Vous êtes sur la page 1sur 30

Series

Mammography X-ray System

Service Manual

Section 3.
Calibration Manual
of the
MX-600

Diagnostic Mammography X-ray System

RADIEQ CO., LTD.

MX-600

Series

Mammography X-ray System

Service Manual

Table of Contents

1. Calibration Overview ....

2. Required Staff and Used Tools ......

....

4. Calibration and Adjustment ....

Block Diagram ....

4.1. Cover Removal of Control Box ....

4.2. Mode Switch of CPU Board ......

....

10

4.4. Selection Method in Adjustment Sub Mode 1.x ....

10

4.5. Initial Data Transmission of CPU Board .....

11

4.6. Connection of Measuring Instrument ......

13

4.7. Setting Adjustment Mode ......

14

4.8. Adjustment of X-ray Tube Current ...

15

4.9. Adjustment of AEC .....

17

4.10 Mode Setting ...................

23

4.11 Data Sheet .....

24

3. Pre-Check list

4.3. Selection Method in Adjustment Mode

.
5. Mechanical Adjustment ...

28

5.1. Adjusting the Light Field ...

28

5.2. Adjusting X-ray exposure field .....

29

RADIEQ CO., LTD.

MX-600

Series

Mammography X-ray System

Service Manual

1. Calibration Overview
All new equipment installations will require that M101-5401 MAIN CPU board be
adjusted. If you are comfortable with your evaluation of kV/mA output findings,
proceed to AEC & mAs procedures.
ALL Radieq Co.,Ltd. EQUIPMENT IS PRE-CALIBRATED AT THE FACTORY. THE
FOLLOWING PROCEDURES ARE PROVIDE ONLY IN EVENT THAT THE UNIT
FAILS TO MEET SPECIFICATIONS.
It is recommended that all electrical and mechanical adjustments be checked
annually.
When performing the following procedures, follow the sequence at which the
procedure is given, as well as its content.
It is suggested that you calibrate the equipment annually, first verify the performance
to determine whether or not a complete calibration is required.
Always begin by checking the kV/mA/mAs DIGITAL REGERENCE SIGNAL that
were recorded on the "DATA SHEET". You can either spot check or verify all
selected ranges for output comparison for what was recorded on the "DATA SHEET"
on the initial installation or previous calibration. Always start with a new copy of
the "DATA SHEET" and add it to the equipment records.
In the event the
CPU(IC1,Main CPU Board) is replaced or the Non-volatile RAM should fail, the
information on the "DATA SHEET" will reduce the number of X-ray's required to
re-calibrate the mammography unit.
By following the procedures, we are actually reprogramming the information fund in
the RAM memory. This allows us to fine tune the broad exposure information
supplied to the buffer from Non-volatile RAM chip. The information in the
Non-volatile RAM will be maintained even with power off by inner the battery
back-up system. Should this system fail for whatever reason, the RAM memory will
be lost and no exposure can take place. In this event, the programming must be
re-supplied to the Non-volatile RAM from the ROM memory and the kV and mA
calibration performed again. Having recorded all of the data on the "DATA SHEET"
and within this section of the manual, you should be able to quickly restore the
conditions within a very short time.

RADIEQ CO., LTD.

MX-600

Series

Mammography X-ray System

Service Manual

Caution

Remember we are dealing with small focal spot size. Take your time to
prevent tube over loading, and you will gain faster result in calibration. Allow
30 seconds between exposures.

RADIEQ CO., LTD.

MX-600

Series

Mammography X-ray System

Service Manual

2. Required Staff and Used Tools


The following personnel and tools are required to Calibration MX-600, and they
must be organized beforehand.
(1) Tools and instruments required
(2) One set of general purpose tools

Screwdriver (small and medium size):


Flat screwdriver for adjustments:
M6 and M10 wrenches:

Each 1
1
Each 1

(3) Unpacking tools:

1 set

(4) AC voltmeters to measure the power supply voltage:

(5) One X-ray tube voltmeter set:


1 set

X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent

X-ray tube voltmeter cable


(6) X-ray tube current meter:
1 set

X-ray tube voltmeter/mAs meter as Dynalizer III or equivalent


(7) Acrylic phantoms for AEC adjustment

10mm thickness (120mm x 120mm):

(8) Density meter:

RADIEQ CO., LTD.

MX-600

Series

Mammography X-ray System

Service Manual

3. Pre-Check List
Date Installed :
Caution

Equipment S/N:

Do not apply power without first checking that all boards are properly
seated and all connectors and wires are connected exactly, damaged
could occur to the unit.

Changing the Power Regulator (PSU1) supply voltages can result in a change in the
calibration of the mammography unit. Record below the initial PWR supply voltage
setting, if a PWR supply is replaced adjust it to the recorded value below to avoid
extensive re-calibration.
1) Turn on the line breaker on the Control Box side.
2) Check that all power supply voltages present, do not change outputs if they are
within the tolerance of 2%:
+5VDC

+12VDC

-12VDC

+24VDC

3) Turn off the line breaker on the Control Box side.

RADIEQ CO., LTD.

MX-600

PSU1

RADIEQ CO., LTD.

ROTOR &
IF BOARD

MAIN CONTROL
BOARD

MAIN OP
BOARD

PSU2

SUB OP1
BOARD

M_INV
BOARD1

INVERTER
BOARD

COM. MOTOR
DRIVER BOARD

MOTOR CONTROL
BOARD

SUB OP1
BOARD

MX-600 BLOCK DIAGRAM


M_INV
BOARD2

HFG

TUBE

Series

Mammography X-ray System

Service Manual

4. Calibration and Adjustment

Fig.1 MX-600 Block Diagram.

MX-600

Mammography X-ray System

Series

Service Manual

4.1. Cover Removal of Control Box


Before starting calibration and adjustment, please remove sideboards of control box
as shown in Fig. 2.

3
1

Fig.2 Cover Removal of Control Box.


1)
2)
3)
4)

At first loosen screws and remove the front cover 1.


Next loosen screws and remove the side cover 2.
Next loosen screws and remove the side cover 3.
Finally loosen screws and remove the rear cover 4.

RADIEQ CO., LTD.

MX-600

Mammography X-ray System

Series

Service Manual

The Boards are showed in Fig. 3 as below.

MAIN CPU
BOARD

I/F&ROTOR
BOARD

C1

PSU1

T1

S2 (MODE S/W)

TB10

HIGHTENSION
TRANSFORMER

INVERTER

HIGHTENSION
TRANSFORMER

FUSE
BOARD

TB2
INVERTER

TB1
Control Box
[ FRONT ]

Control Box
[ BACK ]

Fig. 3 Position of Boards

4.2 Mode Switch of CPU Board


Each following mode is chosen by selection of the number of the MODE switch(S2)
on Main CPU board.
0 - - - - - - - - - - - - - - - Normal mode(Operation mode)
1 - - - - - - - - - - - - - - - Adjustment mode (Calibration mode)

CE
F01

23
RADIEQ CO., LTD.

Mode switch : S2

MX-600

Series

Mammography X-ray System

Service Manual

4.3 Selection Method in Adjustment Mode


The contents of Display in the adjustment mode are displayed by 7 segment Light
Emitting Diode on the left-hand side of kgf.
The adjustment mode is chosen by Density switch
0 Adjusting mode of mA

0.

kgf

1Adjusting mode of AEC

1.

kgf

Setting for a various Mode

2.

kgf

Selection mode of the transmission data

3.

kgf

2
3
4

4.4 Selection Method in Adjustment Sub Mode 1.x


The contents of Display in the adjustment mode are 7 segment LED on the
right-hand side of Density. It is displayed numerically.
The switch of Density chooses the adjustment sub mode.
In the case of AEC adjustment:
0 - - - - - - - - - - - - - - Semi AEC adjustment
1-------------

RADIEQ CO., LTD.

Density Data adjustment

1. 0

kgf

1. 1

kgf

MX-600

10

Mammography X-ray System

Series

Service Manual

4.5 Initial Data Transmission of CPU Board


Initial data other than individual data of the following clause is transmitted to
Non-volatile RAM.
Disconnect the connection cable between Inverter and MAIN CPU
board(M101-5401) at MAINCPU board side.
The MODE switch S2 of CPU board is set as No. 1.
Turn on the circuit breaker

kV

mAs

DENSITY

The display of kV blinks.


NOTE:

is showed a blank mark.

4.5.1
(1)

CPU Board Individual Data Transmission


Set up the adjustment mode Push Switch"" of DENSITY.
Then a display is made into 3. by 7 segment LED on the left side
of Density.

3.
kV

mAs

DENSITY

kgf
mm
DEG.

(2)

Push Switch of kV,


A display of 7 segment LED of kV is made into the following
selection number.
8...CPU RAM to Non-volatile RAM

3.
kV

mAs

DENSITY

kgf
mm
DEG.

(3)

RADIEQ CO., LTD.

Pushing AEC switch is continued for 1 second.


7 segment LED of mAs blinks, about 10 seconds after, a buzzer
sounds for 1 second and completion is told.

MX-600

11

Series

Mammography X-ray System

Service Manual

(4)

Push Switch of kV,


A display of 7 segment LED of kV is made into the following
selection number.
9 .Non-volatile RAM to Buffer

3.
kV

mAs

DENSITY

kgf
mm
DEG.

(5)

Pushing AEC switch is continued for 1 second.


7 segment LED of mAs blinks, about 10 seconds after, a buzzer
sounds for 1 second and completion is told.

NOTE 4.5.1 clauses needs to be operated, only when the Non-volatile RAM for backup
the case where MAIN CPU board is changed into a new , or CPU board is exchanged
Here, since the data transmission from RAM / Buffer finished to Non-volatile RAM, Line
Breaker is turned off, and the cable between an inverter and CPU board is inserted in a
connector.

RADIEQ CO., LTD.

MX-600

12

Mammography X-ray System

Series

Service Manual

4.6 Connection of Measuring Instrument


Connect tube voltmeter and mA or mAs meters as shown in Fig. 4.
High-voltage
transformer
High-voltage cable
for measurement

Control box

Unit high-voltage cable


Anode

TB 2
MA+

X-ray tube

MA-

Cathode

mA meter

kV meter

or

mAs meter

Fig. 3 Connection for kV and mA/mAs meter


X-ray tube voltmeter
Between a high-voltage generator and an X-ray tube, the breeders for X-ray
high-voltage measurement (DYNALYZER, etc.) are connected using the
high-voltage cable adapter for the mammography device.
X-ray tube current
The jumper line between terminal MA+ to MA- of X-ray tube current
measurement terminal stand TB2 is removed, and an mA meter(or mAs meter) is
connected between them.

Caution

If between terminal MA+ to MA- of the terminal stand TB2 become open,
the unusual high voltage is impressed by the X-ray tube, and since it may
be damaged, you be careful of connection of an ammeter enough.

RADIEQ CO., LTD.

MX-600

13

Series

Mammography X-ray System

Service Manual

4.7 Setting Adjustment Mode


(1) The MODE switch S2 of CPU board is set as position 1.
2 Turn on Line breaker.

kV

mAs

DENSITY

Display becomes as it is shown in a figure.


NOTE: The setting method of the MODE switch S2 unites a number with the portion
with a cut of a knob.

Caution

X-ray exposure of continuation is not made. Since there is a possibility that


equipment may be damaged, please take the pause time of 30sec or more
in X-ray exposure.

(3Push On-(internal) switch of Control Panel .


Switch I ON(

The control circuit will be activating.


Display becomes as it is shown in a figure.

Lamp LARGE F.S. lights up.

(4) The switch of DENSITY is pushed and the adjustment mode is chosen.
Display becomes as it is shown in a figure.

2 2 X X.X
kV

mAs

0.
DENSITY

kgf
mm
DEG.

RADIEQ CO., LTD.

MX-600

14

Series

Mammography X-ray System

Service Manual

4.8 Adjustment of X-ray Tube Current


NOTE

There is no adjustment of X-ray tube voltage.

An adjusting point is shown below.

Large Focus22kV, 25kV, 30kV, 35kV, 39kV .

Small Focus22kV, 25kV, 30kV, 35kV


4.8.1Adjustment of mA
NOTE: Record this adjustment data to DATA SHEET which are indicated by
4.12.1 clause.
These data is used as future adjustment and reference data at the time of
maintenance.
1)

A setup in the adjustment mode


Display becomes as it is shown in a figure. Lamp LARGE F.S. lights up.

2 2 X.X X
kV

mAs

0.

kgf

DENSITY

2) The switch "" of mAs and "" are pushed and mA data is inputted.

2 2 7 5.8
kV

mAs

0.

kgf

DENSITY

A switch "AEC" is pushed and data is saved.

3) An X-ray exposure switch is pushed and expose, X-ray tube voltage and
current are measured.

The tube current is displayed on mAs display position during about 5 seconds after
an X-ray exposure end. (At the time of the adjustment mode).

RADIEQ CO., LTD.

MX-600

15

Mammography X-ray System

Series

Service Manual

When measured value does not fulfill specified value, a switch "" or "" of mAs is
pushed, mA setting value is changed, and it measures again.
If standard value is satisfied, switch "AEC" is pushed and data is saved.

4) The setting value of kV is changed by switch "" or "" of kV, and


measurement and adjustment of 3) are performed about each kV.
LARGE Focus :

22KV 70mA
25KV 85mA
30KV 72mA
35KV 60mA
49KV 60mA

5) A switch "HELP" is pushed and a focus is changed to a small focus.

It is made 22kV.
Display becomes as it is shown in a figure. SMALL F.S. lamp turns on.

NOTEIf "HELP" switch is pushed again, it will change to a large focus.


6Adjustment to 1 to 5 clause performed on the large focus is similarly
performed with a small focus.

2 2 X.X X
kV

mAs

0.

kgf

DENSITY

SMALL Focus :

RADIEQ CO., LTD.

22KV 13mA
25KV 15mA
30KV 12mA
35KV 10mA

MX-600

16

Mammography X-ray System

Series

Service Manual

4.9 Adjustment of AEC


The tube voltage set as adjustment is 24kV, 27kV, 30kV, 32kV and 35kV, and a
radiograph subject thick is 20mm, 40 mm, 60mm, and 80mm.
Adjustment of MAIN CPU board of the clause 4.9.1 is done at the time of factory
shipment. When the time of exchange of MAIN CPU board or AEC Detector
About density adjustment of the clause 4.9.2 and 4.9.3 when it is installed or
changed to other combination with film and a screen, please carry out.
4.9.1 AEC Tilt Adjustment of CPU Board
1) The switch "" and "" of DENSITY is pushed and the adjustment mode is
set to 1.1
Display becomes as it is shown in a figure.

2)

24

2.5

kV

mAs

1. 1

kgf

DENSITY

Conditions are set up.


Tube Voltage
24 kV(Fixed)
Radiography conditions
BUCKY device
Compression paddle
50mm acrylics

3The switch "" or "" of mAs is pushed, and mAs display is set to 2.5.
Display becomes as it is shown in a figure.

24

2.5

kV

mAs

1. 1

kgf

DENSITY

4A switch AEC is pushed and data is saved.

RADIEQ CO., LTD.

MX-600

17

Series

Mammography X-ray System

Service Manual

5 An X-ray exposure switch is pushed and X-rays is exposed, then


Voltage/Time is measured at the test pin TP1 on CPU board.

6The adjustment of the relation between the voltage waveform of TP1 and time is
carried out by VR1(at Main CUP board) at approximately 2.5V/1sec as shown in
the following figure.
2.5V

Adjustment of VR1

1.0sec

(Time will be extended if it rotates to clockwise)

4.9.2 100mAs Adjustment of CPU Board


Same condition 4.9.1.(AEC Tilt Adjustment of CPU Board)
Conditions are set up.
Tube Voltage24 kV(Fixed)
Radiography conditions
BUCKY device
Compression paddle
50mm acrylics
The switch "" of DENSITY is pushed and the adjustment mode is set to 1.1.

The adjustment of the relation between the voltage waveform of mAs meter
or Dynalizer III is carried out by VR2(at Main CPU board) at exactly 100mAs.

Adjustment of VR2

RADIEQ CO., LTD.

(Time will be extended if it rotates to clockwise)

MX-600

18

Series

Mammography X-ray System

Service Manual

4.9.3 Density Adjustment in Bucky Device


1) Bucky device is attached in a radiography stand.
Radiography conditions
BUCKY device
Compression paddle
20mm acrylics : 4pcs
Cassette(with a non-exposed film)
2The switch "" of DENSITY is pushed and the adjustment mode is set to 1.0.

2 4 0.1 4
kV

mAs

1.01
1.

kgf

DENSITY

3) The switch of FILM & SCREEN is pushed and "1" is chosen.


1 Factory Setting, 2,3 User Setting.

1: KORDARK FLIM(MIN R) + KORDARK CASSETTE(MIN R)

4) It is set as 24kV by the switch "" or "" of kV. (acrylics 20mm)


It is set as density data by the switch "" or "" of mAs
Display becomes as it is shown in a figure.

2 4 0.1 4
kV

5)

mAs

1.01
1.

kgf

DENSITY

A switch "AEC" is pushed and data is saved.

6An X-ray exposure switch is pushed and X-ray is exposed.

The exposed film is developed and density is measured.


NOTE
1.30~ 1.40 of density is set when the unit released from the factory.

RADIEQ CO., LTD.

MX-600

19

Series

Mammography X-ray System

Service Manual

7 When density is darker, it is smaller in setting value, and setting value is


enlarged when density is lighter.
8The operation from 6) clause to 7) clause is repeated until density suits.

9It is set as 27kV by the switch "" or "" of kV. (acrylics 20mm)
It is set as density data by the switch "" or "" of mAs

2 7 0.1 4
kV

mAs

1.01
1.

kgf

DENSITY

10) The operation from 6) clause to 7) clause is repeated until density suits.
11) It exchanges for an acrylics 40mm board, and the tube voltage is set as 24kV
by the switch "" or "" of kV. (acrylics 40mm)
It is set as density data by the switch "" or "" of mAs

2 4 0.2 4
kV

mAs

1.01
1.

kgf

DENSITY

12The operation from 6) clause to 7) clause is repeated until density suits.


13It is set as 27kV by the switch "" or "" of kV. (acrylics 40mm)
It is set as density data by the switch "" or "" of mAs

2 7 0.2 8
kV

mAs

1.01
1.

kgf

DENSITY

14The operation from 6) clause to 7) clause is repeated until density suits.


15)

It is set as 32kV by the switch "" or "" of kV. (acrylics 40mm)


It is set as density data by the switch "" or "" of mAs

3 2 0.3 8
kV

mAs

1.01
1.

kgf

DENSITY

16) The operation from 6) clause to 7) clause is repeated until density suits.

17)

It is set as 35kV by the switch "" or "" of kV. (acrylics 40mm)


It is set as density data by the switch "" or "" of mAs

3 5 0.5 0
kV

RADIEQ CO., LTD.

mAs

1.01
1.

kgf

DENSITY

MX-600

20

Series

Mammography X-ray System

Service Manual

18) The operation from 6) clause to 7) clause is repeated until density suits.
19) It exchanges for an acrylics 60mm board, and the tube voltage is set as 27kV
by the switch "" or "" of kV.
It is set as density data by the switch "" or "" of mAs

2 7 0.4 8
kV

mAs

1.01
1.

kgf

DENSITY

20) The operation from 6) clause to 7) clause is repeated until density suits.
21)It is set as 30kV by the switch "" or "" of kV. (acrylics 60mm)
It is set as density data by the switch "" or "" of mAs

3 0 0.6 0
kV

mAs

1.01
1.

kgf

DENSITY

22) The operation from 6) clause to 7) clause is repeated until density suits.
23)It is set as 32kV by the switch "" or "" of kV. (acrylics 60mm)
It is set as density data by the switch "" or "" of mAs

3 2 0. 6 4
kV

mAs

1. 10

kgf
kgf

DENSITY

24) The operation from 6) clause to 7) clause is repeated until density suits.
25) It is set as 35kV by the switch "" or "" of kV. (acrylics 60mm)
It is set as density data by the switch "" or "" of mAs

3 5 0.7 4
kV

mAs

1.01
1.

kgf

DENSITY

26) It exchanges for an acrylics 80mm board, and the tube voltage is set as 32kV
by the switch "" or "" of kV.
It is set as density data by the switch "" or "" of mAs

3 2 0.9 4
kV

mAs

1.01
1.

kgf

DENSITY

27) The operation from 6) clause to 7) clause is repeated until density suits.

RADIEQ CO., LTD.

MX-600

21

Mammography X-ray System

Series

Service Manual

28)It is set as 35kV by the switch "" or "" of kV. (acrylics 80mm)
It is set as density data by the switch "" or "" of mAs

3 5 1.0 2
kV

mAs

1.01
1.

kgf

DENSITY

29) The operation from 6) clause to 7) clause is repeated until density suits.

RADIEQ CO., LTD.

MX-600

22

Series

Mammography X-ray System

Service Manual

4.10 Mode Setting


(1) Set up the adjustment mode
Push Switch"" of DENSITY.
Then a display is made into 2. by 7 segment LED on the left side of Density.
Display becomes as it is shown in a figure.

kV

mAs

2.

kgf

DENSITY

(2) Push Switch of kV, select modes. And Push Switch of mAs
Select data.
0Focal spot
1Filter conversion
2Autofilmchanger
3Rh Filter
4REL conversion

1:Manual 0:Auto
1:Manual 0:Auto
1:Install
0: Not Install
1:Rh
0:Al
1:Manual 0:Auto

(3) A switch "AEC" is pushed and data is saved.

4.10.1 Fiducial loading condition

kV
22
25
30
35
36
37
38
39

RADIEQ CO., LTD.

Large focus
mA
mAs(max)
70
600
86
600
72
600
60
600
60
500
60
400
60
320
60
320

kV
22
25
30
35

Small focus
mA
mAs(max)
13
100
15
100
12
100
10
100

MX-600

23

Series

Mammography X-ray System

Service Manual

4.11 Data sheet


4.11.1

Adjustment and measured value of X-ray tube voltage and current

Performed date
Condition
Large Focus
Set Value

Measured Value by mA adjusted


the instruments
Ref. Digital
Code
kV
mA

Specification
Deviation 4% of Set Value

KV

mA

KV

mA

22

70

21.2 22.8

67.2

72.8

25

85

24.0 26.0

81.6

88.4

30

72

28.8 31.2

69.1

74.9

35

60

33.6 36.4

57.6

62.4

39

60

37.4 40.5

57.6

62.4

Small Focus
Set Value

Measured Value by mA adjusted


the instruments
Ref. Digital
Code
kV
mA

Specification
Deviation 4% of Set Value

KV

mA

22

13

21.2 22.8

12.5

13.5

25

15

24.0 26.0

14.4

15.6

30

12

28.8 31.2

11.5

12.5

35

10

33.6 36.4

9.6

10.4

RADIEQ CO., LTD.

KV

mA

MX-600

24

Mammography X-ray System

Series

Service Manual

4.11.2 mAs measurement table


Performed date
Condition
Radiography method:

Set Value

BUCKY + compression paddle + 50mm acrylic

Measured Value by the instruments


L.F. 22kV / 70mA

mAs

S.F.

mAs

22kV / 13mA
mAs

Specification
Deviation 4% of Set
Value

1.00

0.90 1.10

1.30

1.17 1.43

1.60

1.44 1.76

2.00

1.80 2.20

3.20

2.88 3.52

4.00

3.60 4.40

5.00

4.50 5.50

6.40

5.76 7.04

8.00

7.20 8.80

10.0

9.00 11.00

20.0

18.00 22.00

50.0

45.00 55.00

100

90.00 110.00

600

RADIEQ CO., LTD.

540.00 660.00

MX-600

25

Mammography X-ray System

Series

Service Manual

4.11.3 The data sheet for mAs of AEC mode


Performed date
Condition
Radiography method

KV

24

Thick.
of
acrylic
(mm)
20

1
Ref.
of
mAs

Actual
mAs

BUCKY (L.F.)

Magnification(S.F.)

2
Density

Ref.
Actual
of
mAs
mA
s

3
Density

Re
Actual
f.
mAs
of
mA
s

Density

40
20

27

40
60

30

40
60
40

32

60
80
40

35

60
80

RADIEQ CO., LTD.

MX-600

26

Mammography X-ray System

Series

Service Manual

4.11.4 The data sheet for Density of AEC mode


Performed date
Condition
Radiography

Density Position

-9

-45

-8

-40

-7

-35

-6

-30

-5

-25

-4

-20

-3

-15

-2

-10

-1

-5

+5

+10

+15

+20

25

+30

+35

+40

+45

RADIEQ CO., LTD.

27kV

BUCKY + Comp. Paddle + 40mm acrylic

Adjustment
coefficient

Actual mAs

Density

Actual %

MX-600

27

Series

Mammography X-ray System

Service Manual

5. Mechanical Adjustment
5.1. Adjusting the Light Field
Do not attach any units to the mammography table (Bucky unit, etc).
Mount the 24 x 30 cm mask.
Turn on the manual switch for field lamp, then measure dimensions a, b, and
c as Fig.

a
Fig. 5

4)

Check the size of a, b, and c.


If the following conditions are met, adjustment is not necessary.
| a b| 6mm, c = 230 6mm.

RADIEQ CO., LTD.

MX-600

28

Series

Mammography X-ray System

Service Manual

5.2.

Adjusting X-ray exposure field

Preparation
Prepare the cassette shown by Fig. 6.
The edge marker of an aluminum board with a thickness of 0.5mm and a
width of 10mm is stuck on a cassette from edge at about 20mm place.
A marker is given with an aluminum board with a thickness of 0.5mm so
that right and left of a film may be known.

about 20mm
Radiography table

Cassette

Marker
Edge marker
Fig. 6 Test cassette
(a)
(b)
(c)
(d)

Set the cassette as shown in Fig. 6.


Mount the 24 x 30 mask
Exposure X-ray to the film ( large focus, 30kV 20mAs )
Develop the film, and measure any exposure field deviation.

Center of the film


X-ray
exposure field

c
a

Fig. 7
If | a b| 3mm and 1 c 4, adjustment is not required.
RADIEQ CO., LTD.

MX-600

29

Grove
Cassette
Radiography
Enlargement
Filter
Irradiation
Detector
Guide
Patient
Cassette
Fig.
2-9plate
2-19
2-18
2-17
2-16
2-15
2-14
2-13
2-12
2-11
2-10
2-8
2-7
2-6
2-5
2-3
2-1
Radiograp
Foot Switch
Model:
SHIELDPart No: Part
No.:
SPOT
18X24
MASK
222-U542-000
222-U540-000
222-U541-000
PADDLE
Model:
Model:
MAG.
PADDLE
PADDLE
19
18FACE
16 222-U550-000
18X24
24X30
Mode:24X30
15
14
17
table
fixing
selection
lever
radiography
holder
field
limiting
9 BUCKY
8
7
guide
plate
movement
attachment
hy
Table
table
X-RAY TUBE
/ HOUSING
222-U532-000
10X14cm
18X24cm
24X30cm
Part
No.:
222-U530-000
222-U533-000
222-U531-000
CAUTION:
THIS
PATENT
BE ALWAYS
X-ray
FieldNo.:
Size:
SID:GURD SHOULD
60 cm
222-U510-000
222-U511-000
Part
release
plate
section
table
rail
knob lever ATTACHED EXCEPT MAGNIFICATION MODE
X-ray
Field
Size:
18X24
24X30
Max.
comprssion
force:
240N cm
Mammography X-ray System
Series
(center
SID:
60 cm
WARNING
Way of insertion
and
both
This
X-ray unit may be dangerous
Service
Manual
DENSITY
mAs
k
V
Model: MAG. SPOT PADDLE to patient and operator unless safe
sides)
exposure factors and operating
INHIBIT

FILM
SPEED

MAGNIFY

BUKY

1676 SOUTH PIONEER ROAD


SALT LAKE CITY UT 84104-4298

RELEASE

instructions
are ROTATIONAL
observed. C-ARM
+

0.03mm Mo
Max. comprssion force: 240N

INHIBIT

MAGNIFY

13
12
11

TB4

LOW

FILM
Manufacturer
MEDIUM
SPEED
VARIAN
MEDICAL
SYSTEMS
HIGH

CN1

RELEASE

INHIBIT

MAGNIFY

CN2

DENSITY
BACK UP

222-U534-000
-

BUKY

MODEL NO. : RAD-85S / B110


SERIAL NO. :
/
FOCUS :
0.1
0.3

BACK
UP

LOW
MEDIUM
HIGH

AEC DETECTOR

Part No.:

BUKY

TB1

Radiographic Stand
(C-arm)

mAs

kV

MO

BACK UP

ROTOR

BACK
UP

AL

LOW
MEDIUM
HIGH

FILM
SPEED

BACK UP

ANODE

BACK
UP

MO

FILAMENT
(Cathode)

AL

MO

X-ray Tube

AL

MODEL NO. : MXR-200


SERIAL NO. :

TB1

Manufacturer
MedisonAcoma Co., Ltd.
22-8. 3-CHOME, HONGO
BUNKYO-KU TOKYO JAPAN

CN1

Manufacturer
MedisonAcoma Co., Ltd.
22-8. 3-CHOME, HONGO
BUNKYO-KU TOKYO JAPAN

CN7

CPU Board

22-8. 3-CHOME, HONGO


BUNKYO-KU TOKYO JAPAN

Foot switch to compres

CN7

CN6

Manufacturer
MedisonAcoma Co., Ltd.

X-ray exposure inhibit

CPU OUT

CPU INT.

Mask holder

Momentaly : 25 A
Longe time : 3.5 A

CN8

Fig. 8
CN5

AC 13V
AC 32V
AC 100V
DC+5V
DC+12V
DC-12V
DC+24V

Voltage : 230V
Phase : 1
Frequency : 50/60Hz

RADIEQ CO., LTD.

X-ray Exposure
HAND SWITCH

CN1

Control Panel

CN10

(h) After the lamp exposure field has been adjusted to within the conditions above,
tighten the four M3 nuts, exposure X-ray and check the results on the film.
(i) After checking the developed film, make sure that all nuts are securely
tightened, then remount the covers.

AC 230V

Circuit
Braker

Sub-power
circuit

INPUT POWER

Four M3 nuts

Foot switch for Magnet

CN2

CN8

CN10 CN9

CN1

INVERTER
Control Signal

AEC Detector Signal

Control Box
Caution : Do not touch this part

INPUT POWER

Fig. 2-2 Overall BLOCK DIAGRAM

RADIOGRAPHY STAND

CN1

22-8. 3-CHOME, HONGO


BUNKYO-KU TOKYO JAPAN

MODEL NO. : MXR-200


SERIAL NO. :

I/F Signal

CN3

Manufacturer
MedisonAcoma Co., Ltd.

Driving out

10

ROTOR BOARD

High Voltage
Generator

Momentaly : 25 A
Longe time : 3.5 A

I/F BOARD

Voltage : 230V
Phase : 1
Frequency : 50/60Hz

CN2

CN3

TB1

instructions are observed.

High Frequency
DC/AC Inverter

AC/DC Converter

RELEASE

ON/OFF Signal for vertical magnet

If the conditions above are not satisfied, then proceed with (f) below.
(e) Place
0.5mm Al the developed film on the cassette, then the cassette is set by the preDENSITY
mAs
kV
position.
the lamp
exposure
with the X-ray exposure field shown on the film.
WARNING (f) - Align
BOX - field
-CONTROL
+
+
+
MODEL
NO.
:
MXR-200
CONTROL
PANEL move the mask holder by loosening
This X-ray unit may be dangerous
(g) Using theSERIAL
adjusted
lamp field as a guide,
NO. :
to patient and operator unless safe
MODEL NO. : MXR-200
exposure factors and operating
the fourINPUT
M3POWER
nuts fixing it, so
that
SERIAL
NO. : the conditions above are met.

MX-600

30