HumaCount 5L

Automated Hematology Analyzer Laser Technology


> Patented technology
> Cost efficient
> Reliable routine performance

CoreLab DX - Hematology

Movie Flyer

HumaCount 5L HumaCount

Diagnostics Worldwide
HumaCount 5L
Reliable Economical State of the Art

HumaCount 5L Cat. No. 16430

Automated hematology analyzer for


optical 5-part dierentiation of WBC
> 26 parameters, with patented optical 5-part WBC
differential measurement by laser scatter
> WBC, LYM, MON, NEU, BAS, EOS
(LYM%, MON%, NEU%, BAS%, EOS%)
> RBC, MCV, MCH, MCHC HGB, HCT, RDWc, PLT,
MPV, PDWc, PCT, RDWsd, PDWsd, P-LCC, P-LCR
> Flag for immature granulocytes and large blasts
Typical scatter diagram > Impedance method for standard parameters
> 60 samples/hour
> Closed and open tube sampling
> Auto-sampler with mixer and barcode reader (option)
> 110 l sample volume
25 l sample volume with small sample module
> Multi lingual user interface
> 600 x 800 colour LCD touch screen
> External Windows compatible printer via USB
> Memory for 100 000 results incl. histograms
> Built-in QC software

Touch screen with main program window > Closed reagent system
> LIMS ready (HL7)
HumaCount 5L
Optional:
HumaCount 5L
Case Study Book:
| Case study book

Background information on HC5L results,

showing their relation to blood smears,


diagnostic ags and clinic.
CSB

Auto-Sampler
Au
uto
to-S
Sampler Cat. N
No.
o. 16430/
16430/10 Cat No. 164309

HumaCount 5L

Reduce your hematology workload


> Capacity 100 samples (10 racks)
> Built-in mixer Specifications
> Barcode reader for sample identification
> Provides real walk-away capability Typical precision
> Optical sensor for differentiation of primary tubes Parameters CV Range
> Automatic tube type specific alignment WBC CV 3% 4.7 x 103/l WBC 38 x 103/l
> For closed standard primary sample tubes All 5 part parameters
(NEU, LYM, MON, EOS, BAS abs and %)
at 4.7 x 103/l WBC 38 x 10 3/l
NEU% CV 5% 87.4% NEU% 47 %
LYM% CV 8% 35.6% LYM% 15%
MON% CV 20% 16.3% MON% 5.3%
EOS% CV 25% 11% EOS% 1.5%
BAS% CV 40% 2.4% BAS% 1%
NEU CV 5% NEU% 30%
Small Sample Module Cat. No. 16430/12 LYM CV 8% LYM% 15%
MON CV 20% MON% 5%
Only 25 l sample volume required EOS CV 25% EOS% 1.5%
> Automated pre-dilution BAS CV 40% BAS% 1%
> Ideal for pediatric samples RBC CV 1.5% 2.5 x 106/l RBC 5.44 x 106/l
HGB CV 1.5% 78.5 g/l HGB 184g/l
HCT CV 2% 20 49 HCT%
HumaCount 5L System Reagents MCV CV 1% 65 fl MCV 105 fl
Size Cat. No. PLT CV 5% 100 x 10 3/l PLT 492 x 103/l
HC5L-Diluent 20 l 16430/20 MPV CV 5% 5.6fl MPV 11.3fl
Special designed for HumaCount 5L
HC5L-Lyse 5l 16430/30 Linearity
Cyanid-free lyse reagent
Linearity Coefficient of Range
HC5L-Diff 1l 16430/40 determination
Reagent for the differentiation of leukocytes WBC r^2 0.95 1 x 103/l WBC 100 x 10 3/l
CD Diff-Control 2 x 3 x 3 ml 3838 RBC r^2 0.95 0.4 x 106/l RBC 7.5 x 106/l
Multi-parameter control, HGB r^2 0.95 13 g/l HGB 227 g/l
Minimum open-vial stability: 2 weeks PLT r^2 0.95 10 x 103/l PLT 873 x 103/l
3 levels (low, normal, high),(distributed by HUMAN) Carry over < 0.5%
HumaCount 5L
Specifications

Sample volume Closed- and Open-mode: 110 l Fine calibration: With calibration material
(25 l with Small Sample Module) (polystyrene microsphere, 7m)
Sample type Human whole blood (K3-EDTA anticoagulant) Reagents HC5L-Diluent (20 liter)
Tube Identification By means of the front panel keyboard HC5L-Lyse (5 liter)
(enter ID) HC5L-Diff (1 liter)
By means of the barcode labels Sheath fluid Diluent
(manual and/or auto sampler) Quality Control 16- and 64-day Levey-Jennings charts,
Sampling method Ceramic shear valve separate QC database (6 level)
with 3 separated primary loops Flagging Morphological flags
Measured Interpretive flags (Diagnostic flags)
parameters CBC +5DIFF mode (26 parameters): Lab limits (normal ranges)
WBC, LYM, MON, NEU, EOS, BAS, LYM%, MON%, Reagents alert Pre-Alert-online reagent replacement
NEU%, EOS%, PLT, PCT, MPV, PDW, BAS%, RBC, System alert Instrument alerts, self check
HCT, MCV, HGB, MCH, MCHC, RWDcv, RDWsd, Calibration Manual and SW supported automatic mode
PDWcv, PDWsd, P-LCC, P-LCR Languages available UK-English, US-English, Hungarian, German,
Throughput 60 tests/hour Italian, Polish, Russian, Spanish, Turkish, French
Measurement Software upgrade Via USB
method Volumetric impedance change for WBC, RBC, PLT Data storage
Spectrophotometry for HGB capacity 100 000 records including flags,
Light scattering 4-diff measurement: scatter- and histograms
LYM, MON, NEU, EOS Data processing VIA C7 1.8GHz processor
Light scattering BASO measurement Data store Embedded XP
Aperture diameter WBC: 80 m Display 800 x 600 colour graphic LCD, portrait layout
RBC, PLT: 70 m External printing Via USB port, any Windows compatible printer
HGB measurement Light source: green LED with 540nm wavelength External keyboard Via PS/2 or USB
Detector: light to frequency converter Bar-Code reader Optional manual barcode reader via USB
Optical Built in barcode reader in the auto sampler
measurement Light source: semiconductor laser diode with Peripheral ports USB (2.0) 4pc., Ethernet, PS/2
650nm wavelength and 10mW Power requirements 110/230VAC; 47Hz to 63Hz
(Class IIIB laser module) Power Consumption maximum 400VA
Quartz flow cell with hydro-dynamic focusing Main fuse F 10A H 250V
Detector: fiber optic coupled PIN Si photodiodes Operating conditions 15 30C (59 98 F);
Internal safety interlock Maximum relative humidity 80 %
981643/2014-10 2014 Human GmbH

Auto-alignment Weigth Instrument 35 kg, autoloader 10 kg


system Optional horizontal and vertical calibration Size (h x d x w) Instrument 540 x 460 x 450 mm
of laser beam path. Autoloader 180 x 270 x 320 mm
Coarse calibration: With blood