Académique Documents
Professionnel Documents
Culture Documents
and
Assembly/Schematic Drawings
INDEX
The IMMULITE Display Panel buttons are explained in the table below:
Button Function
ALARM MUTE In the Run mode, it silences the audible alarm.
In Diagnostic modes, it is used to pause a motor movement. It
is also used in conjunction with the GO button to cancel a
diagnostic program.
PAUSE Temporarily stops pipetting and Load Chain movements only.
Shuts down the IMMULITE when pressed repeatedly.
Note: Using PAUSE to shut down the instrument ensures all parts
are in the home position.
GO In the Run mode, it activates the initialization (start-up)
procedure and the accompanying instructional messages
appearing on the Display Panel.
It also activates a reading of the Reagent Wedge barcodes if the
instrument is in pause mode when GO is pressed.
In Diagnostic modes, it activates the GO command for the
loaded diagnostic program. It is also used in conjunction with
the ALARM MUTE button to cancel a diagnostic program.
TEST UNIT
WASTE EXIT
Note: Different assay kits are required for each mode. Standard (Non-Turbo) kits cannot be run
while in Turbo mode and Turbo kits cannot be run while in Standard (Non-Turbo) mode.
Symbol Definition
Consult instructions for use
Do not re-use
Batch code
Catalog number
Serial number
Sterile
Biological risk
Non-sterile
Do not re-sterilize
Keep dry
Temperature limitation
Control
Negative control
Positive control
Date of manufacture
Manufactured by
European Community
Representative
Upon receipt of the IMMULITE Analyzer system, visually inspect the exterior of the shipping container
for damage. If there is any external damage to the container, notify your shipping department and contact
Siemens Diagnostics Technical Support.
2.1 Requirements
Electrical
Use of a dedicated, grounded service for the IMMULITE Analyzer, computer, monitor, and printer is
strongly recommended.
Use of an Uninterruptible Power Supply (UPS) or line conditioner is also strongly recommended.
Do not place the system near centrifuges, ultrasound or X-ray machinery, NMR scanners, or other
sources of magnetic fields.
Space
The IMMULITE must be placed on a flat, clean, horizontal surface.
The instrument requires a bench space of 75 inches (191 cm) across by 30 inches (76 cm) deep, with
any overhead cabinets at least 22 inches (56 cm) above the bench.
Environment
Ambient temperature should be between 65F - 86F (18C - 30C), Humidity 20-80% (no
condensation).
Do not place the IMMULITE system in direct sunlight.
Do not install the IMMULITE near heating or air conditioning vents or equipment.
7. Plug the optional Uninterruptible Power Supply (UPS) or line conditioner into the power outlet and
turn it on. NOTE: A dedicated electric power line is strongly recommended to reduce the effects of
electrical noise produced by other instruments.
8. Remove all the shipping tape from the IMMULITE, including the pipettor and its base, the sample
collection tray, the reagent tray, and the main carousel.
9. Install the gas struts by pressing the end of the strut onto the ball studs, then pressing in firmly to lock
them.
10. Install the syringes, with the large syringe on the left, and the small one on the right. Insert each
syringe into the luer lock fitting, and twist them to tighten. Thread in the thumbscrews into the
bottom of the plungers, and tighten.
11. Loosen the screws that hold down the cover for the printed circuit board compartment. Lift the cover,
and gently lift and reseat each of the four (4) printed circuit boards.
12. Turn the IMMULITE power switch to "on. The systems power is brought up safely in stages.
(Initially, the chirp of the +/- 5v & +/- 12v logic will first be heard.) After approximately 15
seconds, the syringes should "home" - that is, they should move to the top. If already there, they will
both move up a bit more to confirm home. This is desired, and confirms initiation of full 24v system
supply. If this does not happen, there are likely troubles of syringe module connections, or the 24v
supply itself.
13. Turn the power on to the computer, monitor, and printer.
14. Once the Start Menu appears on the monitor screen, go to the DOS prompt at CIRRUS/IMM: Type
dir*.cnf and return. Only one file should be listed. It should have a format of IMMXXX.cnf
(where XXX is the serial number of the instrument) and a file size of 488 bytes. Delete any other .cnf
files that may be present.
15. Obtain a formatted, high-density 3.5" floppy disk and perform a backup from the Start Menu.
16. Install the waste tubing and bottle using supplied straight or elbow fitting, while keeping the tubing
straight and to avoid kinks. Route this tubing as needed.
17. Spike a new bottle of substrate and fill the reservoir with substrate up to the fill line using the manual
prime button.
18. Enter diagnostics and load the program SOLE2S.
19. Run SOLE2S to prime the substrate pump and associated tubing.
20. Allow a minimum of 45-50 minutes to ensure full temperature stability of the system before
beginning to run any water and/or chemistry testing. Confirm proper and stable temperatures, using
of Section 3.8 Instrument Temperatures for guidance.
21. Before any initial chemistry testing begins, the proposed water supply source intended for the
IMMULITE must first be tested and properly qualified, as described on the next page.
Use the Water Test Kit (catalog number: LKWT) in conjunction with the WATERTST diagnostic
program to perform the water test.
WATERTST
The WATERTST diagnostic program tests the water from any source, independent of the internal fluidics
of the IMMULITE. Water should be tested before priming water from a new source through the
IMMULITE System, or when contamination is suspected. The procedure for testing the water supply
using the WATERTST program appears below:
1. Before starting, ensure the IMMULITE has been powered on for at least 45 minutes and at a stable
operating temperature.
2. Make sure the printer is on-line.
3. Select two empty Water-Test Test Units.
4. Manually pipette 50 L of water directly from the source or from the water in the water bottle into the
first Test Unit only.
5. Load both Test Units onto the Load Chain. Place the Test Unit containing the water at the beginning
of the Load Chain (so it enters the Barcode Reader first).
6. From the Start-up menu, choose DIAGNOSTICS and then the sub-menu choice of DIAGNOSTICS.
Note: The IMMULITE must be Idle before performing step 6.
7. Select WATERTST and press [Enter].
8. When the loading is complete, press [Enter] again.
9. Press GO (on the IMMULITE) when instructed to do so.
10. Substrate is dispensed into the Test Units. The Test Units are then moved into the Luminometer and
incubated for 10 minutes.
11. Once the Test Units start moving, press [Esc] and choose EXIT TO DOS.
12. At the DOS prompt, type water and press [Enter].
The following screen appears:
If the cps is too high or too low, perform the test again to confirm the results.
If the cps is consistently outside the 4,500 cps to 9,000 cps range, find a new water source and test
this new source.
If the second test unit (substrate only) is > 9,000 cps, the substrate system may be contaminated or the
test units used for testing may not be clean. **The Water result is then not valid, as it is being falsely
biased (high) by the unacceptably high level of this baseline substrate only reading
Once an acceptable water source is obtained, it can be put into the rinsed water bottle, and the probe-wash
solution can be mixed. Once primed into the system using the PRIME common diagnostic program,
clear all air bubbles in the syringes. The instrument is now ready for chemistry testing.
3.1 Introduction
The IMMULITE System is a random access immunoassay Analyzer, built around a proprietary means of
efficiently and rapidly washing the bead solid phase. The 0.25-inch polystyrene bead is captured in a Test
Unit that serves as the vessel for all incubations, washes, and signal development. After incubation of the
sample with alkaline phosphatase-labeled reagent, rapid separation and efficient washing of the Test Unit
and bead are achieved by high-speed spinning of the Test Unit on its vertical axis. The fluid contents are
completely transferred to the coaxial sump chamber of the Test Unit. Multiple washes are accomplished
within seconds, allowing uniform, sequential processing of the Test Units and leaving the bead with no
residual unbound label. The bound label is then quantified, using the dioxetane substrate to produce light.
Light emission from the chemiluminescent substrate, reacting with the bound label, is proportional to the
amount of analyte present in the sample.
Note: The materials needed to operate the IMMULITE are included in the IMMULITE Test Kits.
2. In normal testing, the operator then loads up to five barcoded Test Units after each sample, in any
order, for the tests desired on that sample. (See figure on page 1-4).
Additional samples and Test Units can be added at any time, and Stats can be placed at the
beginning of the Load Platform.
If on-board dilutions are also being used, with the correct diluent wedge resident on the reagent
carousel, the order of Load Platform tube placement becomes: sample cup, dilution cup, 1 empty
space, followed by up to four random test units.
3. The operator presses GO (on the IMMULITE Display Panel) and the Test Units are conveyed into
the Analyzer for barcode identification and then moved onto the Main Incubation Carousel.
4. The pipettor adds sample and alkaline phosphatase-labeled reagent. (The Reagent Carousel
accommodates up to twelve resident assays.)
Note: Reagent changes can be made during instrument operation by pressing the PAUSE button,
which stops the Load Chain and the Pipettor. Reagent Carousel will not go into pause if sequential
assays need to be pipetted. Normal operation resumes when the GO button is pressed.
The IMMULITE's attenuation filter restricts the number of photons entering the PMT. This is used to
prevent high light signals from saturating the PMT. Any readings above saturation are not accurate,
so the IMMULITE uses the attenuated light signal and compensates for the attenuation via software.
2. For each sample, the IMMULITE takes a one-second reading dark count in the closed position and a
one-second reading decision count in the attenuated position.
Note: To view the dark counts while in the run mode, select System Status and View Temps/Dark
Counts.
If the one-second attenuated reading is less than 10,240 cps, the attenuator disk moves to the open
position; otherwise, the attenuator disk remains in the attenuated position while the readings are
taken.
3. The IMMULITE takes twelve one-second readings and returns the attenuator disk to the closed
position.
4. Dark counts are calculated by averaging the last ten reads. This average is subtracted from each of
the twelve readings.
5. To ensure that the final reading is representative, the highest and lowest of the twelve individual
readings are dropped.
6. All PMTs are slightly different in sensitivity. Therefore each PMT has a unique PMT Multiplier
factor. This PMT Multiplier number is multiplied by the cps reading to get the final cps reading.
7. If the counts were measured attenuated, the trimmed mean of the ten individual readings is multiplied
by the instrument-specific attenuation factor to yield the total unattenuated counts per second (cps).
CPS (unattenuated)= CPS (attenuated) x attenuation factor
8. To determine analyte concentrations, the IMMULITE software refers to lot-specific master curve
parameters, which were entered into the system via the kit barcodes.
1 2 3 4 5 6 7
PMT Pre-scaler Mean of last High and Mean X 10 (for Mean X Mean X
Output Output* 10 dark Low count Pre-Scaler*) PMT Attenuation
(CPS/10) count reads removed Multiplier Factor**
(20 CPS then mean Factor** (only if
used here) calculated (.90 used counts were
subtracted here) taken
from each attenuated)
count FINAL CPS
8000 800 780 780
8100 810 790 790
8200 820 800 800
8300 830 810 810
8400 840 820 820
8500 850 830 830
8600 860 840 840
8700 870 850 850
8800 880 860 860
8900 890 870 870
9000 900 880 880
9100 910 890 890
8350 7515 864225
8350/10 =
835
* Original IMMULITE instruments with serial numbers less than approximately 1900 have an external
pre-scaler. Later instruments above this serial number have the pre-scaler built into the PMT.
** The GETDATAT program does not compensate for the PMT Multiplier or the Attenuation factor. It
will factor in a pre-scaler if present.
In the luminogenic reaction (refer to figure below), the substrate (an adamantyl dioxetane phosphate 1 ) is
dephosphorylated into an unstable anion intermediate by the alkaline phosphatase conjugate captured on
the bead. The unstable intermediate emits a photon upon decomposition. The amount of light emitted is
directly proportional to the amount of bound alkaline phosphatase.
Compared to other means of detection, chemiluminescence provides the highest degree of sensitivity
available. In many cases, the sensitivity is orders of magnitude better than that attainable with
radioimmunoassays.
1
LUMIGEN PPD: 4-methoxy-4-(3-phosphatephenyl)-spiro-(1,2-dioxetane-3,2-adamantane). LUMIGEN is a
registered trademark of LUMIGEN, Inc., Southfield, MI.
O O
OCH3
DIOXETANE PHOSPHATE
(STABLE)
OPO3
O O
OCH3
UNSTABLE DIOXETANE
OCH3
O
O
LIGHT (hv)
3.5.1 Introduction
Primary Modules are defined as those modules that include stepper motors and require configuration
parameters and adjustments. Primary Modules require alignment through the configuration table (see
Section 7.2: The Configuration Table). Secondary Modules (discussed in Section 3.6: Secondary
Modules) are modules that require mechanical or electrical adjustments.
The Primary Modules of the IMMULITE are as follows:
Incubator Carousel
Load Chain Platform
Luminometer
Pipettor / Reagent Carousel
4. Heater Pad
5. Position Sensor PCB & rotating tower
During the RUN IMMULITE initialization process, the carousel will index before the run to find the first
position indicator. The carousel then stops just beyond the position indicator. The software then checks
for the passing of each of the three position indicators, 20 positions apart from each other. If the sensor
does not see another position indicator after 20 positions a Main Carousel Positioning Error will be
posted.
Note: When running the diagnostic CARPOS, the main carousel position error-checking sub-routines are
activated. A Main Carousel Positioning Error will post if triggered.
The load chain baffles do not inhibit assay tube travel in the transfer block (the crossover point
with path of Incubator Carousel).
BCODETST
Gearbox
This program will advance a tube into the Barcode Reader, read it 25 Assembly
times, advance the chain looking for another tube to read, then repeat.
This is the most useful test for the barcode reader, and can show useful patterns Motor
for diagnosing intermittent problems.
2
6 3
8
4 4
2
5
7
5
1
10
3.6.1 Introduction
Secondary Modules are defined as those support modules not directly responsible for tube transport,
and/or smaller modules that require mechanical or electrical adjustments.
Prescaler Description
Note: This module is only present on earlier manufactured instruments. At approximately serial number
1900, an integrated PMT was cut-in that contains the Pre-scaler components within the PMT itself.
This module provides the high voltage to the PMT and conditions the
resultant PMT counts. It interfaces with the PMT and CPU/Sensor
PCB. The main areas of interest are:
1. High voltage power supply
2. PMT prescaler
PMT Prescaler/Power Supply Module Adjustments
No adjustments of this module are necessary. The logic divider
switch at the bottom of the module must be set to 10 at all times
All Injector Pump modules automatically pump substrate and water into the assay tube during the
spin/wash cycle. The solenoid pumps interface via tubing to the water and substrate containers and the
Luminometer module. The solenoid pumps interface electrically with motherboard #1. The main areas of
interest are:
1. Solenoid pumps
2. Interface PCB
Substrate and Water Solenoid Pump Adjustments
1. Water or Substrate dispensing volume
The volumes of both solenoid pumps are factory set and should not need adjustment. Use the DISSUB or
DISWAT diagnostic programs (described below) to check the dispensing volumes. The volume should be
200ul 10ul for all pump types.
CAUTION: In regards to pump types #1 and 2, on the rare occasion where the volume of dispense is
incorrect, it can be adjusted within the inner assemblies of the pump module. However, as a general rule it
is better and safer to replace the entire pump assembly as the adjustment balance can easily be disturbed,
possibly leading to worse problems. If an adjustment in the field is absolutely necessary, contact
Technical Support for details.
NOTE: In regards to the newest design of the Linear Actuator Pumps, on the rare occasion when the
volume of dispense is incorrect, it can be adjusted within the inner assemblies of each pump module.
After removing the pump from its platform, the DIP switches inside control both volume and drawback
settings. Refer to Chapter 7: Miscellaneous Procedures for more details.
Note: Refer to section 7.4.4 for more information on drawback. Drawback is not adjustable with the
MGM solenoid pumps. The MGM pump must be replaced, if the drawback is more than .
Note: The temperature of the heater should be at ambient when the program first starts. Multiply the
voltage measured by 100 and this will be the temperature in Centigrade. The stabilized temperature of the
heater should be 36.6 +/- .6C. When done measuring the temperature, be sure to terminate the program
so as not to heat the substrate for a long period of time.
Note: Since this program checks to see if the heater is always activated, it is very important that the heater
be deactivated for at least five minutes before the program is run or a false failure may occur. Remember
that the heater is only activated when a test unit requiring substrate is seven minutes away from the
substrate dispense position. For example, if this program is run successfully, and then run again
immediately afterward, it will fail since the heater is still at temperature.
V4
5
5 V2
V1 6 and 7
7
A B A B
1 and 2 1 and 2
Notes: The older Hamilton 900 series module taps all system voltages (+/- 5, +/- 12, +24v). In the newer
500 series, only the systems +24v is used, and then is reduced as needed by the circuits of the Hamilton
500 logic/driver PCB.
All types of pump modules are interchangeable, as the electrical connections to the modules are identical.
Tubing kits may need to be changed, if going from Hamilton to Euro or from Euro to Hamilton. Refer to
Chapter 10 for a listing of the tubing kits.
The Syringe Module performs precision liquid transfer using stepper motor driven syringes to pipette
water, wash, sample and reagent. It interfaces via tubing to the water and wash containers and to the
pipettor module. The main areas of interest are:
1. "A" and "B" Diluter assemblies
2. "A" syringe (5000ul) and "B" syringe (250ul)
3. Macrovalve
4. Microvalve
5. Solenoid Valves (V1, V2, and V4)
6. Controller PCB (The Euro Syringe Pump combines the Controller and Interface PCBs into one board)
7. Interface PCB (The Euro Syringe Pump combines the Controller and Interface PCBs into one board)
8. Prime Button
Syringe Module Adjustments
No adjustments of this module are necessary or possible. The inner armature lead-screws are lightly
lubricated at the yearly PM.
Changing a Syringe
1. Loosen the thumbscrew under the syringe to be changed to a point at which the syringe plunger is
free from the driver arm.
2. Use the diagnostic program SCHANGE to move the syringes down.
3. Unscrew the syringe from the valve above while taking care not to exert side pressure the valve stem.
4. Install the new syringe onto valve fitting above.
5. Press the "GO" button to move the syringes up and tighten the thumbscrews.
Warning: Over-tightening the fittings into the valve above can deform the inner body of the valve and
cause air and/or water leaks.
Note: The clear acrylic manifold blocks are not regularly replaced. However, regular inspections should
be performed and the valves replaced if damage or cracks are found.
1. Making sure main power is off, disconnect short power wire of the solenoid valve from the rear
Controller PCB behind the acrylic manifold intended for replacement.
2. Remove three hex screws, which mount the acrylic manifold block to the front face of the module.
3. Remove the entire assembly (manifold and solenoid valve).
4. Install new manifold and solenoid valve in reverse order.
Syringe Module Test Programs
SCHANGE
This program moves both syringes down approximately halfway to permit their removal, waits until the
"GO" button is pressed, then moves both syringes up to home.
PRIME
This program primes the two syringes and the associated tubing. It exercises syringes "A and "B", valves
"A" and "B," and solenoid valves 1, 2, and 4. System error checking is active.
Note: Each of the three valve programs below assumes the valves not being tested are working properly.
This means if V1TEST is run and fails, insure V2 and V4 are working properly before replacing V1.
V1TEST
This program should alternately dispense water through the probe and the wash well. The change in
dispense point is altered by the activation or deactivation of V1. If the valve is not working, the water will
always dispense from the probe. If the valve is stuck in the activated position, the water will always come
from the wash well. If one of the two diaphragms in the valve is partially clogged, one port will slowly
drip while the other is dispensing.
V2TEST
This program should alternately dispense water through the probe and the detergent container tubing. The
change in dispense point is altered by the activation or deactivation of V2. If the valve is not working, the
water will always dispense from the probe. If the valve is stuck in the activated position, the water will
always come from the detergent container tubing. If one of the two diaphragms in the valve is partially
clogged, one port will slowly drip while the other is dispensing. Follow instructions on the IMMULITE
display panel.
Another way of testing V2 is to suspend the detergent container above the instrument and check to see if a
drop begins to form on the end of the probe. If a drop has not formed within 10 minutes, the valve is
sealing properly.
V4TEST
This program works exactly like V1TEST. This program should alternately dispense water through the
probe and the wash well. The change in dispense point is altered by the activation or deactivation of V4.
If the valve is not working, the water will always dispense from the probe. If the valve is stuck in the
activated position, the water will always come from the wash well. If one of the two diaphragms in the
valve is partially clogged, one port will slowly drip while the other is dispensing. This program uses V1
to direct the flow in a direction that can be differentiated from the other.
Note: All valve testing should be performed substituting a cut length of probe tubing acquired from a
previous PM service, installed in place of the existing system probe. This is needed in order to observe
and evaluate possible movement of the locked liquid level near the tip of this valve test tubing. The valves
should never be tested using the real system probe, as this liquid level near the tip cannot be seen. For
further details of proper valve testing, refer to the Section 7.4.4 in Miscellaneous Repair Procedures.
1
3
Description 1
This module collects the sample cups and holders ejected from the load chain so the user can recover
them for reuse. It interfaces mechanically with the load chain module. The main areas of interest are:
1. Removable tray
2. Tray housing
3. Tray-In-Place and Tray-Full Buffer Sensors
Sample Collection Tray Module Adjustments
No adjustments of this module are necessary or possible.
3.7.1 Description
Although most parts of the fluidic system are a part of one of the modules already discussed, a section in
this manual has been devoted to addressing the combination of all fluidic parts as a whole. The
components involved in the fluidic system are as follows:
Tubing
An older style of tubing (DIBA), used on Instruments up to L4917, is made where the shape of each piece
and the flare at each end are formed by the application of heat. When damaged, the entire tube assembly
must be replaced.
Each piece of tubing that comes in contact with liquid used for test unit processing is Teflon. The shape
of each piece is formed by the application of heat. The newer style tubes (Upchurch) may be repaired to
create a new clean end and a new ferrule of the appropriate size is mated to the end of the tube (wider
end of ferrule toward the tubes end). This repair work can be done using the Upchurch Tubing Repair
Kit, P/N 420109. Each tube is numbered and has color-coded fittings for easy tube identification. Color
coding also corresponds to the type of fluid that will pass through the tubing:
Blue - Fresh water from the water container
Red - Substrate
Yellow - Detergent or detergent and water mixture
White - Fresh water used to rinse the wash well
Green - Water to be used in the test unit
Each tube has a fitting that is designed to be gently tightened only with the tightening fitting, supplied
with the repair kit. (The probe has a hexagonal fitting and can be tightened with a wrench). The proper
technique for seating the tubing and tightening the fitting is:
1. Move the fitting away from the flared end approximately 15mm.
2. Insert the flared end into the port (of the valve, coupling, etc.) until it "bottoms."
3. Move the fitting into the port and tighten until it "bottoms."
4. For fittings in valves, tighten the fitting another 1/4 of a turn.
5. For fittings elsewhere, tighten until snug.
Note: Over-tightening the fittings can damage the flared end, or ferrule fitting, or the silicone O-ring of
the tubing (where used). It can strip the port of a solenoid valve, or deform the inner shape of a rotary
valve. If any of the mentioned parts are damaged, they must be replaced for proper fluidics operation. If
tubings are of the newer design using ferrules, these tubes can be repaired using the Upchurch Tubing
Repair Kit, p/n 420109.
Minimum
Control Range
Maximum
Operating 37.10 37.50 16.5 N/A N/A
Temperature
Lower Alarm
36.80 36.00 N/A N/A N/A
Temperature
Alarm
Range
Upper Alarm
37.20 38.00 22.00 34.00 40.00
Temperature
Minimum
displayable 34.XX 34.XX 11.60 11.60 11.60
Reportable
temperature
Range
Maximum
displayable 39.XX 39.XX 41.XX 41.XX 41.XX
temperature
6013 ohms @ 6013 ohms 15000 ohms 10000 ohms 10000 ohms
Thermistor Resistance
37.00C @ 37.00C @ 16.00C @ 25.00C @ 25.00C
Note: XX stands for the numbers that will vary depending on the circuit components of each specific
Heater-Motor PCB, and its controller circuitry.
1. The temperature display does not automatically update; the space bar must be pressed for the most
recent information to be displayed. To verify if the displayed data is current, compare the time of the
read with the current time displayed in the upper left corner of the screen.
2. The instrument must be running a program that sends information to the computer in order for the
temperature data to be valid.
When it is desired to print the real time temperature readings in this program, the FSE must let the screen
fill with three fresh blocks of data, then use the Print Screen key. Then let the screen update with three
more fresh readings, and press Print Screen again. In this way, up to six temperature updates can be
captured on one sheet of paper.
The real time temperatures are shown as the bottom line of each of these blocks of raw data.
A depiction of six temperature readings on one sheet of GETDATAT is shown below:
If a condition exists that sends the Temperature Controller system into a full shutdown, a default code of
45.xx will be posted into the address location of the Luminometer temperature. Though the Luminometer
may not necessarily have been the cause of the shutdown, this value will always be posted into the
Luminometer parameter. It is just default display indication of a full shutdown condition.
Note: In the event a short circuit occurs in one of the transistors that controls the power applied to a
heater (resulting in full power applied to that heater), a thermostat will terminate the power to that heater
when the temperature of the heated device reaches 45C. The thermostat will allow current to flow to the
heater when the temperature of the device decreases to 43C. While it is not permissible to process test
units at this elevated temperature, no damage to the Luminometer or Incubator Carousel will occur if
these temperatures are reached.
A full Temperature Controller shutdown occurs under one or more of the following conditions:
Luminometer and/or Carousel temperature is less than 35C for more than 60 minutes
Luminometer and/or Carousel temperature is greater than 39C for more than 10 minutes
Reagent Carousel temperature is greater than 30C for more than 30 minutes
Reagent Carousel temperature is less than 15C for more than 10 minutes
Note: The only way to reset the Temperature Controller circuitry after a full shutdown is to turn the
instrument off, wait ten seconds and then power the instrument back on.
Luminometer
Identity TED (Cooler) Carousel Heater Time Hack
Heater
LED
Lum temp is above Car temp is above 35C
Blinking N/A N/A
35C and below 39C and below 39C
@ 5hz
*** As the software is updated, these characters may also change to reflect the new software version
number. For example, with software version 4.3 the high level program is called HI43.HEX.
Note: For information on utility programs and procedures, refer to section 7.4 Miscellaneous
Procedures. For information on Diagnostic programs, refer to Chapter 8 Diagnostics.
Note: Export is not a menu option. You must exit the IMMULITE software to run this program.
3. Type 3 characters (at least one character must be entered to continue). The following screen appears:
4. At the Enter the starting date prompt, type the starting date (the year must be four digits) for
generating the data or press [Enter] to accept the default start-date.
5. At the Enter the ending date prompt, type the ending date (the year must be four digits) or press
[Enter] to accept the default.
Note: The test time is the time the sample cup was read at the Barcode Reader or entered into the
worklist - not the time the Test Unit was measured by the PMT.
6. At the Enter the starting time prompt, type the starting time for generating the data or press [Enter]
to accept the default start-time.
7. At the Enter the ending time prompt, type the ending time or press [Enter] to accept the default.
8. At the Enter the test type prompt, type the Test Code or press [Enter] to accept the default (which is
ALL test types).
9. Results for controls, Adjustors, and patients can be printed together or separately. Select either:
1 - Control
2 - Adjustors
3 - Patients
4 - All
The following screen appears, giving the option to print, display, or put the data on disk.
CAUTION: Always leave the tutorial by selecting the Exit button or by pressing [F12]. If you exit the
program by any other means, or if a software error causes the program to stop, reboot the system by
turning the computer off, waiting ten seconds, and turning it back on.
Software Overview accesses the IMMULITE Primary screen, which contains information regarding
samples, tests, and reagents on the instrument. From the Primary screen, selecting the Main Menu
options provides an explanation of each option.
CHEMISTRY
The Chemistry selection accesses information about the IMMULITE's enzyme-enhanced
chemiluminescence, master curve generation, two-point adjustment, and internal calculations used to
determine the final patient result. The operator may select the overview or more detailed information on
each topic.
The Checklist selection accesses the steps required to run the IMMULITE during an average day,
including start-up, running, and shutdown activities. The operator may select the overview or more
detailed information on each topic.
INDEX
The Index provides a listing, by category, of all topics contained in the IMMULITE On-Line Reference
Software. For information regarding a specific topic or question, select Index and click on the desired
topic.
4.1 Introduction
This section contains descriptions of the Printed Circuit Boards, or PC Boards. Refer to the Assembly and
Schematic Manual part number 600240 for actual drawings of the boards.
Beginning at power on, the IMMULITE goes through its initialization routines. Program instructions stored in
EPROM #1 are accessed with 15 bit addressing. The lower byte is obtained from port 0 of the Super 8 and
latched from the address/data bus. Port 0 is wired to a multiplexed bidirectional address/data bus. The upper 7
bits of the address are gotten from port 1, which is dedicated to addressing. Program instructions from the
EPROM and data from the non-volatile static RAM are then placed on the data bus to be read by the
processor.
Note: The red DIP switches near the boards top edge do not affect system operation of this PCB, and are
only used for final vendor testing of PCB performance.
Note: The red DIP switches near the boards top edge do not affect system operation of this PCB, and are
only used for final vendor testing of PCB performance.
Circuit Description:
The heater/DC motor board performs the following control functions for the IMMULITE system: (1)
temperature measurement and control of the Luminometer, the Main Carousel Incubator, the Reagent
Carousel, the PC board housing, and the ambient temperature within the instrument package, (2) velocity
control of the reaction unit wash operation, (3) velocity control of the sample/assay bar code motor, (4)
power distribution to the solenoid valves and the wash fluid and substrate pumps.
To provide the temperature control critical to the performance of the IMMULITE system; a dedicated ZILOG
Super 8 microcontroller is used to process temperature measurement data in real time, and in response,
control the amount of heat applied to the assay incubator and luminometer and remove heat from the reagent
carousel. Temperature control begins with the precise temperature measurement at key system positions. A
semiconductor negative temperature coefficient (NTC) thermistor with a guaranteed measurement accuracy of
0.05 oC is embedded in the Luminometer and Main Carousel housings. This assures that the temperature of
measurement is as close as physically possible to that of the reaction temperature. The resistance vs.
temperature curve of the thermistor is linearized with a full bridge network. The bridge components consist of
high precision resistors, 0.1 % tolerance with a temperature coefficient of 25 ppm, and it is driven with a high
The following table describes the performance characteristics and specifications of the IMMULITE
Analyzer.
Specifications
General Standard Mode Throughput: Up to 120 tests per hour
Specifications:
Time to first result: 42 Minutes
(or 72 minutes for two cycle assays)
Tests per sample: Up to five per sample
General Turbo Mode Throughput: Up to 80 tests per hour
Specifications:
Time to first result: Less than 15 minutes for a one-cycle
assay
Less than 22 minutes for a two-cycle
assay
Tests per sample: Up to five per sample
Luminometer Detection method: Photomultiplier Tube
Specifications:
Wavelength of Emitted 425 to 500 nm
Light:
Incubation temperature: 37C +/- 0.1C
Internal Computer Computer: IBM AT Compatible, 486DX
Specifications: 33Mhz or higher
Minimum Req. RAM: 4 MB or higher
Minimum Req. Hard drive: 120 MB or higher
Monitor: Super VGA
Auxiliary Drives: 3.5 Floppy Drive required
declare under sole responsibility that the following equipment to which this declaration relates,
meets the essential health and safety requirements and is in conformity with the relevant sections
of applicable EC standards and other normative documents. If changes are made to the product
which is covered by this declaration of conformity, the declaration of conformity is no longer
valid.
Model: IMMULITE
National and other IEC 1010-1, IEC 801-2,3,4, IEC 1000-4-2, IEC 1000-4-4,
standards and technical ENV50140, ISO 9001:1994 (BS EN ISO 9001:1994),
specifications: 21CFR, Part 820 FDA cGMP, EN46001: 1996;
ISO 13485: 1996; UL 1262, Laboratory Equipment
Notified body according
to Annex VII: Not Applicable employed to hold the documentation
and evaluate the safety issues of this equipment.
Date/Signature of manufacturer
or responsible party: ______________________________________________________
Date/Signature of EU representative
or responsible party: ______________________________________________________
Title of signatory: ______________________________________________________
LIMITED WARRANTY
Siemens Diagnostics warrants that the items delivered hereunder are of good material and
workmanship, and are free from defects in design and manufacture. Siemens Diagnostics
responsibility is limited to repairing or replacing any item or part, for a period of one (1) year
after delivery to the original purchaser. Defects caused by improper operating conditions,
misuse, negligence, or alteration of the product void this warranty. Siemens Diagnostics shall
not be liable for any direct, indirect, incidental or consequential damages arising out of
possession or use of the items. Consumables, as defined in the appropriate Siemens Diagnostics
Price List for Instrument-Related Parts, Racks and Consumables are not covered by this
warranty.
CUSTOMER REMEDIES
Siemens Diagnostics entire liability and your exclusive remedy shall be replacement of the
Software that does not meet Siemens Diagnostics Limited Warranty and which is returned to
Siemens Diagnostics. The Limited Warranty is void if failure of the Software has resulted from
accident, abuse, or misapplication.
NO OTHER WARRANTIES
Because the software is inherently complex and may not be completely free of errors, you are
advised to verify your work. The software and related documentation are provided as is.
Siemens Diagnostics disclaims all other warranties, expressed or implied, including but not
limited to implied warranties of merchantability and fitness for a particular purpose, with respect
to the software and the accompanying written materials. Siemens Diagnostics shall not be liable
for any direct, incidental, or consequential damages arising out of possession or use of this
product.
6.1 Description
Preventive maintenance is an important service function that must be performed at least once a year. This
prevents problems from occurring due to normal wear or adjustments changing over time. The following
Preventive Maintenance Kit part number 420054 is available which contains all the parts that may need to
be replaced:
When performing Preventive Maintenance, be sure to use the IMMULITE Annual Preventive
Maintenance Report, available from Siemens Diagnostics Field Service as a guide to ensure all items are
covered. The rest of this chapter outlines each item in detail.
CAUTION: Make sure the tips are fully seated on the end of the plungers. Gently press the tips vertically
down onto a flat surface until a click is felt. The large syringe should also be gently rolled on its tip to be
sure it is straight. Take extreme care not to bend the shaft of the small syringe during this process.
It is recommended to pre-moisten the new large syringe tip with the IMMULITE water before re-inserting
it into the glass syringe. If resistance is felt, gently pull the plunger back out and re-insert until there is an
even smooth travel upon re-insertion. Do not rotate the plunger when re-inserting, this can cause the
syringe to dislodge again later.
Run the PRIME diagnostic and check for excessive bubbles particularly in the small syringe. If
necessary, disengage the thumbscrew from the plunger during the PRIME diagnostic and manually move
the small plunger abruptly up and down to clear persistent bubbles. Be careful to prevent any side pull as
this can bend or damage the plunger shaft.
Check operation of V1, V2, & V4: Tests the operation of V1, V2, and V4. Refer to Section 7.4.4:
Miscellaneous Procedures for information on checking the operation of the valves.
Check the operation of the micro and macro valves: Run the PRIME diagnostics. Proper operation can be
verified by checking that fluids are running in the proper direction.
Check pipettor level sensor operation: Run the Configurations tests LEVFSTST and LEVSNTST, and
the regular menus LEVFALSE and LEVCRASH diagnostics. Refer to Section 3.5.5: Theory of
Operation, and Sections 8.5 and 8.6: Diagnostics of this manual for more information on the use of
these diagnostics.
Check Pipettor-Z lead screw and slide bearing: Visually check for excessive wear, dust or grime
accumulation. No lubrication is generally necessary as this can cause future dust accumulation. Check for
smooth travel.
Check tip-jam sensor operation: Run the TIPJMTST diagnostic found under the System Configurations
diagnostic menu. After the program checks five times, a pass/fail result is displayed on the control panel.
Check reagent tray positions (Home and Position): Check for the proper position of the home sensor by
placing a wedge in the home position (Position 1) of the Reagent Carousel and run the REAGENTH
diagnostic. The red LED should shine an oval shape centered between the edge of the wedge and the first
barcode line.
Check for the proper position of the individual position sensor using an empty reagent carousel and while
running the REAGENTP diagnostic. Look down directly through the individual notches found around
the edge of the carousel and check that one of the position sensor heads is centered in the notch. Verify all
wedge positions are centered over the sensor head underneath.
Change probe and check dispense: Replace the probe with the new one supplied in the PM kit.
The probe tubing must be placed in the notch of the vanity panel with the small rubber O-ring located
behind the panel. Manually turn the pipettor to ensure the tubing does not twist or kink. Adjust or flip the
tubing as needed to avoid coiling or wrapping around the pipettor tower.
Check the dispense quality of the new probe. Run the PROBETST diagnostic. Look for excess liquid
moving up the tip of the probe shaft, or any splashing out of the side of the home well. When the probe
momentarily lifts and moves to the dump well, look for any dripping out of the probe. Dripping indicates
backpressure or occlusion inside the probe or at its 4-way junction block at the syringes.
CAUTION: Be aware of the small rectangle shim located underneath this slide bearing. Be careful not to
drop or lose it. A dab of grease can be used upon re-assembly to help keep it in place.
Remove the slide bearing and rail. Clean off excess old grease. Use a clean small blade screwdriver to
work new grease into the rails where the ball bearings reside. After re-assembly into the instrument, slide
the bearing back and forth on the rail to work in the grease. Remove the excess grease.
CAUTION: Do not allow the slide railing to completely disengage from the bearing.
Change small idler wheel assembly: Refer to Section 7.3.13: Removal and Replacement of Assemblies
for instructions on the replacement of the small idler wheel assembly. The entire assembly including the
shaft on which the wheel rotates, should be cleaned.
Check the high- speed spinner O-rings: Take a flashlight to visually inspect the O-rings of both the rear
drive wheel and outer idler wheel for cracks or wear. Wipe clean with a damp paper towel or replace the
wheel assemblies if damage is found.
Set High- Speed Spinner speed (5400-5600) RPMs: Run both the configurations menus SPNCKTST,
and the regular menus WASHSPD diagnostics, to monitor and/or adjust the high speed spinner speed,
and its error-checking capabilities. Refer to Section 8.0: Diagnostics for more information on these tests.
Check shuttle positions and verify shuttle sensor flag is not bent: Run the SHUTTLE diagnostic to ensure
the proper positioning of the Shuttle in all positions. Refer to Section 8.0: Diagnostics for more
information. Use a flashlight and manually move the shuttle to check that the flag end moves correctly
and is not physically bent or broken.
Check for proper wash & spin of test unit: Load 15 test units with approximately 200 ul of water on the
carousel one position before the Shuttle position. Run the ATTENTM diagnostic to visually check for
proper wash and spin operation. The test unit should slightly lift on the first spin and the liquid should
quickly clear from the bottom of the tube. Watch that the bead moves off the bottom of the tube and floats
in air, while the tube is spinning. This item should also be checked during the final chemistry run of the
instrument.
Check Luminometer Chain positioning and tension: Run both the LUMCHECK and SHUTTLE
diagnostics to evaluate the Luminometer Chain position versus the shuttle. It is best to remove the
Spinner Platform for direct visual clearance for the checks. Refer to Section 8.0: Diagnostics for more
information on the chain position adjustment. The chain should align so the shuttle passes through the
center of a Luminometer Chain baffle.
Set barcode reader speed to 750-790 ms/rev: Run BCODESPD diagnostic to adjust the barcode reader
speed. Use a minimum of four test units to check all four positions around the idler assembly. Each
position should not vary more than 40 msec of each other. If it does, check and/or replace rollers and pins
or check for even rotation of the drive O-ring.
CAUTION: Do not push the button, as this can cause damage to the button.
6.11 Fluidics
Replace water & probe wash filters: Replace the probe wash and water end-of-line filters (inside the
bottles) while wearing gloves or using a lint-free tissue.
Replace air filter in caps of water and wash containers: Replace with filters from the PM kit.
Replace waste tubing and check waste path for blockage: Replace with waste tubing from the PM kit.
Check all other waste tubing and manifold block for blockage or clogs. Clean and/or rinse as needed.
Check all tubes for kinks and tighten all fittings: Visually check the tubing for any kinks. Check that all
tubes are finger tight. Specifically, check water tubes coming in from the water bottles for any kinks
and/or damage from the top cover edge (crushing the tubes when closed).
Decontaminate system and test for contamination: Refer to Section 7.4: Miscellaneous Procedures for
information on performing system decontamination. Enclose a copy of a printout of the final
WATERTPM test with the PM report. If decontamination of the substrate system is needed, consult
Siemens Diagnostics Technical Support for detailed options and procedures.
Note: Be sure to manually write the instrument serial number, CAF value and PMT value on the printed
output of the configuration table.
6.13 Temperatures
First check the substrate heater temperature. Remove the cover on the substrate heater. Using a DVM set
to DC volts, place the leads on the test points of the Substrate Heater PCB. Run the SUBSHEAT
diagnostic. Record the starting (ambient) temperature and final (heated) temperature on the PM report.
The DVM reading is in direct correlation to the temperature. Multiply the measured voltage by 100 and
this will be the temperature in Centigrade. The stabilized temperature should be 36.6C +/- 6 o C. After
measuring the temperature, be sure to terminate the program so that the substrate is not heated for a long
period of time.
Check all other temperatures while running the final chemistry run (the next step in the PM) by going into
the System Status dropdown menu, and select the Temps and Dark Counts option. Press the space bar
to obtain updates of the temperatures until they stabilize. Record all temperatures on the PM report.
Note: Controls should be run in five replicates of each level control. Look for the majority of replicates to
be in range. Updated control ranges are issued regularly or can be found in the associated package insert
of the CON6 control module.
Check the error log after the adjustment and compare the last adjustment for that assay with the current
one run. Ensure there is reasonable consistency between the two.
Ensure the customer is happy with the final chemistry run results and that all issues and questions have
been addressed. Do not leave until final confirmation has been made with the customer.
Not all of the tools listed are readily available outside the United States however, any or all of the above
can be purchased from Siemens Diagnostics.
The configuration table is a file located in the C:\CIRRUS\IMM directory of the computer and is always
prefixed by the letters IMM followed by an alphanumeric or numeric number with a CNF extension.
For example, the configuration table for IMMULITE serial number D0100 is IMM100.CNF. The
alphanumeric or numeric number within the file name is part of the instrument serial number. (The serial
numbers prefix letter code is not in the file name.)
Parameters
The configuration table has many parameters, each of which represent an adjustment within the
instrument. Each parameter has a parameter number, parameter name, error number, and extra number.
These are defined as:
Parameter Number: The identification number of the parameter.
Parameter Name: The abbreviation of the specific adjustment the parameter represents.
Notes: 1. For reasons too complicated to explain, the error steps in parameters 32, 33, and 34 do not set
the point of corrective action initiation, they set the depth of the probe in their respective
positions.
2. The error steps for parameters 36-40, and 42-48 are not used and set to zero. These parameters
use the error steps (200) from parameter 8.
Parameter Extra Number (commonly referred to as "extra steps"): This is the parameter that sets the
mechanical positioning of all sensors in the instrument. The mechanical components of the instrument
were designed such that when a device driven by a motor triggers a corresponding sensor, that device then
has a small number of steps more to move to be positioned properly. The small number of steps taken
after the sensor has been triggered; signaling a proper position of the device is defined as the extra steps.
Example: The Load Chain motor begins to move the Load Chain to the next position. After 550 steps the
open slot in the sensor disk reaches the optical sensor and allows light to flow through the sensor
triggering its output. The microprocessor then looks at the configuration table for the number of extra
steps needed to properly position the Load Chain and sends that additional number of steps to the motor.
To put the positioning in terms of coarse vs. fine, think of the mechanical placement of sensors as the
Error Step coarse adjustment and the Extra Steps as the fine adjustment.
ERROR EXTRA
# NAME DESCRIPTION RANGE RANGE
0 LcPos Sets the positioning and error detection of the Load Chain. Fixed 35-180
Not currently used (McHome stands for Main Carousel Not Not
5 McHome
home). used used
Error steps are used for error detection of the pipettor when
8 PxPos moving to one of the X-Pipettor positions. Extra steps are Fixed 5-13
only used in a few diagnostics.
The error steps are used to set the minimum sample volume. Not
13 LevSensS 175-205
The extra steps are not used. used
The error steps are used to set the volume (1000 ul) at which
14 LevSensR 420-464 Fixed
the reagent vial is considered "empty" by the instrument.
Sets the horizontal positioning of the probe in the home well Not
36 PX Blind 2-20
of the wash station. used
Note: To keep a proper updated printout in the rear electronics, be sure to replace the existing printout
after any changes are made. Also, obtain the PMT and CAF numbers, and Instrument Serial number, and
write them on the final sheet. Replace into the rear electronics compartment. This printout is insurance of
easy retrieval of all of this information, in case of computer failure.
CAUTION: The power cord must be removed from system before starting removal of any module.
1. Loosen the thumbscrew on the substrate heater and move away from the carousel ring gear.
2. Loosen the ten slotted screws on the main section of the clear plastic carousel cover and remove the
cover.
3. Remove the probe tubing from the notch on black probe vanity panel.
4. Remove the probe by unscrewing it from the upper pipetting arm and unscrewing the fitting at the
syringe module.
5. Loosen two screws holding the probe vanity panel and remove the panel.
6. Loosen three screws holding the Reagent Carousel cover (two located on top and one located on the
left side of cover close to the chassis). Carefully move cover up and away from module.
7. Remove Reagent Carousel Tray.
8. After removing the front lower chassis panel, unplug the two-pin connector (A12-J5) on the Reagent
Carousel Interface PCB that is coming from the TED Cooler below.
9. Loosen the cardboard collar that surrounds the lower part of the TED fan [keeping it in place].
10. Remove two foam pads (do not discard) covering the top screws securing the TED Cooler from
above. Loosen the two screws allowing the TED Cooler to drop into the lower chassis cavity.
11. Manipulate the TED Cooler from out the front of the lower chassis area if replacement or inspection
of the TED Cooler is desired, otherwise the TED Cooler can remain in the lower chassis.
Note: Care must be taken in the TED removal, as space is limited. It may be necessary to shift and/or
remove some of the drain tubing in order to make room. Do not lose the thin black conductive pad
that is sandwiched between the head of the TED. It is required for proper heat transfer.
9. Reassemble in reverse order. Before re-inserting the Main Carousel Ring Gear Assembly, check that
the Main Carousel positioning sensor is still rotated out of the way. Do not over-mesh the gears of the
motor assembly to the Main Carousel Ring Gear. Use moderate pressure only.
1. Remove two 9/64 hex screws that are securing the bracket on the top of the Barcode Idler assembly.
Remove bracket.
2. Remove three Phillips head screws securing the Barcode PCB, do not drop any washers into the
instrument.
3. Manipulate the PCB out from its mounts and remove the J1 connector.
4. Re-assemble in reverse order
1. Remove the Sample/Test Unit Barcode Reader PCB as described in steps 1 through 4 in Section
7.3.7.
2. Lift the entire motor and bracket assembly from the lower pivot point bushing, while holding the
motor return spring, which is inserted towards the rear of the bracket.
3. Loosen the 5/64 locking hex screw of the barcode motor drive wheel and remove the drive wheel
from the motor shaft.
4. Remove the two small 1.5 mm hex screws securing the motor to the top of the bracket.
5. Cut the wire tie securing the motor wires to the motor body. Trim the heat shrink tubing of the wires
to expose the connection point with the power cable, which is routed through the small hole in the
bracket.
6. Cut the motor wires from the power cable. Strip new ends to the power cable. Slip a new piece of
shrink tubing (supplied with kit) onto the power cable and through the bracket hole.
7. Solder the connections of the new motor wires, insulating one from the other with electrical tape.
Slide the new shrink tubing over the solder joints. Ensure connections are protected.
8. Use a heat gun to carefully shrink the tubing around the solder joint and through the bracket.
9. Remove the DC Motor/Heater PCB (A2).
10. Replace the Q17 driver FET with one supplied in the kit with a solder iron.
11. Re-assemble the motor to bracket and secure wires to motor body with new wire tie.
12. Re-assemble the Sample/Test Unit Barcode Reader Motor in reverse order. Carefully re-insert the
drive motor return spring.
CAUTION: When re-securing the drive wheel to the motor shaft, be sure to align the screw with the
flat of the motor shaft. Also ensure the drive wheel is at the proper height so as not to rub the motor
mounting screws below or to contact the Barcode Reader PCB opto-sensor above.
4. Prepare the new O-ring and drive wheel by gently moistening them with water.
5. Use two fingers to partially stretch the O-ring around the drive wheel to insert the new O-ring.
6. Just before the O-ring is completely around the drive wheel, carefully insert a small blade screwdriver
between the O-ring and the far side of the drive wheel. Use the screwdriver to pry the remaining part
of the O-ring onto the drive wheel. Do not remove the screwdriver yet.
Note: Once the screwdriver is removed, it is not possible to re-align the seam of the O-ring. If the
seam is not straight, it must be replaced and the procedure started over again.
7. Gently rotate the screwdriver around the drive wheel 5 to 6 times. Visually check that the O-rings
molding seam is straight around the drive wheel without spiraling. Carefully, remove the
screwdriver.
8. Manually press the O-ring into place around the circumference of the drive wheel. Inspect for a
straight seam with no high or low areas on the O-ring circumference.
9. Re-assemble in reverse order.
1. In the case of Linear Actuator Pumps, carefully remove single connector behind the pump, leading
from the Y cable from rear of each pump. In the older Cirrus Solenoid Pumps, the single cable
connection to the rear PCB is carefully removed, after removal of a snap-on plastic cover.
2. Move the shuttle assembly manually to the far left, to the home position at the Main Carousel.
3. Remove front output fitting of the red substrate side and the two rear input fittings from the Wash
Pump Assembly (keep the Water Pump output fitting at front connected).
Note: After removing tubing from the substrate side rear input, raise the tubing and tape it to the
Substrate Reservoir Assembly or bottle, to prevent leakage of substrate.
4. Using a Phillips screwdriver, loosen the two captive screws at the rear edge of the base of the Wash
Pump Assembly.
CAUTION: Do not let the screwdriver shaft damage the pumps or the PMT behind.
5. Lift the Injector Pump Assembly straight up and out of the instrument.
6. Re-assemble in reverse order.
7.3.9 Substrate Heater Assembly
1. Move the shuttle assembly manually to the far left, to the home position at the Main Carousel.
2. Loosen the screw at the rear of the heater to free the wire clip and the L-bracket securing the tube
fitting.
3. Disconnect all wiring from the rear of the Substrate Heater (A162-J1, J2, and J3).
4. Disconnect red substrate output fitting from the top of the Substrate Heater. Allow a few seconds for
the substrate to drain from the heater into shuttle drain below.
5. Loosen the thumbscrew securing the heater to the platform. Remove the entire assembly.
6. Re-assemble in reverse order.
7.3.10 Substrate Reservoir Assembly
1. Disconnect the reservoir level sensor wire from the rear of the substrate heater at J2.
2. Disconnect the tubing from the Wash Pump Substrate input.
Note: After removing the tubing from the substrate pump input side, raise the tubing and tape it to
the Substrate Reservoir Assembly or bottle, to prevent leakage of substrate.
3. If present, disengage the bottom drain tubing of the Substrate Reservoir Assembly.
4. Loosen the single captive screw at the base of the Substrate Reservoir Assembly.
5. Remove the entire assembly from the platform being aware of the slot that is cut into the base at the
rear of assembly.
6. Re-assemble in reverse order.
1. Remove the Substrate Heater Assembly and the Substrate Reservoir Assembly.
2. Remove the single screw and the tube guide at the top of the black plastic cover of the Luminometer
Module and carefully manipulate the cover off and out of instrument.
3. Manually move the shuttle from the left side, about two inches in toward the Luminometer.
4. Disconnect the 8-pin connector (A10-J4) from the Luminometer Interface PCB.
5. Loosen four captive screws located at the four corners of the High-Speed Spinner Assembly.
6. Remove the entire spinner platform assembly including the Spinner motor and idler wheel.
Motor Assembly removal from platform
7. Remove the three 9/64 hex screws (in a triangle format) holding the High-Speed Spinner Motor and
Drive Wheel.
8. Gently separate the upper cylinder bracket from the motor assembly below. The Spinner platform is
sandwiched in-between.
9. Remove the cable clip holding the cable going to the High Speed Spinner Motor.
Motor removal
10. Orient the assembly so that the drive wheel is facing toward the top.
11. Locate the motors spring coupler in the center of the assembly. Loosen the two .050 hex screws of
the coupler closer to the drive wheel.
CAUTION: Do not loosen the two hex screws closer to the High Speed Spinner Motor.
12. Loosen (but do not remove) the four 5/64 hex screws securing the High-Speed Spinner Motor
Assembly. Access these screws through two holes found in the Drive Wheel Assembly. Drop the
motor out from the bottom.
CAUTION: Keep the assembly upright and do not remove the screws from the bracket since re-
insertion of these screws is very difficult.
6. With the screws loose, carefully remove the slide bearing and rail off of the shuttle assembly.
CAUTION: Do not allow the bearing assembly to completely slide off the rail.
7. Clean and/or lubricate the slide bearing and re-assemble in reverse order.
CAUTION: Do not lose a small thin rectangular shim located towards the rear of the bearing
assembly. The shim is necessary for proper alignment and performance of the upper shuttle
assembly. A dab of grease may be applied to the shim to hold it in place during re-assembly.
1. Remove the Tube Shuttle Upper Slide Bearing Assembly referring to steps 1 through 5 in Section
7.3.12.
2. Using a small-blade screwdriver partially spread the small retaining-clip at the top of the Idler Wheel
shaft, until the clip is removed from the catch groove.
3. Once the tiny clip is removed, the silver cap above the Idler Wheel can be removed and should be
cleaned.
4. Below the silver cap is a small wave-washer. This is replaced (P/N 900261) as part of the Shuttle
Idler Wheel replacement kit (P/N 420001).
Important: The wave washer must be re-installed with the bowl of the washer facing up.
5. After removal of the cap and wave washer, the Idler Wheel can be pulled off the shaft and cleaned or
replaced. Once removed, the opportunity should be taken to carefully clean the entire area of the
upper shuttle body assembly around the idler wheel.
6. Re-assemble in reverse order with the wheel on the shaft and re-capture, using a new retaining-clip
(P/N 900262) of the Shuttle Idler Wheel replacement kit (P/N 420001).
4. Locate the black metal bracket ring located at the end of the PMT. Remove the three 7/64 hex
screws that hold this ring in place. Carefully, slide the ring away from the PMT.
5. Gently shake the PMT back and forth and outward to facilitate removal.
CAUTION: After the PMT is removed, even with the power off, caution must be taken to not
expose the PMT to ambient light, as damage can occur. Carefully, tuck the PMT face into the back
corner of the instrument and cover if necessary. DO NOT TOUCH THE PMT LENS!
6. Remove all the connections from the Luminometer Interface PCB (A10). Release the five plastic
standoffs that secure the A10 PCB to the top of the Luminometer.
7. Lean the PCB back and off the rear of the Luminometer cover. The PCB cannot be totally removed
because the other wires are soldered to the right edge of the PCB.
8. Remove the two sensor wires from the attenuator home and positioning sensors found on the right
side of the Luminometer cover. Properly mark the connectors to facilitate re-assembly.
9. Loosen the five 9/64 hex screws located around the lip of the upper Luminometer cover. There are
three screws along the left edge of the cover, one at the front center and one in the rear right corner.
Tool extensions may be necessary. The rear right corner hex screw has a wire clip under it that
secures a number of sensor wires. Keep the hex screw inserted in the clip for easier re-assembly later.
10. Read all four cautions listed below before proceeding to the next step.
CAUTIONS: 1.When attempting to lift the Luminometer cover off, some resistance may be felt
due to tightness within the locator pins. It may be necessary to gently pry the
cover free of its locator pins with a screwdriver.
2. The Luminometer cover may be trapped along the left edge under the plastic
casing of the lower Main Carousel casting. Gently flex or pry the Main Carousel
casing to free the Luminometer cover.
3. Do not damage the Luminometer Temperature Thermistor, which remains
screwed into the top cover, located behind the Chain Drive Motor Assembly.
4. Do not touch the attenuator lens located on the disk assembly within the
Luminometer cover. Maintenance on this assembly is rarely needed. For further
details on this assembly, please contact IMMULITE Technical Support.
11. Grasp the motor assembly from the top and carefully lift the cover off the module, while pivoting the
Luminometer Interface PCB out of the way. The Luminometer chain is now exposed.
WATERTPM
The WATERTPM diagnostic program tests the water dispensed from the probe and the water pump.
Use this procedure when contamination within the system is suspected.
1. Check to make sure the daily maintenance has been performed and that the substrate and the Wash
Pump water lines have both been sufficiently primed.
2. Make sure the printer is on-line.
3. Select four empty water Test Units.
Note: Do not pipette water into any of the Test Units.
4. Load all four water Test Units onto the Load Chain.
5. From the Start-up screen, choose DIAGNOSTICS and then the submenu DIAGNOSTICS.
6. Select WATERTPM and press [Enter].
7. When the loading is complete, press [Enter] again.
13. Type start and press [Enter] to return to the IMMULITE Start-up menu.
14. Refer to Acceptance Criteria on the next page to determine the outcome of this test.
Unacceptable alkaline
> 9,000 phosphatase contamination is
present
If the CPS is too high or too low, perform the test again to confirm these results.
If the CPS is consistently outside the 4,500 CPS to 9,000 CPS range, find a new water source and
test this new source.
If the second test unit (substrate only) is > 9,000 CPS, the substrate system may be contaminated
or the test units used for testing may not be clean.
Note: The Water result is then not valid, as the unacceptably high level of this baseline substrate only
reading is falsely biasing it.
2. For WATERTPM, refer to the last two lines printed with results:
Probe Water Substrate Only = XXX CPS
Pump Water Substrate Only = XXX CPS
If the value of XXX for either equation above is lower than 650 CPS, insufficient substrate may have
been dispensed into one of the Test Units. This could be due to a clogged nozzle. Visually check all
Test Units to determine if the liquid level is equally distributed between the Test Units. If the value of
XXX above is above 2000 CPS, perform the test again to confirm these results. If the CPS is
consistently above 2000 CPS, decontaminate the system. Refer to section 7.4.2 System
Decontamination on the next page.
17. Refill the water container with fresh IMMULITE acceptable distilled water.
18. Using a sterile cotton swab, wipe the accessible surface of the wash station to help remove any
growth.
19. Run the DECON1 diagnostic program with fresh distilled water:
Note: Prior to filling the container, the water must be tested using the WATERTST diagnostic program.
From the same Diagnostics Menu, choose DECON1, and press [Enter].
When program has finished loading, press [Enter] again.
Press GO on the IMMULITE when instructed to do so.
The program is finished automatically, in approximately fifteen minutes.
20. Re-test the system for contamination using the WATERTPM diagnostic program to verify the
system now meets the proper criteria.
-- When priming before chemistry, be sure to press and hold the prime button down for each full cycle.
This creates a more proper full extension of the bellows, for a better, quicker prime.
-- When running decontamination programs DECON & DECON1, it is suggested to tape down the
water prime button, for the same purpose as above.
-- Be sure to always have the Instruments Power off whenever removing or connecting the power cable
to the pump. -- Doing so with the power on will cause instant and permanent damage to the pumps
electronics!!
1. Insert the larger gauge (representing the test unit) into the carousel ring gear position where the load
chain crosses the carousel ring gear.
2. Manually rotate the carousel ring gear three spaces back towards the shuttle.
3. Initiate and load the CARCHECK diagnostic from the main diagnostic menu. Refer to Chapter 8:
Diagnostics for more information on the initiation and loading of diagnostics.
4. After pressing GO, watch the test unit gauge increment one space at a time alternating with the
shuttle.
5. Press and hold the ALARM MUTE button and then press the GO button when the test unit gauge
arrives at the dispensing position one step beyond the load chain.
6. Initiate and load the LCHAIN diagnostic. Press GO and allow the load chain to increment two steps
to be sure it is properly indexed. Press the ALARM MUTE button and the GO button to stop the
diagnostic.
7. Place the smaller gauge (representing the sample cup) at the left most (number 5) sampling position at
the Star Wheel (See next section for Star Wheel details.)
Leading edge
Will encroach slightly into
Carousel casting
Trailing edge
Should be flush with outer
wall of Carousel casting
Sampling
Position #1
Space of
tube on
chain
The correct position of the Star Wheel is determined by the following two points of reference:
The position of a sample cup at the number 5 sampling position
The position of a sample cup as it loses contact with the Star Wheel and enters the sample
collection tray
4. Manually, place one sample cup in sampling position number 5 and another in the Load Chain
position just past the Star Wheel. Rotate the Star Wheel to the position where contact with the sample
cup in sample position number 5 is with the right Star Wheel tooth, and the other sample cup is
perpendicular with the opposite Star Wheel arm. Refer to the illustration above.
5. Once proper alignment is performed, carefully hold the Star Wheel in position and tighten the nut on
top, being careful not to change the Star Wheel position.
6. Repeat steps 1 and 2 to verify alignment. If necessary, repeat the adjustment.
7. Finally, check the adjustment under a high load condition. Fill the Sample Collection Tray with
sample cup holders to nearly full. Repeat step 2 watching for proper travel with the high load
condition.
pipettor base
1. Using a Phillips-head screwdriver, loosen the two screws securing the inner sample tube guide.
2. Use a standard piece of paper folded in fours as a gap tool. Gently slide the folded piece of paper in-
between the inner sample guide and the pipettor base.
3. Using a gentle even pressure onto the center point of the inner guide, tighten the two screws down
and remove the piece of paper.
4. Inspect for an even gap at each end of the inner sample tube guide against the pipettor base.
5. Using a Phillips-head screwdriver, loosen the two screws securing the outer sample tube guide.
6. Insert sample cup holders upside-down in between the inner and outer sample tube guides at each
end.
7. Apply gentle pressure on the center of the outer sample tube guide against the upside-down sample
cups and tighten the two screws down.
8. Inspect the adjustment by riding an upside-down sample cup through the space of the two sample
tube guides. There should some slight play without any resistance or obstruction throughout the entire
pathway.
9. Perform a final test by placing 5 to 10 sample cup holders on the Load Chain and run the LCHAIN
diagnostic from the Diagnostic Screen. View the sample cups as they move through the sample tube
guide area. There should be no tilting of the sample cups or any resistance or obstruction seen.
4. Press and hold the PAUSE button to index the next test unit. Release the PAUSE button once
indexing starts.
5. Repeat the above adjustment for each of the remaining three positions around the Barcode Idler
Assembly.
Note: The four positions around the Idler Assembly may differ from each other as much as 40
ms/rev. This is acceptable. If the differences are greater than 40 ms/rev, inspect the condition of the
rollers, O-rings, pins, and main drive O-ring. Refer to Section 7.3.7 for more details on the
inspection of these parts.
Cautions: Be aware of false readings. These are tube-specific, based on the different mode in
which this test uses the reader circuit. False readings are ranged 200 440 ms/rev, and are known to
be false, as the reader motor cannot rotate more than 500 ms/rev. When these false readings are seen,
it is suggested to try using a test unit of a different assay, until 4 tubes are found which remain
stable, when rotating in the reader. The barcodes of some assays may randomly hinder accurate
readings of these speed tests.
Also note that if numbers are jumping between real and false, note the audible tone of the barcode
motor. The jumping numbers are known to be false, as usually there is no corresponding variation in
the tone of the motor.
7.4.9 Tuning the Sample/Test Unit Barcode Reader PCB P/N 450056
1. Turn IMMULITE power on.
2. Connect a DVM to TP1 (positive) and TP2 (negative) using alligator clips.
3. Place a test unit on the Load Chain and run the BCODESPD diagnostic from the Diagnostic Screen.
4. Press GO to index the test unit. Once the test unit reaches the barcode reader, press ALARM MUTE
and GO to stop the test unit from indexing.
5. Manually rotate the test unit so that the barcode readers red beam is reading the all-white section of
the test unit.
6. Adjust potentiometer R1 to 3.0 VDC.
$ + T A 3
Default Config.
. +
$ + T A 1
3. The enter key on a personal computer is typically a carriage return <CR>. To set the postamble to
<CR>, scan the barcode below.
Carriage Return
Test 123
T e s t 1 2 3
3. Scan the PC-AT, PS/2 and 50/60/80 barcode (Hex Equivalent is CF):
8. Scan the Confirmation barcode and verify 01234567 appears after the DOS prompt with an error
Bad command or file name. This is normal since DOS doesnt know what to do with this
information:
DBASE317.ISM CNF317.ISM
Patient Results Laboratory Name (PC Screen)
Control Results LIS Configuration
Adjustment Slope and Intercept Parameters Dark Count Limit
Kit Entry Information PMT Multiplication Factor
Patient Normal Ranges Attenuation Factor
Control Ranges
Test Parameters
QCLIS317.ISM PAT30.INI
Control Data Long Report Format
Patient Data
Current Worklist
Current LIS Worklist
CAUTION: Before replacing any file, make a copy of the file to avoid loss of data. In the event of a
problem, or if replacing the file was not the solution, it can be returned to its previous status.
To replace the DBASE317.ISM file from the installation disk, do the following:
1. Insert the version 4.0 software diskette into drive a.
2. At the c:\CIRRUS\PC prompt, type the following:
arj x a:pcver40.exe dbase317.ism then press [Enter].
3. Following the extraction procedure, the system will prompt the following message:
DBASE317.ISM is same or newer, Overwrite?
4. Type y for Yes, then press [Enter].
You may now execute the RUN IMMULITE or START TURBO program from the start-up menu.
To replace other components of the database, simply use any other .ISM filename in the above procedure.
In the vast majority of cases for any drive or motor issue, historically the problem has rarely been the
motor. They are much heartier than the application that are used for. Any strange activity of the motor is
more likely to be a fault for its support diodes on the module interface PCBs.
Testing the resistance of the motor windings is as seen in the small chart above. If readings are in spec, it
is assured the motor has no issue.
The purpose for the diodes is to protect the driver circuits on the Motor Driver PCBs #s 1 and 2 from
spikes that occur from collapsing EMF from the motors.
The purpose of the PTC devices is to protect the drive circuitry in case a short occurs in the motor. The
PTC device will heat up due to higher current, and the resistance will increase, therefore keeping the
current down, limiting damage that can occur.
Note: As of IMMULITE 1000 serial number D0640 produced in January 2005, all these devices have
been removed and hard-jumpered. In older systems where these devices are present, if they break or
become damaged, simply remove the device and solder in a jumper.
8.1.1 Investigate
Discuss past performance with customer: Customer communication ensures that all problems experienced
by the customer are covered. Proper communication helps guarantee customer satisfaction is achieved after
you leave. Ask for any specific situations that the customer may have seen or heard on the IMMULITE.
Check Error Log: Review at least two months of the log. Check for consistent error messages. They could
be warning signs of impending problems. Also, do not be concerned with individual isolated messages.
They may simply indicate an operator error.
Check for IMMULITE upgrades: A list of current upgrades and standards are available from Siemens
Diagnostics Field Service. If necessary, Siemens Diagnostics Field Service can fax a copy to an FSE
onsite. Evaluate the instrument for the presence of all current upgrades and perform any necessary
revisions.
Can the problem be duplicated?: Does the same problem occur every time a certain situation is duplicated?
Does the problem occur randomly?
Follow typical troubleshooting techniques to narrow down the location of the problem: Refer to Section
9.2 for more information.
8.1.2 Power
Check that AC power is within range: 110-volt systems should have incoming VAC voltage of 100 120
volts. 220-volt systems should have incoming VAC voltage of 210 230 volts. Check the intended wall
socket using a line voltage tester (available through Technical Support).
Check voltages at the Syringe Interface PCB: Check DC voltages at the main power connector (A15-J1) on
the Hamilton/Euro Interface PCB. Pins can be accessed through the side of the connector with small pin
type probes. A good chassis ground can be found at the grounding strap connected to the front latch of the
instruments lid. Ensure voltages are in tolerance as indicated below:
5 VDC: (4.9 to 5.2) -5 VDC: (-5.2 to 4.8)
12 VDC: (11.8 to 12.2) -12 VDC: (-12.2 to 11.8)
24 VDC: (23.8 to 24.2)
Check if voltages are in range at the PMT Power Supply: Check voltages for the PMT power supply by
locating the quick-connect connection for the cable of the PMT, and tap into these points to read the +/-
5V supplies dedicated to the PMT. Ensure voltages are in tolerance as shown below.
5 VDC: (4.8 to 5.2) -5 VDC: (-5.2 to 4.8)
15 VDC: (14.8 to 15.2)
The MESS43 (Main Message Table) is loaded automatically with the RUN
IMMULITE/START TURBO command from the start-up menu.
The Main Diagnostic Menu (DIAGNOSTICS) uses whatever message table was last loaded.
The SYSTEM CONFIGS. Menu allows the loading of different message tables upon entry
and exit of the menu.
Note: The rest of this page below applies only to older software versions up to 4.1. For version 4.3 and
later, loading these message files is automatic when navigating between the two menus.
When selecting DIAGNOSTICS from the start-up menu and then SYSTEM CONFIGS., the user must be
careful when answering of questions relating to the message tables. Normal operation is listed below.
1. When entering SYSTEM CONFIGS., the following question should always be answered with
a Y for Yes:
CAUTION: Answering the above questions incorrectly can cause wrong messages or a blank display
while running diagnostics. If this occurs, follow the procedure below.
8.1.4 Loading the Correct Message Tables (versions 4.1 software and earlier)
RUN IMMULITE/START TURBO The Main Message table is automatically loaded when RUN
IMMULITE/START TURBO is activated. For this reason, there is no need to reload message tables in
this mode.
Main Diagnostics (DIAGNOSTICS) If wrong messages or a blank display are seen while running these
diagnostics, simply download the MESS43 message table file (found on this menu) and then another
Go command will also be sent, where the chosen program will now run with correct text on the panel.
Note: Refer to Section 9: Troubleshooting for more troubleshooting information, including a list of error
messages.
Diagnostic programs are used to test particular components of the IMMULITE System. When the cursor is
positioned on the diagnostic program name, a description of the program appears at the bottom of the
screen.
To load and run the diagnostic programs, follow the steps below:
1 Ensure IMMULITE is in Idle. From the Start-up menu, select DIAGNOSTICS.
2 To access the Main Diagnostic Menu, select DIAGNOSTICS and press [Enter].
4 There are also special Configurations diagnostics, which can be accessed by selecting SYSTEM
CONFIGS from the menu choice screen.
Note: Additional diagnostic programs may become available with future software updates.
6 Once you have entered the needed menu, select the desired diagnostic program by pressing the arrow
keys to highlight the program name and press [Enter].
7 When the loading is complete, press [Enter] again (sending the * RUN * command).
8 Press GO (on the IMMULITE) when instructed to do so.
9 Whenever terminating diagnostics that are in motion progress, first press and hold the ALARM
MUTE button until all motion stops, and while still holding the ALARM MUTE button, then press
GO to stop the program, which will set the instrument into Idle mode, ready for the next download.
Note: Some routines stop automatically at the end of the program cycle, while others repeat until the
ALARM MUTE and GO buttons are pressed in sequence described above.
Notes:
Error checking routines are not activated during the use of most main diagnostics except in specially
needed cases.
Error messages are posted during CARPOS, CARSHAKE, MCARTST, TEMPS, PRIME, and
PMTCOUNT main diagnostics, and when using certain System Configs adjustment and logic test
programs.
1. After the diagnostic program begins running (after pressing GO on the IMMULITE), press [Esc].
2. From the Start-up menu, select EXIT TO DOS.
3. Type getdatat and press [Enter]. A menu appears.
4. Press [P] to turn on the printer (if a printout is desired).
Notes: Make sure the printer is on-line and properly loaded with paper.
If a printout of PMT signals only is desired, then activate printing by the step 4 above. It will not include
the bottom line of data depicting real time temperature readings.
If a printout of the temperatures is needed, leave the printer off in this screen, allow the screen
to fill with readings, then use the Print Screen key to print all data including the temperatures.
Key to the labeled items above (all data is in real time as it posts to the screen):
1 This is one single block of the Getdatat raw data readings (three readings are shown here).
2 The exact date and time of the reading.
3 Twelve one-second CPS signals from the PMT (the highest and lowest are dropped, to create #4).
4 A general average of the twelve PMT readings in the line above (termed as Cleaned CPS).
5 The bottom line of each block gives real time readings of all five temperature parameters.
The remaining contents of this data beyond the labeled items above, are erroneous numbers that should be
ignored, and are of no practical use to field repair work.
Note: In some rare circumstances, communication may be lost (PC Communication Failure) during a
chemistry run. If the IMMULITE is still running normally, the PC program may be used to track the
chemistry run progress. If a TURBO chemistry run is in progress, the user must use the PCSTAT
program. Care must be taken to use the correct program PC or PCSTAT depending on the assay run in
progress, otherwise the display and indexing of the instrument motions may be incorrect and mis-
matched.
1. After the diagnostic program begins running (after pressing GO on the IMMULITE), press [Esc].
2. From the Start-up menu, select EXIT TO DOS.
3. Type PC or PCSTAT and press [Enter].
4. Respond to the questions displayed by pressing [Enter] three times.
5. When the following message appears, choose [Y] for yes:
Cannot communicate with the IMMULITE.
Continue running? Y/N
6. To exit, select Log Off and Log Off System.
ATTURN
Use this program to visually check the PMT attenuator disk No PMT Attenuator Disk
motion and to verify sensor operation. The test cycles the
attenuator disk to the home position, attenuated position and
the unattenuated position. This can be monitored by watching
the position of the single sensor notch found on the right side
of the Luminometer cover above the PMT. The single sensor
notch points to a 9:00 position at home, 5:00 in the
attenuated position and 1:00 in the unattenuated position.
BCODE
Verifies proper functionality of the sample/assay barcode No Sample/Test Unit Barcode
reader. Place as many test units as desired onto the load Reader
platform. A test unit is read once every two seconds, then the
next test unit is indexed in front of the barcode reader. Verify Program used to View:
proper barcodes using the PC program. PC
BCODESPD
This program is used to observe and/or adjust the motor No Sample/Test Unit Barcode
speed of the sample/assay barcode reader. Load at least four Reader
test units onto the load platform. Run the program and
observe the motor speed. It should be between 750 to 790
ms/rev. Press the pause button to advance to the next test
unit. Each position should not be more than 40 ms/rev of
each other.
BCODETST
Verifies proper functionality of the sample/assay barcode No Sample/Test Unit Barcode
reader. Place as many test units as desired onto the load Reader
platform. Each test unit is read 25 times, and then the next
test unit is indexed in front of the barcode reader. Verify Program used to View:
proper barcodes using the PC program. PC
BKGNDCNT
This program provides the current background counts for the No PMT
instruments PMT. It will alternate moving the attenuator
between the attenuated and unattenuated position. While at Program used to View:
each position, a set of 12 one-second background counts is GetDataT
taken. The GetDataT program provides the actual raw
background counts.
BPRIME
Used to isolate priming of "B" small syringe for diagnostic No Syringe B Pump Assembly
evaluation. Also, useful when performing a small syringe tip
replacement.
CARBACK
Useful for removing sample cups that have been mistakenly No Incubator Carousel
indexed around the incubator carousel. This program moves
the carousel backward in single index steps by pressing the
syringe prime button momentarily, or indexes continuously by
holding the prime button.
CARCHECK
This program is used to judge the position of the Main No Incubator Carousel
Incubator Carousel in relation to the tube Shuttle. It indexes
the carousel ring gear one position, moves the shuttle into
the space between the carousel baffles, pauses for one
second, moves the shuttle back to home and repeats.
CARPOS
This test checks the indexing of the main incubator carousel. No Incubator Carousel
It indexes the carousel ring gear one position every second.
Also activates system error checking to aid diagnosis.
DARKCNT
This program provides the current dark counts for the No PMT
instrument. It moves the attenuator to the home position, then
takes a set of twelve one second dark counts with the PMT Program used to View:
every fifteen seconds. The GetdataT program provides the GetDataT
actual raw dark counts.
DECON
This program is used to introduce decontamination material No Fluidics (Decontamination)
into the fluidics of the Instrument. Refer to Maintenance
Procedure listed in Section 7.4.2 for further details. This
procedure draws the decontamination fluid into both syringes,
one at a time and then into the water pump five times. This
procedure should be run for at least 15 minutes, followed by
the DECON1 procedure.
DECON1
This program must be used after the DECON procedure. The Yes Fluidics (Decontamination)
purpose of this program is to flush out the decontamination
fluid and replace it with fresh fluids. Refer to Maintenance
Procedure listed in Section 7.4.2 for further details. This test
will automatically stop after approximately twenty minutes.
The test will cycle through filling each syringe twice then the
water pump ten times.
DISSUB
This program tests substrate pump dispensing volume. Place Yes Substrate Pump
five test units on the load platform as per display. If the wrong
number of vessels is used, the program will halt with Test
Failed. The program indexes each test unit around the
Incubation Carousel then shuttles them to the substrate fill
position. A shot of substrate is added, and the test units
moved to the sample collection tray for evaluation.
DISWAT
This program tests water pump dispensing volume. Place five Yes Water Pump
test units on the load platform. If the wrong number of
vessels is used, the program will halt with Test Failed. The
program indexes each test unit around the Incubation
Carousel then shuttles them to the water fill position. A shot
of water is added and the test units moved to the sample
collection tray for evaluation.
LEVCRASH
This program tests for the proper level sensing of the probe No Pipettor (Level Sensing)
in each of the five pipetting positions. Manually, pipette a
minimum of 100 uL of liquid into five sample cups. Place the
five sample cups into the five pipetting positions. During the
test, the Pipettor will level sense each sample cup starting in
position five and then progress cup by cup to position one.
The program will time-out after 20 rotational cycles (~ 100
potential signals) Any failures to level sense the liquid will trip
the tip jam sensor, which in turn will indicate test failed.
LEVFALSE
This program tests for the proper level sensing of the probe No Pipettor (Level Sensing)
in the reagent, five sampling, and assay dispense pipetting
positions. This test requires no sample cups to run. Remove
the Reagent Tray. The Pipettor will start at the home position
and then checks level sensing at the reagent position, each
of the five sample pipetting positions, and then return to the
home position to start again. The program will time-out
automatically after 20 full cycles to home (~ 120 potential
signals). Any false level sense triggers will trip the tip jam
sensor, which in turn will indicate test failed. During the
sequence the technician can also jiggle the wires or cables to
attempt to purposely initiate intermittent problems.
LEVSENSR
This program is used in conjunction with LEVSENSS to No Pipettor
check the auto calibration of the DEADVADJ (automatic), or (Level Sense Reagent Dead
LEVSNADJ (manual set-up) dead volume Configurations Volume Check)
programs. This program should display an SL (steps left)
number between 8 and 12. If the program does not confirm
good SL numbers, repeat the Configurations DEADVADJ
Auto-calibration. Or if the LEVSNADJ test is used, re-check
for proper manual pipetting of the minimum dead volume in
the test reagent vial.
LUMCHECK
This is one of the most useful tools for close inspection of the No Luminometer Chain
crucial crossover alignment of the Shuttle and Luminometer (at the crucial cross-over point
Chain. The program repeatedly inserts the shuttle partially with the shuttle)
into the indexing lum.chain baffles.
Note: It is recommended to remove the entire Spinner
platform assembly to more properly judge this crucial
alignment, sighted from straight up above the area.
LUMTEST 4, LUMTEST 5,
LUMTEST 6, LUMTEST 9
Used in manufacturing only. N/A N/A
PIPXPOS
This program allows for visual inspection of the Pipettor No Pipettor
positioning. It moves the X-Pipettor to each of 10 rotational
positions, starting with home. At each position, it lowers the
Z-Pipettor momentarily. This test is used in conjunction with
the Pip-X Positioning Gauges. Refer to Section 7.4.5 for
more information.
PMTCOUNT
Activates measurement readings by the PMT, but does not No PMT
move the attenuator disk. Use the GetDataT program to see Program used to View:
the current readings. GetDataT
SHUTLOAD
This program allows for the evaluation of the shuttle as it No Shuttle
loads test units into the spin position. Manually load a test
unit onto the carousel one position before entering the
shuttle. The test unit will be loaded into the spin position.
Inspect for proper actuation of the springs of the shuttle, so
that the test unit fits snugly pushed into the spinner wheels.
The test unit should sit on a straight angle (without tilting).
Press GO to have the program eject the test unit for removal.
SHUTSPIN
This program is used to evaluate the high-speed spinner Yes Shuttle, High Speed Spinner
cycle. Manually load a test unit with 200 uL of liquid pipetted Motor
into it, onto the carousel one position before entering the
shuttle. The test unit will be loaded into the spin position and
the high-speed spin motor will activate for 30 seconds. The
test unit should slightly lift up and the liquid should quickly
clear from the bottom of the tube. Watch that the bead moves
off the bottom of the tube and floats in air, while the tube is
spinning. To check the actual spinner speed, use the
WASHSPD diagnostic.
SOLE1W
This program is used to check the drawback adjustment of No Water Pump
the water pump. Disconnect the water nozzle and hold a
sample cup under the water nozzle. Run the SOLE1W
diagnostic to dispense two shots. Use the ALARM MUTE
button to pause and control the shots into the cup. The
approximate level of water should be at the fourth line of the
sample cup (2 shots at 200 ul = 400 ul). Confirm that the
drawback is within specification of , +/- . Also refer to
the end of Section 7.4.4 for further adjustment details.
SOLE2S
This program is used to check the drawback adjustment of No Substrate Pump
the substrate pump. Disconnect the tubing on top of the
substrate heater and hold a sample cup under the substrate
tubing. Run the SOLE2S diagnostic to dispense two shots.
Use the ALARM MUTE button to pause and control the shots
into the cup. The approximate level of substrate should be at
the fourth line of the sample cup (2 shots at 200 ul = 400 ul).
Confirm that the drawback is within specification at , +/-
. Also refer to the end of Section 7.4.4 for further
adjustment details.
SPINCHK
This program allows for a quick visual check that the high- No High Speed Spinner Motor
speed spinner motor is working. Manually load as many test
units as desired onto the carousel and manually rotate to one
position before entering the shuttle. The test unit will be
loaded into the spin position and the spinner motor will turn
on for about 3 seconds.
SPINON
This program turns the high-speed spinner motor on or off, No High Speed Spinner Motor
when pressing the GO button.
TEMPS
This program is used to monitor instrument temperatures. No Temperature Sensors of:
The temperatures are updated every thirty seconds and can Reagent Carousel,
be seen using the GetDataT or PC programs (GetDataT Luminometer, Main Carousel,
gives readings in real time). The screens can also be Ambient System, Ambient
printed. Electrical
Also, refer to Section 3.8 for complete temperature
Program used to View:
information.
GetDataT or PC
Note: The PMT is not addressed in this diagnostic and
therefore will read zero.
UNATTN10
Used in manufacturing only. N/A N/A
UPDOWNA
This program is very useful for checking the centering and No Probe
rough depth of the probe in the test unit and the functions of
the Z-Pipettor home sensor and mid sensor. Load a test unit
on the load platform. Use the EMPTYTUB diagnostic to index
the test unit to the Main Carousel dispense position. The
program will move the X-Pipettor to the test unit dispensing
position, then move the Z-Pipettor to the home sensor, down
to the mid sensor, then down the amount of error steps for
parameter 34 (Z Assay) of the configuration table.
UPDOWNH
This program is very useful for checking the centering and No Probe
rough depth of the probe in the home well and the functions
of the Z-Pipettor home sensor and mid sensor. The program
will move the X-Pipettor to the home well position, then move
the Z-Pipettor to the home sensor, down to the mid sensor,
then down the amount of error steps for parameter 32 (Z
Home) of the configuration table.
V2TEST
This program tests the functionality of Valve 2 (V2) of the No Syringe Valve 2
syringe assembly. This program alternately dispenses water
through the probe and the detergent container tubing. The
change in dispense point is altered by the activation or
deactivation of V2. If the valve is not working at all the water
will always dispense from the probe. If the valve is stuck in
the activated position, the water will always come from the
detergent container tubing. If one of the two diaphragms in
the valve is partially clogged, one port will slowly drip while
the other is dispensing. Another way of testing V2 is to
suspend the detergent container above the instrument and
check to see if a drop begins to form on the end of the probe.
If a drop has not formed within 10 minutes, the valve is
sealing properly.
V4TEST
This program tests the functionality of Valve 4 (V4) of the No Syringe Valve 4
syringe assembly. This program works exactly like V1TEST.
This program should alternately dispense water through the
probe and the wash well. The change in dispense point is
altered by the activation or deactivation of V4. If the valve is
not working at all the water will always dispense from the
probe. If the valve is stuck in the activated position, the water
will always come from the wash well. If one of the two
diaphragms in the valve is partially clogged, one port will
slowly drip while the other is dispensing. This program uses
V1 to direct the flow in a direction that can be differentiated
from the other.
WATERTPM
This program is used whenever system fluidics contamination Yes Fluidics
is suspected in the instrument. Refer to Section 7.4.1: (suspected water contamination
Miscellaneous Procedures for further information. in the instrument)
Note: This program needs 4 Water Test Vessels. If the wrong Program used to View:
number of vessels is placed onto system, the test will stop WATER
and indicate Test Failed on the display panel.
WATERTST
This program is used whenever water contamination is Yes Fluidics
suspected from the source water supply. The program tests a (suspected water contamination
water source independent of the systems fluidics. Refer to from the water supply)
Section 7.4.1: Miscellaneous Procedures for further
information.
Program used to View:
Note: This test needs 2 Water Test Vessels. If the wrong WATER
number of vessels is placed onto system, the test will stop
and indicate Test Failed on the display panel.
PIPXRADJ
This program performs an isolated adjustment of rotation Yes Probe
alignment of probe over reagent wedge. After pre-setting (X-rotational position at Reagent)
an open reagent wedge at the probe position with
REAGENTP, and with the reagent chamber lid open, use
the small opening of the reagent vial as sight judgement
for a centering target of the probes X-Position over this
hole.
PIPXSADJ
This program ensures the probe is centered in sample Yes Probe
cups in the five pipetting positions. To prevent damage to (X-rotational position
the customer probe, install a dummy probe. The smaller at all 5 Sample Cups)
notched Pip-X positioning gauge (from Kit p/n 420078)
is used to precisely target and/or adjust each of the five
sample pipetting positions. The program actions will work
clockwise around from position S5 to S1. Start the target
gauge at S5, and set the probe one at a time, using the
Go button for raising the test probe and shift the target
gauge to next S-position. Ensure the probe comes down
within one probe width of the center target of the gauge.
Refer to Section 7.4.5 for more information.
PIPXTADJ
This program allows the adjustment of the rotational Yes Probe
alignment of the probe at the dispensing position into the (X-rotational position at Assay Test
test unit. Place the larger (without a notch) Pip-X Unit Dispense Position)
positioning gauge (from Kit p/n 420078) in the test unit
dispensing position using CARCHECK. Run the
diagnostic and check for proper positioning either
centered or within one probe width to the rear (counter-
clockwise). Refer to Section 7.4.5 for more information.
This is the preferred diagnostic to test the tip jam circuit. Yes Probe
The program bottoms out the probe five times to trip the (Tip Jam Sensor Function)
tip jam sensor. A Pass/Fail condition will be posted on
the display panel. Can be repeated several times to
ensure proper operation.
9.1 Description
This section is designed to aid in the quick, accurate diagnosis of IMMULITE problems. While many
people can make IMMULITE repairs, accurate troubleshooting is a rare skill for the experienced as well
as the inexperienced.
In its simplest state, troubleshooting is an exercise in logic. It is essential to realize that the IMMULITE is
really composed of a series of modules. Some of these modules are interrelated, others are not. The
instrument operates within a framework of logical rules and physical laws, and the key to troubleshooting
is a good understanding of all the IMMULITE modules.
LCHAIN Intermittent Load Chain errors. The chain has become stiff from
spilled liquids.
EMPTYTUB The position sensor disk has dust
interference.
Luminometer Luminometer Chain is not The "LumcPos" parameters in the
moving. config table are incorrect.
The motor and/or motor drive
circuitry has failed.
The shuttle is mispositioned and
interfering with the chain.
A lumchain baffle has become
dislodged and is interfering with the
LUMCHN chain.
An assay tube or partially dislodged
LUMCHECK Luminometer Chain baffle has
become jammed inside the
MCARTST luminometer, thus jamming the
chain.
SHUTTLE The chain tension has increased
surpassing the torque capabilities of
the motor.
Luminometer Chain is not The "LumcPos" parameters in the
moving to the correct position. config table are incorrect.
The position sensor has become
dislodged.
The slotted sensor disk has become
loose.
The chain tension has increased
causing the motor to miss steps and
"time out" during a move.
Drive Motor coupling inside cover
may be shifting loose.
EMPTYTUB No motors are moving nor is The fuse for the heater/motor power
the main cooling fan moving. supply (F3) has blown.
The heater/motor power supply has
become defective.
DARKCNT The CPS of the test units is The fuse for the PMT power supply
zero. (F1) has blown.
BKGNDCNT The PMT power supply has become
defective.
Linear Actuator Substrate / Air bubbles have appeared in An up-line fitting has loosened.
Water Injector Pumps the tubing. The water or substrate supply is low
enough to allow air to start entering
the system.
DISSUB (substrate) No substrate being dispensed. The substrate nozzle is clogged.
The substrate spike is clogged,
DISWAT (water) causing a vacuum.
The Substrate bottle is empty.
The Substrate actuator pump or
driver circuitry has failed.
Substrate volume dispensed is The substrate nozzle is partially
incorrect. clogged causing the substrate to
SOLE2S (substrate) dispense at an angle partially
missing the assay tube.
SOLE1W (water) Clogged air filter in the substrate
spike.
Reservoir was never primed.
Defective substrate actuator pump.
No water being dispensed. Water container is empty.
A line fitting up-line has become
very loose.
Water actuator pump or driver
circuitry has failed.
Water volume being dispensed The water supply bottle has become
is incorrect. empty.
Defective water actuator pump.
Air leak in the input line to the
pump.
Clogged in-line water filter.
Carousel Temp. High The Incubation Carousel The IMMULITE should correct
[Low] temperature is above or below the itself. If it does not, within 20
Note: The error acceptable range. minutes, troubleshoot using
Section 3.8.
message will read Note: All results generated during
either High or Low. the high or low temperature period If the component feels hot, turn
will be flagged on the printout. the instrument off, allow it to
cool, and reset and evaluate
using Section 3.8.
Collection Tray The sample collection tray is full If the sample collection tray is
Full or missing. full, remove the sample cups
from the collection tray and
Note: An alarm sounds and the replace the sample collection
IMMULITE goes into tray.
AutoPAUSE mode.
If the sample collection tray is
missing, replace it.
Follow the instructions on the
instrument Display Panel to
continue.
Lum. Chain Error. Appears on the IMMULITE Check for Test Units (or other
Display Panel if an error is obstructions, such as
detected and the IMMULITE is Luminometer chain baffles) that
able to correct itself. may interfere with the chain
movement.
Note: If the IMMULITE cannot
correct the error, the instrument Investigate all related positions
automatically shuts down and a and alignments.
message indicates this has Note: Repeat any tests that
occurred. were in progress.
Main Car. Error This message will appear on the Run related diagnostics,
IMMULITE Display Panel if the including the test of the sensor
error corrects itself. hole sizes MCARTST.
Note: If the problem cannot be
corrected, the instrument
automatically shuts down and an
error message indicates this has
occurred. This is often caused by
a sample cup holder jammed on
the Carousel under the drive gear.
Mechanical error Either: Based on the cause, either:
encountered
(pipetting) This message appears in Determine the cause of all bad
conjunction with numerous barcode reads and correct if
bad barcode reads. possible. If persistent,
A problem was encountered troubleshoot related tests for the
with the pipettor. BC Reader area.
Determine the source of the
pipettor interference and correct
if possible.
Message received The computer missed a message The software automatically
out of sequence, from the IMMULITE. requests that the message be re-
Seq = xx sent.
OldSeq = yy If the message persists,
problems are evident for the
computers hard drive.
Message Too Short Usually indicates a bug in the LIS Re-send the results to the LIS.
LIS mesg. Too software; however, could be If the error continues to occur,
short\bad frame caused by a bad message (due to Customer should contact their
line noise, etc.). LIS provider.
Negative Slope Sample cups were identified as low Re-run both levels of Adjustors,
calculated! Are the and high, but either the levels pipetting each level into the correct
Adjustor levels dispensed into the cups were in the sample cup holder.
reversed? reverse order or only one level was
run for both samples.
General Information
Generally, when traveling to an IMMULITE site to replace a part, it should be known before leaving what
part or parts will be needed. However, there have been situations when for a number of possible reasons,
the part that will fix the problem is either incorrect or not on hand. (A few situations have occurred where
a service person has been in the vicinity of a "down" instrument but did not have the correct part to make
a speedy repair). If this is the case when the service person is far away from the office, it could mean
many more hours, possibly another day, before the instrument is able to operate. A list of parts to have on
hand when responding to a trouble call has been developed based on the experience of IMMULITE
service people around the world. Even though some parts on the list have never failed, they should be
carried since they are small, but critical to instrument operation.
The following pages contain the Fluidics, Chassis Wiring Diagrams and Isometric Drawings. For
board drawings refer to the Schematic and Assembly Drawings section.
450029
MAIN CAROUSEL
POSITION PCB
450020
MAIN CAROUSEL
INTERFACE PCB
(A14)
500387-11
MAIN CAROUSEL FOAM KIT
400120
MAIN CAROUSEL HEATER PAD
(LOCATED UNDER MAIN CAROUSEL
FOAM KIT)
400083-01 400064-04
CABLE ASSY,MAIN MAIN CAROUSEL
NOTE:BEARING ASSEMBLY AT CAROUSEL THERMISTOR MOTOR ASSEMBLY
9:00 POSITION IS ADJUSTABLE BY (LOCATED UNDER
USING TWO 1/8" HEX SCREWS CASTING)
LOCATED BELOW.
400142-01
GROUND STRAP
ISOMETRIC DRAWING,
IN C U B A T O R C A R O U S E L
ASSEMBLY P/N 400003
SHEET 1 of 1
500339
LUMINOMETER COVER
400064-06
LUMINOMETER CHAIN
MOTOR ASSEMBLY
450011
LUMINOMETER/WASH
INTERFACE PCB
(A10)
400095
HIGH SPEED SPINNER ASSEMBLY
400084-01
LUMINOMETER
THERMISTOR ASSEMBLY
420040
HIGH SPEED SPINNER IDLER KIT
400064-08
LUMINOMETER
ATTENUATOR MOTOR
ASSEMBLY
420038
HIGH SPEED SPINNER MOTOR
ASSEMBLY
(901132
WITH COUPLING)
450006
SHUTTLE AT HOME
420001 GENERAL PURPOSE
OPTOSENSOR
SHUTTLE IDLER
500110
REPLACEMENT KIT ATTENUATOR MOTOR
SHAFT PULLEY
900262
RETAINING CLIP
500183
SHUTTLE IDLER WHEEL
COVER PLATE 400126
NON-INTERGRATED
900261
900026 PHOTOMULTIPLIER
WAVE WASHER TUBE ASSEMBLY
LUMINOMETER
500577 ATTENUATOR BELT
UPPER SHUTTLE SHIM OR
400164
(LOCATED UNDER TWO 450006
SHUTTLE IDLER WHEEL
REAR SCREWS) ATTENUATOR POSITIONING 400248
900731-16 GENERAL PURPOSE INTERGRATED
UPPER SHUTTLE SPRING OPTOSENSOR PHOTOMULTIPLIER
TUBE ASSEMBLY
450006
900731-32 ATTENUATOR HOME
LOWER SHUTTLE SPRING GENERAL PURPOSE
OPTOSENSOR
900159 450006
SHUTTLE UPPER SHUTTLE AT SPIN &
BEARING ASSEMBLY SUBSTRATE POSITIONS
400190-02 GENERAL PURPOSE
LUMINOMETER CHAIN ASSEMBLY 500048-02 OPTOSENSOR
BAFFLE,MOLDED
900160 LUMINOMETER CHAIN 400064-07
LOWER SLIDE BEARINGS SHUTTLE MOTOR
500506 500507
SIDE PLATE,CHAIN ROLLER,CHAIN
900160
LOWER SLIDE BEARINGS
ISOMETRIC DRAWING,
COMPLETE LUMINOMETER MODULE
SHUTTLE ASSEMBLY P/N 400020-02
SHEET 1 of 1
400185
REAGENT COOLING
CHAMBER ASSY
450014
REAGENT CAROUSEL
HOME & POSITION PCB
450007
REAGENT CAROUSEL
BAR CODE READER PCB
(A49)
400145
REAGENT TRAY
450051
LEVEL SENSING &
PROBE JAM
SENSOR PCB
(REFER TO KIT 420072
INSTRUCTIONS FOR
INSTALLATION)
420072
COMPLETE EURO
LEVEL SENSING KIT
(INCLUDES 450050 PCB,
450051 PCB AND MOUNTING
HARDWARE)
450050
PIPETTE Z AND
LEVEL SENSE
INTERFACE PCB
LED
450017
PIPETTOR X
HOME / POSITION PCB
450010
REAGENT CAROUSEL
INTERFACE PCB
(A12)
400064-01
REAGENT
CAROUSEL 400184
MOTOR SPIRAL CABLE ASSEMBLY
900422
TE D
THERMAL SHEET
400064-03
PIPETTOR "Z" MOTOR
400178
T E D ASSEMBLY 900114
PIPETTOR "X" MOTION DRIVE BELT
(FROM PIP-X MOTOR AROUND
Z-MOTION LEAD SCREW - NOT SHOWN)
ISOMETRIC DRAWING,
420096
T E D KIT P IP E T T O R / R E A G E N T
(INCLUDES THERMAL SHEET
AND FAN COWLING)
CAROUSEL
400064-02 MODULE P/N 400023
PIPETTOR "X" MOTOR
SHEET 1 of 1
CABLE ASSEMBLY TO LOAD CHAIN
POSITION GENERAL PURPOSE
COVER,BARCODE READER PCB OPTO SENSOR
500479 450006
CHAIN BAFFLE,
1.235 INCH
500048-01
}
500536
GROUND STRAP C O M P L E TE B A R C O D E
400142-03 O-RING IDLER ASSEMBLY
900771 400008
ROLLER
500059
(A TTA C H ED W ITH SC R EW
TO IN STR U M EN T LID LA TC H )
O-RING
900771
(BARCODE READER
MOTOR ASSEMBLY
REPLACEMENT KIT
WITH ACCESSORIES
420056)
L O W E R S E C T IO N O F C H A IN
POSITION SENSOR ASSEMBLY
(REFERENCE)
ISOMETRIC DRAWING,
PRINTED CIRCUIT BOARD LOAD PLATFORM
GENERAL PURPOSE
OPTO SENSOR
MODULE P/N 400199
450006
SHEET 1 of 1
IMMULITE Index
Decontamination, 7-30
Diagnostic Error Codes, 8-27
A Diagnostics, 2-3, 2-6
Ambient Temperature GetDataT, 8-6
Electrical, 3-35 Investigation, 8-1
System, 3-35 Level Sensing Tests, 3-16
Assemblies Main Programs, 8-9
Removal, 7-9 Message Tables, 8-2
Attenuator Motor and Belt PC, 8-8
Removal, 7-18 PCSTAT, 8-8
Power, 8-1
Syringe Module Test Programs, 2-6
B Testing Drawback, 2-2
Using, 8-3
Barcode Drive Wheel
DPC History, 1-1
Motor Speed, 3-9
Drive Wheel
Travel, 3-9
Removal, 7-16
Barcode Laser Scanner
PSC Laser, 7-42
Reprogramming, 7-41 E
Welch-Allyn, 7-43
EC Declaration of Conformity, 5-3
Error Log, 7-46
C Historical, 7-47
Error Message Table, 9-14
Chemiluminescence, 3-4
Euro Syringe Pumps, 3-25
Cirrus Injector Pumps, 3-20
Euro Tubing Kit, 3-25
Common Diagnostics, 2-3, 2-6
EuroTubing Kit, 10-5
Configuration Table, 3-6
Export Utlility Program, 3-45
Accessing, 7-8
Changing Parameters, 7-8
Delineation, 7-5 F
Description, 7-2
Parameters, 7-2 Final Checkout, 6-9
Sample Printout, 7-4 Fixit, 7-45
Configurations, 8-20 FIXIT, 7-45
Diagnostic Error Codes, 8-27 Fludics
CPU/Sensor PCB Description, 3-30
Description, 4-1 Fluidic Valve
Theory of Operation, 4-1 Testing, 7-32
Fluidics, 3-30
D
G
Dark Counts
Viewing, 3-2 General Checkout, 6-2
Database GetDataT, 8-6
Fixing, 7-45
Fixit, 7-45 H
Restoring, 7-45
Database Components HamiltonTubing Kit, 10-5
Replacing, 7-44 Heater and Cooler
DECON1, 7-30 Shutdown, 3-38
INDEX
The IMMULITE 1000 Display Panel buttons are explained in the table below:
Button Function
ALARM MUTE In the Run mode, it silences the audible alarm.
In Diagnostic modes, it is used to pause a motor movement. It
is also used in conjunction with the GO button to cancel a
diagnostic program.
PAUSE Temporarily stops pipetting and Load Chain movements only.
Shuts down the IMMULITE 1000 when pressed repeatedly.
Note: Using PAUSE to shut down the instrument ensures all parts
are in the home position.
GO In the Run mode, it activates the initialization (start-up)
procedure and the accompanying instructional messages
appearing on the Display Panel.
It also activates a reading of the Reagent Wedge barcodes if the
instrument is in pause mode when GO is pressed.
In Diagnostic modes, it activates the GO command for the
loaded diagnostic program. It is also used in conjunction with
the ALARM MUTE button to cancel a diagnostic program.
Fan Filter
Monitor Arm
Note: Different assay kits are required for each mode. Standard (Non-Turbo) kits cannot be run
while in Turbo mode and Turbo kits cannot be run while in Standard (Non-Turbo) mode.
Symbol Definition
Consult instructions for use
Do not re-use
Batch code
Catalog number
Serial number
Sterile
Biological risk
Non-sterile
Do not re-sterilize
Keep dry
Temperature limitation
Control
Negative control
Positive control
Date of manufacture
Manufactured by
European Community
Representative
Upon receipt of the IMMULITE 1000 Analyzer system, visually inspect the exterior of the shipping
container for damage. If there is any external damage to the container, notify your shipping department
and contact Siemens Diagnostics Technical Support.
2.1 Requirements
Electrical
Use of a dedicated, grounded service for the IMMULITE 1000 Analyzer, computer, monitor, and printer
is strongly recommended.
Use of an Uninterruptible Power Supply (UPS) or line conditioner is also strongly recommended.
Do not place the system near centrifuges, ultrasound or X-ray machinery, NMR scanners, or other
sources of magnetic fields.
Space
The IMMULITE 1000 must be placed on a flat, clean, horizontal surface.
The instrument requires a bench space of 75 inches (191 cm) across by 26 inches (66 cm) deep, with
any overhead cabinets at least 28 inches (71 cm) above the bench.
Environment
Ambient temperature should be between 65F - 86F (18C - 30C), Humidity 20-80% (no
condensation).
Do not place the IMMULITE 1000 system in direct sunlight.
Do not install the IMMULITE 1000 near heating or air conditioning vents or equipment.
Arm Bracket
Hardware
Monitor Arm Upper Monitor Pole
Note: Do not turn the IMMULITE 1000 power switches on until instructed to do so.
Use the Water Test Kit (catalog number: LKWT) in conjunction with the WATERTST diagnostic
program to perform the water test.
WATERTST
The WATERTST diagnostic program tests the water from any source, independent of the internal fluidics
of the IMMULITE 1000. Water should be tested before priming water from a new source through the
IMMULITE 1000 System, or when contamination is suspected. The procedure for testing the water
supply using the WATERTST program appears below:
1. Before starting, ensure the IMMULITE 1000 has been powered on for at least 45 minutes and at a
stable operating temperature.
2. Make sure the printer is on-line.
3. From the Start-up screen, select the DIAGNOSTICS button. The Common Diagnostics screen
appears:
5. If not already done in the previous instrument setup process, be sure the substrate is well primed. If
needed, load and run the SOLE2S program to fully prime the substrate nozzle. The IMMULITE 1000
must be Idle before performing this step, or before step 9.
6. Select two empty Water-Test Test Units.
7. Manually pipette 50 L of water directly from the source or from the water bottle into the first Test
Unit only.
8. Load both Test Units onto the Load Chain. Place the Test Unit containing the water at the beginning
of the Load Chain (so it enters the Barcode Reader first).
9. From the Diagnostic Menu, load and run the WATERTST program.
10. Press GO (on the IMMULITE 1000) when instructed to do so. Substrate is dispensed into the Test
Units. The Test Units are then moved into the Luminometer and incubated for 10 minutes.
11. Once the Test Units start moving, select the VIEW button. The following screen appears:
1. The cps (counts per second) for each Test Unit should be:
> 4,500 cps and < 9,000 cps
If the cps is too high or too low, perform the test again to confirm the results.
If the cps is consistently outside the 4,500 cps to 9,000 cps range, find a new water source and test
this new source.
If the second test unit (substrate only) is > 9,000 cps, the substrate system may be contaminated or the
test units used for testing may not be clean. **The Water result is then not valid, as it is being falsely
biased (high) by the unacceptably high level of this baseline substrate only reading
CAUTION: If this water source is determined to be unacceptable, and has already been used on the
instrument, the fluidics system must be decontaminated. Refer to Section 7.4.2: System
Decontamination for information.
Once an acceptable water source is obtained, it can be put into the rinsed water bottle, and the probe-wash
solution can be mixed. Once primed into the system using the PRIME common diagnostic program, clear
all air bubbles in the syringes. The instrument is now ready for chemistry testing.
3.1 Introduction
The IMMULITE 1000 System is a random access immunoassay Analyzer, built around a proprietary
means of efficiently and rapidly washing the bead solid phase. The 0.25-inch polystyrene bead is captured
in a Test Unit that serves as the vessel for all incubations, washes, and signal development. After
incubation of the sample with alkaline phosphatase-labeled reagent, rapid separation and efficient
washing of the Test Unit and bead are achieved by high-speed spinning of the Test Unit on its vertical
axis. The fluid contents are completely transferred to the coaxial sump chamber of the Test Unit. Multiple
washes are accomplished within seconds, allowing uniform, sequential processing of the Test Units and
leaving the bead with no residual unbound label. The bound label is then quantified, using the dioxetane
substrate to produce light. Light emission from the chemiluminescent substrate, reacting with the bound
label, is proportional to the amount of analyte present in the sample.
Note: The materials needed to operate the IMMULITE 1000 are included in the IMMULITE 1000
Test Kits.
2. In normal testing, the operator then loads up to five barcoded Test Units after each sample, in any
order, for the tests desired on that sample. (See figure on page 1-4).
Additional samples and Test Units can be added at any time, and Stats can be placed at the
beginning of the Load Platform.
If on-board dilutions are also being used, with the correct diluent wedge resident on the reagent
carousel, the order of Load Platform tube placement becomes: sample cup, dilution cup, 1 empty
space, followed by up to four random test units.
3. The operator presses GO (on the IMMULITE 1000 Display Panel) and the Test Units are conveyed
into the Analyzer for barcode identification and then moved onto the Main Incubation Carousel.
4. The pipettor adds sample and alkaline phosphatase-labeled reagent. (The Reagent Carousel
accommodates up to twelve resident assays.)
Note: Reagent changes can be made during instrument operation by pressing the PAUSE button,
which stops the Load Chain and the Pipettor. Reagent Carousel will not go into pause if sequential
assays need to be pipetted. Normal operation resumes when the GO button is pressed.
The IMMULITE 1000's attenuation filter restricts the number of photons entering the PMT. This is
used to prevent high light signals from saturating the PMT. Any readings above saturation are not
accurate, so the IMMULITE 1000 uses the attenuated light signal and compensates for the attenuation
via software.
2. For each sample, the IMMULITE 1000 takes a one-second reading dark count in the closed position
and a one-second reading decision count in the attenuated position.
Note: To view the dark counts while in the run mode, select System Status and View Temps/PMT
Counts.
If the one-second attenuated reading is less than 10,240 cps, the attenuator disk moves to the open
position; otherwise, the attenuator disk remains in the attenuated position while the readings are
taken.
3. The IMMULITE 1000 takes twelve one-second readings and returns the attenuator disk to the closed
position.
4. Dark counts are calculated by averaging the last ten reads. This average is subtracted from each of
the twelve readings.
5. To ensure that the final reading is representative, the highest and lowest of the twelve individual
readings are dropped.
6. All PMTs are slightly different in sensitivity. Therefore each PMT has a unique PMT Multiplier
factor. This PMT Multiplier number is multiplied by the cps reading to get the final cps reading.
7. If the counts were measured attenuated, the trimmed mean of the ten individual readings is multiplied
by the instrument-specific attenuation factor to yield the total unattenuated counts per second (cps).
CPS(unattenuated)= CPS(attenuated) x attenuation factor
8. To determine analyte concentrations, the IMMULITE 1000 software refers to lot-specific master
curve parameters, which were entered into the system via the kit barcodes.
1 2 3 4 5 6 7
PMT Pre-scaler Mean of last High and Mean X 10 (for Mean X Mean X
Output Output* 10 dark Low count Pre-Scaler*) PMT Attenuation
(CPS/10) count reads removed Multiplier Factor**
(20 CPS then mean Factor** (only if
used here) calculated (.90 used counts were
subtracted here) taken
from each attenuated)
count FINAL CPS
8000 800 780 780
8100 810 790 790
8200 820 800 800
8300 830 810 810
8400 840 820 820
8500 850 830 830
8600 860 840 840
8700 870 850 850
8800 880 860 860
8900 890 870 870
9000 900 880 880
9100 910 890 890
8350 7515 864225
8350/10 =
835
* Original IMMULITE ONE instruments with serial numbers less than approximately 1900 have an
external pre-scaler. Later instruments including all IMMULITE 1000 Instruments have the pre-scaler
built into the PMT.
** The GetdataT program does not compensate for the PMT Multiplier or the Attenuation factor. It will
factor in a pre-scaler if present.
In the luminogenic reaction (refer to figure below), the substrate (an adamantyl dioxetane phosphate 1 ) is
dephosphorylated into an unstable anion intermediate by the alkaline phosphatase conjugate captured on
the bead. The unstable intermediate emits a photon upon decomposition. The amount of light emitted is
directly proportional to the amount of bound alkaline phosphatase.
Compared to other means of detection, chemiluminescence provides the highest degree of sensitivity
available. In many cases, the sensitivity is orders of magnitude better than that attainable with
radioimmunoassays.
1
*LUMIGEN PPD: 4-methoxy-4-(3-phosphatephenyl)-spiro-(1,2-dioxetane-3,2-adamantane). LUMIGEN is a
registered trademark of LUMIGEN, Inc., Southfield, MI.
O O
OCH3
DIOXETANE PHOSPHATE
(STABLE)
OPO3
O O
OCH3
UNSTABLE DIOXETANE
OCH3
O
O
LIGHT (hv)
3.5.1 Introduction
Primary Modules are defined as those modules that include stepper motors and require configuration
parameters and adjustments. Primary Modules require alignment through the configuration table (see
Section 7.2: The Configuration Table). Secondary Modules (discussed in Section 3.6: Secondary
Modules) are modules that require mechanical or electrical adjustments.
The Primary Modules of the IMMULITE 1000 are as follows:
Incubator Carousel
Load Chain Platform
Luminometer
Pipettor / Reagent Carousel
During the RUN IMMULITE initialization process, the carousel will index before the run to find the first
position indicator. The carousel then stops just beyond the position indicator. The software then checks
for the passing of each of the three position indicators, 20 positions apart from each other. If the sensor
does not see another position indicator after 20 positions a Main Carousel Positioning Error will be
posted.
Note: When running the diagnostic CARPOS, the main carousel position error-checking sub-routines are
activated. A Main Carousel Positioning Error will post if triggered.
The load chain baffles do not inhibit assay tube travel in the transfer block. (The cross-over point
with path of the Incubator Carousel)
2
6 3
8
4 4
2
5
7
5
1
10
3.6.1 Introduction
Secondary Modules are defined as those support modules not directly responsible for tube transport,
and/or smaller modules that require mechanical or electrical adjustments.
7
5 4
6
BKGNDCNT
This program will alternate rotating the attenuator disk to the attenuated and then unattenuated
position. While at each position, a set of 12 one-second background counts will be taken. The button
VIEW must be used with this program to view the PMT data. Normal readings are in a range of 0 40
CPS in attenuated (filtered) position, and comparable readings in unattenuated (open aperture)
position, which will have more values in the 30-40 range, along the 12 readings.
PMTCOUNT
Using the VIEW button, the current PMT counts are displayed. Readings are taken in the current position
of the attenuator, and it does NOT move the disk. Judgment of normal readings is dependant on the
current disk position, as above.
Description
These stepper-motor actuated companion Substrate and water side Pump modules automatically pump
substrate and water into the assay tube during the spin/wash cycle. The pumps interface via tubing to the
water and substrate containers and the Luminometer module. The pumps interface electrically with
motherboard #1.
NOTE: In order to prime this pump most effectively, the upper prime buttons should be held manually
during pre-chemistry priming, or when using the priming diagnostic programs. This creates a fuller
compression of the bellows assembly. When using the DECON and DECON1 decontamination
procedures, it is also advised to secure the prime button of the water side pump using tape, to create the
fullest compression of the bellows.
Note: The temperature of the heater should be at ambient when the program first starts. Multiply the
voltage measured by 100 and this will be the temperature in Centigrade. The stabilized temperature of the
heater should be 36.6 +/- .6C. When done measuring the temperature, be sure to terminate the program
so as not to heat the substrate for a long period of time.
SUBHTTST
This program (located in the system configuration diagnostics menu) will first check to see that the heater
is not at temperature (in the event the heater has shorted and is always on). If the heater is not on, the
program will activate the heater and check every 30 seconds to see if it has reached temperature. If the
heater was already at temperature when the program was started or does not reach the proper temperature
within 6 minutes, the program will indicate the test has failed.
If the proper temperature is reached within 6 minutes, the program will indicate the test has passed. The
heater will only come up to the on/off trigger point of approximately 35C - 35.5C and then shut off. It
will not heat to the full temperature of 36.6C +/- 0.6C.
Note: Since this program checks to see if the heater is always activated, it is very important that the heater
be deactivated for at least five minutes before the program is run or a false failure may occur. Remember
that the heater is only activated when a test unit requiring substrate is seven minutes away from the
substrate dispense position. For example, if this program is run successfully, and then run again
immediately afterward, it will fail since the heater is still at temperature.
V4
5
5 V2
V1 6 and 7
7
A B A B
1 and 2 1 and 2
1. The Hamilton 500 Series Syringe Pumps Assembly has been in production since instruments
starting with serial number H2469 and above. It has been implemented into production alternately
(according to stock availability) with the Euro design described below. The complete module is
available under P/N 400273 and individual diluters are available under P/N 901664. It involves a
circuit redesign where the Driver Circuit PCB and Logic PCB were combined into a single
Logic/Driver PCB, P/N 901666. The prime button is also located to the right of the syringes and the
same IMMULITE 1000 Hamilton Tubing Kit, P/N 400360 and Syringe Interface PCB are used. The
cabling for sensors and motors of the rotary valve and lead-screw drives are all brought to the right
side of the Module and connected to the Logic/Driver PCB. This module also utilizes two rotary
valves; a Macrovalve for the A syringe (identified by a symbol M) and a Microvalve for the B
syringe (identified by a symbol ).
2. The Euro Syringe Pumps Assembly, P/N 400263 is available as a complete assembly only. Individual
diluters cannot be replaced separately. The Euro design uses solenoid valves to direct the flow of
liquid in or out of the syringes. This assembly also utilizes a single large PCB which covers the entire
back wall of the module. It includes all the circuitry of the Driver Circuit, Logic and Syringe Interface
PCBs. The prime button is placed between the two syringes and the side fluidics valves have
connectors (labeled V1, V2 and V4) more easily accessible at the top rear of the module. The syringes
are secured into acrylic manifold blocks where the fluid is controlled by two more solenoid valves in
place of the rotary valve assemblies used with the Hamilton Modules above. Tubing is different for
the Euro Module and is available in a Euro Tubing Kit under P/N 400361.
Note: All types of pump modules are interchangeable, as the electrical connections to the modules are
identical. Tubing kits may need to be changed, if going from Hamilton to Euro or from Euro to Hamilton.
Refer to Chapter 10 for a listing of the tubing kits.
Note: Removing a syringe without first moving the syringe arm down can crack the fitting on the valve
above.
3. Unscrew the syringe from the valve above while taking care not to exert side pressure on the valve
stem.
4. Install the new syringe onto valve fitting above.
5. Press the "GO" button to move the syringes up and tighten the thumbscrews.
Changing the Microvalve or Macrovalve (Hamilton 500 units only)
Note: These rotary valves are changed at the yearly PM, ONLY if they are of the black-bodied style. The
newer white-bodied ceramic rotary valves are NOT replaced, and are intended as permanent.
1. Remove the syringe as described above, using the SCHANGE diagnostic.
2. Remove the tubing on either side of the valve.
3. Move the silver locking lever (below valve) to the right position to release the valve barrel.
4. Above the valve is a hex-cap capture setscrew, which must also be slightly backed off, in order to
release the valve barrel.
5. Remove the valve from diluter assembly.
6. When installing the new valve insure the slot in the valve stem is positioned so that it will mate
correctly with the stem inside the valve drive mechanism.
7. While making sure the valve body is flush to the dilutor manifold, retighten the top hex-cap setscrew
from above, until gently snug.
Warning: Over-tightening the fittings into the valve above can deform the inner body of the valve and
cause air and/or water leaks.
Note: The clear acrylic manifold blocks are not regularly replaced. However, regular inspections should
be performed and the valves replaced if damage or cracks are found.
1. Making sure main power is off, disconnect short power wire of the solenoid valve from the rear
Controller PCB behind the acrylic manifold intended for replacement.
2. Remove three hex screws, which mount the acrylic manifold block to the front face of the module.
3. Remove the entire assembly (manifold and solenoid valve).
4. Install new manifold and solenoid valve in reverse order.
Syringe Module Test Programs
SCHANGE
This program moves both syringes down approximately halfway to permit their removal, waits until the
"GO" button is pressed, then moves both syringes up to home.
PRIME
This program primes the two syringes and the associated tubing. It exercises syringes "A and "B", valves
"A" and "B," and solenoid valves 1, 2, and 4. System error checking is active.
APRIME
This program is used to isolate priming of the "A" large syringe for diagnostic evaluation or to aid in
syringe tip replacement. System error checking is not active.
BPRIME
This is used to isolate priming of the "B" small syringe for diagnostic evaluation or to aid in syringe tip
replacement. System error checking is not active.
Note: Each of the three valve programs below assumes the valves not being tested are working properly.
This means if V1TEST is run and fails, insure V2 and V4 are working properly before replacing V1.
V1TEST
This program should alternately dispense water through the probe and the wash well. The change in
dispense point is altered by the activation or deactivation of V1. If the valve is not working, the water will
always dispense from the probe. If the valve is stuck in the activated position, the water will always come
from the wash well. If one of the two diaphragms in the valve is partially clogged, one port will slowly
drip while the other is dispensing.
Note: All valve testing should be performed substituting a cut length of probe tubing acquired from a
previous PM service, installed in place of the existing system probe. This is needed in order to observe
and evaluate possible movement of the locked liquid level near the tip of this valve test tubing. The valves
should never be tested using the real system probe, as this liquid level near the tip cannot be seen. For
further details of proper valve testing, refer to the Section 7.4.4 in Miscellaneous Repair Procedures.
PROBETST
This program will move the probe into the home well, dispense 3000ul at maximum velocity
(1000ul/second), move the probe up and over to the dump position, and repeat. Because it simulates the
washing of the probe during the main program, it is very useful in testing the integrity of the washing. For
example, if the probe were partially occluded or damaged, the depth too far into the home well, or not
centered in the home well, the water coming out of the probe would be uncontrollable and would "splash"
out of the well. Also an occlusion may be present if dripping from the probe occurs when it momentarily
locates over the dump well.
PROBANGL
This program is used to evaluate the dispense angle of the probe. The probe is positioned over the home
well and water is dispensed.
3 1
Description
1
This module collects the sample cups and holders ejected from the load chain so the user can recover
them for reuse. It interfaces mechanically with the load chain module. The main areas of interest are:
1. Removable tray
2. Tray housing
3. Tray-In-Place and Tray-Full Buffer Sensors
Sample Collection Tray Module Adjustments
No adjustments of this module are necessary or possible.
Sample Collection Tray Module Test Programs
SAMPFTST
This program located in the system configuration diagnostics menu is used to check the sample collection
tray full sensor (located under the Sample Collection Tray). Follow instructions on the IMMULITE 1000
display panel.
SAMPTTST
This program located in the system configuration diagnostics menu is used to check the sample collection
tray-in-place sensor (located under the Sample Collection Tray). Follow instructions on the IMMULITE
1000 display panel.
3.6.9 Water Supply Bottle and Probe Wash Bottle Level Indicators
Description
LED sensors detect the level in the bottles for three main levels of operation. The four conditions of the
LEDs are as follows:
No LEDs lit = Bottle removed
Red LEDs lit = Bottle present but empty
Yellow LEDs lit = Bottle approx. full
Green LEDs lit = Bottle above full
Refer to section 7.4.15 for detailed calibration procedure.
3.7.1 Description
Although most parts of the fluidic system are a part of one of the modules already discussed, a section in
this manual has been devoted to addressing the combination of all fluidic parts as a whole. The
components involved in the fluidic system are as follows:
Tubing
Each piece of tubing that comes in contact with liquid used for test unit processing is Teflon. The shape
of each piece is formed by the application of heat. The tubes may be repaired to create a new clean end
and a new ferrule of the appropriate size is mated to the end of the tube (wider end of ferrule toward the
tubes end). This repair work can be done using the Upchurch Tubing Repair Kit, P/N 420109. Each
tube is numbered and has color-coded fittings for easy tube identification. Color coding also corresponds
to the type of fluid that will pass through the tubing:
Blue - Fresh water from the water container
Red - Substrate
Yellow - Detergent or detergent and water mixture
White - Fresh water used to rinse the wash well
Green - Water to be used in the test unit
Each tube has a fitting that is designed to be gently tightened only with the tightening fitting, supplied
with the repair kit. (The probe has a hexagonal fitting and can be tightened with a wrench). The proper
technique for seating the tubing and tightening the fitting is:
1. Move the fitting away from the flared end approximately 15mm.
2. Insert the flared end into the port (of the valve, coupling, etc.) until it "bottoms."
3. Move the fitting into the port and tighten until it "bottoms."
4. For fittings in valves, tighten the fitting another 1/4 of a turn.
5. For fittings elsewhere, tighten until snug.
Note: Over-tightening the fittings can damage the flared end or ferrule, fitting, or the silicone O-ring of
the tubing (where used). It can strip the port of a solenoid valve, or deform the inner shape of a rotary
valve. If any of the mentioned parts are damaged, they must be replaced for proper fluidics operation, but
tubings of the type known as Upchurch can be repaired using the Upchurch Tube Repair Kit, p/n 420109.
Minimum
Control Range
Maximum
Operating 37.10 37.50 16.5 N/A N/A
Temperature
Lower Alarm
36.80 36.00 N/A N/A N/A
Temperature
Alarm
Range
Upper Alarm
37.20 38.00 22.00 34.00 40.00
Temperature
Minimum
displayable 34.XX 34.XX 11.60 11.60 11.60
Reportable
temperature
Range
Maximum
displayable 39.XX 39.XX 41.XX 41.XX 41.XX
temperature
6013 ohms @ 6013 ohms 15000 ohms 10000 ohms 10000 ohms
Thermistor Resistance
37.00C @ 37.00C @ 16.00C @ 25.00C @ 25.00C
Note: XX stands for the numbers that will vary depending on the circuit components of each specific
Heater-Motor PCB, and its controller circuitry.
Note: If a condition exists that sends the Temperature Controller system into a full shutdown, the error
display box will overlay the existing screen, which will specify the area of trouble which actually caused
the controller circuit shutdown. This is a main part of evidence for proper diagnosis.
If a Heaters/Coolers Shutdown or a Full Temperature Controller Failure condition exists, error messages
will be seen on the display and in the error log.
Note: In the event a short circuit occurs in one of the transistors that controls the power applied to a
heater (resulting in full power applied to that heater); a thermostat will terminate the power to that heater
when the temperature of the heated device reaches 45C. This Runaway Thermostat will again allow
current to flow to the heater when the temperature of the device decreases to 43C. While it is not
permissible to process test units at this elevated temperature, no damage to the Luminometer or Incubator
Carousel heater blankets occurs if these temperatures are reached.
A full Temperature Controller shutdown occurs under one or more of the following conditions:
Luminometer and/or Carousel temperature is less than 35C for more than 60 minutes
Luminometer and/or Carousel temperature is greater than 39C for more than 10 minutes
Reagent Carousel temperature is greater than 30C for more than 30 minutes
Reagent Carousel temperature is less than 15C for more than 10 minutes
Note: The only way to reset the Temperature Controller circuitry after a full shutdown is to turn the
instrument off, wait ten seconds and then power the instrument back on.
Luminometer
Identity TED (Cooler) Carousel Heater Time Hack
Heater
LED
Lum temp is above Car temp is above 35C
Blinking N/A N/A
35C and below 39C and below 39C
@ 5hz
Note: After five cycles of the software acknowledging temperature shutdown, it will recognize this as
a fatal condition, and will automatically shutdown the IMMULITE 1000 software to the Windows
desktop. If the user wishes to avoid this, stop the TEMPS diagnostic on the IMMULITE 1000 before
it has the chance to post five fatal shutdown errors.
On the printout, take note of all temperature numbers, as these were the last true readings at the
time of shutdown:
One of the five parameters is likely to report an abnormal reading and that is the likely
parameter which is the cause of the temperature shutdown.
The rear Temperature Indicator LEDs may or may not help reflect the parameter which is
truly at fault. In the case of the heaters, this is dependant upon the specific controller
circuits of that PCB. HOWEVER, if the TED is the cause of shutdown, the rear LEDs
will show numbers 1 (time hack) and 4 (TED) blinking slowly on/off together, or
alternating. This may also be true for the heaters causing shutdown, but in either case,
this unusual LED behavior confirms the circuits are currently in shutdown.
After gathering ALL the available evidence, then reset Instrument power, and observe the
temperatures after a warm-up, to then observe the improper operation of the troubled parameter.
This can confirm proper diagnosis.
REAR
Part Identifications:
Note: For information on utility programs and procedures, refer to Section 7.4: Miscellaneous
Procedures. For information on Diagnostic programs, refer to Chapter 8: Diagnostics.
Note: The test time is the time the sample cup was read at the Barcode Reader - not the time the Test Unit
was measured by the PMT.
Select ALL to export all data. Select TODAY ONLY to export data generated today. Select
SPECIFIC RANGE to export from a specific date and time range.
Click in the From Date field and select a date from the calendar window, followed by OK.
Click in the From Time field and select a time from the clock window, followed by OK.
Click in the To Date field and select a time from the calendar window, followed by OK.
Click in the To Time field and select a time from the clock window, followed by OK.
Select OK.
Use the UP and DOWN arrow buttons to highlight the Test Type to be exported.
PATIENT NAME
ACCESSION #
4.1 Introduction
This section contains descriptions of the Printed Circuit Boards, or PC Boards. Refer to the Assembly and
Schematic Manual for actual drawings of the boards.
Note: The red DIP switches near the boards top edge do not affect system operation of this PCB, and are
only used for final vendor testing of PCB performance.
Theory of Operation:
This board is identical to the Stepper Motor Driver #1 except for the switch settings of the jumpers at the
top edge.
Note: The red DIP switches near the boards top edge do not affect system operation of this PCB, and are
only used for final vendor testing of PCB performance.
Circuit Description:
The Heater/DC Motor board performs the following control functions for the IMMULITE 1000 system:
(1) temperature measurement and control of the Luminometer, the Main Carousel Incubator, the Reagent
Carousel, the PC board housing, and the ambient temperature within the instrument package, (2) velocity
control of the reaction unit wash operation, (3) velocity control of the sample/assay bar code motor and
(4) power distribution to the solenoid valves and the wash fluid and substrate pumps.
To provide the temperature control critical to the performance of the IMMULITE 1000 system, a
dedicated ZILOG Super 8 microcontroller is used to process temperature measurement data in real time
and in response, control the amount of heat applied to the assay incubator and Luminometer and remove
heat from the Reagent Carousel. Temperature control begins with the precise temperature measurement at
key system positions. A semiconductor Negative Temperature Coefficient (NTC) thermistor with a
guaranteed measurement accuracy of 0.05 oC is embedded in the Luminometer and Main Carousel
When the enable signal is removed, the brake signal is applied to the control IC. This causes the motor
windings to be shorted together. This dynamic braking action brings the motor to a stop more quickly
than simply allowing the motor to coast to a stop.
The Sample/Assay Bar Code Motor circuit is also configured as a feedback control network. An oscillator
internal to the control IC sets the PWM (Pulse-Width Modulation) frequency at approximately 4.8 kHz.
When the enable signal is applied to the control IC, the output of the controller turns on the PNP power
transistor, at the same PWM frequency. This supplies current from the power supply to the motor
winding, causing it to rotate. Because of physical constraints, the motor and gearbox are positioned in a
very small space. There is no room for any type of encoder, which could be used to control the motors
velocity. Thus, the motors velocity is controlled by measuring its back EMF (Electro-Magnetic Field), a
voltage generated by the motor when it is rotating and not energized by the power transistor. When the
controller turns the power transistor off, a 555 timer IC used as a one shot is triggered. This produces a
timed pulse used to enable the sample and hold the amplifier to measure the rotating motors back EMF.
Since the motor is rotating and the power transistor is off, this back EMF is proportional to the speed of
the motor. The magnitude of the back EMF is divided by R1 and R2 to limit its value and is then stored
by the sample and hold (S/H) amplifier. The output is then low pass filtered to eliminate noise and sent to
the inverting input to an op amp internal to the control IC. The other input to the op amp is connected to
an adjustable reference voltage which is used to set the motor speed. The output of the amplifier is used to
set the duty cycle of the pulse width modulator. This way, the amount of electrical energy sent to the
motor windings is proportional to the difference between the set point voltage and the back EMF
generated by the motor. If the load on the motor increases, the speed of the motor will slow down,
unbalancing the amplifiers input and will cause the output of the controller to apply more current to the
motors winding, thus causing its speed to increase toward the set point. If the motor load decreases, the
opposite occurs, the pulse width decreases because the motor requires less energy to maintain its angular
velocity.
Additional Circuits:
In addition to the motor and heat controls, the PCB supports the capability of turning on eight general
purpose loads. For the IMMULITE 1000, these loads consist of several solenoid valves, and the two
injector pumps supplying the system with water for bead washing, and the luminagenic substrate. The
main control CPU controls these devices and determines when and for how long to turn them on.
The following is a table describing the performance characteristics and specifications of the IMMULITE
1000 Analyzer.
Specifications
General Standard Mode Throughput: Up to 120 tests per hour
Specifications:
Time to first result: 42 Minutes
(or 72 minutes for two cycle assays)
Tests per sample: Up to five per sample
(four if dilution cup included)
General Turbo Mode Throughput: Up to 80 tests per hour
Specifications:
Time to first result: Less than 15 minutes for a one-cycle
assay
Less than 22 minutes for a two-cycle
assay
Tests per sample: Up to five per sample
(four if dilution cup included)
Luminometer Detection method: Photomultiplier Tube
Specifications:
Wavelength of Emitted 425 to 500 nm
Light:
Incubation temperature: 37C +/- 0.1C
Internal Computer Computer: Intel Motherboard ProX 1750 with
Specifications: on-board audio, video and
networking
Minimum Req. RAM: 256 MB
Minimum Req. Hard drive: 10 GB
Monitor: 15inch Flat with Touch Screen
1024 X 768 resolution
Auxiliary Drives: 3.5 Floppy / 24X CD ROM Drives
declare under sole responsibility that the following equipment to which this declaration relates, meets the essential
health and safety requirements and is in conformity with the relevant sections of applicable EC standards and other
normative documents. If changes are made to the product which is covered by this declaration of conformity, the
declaration of conformity is no longer valid.
Model: IMMULITE1000
National and other IEC 1010-1, IEC 801-2,3,4, IEC 1000-4-2, IEC 1000-4-4,
standards and technical ENV50140, ISO 9001:1994 (BS EN ISO 9001:1994),
specifications: 21CFR, Part 820 FDA cGMP, EN46001: 1996;
ISO 13485: 1996; UL 3101-1, Laboratory Equipment
employed to hold the documentation and evaluate the safety issues of this equipment.
Date/Signature of manufacturer
or responsible party: ______________________________________________________
Date/Signature of EU representative
or responsible party: ______________________________________________________
LIMITED WARRANTY
Siemens Diagnostics warrants that the items delivered hereunder are of good material and
workmanship, and are free from defects in design and manufacture. Siemens Diagnostics
responsibility is limited to repairing or replacing any item or part, for a period of one (1) year
after delivery to the original purchaser. Defects caused by improper operating conditions,
misuse, negligence or alteration of the product void this warranty. Siemens Diagnostics shall not
be liable for any direct, indirect, incidental or consequential damages arising out of possession or
use of the items. Consumables, as defined in the appropriate Siemens Diagnostics Price List for
Instrument-Related Parts, Racks and Consumables are not covered by this warranty.
CUSTOMER REMEDIES
Siemens Diagnostics entire liability and your exclusive remedy shall be replacement of the
Software that does not meet Siemens Diagnostics Limited Warranty and which is returned to
Siemens Diagnostics. The Limited Warranty is void if failure of the Software has resulted from
accident, abuse, or misapplication.
NO OTHER WARRANTIES
Because the software is inherently complex and may not be completely free of errors, you are
advised to verify your work. The software and related documentation are provided as is.
Siemens Diagnostics disclaims all other warranties, either expressed or implied, including but
not limited to implied warranties of merchantability and fitness for a particular purpose, with
respect to the software and the accompanying written materials. Siemens Diagnostics shall not
be liable for any direct, incidental, or consequential damages arising out of possession or use of
this product.
6.1 Description
Preventive maintenance is an important service function that must be performed at least once a year. This
prevents problems from occurring due to normal wear or adjustments changing over time. The following
Preventive Maintenance Kit (p/n 420140) is available which contains all the parts that may need to be
replaced:
When performing Preventive Maintenance, be sure to use the IMMULITE 1000 Annual Preventive
Maintenance Report, available from Siemens Diagnostics Field Service as a guide to ensure all items are
covered. The rest of this chapter outlines each item in detail.
CAUTION: Make sure the tips are fully seated on the end of the plungers. Gently press the tips straight
down onto a flat surface until a click is felt. The large syringe should also be gently rolled on its tip to be
sure it is straight. Take extreme care not to bend the shaft of the small syringe during this process.
It is recommended to pre-moisten the new large syringe tip with the IMMULITE 1000 water before re-
inserting it into the glass syringe. If resistance is felt, gently pull the plunger back out and re-insert until
there is an even smooth travel upon re-insertion. Do not rotate the plunger when re-inserting, this can
cause the syringe to dislodge again later.
Run the PRIME diagnostic and check for excessive bubbles particularly in the small syringe. If necessary,
disengage the thumbscrew from the plunger during the PRIME and manually move the small plunger
abruptly up and down to clear persistent bubbles. Be careful to prevent any side pull as this can bend or
damage the plunger shaft.
Check operation of V1, V2, & V4: Tests V1, V2, and V4 operation. Refer to Section 7.4.4: Miscellaneous
Procedures for information on checking the operation of the valves.
Check the operation of the micro and macro valves: Run the PRIME diagnostics. Proper operation can be
verified by checking that fluids are running in the proper direction.
The probe tubing must be placed in the notch of the vanity panel with the small rubber O-ring located
behind the panel. Manually turn the pipettor to ensure the tubing doesnt twist or kink. Adjust or flip the
tubing as needed to avoid coiling or wrapping around the pipettor tower.
Check the dispense quality of the new probe. Run the PROBETST diagnostic. Look for any excess liquid
moving up the tip of the probe shaft, or any splashing out of the side of the home well. When the probe
momentarily lifts and moves to the dump well, look for any dripping out of the probe. Dripping indicates
back pressure or occlusion inside the probe or at its 4-way junction block at the syringes
Run the PROBANGL diagnostic to evaluate the angle of dispensing. Verify that the dispense does not
exceed 7 - 10 degrees off center (check angle from various sides of the instrument). Also, check for
excess spiraling of the stream. No more than one spiral should occur per inch.
CAUTION: Be aware of the small rectangle shim located underneath this slide bearing. Be careful not to
drop or lose it. A dab of grease can be used upon re-assembly to help keep it in place.
Remove the slide bearing and rail. Clean off any excess old grease. Use a clean small blade screwdriver to
work new grease into the rails where the ball bearings reside. After re-assembly into the instrument, slide
the bearing back and forth on the rail to work in the grease. Wipe off any excess.
CAUTION: Do not allow the slide railing to completely disengage from the bearing.
Change small idler wheel assembly: Refer to Section 7.3.13: Removal and Replacement of Assemblies
for instructions on the replacement of the small idler wheel assembly. The entire assembly including the
shaft on which the wheel rotates, should be cleaned.
CAUTION: Do not push the button, as this can cause damage to the button.
6.11 Fluidics
Replace water & probe wash filters: Replace the probe wash and water in-line filters (clipped to left rear
instrument wall), using the filter assemblies supplied in the PM kit. The blue marked assembly is for
water, and the yellow marked assembly is for probe wash.
Take careful note of the proper orientation of the replacement filter canisters and related tubing, being
sure the filters mount correctly as per instructions. Refer to Section 7.4.2: Miscellaneous Procedures for
more details.
Replace air filter in caps of water and wash containers: Replace with filters from the PM kit.
Replace waste tubing and check waste path for blockage: Replace with waste tubing from the PM kit.
Check all other waste tubing and manifold block for blockage or clogs. Clean and/or rinse as needed.
Check all tubes for kinks and tighten all fittings: Visually check the tubing for any kinks. Check that all
tubes are finger tight.
Note: Be sure to manually write the instrument serial number, CAF value and PMT value on the printed
output of the configuration table.
6.13 Temperatures
First check the substrate heater temperature. Remove the cover on the substrate heater. Using a DVM set
to DC volts place the leads on the test points of the Substrate Heater PCB. Run the SUBSHEAT
diagnostic. Record the starting (ambient) temperature and final (heated) temperature on the PM report.
The DVM reading is in direct correlation to the temperature. Multiply the measured voltage by 100 and
this will be the temperature in Centigrade. The stabilized temperature should be 36.6C +/- 6 o C. After
measuring the temperature, be sure to terminate the program so that the substrate is not heated for a long
period of time.
Check all other temperatures while running the final chemistry run (the next step in the PM) by going into
the System Status menu, and select the View Temp/Dark Counts option. Record all temperatures on the
PM report.
Note: Controls should be run in five replicates of each level control. Look for the majority of replicates to
be in range. Updated control ranges are issued regularly or can be found in the associated package insert
of the CON6 control module.
Check the database logs after the adjustment and compare the last adjustment for that assay with the
current one run. Ensure there is reasonable consistency between the two.
Ensure the customer is happy with the final chemistry run results and that all issues and questions have
been addressed. Do not leave until final confirmation has been made with the customer.
Not all of the tools listed are readily available outside the United States however, any or all of the above
can be purchased from Siemens Diagnostics.
Note: When facing the instrument, the IMMULITE 1000 serial number can be found on the left side of
the instrument in the upper right corner of the instrument label.
The configuration table is a file located in the C:\IMMULITE 1000\DOWNLOAD folder of the computer
and is always prefixed by the letters IMM followed by an alphanumeric or numeric number. The type
of file is known as speedial. This is the Windows recognition of the DOS .cnf file type. For example,
the configuration table for IMMULITE 1000 serial number D0100 is IMM0100.CNF. The alphanumeric
or numeric number within the file name is part of the instrument serial number. (The serial numbers
prefix letter code is not in the file name.)
Parameters
The configuration table has many parameters, each of which represent an adjustment within the
instrument. Each parameter has a parameter number, parameter name, error number, and extra number.
These are defined as:
Parameter Number: The identification number of the parameter.
Parameter Name: The abbreviation of the specific adjustment the parameter represents.
Example: It is known that the number of steps the load chain motor must take to move from one position
to the next is approximately 550. After adding a tolerance factor, we set the error steps for this parameter
to 600. As long as the Load Chain motor reaches the next position within the 600 step limit, no errors are
reported. If for some reason one of the components that are necessary for a successful move has failed
(motor, sensor, cabling, etc.) or the chain has jammed, the number of steps sent by the microprocessor
will exceed 600 and corrective action will occur.
Notes: 1. For reasons too complicated to explain, the error steps in parameters 32, 33, and 34 do not set
the point of corrective action initiation, they set the depth of the probe in their respective
positions.
2. The error steps for parameters 36-40, and 42-48 are not used and set to zero. These parameters
use the error steps (200) from parameter 8. 3.
The error steps for parameter 35 (Diluent) should stay set at 160, and should not change. Extra
steps are not used.
Parameter Extra Number (commonly referred to as "extra steps"): This is the parameter that sets the
mechanical positioning of all sensors in the instrument. The mechanical components of the instrument
were designed such that when a sensor is triggered by a device driven by a motor, that device then has a
small number of steps more to move to be positioned properly. The small number of steps taken after the
sensor has been triggered, signaling a proper position of the device is defined as the extra steps.
Example: The Load Chain motor begins to move the Load Chain to the next position. After 550 steps the
open slot in the sensor disk reaches the optical sensor and allows light to flow through the sensor
triggering its output. The microprocessor then looks at the configuration table for the number of extra
steps needed to properly position the Load Chain and sends that additional number of steps to the motor.
To put the positioning in terms of coarse vs. fine, think of the mechanical placement of sensors as the
Error Step coarse adjustment and the Extra Steps as the fine adjustment.
0 LcPos Sets the positioning and error detection of the Load Chain. Fixed 35-180
Not currently used (McHome stands for Main Carousel Not Not
5 McHome
home). used used
Error steps are used for error detection of the pipettor when
8 PxPos moving to one of the X-Pipettor positions. Extra steps are Fixed 5-13
only used in a few diagnostics.
The error steps are used to set the minimum sample volume. Not
13 LevSensS 175-205
The extra steps are not used. used
The error steps are used to set the volume (1000 ul) at which
14 LevSensR 420-464 Fixed
the reagent vial is considered "empty" by the instrument.
Sets the horizontal positioning of the probe in the home well Not
36 PX Blind 2-20
of the wash station. used
Note: All circuit boards in the instrument have an A number associated with it. All cables are labeled
with the interconnecting circuit board A number and a jack position J number. For instance, the Reagent
Carousel interface PCB is labeled A12 and there is a ribbon cable labeled A12-J8 that connects to jack
position J8.
CAUTION: The power cord must be removed from system before starting removal of any module.
1. Loosen the thumbscrew on the substrate heater and move away from the carousel ring gear.
2. Loosen the ten slotted screws on the main section of the clear plastic carousel cover and remove the
cover.
3. Remove the probe tubing from the notch on black probe vanity panel.
4. Remove the probe by unscrewing it from the upper pipetting arm and unscrewing the fitting at the
syringe module.
5. Loosen two screws holding the probe vanity panel and remove the panel.
6. Loosen three screws holding the Reagent Carousel cover (two located on top and one located on the
left side of cover close to the chassis). Carefully move cover up and away from module.
7. Remove Reagent Carousel Tray.
8. After removing the front lower chassis panel, unplug the two-pin connector (A12-J5) on the Reagent
Carousel Interface PCB that is coming from the TED Cooler below.
9. Loosen the cardboard collar that surrounds the lower part of the TED fan [keeping it in place].
10. Remove two foam pads (do not discard) covering the top screws securing the TED Cooler from
above. Loosen the two screws allowing the TED Cooler to drop into the lower chassis cavity.
11. Manipulate the TED Cooler from out the front of the lower chassis area if replacement or inspection
of the TED Cooler is desired, otherwise the TED Cooler can remain in the lower chassis.
12. Carefully remove the following connectors from the Reagent Carousel Interface PCB (do not remove
any other cables):
Wide main connector (A12-J6)
6-pin connector (A12-J7)
5-pin connector (A12-J4)
Ribbon cable (A12-J8)
13. Locate and remove the wash well supply tubing (#6) fitting from below the wash well. Remove the
two drain tubes from below the module.
14. Loosen the two screws (keeping them in place) holding the inner sample tube guide located at the
base of the pipettor tower, and remove the inner tube guide with screws.
15. Using a long screwdriver or long nose pliers, free the pipettor tower spiral cable from under the main
carousel casting.
16. Loosen the five captive screws (two located on the left side, two located on either side of the
pipettors positioning sensor, and one located in front of the wash station) securing the module to the
chassis.
17. Grasp module carefully by the pipettor tower on the right and at the carousel casting on the left.
Carefully lift the module straight up from the chassis, being careful to avoid brushing the Reagent
Carousel Interface PCB against the chassis opening.
18. Re-assemble in reverse order. Be careful not to damage to the Reagent Carousel Interface PCB upon
re-insertion into the chassis. Also be sure to re-tuck the pipettor spiral cable connector under the main
carousel casting and make sure the wash tubing #6 is re-connected on the bottom of the wash well.
Reconnect the drain tubing.
CAUTION: Use extreme caution when re-inserting power connector (A15-J1). If this connector is
not positioned correctly, irreversible damage can occur to several logic boards and/or the main power
supply.
CAUTION: It is very important that the main carousel casting below the ring gear assembly is not
removed unless absolutely necessary. This is the very first element that is secured into the
instrument. It acts as the point of reference for all other interfacing modules. If it is shifted or
removed, all modular motor movements will have to be re-aligned and the motor configuration
table changed. Consult Siemens Diagnostics Field Service whenever it is deemed absolutely
necessary to remove the main carousel lower base casting.
9. Reassemble in reverse order. Before re-inserting the Main Carousel Ring Gear Assembly, check that
the Main Carousel positioning sensor is still rotated out of the way. Do not over-mesh the gears of the
motor assembly to the Main Carousel Ring Gear. Use moderate pressure only.
8. The module can now be carefully lifted out of the chassis. Note that the left end is very tight. Be
careful to guide the Barcode Reader PCB cable (A44-J1) out, which is soldered to the interface board.
9. Re-assemble in reverse order. Do not attempt to re-connect cables until assembly is nearly inserted all
the way back into the chassis. Ensure the cable (A44-J1) is properly routed through the lower chassis
before full re-insertion of the assembly.
CAUTION: Be careful not to damage the wires or Interface PCB upon re-insertion through the
chassis cutout.
CAUTION: Ensure power is turned off before starting the removal procedure. Damage to the
Barcode Reader PCB can occur easily.
1. Remove two 9/64 hex screws that are securing the bracket on the top of the Barcode Idler assembly.
Remove bracket.
2. Remove three Phillips head screws securing the Barcode PCB, do not drop any washers into the
instrument.
3. Manipulate the PCB out from its mounts and remove the J1 connector.
4. Re-assemble in reverse order
Note: Barcode Motor Replacement Kit 420056 must be on-hand to perform the following.
1. Remove the Sample/Test Unit Barcode Reader PCB as described in steps 1 - 4 in Section 7.3.7.
2. Lift the entire motor and bracket assembly from the lower pivot point bushing, while holding the
motor return spring, which is inserted towards the rear of the bracket.
3. Loosen the 5/64 locking hex screw of the barcode motor drive wheel and remove the drive wheel
from the motor shaft.
4. Remove the two small 1.5 mm hex screws securing the motor to the top of the bracket.
5. Cut the wire tie securing the motor wires to the motor body. Trim back the heat shrink tubing of the
wires to expose the connection point with the power cable, which is routed through the small hole in
the bracket.
6. Cut the motor wires from the power cable. Strip new ends to the power cable. Slip a new piece of
shrink tubing (supplied with kit) onto the power cable and through the bracket hole.
7. Solder the connections of the new motor wires, insulating one from the other with electrical tape.
Slide the new shrink tubing over the solder joints. Ensure connections are protected.
8. Use a heat gun to carefully shrink the tubing around the solder joint and through the bracket.
9. Remove the DC Motor/Heater PCB (A2).
10. Replace the Q17 driver FET with one supplied in the kit with a solder iron.
11. Re-assemble the motor to bracket and secure wires to motor body with new wire tie.
12. Re-assemble the Sample/Test Unit Barcode Reader Motor in reverse order. Carefully re-insert the
drive motor return spring.
CAUTION: When re-securing the drive wheel to the motor shaft, be sure to align the screw with the
flat of the motor shaft. Also ensure the drive wheel is at the proper height so as not to rub the motor
mounting screws below or to contact the Barcode Reader PCB opto-sensor above.
Note: Entire roller assembly should be visually inspected and cleaned. Look for spilled patient samples
throughout the area and on the top of the load platform casting where the sample cups travel.
4. Prepare the new O-ring and drive wheel by gently moistening them with water.
5. Use two fingers to partially stretch the O-ring around the drive wheel to insert the new O-ring.
6. Just before the O-ring is completely around the drive wheel, carefully insert a small blade screwdriver
between the O-ring and the far side of the drive wheel. Use the screwdriver to pry the remaining part
of the O-ring onto the drive wheel. Do not remove the screwdriver yet.
Note: Once the screwdriver is removed, it is not possible to re-align the seam of the O-ring. If the
seam is not straight, it must be replaced and the procedure started over again.
7. Gently rotate the screwdriver around the drive wheel 5 to 6 times. Visually check that the O-rings
molding seam is straight around the drive wheel without spiraling. Carefully, remove the
screwdriver.
8. Manually press the O-ring into place around the circumference of the drive wheel. Inspect for a
straight seam with no high or low areas on the O-ring circumference.
9. Re-assemble in reverse order.
1. Carefully remove single connector behind the pump, leading from the Y cable from rear of each
pump.
2. Move the shuttle assembly manually to the far left, to the home position at the Main Carousel.
3. Remove front output fitting of the red substrate side and the two rear input fittings from the Wash
Pump Assembly (keep the Water Pump output fitting at front connected).
Note: After removing tubing from the substrate side rear input, raise the tubing and tape it to the
Substrate Reservoir Assembly or bottle, to prevent leakage of substrate.
4. Using a Phillips screwdriver, loosen the two captive screws at the rear edge of the base of the Wash
Pump Assembly.
CAUTION: Do not let the screwdriver shaft damage the pumps or the PMT behind.
5. Lift the Injector Pump Assembly straight up and out of the instrument.
6. Re-assemble in reverse order.
Note: After removing the tubing from the substrate pump input side, raise the tubing and tape it to
the Substrate Reservoir Assembly or bottle, to prevent leakage of substrate.
3. If present, disengage the bottom drain tubing of the Substrate Reservoir Assembly.
4. Loosen the single captive screw at the base of the Substrate Reservoir Assembly.
5. Remove the entire assembly from the platform being aware of the slot that is cut into the base at the
rear of assembly.
6. Re-assemble in reverse order.
1. Remove the Substrate Heater Assembly and the Substrate Reservoir Assembly.
2. Remove the single screw and the tube guide at the top of the black plastic cover of the Luminometer
Module and carefully manipulate the cover off and out of instrument.
3. Manually move the shuttle from the left side, about two inches in toward the Luminometer.
4. Disconnect the 8-pin connector (A10-J4) from the Luminometer Interface PCB.
5. Loosen four captive screws located at the four corners of the High-Speed Spinner Assembly.
6. Remove the entire spinner platform assembly including the High-Speed Spinner motor and idler
wheel.
Motor Assembly removal from platform
7. Remove the three 9/64 hex screws (in a triangle format) holding the High-Speed Spinner Motor and
Drive Wheel.
8. Gently separate the upper cylinder bracket from the motor assembly below. The Spinner platform is
sandwiched in-between.
9. Remove the cable clip holding the cable going to the High Speed Spinner Motor.
Motor removal
10. Orient the assembly so that the drive wheel is facing toward the top.
11. Locate the motors spring coupler in the center of the assembly. Loosen the two .050 hex screws of
the coupler closer to the drive wheel.
CAUTION: Do not loosen the two hex screws closer to the High Speed Spinner Motor.
12. Loosen (but do not remove) the four 5/64 hex screws securing the High-Speed Spinner Motor
Assembly. Access these screws through two holes found in the Drive Wheel Assembly. Drop the
motor out from the bottom.
CAUTION: Keep the assembly upright and do not remove the screws from the bracket since re-
insertion of these screws is very difficult.
Note: Use a ball end hex key to remove the two 2 mm hex screws.
6. With the screws loose, carefully remove the slide bearing and rail off of the shuttle assembly.
CAUTION: Do not allow the bearing assembly to completely slide off the rail.
7. Clean and/or lubricate the slide bearing and re-assemble in reverse order.
CAUTION: Do not lose a small thin rectangular shim located towards the rear of the bearing
assembly. The shim is necessary for proper alignment and performance of the upper shuttle
assembly. A dab of grease may be applied to the shim to hold it in place during re-assembly.
1. Remove the Tube Shuttle Upper Slide Bearing Assembly referring to steps 1 - 5 in Section 7.3.12.
2. Using a small-blade screwdriver partially spread the small retaining-clip at the top of the Idler Wheel
shaft, until the clip is removed from the catch groove.
3. Once the tiny clip is removed, the silver cap above the Idler Wheel can be removed and should be
cleaned.
4. Below the silver cap is a small wave-washer. This is replaced (P/N 900261) as part of the Shuttle
Idler Wheel replacement kit (P/N 420001).
Important: The wave washer must be re-installed with the bowl of the washer facing up.
5. After removal of the cap and wave washer, the Idler Wheel can be pulled off the shaft and cleaned or
replaced. Once removed, the opportunity should be taken to carefully clean the entire area of the
upper shuttle body assembly around the idler wheel.
6. Re-assemble in reverse order with the wheel on the shaft and re-capture, using a new retaining-clip
(P/N 900262) of the Shuttle Idler Wheel replacement kit (P/N 420001).
CAUTION: Power must be turned off before attempting to remove the PMT.
4. Locate the black metal bracket ring located at the end of the PMT. Remove the three 7/64 hex
screws that hold this ring in place. Carefully, slide the ring away from the PMT.
5. Gently shake the PMT back and forth and outward to facilitate removal.
CAUTION: After the PMT is removed, even with the power off, caution must be taken to not
expose the PMT to ambient light, as damage can occur. Carefully, tuck the PMT face into the back
corner of the instrument and cover if necessary. DO NOT TOUCH THE PMT LENS!
6. Remove all the connections from the Luminometer Interface PCB (A10). Release the five plastic
standoffs that secure the A10 PCB to the top of the Luminometer.
7. Lean the PCB back and off the rear of the Luminometer cover. The PCB cannot be totally removed
because the other wires are soldered to the right edge of the PCB.
8. Remove the two sensor wires from the attenuator home and positioning sensors found on the right
side of the Luminometer cover. Properly mark the connectors to facilitate re-assembly.
9. Loosen the five 9/64 hex screws located around the lip of the upper Luminometer cover. There are
three screws along the left edge of the cover, one at the front center and one in the rear right corner.
Tool extensions may be necessary. The rear right corner hex screw has a wire clip under it that
secures a number of sensor wires. Keep the hex screw inserted in the clip for easier re-assembly later.
10. Read all four cautions listed below before proceeding to the next step.
CAUTIONS: 1.When attempting to lift the Luminometer cover off, some resistance may be felt
due to tightness within the locator pins. It may be necessary to gently pry the
cover free of its locator pins with a screwdriver.
2. The Luminometer cover may be trapped along the left edge under the plastic
casing of the lower Main Carousel casting. Gently flex or pry the Main Carousel
casing to free the Luminometer cover.
3. Do not damage the Luminometer Temperature Thermistor, which remains
screwed into the top cover, located behind the Chain Drive Motor Assembly.
4. Do not touch the attenuator lens located on the disk assembly within the
Luminometer cover. Maintenance on this assembly is rarely needed. For further
details on this assembly, please contact IMMULITE 1000 Technical Support.
11. Grasp the motor assembly from the top and carefully lift the cover off the module, while pivoting the
Luminometer Interface PCB out of the way. The Luminometer chain is now exposed.
CAUTIONS: 1. The attenuator belt tightness is subjectively a 1/8 to 1/4 deflection when the
central portion of the belt expanse is pressed with moderate pressure. A belt that
is too tight (little or no deflection) can result in premature wear of the bearings in
the attenuator disk shaft within the Luminometer cover. It may also result in a
seizure of the attenuator disk assembly.
2. Before reassembling the Luminometer cover onto the base, check the
alignment of the Luminometer motor coupling with respect to the Luminometer
chain coupling. If these are not aligned, the cover will not fully seat. If upon re-
assembly the Luminometer cover still does not sit down firmly (instead it rocks
back and forth), move the Luminometer chain manually. This will re-engage the
Luminometer drive cog with the drive motor coupling from above. Once re-
engaged, the Luminometer cover should then set into place flatly. Ensure that
there are no loose wires pinched under the cover.
CAUTION: Ensure J1, J2, and J3 wire harnesses are re-routed as they were. When re-inserting the
power supply do not damage any harness wires on sharp chassis edges above. The Power Supply
Module should slide in easily. If there are any problems with re-insertion, pull the Power Supply
Module out and check the wire harness routing.
Note: The 24v module in the center of the drawer can be adjusted using the potentiometer module
p/n 422268. All instruments s/n C0230 and up have this in place. In IMMULITE 1000, this voltage
should be adjusted no lower than 24v.
5. If not already done in the previous instrument setup process, be sure the substrate is well primed. If
needed, load and run the SOLE2S program to fully prime the substrate nozzle.
Note: The IMMULITE 1000 must be Idle before performing this step, or before step 9.
6. Select two empty Water-Test Test Units.
7. Manually pipette 50 L of water directly from the source or from the water in the water bottle into the
first Test Unit only.
8. Load both Test Units onto the Load Chain. Place the Test Unit containing the water at the beginning
of the Load Chain (so it enters the Barcode Reader first).
9. From the Diagnostic Menu, load and run the WATERTST program.
10. Press GO (on the IMMULITE 1000) when instructed to do so.
11. Substrate is dispensed into the Test Units. The Test Units are then moved into the Luminometer and
incubated for 10 minutes.
12. Once the Test Units start moving, select the VIEW button. The following screen appears:
CAUTION: If this water source is determined to be unacceptable, and has already been used on the
instrument, the fluidics system must be decontaminated. Refer to section 7.4.2 System
Decontamination for information.
WATERTPM
The WATERTPM diagnostic program tests the water dispensed from the probe and the water pump.
Use this procedure when contamination within the system is suspected.
1. Check to make sure the daily maintenance has been performed and that the substrate and the Wash
Pump water lines have both been sufficiently primed.
2. Make sure the printer is on-line.
3. Select four empty water Test Units.
4. Load all four water Test Units onto the Load Chain.
5. From the Start-up screen, select the DIAGNOSTICS button. The Common Diagnostics screen
appears.
6. Select Diagnostics, then select WATERTPM and select Run.
7. Press GO (on the IMMULITE 1000) when instructed to do so.
The following process occurs:
Water from the probe is pipetted into the first Test Unit.
Water from the wash line is dispensed into the second Test Unit.
Water from the second Test Unit is pipetted into the third Test Unit.
Substrate is added to the first, third, and fourth Test Units.
The first, third and fourth Test Units are incubated in the Luminometer for 10 minutes and
readings are taken. The second Test Unit is indexed off the Main Carousel and onto the Load
Chain for easy removal.
Note: The first result is for the water from the probe, the second is for the water from the water
pump, and the third is for the Test Unit containing substrate.
10. When the test result appears, the following are further choices for this data:
Select the Print button to print this screen once the data appears.
Select the Save Data button to save the data to a disk.
Select the Retrieve Data button to retrieve data from a disk.
Select GetDataT to obtain more detailed raw WATERTPM data, stored from past tests. As
with the previous test, the following screen appears:
ACCEPTANCE CRITERIA
The criteria for acceptable water tests are listed below.
1. The cps (counts per second) for each Test Unit should be:
> 4,500 cps and < 9,000 cps
CAUTION: If this water source was used, the system must also be decontaminated.
IMMULITE 1000 Service Manual 600797-A1 7-31
2. For WATERTPM, refer to the last two lines printed with results:
Probe Water Substrate Only = XXX CPS
Pump Water Substrate Only = XXX CPS
If the value of XXX for either equation above is lower than 650 CPS, insufficient substrate may have
been dispensed into one of the Test Units. This could be due to a clogged nozzle. Visually check all
Test Units to determine if the liquid level is equally distributed between the Test Units. If the value of
XXX above is above 2000 CPS, perform the test again to confirm these results. If the CPS is
consistently above 2000 CPS, decontaminate the system. Refer to section 7.4.2 System
Decontamination on the next page.
Sample of Acceptable Water Test Results
Acceptable example results for WATERTPM and WATERTST tests are shown below.
1. Put the IMMULITE 1000 in idle by pressing ALARM MUTE and GO.
2. Obtain the following supplies:
Probe Wash (Part No. LPWS2)
2 In-Line Water and Probe Wash Filter Assemblies (Part Numbers 400364 & 400365 )
Gloves
Sterile cotton swabs (Q-Tips)
3. Empty the waste from the waste container.
4. Remove and discard the water filter canisters from the rear left corner wall.
5. Remove, empty and set aside the water container.
6. Remove, empty and clean the Probe Wash Bottle with a flushing of 70% Isopropyl Alcohol, followed
by at least 5 times repeated rinsing with acceptable IMMULITE 1000 water, then a fresh (prepared
10X dilution) solution of probe wash.
7. Swap the blue water lines into the new container containing the probe wash.
8. From the Start-up screen, select the DIAGNOSTICS button. The Common Diagnostics screen
appears.
9. Select Diagnostics, then select DECON and select RUN.
10. Press GO (on the IMMULITE 1000) when instructed to do so.
11. Run the program for ten minutes to ensure the probe wash has thoroughly penetrated the system.
12. Press ALARM MUTE and GO to stop the DECON program. (Best to stop when Syringes are in a
down position, in order to maximize decontamination within the syringe barrels.)
13. Let the system sit for a minimum of thirty minutes. During this time, continue with steps 14 -16.
14. Place 100mL of 70% isopropyl alcohol into the original water container. Place the blue cap and inner
water lines into the original water container and tighten the cap. Shake well to ensure the alcohol
contacts all surfaces in the container, including the cap.
15. Discard the alcohol and rinse the water container and cap with distilled water at least five times (with
the cap on) to ensure all surfaces have been thoroughly rinsed.
16. Install new water filters, as per instructions in the in-line filter replacement kit. Be sure to take note of
the proper orientation of the filters, which both show fluid travel to go upward.
17. Refill the water container with fresh IMMULITE 1000 acceptable distilled water, and replace into
bottle chamber.
18. Using a sterile cotton swab, wipe the accessible surface of the wash station to help remove any
growth.
Note: Prior to filling the container, the water must be tested using the WATERTST diagnostic program.
From the same Diagnostics Menu, select DECON1, and then select RUN.
Press GO on the IMMULITE 1000 when instructed to do so.
The program is finished automatically, in approximately fifteen minutes.
20. Re-test the system for contamination using the WATERTPM diagnostic program to verify the
system now meets the proper criteria.
CAUTION: The O-ring may stick to the bottom of the vial support. If this occurs, be sure to
remove it. (Normally, it rests on the spike tip.)
Note: Keep the tip guard on and do not to touch the tip of the spike, since this could cause
contamination.
Note: All valve testing should be performed substituting a cut length of probe tubing acquired from a
previous PM service, installed in place of the existing system probe. This is needed in order to observe
and evaluate possible movement of the locked liquid level near the tip of this valve test tubing. The valves
should never be tested using the real system probe, as this liquid level near the tip cannot be seen. For
further details of proper valve testing, refer to the Section 7.4.4 in Miscellaneous Repair Procedures.
Remove the probe line fitting from the 4-way block. Attach a discarded probe, having the metal jacket cut
off and the outer tubing removed, to the 4-way block. Run the tests below and look for the appropriate
water movement at the end of the tubing. Water movement that occurs as the valve actuates or de-actuates
is normal, but should be no more than 1/4 inch.
1. Valve #1 Testing
Run the V1TEST diagnostic. The water should alternately dispense from the "probe" and from the
wash well. Except for valve throw, the water should never move in the tubing when the water is
dispensing up from the bottom of the wash well. If water is moving down the probe when it should
not be, this can cause under-aspirating of reagent and/or sample. It can also cause fluid to be drawn
into the probe from the home well resulting in carryover.
2. Valve #4 Testing
Run V4TEST diagnostic. The water should alternately dispense from the probe and from the wash
well. Except for valve throw, the water should never move in the tubing when the water is dispensing
up from the bottom of the wash well. If water is moving down the probe when it should not be, this
can cause under-aspirating of reagent and/or sample. It can also cause fluid to be drawn into the probe
from the home well resulting in carryover.
3. Valve #2 Testing
Run the V2TEST diagnostic. The water should alternately dispense from the probe and the tubing
that goes into the probe wash container (that has been removed from the wash container and placed
into a container or the wash station for this test). Follow the instructions given on the IMMULITE
1000 Display. Except for the valve throw, the water should never move in the probe tubing when the
water is dispensing from the disconnected Probe Wash tubing.
4. Fluid Flow Testing
Run the PRIME diagnostic. During dispense, the 5000 ul syringe stops momentarily to switch a
valve. During this moment the water should come out in drips only since only the 250 ul syringe is
moving. If not, there is back pressure building in the fluidics which is being released during this
moment. This back pressure could be the result of a partially occluded piece of tubing, air in the lines,
or a piece of tubing that is expanding excessively from the pressure during dispensing. When the
water is finished dispensing, make sure no water backs up in the tubing. Water flow back-up in the
tubing is indicative of a solenoid valve failure.
5. Valve Leakage Testing
Secure the tubing to the Main Carousel shroud with tape. Then place the wash bottle on top of the
instrument and check for fluid movement. If none occurs, place the wash bottle at tabletop level and
once again check for fluid movement.
Water movement in the tubing (when the tubing is secured) is indicative of a Valve # 2 failure. This
type of failure usually results in the first test that is run after the pipettor has been idle for a while to
3. Test the volume of the pump by using DISUB. The volume should be 200ul 10% for both pumps.
Volume and Drawback Specifications
The drawback adjustment adjusts the timing between the throw of the valve and the actuation of the
stepper motor-driven bellows. Refer to the instructions below for specific drawback adjustments that
relate specifically to the Linear Actuator Pump Assembly (p/n 420120).
For both the Water and Substrate pumps, the standard Drawback specification is for , +/- .
Important Cautions:
-- When priming before chemistry, be sure to press and hold the prime button down for each full cycle.
This creates a more proper full extension of the bellows, for a better, quicker prime.
-- When running decontamination programs DECON & DECON1, it is suggested to tape down the
water prime button, for the same purpose as above.
-- Be sure to always have the Instruments Power off whenever removing or connecting power cable to
pump. -- Doing so with power on will cause instant and permanent damage to pumps electronics!!
CAUTION: The customers probe must be removed and a straight spare dummy probe must be used
during this procedure. Failure to use a dummy probe may cause irreversible damage to the customers
probe.
1. Insert the larger gauge (representing the test unit) into the carousel ring gear position where the load
chain crosses the carousel ring gear.
2. Manually rotate the carousel ring gear three spaces back towards the shuttle.
3. Initiate and load the CARCHECK diagnostic from the main diagnostic menu. Refer to Chapter 8:
Diagnostics for more information on the initiation and loading of diagnostics.
4. After pressing GO, watch the test unit gauge increment one space at a time alternating with the
shuttle.
5. Press and hold the ALARM MUTE key and then press the GO key when the test unit gauge arrives
at the dispensing position one step beyond the load chain.
6. Initiate and load the LCHAIN diagnostic. Press GO and allow the load chain to increment two steps
to be sure it is properly indexed. Press the ALARM MUTE button and the GO button to stop the
diagnostic.
7. Place the smaller gauge (representing the sample cup) at the left most (number 5) sampling position at
the Star Wheel (See next section for Star Wheel details.)
8. Initiate and load the PIPXPOS diagnostic. After pressing GO the pipettor will automatically begin at
the home position and then move to each position with a brief pause at the bottom (used to judge
alignment). It automatically increments through the three wash positions and the reagent position. In
each position, use the ALARM MUTE key (holding it down) to pause the diagnostic when the probe
is located in the fully down position. Note any positions that are not within one probe width of center.
9. After the reagent pipettor position, the probe will continue clockwise to the S5 position, lower onto
the gauge and pause. Observe the probe position onto the gauge. The pause down onto the gauge is
brief, use the ALARM MUTE button (holding it down) to pause the diagnostic when the probe is
located in the down position. After judging the position, releasing the ALARM MUTE button will
continue the movement clockwise to the S4 position, and so on. Move the gauge between positions as
needed and observe each of the five sampling positions (which are the most crucial of the whole set).
Note any positions that are not within one probe width of the center target of the test gauge.
Leading edge
Will encroach slightly into
Carousel casting
Trailing edge
Should be flush with outer
wall of Carousel casting
Sampling
Position #1
Space of
tube on
chain
Note: Before checking the star wheel position, ensure the Load Chain position is correct as shown on
previous page.
The correct position of the star wheel is determined by the following two points of reference:
CAUTION: Only loosen the nut far enough that the star wheel rotates freely. Do not remove the nut.
Re-assembly after the nut is removed can be very time consuming.
Note: Instrument s/n C0493 and above have a newer design to a hex-screw clamp collar.
4. Manually, place one sample cup in sampling position number 5 and another in the Load Chain
position just past the star wheel. Rotate the star wheel to the position where contact with the sample
cup in sample position number 5 is with the right star wheel tooth, and the other sample cup is
perpendicular with the opposite star wheel arm. Refer to the illustration above.
5. Once proper alignment is performed, carefully hold the star wheel in position and tighten the nut on
top, being careful not to change the star wheel position.
6. Repeat steps 1 and 2 to verify alignment. If necessary, repeat the adjustment.
7. Finally, check the adjustment under a high load condition. Fill the Sample Collection Tray with
sample cup holders to nearly full. Repeat step 2 watching for proper travel with the high load
condition.
pipettor base
1. Using a Phillips-head screwdriver, loosen the two screws securing the inner sample tube guide.
2. Use a standard piece of paper folded in fours as a gap tool. Gently slide the folded piece of paper in-
between the inner sample guide and the pipettor base.
3. Using a gentle even pressure onto the center point of the inner guide, tighten the two screws down
and remove the piece of paper.
4. Inspect for an even gap at each end of the inner sample tube guide against the pipettor base.
5. Using a Phillips-head screwdriver, loosen the two screws securing the outer sample tube guide.
6. Insert sample cup holders upside-down in between the inner and outer sample tube guides at each
end.
7. Apply gentle pressure on the center of the outer sample tube guide against the upside-down sample
cups and tighten the two screws down.
8. Inspect the adjustment by riding an upside-down sample cup through the space of the two sample
tube guides. There should some slight play without any resistance or obstruction throughout the entire
pathway.
9. Perform a final test by placing 5 to 10 sample cup holders on the Load Chain and run the LCHAIN
diagnostic from the Diagnostic Screen. View the sample cups as they move through the sample tube
guide area. There should be no tilting of the sample cups or any resistance or obstruction seen.
Note: If using sample cup holders the range is 800 msec. - 840 msec., since the circumference is
larger.
4. Press and hold the PAUSE button to index the next test unit. Release the PAUSE button once
indexing starts.
5. Repeat the above adjustment for each of the remaining three positions around the Barcode Idler
Assembly.
Note: The four positions around the Idler Assembly may differ from each other as much as 40
ms/rev. This is acceptable. If the differences are greater than 40 ms/rev, inspect the condition of the
rollers, O-rings, pins, and main drive O-ring. Refer to Section 7.3.7 for more details on the
inspection of these parts.
Cautions: Be aware of false readings. These are tube-specific, based on the different mode in
which this test uses the reader circuit. False readings are ranged 200 440 ms/rev, and are known to
be false, as the reader motor cannot rotate more than 500 ms/rev. When these false readings are seen,
it is suggested to try using a test unit of a different assay, until 4 tubes are found which remain
stable, when rotating in the reader. The barcodes of some assays may randomly hinder accurate
readings of these speed tests.
Also note that if numbers are jumping between real and false, note the audible tone of the barcode
motor. The jumping numbers are known to be false, as usually there is no corresponding variation in
the tone of the motor.
Note: Regarding IMMULITE 1000 s/n D0693 and above, computers are going out of
Manufacturing with 1750 Motherboards. The Ghosting CD Kit needs to be for a 1750
Motherboard. Do not Ghost a 1650 Computer with a 1750 Ghost CD. Likewise, do not ghost
1750 Computer with a 1650 Ghost CD. This kit now has both 1750 and 1650 Ghost CDs
Below this level, within the IMMULITE 1000 folder, a sub-folder structure similar to the following is
seen, (the Databases sub-folder is highlighted:)
Also under the previous IMMULITE 1000 level, the Backup folder is seen, where the Retains can be
accessed:
The systems motor configuration table file (a Speedial file) can be found in the IMMULITE
1000\Download folder.
While in Run mode both error logs can be accessed through the System Status drop down menu of the
upper toolbar. The events that are logged to this file are as follows:
1. All Log on with operator initials and all Log off (with NO operator initials)
2. Any mechanical or temperature related errors encountered in the instrument
3. Assay adjustment information (posts only that a kit has been adjusted.) See the adjustment log for the
detailed slope and intercept of the posted kit adjustment.
The full historical event error log is continual and will indefinitely hold the above information even
when the software has been updated.
Note: The historical error log file is copied into the backup folder whenever the auto-backup performs
any time after Log-Off is chosen.
When a Yellow Box or Red Box fatal error occurs, an audible alarm sounds from the IMMULITE
1000 Display Panel, and the message is simultaneously printed, and written to both the Session and
Historical Error Logs.
Select the PREVIOUS and NEXT buttons to page through the errors.
Select the FIRST PAGE and LAST PAGE buttons to move to the beginning and end of the
error list, respectively.
Select the PRINT button to print the Session Error Log.
The Session Error Log is accessible at any time during a chemistry run. It may be used to determine the
status of a ghost Test Unit or many other error conditions.
This log is active from Log-On to Log-Off. All errors are deleted from this log at the time of Log-Off,
but they are not deleted from permanent record. They are also stored permanently in the Historical Event
Viewer.
The procedure on the following pages provides instruction on how to use the Event Viewer screen to
access information regarding specific events.
Select ALL to view the entire event history. Select TODAY ONLY to view the current event
log. Select SPECIFIC RANGE to view the events from a specific date and time range.
o Click the From Time field and select a time from the clock window, followed by OK.
o Click the To Date field and select a date from the calendar window, followed by OK.
o Click the To Time field and select a time from the clock window, followed by OK.
Select OK.
When choosing the Event Number Range, click ALL to view all events or SELECTED
RANGE to view specific events (for example, Pipettor Tip Jam errors).
If SELECTED RANGE is chosen on the event viewer screen, enter the event numbers
which appear on the event messages for the desired events in the From and To fields.
When choosing the Severity Level, select ALL to view all events.
Select the VIEW button. The Event Log screen appears, listing the events which meet the
criteria chosen.
In the vast majority of cases for any drive or motor issue, historically the problem has rarely been the
motor. They are much heartier than the application they are used for. Any strange activity of the motor is
more likely to be a fault for its support diodes on the module interface PCBs.
Testing the resistance of the motor windings is as seen in the small chart above. If readings are in spec, it
is assured the motor has no issue.
The purpose for the diodes is to protect the driver circuits on the Motor Driver PCBs #s 1 and 2 from
spikes that occur from collapsing EMF from the motors.
The purpose of the PTC devices is to protect the drive circuitry in case a short occurs in the motor. The
PTC device will heat up due to higher current, and the resistance will increase, therefore keeping the
current down, limiting damage that can occur.
Note: As of IMMULITE 1000 serial number D0640 produced in January 2005, all these devices have
been removed and hard-jumpered. In older systems where these devices are present, if they break or
become damaged, simply remove the device and solder in a jumper.
After power up, please allow at least 5 minutes for the sensors to be at temperature before calibration.
Also make sure no sunlight is over the bottles. Connect the common lead of a voltmeter to TP5 (ground)
on the display PCB.
Water Bottle Sensors. Place one of the empty water bottles with the empty probe wash bottle into the
housing. Calibrate the lower and upper water sensors (TP1 with R10 and TP2 with R6) at 3.20 volts and
then check that the Empty voltages of the bottom and upper sensors of the second water bottle* read 3.2
+/- 0.4 volts (2.80V to 3.60V). Adjust if necessary. Use table 1.
Fill the water bottles with distilled water and check that the bottom and upper sensors read between 1.1V
to 1.9V. Then remove the water bottle to verify that the no-bottle voltages are less than 0.8V. Place the
water bottle back into the housing. Verify the operation of the water display LEDs.
Probe wash Bottle Sensors. With the empty probe wash bottle, calibrate the lower and upper sensors
using TP4 with R25 and TP3 with R20 to 3.3V. Refer to table 2. Fill the bottle with probe wash and
verify the voltages are between 1.1 to 1.9 volts. Remove the bottle from the housing and verify that the
voltages are less than 0.8 volts. Verify that the operation of the probe wash display LEDs is correct.
Table 1
Water Bottle/Test point/Pot Empty Liquid No-bottle
1 TP1 (bottom sensor) R10 Set at 3.20V ________ (1.1 to 1.9) ______ (< 0.8)
1 TP2 (upper sensor) R6 Set at 3.20V ________ (1.1 to 1.9) ______ (< 0.8)
2* TP1 (bottom sensor) R10 _______ (3.2+/- 0.2) ________ (1.1 to 1.9) ______ (< 0.8)
2* TP2 (upper sensor) R6 _______ (3.2+/- 0.2) ________ (1.1 to 1.9) ______ (< 0.8)
* Second Water Bottle
Table 2
Probe wash Bottle/Pot Empty Liquid No-bottle
TP3 (upper sensor) R20 Set at 3.3V ________ (1.1 to 1.9) _______ (< 0.8)
TP4 (bottom sensor) R25 Set at 3.3V ________ (1.1 to 1.9) _______ (< 0.8)
8.1.1 Investigate
Discuss past performance with customer: Customer communication ensures that all problems experienced
by the customer are covered. Proper communication helps guarantee customer satisfaction is achieved after
you leave. Ask for any specific situations that the customer may have seen or heard on the IMMULITE
1000.
Check Error Log: Review at least two months of the log. Check for consistent error messages. They could
be warning signs of impending problems. Also, do not be concerned with individual isolated messages.
They may simply indicate an operator error.
Check for IMMULITE 1000 upgrades: A list of current upgrades and standards are available from Siemens
Diagnostics Field Service. If necessary, Siemens Diagnostics Field Service can fax a copy to an FSE
onsite. Evaluate the instrument for the presence of all current upgrades and perform any necessary
revisions.
Can the problem be duplicated?: Does the same problem occur every time a certain situation is duplicated?
Does the problem occur randomly?
Follow typical troubleshooting techniques to narrow down the location of the problem: Refer to Section
9.2 for more information.
8.1.2 Power
Check that AC power is within range: 110-volt systems should have incoming AC voltage of 100 120
volts. 220-volt systems should have incoming VAC voltage of 210 230 volts. Check the intended wall
socket using a line voltage tester (available through Technical Support).
Check voltages at the Syringe Interface PCB: Check DC voltages at the main power connector (A15-J1) on
the Hamilton/Euro Interface PCB. Pins can be accessed through the side of the connector with small pin
type probes. A good chassis ground can be found at the grounding strap connected to the front latch of the
instruments lid. Ensure voltages are in tolerance as indicated below:
5 VDC: (4.9 to 5.2) -5 VDC: (-5.2 to 4.8)
12 VDC: (11.8 to 12.2) -12 VDC: (-12.2 to 11.8)
24 VDC: (23.8 to 24.2)
Note: The 24 volt level can be adjusted using the potentiometer wired to the supply. This voltage reading
should never be less than 24 volts.
Check if voltages are in range at the PMT Power Supply: Check voltages for the PMT power supply by
locating the quick-connect connection for the cable of the PMT, and tap into these points to read the +/-
5V supplies dedicated to the PMT. Ensure voltages are in tolerance as shown below.
5 VDC: (4.8 to 5.2) -5 VDC: (-5.2 to 4.8)
Note: Refer to Section 9: Troubleshooting for more troubleshooting information, including a list of error
messages.
Diagnostic programs are used to test particular components of the IMMULITE 1000 System. Click on the
desired diagnostic of the menu, to see a description of the program in the lower text box.
To load and run the diagnostic programs, follow the steps below:
Ensure IMMULITE 1000 is in Idle. From the Home Screen, select DIAGNOSTICS. The Common
Diagnostics screen appears:
Select a desired diagnostic program with the trackball and a brief description of the diagnostics
function can be seen in the Program Description box.
To run any diagnostic, select the desired program, and click the Run button.
Note: The System Configurations menu is mostly used by Field Repair Engineers for individual
instrument position adjustments. The motor configuration table (TABLE) used for all motor movement
step assignments is also accessed and adjusted through this menu.
A desired diagnostic program is selected with the trackball, and a brief description of the diagnostics
function can be seen in the Program Description box.
To run any diagnostic, select the desired program, and click the Run button.
Note: Some routines stop automatically at the end of the program cycle, while others repeat until the
ALARM MUTE and GO buttons are pressed in sequence described above.
If the prior program is not cleared before loading another, the following message appears in a red box
on the PC monitor:
300770 Computer cannot communicate with the Instrument
Programs can be downloaded when a blank display panel or the following message appears:
DPC IMMULITE IDLE
Select RUN IMMULITE to begin.
Notes:
Error checking routines are not activated during the use of most main diagnostics except in specially
needed cases.
Error messages are posted during CARPOS, MCARTST, CARSHAKE, TEMPS, PRIME, and
PMTCOUNT main diagnostics, and when using certain system Configurations adjustment and logic test
programs.
BCODESPD
This program is used to observe and/or adjust the motor No Sample/Test Unit Barcode
speed of the sample/assay barcode reader. Load at least four Reader
test units onto the load platform. Run the program and
observe the motor speed. It should be between 750 to 790
ms/rev. Press the pause button to advance to the next test
unit. Each position should not be more than 40 ms/rev of
each other.
BCODETST
Verifies proper functionality of the sample/assay barcode No Sample/Test Unit Barcode
reader. Place as many test units as desired onto the load Reader
platform. Each test unit is read 25 times, then the next test
unit is indexed in front of the barcode reader. Verify proper
barcodes choosing the VIEW button.
BKGNDCNT
This program provides the current background counts for the No PMT
instruments PMT. It will alternate moving the attenuator
between the attenuated and unattenuated position. While at
each position, a set of 12 one-second background counts is
taken. The GetdataT button provides the actual raw
background counts.
BPRIME
Used to isolate priming of "B" small syringe for diagnostic No Syringe B Pump Assembly
evaluation. Also, useful when performing a small syringe tip
replacement.
CARBACK
Useful for removing sample cups that have been mistakenly No Incubator Carousel
indexed around the incubator carousel. This program moves
the carousel backward in single index steps by pressing the
syringe prime button momentarily, or indexes continuously by
holding the prime button.
CARCHECK
This program is used to judge the position of the Main No Incubator Carousel
Incubator Carousel in relation to the tube Shuttle. It indexes
the carousel ring gear one position, moves the shuttle into
the space between the carousel baffles, pauses for one
second, moves the shuttle back to home and repeats.
CARPOS
This test checks the indexing of the main incubator carousel. No Incubator Carousel
It indexes the carousel ring gear one position every second.
Also activates system error checking to aid diagnosis.
DARKCNT
This program provides the current dark counts for the No PMT
instrument. It moves the attenuator to the home position, then
takes a set of twelve one second dark counts with the PMT
every fifteen seconds. The GetdataT button provides the
actual raw dark counts.
DECON
This program is used to introduce decontamination material No Fluidics (Decontamination)
into the fluidics of the Instrument. Refer to Maintenance
Procedure listed in Section 7.4.2 for further details. This
procedure draws the decontamination fluid into both syringes,
one at a time and then into the water pump five times. This
procedure should be run for at least 15 minutes, followed by
the DECON1 procedure.
DECON1
This program must be used after the DECON procedure. The Yes Fluidics (Decontamination)
purpose of this program is to flush out the decontamination
fluid and replace it with fresh fluids. Refer to Maintenance
Procedure listed in Section 7.4.2 for further details. This test
will automatically stop after approximately twenty minutes.
The test will cycle through filling each syringe twice then the
water pump ten times.
DISSUB
This program tests substrate pump dispensing volume. Place Yes Substrate Pump
five test units on the load platform as per display. If the wrong
number of vessels is used, the program will halt with Test
Failed. The program indexes each test unit around the
Incubation Carousel then shuttles them to the substrate fill
position. A shot of substrate is added, and the test units
moved to the sample collection tray for evaluation.
DISWAT
This program tests water pump dispensing volume. Place five Yes Water Pump
test units on the load platform. If the wrong number of
vessels is used, the program will halt with Test Failed. The
program indexes each test unit around the Incubation
Carousel then shuttles them to the water fill position. A shot
of water is added and the test units moved to the sample
collection tray for evaluation.
LEVCRASH
This program tests for the proper level sensing of the probe No Pipettor (Level Sensing)
in each of the five pipetting positions. Manually, pipette a
minimum of 100 uL of liquid into five sample cups. Place the
five sample cups into the five pipetting positions. During the
test, the Pipettor will level sense each sample cup starting in
position five and then progress cup by cup to position one.
The program will time-out after 20 rotational cycles (~ 100
potential signals) Any failures to level sense the liquid will trip
the tip jam sensor, which in turn will indicate test failed.
LEVFALSE
This program tests for the proper level sensing of the probe No Pipettor (Level Sensing)
in the reagent, five sampling, and assay dispense pipetting
positions. This test requires no sample cups to run. Remove
the Reagent Tray. The Pipettor will start at the home position
and then checks level sensing at the reagent position, each
of the five sample pipetting positions, and then return to the
home position to start again. The program will time-out
automatically after 20 full cycles to home (~ 120 potential
signals). Any false level sense triggers will trip the tip jam
sensor, which in turn will indicate test failed. During the
sequence the technician can also jiggle the wires or cables to
attempt to purposely initiate intermittent problems.
LEVSENSR
This program is used in conjunction with LEVSENSS to No Pipettor
check the auto calibration of the DEADVADJ (automatic), or (Level Sense Reagent Dead
LEVSNADJ (manual set-up) dead volume Configurations Volume Check)
programs. This program should display an SL (steps left)
number between 8 and 12. If the program does not confirm
good SL numbers, repeat the Configurations DEADVADJ
Auto-calibration. Or if the LEVSNADJ test is used, re-check
for proper manual pipetting of the minimum dead volume in
the test reagent vial.
LUMCHECK
This is one of the most useful tools for close inspection of the No Luminometer Chain
crucial crossover alignment of the Shuttle and Luminometer (at the crucial cross-over point
Chain. The program repeatedly inserts the shuttle partially with the shuttle)
into the indexing lum.chain baffles. *It is recommended to
remove the entire Spinner platform assembly to more
properly judge this crucial alignment, sighted from straight up
above the area. Alignment should be inspected so the shuttle
is centered in each baffle of the indexing lum.chain.
LUMCHN
This program checks the movement of the Luminometer No Luminometer Chain
chain by indexing it one position every second.
LUMTEST 4, LUMTEST 5,
LUMTEST 6, LUMTEST 9
Used in manufacturing only. N/A N/A
PIPXPOS
This program allows for visual inspection of the Pipettor No Pipettor
positioning. It moves the X-Pipettor to each of 10 rotational
positions, starting with home. At each position, it lowers the
Z-Pipettor momentarily. This test is used in conjunction with
the Pip-X Positioning Gauges. Refer to Section 7.4.5 for
more information.
PMTCOUNT
Activates measurement readings by the PMT, but does not No PMT
move the attenuator disk. Use the VIEW button to see the
current readings, or the GetdataT button for historical files.
PRIME
This program primes the two syringes and the associated No Syringe Pump Assembly and
tubing. Priming removes excessive air bubbles and allows for Fluidics
accurate dispensing of water and probe wash. It exercises
syringes A and B and solenoid valves 1,2 & 4. The systems
error checking is active, to aid in diagnosis. Also refer to
Section 7.4.4 for more fluidics information.
SHUTLOAD
This program allows for the evaluation of the shuttle as it No Shuttle
loads test units into the spin position. Manually load a test
unit onto the carousel one position before entering the
shuttle. The test unit will be loaded into the spin position.
Inspect for proper actuation of the springs of the shuttle, so
that the test unit fits snugly pushed into the spinner wheels.
The test unit should sit on a straight angle (without tilting).
Press GO to have the program eject the test unit for removal.
SHUTSPIN
This program is used to evaluate the high-speed spinner Yes Shuttle, High Speed Spinner
cycle. Manually load a test unit with 200 uL of liquid pipetted Motor
into it, onto the carousel one position before entering the
shuttle. The test unit will be loaded into the spin position and
the high-speed spin motor will activate for 30 seconds. The
test unit should slightly lift up and the liquid should quickly
clear from the bottom of the tube. Watch that the bead moves
off the bottom of the tube and floats in air, while the tube is
spinning. To check the actual spinner speed, use the
WASHSPD diagnostic.
SHUTTLE
This program is used to check the positioning of the shuttle. No Shuttle
First, the shuttle is moved from the home position to a neutral
Luminometer position. Next the shuttle is moved to the
spinner wheels (which also is the water dispense position), a
brief Sh Push adjustment is applied, then the shuttle is
moved to the substrate dispense position. In some
instruments, if necessary, a slight movement to the
Sh at Lum position is made to align the shuttle with the
Luminometer entrance. The cycle is then repeated.
SOLE2S
This program is used to check the dispensing volume and/or No Substrate Pump
drawback adjustment of the substrate pump. Disconnect the
tubing on top of the substrate heater and hold a sample cup
under the substrate tubing. Run the SOLE2S diagnostic to
dispense two shots. Use the ALARM MUTE button to pause
and control the shots into the cup. The approximate level of
substrate should be at the fourth line of the sample cup (2
shots at 200 ul = 400 ul). Confirm that the drawback is within
specification at , +/- . Also refer to the end of Section
7.4.4 for further adjustment details.
SPINCHK
This program allows for a quick visual check that the high- No High Speed Spinner Motor
speed spinner motor is working. Manually load as many test
units as desired onto the carousel and manually rotate to one
position before entering the shuttle. The test unit will be
loaded into the spin position and the spinner motor will turn
on for about 3 seconds.
SPINON
This program turns the high-speed spinner motor on or off, No High Speed Spinner Motor
when pressing the GO button.
SUBSHEAT
This program is used to monitor the substrate heater. No Substrate Heater
Remove the cover on the substrate heater. Using a DVM set
to DC volts, place the leads on the test points of the Heater
PCB. Run the program and monitor the starting (ambient)
temperature and final (heated) temperature. The DVM
reading is in direct correlation to the temperature. Multiply the
voltage measured by 100 and this will be the temperature in
Centigrade.
The stabilized temperature should be 36.6 .6oC. After
measuring the temperature, be sure to exit the program so
that the substrate is not heated for a long period of time.
UPDOWNH
This program is very useful for checking the centering and No Probe
rough depth of the probe in the home well and the functions
of the Z-Pipettor home sensor and mid sensor. The program
will move the X-Pipettor to the home well position, then move
the Z-Pipettor to the home sensor, down to the mid sensor,
then down the amount of error steps for parameter 32 (Z
Home) of the configuration table.
UPDOWNW
This program is very useful for checking the centering and No Probe
rough depth of the probe in the wash well and the functions
of the Z-Pipettor home sensor and mid sensor. The program
will move the X-Pipettor to the wash well position, then move
the Z-Pipettor to the home sensor, down to the mid sensor,
then down the amount of error steps for parameter 33 (Z
Wash) of the configuration table.
V2TEST
This program tests the functionality of Valve 2 (V2) of the No Syringe Valve 2
syringe assembly. This program alternately dispenses water
through the probe and the detergent container tubing. The
change in dispense point is altered by the activation or
deactivation of V2. If the valve is not working at all the water
will always dispense from the probe. If the valve is stuck in
the activated position, the water will always come from the
detergent container tubing. If one of the two diaphragms in
the valve is partially clogged, one port will slowly drip while
the other is dispensing. Another way of testing V2 is to
suspend the detergent container above the instrument and
check to see if a drop begins to form on the end of the probe.
If a drop has not formed within 10 minutes, the valve is
sealing properly.
V4TEST
This program tests the functionality of Valve 4 (V4) of the No Syringe Valve 4
syringe assembly. This program works exactly like V1TEST.
This program should alternately dispense water through the
probe and the wash well. The change in dispense point is
altered by the activation or deactivation of V4. If the valve is
not working at all the water will always dispense from the
probe. If the valve is stuck in the activated position, the water
will always come from the wash well. If one of the two
diaphragms in the valve is partially clogged, one port will
slowly drip while the other is dispensing. This program uses
V1 to direct the flow in a direction that can be differentiated
from the other.
WASHSPD
This program tests and/or allows adjustment of the speed of Yes High-Speed Spinner Speed
the high-speed spinner motor. Follow the prompts given on
the IMMULITE 1000 display screen. Adjust potentiometer
R107 on the DC Motor/Heater PCB to achieve 5500 RPM.
Press GO to stop the program.
WATERTST
This program is used whenever water contamination is Yes Fluidics
suspected from the source water supply. The program tests a (suspected water contamination
water source independent of the systems fluidics. Refer to from the water supply)
Section 7.4.1: Miscellaneous Procedures for further
information.
PIPXRADJ
This program performs an isolated adjustment of rotation Yes Probe
alignment of probe over reagent wedge. After pre-setting (X-rotational position at Reagent)
an open reagent wedge at the probe position with
REAGENTP, and with the reagent chamber lid open, use
the small opening of the reagent vial as sight judgement
for a centering target of the probes X-Position over this
hole.
PIPXSADJ
This program ensures the probe is centered in sample Yes Probe
cups in the five pipetting positions. To prevent damage to (X-rotational position
the customer probe, install a dummy probe. The smaller at all 5 Sample Cups)
notched Pip-X positioning gauge (from Kit p/n 420078)
is used to precisely target and/or adjust each of the five
sample pipetting positions. The program actions will work
clockwise around from position S5 to S1. Start the target
gauge at S5, and set the probe one at a time, using the
Go button for raising the test probe and shift the target
gauge to next S-position. Ensure the probe comes down
within one probe width of the center target of the gauge.
Refer to Section 7.4.5 for more information.
PIPXTADJ
This program allows the adjustment of the rotational Yes Probe
alignment of the probe at the dispensing position into the (X-rotational position at Assay Test
test unit. Place the larger (without a notch) Pip-X Unit Dispense Position)
positioning gauge (from Kit p/n 420078) in the test unit
dispensing position using CARCHECK. Run the
diagnostic and check for proper positioning either
centered or within one probe width to the rear (counter-
clockwise). Refer to Section 7.4.5 for more information.
This is the preferred diagnostic to test the tip jam circuit. Yes Probe
The program bottoms out the probe five times to trip the (Tip Jam Sensor Function)
tip jam sensor. A Pass/Fail condition will be posted on
the display panel. Can be repeated several times to
ensure proper operation.
9.1 Description
This section is designed to aid in the quick, accurate diagnosis of IMMULITE 1000 problems. While
many people can make IMMULITE 1000 repairs, accurate troubleshooting is a rare skill for the
experienced as well as the inexperienced.
In its simplest state, troubleshooting is an exercise in logic. It is essential to realize that the IMMULITE
1000 is really composed of a series of modules. Some of these modules are interrelated, others are not.
The instrument operates within a framework of logical rules and physical laws, and the key to
troubleshooting is a good understanding of all the IMMULITE 1000 modules.
LCHAIN Intermittent Load Chain errors. The chain has become stiff from
spilled liquids.
EMPTYTUB The position sensor disk has dust
interference.
Luminometer Luminometer Chain is not The "LumcPos" parameters in the
moving. config table are incorrect.
The motor and/or motor drive
circuitry has failed.
The shuttle is mispositioned and
interfering with the chain.
A lumchain baffle has become
dislodged and is interfering with the
LUMCHN chain.
An assay tube or partially dislodged
LUMCHECK Luminometer Chain baffle has
become jammed inside the
MCARTST luminometer, thus jamming the
chain.
SHUTTLE The chain tension has increased
surpassing the torque capabilities of
the motor.
Luminometer Chain is not The "LumcPos" parameters in the
moving to the correct position. config table are incorrect.
The position sensor has become
dislodged.
The slotted sensor disk has become
loose.
The chain tension has increased
causing the motor to miss steps and
"time out" during a move.
Drive Motor coupling inside cover
may be shifting loose.
EMPTYTUB No motors are moving nor is The fuse for the heater/motor power
the main cooling fan moving. supply (F3) has blown.
The heater/motor power supply has
become defective.
DARKCNT The CPS of the test units is The fuse for the PMT power supply
zero. (F1) has blown.
BKGNDCNT The PMT power supply has become
defective.
Linear Actuator Substrate / Air bubbles have appeared in An up-line fitting has loosened.
Water Injector Pumps the tubing. The water or substrate supply is low
enough to allow air to start entering
the system.
DISSUB (substrate) No substrate being dispensed. The substrate nozzle is clogged.
The substrate spike is clogged,
DISWAT (water) causing a vacuum.
The Substrate bottle is empty.
The Substrate actuator pump or
driver circuitry has failed.
Substrate volume dispensed is The substrate nozzle is partially
incorrect. clogged causing the substrate to
SOLE2S (substrate) dispense at an angle partially
missing the assay tube.
SOLE1W (water) Clogged air filter in the substrate
spike.
Reservoir was never primed.
Defective substrate actuator pump.
No water being dispensed. Water container is empty.
A line fitting up-line has become
very loose.
Water actuator pump or driver
circuitry has failed.
Water volume being dispensed The water supply bottle has become
is incorrect. empty.
Defective water actuator pump.
Air leak in the input line to the
pump.
Clogged in-line water filter.
30009 Shuttle The shuttle has not found Look for the cause of
Position Error the sensor at the the jam (Test Units, a
Luminometer position. bent shuttle flag or a
baffle that popped off
Note: This message will
the Luminometer
also appear on the
Chain).
Instrument display panel
when the error is detected. Troubleshoot with all
A notice will appear on related movement and
the display panel if the alignment tests.
IMMULITE 1000 is able
Notes:
to correct itself.
If the problem cannot be
corrected, the instrument
automatically shuts down
and Error Number 30042
will occur.
30011 Main Carousel The Main Carousel has Run related diagnostics,
Error not found a position including the test of the
sensor. sensor holesizes
MCARTST.
Note: This message will
also appear on the Note: If the problem
Instrument display panel. cannot be corrected,
A notice will appear on Error Numbers 30040
the display panel if the and 30041 will occur.
IMMULITE 1000 is able
to correct itself.
30140 Dark Count High The dark count measured Troubleshoot using
xxx exceeds the maximum related PMT tests to
acceptable limit (300 evaluate signals.
CPS).
Note: The preceding
sample may have had very
high CPS.
The following pages contain the Fluidics, Chassis Wiring Diagrams and Isometric Drawings. For
board drawings refer to the Schematic and Assembly Drawings section.
450029
MAIN CAROUSEL
POSITION PCB
450020
MAIN CAROUSEL
INTERFACE PCB
(A14)
500387-11
MAIN CAROUSEL FOAM KIT
400120
MAIN CAROUSEL HEATER PAD
(LOCATED UNDER MAIN CAROUSEL
FOAM KIT)
400083-01 400064-04
CABLE ASSY,MAIN MAIN CAROUSEL
NOTE:BEARING ASSEMBLY AT CAROUSEL THERMISTOR MOTOR ASSEMBLY
9:00 POSITION IS ADJUSTABLE BY (LOCATED UNDER
USING TWO 1/8" HEX SCREWS CASTING)
LOCATED BELOW.
400142-01
GROUND STRAP
ISOMETRIC DRAWING,
IN C U B A T O R C A R O U S E L
ASSEMBLY P/N 400003
SHEET 1 of 1
500339
LUMINOMETER COVER
400064-06
LUMINOMETER CHAIN
MOTOR ASSEMBLY
450011
LUMINOMETER/WASH
INTERFACE PCB
(A10)
400095
HIGH SPEED SPINNER ASSEMBLY
400084-01
LUMINOMETER
THERMISTOR ASSEMBLY
420040
HIGH SPEED SPINNER IDLER KIT
400064-08
LUMINOMETER
ATTENUATOR MOTOR
ASSEMBLY
420038
HIGH SPEED SPINNER MOTOR
ASSEMBLY
(901132
WITH COUPLING)
450006
SHUTTLE AT HOME
420001 GENERAL PURPOSE
OPTOSENSOR
SHUTTLE IDLER
500110
REPLACEMENT KIT ATTENUATOR MOTOR
SHAFT PULLEY
900262
RETAINING CLIP
500183
SHUTTLE IDLER WHEEL
COVER PLATE 400126
NON-INTERGRATED
900261
900026 PHOTOMULTIPLIER
WAVE WASHER TUBE ASSEMBLY
LUMINOMETER
500577 ATTENUATOR BELT
UPPER SHUTTLE SHIM OR
400164
(LOCATED UNDER TWO 450006
SHUTTLE IDLER WHEEL
REAR SCREWS) ATTENUATOR POSITIONING 400248
900731-16 GENERAL PURPOSE INTERGRATED
UPPER SHUTTLE SPRING OPTOSENSOR PHOTOMULTIPLIER
TUBE ASSEMBLY
450006
900731-32 ATTENUATOR HOME
LOWER SHUTTLE SPRING GENERAL PURPOSE
OPTOSENSOR
900159 450006
SHUTTLE UPPER SHUTTLE AT SPIN &
BEARING ASSEMBLY SUBSTRATE POSITIONS
400190-02 GENERAL PURPOSE
LUMINOMETER CHAIN ASSEMBLY 500048-02 OPTOSENSOR
BAFFLE,MOLDED
900160 LUMINOMETER CHAIN 400064-07
LOWER SLIDE BEARINGS SHUTTLE MOTOR
500506 500507
SIDE PLATE,CHAIN ROLLER,CHAIN
900160
LOWER SLIDE BEARINGS
ISOMETRIC DRAWING,
COMPLETE LUMINOMETER MODULE
SHUTTLE ASSEMBLY P/N 400020-02
SHEET 1 of 1
400185
REAGENT COOLING
CHAMBER ASSY
450014
REAGENT CAROUSEL
HOME & POSITION PCB
450007
REAGENT CAROUSEL
BAR CODE READER PCB
(A49)
400145
REAGENT TRAY
450051
LEVEL SENSING &
PROBE JAM
SENSOR PCB
(REFER TO KIT 420072
INSTRUCTIONS FOR
INSTALLATION)
420072
COMPLETE EURO
LEVEL SENSING KIT
(INCLUDES 450050 PCB,
450051 PCB AND MOUNTING
HARDWARE)
450050
PIPETTE Z AND
LEVEL SENSE
INTERFACE PCB
LED
450017
PIPETTOR X
HOME / POSITION PCB
450010
REAGENT CAROUSEL
INTERFACE PCB
(A12)
400064-01
REAGENT
CAROUSEL 400184
MOTOR SPIRAL CABLE ASSEMBLY
900422
TE D
THERMAL SHEET
400064-03
PIPETTOR "Z" MOTOR
400178
T E D ASSEMBLY 900114
PIPETTOR "X" MOTION DRIVE BELT
(FROM PIP-X MOTOR AROUND
Z-MOTION LEAD SCREW - NOT SHOWN)
ISOMETRIC DRAWING,
420096
T E D KIT P IP E T T O R / R E A G E N T
(INCLUDES THERMAL SHEET
AND FAN COWLING)
CAROUSEL
400064-02 MODULE P/N 400023
PIPETTOR "X" MOTOR
SHEET 1 of 1
CABLE ASSEMBLY TO LOAD CHAIN
POSITION GENERAL PURPOSE
COVER,BARCODE READER PCB OPTO SENSOR
500479 450006
CHAIN BAFFLE,
1.235 INCH
500048-01
}
500536
GROUND STRAP C O M P L E TE B A R C O D E
400142-03 O-RING IDLER ASSEMBLY
900771 400008
ROLLER
500059
(A TTA C H ED W ITH SC R EW
TO IN STR U M EN T LID LA TC H )
O-RING
900771
(BARCODE READER
MOTOR ASSEMBLY
REPLACEMENT KIT
WITH ACCESSORIES
420056)
L O W E R S E C T IO N O F C H A IN
POSITION SENSOR ASSEMBLY
(REFERENCE)
ISOMETRIC DRAWING,
PRINTED CIRCUIT BOARD LOAD PLATFORM
GENERAL PURPOSE
OPTO SENSOR
MODULE P/N 400199
450006
SHEET 1 of 1
IMMULITE 1000 Index
L S
Large Syringe Tip Sample Collection, 3-7, 3-8, 3-17, 3-28, 6-6, 7-
Changing, 9-23, 9-39 15, 7-42, 7-43, 8-7, 8-9, 8-18, 8-19, 9-24
Linear Actuator Pump Assembly, 3-20, 3-30, 4- Slope, 9-33
4, 6-7, 7-16, 7-29, 7-30, 8-6, 8-8, 8-13, 8-22 Software Overview, 9-27, 9-29, 9-35
Load Platform Assembly, 7-13 Spare Parts, 3-24, 10-1, 10-5
Luminogenic reaction, 3-4 Inventory, 10-3
Specifications
M Computer, 5-1
Computer, 5-1
Maintenance, 9-16, 9-26 Electrical, 5-2
Daily, 9-16, 9-26 Electrical, 5-2
Error Log, 6-2, 7-50, 7-51, 7-53, 8-1 Electrical, 5-2
Power Log, 2-1, 2-3, 3-17, 3-18, 3-41, 6-2, 7- Environmental, 5-2
20, 9-8, 9-9 Environmental, 5-2
Preventive, 6-1 Fluidics, 5-2
Menu Bar Fluidics, 5-2
Options, 9-28, 9-35, 9-46 General, 5-1
Miscellaneous Procedures, 2-4, 2-5, 2-6, 2-7, 2- Luminometer, 5-1
9, 3-14, 3-23, 6-1, 6-6, 7-26, 7-27, 7-28, 7-29, Luminometer, 5-1
7-30, 7-31, 7-32, 7-34, 7-35, 7-36, 7-39, 7-40, Noise, 5-2
7-43, 8-10, 8-16, 8-19 Physical, 5-2
Physical, 5-2
O Start Up, 3-7, 3-42, 8-5, 9-27, 9-35
Substrate
Operating Modes, 9-35 Reservoir Module, 3-23, 7-16, 7-17, 7-23, 8-
22, 9-10, 10-3
P Temperature, 3-21, 3-22, 6-8, 7-16, 7-17, 7-
23, 8-13, 8-22, 8-24, 9-10, 9-16, 9-26, 10-
Photomultiplier Tube, 1-5, 2-6, 3-2, 3-3, 3-4, 3- 2, 10-3, 10-4
10, 3-11, 3-12, 3-17, 3-18, 3-19, 3-35, 3-44, Syringe Pumps, 3-24, 3-29
4-1, 4-2, 4-4, 6-2, 6-5, 6-8, 7-9, 7-16, 7-19, 7- System Components, 1-3, 2-5, 2-8, 3-1, 3-2, 3-4,
28, 8-1, 8-6, 8-7, 8-8, 8-10, 8-14, 9-6, 9-8, 9- 3-6, 3-8, 3-9, 3-10, 3-11, 3-12, 3-13, 3-14, 3-
9, 9-12, 9-30, 9-33, 9-34 15, 3-19, 3-20, 3-21, 3-30, 3-33, 3-34, 3-36,
Pipetting Sequence, 3-31, 3-32 3-37, 3-39, 3-42, 4-2, 4-4, 4-5, 5-1, 6-1, 6-3,
Power Supply, 3-10, 3-17 6-4, 6-5, 6-6, 6-7, 7-4, 7-6, 7-7, 7-10, 7-11, 7-
17, 7-19, 7-20, 7-21, 7-22, 7-23, 7-24, 7-27,
7-29, 7-31, 7-38, 7-40, 7-41, 7-42, 7-43, 7-44, Error Messages, 9-14
8-6, 8-8, 8-9, 8-10, 8-11, 8-12, 8-14, 8-17, 8- Event Viewer Screen, 7-52
18, 8-20, 8-21, 8-23, 9-2, 9-4, 9-5, 9-6, 9-7, 9- Techniques, 1-2, 9-1
8, 9-12, 9-17, 9-18, 9-19, 9-20, 9-22, 9-23, 9- Typical Problems, 9-4
24, 9-26, 9-27, 9-28, 9-29, 9-30, 9-35, 9-36,
9-38, 9-39, 9-41, 9-42, 9-43, 10-1, 10-2, 10-3, W
10-4
Warranty Statements, 5-4
T Water Acceptance Criteria, 2-8, 7-31
Water Supply
Temperatures Testing, 2-4, 2-5, 2-6, 2-7, 2-8, 2-9, 7-26, 7-
Troubleshooting, 3-13, 3-33, 3-37, 3-38, 3-39, 27, 7-28, 7-29, 7-30, 7-31, 7-32, 7-33, 7-
3-40, 7-10, 7-11, 9-8, 9-29, 9-41, 10-2 34, 7-39, 8-8, 8-16
Testing the Water Supply Water tests
Acceptance criteria, 2-8, 7-31 Acceptance criteria, 2-8, 7-31
From any source, 2-4, 7-26 From any source, 2-4, 7-26
Probe and wash station, 7-29 Probe and wash station, 7-29
Theory of Operation, 3-1, 3-2, 3-4 WATERTPM, 7-29
Troubleshooting, 9-14, 9-43, 9-46 WATERTST, 2-4, 7-26
Error Log, 6-2, 7-50, 7-51, 7-53, 8-1
IMMULITE & IMMULITE 1000
Assembly & Schematic Drawings
Schematic and Assembly drawings are collated in numerical order of Drawing Number.
Note: The revisions listed are the drawing revisions. They do not necessarily reflect the actual PCB
revision. The drawing revisions shown below were the latest available when this manual was released.
Consult Field Services for up to date information on any possible new PCB revisions.
ASSEMBLY DRAWINGS
SCHEMATIC DRAWINGS