Vous êtes sur la page 1sur 1

Tolerance classes for Pt100 thermometers

according to IEC 60751, 2008-07, edition 2.0

A. Graphical form
IEC 60751, 2008, ed. 2

Temperature (
C)
7

Temperature range for film resistors


Temperature range for film detectors
Extended temperature range for wirewound resistors Class C

detectors
Extended range for wirewound detectors

Class C

Tolerance (C)

5
4

Class B
Class B

3
2

Class A

600

550

500

450

400

350

300

200

150

100

50

-50

-100

-150

250

Class
Class
AAAA

0
-200

Class A

Temperature (C)

B. Table
Tolerance class
AA
A
B
C

Temperature range of validity


C
Wire wound resistors
Film resistors
-50 to 250
0 to 150
-100 to 450
-30 to 300
-196 to 600
-50 to 500
-196 to 600
-50 to 600

Tolerance values

C
(0.1 + 0.0017 | t |)
(0.15 + 0.002 | t |)
(0.3 + 0.005 | t |)
(0.6 + 0.01 | t |)

| t | = temperature in
C without regard to sign.
The temperature -196
C is close to the boiling point of nitrogen (N) which is used for calibration.

C. Comments
Compared to the earlier versions of this standard the present one has been adapted to cover more models of
Pt100 detectors (resistors). Where resistors are fulfilling extended or smaller ranges of validity or other
tolerance classes, this must be clearly stated between manufacturer and user.
The tolerance classes apply to all detector values (R 0) at 0
C, i. e. Pt100, Pt1000 and others.
The relationship temperature-resistance has not been changed compared to the previous standard.

2009-06-09