Académique Documents
Professionnel Documents
Culture Documents
INSTRUMENTATION
This chapter describes the instruments: TL glow curve recorder, XRD, TGA (Thermal
gravimetric Analysis), FTIR, ICP-AES (Induction Coupled Plasma Atomic Emission
Spectroscopy) and Laser diffraction particle size analyzer used for the characterization
of collected minerals.
3.1 FURNACE
Thermal annealing treatment for the specimen was carried out in the muffle furnace.
The muffle furnace with maximum 1200C heating capacity and control options was
used for Thermal annealing treatment. The temperature was maintained with 1oC
accuracy using a temperature controller. Alumina crucibles were used in the furnace
3.2 THERMOLUMINESCENCE READER
Nucleonix PC Controlled Thermoluminescence Reader [Type: TL1009] was
employed for the thermoluminescence property analysis of the samples [1]. The
heating rate was set as 1C/sec and samples were heated from room temperature to
400C. Before TL procedure, 5mg sample was exposed to radiations at room
temperature for 5 minutes. After 1-minute retention period, sample was put into TLD
reader.
PC Controlled Thermoluminescence Reader [TYPE: TL 1009]
Thermoluminescence Reader Type TL1009 designed and offered by NUCLEONIX
SYSTEMS is a versatile controller based unit, facilitating the user to subject the TL
sample under study to the desired heating profile, to record the digitized TL glow
curve (Figure-3.1). This unit stores both integral value and digitized glow curve into
EEPROM memory. This unit records the data in 200 channels, temp, TL intensity &
Run time values. Entire electronics including PMT, HV bias, Temperature controller
circuit, Heater transformer heater strip, sample drawer assembly, data acquisition
electronics is all integrated into a single enclosure. The user interface to the unit is
through a powerful software GUI, coded in VB which runs on Windows platform.
This system essentially works as a PC controlled TLD reader with command buttons
and dropdown menus defined for various functions. Built-in small keypad and visual
indication through a 20x4 LCD dotmatrix display provide limited functions to the
40
users which essentially serve the purpose of checking the majority of the unit
hardware functionality. Built-in serial port in the unit facilitates the user to connect it
to a PC for GUI and to achieve full functionality [2].
(i) Features
Software features include glow curve, acquisition, display, filing, printing, processing
etc.
(ii) Hardware specifications
PMT housing and TL sample heating assembly: This unit has a photomultiplier
Tube generally of Hamamatsu / ET make, is used. However, assembly facilitates one
to go in for other photomultiplier tubes also with appropriate modification in the PMT
socket wiring. The cylindrical shell containing the Photomultiplier is fitted on to a
rectangular base drawer block containing a heater arrangement and thermocouple,
heater rods, connected to a power transformer.
Heating Arrangement: Resistive heating method.
Heating Element (Heater Strip): Kanthal strip (72% Fe, 23% Al and 2% Cr or
Nichrome) is used as a heating element. Kanthal Strip has a circular depression of
14mm to hold discs and powder samples.
Heating Process: Programmed heating can be done in two modes i) "PROG MODE"
of Temperature Controller through personal computer program. ii) "ISO MODE"
(Internal mode) of Temperature Controller, by varying the ten turn dial.
Temperature Sensor: Thermocouple Sensor (Cr-Al spot-welded to heater strip).
PMT Cooling: Thermoelectric cooling arrangement for the PMT is provided.
High voltage to PMT: A continuously adjustable HV (0 to -1500V) @1mA is
generated by HV circuits. User can select desired HV from front panel. There is HV
socket (test) for checking HV presence on rear panel.
Auto -Ranging: Current output from the photomultiplier is taken to I-F converter, to
give frequency output proportional to PMT current. This wide dynamic range is
achieved for plotting TL intensity on Y-axis
41
42
provided on the DPM of TC. Also parallel the TC output is sent to TL Acquisition
Module AM 576. Heating is controlled by the TL Acquisition Module as per the
parameter settings. TL Acquisition Module AM 576 reads temperature and TL
intensity (I-F) & displays them on LCD Alphanumeric display. DAC is programmed
to generate different heating profile for heating the sample. The printer attached to a
PC under software control provides plot-out of glow curve & printing of TL intensity,
temperature, glow curve analysis report etc.
3.3 POWDER X-RAY DIFFRACTOMETER
X-ray diffraction (XRD) is a rapid analytical technique primarily used for phase
identification of a crystalline material and can provide information on unit cell
dimensions. The analyzed material is finely ground, homogenized, and average bulk
composition is determined [3].
To characterize the crystal structure of synthesized strontium pyrophosphate and the
doped samples, the powder X-ray diffraction (XRD) measurements were established.
For these measurements, Rigaku MiniFlex X-ray Diffractometer with a radiation
source of CuK was used [4,5].
Basics of X-ray Diffraction:
About 95% of all solid materials can be described as crystalline. When X-rays are
incident on a crystalline substance (Phase), one gets a diffraction pattern. In 1919
A.W.Hull gave a paper titled, A New Method of Chemical Analysis. Here he
pointed out that .every crystalline substance gives a pattern; the same substance
always gives the same pattern; and in a mixture of substances each produces its
pattern independently of the others.
The X-ray diffraction pattern of a pure substance is, therefore, like a fingerprint of the
substance. The powder diffraction method is thus ideally suited for characterization
and identification of polycrystalline phases [4,5,6]. The main use of powder
diffraction is to identify components in a sample by a search/match procedure.
Furthermore, the areas under the peak are related to the amount of each phase present
in the sample.
Theoretical Considerations
The atoms are arranged in a regular pattern, and there is a smallest volume element
that by repetition in three dimensions describes the crystal. This smallest volume
element is called a unit cell. The dimensions of the unit cell is described by three axes
a, b, c and the angles between them alpha, beta, gamma.
44
45
46
The possible d-spacing defined by the indices h, k, l are determined by the shape of
the unit cell. Rewriting Braggs law we get:
sin = /2d
Therefore the possible 2 values where we can have reflections are determined by the
unit cell dimensions. However, the intensities of the reflections are determined by the
distribution of the electrons in the unit cell. The highest electron density is found
around atoms. Therefore, the intensities depend on what kind of atoms we have and
where in the unit cell they are located. Planes going through areas with high electron
density will reflect strongly, planes with low electron density will give weak
intensities.
3.4 FOURIER TRANSFORM INFRARED SPECTROMETER
Fourier transform spectroscopy is a measurement technique where by spectra are
collected based on measurements of the coherence of a radiative source, using timedomain or space-domain measurements of the electromagnetic radiation or other type
of radiation[8]. It can be applied to a variety of types of spectroscopy including
optical spectroscopy, infrared spectroscopy (FTIR, FT-NIRS), nuclear magnetic
resonance.Fourier Transform Infrared Spectroscopy (FTIR) identifies chemical bonds
in a molecule by producing an infrared absorption spectrum. Thus, the presence of
specific functional groups can be monitored. The pellet samples for Fourier
Transform Infrared Spectrometer (FTIR) measurements prepared based on 100mg:
2mg sample to KBr ratio [9,10,11]. The produced samples were studied by VARIAN
1000 FTIR spectrometer and the wave numbers were set between 400 and 4000cm-1.
Infrared Spectroscopy: Infrared spectroscopy has been a workhorse technique for
material analysis in the laboratory for over seventy years. An infrared spectrum
represents a fingerprint of a sample with absorption peaks which correspond to the
frequencies of vibrations between the bonds of the atoms making up the material.
Because each different material is a unique combination of atoms, no two compounds
produce the exact same infrared spectrum. Therefore, infrared spectroscopy can result
in a positive identification (qualitative analysis) of every different kind of material. In
addition, the size of the peaks in the spectrum is a direct indication of the amount of
material present. With modern software algorithms, infrared is an excellent tool for
quantitative analysis.
47
Fig: 3.5
Fig: 3.6
48
49
50
the fast scans enable the coaddition of several scans in order to reduce the random
measurement noise to any desired level (referred to as signal averaging).
Mechanical Simplicity: The moving mirror in the interferometer is the only
continuously moving part in the instrument. Thus, there is very little possibility of
mechanical breakdown.
Internally Calibrated: These instruments employ a He-Ne laser as an internal
wavelength calibration standard (referred to as the Connes Advantage). These
instruments are self-calibrating and never need to be calibrated by the user.
These advantages, along with several others, make measurements made by FT-IR
extremely accurate and reproducible. Thus, it is a very reliable technique for positive
identification of virtually any sample. The sensitivity benefits enable identification of
even the smallest of contaminants. This makes FT-IR an invaluable tool for quality
control or quality assurance applications whether it is batch-to-batch comparisons to
quality standards or analysis of an unknown contaminant. In addition, the sensitivity
and accuracy of FT-IR detectors, along with a wide variety of software algorithms,
have dramatically increased the practical use of infrared for quantitative analysis.
Quantitative methods can be easily developed and calibrated and can be incorporated
into simple procedures for routine analysis. Thus, the Fourier Transform Infrared (FTIR) technique has brought significant practical advantages to infrared spectroscopy. It
has made possible the development of many new sampling techniques which were
designed to tackle challenging problems which were impossible by older technology.
It has made the use of infrared analysis virtually limitless.
3.5 THERMOGRAVIMETRIC ANALYZER (TGA)
Thermogravimetric analysis or thermal gravimetric analysis (TGA) is a type of testing
performed on samples that determines changes in weight in relation to change in
temperature. Such analysis relies on a high degree of precision in three measurements:
weight, temperature, and temperature change [12]. As many weight loss curves look
similar, the weight loss curve may require transformation before results may be
interpreted. A derivative weight loss curve can identify the point where weight loss is
most apparent. Again, interpretation is limited without further modifications and
deconvolution of the overlapping peaks may be required.
51
Thermogravimetric curves (thermograms) provide information regarding polymerization reactions, the efficiencies of stabilizers and activators, the thermal stability of
final materials, and direct analysis.
As many weight loss curves look similar, the weight loss curve may require
transformation before results may be interpreted. A derivative weight loss curve can
identify the point where weight loss is most apparent. Again, interpretation is limited
without further modifications and deconvolution of the overlapping peaks may be
required.
TGA is commonly employed in research and testing to determine characteristics of
materials such as polymers, to determine degradation temperatures, absorbed moisture
content of materials, the level of inorganic and organic components in materials,
decomposition points of explosives, and solvent residues. It is also often used to
estimate the corrosion kinetics in high temperature oxidation.
The analyzer usually consists of a high-precision balance with a pan (generally
platinum) loaded with the sample. A different process using a quartz crystal
microbalance has been devised for measuring smaller samples on the order of a
microgram versus milligram with conventional TGA.The sample is placed in a small
electrically heated oven with a thermocouple to accurately measure the temperature.
The atmosphere may be purged with an inert gas to prevent oxidation or other
undesired reactions. A computer is used to control the instrument.
Analysis is carried out by raising the temperature of the sample gradually and plotting
weight (percentage) against temperature. The temperature in many testing methods
routinely reaches 1000C or greater. After the data are obtained, curve smoothing and
other operations may be done to find the exact points of inflection.
A method known as hi-resolution TGA is often employed to obtain greater accuracy
in areas where the derivative curve peaks. In this method, temperature increase slows
as weight loss increases. This is to more accurately identify the exact temperature
where a peak occurs. Several modern TGA devices can vent burnoff to an infrared
spectrophotometer to analyze composition.
52
THEORY
TGA is used primarily for determining thermal stability of polymers. The most widely
used TGA method is based on continuous measurement of weight on a sensitive
balance (called a thermobalance) as sample temperature is increased in air or in an
inert atmosphere. This is referred to as non-isothermal TGA. Data are recorded as a
thermogram of weight versus temperature. Weight loss may arise from evaporation of
residual moisture or solvent, but at higher temperatures it results from polymer
decomposition. Besides providing information on thermal stability, TGA may be used
to characterize polymers through loss of a known entity, such as HCl from poly (vinyl
chloride). Thus weight loss can be correlated with percent vinyl chloride in a
copolymer. TGA is also useful for determining volatilities of plasticizers and other
additives. Thermal stability studies are the major application of TGA. A variation of
the method is to record weight loss with time at a constant temperature, called
isothermal TGA, this is less commonly used than non-isothermal TGA. Modern TGA
instruments allow thermograms to be recorded on microgram quantities of
material.Some instruments are designed to record and process DSC and TGA data
simultaneously, and may also be adapted for gas chromatographic and/or mass
spectrometric analysis of effluent degradation products.
Advantages of using TGA for Materials Analysis
Thermogravimetric analysis uses heat to force reactions and physical changes in
materials. TGA provides quantitative measurement of mass change in materials
associated with transition and thermal degradation. TGA records change in mass from
dehydration, decomposition, and oxidation of a sample with time and temperature.
Characteristic thermogravimetric curves are given for specific materials and chemical
compounds due to unique sequence from physicochemical reactions occurring over
specific temperature ranges and heating rates. These unique characteristics are related
to the molecular structure of the sample. When used in combination with FTIR,
TGA/FTIR is capable of detailed FTIR analysis of evolved gases produced from the
TGA.
53
Fig; 3.7 a
Fig:3.7 b
Fig:3.7 c
54
features of these results are the Mie resonances, sizes that scatter particularly strongly
or weakly. This is in contrast to Rayleigh scattering for small particles and RayleighGans-Debye scattering for large particles [14]. The existence of resonances and other
features of Mie scattering, make it a particularly useful formalism when using
scattered light to measure particle size.
Using Mie Theory and the Fraunhofer Approximation
When setting up laser diffraction methods, users are faced with the decision as to
whether to use Mie Theory or the Fraunhofer Approximation to calculate the particle
size distribution results.
The Fraunhofer Approximation represents the easiest model to set-up as, in contrast to
Mie Theory, it does not require the user to provide any optical property information.
However, its use can lead to significant errors due to the assumptions it makes
regarding the nature of the materials being measured. As such, users need to consider
the following when selecting the Fraunhofer Model:
Particle Absorption: If the particles show some transparency (absorption < 0.2),
then the Fraunhofer approximation will tend to yield inaccurate results below
50m in size. If the absorption is high (>0.2), good results may be obtained down
to 2m in size, although this does depend on the refractive index.
Particle Refractive Index : If the refractive index different between the particle
and the medium which surrounds it is low, then the Fraunhofer model can shown
errors, even up to very large particle sizes (>200m).
Particle Size: If the particle size distribution contains material less than 2 microns
in size then the Fraunhofer Approximation will lead to an incorrect assessment of
the fine particle fraction.
The nature of the errors observed when using the Fraunhofer Approximation is not
always predictable. In most instances an over-estimation of the fine particle fraction is
observed, as shown here for a pharmaceutical material. However, it is also possible
for the Fraunhofer model to underestimate the fine particle fraction, as is seen for
materials such as calcium carbonate, because it incorrectly predicts the scattering
efficiency of these particles.
56
In the present investigation the Mastersizer Micro system, Version 2.19 was used to
find out the particle size (Figure 3.8). The following are the technical details.
Extensive market research and customer opinion directed the development of the
Mastersizer Micro systems. The result is a small footprint, low cost instrument that
takes laser diffraction out of the specialist laboratory and into the hands of particle
technologists everywhere. The Micro is an instrument system in the clear Malvern
tradition with no compromise on sample handling, Mastersizer Micro performs like a
research machine and yet operates like a cost effective QC tool. The wide measuring
range makes it ideal for application in fundamental studies and quality control of
production lines in diverse industries such as pharmaceuticals, mining, minerals, clays,
food, metal powders, emulsions, ceramics, polymers, pigments, paints and coatings to
name a few. The Block diagram of the detection system is in figure 3.9.
Easy Operation and Maintenance
Single button operation means that inexperienced users can perform repetitive
measurements easily and reproducibly. The stable single lens optics minimizes set up
and routine maintenance.
Rapid data acquisition and analysis allows repeatable measurements to be made in
less than 4 seconds. A unique "dip-in" probe provides sample agitation, sonication
and circulation through the measuring system. By allowing the use of standard
laboratory beakers as the sample tank, variable volumes of material can be analyzed.
Superior Software Complete Control: Malvern software sets an acknowledged
world-wide standard for instrument operation, data acquisition and handling,
reporting and systems integration.
Set up as easy as ABC: A special "ABC" icon guides the user through a set-up
routine ensuring that no parameters are over looked.
57
Fig: 3.8
58
Measurement Principle
: Mie scatteringime
Measurement Time
Light Source
Accuracy
Power
: 100-240V, 50/60Hz
External Dimensions
Weight
: 32Kg
pattern and a
61
The argon gas is ionized in the intense electromagnetic field and flows in a particular
rotationally symmetrical pattern towards the magnetic field of the RF coil. A stable,
high temperature plasma of about 7000 K is then generated as the result of the
inelastic collisions created between the neutral argon atoms and the charged particles
[16,17].
Within the optical chamber(s), after the light is separated into its different
wavelengths (colours), the light intensity is measured with a photomultiplier tube or
tubes physically positioned to "view" the specific wavelength(s) for each element line
involved, or, inperistaltic pump delivers an aqueous or organic sample into a nebulizer
where it is atomized and introduced directly inside the plasma flame. The sample
immediately collides with the electrons and other charged ions in the plasma and is
broken down into charged ions. The various molecules break up into their respective
atoms which then lose electrons and recombine repeatedly in the plasma, giving off
radiation at the characteristic wavelengths of the elements involved.
In some designs, a shear gas, typically nitrogen or dry compressed air is used to 'cut'
the plasma flame at a specific spot. One or two transfer lenses are then used to focus
the emitted light on a diffraction grating where it is separated into its component
wavelengths in the optical spectrometer. In other designs, the plasma impinges
directly upon an optical interface which consists of an orifice from which a constant
flow of argon emerges, deflecting the plasma and providing cooling while allowing
the emitted light from the plasma to enter the optical chamber. Still other designs use
optical fibers to convey some of the light to separate optical cha more modern units,
the separated colours fall upon an array of semiconductor photo detectors such as
CCDs charge coupled devices. In units using these detector arrays, the intensities of
all wavelengths (within the system's range) can be measured simultaneously, allowing
the instrument to analyze for every element to which the unit is sensitive all at once.
Thus, samples can be analyzed very quickly. The intensity of each line is then
compared to previously measured intensities of known concentrations of the elements,
and their concentrations are then computed by interpolation along the calibration lines.
In addition, special software generally corrects for interferences caused by the
presence of different elements within a given sample matrix.
62
Applications:
Examples of the application of ICP-AES include the determination of metals in wine,
arsenic in food, and trace elements bound to proteins. ICP-OES is widely used in
minerals processing to provide the data on grades of various streams, for the
construction of mass balances.
In 2008, the technique was used at Liverpool University to demonstrate that a ChiRho amulet found in Shepton Mallet and previously believed to be among the earliest
evidence of Christianity in England, only dated to the nineteenth century.
ICP-AES is often used for analysis of trace elements in soil, and it is for that reason it
is often used in forensics to ascertain the origin of soil samples found at crime scenes
or on victims etc. Taking one sample from a control and determining the metal
composition and taking the sample obtained from evidence and determine that metal
composition allows a comparison to be made. While soil evidence may not stand
alone in court it certainly strengthens other evidences.
63
REFERENCES
[1]. Nucleonix PC based TL Reader Manual, (2003).
[2]. Murty,K.V.R, Reddy,J.N. Thermoluminescence, Basic Theory & Applications;
(2008)
[3]. B.P.Cullity, Elements of X-ray diffraction, Addison Wesley Inc.,(1978).
[4]. Bish, DL and Post, JE, editors., Modern Powder Diffraction. Reviews in
Mienralogy, v. 20. Mineralogical Society of America(19890).
[5]. Wall, B., Driscoll, C., Strong, J., Fisher, E.: The Suitability of Different
Preparations of Thermoluminescent Lithium Borate for Medical Dosimetry.
Physical Medical biology, 1023-1034,(1982)
[6]. Klug, H. P., and L. E. Alexander.,X-ray diffraction procedures for
polycrystalline and amorphous materials. 2nd ed. Wiley, New York, (1974).
[7]. Moore, D. M. and R. C. Reynolds, Jr. X-Ray diffraction and the identification
and analysis of clay minerals. 2nd Ed. Oxford University Press, New York, (1997).
[8]. Hendra, P.J., Int.J.Vibr.Spec., 5, 5, 2, (2001)
[9]. Capron, M,. Douy, A, J. Am. Ceram. Soc.,85, 12,3036-40,(2002).
[10].T.Katsumata,K.Sasajima,T.Nabae,J.Am.Ceram.Soc.81,413, (1998).
[11].Skoog,Holler and Nieman Principle of instrumental analysis,(2005)
[12].Mansfield,E;Kar,A;Quinn,T.P;Hooker,S.A. Analytical Chemistry,82,24, (2010).
[13].Rod M. Jones, Particle size analysis by laser diffraction:, standard operating
procedures,and Mie theory, Reprinted from American Laboratory,(2003)
[14].Stefnsson.A, Gunnarsson.I,Giroud.N "New methods for the direct determination
of dissolved inorganic, organic and total carbon in natural waters by Reagent-Free Ion
Chromatography and inductively coupled plasma atomic emission spectrometry".
Anal. Chim. Acta,582, (2007).
[15].Mermet, J.M., J. Anal. At. Spectrom. (2005).
[16].Aceto.M, Abollino.O, Bruzzoniti.MC, Mentasti.E, Sarzanini.C,Malandrino.M
Food additives and contaminants, (2002).
[17].Benramdane,L, Bressolle F,Vallon JJ, Journal of chromatographic science,37,
(1999).
64