EON Instrumentation, Inc.

Standard Operating Procedure

Document Number:
Revision Level:

SOP-006
NC

Page:
DCN Number:

1 of 5
DCN-001

CALIBRATION AND VERIFICATION OF EQUIPMENT


1.0

PURPOSE
1.1 The purpose of this procedure is to define the Calibration and Verification
process.

2.0

SCOPE
2.1 This procedure applies to those areas within EON INSTRUMENTATION, INC.s
Quality Management System.

3.0

RESPONSIBILITIES
3.1 The President has assigned the President with the responsibility of
maintaining the Calibration and Verification Process.
3.2 EON INSTRUMENTATION, INC. has determined the monitoring and
measurement to be undertaken and the monitoring and measuring
equipment needed to provide evidence of conformity of product to
determined requirements.
3.3 Employee equipment, tools, or measuring devices are not allowed to be
used within EON INSTRUMENTATION, INC., unless they are placed within the
Calibration and Verification Process.

4.0

DEFINITIONS
Calibration - requirements: The process of validating specific monitoring
and measuring devices to measurement standards traceable to international
or national measurement standards.
Verification - requirements: The process of validating specific monitoring
and measuring devices to documented instructions and when applicable
traceable to international or national measurement standards.
Reference

Only

requirements:
Equipment owned by EON
which does not require Calibration or Verification.
Reference Only - can be defined as No Calibration Required/Necessary.
INSTRUMENTATION, INC.

Equipment: Any monitoring and measuring equipment (Test Equipment,


Calipers, Micrometers, Scales, etc.) used to provide evidence of conformity of
deliverable product to determined requirements.

EON Instrumentation, Inc.


Standard Operating Procedure

Document Number:
Revision Level:

SOP-006
NC

Page:
DCN Number:

2 of 5
DCN-001

CALIBRATION AND VERIFICATION OF EQUIPMENT


5.0

PROCEDURE
5.1

Equipment Calibration and Verification Process

5.1.1 Equipment requiring Calibration or Verification will be recorded into the


Calibration and Verification Log, F_004. This log contains at a minimum:
serial number, description, next calibration or verification frequency.
5.1.2 Equipment shall:

Be calibrated or verified at specified


intervals (EON INSTRUMENTATION, INC. will maintain a +/- 60 Day period
for Calibration / Verification Due Dates based on company needs or
unless specified by a Customer Requirement), or prior to use, against
measurement standards traceable to international or national
measurement standards; where no such standards exist, the basis
used for calibration or verification shall be recorded;

Be adjusted or re-adjusted as necessary (by Calibration Service);

Be identified to enable the calibration status to be determined


(Calibration Service or EON INSTRUMENTATION, INC.s Calibration
Sticker);

Be safeguarded from adjustments (e.g. Tamper Free Sticker) that


would invalidate the measurement result (by Calibration Service);

Be protected from damage and deterioration during handling,


maintenance and storage within EON INSTRUMENTATION, INC.;

Recalled to the requirements of the Calibration and Verification Log,


F_004.

5.1.3 The President is responsible for the following specific Calibration


requirements:
EON INSTRUMENTATION, INC. does not perform in-house Calibration;
Use Calibration sub-contractors that are either ISO17025 accredited
suppliers or equivalent;
Review Calibration Certificate results. In the event of that Calibration
Certifications indicate a failure refer to Step 5.2 Calibration/Verification
Alert.
Maintain records of calibration certificates per SOP-002, Control of
Records;
Ensure only Calibrated items shall be used to perform 1st Article or
Final Inspection or any other inspection of deliverable product.
Increase Calibration Intervals:

EON Instrumentation, Inc.


Standard Operating Procedure

Document Number:
Revision Level:

SOP-006
NC

Page:
DCN Number:

3 of 5
DCN-001

CALIBRATION AND VERIFICATION OF EQUIPMENT

Calibration intervals can be increased to double (twice) the


interval period based on three periods of previous acceptable
calibration results in a row.

Example: Gage #123, which is normally calibrated every


year, has not failed for 3 years in a row. Therefore, Gage
#123 new calibration interval is 2 years, etc.

Decrease Calibration Intervals:

Calibration intervals shall be decreased to one half (1/2) of


interval period based on three periods of previous unacceptable calibration results in a row. Note: Unacceptable
results have to be corrected by External Service before
equipment can be used by EON INSTRUMENTATION, INC..

Example: Gage #123, which is normally calibrated every


year, has failed 3 years in a row but, corrected.
Therefore, Gage #123 new calibration interval is 6
months.

5.1.4 The President is responsible for the following specific Verification


requirements:
Verification is to be conducted using Calibration and Verification Log,
F_004 and identified with a VERIFICATION label.
Equipment verified CAN be used to perform 1 st Article or Final
Inspection or any other inspection of deliverable product;
Verification is to be performed per EON INSTRUMENTATION, INC.
instructions documented within the Calibration and Verification Log,
F_004;
In the event of that verification results indicate failure refer to Step 6.2
Calibration/Verification Alert.
Maintain records of verification results per SOP-002, Control of
Records.
Increase Verification Intervals:

Verification intervals can be increased to double (twice) the


interval period based on three periods of previous acceptable
verification results in a row.

Example: Gage #123, which is normally verified every


year, has not failed for 3 years in a row. Therefore, Gage
#123 new verification interval is 2 years, etc.

Decrease Verification Intervals:

EON Instrumentation, Inc.


Standard Operating Procedure

Document Number:
Revision Level:

SOP-006
NC

Page:
DCN Number:

4 of 5
DCN-001

CALIBRATION AND VERIFICATION OF EQUIPMENT

Verification intervals shall be decreased to one half (1/2) of


interval period based on three periods of previous unacceptable verification results in a row. Note: Unacceptable
results have to be corrected by External Service before
equipment can be used by EON INSTRUMENTATION, INC..

Example: Gage #123, which is normally verified every


year, has failed 3 years in a row but, corrected.
Therefore, Gage #123 new verification interval is 6
months.

5.1.5 The President is responsible for ensuring that computer software used to
monitor or measure deliverable product requirements be confirmed prior
to initial use and reconfirmed as necessary.
Such computer software shall also be recorded into the Calibration and
Verification Log, F_004.
5.2

Calibration/Verification Alert

5.2.1 The President is responsible for taking the appropriate action (both on
equipment and product affected) when equipment is found not to conform
to requirements.
5.3

Reference Only - Equipment

5.3.1 Equipment will be identified with a REFERENCE ONLY label.


5.3.2 Equipment used as REFERENCE ONLY does not require either
Calibration or Verification and therefore excluded from the Calibration or
Verification process.
5.3.3 REFERENCE ONLY equipment shall not be used to perform: 1st Article
Inspection, Final Inspection or any other inspection of deliverable product.
5.3.4 REFERENCE ONLY equipment can be identified in any method (label,
sticker, permanent marker, etc.) that indicates REFERENCE ONLY.
5.3.5 REFERENCE ONLY equipment is NOT required to be recorded onto the
Calibration and Verification Log, F_004.

EON Instrumentation, Inc.


Standard Operating Procedure

Document Number:
Revision Level:

SOP-006
NC

Page:
DCN Number:

5 of 5
DCN-001

CALIBRATION AND VERIFICATION OF EQUIPMENT


6.0

REFERENCES
Procedures / Policies:

Quality System Manual, QSM_001: Section: 7.6 Control of Monitoring and


Measuring Equipment

SOP_002, Control of Records

Instructions:

None

Forms:

Calibration and Verification Log, F_004.