Académique Documents
Professionnel Documents
Culture Documents
com
ASTM D 97 - D 2500 - ISO 3015 & ISO 3016 - IP 15 & IP 219
SCOPE:
The Pour point describes a procedure for testing the fluidity of a
petroleum product at a specified temperature. The Cloud point is
defined as the temperature of a liquid specimen when the smallest
observable cluster of wax crystals first appears upon cooling under
prescribed conditions.
APPLICATIONS:
Lubricating oils, distillate and residual oils,
Biodiesels.
FEATURES
Mimics the standard methods
Detection by specific optical fibre for CLOUD Point
Detection by ultra sonic sensor for POUR Point
Visualisation of the different cooling steps
detection every 3C or every 1C
Programmable cooling slope
Quick fixation on measuring head
Stand alone or network multiple
Easy access to maintenance parameters by
software
Easy access to hardware maintenance
TOUCHSCREEN
interface
Storage: 200 results
SOFTWARE
9 Computer interface through a user friendly software
accessible by touch screen
9 Connexion of external PC by Ethernet cable
9 Storage of 200 results in stand alone
SCOPE OF DELIVERY
REF 60300 NTE 450 CPP for determination of CLOUD
AND POUR POINT delivered complete with Printer, 1 cloud
and 1 Pour point tubes, instruction manual.
CRYOSTAT to be ordered separately
COOLING BY CRYOASTAT
ZOOM on
Detection Head
DISTRIBUTED BY
www.normalab.com
HIGHLY COMPREHENSIVE SOFTWARE
MAIN MENU
ALTERNATIVE MODELS
PROBE CALIBRATION
TECHNICAL SPECIFICATIONS
GENERAL
9 Visualisation of the detection level
9 Adjustable bath temperature at stand by position
9 Pt 100 sample temperature
9 Jacket temperature
9 Visualisation of the cooling levels
9 Apparatus and sensors status
9 Easy access to the service parameters
9 Storage of the 200 last tests results
9 Sample identification
9 Operators name
9 Maintenance and calibration protected by
password
9 Programme files: 15 (5 standards and 10
parametrable)
POUR POINT
9 Detection by ultrasonic sensor
9 Temperature range from -75C to 51C (Lowest
temperature depends on the cryostat type)
9 Temperature measurement resolution : 1C
9 Tilting intervals : every 3C or 1C (parametrable)
CLOUD POINT
9 Detection by optical fibre
9 Intervals of 1C
9 Temperature measurement resolution : 0,1C
9 Temperature range from -75C to 49C (Lowest
temperature depends on the cryostat type)
DISTRIBUTED BY