Just a way to keep the prism:-

Setting of the collimator and the


telescope
Setting for prism table , collimator, and prism.
Measuring the prism angle:-
Finding the angle of prism.
1) Place the prism on the turn table in such a way that the refracting edge A coincide with the centre
of the table and one of the refracting or polished face say AB remains perpendicular to the line joining the
screws Q and R.
2)Now , rotate the prism table to bring the refracting edge A of the prism towards the collimator so
that the light from the collimator falls equally on both faces AB and AC bounding the refracting angle A. the
prism angle is then clamped.
3) determine the least count of the spectrometer.
4) rotate the telescope to the right side to receive the light reflected from the face AB and focus the
cross wire on the image of the slit . Now by tangent screw , the telescope is moved slowly until its vertical
cross wire coincide with the image . The readings of the two verniers of the spectrometer are taken.
5) Now rotate the telescope to the left side to receive the light reflected from the face AC and again
, focus the cross wire on the image of the slit by means of the tangent screw . The readings of the two
verniers are again noted.
6) the difference of the two vernier readings of the same vernier for two positions of the telescope
gives twice the refracting angle A of the prism . Thus , half of this angle will give the value of the angle A of
the prism.
Setting for the minimum deviation:-
deviation:-
Angle of minimum deviation:-
deviation:-

The prism is placed on the turned table G such that the axis of rotation of the turn
table passes through the centre of the prism one of its polished face AB is turned
towards the collimator in such a way that light after refraction emerges out from the
other polished face AC and the dispersed light is received in the telescope .Now the turn
table is slowly rotated in such a direction that the deviation of the rays decreases . The
telescope is also rotated so as to keep the spectrum in the field of view . The rotation is
continued until the spectrum comes to momentary rest and then turns back.

the position of the prism at which the spectrum is just at rest is the position of the
minimum deviation . The turn table is clamped in this position.

the position of the telescope is so adjusted by means of the tangent screw that the
intersection of the cross wires is exactly at the centre of the spectral line under
investigation. the position of the telescope is read on both the verniers. Now the prism is
removed and the telescope is turned to receive the direct image of the slit on the cross
wires and its position is again read. The difference between the readings in the two
positions of the telescope gives the angle of minimum deviation.