Vous êtes sur la page 1sur 11

239-

Review articles

Extracorporeal membrane oxygenation (ECMO):


prolonged bedside cardiopulmonary bypass
JM Sinard and RH Bartlett University of Michigan Medical Center, Ann Arbor

Introduction ECMO specialists. A brief description of the


technique of ECMO support follows, in which
Extracorporeal membrane oxygenation (ECMO) similarities and differences between it and
is a unique form of cardiopulmonary bypass which operative cardiopulmonary bypass are noted.
has been in clinical use for 20 years. It has become
standard therapy in the management of neonatal
respiratory failure in the past decade, and is also Background
being used selectively in the support of
respiratory and cardiac failure in the paediatric The beginning of extracorporeal circulation dates
and adult populations. New applications are back to 1936 when John Gibbon first developed
constantly being evaluated, considerably a means by which haemodynamic and pulmonary

expanding the range of patient care that can be support could be delivered using external (ex-
provided in an intensive care unit. Depending tracorporeal) circulating devices. His first heart-
upon the vascular access and intent, extra- lung machine consisted of a roller pump, which
corporeal life support may be called ECMO, provided a mechanism for delivering flow, and
ECC02R (extracorporeal C02 removal), ECLA a vertically-oriented revolving cylinder, over

(extracorporeal lung assist) and CPS (cardiopul- which blood would drip.2 Oxygenation occurred
monary support) - all are essentially synonymous. through the direct exposure of red blood cells to
ECMO has brought the technology of cardiopul- ambient oxygen. Heparin, discovered by McLean
monary bypass to the bedside under the in 1917 and introduced into clinical practice in
management of intensive care physicians and 1938,3 permitted blood to be delivered through
artificial conduits without clotting. With the
advent of these basic components, the technique
Address for correspondence: Robert H Bartlett, Professor of
Surgery, University of Michigan Medical Center, Ann Arbor, of extracorporeal circulation began to evolve.
Michigan 48109, USA. The design of the original oxygenators was not
240

suitable for the prolonged support provided by did persist, however, with a new focus on cardiac
ECMO, since the direct exposure of blood cells and respiratory support in neonates.
to air or oxygen required by these oxygenators Most forms of neonatal pulmonary pathology
caused extensive protein denaturation after and subsequent respiratory insufficiency follow a
several hours. 4,5 Changes in oxygenator design, clinical course that lends itself readily to support
however, eventually occurred. In 1956, Clowes with ECMO. A common denominator seen in
introduced a plastic membrane into his neonatal pulmonary failure is a condition known
oxygenator in order to separate the blood and as persistent fetal circulation (PFC). PFC denotes
6
gas phases,6 and Kammermeyer, in 1957, an abnormal shunting of deoxygenated blood
discovered that gas exchange could occur more from the right to the left heart across a patent
effectively if a thin membrane of ductus arteriosus or foramen ovale. PFC may
dimethlysiloxane, commonly known as silicone result from deficiencies in alveolar surfactant,
rubber, was used instead of plastic? In the early hyper-reactivity of the pulmonary vasculature,
1960s, Kolobow incorporated both these features infection or from a toxic insult such as meconium
into his membrane lung, which permitted blood or blood aspiration. Unlike several adult
and oxygen to flow separately between thin layers pulmonary diseases, PFC is a reversible
of silastic-coated Dacron and fibre-glass screens.8 phenomenon. The goal, therefore, in using
The layers of silicone membranes were spirally ECMO in the neonatal population is to provide
wrapped to form a compact unit resembling the temporary respiratory support during the acute
appearance of modern membrane lungs. disease process until the PFC is resolved.
Eliminating the direct interface between blood Conventional mechanical ventilation is often
and oxygen significantly decreased protein inadequate in providing sufficient oxygenation
denaturation and permitted prolonged support during the acute phase. The high ventilatory
with an external oxygenator. Kolobow tested his associated pressures cause significant barotrauma
lung in lambs and was able to support them on with immediate and long-term pulmonary -
partial bypass for up to seven days.9 morbidity. ECMO obviates the need for such
The first successful case of prolonged high ventilatory pressures during the acute
extracorporeal support in humans was reported disease process and enables adequate
by Hill in 1972.1 Several other investigators oxygenation to occur without the attendant
began applying this new method of pulmonary barotrauma. When the acute process resolves,
support to moribund adults with respiratory ECMO can be discontinued and ventilation
failure.11,12 In order to assess the efficacy of resumed at reasonable pressure levels.
ECMO, a prospective randomized trial, funded In 1975, Bartlett et al. reported the first
by the National Institutes of Health, was neonatal survivor of acute respiratory
instituted to compare survival in adults with insufficiency from meconium aspiration in which
respiratory insufficiency treated by conventional ECMO was used. 16 With this encouraging report,
therapy to those treated with ECMO. 13 Although two prospective randomized trials were launched
the survival rate with ECMO was 9.5% compared which ultimately demonstrated the safety,
to 8.3% using traditional management, 14 the lack efficacy and benefit of ECMO support in neonatal
of a statistically significant difference between respiratory failure. 17,18
the two survival rates dampened enthusiasm to Presently, over 3300 patients have been treated
pursue further clinical trials in adults. A with ECMO with a neonatal survival rate of
retrospective assessment of these results 83% .19 Using previously established criteria,
suggested that the entrance criteria into the study survival without ECMO would be only about
were too stringent, as most patients who were 15%.2-22 The present applications of ECMO
enrolled already had permanent and irreversible continue to expand and include postoperative
lung injury, caused either by the disease process cardiac support in the paediatric population,
itself or by prolonged aggressive ventilatory perioperative support in cardiac transplantation
support 15 - the benefits of ECMO could not be and paediatric and adult respiratory support
realized in such circumstances. Continued from viral, bacterial, toxic or traumatic lung
research into the clinical application of ECMO injury.
241

ECMO circuitry . the common femoral or axillary arteries. The


largest-sized catheter that can be introduced into
The circuitry of ECMO represents a simplified the target vessel is selected. For neonates,
modification of that used during cardiopulmonary adequate flow rates (200-500cc/min) and
bypass. Certain obvious differences are readily pressures (100-250mmHg preoxygenator) are
apparent to the perfusionist who is unfamiliar best achieved with a 14 French (F) venous
with an ECMO circuit: perfusion and siphon catheter placed in the right internal jugular vein
cannulae do not exit the chest through a and a 10-12F arterial catheter in the right
sternotomy; there is no reservoir to transfuse or common carotid artery. The venous catheter has

haemodeplete a patient; no coronary suction several side-holes which improve the siphoned
device exists; blood filters are eliminated; flow within it. Using smaller catheters may cause
stagnant areas are avoided; and servoregulation insufficient venous drainage or prohibitively high
of pump flow is present. Management issues in circuit pressures when flow rates greater than
caring for the patient on bypass are likewise 300cc/min are required. VA bypass in ECMO is
different: the level of anticoagulation appears similar physiologically to operative cardiopulmo-
unusually low; variations in pump flow rates do nary bypass in that it provides both cardiac and
not universally improve tissue perfusion; pulmonary support for the patient. Increasing the
ventilatory support is titrated with and not flow rate through the ECMO circuit will improve
completely replaced by bypass; patients are whole-body perfusion, elevate the mean arterial
awake and moving about; and care is provided blood pressure and ensure haemodynamic
for several days rather than several hours. To stability while providing adequate gas exchange
present the significance of the differences for the blood.
between ECMO and operative cardiopulmonary In VV bypass, blood is both withdrawn from
bypass more clearly, each component of the and reinfused into the central venous system.
ECMO circuit will be addressed individually. The Blood is siphoned from the right atrium in a
circuitry discussed will apply specifically to manner similar to that in VA bypass but is
neonates, in whom the majority of the ECMO returned, after being oxygenated, into either the
cases have been performed. Similar comparisons femoral vein (or other suitably sized peripheral
also apply to the circuitry for paediatric and adult vein) or back into the right atrium by a newly
patients except that larger cannulae, tubing and designed catheter which has two channels.23 This
circuit components are used. double lumen venous catheter, recently
developed and tested at our facility, has a 14F
Cannulation and the two methods of ECMO outer diameter. Its cross-sectional area is divided
bypass disproportionately to enable the larger lumen to
The most striking initial difference between siphon blood from the catheter tip and the smaller
ECMO and bypass for cardiac surgery is that lumen to perfuse through side-holes directed into
cannulation for ECMO is performed extrathorac- the right ventricle. With VV bypass 15-50% of
ically and at the bedside in an intensive care unit. the central venous blood siphoned into the
Perfusion can be accomplished by two forms of ECMO circuit represents recirculated blood from
bypass: veno-arterial (VA) or veno-venous the venous perfusion cannula. The amount of
(VV). For either form of bypass, an incision is recirculation increases as the flow rate increases.
made over the target vessels to be cannulated. Despite recirculation, sufficient oxygenation of
In VA bypass, blood is typically siphoned from blood occurs to compensate for the failing lungs.
a cannula placed through the right internal The primary advantage of VV bypass is that it
jugular vein into the right atrium and is returned avoids ligation of a major artery; sacrificing such
to the patient by a catheter placed through the an artery can cause distal ischaemic symptoms if

right common carotid artery into the aortic arch insufficient collaterals are present. VV bypass is
after being oxygenated. Venous drainage may ideally suited for the patient who has isolated
also occur from cannulae placed in the common respiratory failure. Unlike VA bypass, VV bypass
femoral or external iliac veins; likewise, has no role in cardiac surgery. It does not divert
reinfusion can be accomplished through access in blood away from either the heart or the lung,
242

making most modern cardiac surgical procedures bypass this value actually represents the patients
impossible. Furthermore, it provides no mixed venous saturation. Variance in the mixed
haemodynamic support; adequate tissue venous saturation reflects changes in the
perfusion is entirely dependent upon the patients haematocrit, haemoglobin saturation, bypass flow
own cardiac function. rate, the patients own cardiac output or the rate
of tissue metabolism. In VV bypass the saturation
Priming the circuit value is of less importance as an absolute number
While a surgical team cannulates the patient, but is helpful when considered as a trending
ECMO specialists construct and prime the circuit. value. Changes in the saturation during VV
The individual steps involved in priming an bypass are likely to reflect similar changes
ECMO circuit closely parallel those in the observed in VA bypass, but the magnitude of
priming of a circuit for operative bypass. The these changes may be masked because of
circuit is first constructed using standard tubing, recirculation effects.
connectors and components with the drainage
and infusion tubing leading to a priming reservoir. Servoregulation
Carbon dioxide is flushed through the circuit to The bladder reservoir is the component of the
displace atmospheric oxygen and nitrogen. Any circuit that lies in the most dependent portion,
CO, remaining in the circuit after the prime is since siphoned flow to it is determined by its
completed will dissolve in blood more readily distance from the right atrium. The reservoir is
than will either oxygen or nitrogen, thereby made of silicone rubber which permits small
reducing the chance of inadvertent gas embolism. changes in its volume to occur. The reservoir
Once the circuit is filled with CO~, suction is servoregulates a double occlusion roller pump
applied to the gas phase of the membrane lung, which drives flow within the system. A pressure
removing much of the gas within the circuit. The sensor which is in continuous contact with the
ECMO specialist then adds, sequentially to the elastic wall monitors the bladder volume. When
circuit, a balanced crystalloid prime, albumin to the volume falls and the bladder walls begin to
coat the prosthetic surfaces and decrease platelet collapse, a pressure change is detected and power
and fibrinogen adherence and, finally, blood to to the roller pump is temporarily interrupted,
displace the crystalloid. A neonatal circuit stopping flow. Continuous suctioning by the roller
contains approximately 500cc of volume, which pump against inadequate venous return would
-- is considerably more than the blood volume of cause haemolysis, cavitation of dissolved oxygen
a neonate. Initiating bypass without a blood and damage to the endothelium at the side-holes
prime would cause significant haemodilution and of the venous catheter. A momentary
haemodynamic instability. Adjustments of the interruption of flow permits additional blood to
,
primed bloods pH and temperature frequently enter the bladder reservoir, replenish its volume
need to be made before connecting the patient and displace the pressure sensor of the reservoir
to the circuit. back to its normal position. Power returns to the
roller pump which restores flow in the circuit.
Blood saturation monitoring Adjustments in flow rate or catheter position can
After the vessels are cannulated, the catheters then be made to ensure continuous,
are connected to the polyvinylchloride (PVC) uninterrupted flow. The servoregulation between
tubing of the primed ECMO circuit. Figure 1 the bladder and the roller pump replaces the
depicts the individual components of the system open or bag blood reservoir used during bypass
diagrammatically. As blood is siphoned from the for cardiac surgery. Removing this component
right atrium, it drains past a fibre-optic probe eliminates any direct interface between blood and
into a bladder reservoir. The fibre-optic probe is ambient gases and reduces protein denaturation
attached to the circuit through a Tuohy-Borst during prolonged bypass. Maintaining a closed
connector and is in constant contact with the system further decreases the possibility of
flowing blood. This probe provides a continuous inadvertent air embolism.
on-line measurement of the saturation of blood Other centres employ different means of
being siphoned from the right atrium. In VA controlling ECMO bypass flow. Shiley (Irvine,
243

California) has introduced a system in which in a similar fashion. In Europe, a unique flow
changes in luminal pressure control pump flow: device called the Rhone-Poulenc pump (Paris,
a sensitive transducer on the venous end detects
France) does not require an external device to
pressure drops below a preset level; as the servoregulate its drive mechanism. This pump
pressure falls, the transducer effects a decrease contains three vertical posts around which a
in the pumps flow rate and therefore causes a distensible silastic tube connected in series with
rise in venous pressure towards a baseline value. the ECMO circuit is stretched. Flow is delivered
To prevent excessive pressurization within the when the rotating posts pinch off segments of
circuit, transducers placed on the high-pressure the tubing and deliver the contained blood. If
side of the pump can slow down the flow rate the rate of blood siphoning cannot keep pace
244

with the rate of attempted delivery, the amount past, when this tubing was used throughout the
of blood contained between each pinched off circuit, raceway rupture was a not uncommon
segment of raceway tubing will fall, decreasing event and the raceway tubing would be regularly
the effective blood flow. When venous pressure walked or advanced to avoid the continuous
rises and the siphon rate increases, blood flow stress placed by the roller head on the raceway.
within the circuit will increase.24 This problem is now averted with the use of more
durable PVC tubing (Supertygon@, Norton
Pumps Plastics, Akron, Ohio). The bypass circuitry used
In most ECMO centres, roller pumps rather during cardiac surgery does not need to
than centrifugal pumps are used to deliver flow. incorporate Supertygon@ tubing because the
Roller pumps offer several advantages, making duration of bypass is relatively short.
them more suitable for long-term bypass. Studies An ECMO circuit contains a relatively fixed
performed in our laboratory with different volume of blood, the silicone bladder being the
centrifugal pumps and our standard double only element of volume variability. Volume
occlusion roller pump demonstrate significantly cannot be transferred from a patient to a
less haemolysis with the latter pumps In component in an ECMO circuit as can be done
addition, the mechanics of a centrifugal pump with a reservoir in an operative bypass circuit.
are not as tightly responsive to servoregulation The patient is the only compliant component of
as those of a roller pump. When power to a an ECMO circuit, and excess volume in this
roller pump is interrupted, as occurs with closed system is seen as third-spaced oedema and
insufficient venous return, the heads of the pump intravascular volume expansion in the patient.
stop turning and flow in the circuit ceases Reduction in the total fluid requires diuresis,
immediately. The rotating head of a centrifugal haemofiltration or actually removing blood from
pump, on the other hand, will continue to spin the circuit.
for a brief time after servoregulation interrupts
its power. The negative pressure generated Oxygenators
(relative to atmospheric pressure) can lead to gas After leaving the roller pump, blood flows
cavitation, vessel endothelial damage and under positive pressure to the membrane lung.
- additional haemolysis from erythrocyte At the University of Michigan Medical Center
deformation. Other features of a roller pumps (UMMC) we use Sci-Med (Minneapolis,
simple design make it a preferred flow-delivering Minnesota) lungs routinely. Neonates require
device: it contains no valves, chambers or lung sizes ranging from 0.4 to 0.8m2, children
complex moving parts to malfunction; flow rate from 1.5 to 2.5m2 and adults from 3.5 to 4.5m2.
can be directly calculated from the RPM of the Selection of a properly sized membrane lung
pump head, obviating the need for an in-line flow requires knowledge of its blood and gas flow
probe; the chamber in contact with the roller properties. As the surface area of a membrane
pump is a component of the bypass circuit, i.e. lung increases, the pressure drop across the lung
the PVC tubing; and it does not require cleaning, becomes less for any given flow rate. The internal
replacement or adjustments. Disadvantages of resistance of each lung is a function of both its
roller pumps include tubing rupture in the surface area and the volume in which it is
raceway, gas embolization and over-pressurizing wrapped. If high flow rates are anticipated a
the circuit. Flow is calculated from RPM of a larger lung should be incorporated into the
roller pump; if the raceway tubing deforms during circuit. Excessively high circuit pressures
prolonged bypass, the calculated flow will be (>400mmHg preoxygenator) may cause circuit
inaccurate. rupture or premature lung failure. In addition to
internal resistance, each lung is characterized by
Tubing, raceway and compliance its ability to transfer gases at specific blood flow
The tubing used in the raceway of the roller rates. The rated flow of a lung describes the
pump is different from that in the remainder of maximal flow at which blood, entering a lung
the circuit. A neonatal circuit is made of PVC with a saturation of 75%, can exit the lung at
tubing with a wall thickness of 1/16 inch. In the 95% saturation; selection of an appropriately
245

sized lung, therefore, requires an estimate of the heat exchanger in an ECMO circuit lies in series
anticipated flow through it. Cardiac outputs in after the membrane lung. Exposure of the blood
resting patients range from 70-90cc/kg/min in tubing and circuit components to ambient
adults, 80-100cc/kg/min in children and 120- temperature causes rapid cooling of the patient,
170cc/kg/min in neonates. Lungs incorporated especially the neonate who lacks mature
into bypass circuits need to transfer both 02 and thermogenesis capabilities. A membrane lung
C02 adequately at the anticipated flow rates with its large surface area will cause the greatest
required by the patient. temperature drop in the circuit. High gas flow
Oxygen, carbogen or a mixture of the two is rates through the membrane lung will further
ventilated through the membrane lung in a accentuate this temperature drop. Placing the
counter-current fashion at a rate sufficient to heat exchanger immediately after the membrane
maintain the PC02 below
postoxygenator lung will warm the blood just before it returns
40mmHg. Carbogen is commonly used in the to the patient. This design ensures that minimal
neonatal population. At lower blood flow rates heat loss will occur between the heat exchanger
the removal of C02 from the blood is more and the patient. The theoretical problem of
effective than is the transfer of oxygen to the generating oxygen bubbles by heating saturated
blood. A marked respiratory alkalosis can occur blood has not proved to be a significant
if pure oxygen is ventilated through the consideration.
membrane lung. When carbogen is added the The PVC tubing connecting the membrane lung
resulting PC02 will rise, but not by more than to the heat exchanger lies at the highest part of
40mmHg. When pure carbogen cannot reduce the bypass circuit, arching between these two
the PC02 of exiting blood sufficiently and pure components. This area serves as the last bubble
oxygen causes excessive C02 removal, oxygen trap within the circuit and helps to minimize the
and carbogen can be blended together in the risks of gas embolism.
ventilating gas to create a mixture that produces
a normal arterial PC02 Bridge
Several monitoring features are attached to From the heat exchanger, blood travels directly
components of the membrane lung. Pressure to the reinfusion cannula. A bridge of PVC tubing
transducers connected immediately before and connects the arterial and venous lines. Flow
after the oxygenator provide a continuous through the bridge is prevented by a tubing clamp
assessment of oxygenator function in terms of its which occludes its lumen. Periodic flushing of
blood flow characteristics. A high pressure drop the bridge by releasing the clamp avoids
across the membrane lung indicates impending prolonged periods of blood stagnation and
lung failure and a need for replacement. High reduces the incidence of clot formation. This
postoxygenator pressures warn of prohibitively bridge permitsrecirculation of blood within the
high blood flow rates through a narrow reinfusion ECMO circuit when the patient-connecting lines
catheter or of a kink within the system. When are clamped during cannulation and trials off
the rate of gas flow through a membrane lung is bypass.
high, over-pressurization of the gas phase and
subsequent gas embolization could occur.
Pressure pop-off valves are therefore included in Managing a patient on ECMO
the sweep gas line to guard against inadvertent
and excessive pressure build up. Typically, the care of a patient on bypass requires
two persons, a nurse and an ECMO specialist.
Heat exchangers Nurses generally perform bedside patient care,
After leaving the oxygenator, blood flows dressing changes and medicine acquisition.
through a separate heat exchanger, where it is ECMO specialists are trained nurses, respiratory
warmed to 37C. The heat exchanger used at technicians, perfusionists or physicians who have
UMMC is made by Sci-Med and is connected to completed an extensive course dealing with the
a heating unit made by Seabrook (Cincinnati, physiology and technology of prolonged
Ohio). Unlike the circuits for bypass surgery, the extracorporeal support.26 Their course work is
246

supplemented with hours of bedside instruction. provides both cardiac and pulmonary support.
ECMO specialists maintain circuit integrity, With neonates, full cardiac output is achieved
administer medication and blood products into with flows approaching 120-170cc/kg/min. At
the circuit and regulate the blood flow rate. All these levels the pulse contour of the arterial wave
interventions are made within the boundaries of form will be markedly dampened. The adequacy
prewritten orders provided by a physician. of the flow is measured physiologically by the
Numerous details are addressed by the ECMO patients mean arterial blood pressure, urine
specialist, the most time-consuming of which is output and mixed venous saturation. Mean
an accurate record of fluid balance. The bedside arterial blood pressure (MAP) can be affected
flow sheet details an hourly and cumulative fluid by administering volume to ensure adequate
balance, including the volume of blood sampled, vascular capacitance and by regulating pump
volume of medications given and fluid losses blood flow to act against a patients vascular
through the dressing, chest tubes and drains. This resistance. Correct selection of the proper
balance, along with daily weights, provides intervention depends upon the information
valuable information in assessing a patients available to the physician or ECMO specialist at
ability to wean from bypass. the patients bedside. Urine output is traditionally
Significant differences exist in the level of regarded as the best physiological parameter of
anticoagulation needed for ECMO compared to adequate perfusion. Unfortunately, it is not
bypass for cardiac surgery. The initial loading always reliable early in the patients course
dose of heparin is only 100U/kg. Thereafter, on bypass. Not infrequently, patients are
heparin is administered (usually 30-60U/kg/hr) placed on ECMO after a prolonged period
to maintain the activated clotting time (ACT) of hypotension, anoxia or both during which
between 180-200 seconds under normal attempts at more conventional management
circumstances. The ACT is measured hourly. have failed. Renal tubular damage occurs
Diligent attention to the urine output, volume of and can cause a low urine output despite
plasma transfused and rate of heparin infusion apparently adequate flow rates and mean arterial
permits very precise control of the ACT. The pressures.
infinite ACTs maintained during operative car- Mixed venous saturation (SV02) at the UMMC
--

diopulmonary bypass are not needed while on is the parameter in VA bypass most commonly
ECMO and would greatly increase the incidence used to assess the adequacy of flow and total
of bleeding complications. ACTs for ECMO were oxygen delivery. In healthy adults the SV02 is
formerly maintained near 250 seconds, but as our approximately 75% at a resting state. Given an
experience has evolved we have progressively arterial saturation near 100%, oxygen delivery is
reduced ACTs to present levels without four times greater than oxygen consumption. If
developing any new clotting problems. bypass is able to achieve near 100% arterial
Occasionally, because of extraordinary clinical oxygen saturation in the patient, flow rates will
circumstances, ECMO bypass has been run with be adjusted to maintain an SV02 of
little or no heparinization for periods longer than approximately 75%. A fall in SV02 will reflect
48 hours. If flow rates are great enough, clotting either a decrease in haematocrit, arterial
problems do not usually occur. , haemoglobin saturation or total cardiac output
Platelets are administered as necessary to or an increase in tissue metabolism. In practice,
maintain a level greater than 100 000/mm3. several variables are assessed and treated when
Because of their adherence to the membranes, a low SV02 occurs: first, blood should be

platelets are infused into the circuit after the transfused if the haematocrit is less than 45%;
oxygenator. Elimination of filters on the infusion secondly, the function and adequacy of the size
side of the circuit helps to prevent the of the oxygenator must be addressed, especially
consumptive thrombocytopenia seen during if the postoxygenator haemoglobin saturation is
bypass. less than 100%; thirdly, the bypass flow rate may
The flow rates used during bypass depend upon need to be increased if the first two variables are
the type of bypass being employed and the optimized; and fourthly, sedation, paralysis, or
reasons for instituting bypass. VA bypass modest cooling of the patient should be instituted
247

if the patient is hypermetabolic. Complications


In VV bypass, management issues are slightly
different. The primary goal of VV bypass is to Bleeding is the most common complication seen
permit a reduction of the high pressures delivered in patients on ECMO. 27 Bleeding has been
by mechanical ventilators by oxygenating the reported from operative sites, mucous
central venous blood. Bypass flow rates are membranes, stomach and distal gastrointestinal
increased to effect an arterial haemoglobin tract, and miscellaneous locations within the
saturation greater than 90% if possible. Unlike cranium, pleura, pericardium or
VA bypass, VV bypass provides no direct retroperitoneum. Conservative techniques used
improvement in cardiac output. The saturation to manage bleeding complications include
of siphoned venous blood will not represent the maintaining the platelet count at above 150 000/
true SVO, because of recirculation effects, and mm3 and lowering the ACT levels. ACTs have
the adequacy of tissue perfusion is therefore been lowered into the 120-140 second range
assessed based on clinical and laboratory without apparent clotting problems. On rare
observations. Usual clinical measures of tissue occasions, if flow rates are high, heparinization
perfusion include urine output, time for capillary can be discontinued with caution. All surgical
refill, quality of peripheral pulses and blood sites should be re-explored, and other more
pressure, while comparable biochemical measures aggressive operative interventions may be
include pH, lactate levels, creatinine, etc. required. If bleeding is severe, ECMO bypass
Patients will retain the typical wave contour of may need to be temporarily or permanently
arterial blood pressure which will not be discontinued. Of all potential bleeding sites,
dampened as in VA bypass. Enhancing cardiac intracranial haemorrhage is most serious.
output in patients on VV bypass requires the Fortunately it occurs in only 14% of neonates
administration of inotropic and chronotropic with gestational ages greater than 35 weeks.28
agents. The function of the patients own heart Circuit-related problems are occurring less
is the sole determinant of cardiac output, and frequently as modifications in design and better-
changes in bypass flow will not directly alter tested equipment are incorporated. Technical
cardiac function or tissue perfusion. problems include oxygenator failure, roller pump
Airway management during bypass requires or heat exchanger malfunction, raceway or other
individualized attention based upon a patients tubing rupture and air leakage into the circuit.
specific disease process and the level of ECMO Most of these complications cause temporary
support provided. When full bypass flows are physiological instability while the circuit
reached, the amount of ventilator support can components are changed and rarely cause any
be reduced to rest settings. In the neonate these permanent morbidity.
generally include a peak inspiratory pressure of Other complications seen include tension
20cm H20, a positive end-expiratory pressure pneumothorax, pericardial tamponade, seizures
(PEEP) of 4cm H20, respiratory rate of 10-20 and organ failure; surprisingly, catheter-related
breaths per minute and an inspired Fio2 of 30%. sepsis is vanishingly rare.
Reducing airway pressures avoids barotrauma
and reduces the incidence of late development
&dquo;

of bronchopulmonary dysplasia. Other means of Weaning the patient off ECMO


airway management are possible and may include
extubation, low continuous positive airway The duration of ECMO support in the neonatal
pressure (CPAP) in patients with a bronchopul- population ranges from four to seven days
monary fistua, high PEEP or frequent pulmonary (average 126 hours). The longest length of time
lavages. spent on ECMO at UMMC as of January 1990
is 27 days for a child with severe viral respiratory
illness; he subsequently recovered and suffers no
apparent long-term sequelae.
After cannulation, the majority of ECMO
patients are allowed to stabilize for 24-48 hours
248

at a flow rate sufficient to provide adequate Summary


oxygenation. If patients receive large volumes of
fluid during their resuscitation prior to initiating The technology of ECMO support continues to
ECMO, aggressive diuresis with frusemide or evolve. New advances presently in development
mannitol is begun. Haemofiltration is used to include the manufacturing of paediatric- and
remove excess volume when the kidneys are adult-sized double lumen catheters, the use of
unresponsive to diuretics. heparin-coated tubing and the automation of
Signs of improvement are assessed daily while bedside ECMO control. As the applications of
on ECMO bypass. These include parenchymal ECMO branch deeper into the area of cardiac
clearing on serial chest X-rays, rising pulmonary transplantation and support, the interface
compliance, increasing end-expiratory CO, between the ECMO specialist and perfusionist
production measured in the airway, rising pO, will become less distinct. Both the ECMO
in the patients arterial blood gases (ABG) in specialist and perfusionist will provide a
addition to mobilization of third-spaced fluid, continuum of patient care, spanning preoperative
rising SVOz or MAP. With VA bypass the flow cannulation in the intensive care unit,
rate is progressively reduced to maintain MAP, intraoperative management during corrective
SV02 and ABGs at acceptable levels. When it cardiac surgery and postoperative cardiac
appears that a patient can support himself recovery back in the ICU. As bypass technology
haemodynamically and maintain adequate advances and its applications broaden, the
oxygenation with minimal ventilator support, a challenges faced by perfusionists will continue to
trial off bypass is attempted. This necessitates grow.
clamping the patient-connecting lines and
removing the clamp on the bridge to allow
recirculation. Ventilatory support needs to be References ...1

increased from rest settings to those sufficient


to provide adequate pulmonary gas exchange. 1 Bartlett RH, Harken DE. Instrumentation
Decannulation can proceed if the patient remains for cardiopulmonary bypass: past, present
stable for 45-60 minutes. and future. Med Instrum 1976; 10: 119-24.
In theory, weaning off VV bypass is simpler 2 Gibbon JH Jr. Artificial maintenance of
since VV bypass provides no cardiac support. circulation during experimental occlusion of
- During a course on VV bypass the ECMO flow pulmonary artery. Arch Surg 1937; 34: 1105-
rate is progressively reduced as pulmonary 31.
function improves. This improvement is 3 McLean J. The discovery of heparin.
evidenced by a rising arterial haemoglobin Circulation 1959; 19: 75-78.
saturation seen on a transcutaneous sensor. The 4 Lee WH Jr, Krumhar D, Fonkalsrud EW et
bypass flow rate can then be decreased to al. Denaturation of plasma proteins as a cause
maintain the arterial haemoglobin saturation of morbidity and death after intracardiac
greater than a suitable minimum value, usually operations. Surgery 1961; 50: 29-39.
90%. When the flow rate reaches a low level 5 Dobell ARC, Mitri M, Galva R et al.
(30cc/kg/min), a trial off bypass can be attempted. Biological evaluation of blood after prolonged
This trial involves removing the ventilating gas recirculation through film and membrane
line from the membrane lung and capping off oxygenators. Ann Surg 1965; 161: 617-22.
the gas inlet and outlet ports. Blood is still 6 Clowes GHA Jr, Hopkins AL, Neville WE.
siphoned from and returned to the patient, but An artificial lung dependent upon diffusion
no gas exchange occurs. The bridge remains of oxygen and carbon dioxide through plastic
clamped. The settings on the ventilator are membranes. J Thorac Surg 1956; 32: 630-37.
increased to levels commensurate with 7 Kammermeyer K. Silicone rubber as a
physiological support. A successful trial off selective barrier. Ind Eng Chem 1957; 49:
bypass for 30-45 minutes demonstrates sufficient 1685.
lung recovery to permit decannulation. 8 Kolobow T, Bowman RL. Construction and
evaluation of an alveolar membrane artificial
249

Trans Am Soc Artif Intern Organs


heart-lung. persistent pulmonary hypertension of the
9: 238-43.
1963; newborn: a prospective randomized study.
9 Kolobow T, Zapol W, Sigman RL et al. Pediatrics 1989; 84: 957-63.
Partialcardiopulmonary bypass lasting up to 19 Extracorporeal Life Support Organization.
seven days in adult lambs with membrane
Extracorporeal Life Support Organization
lung blood oxygenation. J Thorac Cardiovasc (ELSO) Registry. Ann Arbor, Michigan:
Surg 1970; 60: 781-88. University of Michigan (as of December
10 Hill JD, OBrien TG, Murray JJ et al. 1989).
Prolonged extracorporeal oxygenation for 20 Bartlett RH, Andrews AF, Toomasian JM et
acute post-traumatic respiratory failure al. Extracorporeal membrane oxygenation for
(shock-lung syndrome): use of the Bramson newborn respiratory failure: forty-five cases.
membrane lung. N Engl J Med 1972; 286: Surgery 1982; 92: 425-33.
629-34. 21 Zwischenberger JB, Bartlett RH.
11 Zapol W, Pontoppidan H, McCullough N et Extracorporeal circulation for respiratory or
al. Clinical membrane lung support for acute cardiac failure. In: Civetta, Taylor, Kirby eds,
respiratory insufficiency. Trans Am Soc Artif Critical Care, Philadelphia: JB Lippincott,
Intern Organs 1972; 18: 553-60. 1988: 1629-37.
12 Bartlett RH, Gazzaniga AB, Fong SW et al. 22 Krummel TM, Greenfield LJ, Kirkpatrick
Prolonged extracorporeal cardiopulmonary et al. Alveolar-arterial oxygen gradients
BV
support in man. J Thorac Cardiovasc Surg versus the neonatal pulmonary insufficiency
1974; 68: 918-32. index for prediction of mortality in ECMO
13 Zapol WM, Snider MT, Hill JD. candidates. J Pediatr Surg 1984; 19: 380-84.
Extracorporeal membrane oxygenation in 23 Anderson HL, Otsu T, Chapman R, Bartlett
severe acute respiratory failure: a randomized RH. Venovenous extracorporeal life support
JAMA 1979; 242: 2193-96.
prospective study. in neonates using a double lumen catheter.
14 Extracorporeal support for respiratory Trans Am Soc Artif Intern Organs 1989; (in
insufficiency, a collaborative study. Prepared press).
in response to RFP-NHLI-73-20. Bethesda, 24 Durandy Y, Chevalier JY, Lecompte Y.
MD, United States Department of Health, Veno-venous extracorporeal lung support:
Public Health Service, National Institutes of initial experience in paediatric patients. In:
Health, December 1979. Gille JP ed, Neonatal and adult respiratory
15 Hirschl RB. Bartlett RH. Extracorporeal failure. Mechanisms and treatment. Paris:
membrane oxygenation support in Elsevier, 1989: 159-72.
cardiorespiratory failure. Adv Surg 1987; 21: 25 Montoya P, Merz S, Bartlett RH.
189-212. Unpublished data.
16 Bartlett RH, Gazzaniga AB, Jefferies MRet 26 Chapman RA, Toomasian JM, Bartlett RH.
al. Extracorporeal membrane oxygenation Extracorporeal membrane oxygenation
(ECMO) cardiopulmonary support in infancy. technical specialist manual, ninth edition,
Trans Am Soc Artif Intern Organs 1976; 22: 1988.
80-93. 27 Bartlett RH, Gazzaniga AB, Toomasian JM
17 Bartlett RH, Roloff DW, Cornell RG et al. et al. Extracorporeal membrane oxygenation
Extracorporeal circulation in neonatal (ECMO) in neonatal respiratory failure: 100
respiratory failure: a prospective randomized cases. Ann Surg 1986; 204: 236-45.
study. Pediatrics 1985; 76: 479-87. 28 Cilley RE, Zwischenberger JB, Andrews AF
18 ORourke PP, Krone R, Vacanti J et al. et al. Intracranial hemorrhage during
Extracorporeal membrane oxygenation and extracorporeal membrane oxygenation in
conventional medical therapy in neonates with neonates. Pediatrics 1986; 78: 699.

Vous aimerez peut-être aussi