Vous êtes sur la page 1sur 116

Operator Manual

POLYMETRON 9210

221=192=010 - Revision J - 20/11/2007

SILKOSTAT 9210

Notice for readers

The information contained in this manual was as


complete and exact as possible at the time of printing. If
the behaviour of this product during its use is different
from that described in the documentation, it is perhaps
because your manual is out of date. In this case, please
contact your Hach Ultra Analytics agent immediately to
remedy the problem.
If the instrument is used in a way not specified by the
manufacturer, the protection ensured by the instrument
could be compromised.
Hach Ultra Analytics reserves the right to make
modifications and changes to the equipment and
software comprising the product described in this manual.
Thank you.

Restriction of Hazardous Substances


Note:
The following only applies to exports of the product into the Peoples Republic of China.

Marking
Products contain toxic or hazardous substances or elements.

Environment Protection Use Period Marking (years).

Toxic or Hazardous Substances and Elements

Part Name

Hexavalent
Lead Mercury Cadmium Chromium
(Cd)
(Pb)
(Cr VI)
(Hg)

Polybrom
Biphenyls
(PBB)

Transmitter box

CPU PCB
(with battery)

Power PCB

RS 485

Profibus

PCB CAN

PCB Local controller

PCB Measure

PCB Pumps

PCB Mixer

PCB Channel

Polybrom
Diphenyls
(PBDE)

O: Indicates that this toxic or hazardous substance contained in all homogeneous material for
this part is below the limit requirement
X: Indicates that this toxic or hazardous substance contained in at least one of the
homogeneous materials used for this part is above the limit requirement

SILKOSTAT 9210

Safety Precautions
Please read this entire manual before unpacking, setting up, or operating this
instrument. Pay particular attention to all danger and caution statements.
Failure to do so could result in serious injury to the operator or damage to the
equipment.
To ensure the protection provided by this equipment is not impaired, do not
use or install this equipment in any manner other than that which is specified
in this manual.

Use of Hazard Information


If multiple hazards exist, this manual will use the signal word (Danger,
Caution, Note) corresponding to the greatest hazard.
DANGER
Indicates a potentially or imminently hazardous situation which, if not
avoided, could result in death or serious injury.
CAUTION
Indicates a potentially hazardous situation that may result in minor or
moderate injury.
NOTE
Information that requires special emphasis.

Precautionary Labels
Read all labels and tags attached to the instrument. Personal injury or
damage to this instrument could occur if not observed.

This symbol, if noted on the instrument, references the


instruction manual for operation and / or safety information.
This symbol, if noted on the instrument, marks the wiring
connection point for Protective Earth.
This symbol, if noted on the product, indicates the need for
protective eye wear.

Electrical equipment marked with this symbol may not be


disposed of in European public disposal systems after
12 August of 2005. In conformity with European local and
national regulations (EU Directive 2002/96/EC), European
electrical equipment users must now return old or end-of life
equipment to the Producer for disposal at no charge to the
user.
Note: For return for recycling, please contact the equipment
producer or supplier for instructions on how to return end-oflife equipment for proper disposal.
Important document. Retain with product records.

ii

SILKOSTAT 9210

!
!

Power Connections. on page 20.

Wiring the Instrument for Power on page 17.

Reagents preparation on page 23.

Maintenance on pages 9 and A2.

Maintenance
The integrity of electrical connections and protection instruments of the
system should be checked periodically. All necessary measures should be
taken to remedy any inadequate condition.

Service and repairs


None of the instruments components can be serviced by the user. Only
personnel from Hach Ultra Analytics or its approved representative(s) is
(are) authorised to attempt repairs to the system and only components
formally approved by the manufacturer should be used. Any attempt at
repairing the instrument in contravention of these principles could cause
damage to the instrument and corporal injury to the person carrying out the
repair. It renders the guarantee null and void and could compromise the
correct working of the instrument and the electrical integrity or the CE
compliance of the instrument.
If you have any problems of installation, starting or using the instrument
please contact the company that sold it to you. If this is not possible, or if the
results of this approach are not satisfactory, please contact the manufacturers
Customer Service.

Hach Ultra Analytics


6 route de Compois CP212
CH-1222 Vesenaz Geneva
Tl. + 41 22 855 91 00
Fax + 41 22 855 91 99

iii

SILKOSTAT 9210

CONTENTS
Page
1.

Presentation of the instrument............................................................................................................... 5


Operation ................................................................................................................................................... 6
Calibration .................................................................................................................................................. 6

2.

Technical specifications ......................................................................................................................... 7


Specifications ............................................................................................................................................. 7
Dimensions of the analyser...................................................................................................................... 10

3.

Installation .............................................................................................................................................. 13
Unpacking ................................................................................................................................................ 13
Inspection................................................................................................................................................. 13
Pre-installation ......................................................................................................................................... 13
Presentation of the analyser .................................................................................................................... 14
Positioning the canisters .......................................................................................................................... 15
Connecting the sample ............................................................................................................................ 16
Connecting the evacuation (drain) tube................................................................................................... 16
Mains power connection .......................................................................................................................... 17
RS485 connections (optional extra)......................................................................................................... 19
User signals connections ......................................................................................................................... 20

4.

Starting-up .............................................................................................................................................. 23
Reagent preparation ................................................................................................................................ 23
Reagent concentrations (by type).................................................................................................. 23
Reagent preparation (makes 2 litres) ............................................................................................ 23
Calibration solution ........................................................................................................................ 24
Positioning the canisters .......................................................................................................................... 25
Using the menu........................................................................................................................................ 27
Language programming........................................................................................................................... 28
Time and date programming.................................................................................................................... 28
Starting up the analyser ........................................................................................................................... 29
Starting a measurement........................................................................................................................... 30

5.

Analyser programming.......................................................................................................................... 31
Analyser configuration flow chart............................................................................................................. 31
User Interface .......................................................................................................................................... 32
Main display................................................................................................................................... 32
Definition of the display screen 2 (Disp2) ...................................................................................... 33
Definition of the display screen 3 (Disp3) : alarms ........................................................................ 34
Definition of the display 4 (Disp 4) : trend curve............................................................................ 35
Modification of a value ................................................................................................................... 35
Main menu ............................................................................................................................................... 36
Calibration menu............................................................................................................................ 37
Execution primary sub menu ................................................................................................. 38
Execution manual sub menu.................................................................................................. 39
Programming sub menu (automatic execution) ..................................................................... 40
Calibration parameters sub menu.......................................................................................... 42
Calibration historic sub menu................................................................................................. 42
Manual automatic calibration results sub menu.................................................................. 43
Maintenance menu ........................................................................................................................ 45
Start up sub menu.................................................................................................................. 45
Reagents refill sub menu ....................................................................................................... 46
Adjust volumes....................................................................................................................... 46
Tubes priming ........................................................................................................................ 47
Extended stop sub menu ....................................................................................................... 47

Page 1

SILKOSTAT 9210

Tubes cleaning sub menu...................................................................................................... 48


Measure / Photometer test sub menu.................................................................................... 48
Measure / Photometer test sub menu.................................................................................... 49
Pumps-others sub menu........................................................................................................ 49
Solenoid valves sub menu ..................................................................................................... 50
Relays sub menu ................................................................................................................... 50
Logical inputs sub menu ........................................................................................................ 50
Measure Menu............................................................................................................................... 51
Measuring mode sub menu ................................................................................................... 51
Grab sample sub menu.......................................................................................................... 52
Historic sub menu .................................................................................................................. 52
Programming.......................................................................................................................... 53
Read ..................................................................................................................................... 53
ADC values sub menu ........................................................................................................... 53
Alarms Menu.................................................................................................................................. 54
Alarms parameters 1 to 6 sub menu...................................................................................... 54
Warning alarm sub menu....................................................................................................... 56
System alarm sub menu ........................................................................................................ 56
mA outputs menu........................................................................................................................... 58
Outputs affect 1 to 6 sub menu.............................................................................................. 58
Special prog sub menu .......................................................................................................... 60
Test sub menu ....................................................................................................................... 60
Communication menu.................................................................................................................... 61
Choice of the measure sequence menu........................................................................................ 61
Channel activation sub menu................................................................................................. 61
Sequence sub menu .............................................................................................................. 62
Service menu................................................................................................................................. 63
Time sub menu ...................................................................................................................... 63
Display sub menu .................................................................................................................. 63
Customize ......................................................................................................................... 64
Service / Code sub menu....................................................................................................... 64
Soft issues sub menu............................................................................................................. 64
Default / values sub menu ..................................................................................................... 65
Adjust mA sub menu.............................................................................................................. 65
Factory sub menu .................................................................................................................. 66
Solenoid valves................................................................................................................. 66
Pumps............................................................................................................................... 67
Measure ............................................................................................................................ 67
Configuration sub menu......................................................................................................... 67
6.

Maintenance and exploitation............................................................................................................... 69


Reagent refills .......................................................................................................................................... 69
Prolonged stopping .................................................................................................................................. 71
Restarting after a prolonged stopping...................................................................................................... 72
Tube cleaning .......................................................................................................................................... 73

7.

Troubleshooting..................................................................................................................................... 75
Possible sources of the problem.............................................................................................................. 75
Possible solutions .................................................................................................................................... 76
Typical measurement values during calibration....................................................................................... 78

Page 2

SILKOSTAT 9210

Appendix A : Spare parts of the 9210 SILKOSTAT ....................................................................................A1


Appendix B : Safety Precautions .................................................................................................................B1
Appendix C : Table of users configuration ................................................................................................C1
Appendix D : Pressure range change..........................................................................................................D1
Appendix E : Preparation of the tubes for insertion in the fast connections ..................................... E1

Page 3

SILKOSTAT 9210

1.

Presentation of the instrument

Figure 1

Page 5

SILKOSTAT 9210

Operation
The 9210 can analyse up to six different samples. The sample to be
analysed circulates in a fast loop permitting rapid renewal of the sample.
Adjustment of the flow is carried out with the help of a needle valve (2). At
the beginning of the analysis, the sample is introduced into the
measurement cell (4) with the help of a solenoid valve (5).
The 1st reagent is added with the pump (9). The silica contained in the
sample now reacts with the molydate and forms the silicomolybdic
complex. The reagent time is relatively long.
Oxalic acid is then added to avoid the interference of phosphates and to
intensify the colour.
The silicomolybdic complex is reduced to a blue molybdenum complex by
means of ferrous ions. A photometric measurement is carried out at the
end of the reaction.
Figure 2

Working principal

3
3

3
3
3

9
5
11
12

4
7

13

10
8
12345-

Overflow tank with level detector


Flow adjustment valve
Sample input 1 to 6
Photometric measurement cell
Measuring solenoid valve

6789-

Mixing motor
Magnetic bar
Drain
Reagent pump

10 11 12 13 -

Calibration solution canister


Calibration pump
Photometer
Drain pump of measurement cell

Calibration
In order to ensure correct functioning, the analyser should be calibrated
periodically. The zero is realised chemically and internally in the analyser.
The system slope is controlled in comparison to a standard solution of
known concentration.

Page 6

SILKOSTAT 9210

2.

Technical specifications

Specifications
Measurement range (two different versions) :
- 0 1000 ppb
- 0 5000 ppb
This instrument complies with the following standards :
- Immunity to
interference

electromagnetic EN 50082-2 et EN 50082-1

- Electromagnetic emissions

EN 50081-1 et EN 50081-1

- Low voltage standard

IEC61010-1

SAMPLE
Number of channels

16

Measurement cycle

< 10 min / channel

Sample pressure

0.2 to 6 bar
(3 to 87 psi)

Temperature

5 - 50 C (41 122F)

Sample flow

10 to 20 l / hour

CONNECTIONS
Sample admission

tube PE 4/6

Sample evacuation

tube PE 12/17 silicone

CONDITIONS OF USE
Ambient temperature

5 to + 45C (41 to 113F)

- use inside
- fluctuations 10 % of the power
supply voltage
- overvoltage category 2
- degree of pollution 2 (according
to CEI 664)
Altitude

2000 m

Relative humidity

10 to 80 %

WEIGHT
Panel analyser without canister

13 kg

Cabinet analyser without canister

65 kg

Full canisters

8 kg

Page 7

SILKOSTAT 9210

ANALYSIS
Value measured

dissolved SiO2

Measurement field

0 1000 ppb

0 5000 ppb

Repeatability

2 % or 0,5 ppb

2 % or 2 ppb

Detection limit

0,5 ppb

2 ppb

Cycle time

10 minutes per channel

INPUTS / OUTPUTS
Alarms

Number : 8
Functions :

- 6 threshold relays
/ lack of sample /
active channel
- System alarm
3 contacts
- 1 warning relay

Cut-off power : 30 VDC, 0.5 A


maximum
Logic inputs

Number : 12
Remote clearance
Inhibition of samples

Serial input interface

RS 485, Modbus communication


protocol

Outputs 4 / 20 mA

Number : 8
4/20 mA / 0/20 mA
programmable.
Galvanically insulated
Maximum charge of 800 ohms

MATERIALS & PROTECTIONS

Page 8

Transmitter

Aluminium with polyester paint

Panels and other boxes

Polystyrene

Protection transmitter box

IP 65

Protection cabinet (optional)

IP 54

Cabinet front door

Plexiglas

SILKOSTAT 9210

MAINTENANCE

Calibration

Chemical zero, slope with


calibration solution

Maintenance

No particular maintenance is
necessary. Cleaning can be done
with a soft non-aggressive cloth.

Reagent consumption

Approx. 1 l per month and per


reagent

Calibration solution consumption

Approx. 200 ml / calibration

Time safeguard lithium battery

CR 1220 (3 V)

Caution :
There are risks of explosion if the replaced battery does not
correspond to the correct model.
Do not throw away used batteries in the dustbin. Send to a
recycling centre.
TRANSMITTER
Display

Programming by menu.
Concentration of 6 samples.
Analyser status.
Alarms.
Trend curves.
Concentration historic
Calibration parameters historic
(date, slope and offset).

MAINS POWER SUPPLY


Mains

100- 240 VAC 50 - 60 Hz.


automatic switching.
Max. consumption: 80 VA.

Page 9

SILKOSTAT 9210

Dimensions of the analyser


Figure 3

Dimensions of the analyser Panel version

A : Mains power supply cable gland, drilled 11, 5 to 10mm, 100240V, 50-60Hz, 25VA
B : Field bus cable gland, drilled 13, 6 to 12mm
C : 5 to 7 mm cable gland
* Alarm signals : 30V, 0.5A max.
* Logic inputs : Inhibition on channels 1 to 6, acknowledgement of alarm signals
* Analogue outputs : 0-20 mA or 4-20 mA
D : Samples inlets, (channels 1 to 6), QR coupling for semi-rigid tubing, 6mm
5 to 50 C (40 to 120 F), P 0.2 to 6bar (3 to 87 PSI)
E : Drain tube : silicon tubing, 12 x 17mm, at atmospheric pressure

Page 10

SILKOSTAT 9210

Figure 4

Dimensions of the analyser Cabinet version

A : Cable gland, 5 to 7mm


* Mains power supply cable gland, drilled 11, 5 to 10mm, 100240V, 50-60Hz, 25VA
* Field bus cable gland, drilled13, 6 to 12mm
* Alarm signals : 30V, 0.5A max.
* Logic inputs : Inhibition on channels 1 to 6, acknowledgement of alarm signals
* Analogue outputs : 0-20mA or 4-20mA
B : Samples inlets (channels 1 to 6), QR coupling for semi-rigid tubing, 6mm
5 to 50C (40 to 120F), P 0.2 to 6 bars (3 to 87 PSI)
C : Drain tube : silicon tubing, 12 x 17mm, at atmospheric pressure.

Page 11

SILKOSTAT 9210

3.

Installation

Unpacking

The instrument should be unpacked with care. Be careful not to mislay


the accessories during unpacking.
Caution :
Assembly should not be carried out by non-qualified personnel.
The power supply should not be connected until the installation
has been completed and checked.

Inspection
The instrument has been tested and checked at the factory before
shipping, however it is advisable to carry out a visual inspection to be
sure that it has not suffered any damage. Damaged packaging could
mean the presence of damage not immediately visible. Keep the
packaging in case of complaints. For any missing parts or accessories,
please contact your distributor.

Pre-installation
Before installing the analyser, think about the following precautions:
Place the analyser close to a sample point. The sample should be
homogenous and representative. This will allow the response time to
be reduced. The temperature of the sample should be between 5 and
50C. The pressure of the sample should be between 0.2 and 6 bar
and remain relatively stable. The solution should be free of particles.
The sample lines should be in PE/PTFE/FEP ( 4 x 6 mm).
Avoid any location with a corrosive atmosphere or subject to liquid
spills.
Chose a dry and dust-free location.
The ambient temperature of the analyser should not exceed 45C. If
the temperature is below 5C, the analyser should be installed in a
heated cabinet.

Page 13

SILKOSTAT 9210

Presentation of the analyser


Figure 5

Analyser front panel

12345-

Figure 6

Monec transmitter box


Sample level detector
Overflow vessel
Sample 1 to 6 flow adjustment valves
Photometric measurement cell

Analyser rear panel

6 - Local controller box


7 - Measurement solenoid valve
9 - Measurement box
10 - Sample inlet valves
11 - Drain

Page 14

SILKOSTAT 9210

Positioning the canisters


A canister holder is supplied with the analyser, it is fixed as shown in the
figure below.

Figure 7

Canister replacing

In all cases, respect the maximum height between the measurement cell
and the canister.

Page 15

SILKOSTAT 9210

Connecting the sample


Use new pipes for connecting the installation

exterior :
Material :
Pressure :
Temperature :

6 mm imperatively (or '')


polyethylene or PTFE or FEP
0,2 to 6 bars
5 to 50C

At this stage of the installation, make sure that the flow valves (2) are all
closed.
The connection of pipes is done by insertion into the quick release
connections found under the sampling block.

Figure 8

Direct sample connections

123-

Solenoid valve
Solenoid valve flow adjustment screw
Sample input 1 to 6

Connecting the evacuation (drain) tube


The connection of the pipe is located on the bottom of the analyser and is
carried out by the user with the help of a 12 x 17 mm pipe delivered with
the instrument. It should be at atmospheric pressure and not be looped.
Caution :
Evacuation should be free of any mechanical constraints, and free
of any counter-pressure.

Page 16

SILKOSTAT 9210

Mains power connection


Figure 9

Layout of electronic boards in the datalyt box

4
1

1234-

Mains power connection terminals


Real time clock backup battery
RS 485 (or equivalent) communications module
Internal communications and low voltage power supply connector (for
additional electronics modules).
The electrical installation should be carried out by duly qualified
personnel. A supply voltage of 100-240 VAC is acceptable without
changing the configuration. The power supply terminals can be removed
from their housing to make connection easier.
For safety reasons, it is imperative to respect the working procedure
below :
Use a three-wire power lead (line + neutral + earth) sized for
supplying the required power.
The instrument should be connected to the mains via a circuit-breaker
or fuse whose value should be less or equal to 20 A. It should be
located in proximity and be identified.
This connection should cut-off the phase and the neutral when
electrical problems occur or when the user wishes to intervene inside
the instrument. On the other hand, the earth conductor should always
be connected.

Danger :
No intervention should be carried out on the instrument without
first switching off the power.

Page 17

SILKOSTAT 9210

The connection of mains is found in the transmitter box.


To gain access :
Unscrew the 4 screws from the front panel.
Rotate the front panel.
Remove the shielding plate.
Run the cable through the transmitter gland.
Remove the terminals and screw up the wires.

Figure 10

Position of mains connection and the fuse

earth
N
L

Fuse
T2 AL 250 V
according to CEI 127

Page 18

neutral
line

In
mains

SILKOSTAT 9210

Figure 11

Shieldied plate of the 9210

RS485 connections (optional extra)


Unscrew the 4 fixing screws from the front panel and turn it round.
Run the cable through the transmitter gland.
connect RX/TX+ and RX/TX- (connector is on the optional board
figure 3.5 item 3).

Page 19

SILKOSTAT 9210

User signals connections


This box is found on the back of the SILKOSTAT on the top (see
figure 6).
Unscrew the 4 screws of the cover to gain access to the terminals.
Note :
The cable glands mounted on the local controller box should be
perforated with a screwdriver before being used.

To ensure a good seal, the external diameter of cables should be


between 5 and 7 mm.

Rel 12

Rel 11
Rel 10
Rel 9
Rel 8
Rel 7
Rel 6
Rel 5
Rel 4
Rel 3
Rel 2
Rel 1

Local controller connections

Iout 1
Iout 2
Iout 3
Iout 4
Iout 5
Iout 6
Iout 7
Iout 8

Vout 12
Vout 11
Vout 10
Vout 9
Vout 8
Vout 7
Vout 6
Vout 5
Vout 4
Vout 3
Vout 2
Vout 1

Figure 12

+
+

+
+
+

+
+

GND
TDI
TMS
TCK
TDO
Vcc

!
Page 20

Heater
Mixer

In 1
In 2
In 3
In 4
In 5
In 6
In 7
In 8
In 9
In 10
In 11
In 12
Pt 100

FIX2

Danger :
The relay outputs can only supply power in very low safety voltage
(30 VAC Or 42.4 Vdc maximum) and limited to 0,5 A.

SILKOSTAT 9210

Max.
voltage

Max.
current

Mixer control

24 Vac

1A

Heater control

24 Vdc

1A

Cell temperature measurement

5 Vdc

1 mA

30 Veff ou
42.4 Vdc

0,5 A
resistive
charge

5 Vdc

5 mA

24 Vdc

23 mA

30 Vdc

1A

Connections

Re1 to Re6
Re7 to Re10
Re11
Re12
In1 to In6

Function

In7 to In10
In11
In12
Iout1 to Iout6
Iout7
Iout8

User relay of threshold 1 to 6


Not used in this application
Warning alarm
System alarm (NO and NF switch are available)
Inhibition input (channels 1 to 6)
For input In1, use terminals in and "- .
Terminals + and out are not used.
Input not used in this application
Remote acknoledgment of alarms
Level sample detector
Analog output 0-20 or 4-20 mA (channels 1 to 6)
Analyser status
Not used in this application

Vout1
Vout2 to Vout7
Vout8
Vout9 to Vout11
Vout12

Sampling electrovalve control


Sample selection electrovalves control (channels 1 to 6)
Calibration pump
Reagent pump P1 to P3
Drain pump

Note :
The system alarm relay (12) is in the positive safety position for
informing in case of a cut-off or an analyser error.

Page 21

SILKOSTAT 9210

4.

Starting-up

Reagent preparation
Danger :
Read the precaution and security appendix and wear protective
goggles and clothes before use chemical products.

Reagent concentrations (by type)


Reagent 1 - Molybdate reagent
Sodium dihydrate molybdate

50 g/L

Concentrated sulphuric acid

45 mL/L

Reagent 2 - Oxalic reagent


Oxalic acid dihydrate

40 g/L

Reagent 3 - Reducing reagent


Concentrated sulphuric acid

12.5 mL/L

Ammonium ferrous (II) sulphate hexahydrate

20 g/L

Reagent preparation (makes 2 litres)


Reagent 1 - Molybdate reagent
Sodium dihydrate molybdate, Na2MoO4.2H2O, >99%, neat (analytical quality)

100 g

(FLUKA# 71756, MERCK# 106521 or equivalent)


Concentrated sulphuric acid, H2SO4, 95-97%, neat (analytical quality)

90 mL

(FLUKA# 84721, MERCK# 100731 or equivalent)


1) Fill a 2-litre graduated flask to 2/3rds capacity with demineralised water (with the lowest
possible silica content). Add the sodium dihydrate molybdate whilst stirring the solution.
2) Continue to stir until the molybdate is completely dissolved (<5minutes).
3) Still stirring, slowly add the concentrated sulphuric acid in 3 or 4 parts (measured out from
a test tube).
4) Keep stirring until completely dissolved (<5minutes).
5) Make up the volume to the gauge line (2 litres).
6) Leave to cool before transferring to the 2-litre container provided for this reagent.

Page 23

SILKOSTAT 9210

Reagent 2 - Oxalic reagent


Oxalic acid dihydrate, C2H2O4.2H2O, >99.5%, neat (analytical quality),

80 g

(FLUKA# 75700, MERCK# 100495 or equivalent)


1) Fill a 2-litre graduated flask to 2/3rds capacity with demineralised water (with the lowest
possible silica content). Add the oxalic acid whilst stirring the solution.
2) Keep stirring until completely dissolved (<5minutes).
3) Make up the volume to the gauge line (2 litres).
4) Transfer to the 2-litre container provided for this reagent.
Reagent 3 - Reducing reagent
Concentrated sulphuric acid, H2SO4, 95-97%, neat (analytical quality)

25 mL

(FLUKA# 84721, MERCK# 100731 or equivalent)


Mohr's Salt = Ferrous (II) ammonium sulphate hexahydrate,
(NH4)2Fe(SO4)2,6H2O, >99%, neat (analytical quality)

40 g

(FLUKA# 09720, MERCK# 103792 or equivalent)


1) Fill a 2-litre graduated flask to 2/3rds capacity with demineralised water (with the lowest
possible silica content). Add the concentrated sulphuric acid whilst stirring the solution
(measured out from a test tube).
2) Still stirring, add the ferrous ammonium sulphate hexahydrate.
3) Keep stirring until completely dissolved (<5minutes).
4) Make up the volume to the gauge line (2 litres).
5) Leave to cool before transferring to the 2-litre container provided for this reagent.

Calibration solution
CONCENTRATED SOLUTION
Use of Titrisol (Merck) cartridge
Dillute one cartridge of Titrisol Merck 109947 solution into one litre of demineralised water
will give a concentration of 1000 mg/litre of Si, ie, 2139 mg/litre of SiO2.
Dillute this solution, 100 times, ie, 10 ml in one litre. This will give a concentration of
10 mg/litre of Si, ie, 21.39 mg/litre SiO2.

Page 24

SILKOSTAT 9210

CALIBRATION SOLUTION
By diluting an appropriate volume of the concentrated solution (at 21.39 mg/litres) with
demineralised water, various calibration solutions with a known concentration of SiO2 may
be prepared :
Suggestion for calibration solution preparation (for 2 litres)
ml of concentrated
solution 21.39 mg/l

calibration
solution
ppb Si

calibration
solution
ppb SiO2

Instrument
Measuring range

10

50

107

0-200 ppb

20

100

213.9

0-200 ppb

25

125

267.4

0-500 ppb

50

250

534.8

0-1000 ppb

Positioning the canisters


Reagent tubings are tagged as 1 to 3. They come with an analyser and
are fed through the special stoppers on the reagent canisters. One end of
these pipes is fitted with a chemically inert, PVC strainer that is used to
position the pipes in the bottom of the canisters.
Before starting analyser, it is necessary to connect each cap on its
canister :
Tubing N 1 on the reagent canister R1: Sodium molybdate and
sulphuric acid.
Tubing N 2 on the reagent canister R2: Oxalic acid reagent.
Tubing N 3 on the reagent canister R3: Sulphuric acid and ferrous
ammonium sulphate.
Screw the cap on the canisters. Then, check that the strainer is resting in
its proper position at the bottom of the canister. Put the canister on the
canister-stand to have cap towards interior as shown figure 7

Page 25

SILKOSTAT 9210

In the panel version, the canister-stand must be under the analyser


panel. In the cabinet version, it is in the back part of the cabinet.
Note :
The canisters must not be placed more than 40 cm from the
lower part of the down of the panel to have a good reagent
injection in the cell.

Page 26

SILKOSTAT 9210

Using the menu


c
Sample 1

Name of last measured sample (here : sample 1)

Current time (XX:XX).

Concentration of last measured sample.

Unit concentration.

10

Sample is being analysed : SAMPLE 5.


Analysis status : sampling, conditioning, measure
Bargraph indicates the state of the cycle.

Display of the last calibration, date and hour.


Default values for dates and hours are 09/09/09 00/00/00.

Function key
ESC :
cancels a data input or goes back to the
previous screen
Enter :
validate and go to the next step

14:00

10.00 ppb

Last calibration

Sample 5

P:date

Sampling

O:date

j
DISP2

START

MENU

Esc

function keys

Access to Programming mode.

Go to display 2.

10

Start / stop the analyser.

Modification of a value
The highlighted digit may be modified with the key
. Each digit can
be modified by pressing ENTER. Repeat both operations for each digit.
On the first digit : possibility to display the sign "-"
On the others digits : possibility to display the sign "."

Note :
If you do not use the keyboard for at least 10 minutes, the
instrument returns to the measuring mode except for the
calibration and maintenance mode.
An access code may be required for the calibration, programming
and service menu (see CODE menu).

Page 27

SILKOSTAT 9210

Language programming
SERVICE
DISPLAY
TIME
AFFICHAGE
CODE
SOFT ISSUE
DEFAUT VALUES
ADJUST mA

Access to the menu :


Press the "MENU" key.
Choose "SERVICE" menu with the "select" function key and
validate by "Enter".
Choose "DISPLAY" menu with the "select" function key and
validate by pressing "Enter".

CONFIGURATION
FACTORY
select

Esc

DISPLAY
LANGUAGE : F
CONCENTRATION : ppb/ppm
CUSTOMIZE

Choice of the language, measures units and samples


customize.
Use the key
to change the selected digit and press the
"Enter", key to validate the modification and select the next digit.
Languages available are :

Select

Esc

French

English

GB

German

Spanish

Sp

Italian

Unit choice is ppb or g / l.

Time and date programming


TIME
DAY : Monday
DATE (JJ MM AA) : 09/05/01
TIME : 14:43:16

Select

Access to the menu :


Choose the sub menu "TIME" in the menu "SERVICE" and
validate by pressing "Enter".
Use the key
to change the selected digit and press the
"Enter" function key to validate the modification and select
the next digit.

Esc

Note :
See menu "customize" for the customization of the name of
the measuring channels.

Page 28

SILKOSTAT 9210

Starting up the analyser

MAINTENANCE

Access to the menu :


Stop the analyser if it is analysing with the help of
START/STOP function key.

START UP
DEMARRAGE
REAGENTS REFILL
EXTENDED STOP

Press the "MENU" key.

TUBES CLEANING
PHOTOMETER TEST

Choose "MAINTENANCE" menu with the "select" function


key and validate by pressing "Enter".

PUMPS-OTHERS
SOLENOID VALVES
RELAYS

Choose "START UP" sub menu, press "Enter".

LOGICAL INPUTS

Select

This menu allows two choices :

Esc

Adjustment of the samples flow

START UP
Adjust flow
Sample x
OK

This adjustment is used to set up the flow-rates for the various


channels. The sampling solenoid valve is open during this
stage and agitation is continuous. The flow rate may be set by
adjusting the screws on the front panel so as to fill the overflow
vessel and guarantee a slight overflow.
The "Enter" key allows to access to the next sample.

Esc

x : n of the sample (1 to 6).

Caution :
Dont completelly unscrew the flow adjust screws. There is no
stop screw, so it will provoke a sample leak
Pumps priming :

START UP

Imerge the different reagent tubings in the canisters.

Pumps priming

Press OK when pipes of the different pumps are full of reagent.


Go back to "maintenance" menu.

Verify immersion
OK

Esc

Page 29

SILKOSTAT 9210

Starting a measurement

Sample 1

14:00

10.00 ppb
CAL

Sample 5

P : 02/02/02

Sampling

14:30:00
Z : 03/02/02
14:30:00

DISP2

Esc

Page 30

START

MENU

With the help of the "Esc" function key, go back to the main
screen and press "START" key.

SILKOSTAT 9210

5.

Analyser programming

Analyser configuration flow chart


MENU
CALIBRATION

MAINTENANCE

CALIBRATION
PROGRAMMING
EXECUTION PRIMARY
EXECUTION MANUAL
PARAMETERS
HISTORIC

EXECUTION MANUAL
EXECUTION ZERO
EXECUTION SLOPE
EXECUTION ZERO+SLOPE

PROGRAMMING
OFFSET INTERVAL : xxx h
AUTOMAT SLOPE CAL. : YES / NO
CAL. SOL : xxx,x ppb
AUTOMAT CAL: NO

MAINTENANCE
START UP
REAGENTS REFILL
TUBES CLEANING
EXTENDED STOP
PUMPS-OTHERS
SOLENOID VALVES

REAGENTS REFILL
ADJUST VOLUMES
TUBES PRIMING
BOTTLES FULL : yes/no

ADJUST VOLUMES
REAGENT 1 VOL : x.xx l
REAGENT 2 VOL: x.xx l
REAGENT 3 VOL: x.xx l
REAGENT 4 VOL: x.xx l
PUMPS
PUMP 01 : off/on

SOLENOID VALVES
VALVE 01 : off/on
VALVE 07 : off/on

PUMP 04 : off/on
Drain : on/off
Mixer : on/off
Heater : on/off

RELAYS

RELAYS
RELAY 01 : off/on

PHOTOMETER TEST
LOGICAL INPUTS
MEASURE

MEASURE
MEASURING MODE
GRAB SAMPLE

PROGRAMMING
MODE : Normal
CYCLE : IN LOOP
HISTORIC
PROGRAMMING
READ

HISTORIC
ADC VALUES
ALARMS

ALARMS

WARNING ALARM
SYSTEM ALARM
ALARM : yes/no

SYSTEM ALARM
mA OUTPUTS

WARNING ALARM
ALARM : yes/no
ACCEPT : manual
RELAYS : NO/NC

mA OUTPUTS
AFFECT

AFFECT

OUTPUT X

OUTPUT 1
OUTPUT 2
OUTPUT 3
OUTPUT 4
OUTPUT 5
OUTPUT 6

AFFECT : Channel x
TYPE: 0-20
MODE : Linear
LOW : xxx.x ppb
UP : xxx. ppb

SPECIAL PROG.
MAINTENANCE
CALIBRATION
GRAB SAMPLE
SYSTEM ALARM
WARNING ALARM

SPECIAL PROG.

CALIBRATION
MODE : Last

TEST

TEST

SEQUENCE

PROGRAMMING
FROM : xx/xx/xx
AT : xx:xx:xx
CH : A11
ALARMS : yes

ALARM X
MOD : limit
AFFECT : Channel 1
LIMIT : xxx.x ppb
DIRECTION : Down
HYSTERESIS : xx%
RELAYS : NO/NC

ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARM 5
ALARM 6

COMMUNICATION

RELAY 12 : off/on

VALUE : xx mA

COMMUNICATION
N : xx
BAUD : 300/600/1200/2400/4800/9600
PARITY : no/even/odd
BIT STOP : 1/2

SERVICE

SERVICE
TIME
DISPLAY
CODE
SOFT ISSUES
DEFAULT VALUES
ADJUST mA
FACTORY

SEQUENCE
CHANNEL ACTIVATION
SEQUENCE
TIME
DAY : xxxxx
DATE : JJ/MM/AA
TIME : xx/xx/xx
CODE
PROGRAMMATION: xxx
CALIBRATION : xxxx
SERVICE : xxxx
CONFIGURATION

CHANNEL ACTIVATION
CHANNEL 1 : On / Off
...
CHANNEL 6 : On / Off

DISPLAY
LANGUAGE : F/GB/D/Sp/I
CONCENTRATION : ppb-ppm / ug-mg/l
CUSTOMIZE
mA ADJUST
OUTPUT 1
OUTPUT 8

...

Page 31

SILKOSTAT 9210

User Interface
The interface consists of 5 keys and a graphic backlet display screen
providing a display of the data, operational settings and trend curves.
Informations are available from several screens.

Main display

Sample 1

Last calibration

Sample 5

P : 02/02/02

Sampling

Current time (XX:XX).

Concentration of last measured sample.

Unit concentration.

Sample is being analysed : SAMPLE 5.


Analysis status : sampling, conditioning, measure
Bargraph indicates the state of the cycle.

Last slope and offset calibrations, date & time.

Function key
ESC :
cancels a data input or goes back to the
previous screen
Enter :
validate and go to the next step

14:30:00

10

Z : 03/02/02

14:30:00

14:00

10.00 ppb

Name of last measured sample (here : sample 1)

DISP2

START

MENU

Esc

Programming mode.

Go to display 2.

10

Page 32

function keys

Start / stop the analyser.

SILKOSTAT 9210

Definition of the display screen 2 (Disp2)

d
Last measures with next indications:

e
Sample 1 : XXX XX:XX:XX
.
.
.
.

.
.
.
.

.
.
.
.

.
.
.
.

Sample 6 : XXX XX:XX:XX


Extern :

DISP3

Sample name.

The last concentration value of the sample.

Time when last measurement was performed

Function key Disp3 (alarms) allows to access to the


display 3.

Esc

Page 33

SILKOSTAT 9210

Definition of the display screen 3 (Disp3) : alarms


This screen displays the alarms status as well as the following
indications :
ALARM

S1S6 :

Threshold function
parameters,

S1 : OK
.
.
.
.

.
.
S6 : OK

Esc

following

OK :

Alarm condition is not met.

Sample X > XXX :

Alarm is actived and the alarm condition is


clearly mentioned. For example, the
sample X measurement exceeds xxx ppb.

Disable:

Alarm function is disactivated.

: OK

the

W : disable

DISP4

with

Warning alarm.

System alarm.

DISP4 :

Go to the display 4.

The following table indicates the different messages which may appear in
case of default.
Displayed message
(english)

System
System
System
System
System
System
System
Warning
Warning
Warning
Warning

Cal.time over

Warning

Calib < 300 ml


No reagent
Slope drift
Zero drift
Zero error
Slope error
Sample x < C
Sample x < 0 ppb
Sample x > 1000
5000 ppb
Reference error

Reset of the measure module


Reset of the local controller module.
Local controller module error
Measure module error
CAN communication error
Error during table transmission
Task execution error
Measurement error
LED light ajustment problem
Black current problem
Maximum cycle time for the measure
has been exceeded
Maximum cycle time for calibration has
been exceeded
No sample available
Sample x missing
Low level of the reagent left (2.5 days
before no reagent alarm)
Low level of the calibration solution (1
calibration left before no calibration
solution alarm)
Empty reagent canisters <150 ml
Slight drift of slope calibration
parameter (Confirmation zone)
Slight drift of zero offset calibration
parameter (Confirmation zone)
Zero calibration major error
Slope calibration major error
Sample concentration x is lower than
C
Sample concentration x is negative
ppb ou > Sample concentration over range

Category

Reset mod. Measure


Reset module LC
Error lc
Err. measure mod.
Com error
Table error
Task error
Measure error
Reference error
Extinction error
Meas.time over

No sample
Sample x missing
Reagent < 250 ml

Page 34

Description

The reference measurement lies


outside its limiting values

System
Warning
Warning
Warning
System
Warning
Warning
System
System
Threshold
Warning
Warning
Warning

SILKOSTAT 9210

Definition of the display 4 (Disp 4) : trend curve


e

This screen allows to display the trend curves.


Sample 4 (X)

30/08/01

5,00

Constant time scale which corresponds to 24 hours.

concentration scale
programmed scale.

Sample name.

Number of the associated output.

Date of the trend curve.

equals

to

analogue

output

0,00

c0

24

Main

Graph+

Esc

Function key :
Main :

Return to the main screen.

Graph+ :

Visualisation of the 6 curves associated with the


6 analogue outputs.

Modification of a value
The highlighted digit may be modified with the key
. Each digit can
be modified by pressing ENTER. Repeat both operations for each digit.
On the first digit : possibility to display the sign "-"
On the others digits : possibility to display the sign "."

Note :
If you do not use the keyboard for at least 10 minutes, the
instrument returns to the measuring mode except for the
calibration and maintenance mode.
An access code may be required for the calibration, programming
and service menu (see CODE menu).

Page 35

SILKOSTAT 9210

Main menu

Sample 1

14:00

10.00 ppb

You have 2 possibilities to access to the programming menu :


Access to the menus which are not protected by a code :
Simply press "MENU".

CAL

Sample 5

P = 02/02/02

Sampling

14:30:00
Z = 03/02/02
14:30:00

DISP2

STOP

MENU

Esc

Sample 1

14:00

10.00 ppb

Access to the menus which are protected by a code :


Press "STOP" key.
Confirm by "YES" key.

CAL

Sample 5

P = 02/02/02

Sampling

Press "MENU"

14:30:00
Z = 03/02/02
14:30:00

ARE YOU SURE ?


YES / NO

Esc

Sample 1

14:00

10.00 ppb
CAL

ANALYSER

P = 02/02/02

STOPPED

14:30:00
Z = 03/02/02
14:30:00

DISP2

Esc

Page 36

STOP

MENU

Once the programming operation is completed and if the


analyser is stopped, press "START" to restart it.

SILKOSTAT 9210

MENU
CALIBRATION
MAINTENANCE
MEASURE
ALARMS
mA OUTPUTS
COMMUNICATION

"MENU" allows to access to the programming of the instrument.


This screen allows to choose between different programming
sub-menus.
Press the "Select" function key to select the programming menu
you wish to have access to and press "Enter" to validate the
choice.

SEQUENCE
SERVICE
Select

Esc

Calibration menu

CALIBRATION
PROGRAMMING
EXECUTION PRIMARY CAL
EXECUTION MANUAL CAL
PARAMETERS

Esc

To get a good precision from the instrument, you should calibrate


it on a regular basis.
Calibrations allow to adjust :
ZERO of the system.

HISTORIC

SLOPE of the system.

Select

Both ZERO and SLOPE of the system.


The
ZERO
calibration
SILKOSTAT/chemically.

is

performed

by

the

In order to avoid the use of water free of SiO2, the SILKOSTAT


carries out a measurement without a colorimetric reaction.
The slope of the system is calibrated with a known concentration
of SiO2 standard solution.
Three types of calibration are available :
Primary calibration.
Manual execution.
Automatic execution.

Page 37

SILKOSTAT 9210

Execution primary sub menu


CALIBRATION
PROGRAMMING
EXECUTION PRIMARY CAL
EXECUTION MANUAL CAL
PARAMETERS
HISTORIC

Select

A primary calibration is carried out when operating the instrument


for the first time and is used as a reference for the following
calibrations. Generally, it should be carried out only after a
consequent modification.
It allows to fix the zero and slope reference.
All the results of the further manual or automatic calibrations will
be compared to the zero and slope references set by this
calibration.
A more important shift will allow to detect an error calibration
which will be signified by an error message.

Esc

This calibration can be realized only in the manual way.


Access to the main "MENU".
Choose "CALIBRATION" menu with the "select" function key
and validate by pressing "Enter".
Choose "EXECUTION PRIMARY" sub menu and validate by
pressing "Enter".
The slope calibration is going on, followed by the zero.
PRIMARY CALIBRATION

The upper bar indicates the type of calibration : primary, manual


or automatic.

Slope calibration
Cal sol sampling

Press "Esc" if you want to quit the calibration menu : and go back
to the "CALIBRATION" menu.

CALIBRATION

The calibration stages are displayed as follow :


Esc

Sample preparation.
Conditioning
Measure.
Rinse.

A bargraph indicates the calibration state.


At the end of the calibration, results are displayed on screen.

Page 38

SILKOSTAT 9210

PRIMARY CALIBRATION

Default values :
S = 760.0
O = 0.0
Primary calibration

Primary calibration results:


S = value of the slope (ppb / Abs)
O = value of the zero shift in milli-Abs
The default values show a general idea of standard values
according to the range (0-1000 or 0-5000 ppb).

S = XXX
O = XXX

Esc

Execution manual sub menu


CALIBRATION
PROGRAMMING
EXECUTION PRIMARY CAL
EXECUTION MANUAL CAL
PARAMETERS
HISTORIC

Select

Choice of the calibration.


Access to the menu :
Press "MENU" key.
Choose "CALIBRATION" menu with the "select" function key
and validate by pressing "Enter".
Choose "EXECUTION MANUAL CAL" with
function key and validate by pressing "Enter".

the

"select"

Esc

CALIBRATION
EXECUTION ZERO
EXECUTION SLOPE
EXECUTION ZERO + SLOPE

This screen allows a manual calibration of the SILKOSTAT 9210.


Press "select" key.
Choose "EXECUTION ZERO" or "SLOPE or "ZERO + SLOPE"
sub-menu and validate by pressing Enter" key.

Select

Esc

Page 39

SILKOSTAT 9210

Programming sub menu (automatic execution)


CALIBRATION
PROGRAMMING
EXECUTION PRIMARY
EXECUTION MANUAL
PARAMETERS
HISTORIC
Select

The calibration of the zero and the slope can be carried out
automatically at adjustable frequencies.
Press "CALIBRATION" menu.
Sub-menu "PROGRAMMING" and validate by pressing
"Enter".
Choose "PROGRAMMING" sub-menu and validate by pressing
Enter.

Esc

PROGRAMMING
OFFSET INTERVAL : 200 h
CAL. SOL.: 010.0 ppb
AUTO. CAL. : Non

Select

Esc

Page 40

This screen allows to program the zero calibration frequency, the


calibration concentration solution and to choose an automatic
calibration or manual calibration. The following table indicates the
zero calibration frequency controlled in terms of the minimum
concentration of the samples.
Sample minimum concentration

Zero frequency control

< 5 ppb

24 hours

5 < x < 50 ppb

48 hours

> 50 ppb

> 72 hours

SILKOSTAT 9210

PROGRAMMING
OFFSET INTERVAL : 200 h
CAL.: 010.0 ppb
AUTO SLOPE CAL. : Yes
BASE : Day
WEEK : SMTWTFS
HOUR : 15:00:00

Select

In case of an automatic slope calibration, this screen allows to


adjust the frequency.
When the "BASE" menu is selected, the "WEEK" menu allows
to choose the calibration day (s).
Ex :

SMTWTFS, daily calibration.


SM*W*FS, calibration every day except Tuesday and
Thursday.

"HOUR" menu allows to adjust the hour to start calibration.

Esc

OR
PROGRAMMING
OFFSET INTERVAL : 200 h
CAL. : 010.0 ppb
AUTO SLOPE CAL.. : Yes

When the hour base is chosen , the "INTERVAL" menu allows to


adjust the periodicity of the calibration slope.
The required slope calibration frequency is about 1 week
(168 hours).

BASE : Hour
INTERVAL : XXXX

Select

Esc

Page 41

SILKOSTAT 9210

Calibration parameters sub menu


MANUAL / AUTOMATIC
CALIBRATION

This screen allows to visualize the primary calibration parameters


and the last calibration parameters.

Default values :
S = 760.0
O = 0.00
Primary calibration :
S = 694.0

01/01/01

O = 0.00

01/01/01

Last calibration :
S = 694.0

01/01/01

O = 0.00

01/01/01

Esc

Calibration historic sub menu


HISTORIC
31/08/00

auto

09:18:01

O = -5

19:28:51

O = -5

02/09/00

manu

10 :01 :05

O = -4
S = 560

12/09/00

manu

15 :01 :05

o = -6
S = 561

Select

Esc

Page 42

This screen allows to visualize the calibrations historic with the


following information :
Date and hour.
Type : manual or automatic.
Slope (S) and offset (O).

SILKOSTAT 9210

Manual automatic calibration results sub menu


At the end of a manual or automatic calibration, the analyser compares
the results with the primary calibration master.
According to the calibration results, you can validate, restart or quit the
calibration.
There are 3 zones :
20 %

Confirmation
zone

Rejection zone
O : (mAbs)

- 50

- 30

150 %

180 %
Confirmation
zone

Zone ok
30

Rejection zone
50

Zone ok

XXX

Primary calibration :
S = XXX

XX:XX:XX

O = XXX

XX:XX:XX

Last calibration
S = XXX

50 %

Primary
calibration master

S : (%)

XX:XX:XX

O = XXX

XX:XX:XX

C = XXX

XX:XX:XX

Actual calibration

Normal variations around the reference values. The new


parameters are considered.
In automatic mode, the analyser returns to its measurement cycle.
In manual mode, validate with OK and restart the instrument
(START).
P:XXXXX DD:MM:YY slope
Z:XXXXX DD:MM:YY offset

S = XXX

C:XXXXX DD:MM: concentration of the sample solution just


measured, calculated using the calibration values from the "Last
calibration".

O = XXX
Calibration OK
OK

Esc

Page 43

SILKOSTAT 9210

Confirmation zone

XXX

Primary calibration :
S = XXX

XX:XX:XX

O = XXX

XX:XX:XX

Last calibration
S = XXX

XX:XX:XX

O = XXX

XX:XX:XX

C = XXX

XX:XX:XX

Abnormal gap compare to the reference. So it allows to have


correct results and a warning is displayed. In automatic mode, the
analyser returns to the measure mode with the previous
parameters.
In manual calibration, you must confirm (Conf) or cancel (Annul)
the new parameters.

Actual calibration
S = XXX
O = XXX
Calibration error
Conf

Cancel

Esc

XXX

Primary calibration :
S = XXX

XX:XX:XX

O = XXX

XX:XX:XX

Last calibration
S = XXX

XX:XX:XX

O = XXX

XX:XX:XX

C = XXX

XX:XX:XX

Rejection zone
Parameters are abnormal. There is material problem or an
important error like bad calibration solution preparation. (see
troubleshooting).
In automatic mode, the analyser stops and emits an alarm system.
In manual mode, you can cancel (turn to the measure with
previous parameters) or restart a calibration.

Actual calibration

Note :

S = XXX
O = XXX
Calibration error
Cancel Remake

Esc

Page 44

The reference measurement lies outside its limiting


values.
In auto mode, where the calibration has not been made
correctly, the analyser takes one measurement and then
recalibrates. After 3 successive faulty calibrations, the
9210 waits for the next programmed date.

SILKOSTAT 9210

Maintenance menu
MAINTENANCE
START UP

This screen allows to choose the proposed maintenance help


procedure :
Start up

REAGENTS REFILL

Starting procedure.

EXTENDED STOP
TUBES CLEANING
PHOTOMETER TEST

Reagents refill

PUMPS / OTHERS

The procedure is applicable to each reagent refill.

SOLENOID VALVES
RELAYS

Extended stop

LOGICAL INPUTS

Allows to condition the instrument before an extended stop.

Select

Tubes cleaning
In case of reagent tubings clogging.

Esc

Photometer test : Check the photometer.


Pumps, Solenoid valves, relays, others
Working test of this devices.
Logical inputs
Logical inputs test.

Note :
When one of the maintenance procedure is released, the analyser
automatically stops
Dont forget to restart
operations are realized.

the

instrument

when

maintenance

Start up sub menu


This screen allows to adjust the sample flows

START UP

Adjust the first sample flow with the help of the screws placed in
the front panel of the face (see figure 3.4 item. 2) to have a
perpetual overflowing in the top of the overflow vessel.

Adjust flow
Sample X
OK

Press OK to go to the next sample.

Esc

Warning :
Do not unscrew completely the flow rate set screw. It has no
stop and may get out of its housing and provoke a sample leak.

Page 45

SILKOSTAT 9210

Pump priming :

START UP

Press OK when the pump tubing are full of reagent. The display
goes back to the menu Maintenance.

Pump priming

OK

Esc

Reagents refill sub menu


REAGENTS REFILL
ADJUST VOLUMES
TUBING DRAINING
CANISTERS FULL yes/no

Select

Every reagent and calibration solution renewal, access this menu


for.
This menu allows to choose between :
ADJUST VOLUMES : the 4 canisters volumes are displayed and
can be modified.
TUBES PRIMING : start the priming / renewal procedure of the
reagents in the tubes. "Esc" allows to return to the previous
screen or to stop the procedure.

Esc

CANISTERS FULL : choose this option each time than canisters


are full. The volumes of the 4 canisters are automatically
programmed with 2 litres.
Adjust volumes
ADJUST VOLUMES
REAGENT VOL 1 : 2.00
REAGENT VOL 2 : 2.00
REAGENT VOL 3 : 2.00
REAGENT VOL: 2.00

This menu allows to visualize the volumes of the reagent canisters


and the calibration solution. You eventually can adjust these
volumes. The FULL CANISTERS menu put these volumes to
2 litres.
The analyser controls the calibration solution and reagent
consumption and emits alarm system according to the level of the
canisters.

Select

Esc

Page 46

Warning

Alarm system

Reagents

< 250ml

< 150ml

Cal solution

< 300ml

SILKOSTAT 9210

Tubes priming
REAGENTS REFILL

The tubes priming is going on ; the reagent pumps and the


calibration solution are running at the same time than the mixing
motor and the drain.

Please wait

Esc

REAGENTS REFILL

At the end of the tubes priming you can quit the procedure by
pressing "ESC".
After the pump operation, check visually the tubings are full. If it is
not the case, return to the menu "pump".

ESC = quit

Esc

Note :
It is recommended to calibrate the zero after each reagent
refill.

Extended stop sub menu


EXTENDED STOP

Put tubes in
demineralised water
Accept with OK

This screen shows the procedure to follow in case of a long stop


(1 or more than a week).
Provide 250 ml of demineralised water.
Press "OK" to start each stage of the procedure.

OK

Esc

Follow the instructions and validate each stage with "OK".


At the end of the stop, restart the analyser with the help of the
MAINTENANCE / START UP menu.

Page 47

SILKOSTAT 9210

Tubes cleaning sub menu


TUBES CLEANING

Put tube R1 in NH4-OH


solution
Accept with OK

OK

Esc

Tubes cleaning procedure in case of depositing.


Have ready a beaker containing 250ml of demineralised water and
a second beaker containing 100ml of ammonia at 10% of normal
commercial concentrations by volume, ie, approximately 3% v/v.
Immerse tubing R1 in the ammoniac solution.
Press "OK" for the cleaning and press "Esc" to stop and come
back to the previous screen.
Follow the instructions and validate by "OK".

TUBES CLEANING
Put pumps tubes
In demineralised water
Accept with OK..

OK

Esc

Page 48

SILKOSTAT 9210

Measure / Photometer test sub menu


PHOTOMETER TEST
Dark current : 0.008

The mixer is operating and the transfer valve is open.

Reference : 3.50
Led current : 9000
Measure : 3.10
Absorbance : 52.7

Bck

ref

With this menu, technicians can verify the good running order of
the photometer.

meas

Displayed parameters
Dark current :
Measurement value which corresponds to the
emitter LED.
Reference :
reference signal after LED current adjustment.
Led current :
current of the LED.
Measure :
Continuous photometric measure signal.
Absorbance :
Absorbance in milli-Abs.

Esc

Function key
Bck :
allows to begin the measure which corresponds to the
switching off of the LED.
Ref. :
allows to adjust the led current and to measure the
reference voltage.
Mes :

starts the continuous photometric measure.

Pumps-others sub menu


MAINTENANCE
Pump 01

> OFF

Pump 02

OFF

Pump 03

OFF

Pump 04

OFF

Drain

OFF

Mixer

OFF

Heater

OFF

Select

Esc

On

Off

This screen allows to test the good running order of the pumps,
the mixer, the drain and the heater.
Utilize the "Select" key to control an element and press the "On"
key to operate it or "Off" key to stop it.
The following table shows the nominal flow rates for each pump :

Pumps 01
Pumps 02
Pumps 03
Pumps 04

Reagent 1
Reagent 2
Reagent 3
Cal. solution

1,5ml/mn
1,5ml/mn
1,5ml/mn
10l/h

Page 49

SILKOSTAT 9210

Solenoid valves sub menu


MAINTENANCE
Solenoid valve 01 > OFF
Solenoid valve 02

OFF

Solenoid valve 03

OFF

Solenoid valve 04

OFF

Solenoid valve 05

OFF

Solenoid valve 06

OFF

Solenoid valve 07

OFF

Select

Off

On

This screen allows to test the correct operating of the solenoid


valves.
Utilize the "Select" key to select a valve and press the "On" key to
operate it or "Off" key to stop it.
Press "Esc" to come back to the previous screen.
Solenoid valves 01 to 06 : sample solenoid valves
( measuring channel 01 to 06).
Solenoid valves 07 : measure solenoid valves (sample
transfer from overflow tank to measuring cell.).

Esc

Relays sub menu


MAINTENANCE
Relay 01

> OFF

Relay 02

OFF

Relay 03

OFF

Relay 04

OFF

Relay 05

OFF

Relay 06

OFF

Relay 07

OFF

Relay 08

OFF

Relay 09

OFF

Relay 10

OFF

Relay 11

OFF

Relay 12
Select

This screen allows to operate or to stop the relays.


Utilize the "Select" key to select a relay and press the "On" key to
operate it or "Off" key to stop it.

Note :
See the relays functions on page 21.

OFF
On

Off

Esc

Logical inputs sub menu


LOGICAL INPUTS
I/O

0 0 0 0 1 1
2 4 6 8 0 2
------------------------Status * 1 * 1 * * * * * * *

ESC : exit

This screen displays the status of the logical inputs and allows to
test the good running order of the inputs In1 to In12.
Status : 1 : short circuited input
* : open input
Example : in the above example, measurements on channels
3 and 6 are inhibited.
The analyser goes directly to the next channel

Esc

Page 50

SILKOSTAT 9210

Measure Menu
MEASURE
MEASURING MODE
GRAB SAMPLE
HISTORIC
ADC values

This screen allows to accede to the measure parameters ; the


mode and the cycle of measure, the historic, the external sample
measure (manual analysis) and the photometer parameters (ADC
values).
The measure historic allows to visualize the data during 5 days.

Select

Esc

Measuring mode sub menu


PROGRAMMING
MODE : Normal
CYCLE : In loop
INTERVAL : XXX mn

Select

This screen allows to program the measuring mode.


If the interval cycle is selected, the analysis are made every 10
minutes.
INTERVAL : XXX mn, you must adjust the measuring cycle time.
Example : if 030 mn is displayed, the instrument does a measure
every 30 minutes.

Esc

OR
PROGRAMMING

If the In loop cycle is chosen , measures are made one after


one without waiting time.

MODE : Normal
CYCLE : In loop

Select

Esc

Page 51

SILKOSTAT 9210

Grab sample sub menu


GRAB SAMPLE

Allows to carry out the measure with a grab sample.


Before measuring, take 200 ml of the sample to analyse.

Please wait

The analyser begins to rinse the measuring cell.

Esc

GRAB SAMPLE

The analyser complains the sample to analyse :


Remove the level probe of the overflow vessel (figure 3.1,
item 2), refill with the sample in 3 or 4 times.

Fill sample
About 200 ml

Replace the level probe.


OK

Press "Ok" to operate the measure.


To return to the programming, wait the end of the measure
(display of the results and terminate message).

Esc

"Esc" key allows to stop the measure and comes back to the
previous screen.
Historic sub menu
HISTORIC
PROGRAMMING
READ

Select

Esc

Page 52

This screen allows to read the stored data (concentration and


alarm values).

SILKOSTAT 9210

Programming
PROGRAMMING
FROM :

01/01/01

AT :

00/00/00

CH :

all

ALARMS :

yes

Allows to choose the date and the time when the historic will be
displayed and the measure channels desired.
CH : all or individually (1 to 6)
ALARMS (yes or no) : display of the system and
warning alarms .

Select

Esc

Read
HISTORIC

"READ" allows to read the historic.


Use the

18/05/01
13:30 # 1

12.6

13:40 # 2

103

scroll up key to choose the data format.

Date : day/month/year.
Hour, sample n (#n) and concentration or default.

Select

Esc

ADC values sub menu


ADC VALUES
PRIMARY CALIBRATION
ref :

meas :

S : XXXX

XXXX

O : XXXX

XXXX

This screen displays the parameters of the photometer for the


primary calibration, the last calibration and the last measurement.
The information stored in this menu must be utilized only at the end
of the diagnostic in case of incorrect operation.

LAST CALIBRATION
Ref:

meas :

S : XXXX

XXXX

O : XXXX

XXXX

LAST MEASUREMENT
ref :
XXXX

meas :
XXXX

Esc

Page 53

SILKOSTAT 9210

Alarms Menu
This screen allows to assign alarms and parameters to each relay.

ALARMS
ALARM 1
ALARM 2
ALARM 3
ALARM 4
ALARM 5
ALARM 6
WARNING ALARM
SYSTEM ALARM

Select

Esc

Alarms parameters 1 to 6 sub menu

MODE

Value

Description

Limit
Active
channel
Level
sample.

Allows to choose the type of relay : threshold (limit) , active channel


or level sample.

AFFEC

Channel 1 to 6

Allows to assign the alarm (1 to 6) to the measure channel of our


choice.

LIMIT

XXXX

Allows to adjust the limit value (ppb)

DIRECTION

Allows to adjust the relay direction interlocking

HYSTERESIS

XX %

Allows to adjust the hysteresis in %

RELAYS

N.O
N.C

Relay type

Up
Down

N.O : normally open


N.C : normally closed

Page 54

SILKOSTAT 9210

ALARM 1
MODE : Limit
AFFECT : Channel 2
LIMIT : 100.0 ppb
DIRECTION : Up
HYSTERESIS : 00 %
RELAY : N.O

If MODE = Limit (concentration threshold)


Example :
Affects the alarm contact 1 to channel 2
Adjust of the threshold value 100.0 ppb.
Relay activating direction towards up (when C > 100 ppb).
No hysteresis : 00 %.

Select

Relay type N.O. (normally open)


Esc

OR
ALARM X
MODE : active channel
RELAY : NO

If MODE = Active channel


The alarm NX is activated when the sample NX is analysing.
Choice N.O (normally open) or N.C (normally closed).

Select

Esc

OR
ALARM X
MODE : level sample
RELAY : NO

If MODE = Level sample


The alarm NX is activated when the sample NX is absent.
Choice N.O (normally open) or N.C (normally closed).

Select

Esc

Page 55

SILKOSTAT 9210

Warning alarm sub menu

WARNING ALARM

Allows to adjust the WARNING alarm parameters.


Value

ALARM : yes
ACCEPT : Manual

Description

ALARM

Yes
No

Warning alarm in operating or not.

ACCEPT

Manual

Press enter to accept the alarm


(manual).

Auto

Automatic accept of the alarm


(disappear when the alarm
disappear).

RELAY : N.O

Select

Esc

RELAY

Type of relay
N.O

N.C

System alarm sub menu


ALARM 1

" Normally excited (contact open


between 1 and 2 contact closed
between 3 and 4)

Allows to adjust the system alarm

ALARM : yes

Value

RELAY : N.O

ALARM
Select

"Normally not excited (contact


closed between 1 and 2 contact
open between 3 and 4).

Yes
No

RELAY

Activates / Deactivates the


warning
Type of relay

N.O

Esc

Description

N.C

"Normally not excited (contact


closed between 1 and 2 contact
open between 3 and 4).
" Normally excited (contact open
between 1 and 2 contact closed
between 3 and 4)

NB:
The alarm system can only be acknowledged manually, either
by holding down the <Enter> key or, remotely, by using the
digital input, In 11, on the I/O card (see Section 3, figure 3.12).

Page 56

SILKOSTAT 9210

The following table shows the different alarms :


Warning : the analyser emits a warning at the display and on the warning
relay.
System alarm : the analyser locks the system alarm relay, displays a
message ant stops.

Displayed message
(english)
Reset mod. Measure
Reset module LC

Description

Reset of the measure module


Reset of the local controller
module.
Error of the local controller
Error lc
module
Err. measure mod.
Error of the measure module
Com error
Error CAN communication
Error during transmission the
Table error
table
Error task
Error of the task execution
Measure error
Error of the measure
Reference error
Curent del adaptation problem
Extinction error
Extinction del problem
Cycle time of the measure is
Meas.time over
very long
Cycle time of the calibration is
Cal.time over
very long
No sample
No sample here
Sample default x
Short sample x
Reagent < 250 ml
Low level of the reagent
Low level of the calibration
Cal.sol < 300 ml
solution
Low reagent
Empty reagent canisters <150
Slope drift
Confirmation zone
Zero drift
Confirmation zone
Rejection zone for the zero
Zero default
calibration
Offset zero
Rejection zone for the slope
Sample concentration x is < to
Sample x < C
threshold C
Sample concentration x is
Sample x < 0 ppb
negative
Sample x > 1000 ppb ou Sample concentration too high
> 5000 ppb

Reference error.

Incorrect reference
measurement

Category
System
System
System
System
System
System
System
Warning
Warning
Warning
Warning
Warning
System
Warning
Warning
Warning
System
Warning
Warning
System
System
Warning
Warning
Warning
Warning

Page 57

SILKOSTAT 9210

mA outputs menu
mA OUTPUTS

This screen allows to access to the analogue outputs parameters.

AFFECT
SPECIAL PROG
TEST

select

Esc

AFFECT :

Allows to assign the output (1 to 6) to the


channel desired.

SPECIAL PROG:

Allows to program 0/4-20mA outputs in


special cases.

TEST :

Allows to test the correct operating of the


analogue outputs.

Outputs affect 1 to 6 sub menu


AFFECT
OUTPUT 1
OUTPUT 2
OUTPUT 3
OUTPUT 4
OUTPUT 5
OUTPUT 6

Select

Esc

Page 58

This screen allows to select an output and to display its


parameters.

SILKOSTAT 9210

OUTPUT 1 to 6
AFFECT : X
TYPE

: 0-20

MODE

: Linear

LOW

: 000.0 ppb

UP

: 01.00 ppm

Select

Esc

This screen displays the parameters of each 6 output.


AFFECT :

Choice of the measure channel. The


concentration of this channel will be carried
over the output X.

TYPE :

0-20 mA or 4-20 mA, choice of the analogue


output scale.

MODE :

Linear or dual (see curve)

FIRST :

First scale value.

MIDDLE :

Middle scale value.

END :

End scale value.

OR
OUTPUT 1 to 6
AFFECT :

Channel 1

If the dual mode is chosen, you must program the concentration


value corresponding to the middle of the scale.

With output 1
TYPE

: 0-20

MODE

: Linear

FIRST

: 000.0 ppb

MIDDLE : 000.0 ppb


END

: 000.0 ppm

Select

Esc

The following curve indicates the result of the 4-20 current in terms of the
scale programming type and the first, middle and end value of the scale.
Current
0..20mA/4..20m
Linear
20 / 20 mA

Dual

12 / 10 mA
4 / 0 mA

FIRST

MIDDLE

END

Concentration

Page 59

SILKOSTAT 9210

Special prog sub menu

SPECIAL PROG
MAINTENANCE
CALIBRATION
GRAB SAMPLE
SYSTEM ALARM
WARNING ALARM

This screen allows to program the status of the output n7 during


special events :

Maintenance.
Calibration.
Grab sample analysis.
System alarm.
Warning alarm.

With the help of the CHOICE function, choose the event for which
the output n7 can be programmed.

Select

Esc

Storage :

The output n7 is not modified.

MAINTENANCE
MODE : Last

Select

Esc

OR
SPECIAL PROG

Preset : The analogue output n 7 is set to the programmed


value in the VALUE parameter between 0 and 21 mA.

MODE : Preset
VALUE : XXX mA

Note :

Select

In case of system alarm, all the analogue outputs (1 to 7)


are fixed to the preset value.

Esc

According to the programmed choice selected in the menu


mA outputs / special prog / system alarm.

Test sub menu


TEST
VALUE : 00 mA

Allows to test all the loops 0.20 and 4.20.


scroll up key allows to adjust the current value
The
generated by each output between 0 and 21 mA.
(See paragraph Service/mA adjustment).

Esc

Page 60

SILKOSTAT 9210

Communication menu
COMMUNICATION
N : 00
BAUD : 9600
PARITY : No
STOP BIT : 1

This screen allows to adjust the communication parameters like


RS 485 of the SILKOSTAT. The communication protocol is
MODBUS-JBUS.
scroll up key to change the selected digit and
Utilize the
press the Enter key to validate and go to the next digit.

SWAP WORD : No

Select

Esc

N :

Monec number (0 33).

Baud :

Transmission speed in bauds (300, 600 1


200, 2 400, 4 800, 9 600).

Parity :

odd, even, no.

Stop bit:

1 or 2

Swap word

Allows to reverse the send order of the


data.

Choice of the measure sequence menu


SEQUENCE

The "SEQUENCE" menu allows to run or stop the different


measurement channels and to set up the analysis samples order.

CHANNEL ACTIVATION
SEQUENCE

Select

Esc

Channel activation sub menu


CHANNEL ACTIVATION

The CHANNEL ACTIVATION allows to run or to stop the


measurement channel.

CHANNEL 1 : On

Select : allows to choose the channel.

CHANNEL 2 : Off
CHANNEL 3 : On

allows to run / stop a channel

CHANNEL 4 : On
CHANNEL 5 : Off

Enter : allows to validate a choice.

CHANNEL 6 : On

Select

Esc

Page 61

SILKOSTAT 9210

Sequence sub menu


SEQUENCE

SEQUENCE : 123562141131

The "SEQUENCE" menu displays the configured sequence and


allows to change it.
The sequence is set up on 12 steps.
scroll up key to change the number of the channel
Utilize the
as the channel number as possible.

Delete

Esc

The "Delete" key replaces the digit by the symbol "*". This symbol
indicates the inhibition of the step in the sequence.
Press "Enter" to validate and go to the next digit.
Sequences examples:
1)

123456******
456321******
These two sequences allow to analyse samples (1 to 6) in
different order.

2)

12 12 12 12 12 12
or 12 ** ** ** ** **
or 1* 2* ** ** ** **
These 3 sequences are the very same and allow to
analyse alternately the samples 1 and 2.

Page 62

SILKOSTAT 9210

Service menu
SERVICE

This screen gives access to the parameters of the 9210


SILKOSTAT hardware.

TIME
DISPLAY
CODE
SOFT ISSUES
DAFAULT VALUES
mA ADJUST
FACTORY

Select

Esc

Time sub menu


TIME

This screen allows to change the date and the time of the
instrument.

DAY : Monday
DATE (JJ MM AA) : 09/05/01
TIME: 14:43:16

Select

Esc

Display sub menu


DISPLAY
LANGUAGE : F
CONCENTRATION : ppb/ppm
CUSTOMIZE

Select

Esc

This screen allows to change the programming language, the


concentration unit and to customize the samples displays (you can
name the samples).
Languages : choice of the message language
French :

English :

GB

German :

Spanish :

Sp

Italian :

The unit choice is ppb / ppm or g / mg/l

Page 63

SILKOSTAT 9210

Customize
Allows to choose the letter

CUSTOMIZE

"Enter" allows to validate this letter and go to the next character.

Ch 1 : anion1
Ch 2 : anion2
Ch 3 : mixed bed
Ch 4 : ball
Ch 5 : condens

Note

Ch 6 : vapour

Customizing a sample name is possible with the lower-case


letter a...z, with the upper-case letters A...Z, numbers
0...9 and punctuation signs.

Select

Esc

Service / Code sub menu


CODE
PROGRAMMATION : 0000
CALIBRATION : 0000
SERVICE

: 0000

Use this screen to enter the analyser's calibration or service


menu programming access codes. These codes are
independent of one another.
Enter 0000 if there is no access code.

Select

Esc

Soft issues sub menu


SOFT ISSUES
SILKOSTAT

XX

I/O Module

XX

Module Colo.

XX

Esc

Page 64

This screen displays the different software issues installed on


the different electronic boards of the 9210 SILKOSTAT.

SILKOSTAT 9210

Default / values sub menu


DEFAULT VALUES

This screen allows to load default values. (See appendix C).

Loading

Note :
Default values ?

YES

Esc

This operation will erase all the parameters (calibration,


420 mA scale, alarms, sequence,), and the default
values will be loaded and will erase the measurement values
and the historic alarms.

Adjust mA sub menu


Select the output with the "Select " key and validate by the "Enter"
key to display the next screen.

ADJUST mA
OUTPUT 1
OUTPUT 2

Each 420 mA output can be individually calibrated on 2 points.

OUTPUT 3
OUTPUT 4

Connect a mill ammeter to the 1 to 8 outputs.

OUTPUT 5
OUTPUT 6
OUTPUT 7
OUTPUT 8

Select

Esc

Utilize the "-" and "+" key to adjust the current value to 4 and 20
mA, validate (press Enter ) and press Esc to come back to the
previous screen.

ADJUST mA
4 mA
VALUE : 0000

Esc

OR
ADJUST mA
20 mA
VALUE : 0000

Esc

Page 65

SILKOSTAT 9210

Factory sub menu


FACTORY
CODE : 0000

Note :
Personal constructor menu. An access code is required
to enter . Switch off the instrument to have access.

Esc

FACTORY
CHANNEL NUMBER : X

This menu enables to adjust the channel number between 1 and


6, the operating and waiting time of the sample valves, pumps and
measurement parameters.

SOLENOID VALVES
PUMPS
MEASURE
CONFIGURATION

Esc

Solenoid valves
SOLENOID VALVES
Ton SAMPLE

: XXX

Toff SAMPLE

: XXXs

SOL. CAL. Ton

: XXX

SOL. CAL. : Toff

: XXXs

Esc

Page 66

Adjustment of the opening time of the sample solenoid valves and


calibration solution pump (Ton) and waiting after action (Toff).

SILKOSTAT 9210

Pumps
Ton : number of impulses.

PUMPS
PUMP 1 Ton

: XXX

PUMP 1 Toff

: XXXs

PUMP 2 Ton

: XXX

PUMP 2 Toff

: XXXs

PUMP 3 Ton

: XXX

PUMP 3 Toff

: XXXs

Toff : waiting time.

Esc

Measure

MEASURE

RANGE

Select the measuring range between 0-1000


ppb or 0-5000 ppb

LINEAR COEFF.

Correctness coefficient of the non-linearity


(the sub-menu is displayed only if the range
0-5000 ppb is chosen).

WAITING TIME

Waiting time before measure

MEASURE TIME

Measure time

ZERO CONC

reserved

FILTER

reserved

NEGATIVE VAL.

Choice to display the negative values

HEATER

Heating on/heating off

RANGE : XXX
LINEAR COEFF. : XXX
WAITING TIME

: XXXs

MEASURE TIME

: XXXs

ZERO CONC : XXX


NEGATIVE VAL. : XXX
FILTER : XXX
HEATER : XXX

Esc

Configuration sub menu

CONFIGURATION
MAINS FREG

: 50 Hz

This screen allows to change the frequencies


MAINS FREG.:

50 Hz or 60 Hz (it depends on the


country frequency).

REFERENCE FREQ :

Current frequency adjustment of the


LED-level reading.

BLACK FREQ :

Current frequency measurement of the


black-level reading.

REFERENCE FREQ : 0020


BLACK FREQ

: 0020

Select

Esc

Page 67

SILKOSTAT 9210

6.

Maintenance and exploitation


A periodical maintenance will allow to have a good running and an exact
analysis. Verify the level of the calibration solution and the reagents and
refill if it is necessary.

Reagent refills
The analyser monitors the reagents and calibration solution consumption.
Messages are displayed when the canister levels are low.
Messages

Signification

Relays on

Reagent < 250ml

Low reagent level


(range < 10 days)

Warning (Re 11)

Cal. Sol < 300ml

Calibration low level


(range < 2 calibration)

Warning (Re 11)

Short reagent

Reagent canister
volume < 150ml

System (Re 12)

Short cal.

Calibration solution
canister volume
< 150ml

It is necessary to program the new canister volume at each refill for a


good detection of the short reagent.
Stop the instrument "STOP".
Refill the reagent canisters.
Press the "MENU" key.
Choose the "MAINTENANCE" menu.
Choose the "REAGENTS REFILL" menu.
REAGENTS REFILL

ADJUST VOLUMES : the 4 canisters volumes are displayed and


can be modified.

ADJUST VOLUMES
TUBES PRIMING
CANISTERS FULL yes/no

Select

Esc

TUBES PRIMING : start the priming / renewal procedure of the


reagents in the tubes. "Esc" allows to return to the previous
screen or to stop the procedure.
CANISTERS FULL: choose this option each time than canisters
are full. The volumes of the 4 canisters are automatically
programmed with 2 litres.

Choose CANISTERS FULL to program all the canisters


volumes to 2 litres.

Page 69

SILKOSTAT 9210

ADJUST VOLUMES
REAGENT 1 VOL : 2.00
REAGENT 2 VOL : 2.00
REAGENT 3 VOL : 2.00
SOL CAL

VOL : 2.00

Or ADJUST VOLUMES to visualize or program another


value than 2 litres.
This menu allows to visualize the volumes of the reagent
canisters and the calibration solution. You eventually can
adjust these volumes. The FULL CANISTERS menu put
these volumes at 2 litres.
The analyser controls the calibration solution and reagent
consumption and emits system alarm as the level of the
canisters is going too low.

Select

Esc

REAGENTS REFILL

Warning

Alarm system

Reagents

< 250 ml

< 150 ml

Calibration
solution

< 300 ml

Choose REAGENTS REFILL to start the reagents refill


procedure.

Please wait

The tubes priming is going on ; the reagents pumps and the


calibration solution are running at the same time than the
mixing motor and the drain.
Esc

REAGENTS REFILL

At the end of the tubes priming you can quit the procedure by
pressing "ESC".

ESC = quit

Esc

Note :
It is recommended to calibrate zero after each reagent
refill.

Page 70

SILKOSTAT 9210

Prolonged stopping
In case of an extended stop longer than 1 week, it is recommended
to clean the tubes and the measuring cell.
Provide 250 ml of demineralised water.
Submerge the 3 tubes in the beaker and start the rinsing operation:
Stop the instrument by pressing "STOP".
Press the "MENU" key.
EXTENDED STOP

Put tubes in
demineralised water
Accept with OK

Choose the "MAINTENANCE" menu.


Choose the sub menu "EXTENDED STOP".
The measuring cell is empty.
Disconnect the instrument.
Replace the reagents tubes in their canisters.

OK

Esc

Note : Dont reverse the tubes.

In case of an extended stop over 6 weeks, you must throw the


reagents and prepare another set for the restart.

Page 71

SILKOSTAT 9210

Restarting after a prolonged stopping


Connect the analyser to the main.
Start the instrument by pressing "START".
Press the "MENU" key.
Choose the "MAINTENANCE" menu.
Choose the sub menu "START UP".

MAINTENANCE / START UP
This screen allows to adjust the samples flow.

START UP

Set the flow-rate for the first sample using the adjusting screws on
the front panel (see Figure 3.3, serial 4) in such a way that a
constant overflow takes place over the top of the overflow vessel.

Adjust flow
Sample 1 (to 6)
OK

Warning :
Esc

Do not unscrew completely the flow rate set screw. It has


no stop and may get out of its housing and provoke a
sample leak.

This screen enables to prime the pumps.

START UP

After the pump operation, check visually the tubings are full. If it is
not the case, return to the menu pump ".

PIPE START
Verify immersion
OK

Esc

Page 72

SILKOSTAT 9210

Tube cleaning
TUBES CLEANING
Put pumps tubes
in demineralised water

Cleaning procedure of the reagent tubings in case of deposit.


Use a beaker with 250 ml of demineralised water and a beaker
with 100 ml of 10% ammoniac, about 3%. Immerse the tubing R1
in an ammoniac solution.
Press OK to launch the cleaning and esc to stop the
procedure and to come back to the preceding menu.

Accept with OK

OK

Follow the instructions and press OK at each step


The procedure is the following :

Esc

TUBES CLEANING
Put pumps tubes
In demineralised water
Accept with OK..

Rinse the tube with demineralised water.

Clean with a 10% ammonia solution.

Rinse again the tube with demineralised water.

During these stages :


In special cases you must clean the reagents tubes. Follow the
procedure to clean the tubes with demineralised water and a 10%
ammonia solution (NH4-OH) and rinse with demineralised water.
During the cleaning the sample runs into the cell with the mixer
running.

OK

Esc

Page 73

SILKOSTAT 9210

7.

Troubleshooting

Possible sources of the problem


The list gives you an idea to inspect different parts of the 9210
SILKOSTAT when it doesnt normally run.

Problem

Possible cause

Problem
N

The system doesnt run

Start / Stop Datalyt

The measure is not stable, too high or too low.

Cal
Tubes
Photometer
Valves
Pump

5
3
8
4
2

Wrong calibration

Pump
Tubes
Valves
Cal solution
Calibration program
Reagents
Photometer

2
3
4
6
7
7
8

The photometer doesnt run or doesnt


correctly measure.

Photometer

Page 75

SILKOSTAT 9210 - MANUEL D'UTILISATION

Possible solutions
Problem

Symptom

Solution

1. START / STOP

Analyser stopped

2. PUMP

System alarm : Rectify the problem according to the


displayed message (see System Alarm Messages).
Start up the analyser (START key).

Mains power cut


off

Connect the analyser to the mains power.


Verify the fuse.

Doesnt run

Test the pump : in the MAINTENANCE / PUMP menu


start pumps one after one and verify than their flow is
about 1,5ml / mn (30 pulses).
Change the pump.

Irregular

3. TUBES

4. SOLENOID
VALVES

5. CAL.

Page 76

The analyser may have been stopped without tubes


cleaning. Send water in the pump during its work.
Change the pump.

Defective tube

Change the tube

Clogged tube

Change the tube

Solenoid valves
leak

Remove, test and clean.

The solenoid
valves cant open

Sticked collar, must be clean

The solenoid
valves cant close

Locked spring ; necessary cleaning

Dirty

Disconnect the samples from the sample block.

Diminution of the
flow

Verify and adjust the flow (10 l/h), change the sample tube if it
is clogged.

SILKOSTAT 9210

Problem

Symptom

Solution

6. CALIBRATION

Wrong configuration of the calibration


solution concentration

Reprogram
the
concentration (SiO2)

Wrong concentration of the standard


solution

Prepare a new standard solution.

Wrong laboratory analysis of the


calibration solution concentration

Verify the SiO2 concentration


laboratory calibration solution.

Wrong chemicals products or olds or


contaminated

Wrong purity ? New supplier ? Clean the


reagents bottles.

Reagents canisters connected in a


wrong order.

Verify
the
flow
of
the
circuit
"MAINTENANCE/PUMPS-OTHERS" menu.

Reagents tubes connected in a wrong


order.

Verify the flow diagram

Wrong concentration

Prepare a new reagents batch

Wrong chemical products or old or


contaminated

Wrong purity ? New supplier ?

Wrong reagents proportional


Bacterial developpement in reagent R3
(greenish deposit)
White deposit in the reagent tubing R1

Verify the good running order of the pumps.


Add 40 ml of sodium benzoate at 1 g/l

7. REAGENTS

8. PHOTOMETER

Wrong or irregular measure

calibration

solution

of

the

Immerse the tubing in an ammoniac solution


according to the procedure described in the
menu MAINTENANCE/TUBING CLEANING

Verify the photometer "MAINTENANCE /


PHOTOMETER TEST " . In this menu
measurement values of reference and
measurement during the last calibration
of the measurement cycle appears. The
typical values of these values are in the
following table.
Is there an error message ?

Page 77

SILKOSTAT 9210 - MANUEL D'UTILISATION

Typical measurement values during calibration


These values are given as an indication and are dependant on the ambient temperature. The real values
may deviate from more than one volt without having an effect on the measurements.

Page 78

Sample concentration
107 ppb
214,9 ppb
267,4 ppb
534,8 ppb

Version 01000 ppb


Reference (mV)
3500
3500
3500
3500

Measure (mV)
2515
1754
1472
603

Sample concentration
1000 ppb
3000 ppb

Version 05000 ppb


Reference (mV)
3500
3500

Measure (mV)
1400
300

SILKOSTAT 9210

Appendix A : Spare parts of the 9210 SILKOSTAT


DESCRIPTION

Part Number

Accessories / Options :
0,5 6 bar adaptation kit for six sample selection solenoid
valves

09210=A=0100

9210 Instruction manual in English

621=192=010

9210 Instruction manual in French

621=092=010

9210 Instruction manual in German

621=292=010

9210 Instruction manual in Italian

621=492=010

9210 Instruction manual in Spanish

621=592=010

Spare parts - Electronics


Users interface / Interface board

09200=A=5000

CPU board with display for Monec E9200, non programmed

09200=A=1001

CPU board with display for Monec E9200, programmed for


SILKOSTAT

09200=A=1010

CPU board with display for Monec E9200, programmed for


PHOSPHAMAT

09200=A=1011

Power supply board for Monec E9200

09200=A=2000

RS485 board (MODBUS) + Manual

09125=A=0485

92XX Internal bus communication module

09200=A=5500

2 wires interconnection cable (per m)

370=302=034

4 wires interconnection cable (per m)

370=304=034

7 wires interconnection cable (per m)

370=307=034

Analogue board 92XX 820 nm

09210=A=1500

Analogue board 92XX 525 nm

09210=A=1510

Analogue board 92XX 540 nm

09210=A=1520

Spare parts Sampling


4x6 mm PE tubing (per m)

151575,00006

Sampling solenoid valve, NPS 1.6 (0.2-2 bars)

689=118=016

Sampling solenoid valve, NPS 0.8 (0.5-6 bars)

689=118=008

O-ring, 4 x 1

356099,05040

Sampling flow-rate adjustment screw

09210=A=0105

Sample level sensor for 921 X

09210=A=0250

Page A1

SILKOSTAT 9210

DESCRIPTION

Part Number

Viton O-ring 38 x 2 mm

356099,35380

Sampling solenoid valves (overflow vessel)

689=118=024

Spare parts Photometric cell


Connection board for 921 X mixer/heater

09210=A=1234

Complete mixer for 921X analysers

09210=A=0400

Teflon coated magnetic stirrer

226=003=013

Heating element

09097=C=0560

Pt100 temperature sensor

359000,40780

Fibre optics for 9210 Silkostat 921X

09210=A=0500

Mirror for 9210 Silkostat 921X

09210=C=0340

PG21 tap PA black

425=210=022

Spacer ring for 921X mirror

09210=C=0330

O-ring Viton 22x1 mm

356099,05220

Square cover for 921 X photometric cell

09210=C=0310

Spare parts Reagents and calibration


Chemical kit for silkostat 9210

09210=C=7000

PE tubing 1.6x3.2 mm (per m)

590=050=060

PTFE tubing 0.8x1.6 mm (per m)

151065.08699

PE tubing 4x6 mm (per m)

151575.00006

Reagent pulse pump

695=114=001

Calibration / flush pump

695=004=004

2 litres PE cap

495=020=001

Strainer

09210=A=0600

Connection for PE tube, 4 x 6

589=011=040

Connection for PE tube, 1.6 x 3.2

589=010=015

Caution :

Page A2

The chemical products kit of the SILKOSTAT is compounded with all


the reagents except the sulphuric acid for 6 weeks.

SILKOSTAT 9210

Appendix B : Safety Precautions


HEALTH AND SAFETY INFORMATIONS
PRODUCT : OXALIC ACID
The health and safety information on oxalic acid is as follows :
Name :
OXALIC ACID

1. dihydrate
2. anhydrous

Formula : 1. (COOH)2. 2H2O


Formula : 2. (COOH)2.

CAS N144-62-7
PHYSICAL PROPERTIES
1. 101 C

MELTING POINT

2. 190 C

SOLUBILITY IN WATER

About 14%

FIRE AND EXPLOSION HAZARD

Decomposition products include carbon monoxide


(highly flammable and toxic) and carbon dioxide.

HEALTH HAZARD

Classification of risk Harmful


Harmful in contact with skin
Harmful if swallowed
Harmful if inhaled
Minimum letal dose (LD50) = 275 mg/Kg
SAFETY PRECAUTIONS

EYES

Irrigate thoroughly
attention

with water ;

obtain

medical

LUNGS

Remove exposure, rest and keep arm ; obtain


medical attention

INGESTION

Wash out thoroughly with water ; give plenty of water


of drink ; obtain medical attention
CHEMICAL HAZARDS AND STABILITY

STABILITY

Stable

SPECIAL HAZARDS

Mixtures with sodium chlorite can explode on the


addition of water.

WASTE DISPOSAL

Dissolve or mix the material with a combustible


solvent and burn in a chemical incinerator equipped
with an after burner and scrubber. Dispose in safe
manner consistent with Local Authority chemical
waste regulations. For guidelines on disposal consult
the requirements of the Control of Pollution Act
(1974) and the Control of Pollution (Special Wastes)
Regulations (1980 S1 1709).

Page B1

SILKOSTAT 9210

PROTECTIVE MEASURES
RESPIRATOR

Dust mask or positive pressure hood if handling


large quantities.

GLOVES

Rubber or plastic

EYE PROTECTION

Goggles
SPILLAGE HANDLING

SMALL SPILLAGE

Mop up with plenty of water and run to waste, diluting


greatly with running water.

LARGE SPILLAGE

Transfer to salvage canister and treat residue as for


small spillage.

The information on this sheet is believed to be correct at the time of collation. The details should be reviewed
periodically and should be checked for special uses.

Page B2

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : SULPHURIC ACID (98%)
The health and safety information for sulphuric acid is as follows :
Name :
SULPHURIC ACID

Formula : H2 SO4

Hazard class : 8

UN N : 1830

CAS N 7664-93-9
PHYSICAL PROPERTIES
MELTING POINT

10 C

BOILING POINT

330C

SPECIFIC GRAVITY

1,8

MISCIBLE WITH WATER


FIRE AND EXPLOSION HAZARD

May evolve toxic fumes in a fire.

HEALTH HAZARD

Classification of risk Corrosive.


Causes severe burns to eyes and skin.
If indigested causes severe internal irritation and
damage.
Dilute acid irritates the eyes and skin and can cause
burns and dermatitis.
Minimum lethal dose, D50 = 2140 mg/Kg, par voie
orale.
Occupational exposure standard, long term exposure
limit (8 hours TWA) mg-3 :
Keep out of reach of children.
SAFETY PRECAUTIONS-FIRST AID

EYES

Irrigate immediately and thoroughly with plenty of


water at least of 10 minutes. Obtain medical attention.

LUNGS

Remove from exposure, rest and keep warm, or if


exposure has been great, obtain medical attention.

MOUTH

Wash out mouth thoroughly with water ; give plenty of


water to drink ; obtain medical attention.

SKIN

Drench the skin thoroughly with water and give plenty


of water to drink. Obtain medical attention.

Page B3

SILKOSTAT 9210

CHEMICAL HAZARDS AND STABILITY


Reacts vigorously with evolution of heat. Never add
water to this product.
Will react violently with the following :

Acetonitrile.

Perchlorates.

Nitro-organic compounds.

Alkali metals.

Potassium permanganate

Metal halogenates.

Avoid contact with bases.


PROTECTIVE MEASURES
Wear chemical-resistant gloves (nitrile), safety
goggles or face shield and use fume cupboard. For
large quantities use approved breathing apparatus,
plastic apron and sleeves and wear boots.
Ensure that safety shower and eye bath are available
in the locality.
SPILLAGE HANDLING

Wear appropriate protective clothing.


Cover the spillage with soda ash. If local regulations
permit, cautiously add plenty of water and mop up
while diluting with more water and run to waste. Wash
site of spillage thoroughly with water. Otherwise cover
with dry lime, sand or soda ash and place in covered
canisters. Arrange removal by disposal company.
Ventilate area and wash spill site with water after
material removal is complete.
If material has entered drains it may be necessary to
inform the Local Authorities.
WASTE DISPOSAL
Dispose in safe manner consistent with Local
Authority chemical waste regulation. For guidelines on
disposal consult the requirements of the Control
Pollution Act (1974) and the Control of Pollution
(special wastes) Regulations (1980 SI 1709).

The information on this sheet is believed to be correct at the time of collation. The details should be reviewed
periodically and should be checked for special uses.

Page B4

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : SODIUM HEXAFLUOROSILICATE
The health and safety information on this product is as follows :
SODIUM HEXAFLUOROSILICATE
Hazard class : 6.1No. UN : 2674

(sodium fluorosilicate , disodium hexafluorosilicate disodique)


Formula : Na2SiF6

CAS N 16893-85-9

PHYSICAL PROPERTIES
White granular powder
Density (20 C) approx. 2,7 g cm-3
Slightly soluble in water (6g 1-1 to 20C)
FIRE AND EXPLOSION HAZARD

When heated to decomposition it may emit hydrogen


fluoride.

HEALTH HAZARD

Classification of risk toxic.


Avoid contact with skin.
Avoid contact with eyes.
Keep locked up.
Keep out of reach of children.
Toxic by ingestion and skin contact
Toxic by inhalation
Irritating to skin and eyes, may cause damage if
exposure is prolonged.
Occupational exposure standard,
exposure limit (8h TWA) 2,5 mg m-3

long

term

SAFETY PRECAUTIONS FIRST AID

If the casualty feels unwell, seek medical advice and


show label where possible.

EYES

Rinse immediately and thoroughly with water for at


least 10 minutes. Obtain medical attention.

LUNGS

Remove from exposure, rest and keep warm. In


severe cases, or if exposure has been great obtain
medical attention.

MOUTH

Wash out mouth thoroughly with water and give


plenty of water to drink. Obtain medical attention.

SKIN

Wash skin thoroughly with water and give plenty of


water to drink. Obtain medical attention.

CHEMICAL HAZARD AND STABILITY

Contact with acids liberates toxic gas.

Page B5

SILKOSTAT 9210

PROTECTIVE MEASURES (as appropriate to quantity handled)


RESPIRATOR

Dust respirator.

VENTILATION

Extraction hood.

GLOVES

Rubber or plastic.

EYE PROTECTION

Goggles or face shield.

OTHERS MEASURES

Plastic apron, sleeves, boots if handling large


quantities.
SPILLAGE HANDLING

PRECAUTIONS

Wear appropriate protective clothing.


Mix with sand, transfer carefully to canister and
arrange removal by disposal company. Wash site of
spillage thoroughly with water and detergent.
For large spillages liquids should be contained with
sand or earth and both liquids and solids transferred
to salvage canisters. Any residues should be treated
as for small spillages.
If material has entered surface drains it may be
necessary to inform Local Authorities, including fire
services flammable.

WASTE DISPOSAL

Dispose in safe manner consistent with Local


Authorities or national chemical Waste Regulation. For
guidelines for disposal in the UK consult the
requirements of the Control of Pollution (special
wastes) regulations (1980 SI 1709).

The information in this sheet is believed to be correct at the time of callation. The details should be reviewed
periodically and should be checked for special uses.

Page B6

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : FERROUS AMMONIUM SULPHATE HEXAHYDRATE
The table below shows the health and safety (MSDS) data for Ferrous ammonium Sulphate:
PRODUCT IDENTIFICATION
PRODUCT NAME

FERROUS AMMONIUM SULPHATE


HEXAHYDRATE

CAS No. (CE)

7783-85-9 233-151-8

FORMULA

(NH4)2Fe(SO4)2, 6H2O

MOLECULAR WEIGHT

392,14 AMU

ALTERNATIVE NAMES

Ferrous ammonium sulfate hexahydrate *


Ammonium ferrous sulfate, hexahydrate * Iron
ammonium sulfate hydrate * Sulphuric acid,
ammonium iron (2+) salt, hexahydrate
PHYSICAL PROPERTIES

FORM

Solid

COULOUR

Greenish blue

ODOUR

Nil

MELTING POINT

100C

DENSITY

1.86 g/cm
Non-combustible.
Likelihood of dangerous vapours being given off if
close to a fire.

FIRE AND EXPLOSION HAZARD

In fire situations it may form nitrogen oxides and


sulphur oxides.
HEALTH RISKS
SPECIFIC ENVIRONMENTAL AND HEALTH
HAZARDS

Irritant for the eyes, respiratory tract and skin.

SAFETY MEASURES
EYES SKIN

Flush copiously with water.


Remove any contaminated clothing

RESPIRATORY TRACT

Avoid continued exposure, breathe in fresh air; seek


medical attention.

IF SWALLOWED

Drink plenty of water, provoke vomiting, and seek


medical advice.
CHEMICAL RISKS AND STABILITY

STABILITY

Conditions to avoid - high temperatures

CONDITIONS TO AVOID

Sensitive to air, sensitive to light.

PRODUCTS TO AVOID

Powerful oxidising agents and acids.

Page B7

SILKOSTAT 9210

Iron oxides.
DANGEROUS DECOMPOSITION PRODUCTS

Ammonia.
Sulphur oxides.

PREVENTIVE MEASURES
Safe Handling Instructions: Do not breathe the dust.
Avoid contact with the eyes, skin or clothing. Avoid
prolonged or repeated exposures.

HANDLING

UK:
Source - category - value
OEL - LTEL - 1 mg (Fe)/m
OEL - STEL - 2 mg (Fe)/m

EXPOSURE LIMITS

Use dust-masks or over-pressure breathing gear if


dust formation is likely.

BREATHING EQUIPMENT

Essential.

GLOVES

Wear chemical resistant gloves (nitrile rubber, NBR).


Protective glasses/goggles are essential.

EYE PROTECTION

Wear protective clothing and work in a well-ventilated


area.
OTHER

Remove any contaminated clothing and wash hands


thoroughly after working with the product.
DISPOSAL MEASURES
Contact a specialist waste disposal organisation
where disposal of this product is required. Dissolve,
or mix, the product with a combustible solvent and
burn in a chemical incinerator fitted with postcombustion and scrubber systems. Observe all
applicable local, national and international
environmental regulations.
FIRST AID MEASURES

FOLLOWING INHALATION

In the event of inhalation, seek clean air. Where


breathing has stopped, apply artificial respiration.
Where breathing is difficult, give oxygen.

FOLLOWING CONTACT WITH SKIN

In the event of skin contact, wash the skin


immediately using soap and plenty of water.

FOLLOWING CONTACT WITH THE EYE

In the event of contact with the eyes, rinse out


immediately with plenty of water for at least
15 minutes.

FOLLOWING INGESTION (IF SWALLOWED)

If swallowed, rinse out the mouth with water if the


patient is still conscious. Call medical assistance.

Page B8

SILKOSTAT 9210

FIRE FIGHTING - EXTINCTION METHODS


SUTABLE EXTINGUISHERS

Water spray. Carbon dioxide, dry chemical powder or


suitable foam.

SPECIAL RISKS

Specific Danger(s): Gives off toxic fumes if burnt.

SPECIAL PROTECTIVE EQUIPMENT NEEDED


FOR FIRE FIGHTING

Wear individual breathing apparatus and protective


clothing to prevent any contact with the eyes or skin.

The information contained in this document is reviewed at regular intervals and checked for special cases.

Page B9

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : SODIUM MOLYBDATE DIHYDRATE
The health and safety information for Sodium Molybdate Dihydrate is as follows :
Name :
SODIUM MOLYBDATE DIHYDRATE

Formula: Na2MoO4.2H2O

CAS N 10102-40-6
PHYSICAL PROPERTIES
MELTING POINT

687C

SOLUBILITY IN WATER

840 g/L

FIRE AND EXPLOSION RISK

No information available

HEALTH RISKS

Not hazardous according to Directive 67/548/EEC


SECURITY MEASURES

EYES

Rinse immediately with plenty of water, keeping the


eyelids open as much as possible

LUNGS

In case of inhalation, remove from exposure and


move to fresh air immediately.

INGESTION

Drink plenty of water. If feeling unwell seek medical


aid
CHEMICAL DANGER AND STABILITY

STABILITY

Stable

SPECIFIC RISKS

No information available

DISPOSAL

Observe all federal, state, and local environmental


regulations for disposal of chemical products
PERSONAL PROTECTION

RESPIRITORY

Wear appropriate breathing apparatus due to the


formation of dust

GLOVES

Not necessary

EYES

Wear appropriate safety goggles


MEASURES TO TAKE IN CASE OF ACCIDENTAL SPILLAGE

PERSONAL

Avoid inhalation of any dust

ENVIRONMENT

Do not dispose of in any standard waste disposal


systems such as sinks or sewers

CLEANING

Vacuum or sweep up material and place into a


suitable disposal container. Avoid generating dusty
conditions.

The information listed above must be reviewed periodically and should not be considered a substitute for the
MSDS for this product.

Page B10

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : 9210 R1 REAGENT
PRODUCT AND COMPANY IDENTIFICATION
Product code : R1

SUBSTANCE

Name of product : 9210 R1 Reagent


Manufacturer Name : HACH SAS
Address : 29 rue de Bures
14670 Troarm

SUPPLIER

Phone
: (33) 02.31.15.37.10
Fax N
: (33) 02.31.15.37.30
Emergency : (33) 02.31.15.37.35 (Unit of production)
: (33) 01.45.42.59.59 (N ORFILA)
COMPOSITION / INFORMATION ABOUT THE COMPONENTS
Preparation

CHEMICAL PROPERTIES
HAZARDOUS COMPONENTS
CAS NUMBER
213-653-7
10102-40-6
7664-93-9

SOLUTION
Water
Sodium molybdate
Sulphuric acid

QUANTITY
90 %
5%
5%

CLASSIFICATION

C ; R35 ; S26 -30 - 45

IDENTIFICATION OF HAZARDS
Product not considered as dangerous according to the 67/548/CEE directive
FIRST AID
SKIN CONTACT

Rinse affected area with plenty of water. Remove


contaminated clothing. Swab with polyethylene glycol
400

EYE CONTACT

Immediately flush eyes with plenty of water with


eyelids open for 10 minutes. Get medical aid if
necessary

IF SWALLOWED IN LARGE QUANTITIES

Give plenty of water to drink. In case of indisposition,


take medical advice. Do not try to neutralize

FIRE FIGHTING MEASURES


EXTINGUISHING MEANS

Adapt the extinguishing agent to the products stored


in the immediate vicinity

SPECIFIC HAZARDS

Non combustible. Risk of forming hydrogen if in


contact with light metals

Page B11

SILKOSTAT 9210

MEASURES IN CASE OF SPILLAGE


PRECAUTIONS FOR PEOPLE

Avoid contact with skin and eyes

ENVIRONMENT PROTECTION

Do not pour in sewers or rivers

CLEANING

Collect in specific containers. Clean soiled materials


HANDLING AND STORAGE

PRECAUTIONS TO TAKE

Avoid contact with metals

USEAGE ADVICE

Respect conditions of use

STORAGE RECOMMENDATIONS

Store in hermetically sealed containers

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION


RESPIRATORY PROTECTION

Required if steam / aerosol

HAND PROTECTION

Wear latex gloves

EYE PROTECTION

Safety glasses are recommended

OTHER PROTECTION

Acid-proof clothing

HYGIENE MEASURES

Wash hands after contact


PHYSICAL AND CHEMICAL PROPERTIES

FORM

Liquid

COLOR

Light yellow

ODOR

Odorless

pH

<1
STABILITY AND REACTIVITY

CONDITIONS TO BE AVOIDED

Heating

SUBSTANCES TO BE AVOIDED

Metals, base, acid

DANGEROUS BY-PRODUCTS

In case of fire, see FIRE FIGHTING MEASURES


SPECIFIC HAZARDS above
TOXICOLOGICAL INFORMATION

ACUTE TOXICITY

CL50 (oral, rat) : 3250 mg/Kg

IN THE EVENT OF SKIN CONTACT

Causes burns

IN THE EVENT OF EYE CONTACT

Causes burns

IF SWALLOWED IN LARGE QUANTITIES

Strong pains with nausea, vomiting and diarrhea

Page B12

SILKOSTAT 9210

ECOLOGICAL INFORMATION
ECOTOXIC

No specific indications

OTHER ECOLOGY-RELATED INFORMATION

In the event of proper handling and use, there is no


risk of ecological problems.
Injurious on aquatic life, by pH modification

DISPOSAL CONSIDERATIONS
There are as yet no uniform regulations concerning
the disposal of chemicals and their residues within
the European Union.

PRODUCT

The chemicals and their residues must be considered


as special waste; their disposal is regulated by each
EEC member state.
We advise you to contact the competent authorities
(administration or specialist waste disposal
companies) that will inform you of the disposal
measures to take
Disposal in accordance with prevailing legislation.
Contaminated containers must be processed in the
same manner as the chemical itself.

PACKAGING

Unless specifically regulated, non-contaminated


containers may be disposed of as household waste,
or be recycled
TRANSPORT INFORMATION
Not subject to transport provisions
REGULAR INFORMATION
N CE : 231-211-8

LABELLING TO EEC DIRECTIVES


HAZARDOUS PRODUCT IDENTIFICATION
C

Corrosive

R Sentences

35

Causes burns

S Sentences

26 - 30 - 36/37/39-45

If in contact with the eyes, flush


eyes immediately with plenty of
water.
Wear protective clothes, gloves,
goggles and a face protection
device. In case of accident or
feeling faint, contact a doctor

OTHER INFORMATION
Updated version : November 2005 Issue A

Page B13

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : 9210 R2 REAGENT
PRODUCT AND COMPANY IDENTIFICATION
Product code : R2

SUBSTANCE

Name of product : 9210 R2 Reagent


Manufacturer Name : HACH SAS
Address : 29 rue de Bures
14670 Troarm

SUPPLIER

Phone
: (33) 02.31.15.37.10
Fax N
: (33) 02.31.15.37.30
Emergency : (33) 02.31.15.37.35 (Unit of production)
: (33) 01.45.42.59.59 (N ORFILA)
COMPOSITION / INFORMATION ABOUT THE COMPONENTS
Preparation

CHEMICAL PROPERTIES
HAZARDOUS COMPONENTS
CAS NUMBER
213-653-7
6153-56-6

SOLUTION
Water
Oxalic acid

QUANTITY
95 %
5%

CLASSIFICATION
Xn ; R21/22

IDENTIFICATION OF HAZARDS
Injurious on contact with skin or by ingestion
FIRST AID
SKIN CONTACT

Rinse affected area with plenty of water. Remove


contaminated clothing.

EYE CONTACT

Immediately flush eyes with plenty of water with


eyelids open

IF SWALLOWED IN LARGE QUANTITIES

Give plenty of water to drink. In case of vomitting,


take medical advice.

FIRE FIGHTING MEASURES


EXTINGUISHING MEANS

Adapt the extinguishing agent to the products stored


in the immediate vicinity

SPECIFIC HAZARDS

Non combustible. Risk of forming hydrogen if in


contact with light metals

Page B14

SILKOSTAT 9210

MEASURES IN CASE OF SPILLAGE


PRECAUTIONS FOR PEOPLE

Avoid contact with skin and eyes

ENVIRONMENT PROTECTION

Do not pour in sewers or rivers

CLEANING

Collect in specific containers. Clean soiled materials


HANDLING AND STORAGE
Store in hermetically sealed containers

STORAGE RECOMMENDATIONS

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION


RESPIRATORY PROTECTION

Required if steam / aerosol

HAND PROTECTION

Wear nitrile gloves

EYE PROTECTION

Safety glasses are required

HYGIENE MEASURES

Wash hands after contact


PHYSICAL AND CHEMICAL PROPERTIES

FORM

Liquid

COLOR

Colorless

ODOR

Odorless

pH

<1
STABILITY AND REACTIVITY

CONDITIONS TO BE AVOIDED

Strong heating

SUBSTANCES TO BE AVOIDED

Bases

DANGEROUS BY-PRODUCTS

No available information
TOXICOLOGICAL INFORMATION

ACUTE TOXICITY

DL50 (oral, rat) : 7500 mg/Kg

IN THE EVENT OF SKIN CONTACT

Slight irritation

IN THE EVENT OF EYE CONTACT

Slight irritation

IF SWALLOWED IN LARGE QUANTITIES

Injurious

Page B15

SILKOSTAT 9210

ECOLOGICAL INFORMATION
ECOTOXIC

Biologic : Fish CL50 : 325 mg/l (anhydrous liquid)

OTHER ECOLOGY-RELATED INFORMATION

Damage :
DCO : 0.18 g/g (compared to the pure substance)
DThO : 0.18 g/g (compared to the pure substance)
DBO : 0.16 g/g (compared to the pure substance)
Do not pour in sewer or in the rivers.

DISPOSAL CONSIDERATIONS
There are as yet no uniform regulations concerning
the disposal of chemicals and their residues within
the European Union.

PRODUCT

The chemicals and their residues must be considered


as special waste; their disposal is regulated by each
EEC member state.
We advise you to contact the competent authorities
(administration or specialist waste disposal
companies) that will inform you of the disposal
measures to take
Disposal in accordance with prevailing legislation.
Contaminated containers must be processed in the
same manner as the chemical itself.

PACKAGING

Unless specifically regulated, non-contaminated


containers may be disposed of as household waste,
or be recycled
TRANSPORT INFORMATION
Not subject to transport provisions
REGULAR INFORMATION
LABELLING TO EEC DIRECTIVES
HAZARDOUS PRODUCT IDENTIFICATION
Xn

Injurious

R Sentences

21/22

Injurious in contact with skin or if


ingested

S Sentences

36/37

Wear protective clothes and


appropriate gloves

OTHER INFORMATION
Updated version : November 2005 Issue A

Page B16

SILKOSTAT 9210

HEALTH AND SAFETY INFORMATION


PRODUCT : 9210 R3 REAGENT
PRODUCT AND COMPANY IDENTIFICATION
Product code : R3

SUBSTANCE

Name of product : 9210 R3 Reagent


Manufacturer Name : HACH SAS
Address : 29 rue de Bures
14670 Troarm

SUPPLIER

Phone
: (33) 02.31.15.37.10
Fax N
: (33) 02.31.15.37.30
Emergency : (33) 02.31.15.37.35 (Unit of production)
: (33) 01.45.42.59.59 (N ORFILA)
COMPOSITION / INFORMATION ABOUT THE COMPONENTS
Preparation

CHEMICAL PROPERTIES
HAZARDOUS COMPONENTS
CAS NUMBER
213-653-7
7783-85-9
7664-93-9

SOLUTION
Water
Ammonium iron (III) sulphates
Sulphuric acid

QUANTITY
96 %
2%
2%

CLASSIFICATION

C ; R35 ; S26 -30 - 45

IDENTIFICATION OF HAZARDS
Product not considered as dangerous according to the 67/548/CEE directive
FIRST AID
SKIN CONTACT

Rinse affected area with plenty of water. Remove


contaminated clothing. Swab with polyethylene glycol
400

EYE CONTACT

Immediately flush eyes with plenty of water with


eyelids open for 10 minutes. Get medical aid if
necessary

IF SWALLOWED IN LARGE QUANTITIES

Give plenty of water to drink. In case of indisposition,


take medical advice. Do not try to neutralize

FIRE FIGHTING MEASURES


EXTINGUISHING MEANS

Adapt the extinguishing agent to the products stored


in the immediate vicinity

SPECIFIC HAZARDS

Non combustible. Risk of forming hydrogen if in


contact with light metals

Page B17

SILKOSTAT 9210

MEASURES IN CASE OF SPILLAGE


PRECAUTIONS FOR PEOPLE

Avoid contact with skin and eyes

ENVIRONMENT PROTECTION

Do not pour in sewers or rivers

CLEANING

Collect in specific containers. Clean soiled materials


HANDLING AND STORAGE

PRECAUTIONS TO TAKE

Avoid contact with metals

USEAGE ADVICE

Respect conditions of use

STORAGE RECOMMENDATIONS

Store in hermetically sealed containers

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION


RESPIRATORY PROTECTION

Required if steam / aerosol

HAND PROTECTION

Wear nitrile gloves

EYE PROTECTION

Safety glasses are required

OTHER PROTECTION

Acid-proof clothing

HYGIENE MEASURES

Wash hands after contact


PHYSICAL AND CHEMICAL PROPERTIES

FORM

Liquid

COLOR

Colorless

ODOR

Odorless

pH

<1
STABILITY AND REACTIVITY

CONDITIONS TO BE AVOIDED

Strong heating

SUBSTANCES TO BE AVOIDED

Metals, base, acid

DANGEROUS BY-PRODUCTS

In case of fire, see FIRE FIGHTING MEASURES


SPECIFIC HAZARDS above
TOXICOLOGICAL INFORMATION

ACUTE TOXICITY

DL50 (oral, rat) : 7500 mg/Kg

IN THE EVENT OF SKIN CONTACT

Causes burns

IN THE EVENT OF EYE CONTACT

Causes burns

IF SWALLOWED IN LARGE QUANTITIES

Strong pains with nausea, vomiting and diarrhea

Page B18

SILKOSTAT 9210

ECOLOGICAL INFORMATION
ECOTOXIC

No specific indications

OTHER ECOLOGY-RELATED INFORMATION

In the event of proper handling and use, there is no


risk of ecological problems.
Injurious on aquatic life, by pH modification

DISPOSAL CONSIDERATIONS
There are as yet no uniform regulations concerning
the disposal of chemicals and their residues within
the European Union.

PRODUCT

The chemicals and their residues must be considered


as special waste; their disposal is regulated by each
EEC member state.
We advise you to contact the competent authorities
(administration or specialist waste disposal
companies) that will inform you of the disposal
measures to take
Disposal in accordance with prevailing legislation.
Contaminated containers must be processed in the
same manner as the chemical itself.

PACKAGING

Unless specifically regulated, non-contaminated


containers may be disposed of as household waste,
or be recycled
TRANSPORT INFORMATION
Not subject to transport provisions
REGULAR INFORMATION
N CE : 23-211-8

LABELLING TO EEC DIRECTIVES


HAZARDOUS PRODUCT IDENTIFICATION
C

Corrosive

R Sentences

35

Causes burns

S Sentences

26 - 30 - 36/37/39-45

If in contact with the eyes, flush


eyes immediately with plenty of
water.
Wear protective clothes, gloves,
goggles and a face protection
device. In case of accident or
feeling faint, contact a doctor

OTHER INFORMATION
Updated version : November 2005 Issue A

Page B19

SILKOSTAT 9210

LITHIUM BATTERY SAFETY PRECAUTIONS

The lithium battery on the Datalyt Microprocessor p.c.b does not constitute a serious hazard
providing the following instructions are observed :

NEVER :
Attempt to recharge the battery.
Crush, puncture, open or dismantle the battery.
Store or operate equipment above the temperature specified.
Short-circuit the battery.
Dispose of the battery in a fire.
Try to repair faults on the control p.c.b.
In the event of a p.c.b fault or battery leakage :

ALWAYS :
Wear suitable protective clothing.
Remove the equipment to a well ventilated area.
Contact your local agent.
Refer to Health and Safety Notes on Lithium batteries.

Page B20

SILKOSTAT 9210

Appendix C : Table of users configuration


Users configuration
Order

Function

offset interval
cal. sol
auto. cal
base
week
hour

Frequency
Concentration
Auto calibrate

Default/Values

Possible values

Selected values

Date

CALIBRATION / PROGRAMMING
24 h
214 ppb
No
Day
Mtwtfs
00:00:00

16
Xxxxx
yes/no
Day/hour
Calendar Mtwtfs
Hours

MAINTENANCE/REAGENTS REFILL
Reagents vol. 1
Reagents vol. 2
Reagents vol. 3
Reagents vol. 4

Used reagent volume


Used reagent volume
Used reagent volume
Used reagent volume

2,00 l
2,00 l
2,00 l
2,00 l

Xx
Xx
Xx
Xx

MEASURE/MEASURE MODE
Cycle
Interval

Interval
15 mn

interval/in loops
xx hour/minute

ALARMS / ALARM 1
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 1
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

ALARMS / ALARM 2
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 2
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

ALARMS / ALARM 3
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 3
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

ALARMS / ALARM 4
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 4
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

Page C1

SILKOSTAT 9210

Users configuration
Order

Selected values

Selected values

Possible values

ALARMS / ALARM 5
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 5
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

ALARMS / ALARM 6
Mode
Affect
Limit
Direction
Hysteresis
Relay

Limit
Channel 6
1000 ppb
Low
0%
N.O.

limit/active channel/level sample.

Conc
Xxx
low/high
Xx
N.O./N.C.

ALARMS / WARNING ALARM


Alarm
Accept
Relay

Yes
Manual
N.O.

yes/no
Manual/automatic
N.O./N.C.

ALARMS / SYSTEM ALARM


Alarm
Accept
Relay

Yes
Manuel
N.C.

Oui/non
Manual/automatic
N.O./N.C.

mA OUTPUTS/ AFFECT / OUTPUT 1


Affect
Type
Mode
Low
Middle
Up

Channel 1
0-20 mA
Linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

mA OUTPUTS/ AFFECT / OUTPUT 2


Affect
Type
Mode
Low
Middle
Up

Channel 2
0-20 mA
linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

mA OUTPUTS/ AFFECT / OUTPUT 3


Affect
Type
Mode
Low
Middle
Up

Page C2

Channel 3
0-20 mA
Linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

Selected values

Date

SILKOSTAT 9210

Users configuration
Order

Selected values

Selected values

Possible values

Selected values

Date

mA OUTPUTS/ AFFECT / OUTPUT 4


Affect
Type
Mode
Low
Middle
Up

Channel 4
0-20 mA
Linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

mA OUTPUTS/ AFFECT / OUTPUT 5


Affect
Type
Mode
Low
Middle
Up

Channel 5
0-20 mA
Linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

mA OUTPUTS/ AFFECT / OUTPUT 6


Affect
Type
Mode
Low
Middle
Up

Channel 6
0-20 mA
Linear
0 ppb
100 ppb
1 ppm

Channel 1Channel 6
0/20 mA ou 4/20 mA
Linear / dual
XXX
XXX
XXX

mA OUTPUTS / SPECIAL PROG / MAINTENANCE


Mode
Value if preset mode selected :

Last
0 mA

Last/preset/live

Xx

mA OUTPUTS / SPECIAL PROG / CALIBRATION


Mode
Value if preset mode selected:

Last
0 mA

Last/preset/live

XX

mA OUTPUTS / SPECIAL PROG / GRAB SAMPLE


mode
Value if preset mode selected:

Last
0 mA

Last/preset/live

XX

mA OUTPUTS / SPECIAL PROG / SYSTEM ALARME


mode
Value if preset mode selected:

Last
0 mA

Last/preset/live

XX

mA OUTPUTS / SPECIAL PROG / WARNING ALARME


mode
Value if preset mode selected:

Last
0 mA

Last/preset/live

Xx

COMMUNICATION
N
Baud
Parity
Stop bit

n of Monec
Speed communication

0
9600
No
1

0 33
300,600,1200,2400,4800,9600
No, odd, even
1 or 2

Swap word

Word send order

No

Yes/no

Page C3

SILKOSTAT 9210

Users configuration
Order

Selected values

Selected values

Possible values

SEQUENCE / CHANNEL ACTIVATION


Channel 1
Channel 2
Channel 3
Channel 4
Channel 5
Channel 6

On
On
On
On
On
On

on / off
on / off
on / off
on / off
on / off
on / off

SEQUENCE / PROGRAMMING
Sequence

Sequence programming

123456123456

123456*

SERVICE/TIME
Day
Date
Time

Day of the week


Date of the day
Hour

Monday
01/01/01
09:25:55

Monday Friday

SERVICE/DISPLAY
Language
concentration

GB
Ppb

F/GB/D/SP/I
ug/l / ppb

SERVICE/CUSTOMIZE
Sample 1
Sample 2
Sample 3
Sample 4
Sample 5
Sample 6

name
name
name
name
name
name

Sample 1
Sample 2
Sample 3
Sample 4
Sample 5
Sample 6

a..z and 0..9


a..z and 0..9
a..z and 0..9
a..z and 0..9
a..z and 0..9
a..z and 0..9

SERVICE / CONFIGURATION
Mains freq.
Reference freq.
Freq. noir

60 Hz
0
9999

50/60 Hz
Xx
Xx

SERVICE/ FACTORY
Channel number

1 to 6

SERVICE/ FACTORY/SOLENOID VALVES


Sample N
Toff sample
Cal. Sol. N
Toff cal. sol.

Open time adjust


Close time adjust
Open time adjust
Close time adjust

3
180 s
3
180 s

Pump 1 Ton
Pump 1 Toff
Pump 2 Ton
Pump 2 Toff
Pump 3 Ton
Pump 3 Toff

Impulses number
Waiting time
Impulses number
Waiting time
Impulses number
Waiting time

Type

Choice of the range


0-1000 or 0-5000 ppb

0-1000 ppb

Linear coeff.
Waiting time
Measure time
Conc. zero
Filter
Heating
Negative values

Coefficient of non-linearity

0
1s
10 s
0 ppb
50 mV
Yes
no

xx
xx s
xx
xx s

SERVICE/ FACTORY/ PUMPS


5
240 s
5
15 s
5
60 s

Xx
xx s
Xx
xx s
Xx
xx s

SERVICE/ FACTORY/MEASURE

Page C4

Before the measure


Extended measure
Reserved
Reserved
Heating activation
Display or no display of
the negative values

0-1000 ppb
0-5000 ppb
0---XXX
xx s
xx s

Yes/no
Yes/no

Selected values

Date

SILKOSTAT 9210

Appendix D : Pressure range change


The battery of the input solenoid valves works with pure water sample.
The pressure range is about 0,2 to 2 bars. If the pressure is about 0,5
to 6 bars, it is required to change the solenoid valves collar (see figure
1 item 6). So a High pressure kit contains 6 collars (0,8 diameter)
and a TORX key.
Figure 1 Mounting of the solenoid valves

1 Hold down screw of the solenoid valves


2 Order coil
3 Body fixing screw
4 Solenoid valves main part
5 Flow direction
6 Solenoid valves collar
7 Seal
8 Calibration steel block
9 Sample output
10 Undeceiver

Page D1

SILKOSTAT 9210

To change the solenoid valves collar, proceed as follows :


Unscrew the 2 superior screw (1) of the solenoid valves.
Move the solenoid valves coil
Unscrew the 2 screws (3) of the main part fixing (4).
Move the collar (6) of the body (4).
Change the 2 O-rings (7) of the collar (6).
Replace the joints to the new collar.
Put the collar in position on the steel block (8) in the good direction.
Place the undeceiver (10) in the indicated position in terms of the
cal output (9).
Replace the main part (4) in position ; mark NC (5) and sample
output (9) must be placed in the same side.
Re-screw the fixing screw (3).
Replace the coil (2) and tighten with help of the 2 screws (1).

Page D2

SILKOSTAT 9210

Appendix E : Preparation of the tubes for insertion in the fast


connections
Figure 1 Slot marking

Locate the slot to realise (see the table below)


Tube length

6 mm

1/4

5/16

7,3 0,5

0,33 0,02

0,36 0,02

Figure 2 Fasten the knurl

Place the tube in the tube-cutter and set the knurl on the mark.
Fasten the knurl a quarter turn.
Turn the tool around the tube and check the presence of a slot all
around the turn.
Remove the tube-cutter.

Page E1

Vous aimerez peut-être aussi