
0.08
0.01
0.09
0.01
0.04
0.02
0.09
0.02
0.09
0.02
0.04
0.02
0.07
0.02
0.06
0.01
0.04
0.00
0.00

0.01
0.03
0.01
0.08
0.01
0.09
0.01
0.04
0.02
0.09
0.02
0.09
0.02
0.04
0.02
0.08
0.02
0.07
0.01
0.05
0.01
0.01

0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

3 (6)

Reference

MTePX22199-K03 rev2

Harmonics, levels 100 % of class 3 in accordance with of IEC 61000-2-4


Set voltage

L1
Error

L2
Error

L3
Error

Expanded
uncertainty

/Harm. nbr

/V

/mV

/mV

/mV

/mV

60

-19

-2

-2

/Harm. nbr

/%

/%

/%

/%

/%

2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
23
33
38

3.00
6.00
1.50
8.00
1.00
7.00
1.00
2.50
1.00
5.00
1.00
4.50
1.00
2.00
1.00
4.00
1.00
3.50
1.00
2.80
1.00
1.00

0.00
0.01
0.01
0.04
0.01
0.05
0.01
0.02
0.01
0.05
0.01
0.04
0.01
0.02
0.01
0.04
0.01
0.03
0.01
0.02
0.00
0.00

0.00
0.01
0.01
0.04
0.01
0.05
0.01
0.02
0.01
0.05
0.01
0.04
0.01
0.02
0.01
0.04
0.01
0.03
0.01
0.02
0.00
0.00

0.00
0.01
0.01
0.04
0.01
0.05
0.01
0.02
0.01
0.05
0.01
0.04
0.01
0.02
0.01
0.04
0.01
0.03
0.01
0.02
0.00
0.00

0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01
0,01

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

4 (6)

Reference

MTePX22199-K03 rev2

Harmonics, levels 40 % of class 3 in accordance with of IEC 61000-2-4

Set voltage

L1
Error

L2
Error

L3
Error

Expanded
uncertainty

/Harm. nbr

/V

/mV

/mV

/mV

/mV

60

-15

-15

/Harm. nbr

/%

/%

/%

/%

/%

3
4
7
9
11
16
23

2.40
0.60
2.80
1.00
2.00
0.40
1.12

0.01
0.00
0.02
0.01
0.02
0.00
0.01

0.01
0.00
0.02
0.01
0.02
0.00
0.01

0.01
0.00
0.02
0.01
0.02
0.00
0.01

0,01
0,01
0,01
0,01
0,01
0,01
0,01

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

5 (6)

Reference

MTePX22199-K03 rev2

Harmonics, levels 10 % of class 3 in accordance with of IEC 61000-2-4

Set voltage

L1
Error

L2
Error

L3
Error

Expanded
uncertainty

/Harm. nbr

/V

/mV

/mV

/mV

/mV

60

-15

-3

-3

/Harm. nbr

/%

/%

/%

/%

/%

2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
23
33
38

0.30
0.60
0.15
0.80
0.10
0.70
0.10
0.25
0.10
0.50
0.10
0.45
0.10
0.20
0.10
0.40
0.10
0.35
0.10
0.28
0.10
0.10

-0.002
0.000
0.001
0.004
0.001
0.006
0.001
0.003
0.001
0.007
0.001
0.005
0.000
0.002
0.001
0.004
0.001
0.002
0.001
0.004
0.001
0.001

-0.001
0.001
0.001
0.004
0.001
0.006
0.001
0.003
0.001
0.007
0.001
0.005
0.001
0.002
0.001
0.004
0.001
0.003
0.001
0.004
0.001
0.001

-0.001
0.001
0.001
0.005
0.001
0.006
0.000
0.003
0.001
0.007
0.000
0.005
0.001
0.002
0.002
0.004
0.001
0.002
0.001
0.004
0.001
0.000

0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005
0,005

Summary: In all the above tested points the results are in compliance with IEC 61000-430 Class A.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

6 (6)

Reference

MTePX22199-K03 rev2

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of the Swedish government. SP realizes fundamental units such as volt, ohm and
second from primary standards. Traceability for other units is established from these
realizations by means of in-house calibrations and scientific analyses. To ensure international equivalence and acceptance of the established traceability, interlaboratory
comparisons are made between national laboratories.

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
Voltage amplifier Fluke 5215A s/n 455014
DAC Iotech DAC488HR/2, SPno 502132

SP Technical Research Institute of Sweden


Measurement Technology
Performed by

Signed by: Stefan Svensson


Reason: I am the author of this document
Date & Time: 2012-04-23 14:23:25 +02:00
__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2014-07-01

1 (3)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTe4P04418-K03
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Harmonics
Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2014-07-01

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature

23.0 1.0C

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (2011:791) and the Decree (2011:811) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-07-01

2 (3)

Reference

MTe4P04418-K03

Harmonics, levels 10 % of class 3 in accordance with of IEC 61000-2-4


Set voltage

/Harm. nbr
41
45
50

/V

0.08
0.06
0.06

Error

Error

Error

0.000
0.000
0.000

0.000
0.000
-0.001

0.001
0.000
0.000

/V

/V

/V

Expanded
uncertainty
/mV
0.006
0.006
0.006

Harmonics, levels 40 % of class 3 in accordance with of IEC 61000-2-4

Set voltage

/Harm. nbr
41
45
50

/V

0.33
0.24
0.24

Error

Error

Error

0.000
-0.001
-0.002

0.001
0.000
-0.004

0.001
0.000
-0.001

/V

/V

/V

Expanded
uncertainty
/mV
0.006
0.006
0.006

Harmonics, levels 100 % of class 3 in accordance with of IEC 61000-2-4

Set voltage

/Harm. nbr
41
45
50

/V

0.8
0.6
0.6

Error

Error

Error

-0.0003
-0.0017
-0.0040

0.0010
-0.0014
-0.0098

0.0013
-0.0006
-0.0011

/V

SP Technical Research Institute of Sweden

/V

/V

Expanded
uncertainty
/mV
0.006
0.006
0.006

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-07-01

3 (3)

Reference

MTe4P04418-K03

Harmonics, levels 200 % of class 3 in accordance with of IEC 61000-2-4

Set voltage

/Harm. nbr
41
45
50

/V

1.6
1.2
1.2

Error

Error

Error

-0.001
-0.003
-0.007

0.001
-0.004
-0.020

0.002
-0.001
-0.001

/V

/V

/V

Expanded
uncertainty
/mV
0.006
0.006
0.006

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
DAC Iotech DAC488HR/2, SPno 502132
Voltage amplifier Fluke 5215A s/n 455014

SP Technical Research Institute of Sweden


Measurement Technology - Electricity
Performed by

__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2012-03-23

1 (3)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX22199-K05 rev2
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Interharmonics, Signaling


Rev1: Summary included, calculation error amended. Rev 2: Editorial.

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2012-02-16

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature
Warm-up time

23.0 2.0 C
> 12 h (connected to mains)

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)/true value*100 %.
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

2 (3)

Reference

MTePX22199-K05 rev2

Signalling, at 60 V

Signal
frequency
(Hz)

Set signal
level
(V)

Reading EUT
(V)

Expanded
uncertainty
(V)

620
1070
1070

3
6
9

3.020
5.857
9.042

0.001
0.001
0.001

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.
Interharmonics, at 60 V

Signal
frequency
(Hz)

Set singal
level
(V)

Reading
EUT U1
(V)

Reading
EUT U2
(V)

Reading
EUT U3
(V)

Expanded
uncertainty
(V)

83.33
175
375

3
1.8
1.2

2.986
1.802
1.204

2.945
1.802
1.205

2.945
1.802
1.205

0.001
0.001
0.001

Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2012-03-23

3 (3)

Reference

MTePX22199-K05 rev2

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
Voltage amplifier Fluke 5215A s/n 455014
DAC Iotech DAC488HR/2, SPno 502132

SP Technical Research Institute of Sweden


Measurement Technology
Performed by
Signed by: Stefan Svensson
Reason: I am the author of this document
Date & Time: 2012-04-17 11:10:39 +02:00
__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2014-07-01

1 (3)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTePX04418-K05
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Interharmonics, Signaling


Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2014-06-11

Measurement methods and procedures


The instrument was calibrated by comparison with the electrical voltage and frequency
standards of SP.

Measurement conditions
Room temperature

23.0 2.0 C

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)/true value*100 %.
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (1992:1119) and the Decree (1993:1065) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-07-01

2 (3)

Reference

MTePX04418-K05

Signalling, at 60 V

Signal
frequency
(Hz)

Set signal
level
(%)

Reading EUT
(%)

Expanded
uncertainty
(%)

315
315

15
5

15.049
5.022

0.001
0.001

For sinusoidal signals the reported levels S1-S25 (0 -2500 Hz) were all < 0,095 %
Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.
Interharmonics, at 60 V

Signal
frequency
(Hz)

Set singal
level
(V)

Reading
EUT U1
(V)

Reading
EUT U2
(V)

Reading
EUT U3
(V)

Expanded
uncertainty
(V)

25
225
1075
1725
2475

3.6
3.6
3.6
3.6
3.6

3.580
3.601
3.611
3.576
3.475

3.586
3.607
3.618
3.583
3.444

3.582
3.603
3.613
3.579
3.491

0.001
0.001
0.001
0.001
0.001

For sinusoidal 60V signals the reported levels were all < 0,03 V
Summary: In the tested points the results are in compliance with IEC 61000-4-30 Class A.

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-07-01

3 (3)

Reference

MTePX04418-K05

Equipment
Digital wattmeter DSWM, SP, SPno 503 243
Voltage amplifier Fluke 5215A s/n 455014
DAC Iotech DAC488HR/2, SPno 502132

SP Technical Research Institute of Sweden


Measurement Technology
Performed by

__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden

__Signature_2

KALIBRERINGSBEVIS
utfrdat av riksmtplats 01
CALIBRATION CERTIFICATE issued by a Swedish National Laboratory
Contact person

Date

Page

Stefan Svensson

2014-08-28

1 (6)

Measurement Technology
+46 10 516 54 15
stefan.svensson@sp.se

Reference

MTeP05735-K01 Rev1
Unipower AB
P. O. Box 411
SE-441 28 Alingss
Sweden

Calibration of Power
Rev1:Editorial

Identification
Object
Object state
Calibration location
Calibration date

Power Quality meter, Unipower UP-2210R, Serial number


22120402
Upon arrival the object had no visual damage.
Bors
2014-08-28 and 2014-06-11

Measurement methods and procedures


The instrument was calibrated by comparison with the standards of SP.

Measurement conditions
Room temperature

23.0 1.0C

Results
The result is presented below as the error expanded uncertainty, or as set value, measured
value and expanded uncertainty. The reported expanded uncertainty of measurement is stated
as the standard uncertainty of measurement multiplied by the coverage factor k = 2, which for
a normal distribution corresponds to a coverage probability of approximately 95%. The
standard uncertainty has been determined in accordance with EA Publication EA-4/02. The
long term stability of the calibrated object is not included in the reported expanded uncertainty
of measurement.
The error is calculated as (measured value true value)/true value*100 %.
Tables of the result are found on following pages.

SP Technical Research Institute of Sweden


Postal address

Office location

Phone / Fax / E-mail

SP
Box 857
SE-501 15 BORS
Sweden

Vstersen
Brinellgatan 4
SE-504 62 BORS

+46 10 516 50 00
+46 33 13 55 02
info@sp.se

National Laboratories are designated by the Swedish Government


according to the Act (2011:791) and the Decree (2011:811) concerning
testing and metrology. SP operates under ISO 17025, supervised by
SWEDAC. This document may not be reproduced other than in full, except
with the prior written approval of SP.

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-08-28

2 (6)

Reference

MTeP05735-K01 Rev1

Power at 60 V, 50 Hz, positive energy direction


Three-phase connection, symmetrical load,

Set
Set
Current power
factor
(A)
(-)

Error
Ptot

Error
Qtot

Error
PF

(%)

(%)

(-)

Expanded Expanded Expanded


uncertainty uncertainty uncertainty
Ptot
Qtot
PF
(%)
(%)
(-)

5
5
5

1
0.5i
0.8c

-0.022
-0.168
0.038

0.025
0.140

0.0000
-0.0007
0.0004

0,020
0,020
0,020

0,020
0,020

0,001
0,001
0,001

1.2
1.2
1.2

1
0.5i
0.8c

-0.054
-0.066
-0.012

-0.014
-0.115

0.0000
-0.0002
0.0003

0,020
0,020
0,020

0,020
0,020

0,001
0,001
0,001

1.0
1.0
1.0

1
0.5i
0.8c

-0.038
-0.116
-0.006

-0.010
-0.049

0.0000
-0.0004
0.0001

0,020
0,020
0,020

0,020
0,020

0,001
0,001
0,001

0.5
0.5
0.5

1
0.5i
0.8c

-0.042
-0.086
-0.027

-0.016
0.083

0.000
0.000
0.000

0,020
0,020
0,020

0,020
0,020

0,001
0,001
0,001

0.1
0.1
0.1

1
0.5i
0.8c

-0.032
-0.096
-0.027

0.067
-0.053

-0.0010
-0.0009
-0.0006

0,020
0,020
0,020

0,020
0,020

0,001
0,001
0,001

0.05
0.05

1
0.5i

0.095
0.090

0.150

-0.001
-0.001

0,020
0,020

0,020

0,001
0,001

0.02
0.02

0.5i
0.8c

-0.040
0.075

0.431
-0.002

-0.0060
-0.0070

0,020
0,020

0,020
0,020

0,001
0,001

0,01

-0.0339

0,020

0,001

-0.065

Summary: In the tested points the results are in compliance with IEC 62053-22, cl 0,2 S.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-08-28

3 (6)

Reference

MTeP05735-K01 Rev1

Power at 60 V, 50 Hz, negative energy direction


Three-phase connection, symmetrical load,

Set
Set
Current power
factor
(A)
(-)

Error
Ptot

Error
Qtot

Error
PF

(%)

(%)

(-)

Expanded Expanded Expanded


uncertainty uncertainty uncertainty
Ptot
Qtot
PF
(%)
(%)
(-)

5
5
5

1
0.5i
0.8c

-0.073
-0.223
-0.012

-0.024
-0.184

0.0000
0.0008
-0.0005

0,020
0,020
0,020

0,020
0,020

0,0010
0,0010
0,0010

1.2
1.2
1.2

1
0.5i
0.8c

-0.054
-0.145
-0.021

-0.027
-0.120

0.0000
0.0004
-0.0003

0,020
0,020
0,020

0,020
0,020

0,0010
0,0010
0,0010

1.0
1.0
1.0

1
0.5i
0.8c

0.023
-0.051
0.092

0.026
-0.034

0.0000
0.0001
-0.0003

0,020
0,020
0,020

0,020
0,020

0,0010
0,0010
0,0010

0.5
0.5
0.5

1
0.5i
0.8c

-0.039
-0.104
-0.020

-0.030
-0.078

0.0000
0.0000
-0.0001

0,020
0,020
0,020

0,020
0,020

0,0010
0,0010
0,0010

0.1
0.1
0.1

1
0.5i
0.8c

-0.051
-0.119
-0.043

0.013
0.044

0.0011
0.0008
0.0010

0,020
0,020
0,020

0,020
0,020

0,0010
0,0010
0,0010

0.05

-0.093

0.0023

0,020

0,0010

0.02
0.02

0.5i
0.8c

-0.188
-0.017

0.085
0.493

0.0052
0.0089

0,020
0,020

0,020
0,020

0,0010
0,0010

0.01

-0.178

0.0333

0,020

0,0010

Summary: In the tested points the results are in compliance with IEC 62053-22, cl 0,2 S.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-08-28

4 (6)

Reference

MTeP05735-K01 Rev1

Power at 60 V, 50 Hz, positive energy direction


Single-phase load, three-phase voltages connected

Set
Set
Current power
factor
(A)
(-)

Phase

Error
Ptot

Error
Qtot

Error
PF

(-)

(%)

(%)

(-)

Expanded Expanded Expanded


uncertainty uncertainty uncertainty
Ptot
Qtot
PF
(%)
(%)
(-)

1.2
1.2
1.2
1.2
1.2
1.2

1
1
1
0.5 ind
0.5 ind
0.5 ind

L1
L2
L3
L1
L2
L3

-0.050
-0.041
-0.107
-0.038
-0.012
-0.378

-0.051
-0.039
0.006

-0.0037
-0.0055
-0.0042
-0.0049
-0.0026
-0.0034

0,020
0,020
0,020
0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010
0,0010
0,0010
0,0010

1
1
1
1
1
1

1
1
1
0.5 ind
0.5 ind
0.5 ind

L1
L2
L3
L1
L2
L3

-0.030
-0.023
-0.082
-0.028
-0.047
-0.347

-0.021
-0.024
0.023

-0.0043
-0.0065
-0.0052
-0.0022
-0.0031
-0.0038

0,020
0,020
0,020
0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010
0,0010
0,0010
0,0010

0.1
0.1
0.1

0.5 ind
0.5 ind
0.5 ind

L1
L2
L3

0.145
0.168
-0.226

0.120
0.124
0.166

-0.0182
-0.0267
-0.0254

0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010

0.05
0.05
0.05

1
1
1

L1
L2
L3

0.094
0.093
0.027

-0.0703
-0.1031
-0.0975

0,020
0,020
0,020

0,0010
0,0010
0,0010

Summary: In the tested points the results are in compliance with IEC 62053-22, cl 0,2 S.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-08-28

5 (6)

Reference

MTeP05735-K01 Rev1

Power at 60 V, 50 Hz, negative energy direction


Single-phase load, three-phase voltages connected

Set
Set
Current power
factor
(A)
(-)

Phase

Error
Ptot

Error
Qtot

Error
PF

(-)

(%)

(%)

(-)

Expanded Expanded Expanded


uncertainty uncertainty uncertainty
Ptot
Qtot
PF
(%)
(%)
(-)

1.2
1.2
1.2
1.2
1.2
1.2

1
1
1
0.5 ind
0.5 ind
0.5 ind

L1
L2
L3
L1
L2
L3

-0.048
-0.039
-0.100
-0.050
-0.011
-0.396

-0.055
-0.062
-0.008

0.0043
0.0052
0.0041
0.0020
0.0023
0.0035

0,020
0,020
0,020
0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010
0,0010
0,0010
0,0010

1
1
1
1
1
1

1
1
1
0.5 ind
0.5 ind
0.5 ind

L1
L2
L3
L1
L2
L3

-0.049
-0.039
-0.083
-0.041
-0.005
-0.357

-0.057
-0.055
-0.010

0.0050
0.0062
0.0051
0.0023
0.0029
0.0039

0,020
0,020
0,020
0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010
0,0010
0,0010
0,0010

0.1
0.1
0.1

0.5 ind
0.5 ind
0.5 ind

L1
L2
L3

0.044
0.088
-0.272

-0.072
-0.094
-0.041

0.0191
0.0252
0.0263

0,020
0,020
0,020

0,020
0,020
0,020

0,0010
0,0010
0,0010

0.05
0.05
0.05

1
1
1

L1
L2
L3

-0.026
0.009
-0.062

0.0786
0.0990
0.0944

0,020
0,020
0,020

0,0010
0,0010
0,0010

Summary: In the tested points the results are in compliance with IEC 62053-22, cl 0,2 S.

SP Technical Research Institute of Sweden

KALIBRERINGSBEVIS
CALIBRATION CERTIFICATE

Date

Page

2014-08-28

6 (6)

Reference

MTeP05735-K01 Rev1

Traceability
SP is National Laboratory for electrical quantities and time and frequency by appointment of
the Swedish government. SP realizes fundamental units such as volt, ohm and second from
primary standards. Traceability for other units is established from these realizations by means
of in-house calibrations and scientific analyses. To ensure international equivalence and
acceptance of the established traceability, interlaboratory comparisons are made between
national laboratories.

Equipment
Three phase comparator Zera Com303-3, SPno 503044

SP Technical Research Institute of Sweden


Measurement Technology - Electricity
Performed by

__Signature_1

Stefan Svensson

SP Technical Research Institute of Sweden