Académique Documents
Professionnel Documents
Culture Documents
Monitoring by
Data Logging
Data are usually collected simultaneously at multiple locations within the area being tested. Locations are carefully chosen so that the entire area is
mapped. The data are then analyzed to determine
the worst-case locations within the area or system
(i.e., the hottest or coldest locations). The data
are then used to perform additional calculations
including the following:
Minimums
Maximums
Averages
Standard deviations
Lethality of a sterilization cycle (F-sub-zero).
Because a complete data-logging event contains
hundreds to thousands of data points, these calculations make it possible to draw conclusions about the
system from the data collected.
Data logging is critical to obtaining an accurate
and detailed representation of how a system operates.
Some systems, such as a sterilizer, change temperature very fast. Others, such as a refrigerator, change
temperature slowly. These systems may also have
long cycle times that make it impractical for manual
data recording.
Data logging also enables a significant number of
measurements to be collected over short or long periods of time. The large number of measurements can
easily surpass the quantity of points required to be
statistically significant.
This discussion of data loggers will address the
following topics:
Types of data loggers, including stand-alone and
computer-based instruments
Temperature elements used in data loggers,
including thermocouples, thermistors, and resistance temperature detectors (RTDs)
The configuration of data loggers, including
wiring, multiple channels, and self-contained
instruments
Calibration of data loggers
Use of data loggers in validation (qualification),
including environmental chambers, steam sterilizers, warehouses, and production rooms
C
ollecting data, including probe placement,
Spring 2009 Volume 13 Number 2
33
E n g i n ee r i n g f o r N o n - E n g i n ee r s
Figure 1:
Kaye Validator 2000.
Figure 2:
ACR Systems SmartReader.
(Reprinted with permission from ACR Systems)
Figure 3:
Figure 4:
Veriteq data logger.
35
E n g i n ee r i n g f o r N o n - E n g i n ee r s
assessment should be performed to determine the required calibration tolerance and accuracy; this assessment should be documented and include the downstream impact of an OOT or inaccurate measurement.
This risk assessment can refer to any critical process
parameter (CPP) justifications because similar information and logic is required. The calibration must
also be performed against a National Institute of
Standards and Technology (NIST) traceable standard
that has a tolerance less than or equal to the accuracy
needed for the instrument.
Range
Instruments cannot be assumed to provide accurate
readings over their full operating range. The calibration must be performed over the intended measuring range. If the measurements are expected to be
in the range of 20-40C, then the calibration at a
minimum should include this range. A wider range
could also be used, such as 10-50C. This allows for
temperature excursion that may not be anticipated.
This specific instrument should not be used for taking measurements in a freezer that has an operating
range of -40C to -20C.
Calibration Checks
Prior to starting a data logging study, the instrument
should be precalibrated with a full calibration. A full
calibration includes multiple data points that bracket
the range of measurements that will be collected. Following completion of the study, the instrument should
be post calibrated with either a full calibration or a
calibration check. A calibration check is a single-point
calibration that is usually performed at the setpoint,
or middle point, of the systems operating range. This
helps reduce the cost of calibration activities.
QUALIFICATIONS USING DATA LOGGERS
The following are several basic concepts common to
all qualifications that utilize data loggers:
Generating a temperature map of the equipment or room during production
Creating a temperature map by placing probes at
various locations throughout the space to identify
the hottest and coldest locations within the space
V
erifying that the temperature cycles of the
equipment or room remain within the required
temperature ranges by analyzing the continuous
data measurements.
Environmental Chambers
Environmental chambers include any equipment
that is designed to maintain certain environmental
conditions, such as temperature, humidity, or CO2
level. Equipment in this category includes, but is not
limited to the following:
Refrigerators
Freezers
Ultra-freezers
Cryogenic tanks
Incubators.
Operational and/or performance qualification of
these chambers involves temperature mapping the
chamber in reasonable worst-case conditions. The
qualification is written to ensure that the entire volume
of the chamber is maintained within a specified operating temperature range for a given duration.
Steam Sterilizers
Steam sterilizers are used to keep items free of living microorganisms and bacterial endospheres prior to transfer
into a sterile production room or for products prior to
shipment. Steam sterilizers operate at temperatures of
at least 121.1C. Data logging is crucial to identifying
whether or not all locations in the chamber reach and
maintain 121.1C for the duration of the sterilization
cycle. This test is an operational qualification of the
sterilizer and is performed with an empty chamber.
Because the sterilizer monitors temperate at the
drain, which is the worst-case location, a thermocouple
should also be placed at this location. Failure at any
of the data points in the sterilizer could indicate a
mechanical problem with the sterilizer. There may be
a leak in the door seal, inadequate insulation, or insufficient steam supply.
Performance qualification of the sterilizer is executed with a worst-case load using actual items that
will be processed. Thermocouples are inserted into
the items in the sterilizer at the worst-case locations.
Spring 2009 Volume 13 Number 2
37
E n g i n ee r i n g f o r N o n - E n g i n ee r s
39
E n g i n ee r i n g f o r N o n - E n g i n ee r s
Figure 5:
Environmental chamber with four storage levels.
Figure 6:
Freezer 24-hour temperature map.
Degrees celcius
-24.0
'2-01
-26.0
'2-02
'2-03
'2-04
'2-05
-28.0
'2-06
'2-07
'2-08
'2-09
'2-10
-30.0
'2-11
'2-12
-32.0
-34.0
0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
:0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0 :0
00 :50 :40 :30 :20 :10 :00 :50 :40 :30 :20 :10 :00 :50 :40 :30 :20 :10 :00 :50 :40 :30 :20 :10 :00 :50 :40 :30 :20
:
7 7 8 9 10 11 12 12 13 14 15 16 17 17 18 19 20 21 22 22 23 0 1 2 3 3 4 5 6
Time
Spring 2009 Volume 13 Number 2
41
E n g i n ee r i n g f o r N o n - E n g i n ee r s
QUALITY CONCERNS
The wide use of data logging in the pharmaceutical industry makes it very likely that most people
will encounter this activity in some way or another
during their careers. Quality organizations may be
approached to review or approve these activities.
This helps to ensure that the tests are pre-approved
and executed in such a way that the data is valid and
meets CGMP requirements. This article is merely
an introduction to the concepts of data logging, and
should provide a basic understanding of what is involved and some of the important factors to consider.
When evaluating a test plan, the following items
could impact quality if not executed properly:
Data loggers are calibrated before and after
the qualifications. Calibrations bracket the
expected operating range of the equipment
or room. The calibrated tolerance of the data
loggers is tighter than the specifications for the
system being tested.
Probes are placed in worst-case locations with
supporting justification. The author of the test
plan should exhibit a general knowledge of the
system through the justifications provided in the
test plan.
Extra probes are available in case of a malfunction or calibration failure. If no additional
probes are included, the qualification should
allow for the exclusion of one or two data loggers in the event of a malfunction or calibration
failure.
S
ampling interval and test duration are appropriate. The sampling interval should be relatively short for dynamic systems. The test duration
should capture several heating or cooling cycles.
P
roper handling of the original, transcribed,
and printed data. Electronic files are saved to a
CD, original data is printed and verified, and any
transcriptions to Excel and subsequent calculations are verified by a second person.
CONCLUSIONS
This discussion of data logging has addressed several
general topics, including equipment, calibration,
configuration, probe placement, data handling, and
REFERENCES
1. Information about the Kaye Validator 2000 can be obtained
online at http://www.gesensing.com and http://www.kayeinc.
com.
2. Information about ACR Systems SmartReaders can be obtained online at http://www.acrsystem.com.
3. Information about TempTale data loggers is available online at
http://www.sensitech.com.
4. Information about Veriteqs data logger can be found online at
http://www.veriteq.com.
5. Veriteq, Temperature Measurement: Thermocouples vs.
Thermistors, Veriteq.com, http://www.veriteq.com/validation/
thermocouples-vs-thermistors.htm, retrieved March 3, 2009.
6. Veriteq, Thermal Validation in the Pharmaceutical Industry:
An argument against the use of thermocouples, Veriteq.com,
http://www.veriteq.com/thermistor/, retrieved March 3, 2009.
7. Gesensing, Thermometrics Products, Gesensing.com,
http://www.gesensing.com/thermometricsproducts, retrieved
March 3, 2009.
43