Technical

Information
TI 094D/24/ae

Electromagnetic flow
measuring system
PROline promag 10 P
Flow measurement in chemical
or process applications

Application
Electromagnetic flowmeter for measuring
liquids with a minimum conductivity of
50 S/cm, such as:

Acids or alkalis
Chemically loaded wastewater
Paints
Sewage sludge

Nominal diameters from 1 to 12


Maximum fluid temperature:
PTFE: 265F (130C)
Sensor fitting lengths to ISO and DVGW

PTFE lining guarantees maximum


resistance in the case of aggressive
fluids and temperatures up to 265F
Used in difficult conditions (in shafts,
areas with vibration, or permanently
under water) - the sensor can be
installed separate from the transmitter
and is availalbe with NEMA 6P (IP 68)
protection.
Service friendly - functional design
makes maintenance work easier HART, Fieldtool and FieldCheck
provide support right down to device
verification without the need for sensor
removal.

Features and benefits


Ideal for simple measuring tasks current output for displaying the
current flow, pulse output for driving
an external totalizer or as status
output
Guaranteed safety of application Promag 10 offers all basic functions
to guarantee a high degree of
reliability and measured value stability
Optimum process control based on an
accuracy of 0.5%

The Power of Know How

PROline Promag 10 P

Function and system design


Measuring principle

Faradays law of induction states that a voltage is induced in a conductor moving in a


magnetic field. In electromagnetic measuring, the flowing medium corresponds to
the moving conductor. The induced voltage is proportional to the flow velocity and is
detected by two measuring electrodes and transmitted to the amplifier. Flow volume
is computed on the basis of the pipes diameter. The constant magnetic field is
generated by a switched direct current of alternating polarity.

Ue = B L v
Q=Av
Ue = induced voltage
B = magnetic induction (magnetic field)
L = electrode gap
v = flow velocity
Q = volume flow
A = pipe cross-section
I = current strength

Measuring system

The measuring system consists of a transmitter and a sensor.


Two versions are available:
Compact version: transmitter and sensor form a single mechanical unit.
Remote version: transmitter and sensor are installed separately.
Transmitter:
Promag 10 (user interface with push buttons for operation, unilluminated two-line
display)
Sensor:
Promag P, 1 to 12 (DN 25 to 300)

Input
Measured variable

Flow rate (proportional to induced voltage)

Measuring range

Typically v = 0.033 to 33 ft/s (0.01 to 10 m/s) with the specified measuring accuracy

Operable flow range

Over 1000 : 1

Endress+Hauser

PROline Promag 10 P

Output
Output signal

Current output:
Active, galvanically isolated, full scale value selectable temperature coefficient:
typically 1.1 A/F (2 A/C)
Resolution: 1.5 A
Active: 4 to 20 mA, RL < 700 (HART, RL 250)
Pulse output:
Passive, open collector, 30 VDC, 250 mA, galvanically isolted.
Alternatively configurable as:
- Pulse output:
Pulse value and pulse polarity selectable, maximum pulse width configurable
(5 to 2000 ms), pulse frequency maximum 100 Hz
or
- Status output:
Configurable e.g. for error messages, Empty Pipe Detection (EPD), recognition
of flow direction, limit values.

Signal on alarm

Current output failure response selectable


Pulse output failure response selectable
Status output non-conductive by fault or power supply failure

Load

See Output signal

Low flow cutoff

Switch point for low flow cutoff are selectable

Galvanic isolation

All circuits for inputs, outputs, and power supply are galvanically isolated from each
other.

Power supply
Electrical connection
measuring unit
a

b
c
d
e
f
g

Cable for power supply: 85 to 250 VAC,


20 to 28 VAC, or 11 to 40 VDC
Terminal No. 1: L1 fo AC, L+ for DC
Terminal No. 2: N for AC, L- for DC
Signal cable: terminals No. 24 to 27
Cover of the electronics compartment
Terminals for the power supply
Terminals for the current output
Service adapter for connecting service interface
FXA 193 (FieldCheck, FieldTool)
Grounding terminal for signal-cable shield

f
e

Maximum cable size, 16 AWG (2.5 mm2)

24 25 26 27 1 2
+ + L1 N
(L+) (L-)

Order variant

Terminal No. (outputs)


24 (+) / 25 (-)

10***-*********A

Pulse output

26 (+) / 27 (-)
Current output
HART

Pulse output (passsive)


Open collector, maximum 30 VDC / 250 mA, galvanically isolated, configurable
Operating mode pulse: pulse frequency maximum 100 Hz
Operaing mode status: yes, configurable
Current output active
Galvanically isolated, active: 4 to 20 mA, RL < 700 (HART, RL 250)
Ground connection, power supply, see diagram above.

Endress+Hauser

PROline Promag 10 P

Electrical connection, remote


version
Coil circuit

Electrode circuit
Esc

Meas.signal Pipe EPD


S1 E1 E2 S2GND E S
6

4 37 36

42 41

yel

grn

brn

wht

b
n.c.

n.c.
5

n.c.
4 37

42 41

E1 E2 GND E

a
b
c
d
n.c.

Cover of the connection compartment


Cover of the sensor connection housing
Signal cable
Coil current cable
Not connected, isolated cable shields

Supply voltage

85 to 250 VAC, 45 to 65 Hz
20 to 28 VAC, 45 to 65 Hz / 11 to 40 VDC

Cable entry

Power supply and signal cable (outputs):


1/2 NPT
Connecting cable for remote version:
1/2 NPT

Cable specifications

Coil cable:
2 x 18 AWG (0.75 mm2) PVC cable with common, braided copper shield,
approximately 0.28 diameter ( approx. 7 mm)
Conductor resistance: 0.011 /ft (37 /km)
Capacitance: core/core, shield grounded: 36 pF/ft (120 pF/m)
Permanent operating temperature: -5 to +180F (20 to +80C)
Signal cable:
3 x 20 AWG (0.38 mm2) PVC cable with common, braided copper shield,
approximately 0.28 diameter ( approx. 7 mm) and individually shielded cores.
With Empty Pipe Detection (EPD): 4 x 20 AWG (0.38 mm2) PVC cable with
common, braided copper shield, approximately 0.28 diameter ( approx. 7 mm)
and individually shielded cores.
Conductor resistance: 0.015 /ft (50 /km)
Capacitance: core/shield: 128 pF/ft (420 pF/m)
Permanent operating temperature: -5 to +180F (-20 to +80 C)
Conductor cross-section: maximum 0.004 in2 (2.5 mm2)

Endress+Hauser

PROline Promag 10 P

1
2
3
4
5
6
7

a = signal cable, b = coil current cable (maximum 16 AWG (2.5 mm2)


1 = Core
5 = Core strengthening
2 = Core insulation
6 = Cable shield
3 = Core shield
7 = Outer jacket
4 = Core jacket

Operation in areas of severe electrical interference:


The measuring device complies with the general safety requirements in accordance
with EN 61010, the EMC requirements of EN 61326, and NAMUR recommendation
NE 21.
Caution:
Grounding is by means of the ground terminals provided for that purpose inside the
connection housing. Keep the stripped and twisted lengths of cable shield to the
terminals as short as possible.
Power consumption

85 to 250 VAC: < 12 VA (including sensor)


20 to 28 VAC: < 8 VA (including sensor)
11 to 40 VDC: < 6 W (including sensor)
Switch-on current:
maximum 3.3 A (< 5 ms) at 24 VDC
maximum 5.5 A (< 5 ms) at 28 VAC
maximum 16 A (< 5 ms) at 250 VAC

Power supply failure

Lasting minimum 1/2 power cycle:


EEPROM retain the measuring system data in the event of a power supply failure

Potential equalization

Standard case
Perfect measurement is only ensured when the medium and the sensor have the
same electrical potential. Most Promag sensors have a standard installed reference
electrode which guarantees the required potential matching. This usually means that
additional potential matching measures are unnecessary.
NOTE: For installation in metal pipes, it is advisable to connect the ground terminal
of the transmitter housing to the piping.

Caution:
For sensors without reference electrodes or without metal process terminals, carry
out potential matching as per the instructions for special cases described below.
These special measures are particularly important when standard grounding practice
cannot be ensured or extremely strong matching currents are expected.
Endress+Hauser

PROline Promag 10 P

Metallic, ungrounded pipes


In order to avoid errors in measurement, use grounding cables to connect each
sensor flange to its corresponding pipe flange and ground the flanges. Connect the
transmitter or sensor connection housing, as applicable, to ground by means of the
grounding terminal provided for that purpose.
Note:
The grounding cable for flange-to-flange connections can be ordered separately as
an accessory from Endress+Hauser.
he grounding cable is in direct connection with the conductive flange
coating and is secured by the flange screws.

9 AWG Cu

Potential equalization with equalizing currents in metallic, non-grounded piping systems

9 AWG Cu

Plastic pipes and lined pipes


Normally, potential is matched using the reference electrodes in the measuring tube.
However, in exceptional cases it is possible that, due to the grounding plan of a
system, large matching currents flow over the reference electrodes. This can lead to
destruction of the sensor, e.g. through electrochemical decomposition of the electrodes. In such cases, e.g. for fiberglass or PVC piping, it is recommended that you
use additional ground disks for potential matching.
When using ground disks, note the following points:
Ground disks can be ordered from Endress+Hauser as an accessory
Ground disks (including seals) increase the installation length. Refer to the
disk dimensions.
Caution:
Risk of damage by electro-chemical corrosion. Note the electro-chemical series of
metals, if the grounding rings and measuring electrodes are made of different
materials.
6

Endress+Hauser

PROline Promag 10 P

Pipes with cathodic protection


If the medium cannot be grounded for process-related reasons, the measuring
device must be installed in such a way as to be potential-free:
When installing the measuring device, make sure that there is an electrical
connection between the two piping runs, 9 AWG copper wire (6 mm2).
Make sure that the installation materials do not establish a conductive connection
to the measuring device and that the installation materials withstand the
tightening torques applied when the threaded fasteners are tightened.
Also comply with the regulations applicable to potential-free installation.
1
2

9 AWG Cu
1 = isolating transformer, 2 = electrically insulated

Performance characteristics
Reference conditions

To DIN 19200 and VDI/VDE 2641:


Medium temperature: +28 C 2 K
Ambient temperature: +22 C 2 K
Warm-up time: 30 minutes
Installation:
Inlet run >10 x Dia
Outlet run > 5 x Dia
Sensor and transmitter grounded.
Sensor centered relative to the pipe.

Measured error

Pulse output: 0.5% o.r. 0.02% of max. full scale (o.r. = of reading)
Current output: plus typically 5 A
Supply-voltage fluctuations have no effect within the specified range.
[%]
2.5
2.0
0.5 %
1.5
1.0
0.5
0
0

3
(1)

6
(2)

13
(4)

20
(6)

26
(8)

33
(10)

v ft/s (m/s)
Measured error in [%] of reading

Repeatability

Endress+Hauser

Maximum 0.2% o.r. 0.02% of maximum full scale (o.r. = of reading)

PROline Promag 10 P

Operating conditions: Installation


Installation instructions

Mounting location
Correct measuring is possible only if the pipe is full. Avoid the following locations:
Highest point in the run. Risk of air accumulating
Directly upstream from an open pipe outlet in a down pipe

Installation of pumps
Do not install the sensor on the inlet side of a pump. This precaution is to avoid low
pressure and the consequent risk of damage to the lining of the measuring tube.
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or peristaltic pumps. The sensor is shock and vibration resistant for
acceleration up to 2 g in accordance with IEC 68-2-6.

Partially filled pipes


Partially filled pipes with gradients necessitate a drain-type configuration. The Empty
Pipe Detection (EPD) function offers additional protection by detecting empty or
partially filled pipes.
Caution:
To avoid the risk of solids accumulating, do not install the sensor at the lowest point
in the drain. It is advisable to install a cleaning valve.

2xD

5xD

Endress+Hauser

PROline Promag 10 P

Vertical pipes
Install a siphon (2) or a vent valve (1) downstream of the sensor in vertical pipes
longer than 15 feet (5 meters). This precaution is to avoid low pressure and the
consequent risk of damage to the lining of the measuring tube. These measures
also prevent the system losing prime, which could cause air inclusions.
1

> 15 ft
(5 m)

1 = vent valve, 2 = siphon

Orientation
An optimum orientation helps avoid gas and air accumulations and deposits in the
measuring tube. Promag, nevertheless, supplies a range of options and accessories
for correct measuring of problematic mediums.

Vertical orientation:
This orientation is ideal for self-emptying piping systems and for use in conjunction
with Empty Pipe Detection.

Horizontal orientation:
The measuring electrode-plane should be horizontal. This prevents brief insulation of
the two electrodes by entrained air bubbles.
Caution:
Empty Pipe Detection functions correctly only when the measuring device is installed
horizontally and the transmitter housing is facing upward. Otherwise, there is no
guarantee that Empty Pipe Detection will respond if the measuring tube is only
partially filled or empty.

1 = EPD electrode (Empty Pipe Detection)


2 = Measuring electrodes (signal detection)
3 = Reference electrode (potential equalization)

Endress+Hauser

PROline Promag 10 P

Vibrations
Secure the piping and the sensor if vibration is severe.
Caution:
It is advisable to install the sensor and transmitter separately if vibration is excessively severe. The sensor is shock and vibration resistant for acceleration up to 2 g
in accordance with IEC 68-2-6.
> 30 ft
(10 m)

Inlet and outlet runs

If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows,
etc. Compliance with the following requirements for the inlet and outlet runs is
necessary in order to ensure measuring accuracy:
Inlet run 5 x pipe diameter
Outlet run 2 x pipe diameter

5xD

Adapters

10

2xD

With the help of the appropriate adapters (reducers and expanders) the sensor can
be mounted in a pipeline of a larger diameter. For slowly flowing fluids, the resulting
higher velocity increases the measuring accuracy.

Endress+Hauser

PROline Promag 10 P

Length of connecting cable

Permissible cable length Lmax depends on the conductivity of the medium. A


minimum conductivity of 50 S/cm is required.

Esc

200

L max

100
50
Ft (m)
165
(50)

330
(100)

650
(200)

L max

Gray shaded area = permissible range for medium conductivity


Lmax = length of connecting cable in feet (m)
Medium conductivity in S/cm

In order to ensure measuring accuracy, comply with the following instructions when
installing the remote version:
Secure the cable run or route the cable in a conduit. Movement of the cable
can falsify the measuring signal, particularly if the conductivity of the medium
is low.
Route the cable well clear of electrical machines and switching elements.
Ensure ground equalization between the sensor and transmitter, if necessary.
When the Empty Pipe Detection function is switched on, the maximum connecting
cable length is 33 ft. (10 m).

Operating conditions: Environment


Ambient temperature

-5 to +140F (-20 to +60C) for sensor and transmitter


Install the device at a shady location. Avoid direct sunlight, particularly in warm
climatic regions.

Storage temperature

15 to + 120F (-10 to +50C), 70F (20C) is the preferable storage temperature.

Degree of protection

Standard: NEMA 4X (IP 67) for transmitter and sensor


Optional: NEMA 6P (IP 68) for Promag P sensor, remote version

Shock and vibration resistance

Acceleration up to 2 g in accordance with IEC 68-2-6

Electromagnetic compatibility
(EMC)

To EN 61326
Emission according to limit value for industrial environments EN55011

Process conditions
Medium temperature range

The -40 to +265F (-40 to +130C) for PTFE liner. Refer to diagrams below for
restrictions

265

265

160

160

140

140

120

120
Esc

TA [F]

80
60
40

32

20

Endress+Hauser

80

40

32

Compact version Promag P


TA = ambient temperature
TF = fluid temperature

PTFE

PTFE

TA [F]

100

0
-10

20

60

100

140

180 220

-5

230

TF [F]

260

300

Remote version Promag P


TA = ambient temperature
TF = fluid temperature

0
-10
-5

20

60

100

140

180

220 260

300

TF [F]

11

PROline Promag 10 P

Conductivity

Minimum conductivity: 50S/cm


Note that in the case of the remote version, the minimum conductivity is also
influenced by the length of the connecting cable. Refer to Length of connecting
cable on Page 11.

Medium pressure range


(nominal pressure)
Pressure tightness (liner)

ANSI B 16.5:
Class 150, 1 to 12
Size

Measuring Resistance to partial vacuum


tube lining of measuring tube lining
Limit values for abs. pressure (psia)
at various fluid temperatures
75F
175F
212F
265F
1
PTFE
0
0
0
1.4
1-1/2 PTFE
0
0
0
1.4
2
PTFE
0
0
0
1.4
3
PTFE
0
*
0.6
1.9
4
PTFE
0
*
1.9
2.5
6
PTFE
1.9
*
3.5
5.6
8
PTFE
2.9
*
4.2
5.9
10
PTFE
4.8
*
5.8
7.7
12
PTFE
5.8
*
7.3
9.1
* No value can be specified.

Flow limit

As a rule, the pipeline diameter determines the sensor diameter. With a known
flowrate and the help of the Endress+Hauser Flow Applicator, the best size can be
selected. Generally, a velocity range of 6 to 10 ft/sec should be selected.
A necessary increase in velocity can be achieved through a reduction of the sensor
diameter. The higher installation expense is normally balanced by the lower sensor
cost.
The flow velocity should also be determined by the fluids physical properties:
v < 6 ft/sec, for abrasive mediums such as potters clay, lime milk, ore slurry, etc.
v > 6 ft/sec, for fluids causing build-up such as wastewater sludge, etc.
Nominal
diameter
inch
1
1-1/2
2
3
4
6
8
10
12

Pressure loss

12

mm
25
40
50
80
100
150
200
250
300

Recommended
flow rate
Min./max. full scale value
(v ~ 1.0 or 33 ft/s)
2.5 to 80 gpm
7 to 190 gpm
10 to 300 gpm
24 to 800 gpm
40 to 1250 gpm
90 to 2650 gpm
155 to 4850 gpm
250 to 7500 gpm
350 to 10,600 gpm

Factory settings
Full scale value
(v ~ 8 ft/s)
18 gpm
50 gpm
75 gpm
200 gpm
300 gpm
600 gpm
1200 gpm
1500 gpm
2400 gpm

Pulse weighting
(~ 2 pulses/s)
0.20 gal
0.50 gal
0.5 gal
2 gal
2 gal
5 gal
10 gal
15 gal
25 gal

Creepage
(~ 0.1 ft/s)
0.25 gpm
0.75 gpm
1.25 gpm
2.5 gpm
4.0 gpm
12 gpm
15 gpm
30 gpm
45 gpm

No pressure loss if the sensor is installed in a pipe of the same nominal diameter.
Pressure losses occur in configurations incorporating adapters (reducers, expanders). Contact your Endress+Hauser representative or Endress+Hauser for information.

Endress+Hauser

PROline Promag 10 P

Mechanical construction
Design, dimensions

Remote transmitter dimensions


Dimensions are in inches (mm)
6.34 - 7.13
(161 - 181)
7.01
(178)

0.78 - 1.18
(20 - 30)

4.76
(121)

Esc

12.9
(327.5)

6.99
3.94 5.24 (177.5)
(100) (133)

ANSCHLUSSKLEMMEN - FIELD TERMINALS

4.49
(113)

5.91
(150)

5.31
(135)

0.34
(8.6)
M8

3.94
(100)
4.84
(123)

0.98
(25)

Promag 10 remote transmitter

A
9.76 0.08
(248 2)

Esc

ANSCHLUSSKLEMMEN - FIELD TERMINALS

B
9.37
(238)

Esc

ANSCHLUSSKLEMMEN - FIELD TERMINALS

Promag 10 remote transmitter mounting


A Direct wall mounting
B Pipe mounting

Endress+Hauser

13

PROline Promag 10 P

Promag 10 P (compact version)


6.34 - 7.13
(161 - 181)
7.01
(178)

4.76
(121)

0.78 - 1.18
(20 - 30)

5.91
(150)

Esc

B
A

E
L
(DVGW)

Nominal
Size

inches

mm

in

1
1-1/2
2
3
4
6
8
10
12

25
40
50
80
100
150
200
250
300

7.87
7.87
7.87
7.87
9.84
11.8
13.8
17.7
19.7

mm
200
200
200
200
250
300
350
450
500

in
13.0
13.0
13.0
15.0
15.0
18.2
20.3
22.3
24.3

B
mm
331
331
331
381
381
462
517
567
617

in

mm

in

mm

9.72
9.72
9.72
10.7
10.7
12.3
13.3
14.2
15.2

247
247
247
272
272
312
337
362
387

3.31
3.31
3.31
4.29
4.29
5.90
7.08
8.07
9.05

84
84
84
109
109
150
180
205
230

in
4.72
4.72
4.72
7.08
7.08
10.2
12.7
15.7
18.1

E
mm
120
120
120
180
180
260
324
400
460

in

Weight *
mm
94
94
94
94
94
140
156
156
166

3.70
3.70
3.70
3.70
3.70
5.51
6.14
6.14
6.53

lb

kg

7.5
16
9.6
21
10.8
24
14.2
31
16.2
35
25.7
57
45.2
99
166 75.2
243 110.2

NOTE: The fitting length (L) is always the same, regardless of the pressure rating.
* Weight data valid for 150 lb ANSI flanges, standard pressure ratings and without
packaging material

14

Endress+Hauser

PROline Promag 10 P

Promag 10 W (remote version)

5.08"
(129)

6.73"
(171)
5.63"
(143)
4.01"
(102)
B
A

E
L

Nominal
Size

inches

mm

in

1
1-1/2
2
3
4
6
8
10
12

25
40
50
80
100
150
200
250
300

7.87
7.87
7.87
7.87
9.84
11.8
13.8
17.7
19.7

mm
200
200
200
200
250
300
350
450
500

in

mm

in

mm

in

mm

in

mm

in

11.2
11.2
11.2
13.2
13.2
16.4
18.6
20.5
22.5

286
286
286
336
336
417
472
522
572

7.95
7.95
7.95
8.94
8.94
10.5
11.5
12.5
13.5

202
202
202
227
227
267
292
317
342

3.31
3.31
3.31
4.29
4.29
5.90
7.08
8.07
9.05

84
84
84
109
109
150
180
205
230

4.72
4.72
4.72
7.08
7.08
10.2
12.7
15.7
18.1

120
120
120
180
180
260
324
400
460

3.70
3.70
3.70
3.70
3.70
5.51
6.14
6.14
6.53

Weight *
mm
94
94
94
94
94
140
156
156
166

lb

kg

12
16
19
24
31
52
95
161
238

5.3
7.4
8.6
12
14
23.5
43
73
108

NOTE: The fitting length (L) is always the same, regardless of the pressure rating.
* Weight data valid for 150 lb ANSI flanges, standard pressure ratings and without
packaging material.
Remote transmitter weighs 6.8 lbs (3.1 kg).

Endress+Hauser

15

PROline Promag 10 P

Ground disk, 1 to 12

B
0.26"
(6.5)
0.59"
(15)

0.39"
(10)
D

0.08"
(2)
A

Nominal
1)
size
in
1
1-1/2
2
3
4
6
8
10
12 2)
1)

2)

Materials

16

1.18
1.75
2.22
3.35
4.33
6.42
8.29
10.4
12.5

mm

in

mm

in

30
44.5
56.5
85
110
163
210.5
265
317

2.44
3.23
3.98
5.16
6.14
8.54
10.5
12.9
14.8

62
82
101
131
156
217
267
328
375

3.05
3.98
4.55
6.08
7.34
10.1
11.3
14.1
16.3

H
mm
77.5
101
115.5
154.5
186.5
256
288
359
413

in
3.44
4.06
4.25
5.31
6.02
7.24
8.07
9.45
10.7

mm
87.5
103
108
135
153
184
205
240
273

Ground disks, apart from those for 12", can be used for all
supplied flange standards / pressure ratings
Class 150

Transmitter housing:
Compact housing: powder coated die-cast aluminum
Remote version housing: powder coated die-cast aluminum
Sensor housing:
1 to 12: powder coated die-cast aluminum
Measuring tube:
304 SS or 304L SS; flange material with Al/Zn protective coating
Flanges:
ANSI: A105 CS with Al/Zn protective coating;
Electrodes:
316L SS standard, optional Alloy C-22
Seal:
Seals to DIN 2690
Ground disks:
316L SS or Alloy C-22

Endress+Hauser

PROline Promag 10 P

Material load diagram

Flange material: carbon steel A 105


to ANSI B 16.5
[psi]

[bar]

870

60

725

50

580

40

435

30
20

290
Class 150

145

10
0

0
-76
-60

-40
-40

-4
-20

32
0

68
20

104
40

140
60

176 212
80 100

248
120

284
140

320 [F]
160 [C]

Fitted electrodes

Measuring, reference and Empty Pipe Detection electrodes:


316L SS, standard
Alloy C-22

Process connection

Flange connections, ANSI (DIN and JIS flanges also available)

Surface roughness

Electrode portion in contact with medium:


316L or Allow C-22: 16 inch, 240 grit ( 0.4 m)

Human interface
Display elements

Liquid crystal display, unilluminated, two-line, 16 characters per line


Display (operating mode) preconfigured: volume flow and totalizer status
1 totalizer (7-digit plus 7-digit overflow with sign and units)

Operating elements

Local operation with three push buttons (-, +, E)

Remote operation

Remote control via HART and FieldTool

Certificates and approvals


CE mark

By attaching the CE mark, Endress+Hauser confirms that the instrument fulfills all
the requirements of the relevant EC directives.

Other standards and guidelines

EN 60529:
Degrees of protection by housing (IP code)
EN 61010:
Protection Measures for Electrical Equipment for Measurement, Control, Regulation
and Laboratory Procedures.
EN 61326 (IEC 1326):
Emmision to class A requirements Electromagnetic compatibility (EMC requirements).
ANSI / ISA-S82.01:
Safety Standards for Electrical and Electronic Test, Measuring, Controlling and
related Equipment
- General Requirements. Pollution degree 2, Installation Category II.
CAN / CSA-C22.2 No. 1010. 1-92:
Safety requirements for Electrical Equipment for Measurement and Control and
Laboratory Use. Pollution degree 2, Installation Category II.

Pressure measuring device


approval

Devices with a nominal diameter smaller than or equal to 1 (DN 25) correspond to
Article 3 (3) of the EC directive 97/23/EC (Pressure Equipment Directive). For larger
nominal diameters, certified flowmeters to Category III are available as an option if
necessary (depends on fluid and operating pressure). All devices are applicable for
all fluids and instable gases on principle and have been designed and manufactured
in accordance to sound engineering practice.

Endress+Hauser

17

PROline Promag 10 P

Accessories

Supplemental documentation

18

Various accessories, which can be ordered from Endress+Hauser, are available for
the transmitter and the sensor. Contact your local representative or Endress+Hauser
for detailed information.

Technical Information Promag 10 W (TI 093D/24/ae)


Technical Information Promag 10 H (TI 095D/24/ae)
Operating instructions Promag 10 (BA 082D/06/en)

Endress+Hauser

PROline Promag 10 P

Ordering information
1

10

11

12

13

Promag 10 P 1

2
3
4

5
6

10
11

12

13

Endress+Hauser

Nominal diameter / maximum full scale


25 1 / 80 gpm
40 1-1/2 / 200 gpm
50 2 / 290 gpm
80 3 / 790 gpm
1H 4 / 1240 gpm
1F 6 / 2640 gpm
2H 8 / 5280 gpm
2F 10 / 7930
3H 12 / 10,600 gpm
Liner
E
PTFE
Process connection
L
Class 150 ANSI B16.5 A105 flange
Electrodes / material
0
Measuring, reference and EPD electrodes / 316L SS
1
Measuring, reference and EPD electrodes / Alloy C-22
G
Measuring, reference and EPD electrodes, bullet nose / 316L SS
H
Measuring, reference and EPD electrodes, bullet nose / Alloy C-22
Calibration
A
0.5% calibration
Certificates
1
Standard, without certificate
p
Approval according to PED Cat. II;III (nor for 1, only for ANSI/DIN flanges,
only for approval A)
Approvals
A
For use in nonhazardous area
R
FM approved non-incendive Class I, Division 2 / CSA Class I, Division 2
Protection type
A
NEMA 4X (IP 67), compact, aluminum field housing
G
NEMA 4X (IP 67), remote, aluminum field housing
N
NEMA 6P (IP 68), remote, aluminum field housing
Cable for remote version
0
Without cable
2
30 ft (10 m) coil and signal cable
4
Specify length in meters, coil and signal cable (priced per meter)
5
Specify length in feet, coil and signal cable (priced per foot)
Cable entries
B
1/2 NPT threaded
Display / power supply / operation
0
Without display / 85 to 250 VAC
1
Without display / 20 to 28 VAC, 11 to 40 VDC
4
With display / 85 to 250 VAC / push-button operation
5
With display / 20 to 28 VAC, 11 to 40 VDC / push-button operation
X
Sensor only, without transmitter (only up to 8 size)
Software
A
Standard software
X
Sensor only
Outputs
A
Current HART (active), pulse (passive)
X
Sensor only

19

For application and selection assistance,


in the U.S. call 888-ENDRESS
For total support of your installed base, 24 hours
a day, in the U.S. call 800-642-8737
Visit us on our web site, www.us.endress.com
United States

Canada

Mexico

Endress+Hauser, Inc.
2350 Endress Place
Greenwood, IN 46143
Phone: (317) 535-7138
888-ENDRESS
FAX: (317) 535-8498

Endress+Hauser Canada
1075 Sutton Drive
Burlington, ON L7L 5Z8
Phone: (905) 681-9292
800-668-3199
FAX: (905) 681-9444

Endress+Hauser
Paseo del Pedregal No. 610
Col. Jardines del Pedregal
01900, Mexico D.F.
Mexico
Phone: (525) 568-2405
FAX: (525) 568-7459

Endress+Hauser
The Power of Know How

TI 094D/24/ae/05.04
2004 Endress+Hauser, Inc.