Académique Documents
Professionnel Documents
Culture Documents
2
HgbA1c_VIITurbo_2.0_7-9-12.doc ..................................................................................................................................2
1 of 21
II. POLICY/SCOPE
This is intended for the China Basin Chemistry section of the Clinical Laboratories and
intended for testing by licensed Clinical Laboratory Scientists and Clinical Laboratory
staff.
III. SPECIMEN REQUIRMENTS
a. Blood is collected in a lavender top tube (EDTA) and refrigerated at 2-8C.
Whole blood is stable 7 days at 2-8 C or 24 hours at room temperature (15 30 C).
Lipemia up to a level of 6000 mg/dL of triglycerides does not interfere.
Icterus up to a level of 20 mg/dL does not interfere.
Hemolysis of the sample is not relevant, as whole blood is hemolyzed in the course
of the analysis.
b. Acceptable container sizes are 5 mL, 7 mL and 10 mL
c. Samples with volume < 2.0 mL (or height less than 25 mm), or clotted samples, require
pre-dilution before being placed on the VARIANT II TURBO.
d. Allow sample tubes to reach room temperature (1530 C) before performing the assay.
2 of 21
3 of 21
QUALITY CONTROL
a. BioRad LYPHOCHEK Diabetes Control Levels 1 and 2 (Cat. No. 740), stored at 2-8C,
is stable until the manufacturers expiration date. Reconstitute each vial with 0.5 mL of
CLRW. Reconstituted controls are stable for 7 days at 2-8o C.
b. Each run should be bracketed by a set of controls, one at the beginning and end of each
run.
VIII.
PROCEDURE
a. Refer to Appendix B BioRad Variant II Turbo Maintenance for instructions on daily,
monthly and as needed instrument maintenance.
b. VARIANT II TURBO requires a warm-up, from the inactive state, once per day prior to
beginning a procedure. Cartridge temperature must be within specifications before
starting the days run.
c. Allow patient samples to reach room temperature before loading onto the VARIANT II
TURBO. No sample preparation is required. Mixing the tubes prior to loading is not
necessary.
d. Two levels of quality control material should be included at the beginning and end of the
each patient run to check the performance of the assay.
i. Load a sample rack with microtube adapters, barcode labeled for Blank, C-DB1
(Level 1 Control) and C-DB2 (Level 2 Control).
ii. Fill a sample vial with a 1:300 dilution of control (either DB1, DB2 or whole
blood) and place into the Blank adapter.
iii. Make a 1:300 dilution, into sample vials, of both control levels using 5uL of
control to 1.5 mL of Wash/Diluent. Place controls into corresponding barcode
labeled adapters.
4 of 21
iv. Small volume patients (less than 2.0 mL or 25 mm sample height) are also
diluted 1:300 in sample vials, with Wash/Diluent, and place into non-barcoded
microvial adapters labeled with patient ID before being placed into the sample
rack.
Sample #
1
2
3
4 to N*
N+1
N+2
Sample
BLANK (QC sample)
C-DB1 (QC Level 1)
C-DB2 (QC Level 2)
Patient Samples
C-DB1 (QC Level 1)
C-DB2 (QC Level 2)
N+3
5 of 21
CRITERIA
ACTION
OK to report
Outside range
OK to report
3.5 19.0 %
OK to report
Report as <3.5%
OK to report
Correctly identified
OK to report
Repeat analysis
OK to report
Inconsistent
Repeat analysis
OK to report
Repeat analysis
Properly constructed
(i.e. stable, not drifting)
OK to report
Repeat analysis
OK to report
Quality Control
HbA1c
Reportable
Range
All normal
peaks are
present
A1c and Ao
peaks
A1c and Ao
retention times
Baseline
HbF (Fetal
Hemoglobin)
6 of 21
ITEM
CRITERIA
ACTION
OK to report
Heterozygous
hemoglobins E,
D, S, and C
P3 or P4 Peak
Variant and/or C
windows
Unknown
peak following
Ao peak
OK to report
OK to report
P4 peak 10%
OK to report
7 of 21
b. All samples that were not flagged by CDM can be released using the OEM function in
Sunquest after the tech has reviewed the chromatograms.
c. For results on samples that were prediluted, did not have a barcode read or flagged by
CDM, the OEH or MEH function should be used in Sunquest to report these values.
d. The reportable range for VARIANT II TURBO Hemoglobin A1c is 3.5 to 19.0%.
e. Results greater than 19.0% are reported as >19.0%.
f.
X. EXPECTED VALUES
The non-diabetic reference range for VARIANT II TURBO Hemoglobin A1c is 4.3 to 5.6%.
All hemoglobin A1c results will automatically be appended with the following table:
_______________________________________
HbA1c cutoffs for diagnosing diabetes:
4.3% - 5.6% = normal
5.7% - 6.4% = increased risk for diabetes
>6.4% = diabetes
___________________________________________
HbA1c goals in treatment of diabetes:
Ages 0-6 years: 7.6% - 8.4%
Ages 6-12 years: <8%
Ages 13-19 years: <7.5%
Adults: <7%
8 of 21
PERFORMANCE
a. SPECIFICITY/INTERFERENCES
b. PRECISION
The precision of the VARIANT II TURBO HbA1c Kit - 2.0 was evaluated in a study
based on the Clinical and Laboratory Standards Institute (CLSI) EP5-A2 guideline
Evaluation of Precision Performance of Quantitative Measurement Methods. The study
design was modified to include two different VARIANT II TURBO instruments at each
of three different laboratories, for a total of six instruments. The same set of normal and
diabetic samples were run in duplicate, in each of 2 runs per day, for 10 days, on each
instrument, using a single kit lot. A single calibration was performed on each instrument,
during the first run of the study. The results of the precision study are summarized in the
table below:
Normal Patient
Diabetic Patient
Mean (% A1c)
5.6
11.4
Within-Run (% CV)
0.78
0.39
Between-Day (%CV)
0.66
0.69
Between-Run (% CV)
0.53
0.45
Within-Device (% CV)
1.15
0.91
c. ACCURACY
The VARIANT II TURBO HbA1c Kit 2.0 was compared to the previous VARIANT II
TURBO Hemoglobin A1c Program. 40 whole blood patient samples were run using the
VARIANT II TURBO HbA1c Kit 2.0 and the previous VARIANT II TURBO
Hemoglobin A1c program. The range of values on the VARIANT II TURBO HbA1c Kit
2.0 was 4.8 12.5% HbA1c. The correlation results are as follows:
n = 40
slope = 1.0957
intercept = -0.7375
R2 = 0.9952
9 of 21
To demonstrate accuracy within the linear range of 3.5 19.0% HbA1c, the VARIANT
II TURBO HbA1c Kit 2.0 was compared to the VARIANT II Hemoglobin A1c
Program. 40 EDTA whole blood patient samples and 12 samples prepared by mixing
EDTA whole blood patient samples with Lyphochek Hemoglobin A1c Linearity Set
samples in various ratios were run using the VARIANT II TURBO HbA1c Kit 2.0 and
the VARIANT II Hemoglobin A1c Program. The range of values on the VARIANT II
TURBO HbA1c Kit 2.0 was 2.6 19.0% HbA1c. The correlation results are as
follows:
n = 52
slope = 0.9621
intercept = 0.4443
R2 = 0.994
d. LINEARITY
To demonstrate the linearity of the HBA1c measurement throughout the reportable range, a
normal and a diabetic HbA1c whole blood patient sample were used to prepare dilutions, and
the diluted samples were analyzed with the VARIANT II TURBO HbA1c Kit 2.0. The
linearity was assessed following the CLSI EP6-A guideline Evaluation of the Linearity of
Quantitative Measurement Procedures: A Statistical Approach. The results of the study
demonstrate HbA1c linearity from 3.5 19.0% within a maximum measured difference of
0.24% in this interval.
Sample Pool
1
2
3
4
5
6
7
8
9
XIII.
Predicted 1st
Order
2.82
5.26
7.28
9.66
11.20
13.70
15.59
17.61
19.56
Predicted 3rd
Order
2.58
5.34
7.45
9.79
11.26
13.64
15.47
17.52
19.63
Difference
0.24
-0.08
-0.17
-0.13
-0.06
0.06
0.12
0.09
-0.07
TECHNICAL NOTES
a. Samples from patients with hemolytic anemias will exhibit decreased glycosylated
hemoglobin values due to the shortened life span of the red cells. Samples from patients
with polycythemia or post-splenectomy may exhibit increased glycosylated hemoglobin
values due to a somewhat longer life span of the red cells.
b. Each analytical cartridge is good for about 2500 injections.
c. Column temperature is approx. 35 2 C.
d. -thalassemia trait, as indicated by increased HbA2 concentrations up to 10%, does not
interfere with the assay.
e. No significant interference from variant hemoglobin was observed at the following
concentrations: HbS 67%; HbC 72%; HbD 55%; HbE 41%.
10 of 21
XIV.
ALTERNATE METHODS
Hemoglobin A1c with eAG (Immunoturbidmetry) performed by Quest Diagnostics.
REFERENCES
a. VARIANT II TURBO HbA1c Kit 2.0 Instruction Manual: Bio-Rad Laboratories,
effective November 2012.
b. American Diabetes Association Home Page. http://www.diabetes.org (accessed April
2012).
c. Fowler, M.J. Microvascular and Macrovascular Complications of Diabetes. Clin.
Diabetes 2008, 26 (2), 77-82.
d. Whiting, D.R.; Guariguata, L.: Weil, C.; Shaw, J. IDF Diabetes Atlas: Global Estimates
of the Prevalence of Diabetes for 2011 and 2030. Diabetes Res. Clin. Pract. 2011, 94,
311-321.
e. Forsham, P. H. Diabetes Mellitus: A Rational Plan for Management. Postgrad. Med.
1982, 71, 139-154.
f.
Hollander, P. The Case for Tight Control in Diabetes. Postgrad. Med. 1984, 75, 80-87.
g. Baynes, J. W.; Bunn, H. F.; Goldstein, D.; Harris, M.; Martin, D. B.; Peterson, C.;
Winterhalter, K. NationalDiabetes Data Group: Report of the Expert Committee on
Glucosylated Hemoglobin. Diabetes Care 1984, 7, 602-606.
h. Nathan, D. M.; Singer, D. E.; Hurxthal, K.; Goodson, J. D. The Clinical Information
Value of the Glycosylated Hemoglobin Assay. N. Engl. J. Med. 1984, 310, 341-346.
i.
j.
Rohlfing, C. L.; Little, R. R.; Wiedmeyer, H. M.; England, J. D.; Madsen, R.; Harris, M.
I.; Flegal, K. M.; Eberhardt, M. S.; Goldstein, D. E. Use of GHb (HbA1c) in Screening
for Undiagnosed Diabetes in the U.S. Population. Diabetes Care 2000, 23, 187-191.
Hoelzel, W.; Weykamp, C.; Jeppsson, J. O.; Miedema, K.; Barr, J. R.; Goodall, I.;
Hoshino, T.; John, W. G.; Kobold, U.; Little, R.; Mosca, A.; Mauri, P.; Paroni, R.;
Susanto, F.; Takei, I.; Thienpont, L.; Umemoto, M.; Wiedmeyer, H. M.; IFCC Working
Group on HbA1c Standardization. IFCC Reference System for Measurement of
Hemoglobin A1c in Human Blood and the National Standardization Schemes in the
United States, Japan, and Sweden: A Method-Comparison Study. Clin. Chem. 2004, 50
(1), 166-174.
11 of 21
Bain, B.J. Haemoglobinopathy Diagnosis; Blackwell Science, Ltd.: Malden, MA, 2001;
pp 154 and 164.
12 of 21
13 of 21
3) When a new buffer lot is installed, a short system flush has to be performed. When new
buffers of the same lot are installed, a system flush is not necessary. PROCEED TO STEP
#6.
4) To perform a System Flush:
Under the Setup Icon,
a) Select the Test screen
b) Select REAGENTS
c) Select START SYSTEM FLUSH
d) The screen will prompt:
Choose one of the System Flush Types: Use the Short flush when changing reagents
for the same test. Use the extended flush when changing test.
e) Select SHORT
f) The screen will prompt:
The first step of the System Flush action is over.
Put reagent lines in the new buffer and continue
g) Follow the prompts and replace the new buffers
h) Select OK
i)
The purge valve will open automatically and the reagent lines will be flushed with the new
buffers at the rate of 6 mL/min for 20 minutes (an onscreen timer will count down remaining
time). The buffers do not go through the cartridge.
j)
k) Select DONE
l)
14 of 21
d) Grasp the PEEK Housing and turn it counterclockwise to remove it from the top (outlet cap)
of the holder. Remove the used analytical cartridge.
e) Remove a new prefilter from the package. The prefilter can be installed in either direction.
f) Push the prefilter firmly onto the end WITH THE O-RING of the Stainless Steel Prefilter
Adapter . Insert the adapter into the bottom (inlet) cap, prefilter facing upward.
g) Place the PEEK Housing into the inlet cap with the arrow pointing in the direction of flow
(bottom to top). Turn it clockwise until it is secured.
h) Remove the end caps from a new analytical cartridge. Position the cartridge with the arrow
pointing in the direction of flow (bottom to top). Push the cartridge firmly into the PEEK
Housing. Connect the PEEK Housing to the outlet cap by turning the outlet cap clockwise
until it is secured.
i)
j)
k) Place the assembled cartridge holder into the clip inside the cartridge thermal block.
l)
Close the cartridge thermal block cover, turning the lever clockwise to lock it.
m) After the startup actions are completed, return the instrument to Ready state.
Cartridge
Prefilter
Prefilter Adapter
15 of 21
ii) Change the In Use column entry from Yes to No for the current prefilter; a new line will
automatically be generated for the new prefilter.
iii) In the new prefilter entry, enter its lot number (if applicable) in the Lot # column, and type
500 in the Inj. Limit column.
iv) In the In Use column, select Yes.
6) Priming and calibration:
a) A new Analytical cartridge is primed twice before the calibration run.
b) Fill 2 microvials with 1 ml Whole Blood Primer in each and 3 vials with 1 ml of CLRW in
each.
c) Place the vials in barcoded microvial adapters in a sample rack as indicated below (using
BLANK adapters for the CLRW will not count against the number of injections on the
cartridge, nor will results print out):
1. PRIMER
2. PRIMER
3. BLANK (CLRW)
4. BLANK (CLRW)
5. BLANK (CLRW)
6. STOP tube or empty rack
d) Place Blank, Calibrator 1 and 2, and QC in barcoded microvial adapters in a second rack as
indicated below, and place this rack behind the primer rack on the VSS.
NOTE: Allow the instrument to return to READY after the prime is complete before
performing calibration, to eliminate any carryover from the primer:
1. BLANK (QC material or patient sample)
2. CALIBRATOR LEVEL 1
3. CALIBRATOR LEVEL 2
4. CONTROL LEVEL 1 (DB1)
5. CONTROL LEVEL 2 (DB2)
6. STOP tube or empty rack
e) Start the run. When the calibration is complete, check the Slope and Intercept of the
Calibrators on the printout. Slope and intercept are calibrator lot-specific and are updated via
the CD-ROM. If the calibration failed and delta factors are enabled in test setup, the slope
and/or intercept will flag as out of range and the analyzer will stop; if the calibration passed,
analysis will continue. Check current QC ranges for acceptability of results before running
patient samples.
16 of 21
17 of 21
18 of 21
MONTHLY MAINTENANCE
1) Exterior and interior surface cleaning
a) Use gauze moistened with CLRW to wipe the exterior surface of the system.
b) Wipe up any fluid inside the chromatography station. Tighten any leaking connections.
c) Clean the interior base with gauze moistened with CLRW.
2) Clean the conveyor belt
a) Under the Maintain Icon, select the Instruments screen
b) Under Execute commands, select CLEANING BELTS
c) Select START
d) When the belts start moving, use gauze moistened with CLRW to wipe the belts.
e) Select STOP when both belts are clean.
3) Clean Sample Needle and Barcode Reader
a) Under the Maintain Icon, select the Instruments screen
b) Under Sampler, select REPLACE NEEDLE
c) After the needle is repositioned, open the automatic sampler cover of the Sampling Station
(when the conveyor belt stops moving).
d) The screen will prompt:
Variant Front End door is open
Select OK
e) Clean the needle with a Kimwipe moistened with CLRW, removing debris. DO NOT USE
BLEACH. Be careful not to bend the needle.
f) Using Kimwipe or lens paper dampened with CLRW, gently wipe the barcode reader face.
BE CAREFUL NOT TO SCRATCH THE BARCODE READER FACE.
g) Replace the sampler cover. Power should be restored to the unit. A self-check is initiated.
h) Click on the Instrument #1 Variant II Turbo icon.
i)
j)
Select OK
19 of 21
n) Select YES
o) When process is complete, return the instrument to READY.
4) Clean the Dilution Well
a) Power off the Variant II TURBO Sampling Station (VSS) using the toggle switch on the right
side of the VSS.
b) Remove the needle cover door.
c) Push the needle gantry back, away from the dilution chamber.
d) Clean the dilution chamber using cotton swabs and CLRW. Remove particulate by filling the
tower with CLRW, then using cotton swabs to clean and soak up loose particulate.
e) Replace the needle cover door.
f) Power on VSS, wait for sampler to complete initialization and flush cycle.
g) Select OK to CDM communication error box, sampler resetting box and rack resetting box,
returning the system back to READY state.
AS NEEDED MAINTENANCE
1) Clean and decontaminate the probe and sampling fluid path
This procedure is performed as needed, generally for troubleshooting purposes. Perform the
decontamination procedure when any of the following symptoms occur:
Carryover from a hemoglobin variant sample generates an identified peak in the next
injection.
High total area counts occur across multiple samples.
Buildup is visible inside the injection valve tubing.
Blood is visible in the dilution chamber.
NOTE: The prefilter must be discarded after the decontamination procedure is
completed.
a) Remove the Analytical cartridge and replace it with the plastic PEEK dummy cartridge. If
the Analytical cartridge is not due for replacement, place green end caps on cartridge and
store for reinstallation after procedure completed. If the Analytical cartridge is due for
replacement, it may be discarded.
b) Disconnect cartridge-to-detector tubing from the cartridge holder. Install decontamination
tubing to the outlet of the cartridge holder.
c) Place outlet side of decontamination tubing into a beaker to collect waste from the
decontamination procedure.
20 of 21
Place the STOP adapter with an empty sample microvial into position six of the second
sample rack, and place both sample racks on the Sampling Station.
j)
Select START
iii) Select OK
k) When the status returns to READY, remove the plastic PEEK dummy cartridge. Clean the
Analytical cartridge and prefilter holders with a cotton swab moistened with DI water. The
prefilter must be discarded and replaced after the decontamination procedure is completed.
If Analytical cartridge injection limit has been reached, install a new Analytical cartridge and
perform required prime and calibration.
l)
Remove decontamination tubing from the cartridge holder. Discard beaker waste. Reconnect
the cartridge-to-detector tubing, securing both tubing end connections.
21 of 21