TROUBLESHOOTING GUIDE
Abbott Diagnostics
A Division of Abbott Laboratories
Abbott Park, IL 60064
CELL-DYN 3000/3500/3700/3200 Troubleshooting Guide
ii 67483-101
REVISION STATUS
CONTENTS
RBC Diluent Syringe Overpressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
HGB Lyse Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
WBC Lyse Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Lyse Empty (CELL-DYN 3500/3700) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Diluent/Sheath Empty (CELL-DYN 3200) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Reagent Empty - Diluent, Sheath, HGB Lyse (CELL-DYN 3000) . . . . . . . . . . . . . . . . . . . . . . . . 9
Reagent Empty - Diluent, Sheath, Detergent (CELL-DYN 3500/3700) . . . . . . . . . . . . . . . . . . . 11
Shear Valve Position Fault . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Vacuum Accumulator Wet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Clog Error (RBC or WIC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Flow Error (RBC or WIC) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
WOC / WBC Flow Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Elevated PLT Backgrounds / LRI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Controls Out of Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Controls Out of Range - WBC / WOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Controls Out of Range - NOC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Controls Out of Range - WIC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Controls Out of Range - RBC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Controls Out of Range - HGB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Controls Out of Range - MCV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Check center
section of shear
valve
Yes
Check RBC
dilution pathway Run Does error
No Problem resolved
for pinched/ backgrounds recur?
crimped tubing
Yes
Pinched/ Refer to HS or
crimped tubing dispatch service
seen?
Yes
Remove crimps/
massage tubing
Go to Diagnostics,
More, Initialization
Run backgrounds
No
Does error
No Problem resolved
recur?
NOTE:
Remove RBC diluent syringe from
bracket (do not disconnect tubing)
Yes
y Pull down slightly on plunger
y Reinitialize analyzer by going
to: Remove, clean, and
- Main reinstall shear valve &
- Diagnostics release pressure from Does error
No Problem resolved
- More RBC diluent syringe recur?
- Initialization (See Note) Run
y If cannot initialize, cycle power backgrounds
and reinitialize
Yes
Clean/replace
Refer to HS or Does error
Yes RBC Diluent
dispatch service recur?
Syringe
No
Problem resolved
Press CLEAR
FAULT and
observe HGB Lyse
Syringe
Is syringe
Clean or replace Does error
moving No No Problem resolved
syringe recur?
properly?
Yes
Yes
Check reagent cube and inlet lines:
- cube empty?
- reagent correct?
- inlet line reaches to bottom? Dispatch Service
- vent hole in cap is open?
- foam seal removed from under cap?
- lines not crimped?
- ports not plugged?
- check fluid level in syringe
Disable analyzer,
Is syringe filled exercise valve 24
properly? Are using valve puller Does error
No
reagent lines not recur?
crimped? Massage and reseat
tubing in valve 28
Yes
Yes
Does error
Dispatch Service Yes No Problem resolved
recur?
Press CLEAR
FAULT and
observe WBC
Lyse Syringe
Is syringe
Clean or replace Does error
moving No No Problem resolved
syringe recur?
properly?
Yes Yes
Yes
Yes
Refer to HS or
dispatch service
Cycle power
Does error
No Problem resolved
recur?
Yes
Dispatch Service
C h e c k r e a g e n t c u b e a n d in le t lin e s :
- c u b e e m p ty ?
- r e a g e n t c o r r e c t?
- in le t lin e r e a c h e s to b o tto m ?
- v e n t h o le in c a p is o p e n ?
- fo a m s e a l r e m o v e d fr o m u n d e r c a p ?
- lin e s n o t c r im p e d ?
- p o r ts n o t p lu g g e d ?
- c h e c k flu id le v e ls in s y r in g e
P r im e ly s e t o w a s t e c h a m b e r
2 b y r e m o v in g t u b in g f r o m
P re s s C L E A R F A U L T V 3 6 (N C V )
F lo s s tu b in g in to N C V 3 6
E x e r c is e s o le n o id s D o e s e rro r
Yes
42, 47 re c u r?
No
D o e s e rro r
No P r o b le m r e s o lv e d
re c u r?
Yes
C h e c k W IC /H G B L y s e S y r in g e :
- s e a te d a n d m o v in g c o r r e c t ly ?
- tu b in g c r im p e d ?
C le a n /r e s e a t s y r in g e a s n e e d e d
a n d r e m o v e c r im p s
D o e s e rro r
No P r o b le m r e s o lv e d
re c u r?
Yes
C h e c k ly s e in le t lin e fo r c r im p s No
o r p lu g s
R e m o v e ly s e in le t tu b in g & f lu s h
lin e w ith w a rm D I w a te r
D o e s e rro r D o e s e rro r
P r o b le m r e s o lv e d No Yes C y c le P o w e r
re c u r? re c u r?
Yes
R e fe r to H S o r
d is p a tc h s e r v ic e
Run background
Does error
Problem resolved No
recur?
Yes
Yes
Check sensor
contacts at top of
Refer to HS or
reservoirs are
dispatch service Cracks / leaks
No seated properly.
seen?
Reseat if
indicated. Run
background count.
Yes
Dispatch service
Does error
No Problem resolved
recur?
Yes
Page 1 of 2
Go to
A
Does error
Problem resolved No
recur?
Yes
Run background
Does error
Problem resolved No
recur?
Yes
Cycle power
Does error
Problem resolved No
recur?
Yes
Refer to HS or
dispatch service
Page 2 of 2
Clear fault
Clean sensor
Run background
contacts
Ensure sensor is
seated correctly
Does error
No Problem resolved
recur?
Does error
Problem resolved No
recur?
Yes
No
Yes
Check reservoir for
Does error cracks
Cycle power
recur?
Yes
Floss NCV tubing
Refer to HS or (V6-1 (Diluent), V6-2 (Sheath),
dispatch service V6-3 (HGB Lyse))
No Cracks seen?
for affected reservoir
or
replace check valve
Yes
Refer to HS or
Dispatch Service
Does error
Problem resolved No
recur?
Yes
Page 1 of 2 Go to
A
Does error
Problem resolved No
recur?
Yes
Cycle power
Does error
Problem resolved No
recur?
Yes
Page 2 of 2 Refer to HS or
Dispatch Service
Clear fault
Clean sensor
contacts
Ensure sensor is
seated correctly
Does error
No Problem resolved
recur?
Does error
Problem resolved No
recur?
Yes
No
Yes
Check reservoir for
Does error
Cycle power cracks
recur?
Yes
Yes
Refer to HS or
Does error Dispatch Service
Problem resolved No
recur?
Yes
Go to
Page 1 of 2 A
Does error
Problem resolved No
recur?
Yes
Cycle power
Does error
Problem resolved No
recur?
Yes
Refer to HS or
Page 2 of 2 Dispatch Service
S h e a r V a lv e
P o sSitio
h ena rF V
aua lt
lv e
P o s itio n F a u lt
D id e rro r o c c u r Is s h e a r v a lv e
C le a n a n d in s p e c t R e fe r to H S o r
a fte r c le a n in g s h e a r No d a m a g e d /p in s Yes
s h e a r v a lv e D is p a tc h s e rv ic e
v a lv e ? b ro k e n ?
Yes No
R e m o v e a n d re in s ta ll
R e in s ta ll s h e a r
s h e a r v a lv e c o rre c tly
v a lv e c o rre c tly
(C E L L -D Y N 3 0 0 0 :
u s e lu b ric a n t)
R e in itia liz e s y s te m
D o e s e rro r
No P ro b le m re s o lv e d
re c u r?
Yes
C y c le p o w e r
D o e s e rro r
No P ro b le m re s o lv e d
re c u r?
Yes
R e fe r to H S o r
d is p a tc h s e rv ic e
Vacuum
Vacuum
Accumulator W et
Accumulator W et
Observe waste
line for
obstructions or
correct draining
Yes
Drain accumulators
Reinitialize system
Does error
No Problem resolved
recur?
Yes
Cycle power
Does error
No Problem resolved
recur?
Yes
Refer to HS or
dispatch service
C log E rror (R B C or W IC )
C log E rror (R B C or W IC )
R un background
D oes error
P roblem resolved No
recur?
Y es P roblem resolved
R B C C log:
- clean/replace 10 m L syringe
- clean air filter
D oes error - run autoclean
P roblem resolved No
recur?
V erify aperture plate(s) are correctly
installed and check for cracks
R eseat and clean per O P S m anual
Y es
NOTE:
S olenoids D oes error
C E LL-D Y N 3000: No P roblem resolved
recur?
3-1, 3-2, 3-3, 3-6, 3-7, 3-8
C E LL-D Y N 3500/3700:
R B C : 3-1, 3-2, 3-3, 3-4, 3-5, 3-7, 3-8 Y es
W IC : 8-1, 8-2, 8-3, 8-4, 8-5, 8-7, 8-8
G o to
A
P age 1 of 2
Check
transducer(s) and
metering tubes for
cracks or leaks
Check quality of
Cracks/leaks
meniscus in No
seen?
metering tube
Yes
Check fitting between top
of metering tube and V 3-
7 (RBC) or V 8-7 (W IC) Good
No Dispatch Service
meniscus?
Clean or flush metering
tube if needed
Yes
Does error
Problem resolved No Yes Cycle power
recur?
Run background
Does error
Problem resolved No
recur?
Yes
Does error
Problem resolved No
recur?
Yes
Does error
Problem resolved No
recur?
Yes
Check
transducer(s) and
metering tubes for
cracks or leaks
NOTE:
Solenoids
CELL-DYN 3000:
3-1, 3-2, 3-3, 3-6, 3-7, 3-8
Cracks/leaks Go to
CELL-DYN 3500/3700: No
seen? A
RBC: 3-1, 3-2, 3-3, 3-4, 3-5, 3-7, 3-8
WIC: 8-1, 8-2, 8-3, 8-4, 8-5, 8-7, 8-8
Yes
Dispatch Service
Page 1 of 2
Run background
Does error
Problem resolved No
recur?
Yes Note:
To clean RBC or WIC metering tube:
Check quality of CELL-DYN 3000: V38 (RBC)
meniscus in CELL-DYN 3500/3200: V38 (RBC), V88 (WIC)
metering tube -- it
should look like an 1. Disconnect tubing above V38 or V88.
upside down U. 2. Connect 20 mL syringe filled with 20% bleach
to tubing.
3. Open V38 or V88 until syringe is half empty.
Check fitting between top of 4. Release solenoid and wait 5 minutes.
metering tube and V 3-7 5. Open V38 or V88 until syringe is empty and
(RBC) or V 8-7 (WIC)and Good release solenoid.
No 6. Fill syringe with HGB Lyse reagent and repeat
exercise valves if indicated meniscus?
Clean metering tubes step 5.
(See Note) 7. Run background.
Yes
Does error
Problem resolved No Yes Run autoclean
recur?
Does error
No Problem resolved
recur?
Yes
Cycle power
Page 2 of 2
WOC/WBC Flow
WOC/WBC
Errors Flow
Errors
Rerun sample
Does error
Yes recur? No
Does error
Problem resolved No Drain/Fill reagents Replace reagents
recur?
Yes
Does error
No Problem resolved
Verify Results by an recur?
alternate method
Yes
Verify syringes:
- seated correctly
- moving correctly
- bubble free
See NOTE
Clean or replace
Syringe OK? No
as needed
Yes
Does error
NOTE: No Problem resolved
recur?
Check the following syringes:
CELL-DYN 3000: WBC metering syringe
CELL-DYN 3200: Sample injection syringe
CELL-DYN 3500/3700: WOC metering sheath syringes Go to
Yes
A
Page 1 of 2
Check transfer
peristaltic pump tubing
for wear or holes
Replace as needed
SEE NOTE
Does error
No Problem resolved
recur?
Yes
Yes
Check channel
offsets
Run background
(<2.0, N/A on
CELL-DYN 3200)
No
Yes Yes
Yes
Does error
Cycle power
Clean Shear Valve recur?
Yes
No
NOTE:
Problem resolved CELL-DYN 3000: WBC transfer peripump tubing
CELL-DYN 3200: Sample transfer peripump tubing
Page 2 of 2 CELL-DYN 3500/3700: WOC transfer peripump tubing
Does problem
Problem resolved No
recur?
Yes
Is electrical
Go to Does problem
No background = No Problem resolved
C recur?
0?
Yes Yes
Does problem
Problem resolved No
recur?
Yes
Run autoclean
Does problem Go to
Problem resolved No Yes
recur? A
Page 1 of 4
R u n b a c k g ro u n d
D o e s p ro b le m
P ro b le m re s o lv e d No
re c u r?
Yes
C le a n R B C /P L T A p e rtu re P la te
and
(C E L L -D Y N 3 0 0 0 /3 5 0 0 /3 7 0 0 o n ly )
s h e a r v a lv e a s p re c a u tio n
R u n b a c k g ro u n d
D o e s p ro b le m
P ro b le m re s o lv e d No
re c u r?
Yes
C E L L -D Y N 3 0 0 0 /3 2 0 0 : C h e c k /
re p la c e (if in d ic a te d ) d ilu e n t/
s h e a th filte r
R u n b a c k g ro u n d
D o e s p ro b le m
P ro b le m re s o lv e d No
re c u r?
Yes
D ra in /fill b a th s a n d re s e rv o irs
(flo w c e ll o n C E L L -D Y N 3 2 0 0 )
D o e s p ro b le m G o to
P ro b le m re s o lv e d No Yes
re c u r? B
P age 2 of 4
Perform prepare
for shipping
procedure
Run 3-5
backgrounds
Does problem
Problem resolved No
recur?
Yes
Clean or drain
vacuum
accumulators
Run background
Does problem
Problem resolved No
recur?
Yes
Yes
No
Page 3 of 4
Run background
Does problem
Problem resolved No
recur?
Yes
Does problem
Problem resolved No
recur?
Yes
Run background
Does problem
Problem resolved No
recur?
Yes
Check environment:
- printer/other devices on or near CELL-DYN? (e.g.: centrifuge, air conditioning)
- flickering/fluorescent lights?
- humidity?
Problem resolved
- CELL-DYN connected to UPS?
Yes
Refer to HS or
Dispatch Service
Page 4 of 4 Yes
- Verify QC means/limits
- Check storage/handling
- Condition of QC when
received
- Rerun controls
Does problem
No Problem resolved
recur?
Check reagent cube and inlet lines:
- cube em pty?
Yes - reagent correct?
- inlet line reaches to bottom ?
- vent hole in cap is open?
- foam seal removed from under cap?
Are backgrounds Troubleshoot
No - lines not crimped?
within specs? backgrounds No - ports not plugged?
- check fluid level in syringe
Yes
Does problem
occur in one mode
Run Fresh
only?
Controls
Does problem
No Problem resolved
Yes Yes recur?
Troubleshoot aspiration
Does problem system of affected system Yes
recur?
Yes
Are sam ples Go to Incomplete
No
aspirated correctly? Aspiration Tree
Run precision
Yes
Yes
Yes
Does problem
Problem resolved No
recur? Go to OPEN
/ CLOSED
m ode
Page 1 of 10
Issue resolved
OPEN /
CLOSED
No
No Yes
Yes
Issue resolved
Were controls mixed by
hand, warmed to room
Check affected temperature for 15
Yes
parameters minutes, and stored
No inside refrigerator shelf?
No
Are multiple
parameters out of No Redo controls
range? mixed by hand
WOC/
NOC WIC RBC HGB MCV PLT
WBC
Yes
Go to Go to Go to Go to Go to Go to Go to
A B C D E F G
Does problem
Problem resolved No
recur?
Yes
Issue resolved
Yes
Check syringes:
- seated/moving
properly?
- filling?
Clean/replace as
needed
Does problem
Problem resolved No
recur?
Page 2 of 10
Run precision
Run normal
No
patient sample
Is placement of
Replace WOC
neutrophil population
No peripump tubing
correct on 0/10
and rerun controls
scattergram?
Yes
Are channel
Dispatch service No
offsets < 2.00?
Yes
NOTE:
Check the following syringes:
CELL-DYN 3000: WBC metering syringe & WOC sheath syringe
CELL-DYN 3200: Sample injection syringe
CELL-DYN 3500/3700: WOC metering syringe & WOC sheath Does problem
No Problem resolved
syringe recur?
Yes
Page 3 of 10
Go to
AA
AA
Is mixing chamber
Dispatch service No draining/filling
correctly?
Yes
Run controls
Is the WOC
Problem resolved No
within ranger?
Yes
Dispatch service
Page 4 of 10
Yes Yes
Yes
Does problem
Problem resolved No Check sample
recur?
injection syringe
Clean/replace as
Yes needed
Check HGB Lyse Syringe:
- seated/moving properly?
- filling?
Clean or replace as Does problem
needed No Problem resolved
recur?
Yes
Does problem
Problem resolved No
recur?
Verify NOC calibration by
performing a calibration
verification and
Yes calibrate if indicated
Rerun controls
Check staging
pathway from
shear valve to
HGB flow cell for
obstructions Does problem
No Problem resolved
recur?
Yes
Does problem
Problem resolved No
recur?
Refer to HS or
dispatch service
Yes
Refer to HS or
Page 5 of 10 dispatch service
Yes
Rerun
Yes
controls
Is WIC within
Problem resolved Yes
range?
No
Is transducer
Refer to HS or
No operating
dispatch service
correctly?
Yes
Does problem
Problem resolved No
recur?
Yes
Page 6 of 10
Run precision
Yes
Does problem
Clean RBC aperture plate No Problem resolved
recur?
(except CELL-DYN 3200)
Yes
Yes
Is transducer
Refer to HS or
No operating
dispatch service CELL-DYN 3200:
correctly?
check vacuum #2
(3.5" open,
5.0" closed)
Yes
CELL-DYN 3200
Verify RBC mixing chamber
for proper fill/drain; swirl mix
Vac #2 within Refer to HS or
No
specs? dispatch service
Page 7 of 10
Yes
HGB reference
within specs?
No
HGB Reference
Clean HGB flow >2200 for CELL-DYN 3000/ Refer to HS or
No Yes
cell 3500/3700 dispatch service
>2050 for CELL-DYN 3200
Yes
Page 8 of 10
Yes
Does problem
Problem resolved No
recur?
Yes
Yes
Page 9 of 10
Yes
Yes Yes
Yes
Yes Is transducer
Refer to HS or Does problem
No operating No Problem resolved
dispatch service recur?
correctly?
Yes Yes
Issue resolved
Does problem
Problem resolved No
recur?
Page 10 of 10
Run controls in
open m ode
Are W IC and
No Is W IC correct? No Troubleshoot W IC
W OC correct?
Yes Yes
Go to
D
Troubleshoot Troubleshoot
closed m ode W OC
Go to
A
Run precision
Yes
Check closed
m ode aspiration
Yes
Does problem
Problem resolved No
recur?
Yes
Verify m ode to
m ode calibration
Calibrate if
indicated
Does problem
Problem resolved No
recur?
Page 1 of 6
Run precision
Does problem
Problem resolved No
recur?
Yes
Clean/replace if needed
Does problem
Problem resolved No
recur?
Yes
Check sample
aspirations/shear
valve
Is sample
Go to Incomplete
No aspirating
Aspiration tree
correctly?
Yes
Go to
Page 2 of 6 B
Check WOC
mixing chamber
Check solenoids
Does chamber
1-7, 2-3, 2-4
No drain/fill and bubble
Massage and
mix properly?
reseat tubing
Yes
Replace WOC
Does problem
Problem resolved No Yes peristaltic pump
recur?
tubing
Does problem
Problem resolved No
recur?
Yes
Does problem
Problem resolved No
recur?
Yes
Does problem Go to
Problem resolved No
recur? C
Page 3 of 6
Check channel
offsets
Are channel
Dispatch service
offsets less No
or refer to HS
than 2.00?
Yes
Verify WOC
calibration
Recalibrate if
indicated
Does problem
Problem resolved No
recur?
Yes
Dispatch Service
Page 4 of 6
Run Precision
Does problem
Problem resolved No
recur?
Yes
Clean/replace if needed
Does problem
Problem resolved No
recur?
Yes
Clean WIC
Aperture Plate
Does problem
Problem resolved No
recur?
Yes
Go to
E
Page 5 of 6
Check W IC
transducer
Does problem
Problem resolved No
recur?
Yes
Is sam ple
Go to Incom plete
No aspirating
Aspiration tree
correctly?
Yes
Verify W IC
calibration
Recalibrate if
indicated
Does problem
Problem resolved No
recur?
Yes
Dispatch Service
Page 6 of 6
H & H MISMATCH
H & H Mismatch
H & H Mismatch
Run controls to
determine affected
parameters
Run precision
Is precision Troubleshoot
No
within specs? imprecision
Yes
Affected
parameter
Go to Go to Go to
A B C
Page 1 of 5
H & H MISMATCH
C heck reagent cube and inlet lines:
- cube em pty?
- reagent correct?
- inlet line reaches to bottom ?
A - vent hole in cap is open?
- foam seal rem oved from under cap?
- lines not crim ped?
- ports not plugged?
- check fluid level in syringe
D oes problem
P roblem resolved No
recur?
Y es
R un autoclean
D oes problem
P roblem resolved No
recur?
Y es
D oes problem
P roblem resolved No
recur?
Y es
D oes problem
P roblem resolved No
recur?
Y es
D oes problem
P roblem resolved No Y es D ispatch S ervice
recur?
P age 2 of 5
H & H MISMATCH
B
Does problem
Problem resolved No
recur?
Yes
Run autoclean
Does problem
Problem resolved No
recur?
Yes
Clean/replace syringe
Perform
Calibration
Does problem
Problem resolved No verification
recur?
Calibrate if
indicated
Yes
Yes
Does problem
Problem resolved No
recur?
Dispatch Service
Page 3 of 5
H & H MISMATCH
C
Does problem
Problem resolved No
recur?
Yes
Run autoclean
Does problem
Problem resolved No
recur?
Yes
NOTE:
HGB References:
CELL-DYN 3000 1850-2250 Check HGB
CELL-DYN 3500 1850-2250 Reference
CELL-DYN 3700 1800-2300 See Note
CELL-DYN 3200 1800-2300
Yes
Does problem
Problem resolved No
recur?
Does problem
No Problem resolved
recur?
Yes
Go to
Page 4 of 5 D
H & H MISMATCH
Exercise solenoids:
CELL-DYN 3000/3500/3700: V 1-3
CELL-DYN 3200: V2-4
Massage/reseat tubing
Does problem
Problem resolved No
recur?
Yes
CELL-DYN 3500/3700:
Exercise WIC transducer drain
solenoids V8-2 and V8-3
Massage and reseat tubing
Does problem
Problem resolved No
recur?
Yes
Perform calibration
verification
Calibrate if indicated
Does problem
Problem resolved No
recur?
Yes
Dispatch Service
Page 5 of 5
INCOMPLETE ASPIRATION
Incomplete
Incomplete
Aspiration
Aspiration
(multiple samples)
(multiple samples)
Does error
Go to
occur in both No
A
modes?
Yes
Inspect/change
sample peristaltic Check vacuum #2
pump tubing
Yes Yes
Check sample
Flush Y-valve
sensor
Yes
Inspect/clean
shear valve and
Yes
flush Y-fitting with
10% bleach
Page 1 of 2
INCOMPLETE ASPIRATION
A
Perform Sam ple path cleaning and include: Perform Sam ple path cleaning and include:
CELL-DYN 3000/3500/3700 : clean Y-fitting CELL-DYN 3000/3500/3700 : clean Y-fitting
CELL-DYN 3200: clean Y-valve CELL-DYN 3200: clean Y-valve
Yes
Yes
Check and exercise solenoids:
CELL-DYN 3000/3500/3700: 9-5 (open
Check/clean m ode)
sam ple/vent CELL-DYN 3200: Exercise V1-7 and verify
needles no restrictions between Y-valve and V1-7
Massage/reseat tubing
Does error
No Problem resolved
Does error recur?
Problem resolved No
recur?
Yes
Yes
Replace aspiration
peripum p tubing
Check SL LED (for CELL-DYN
3000/3500/3700
only)
No
No
Refer to HS or
Issue resolved
dispatch service
Page 2 of 2
Does problem
Problem resolved No
recur?
Yes
Does problem
Problem resolved No
recur?
Yes
Replace W OC
peristaltic pump
tubing
Does problem
Problem resolved No
recur?
Yes
Clean/replace if needed
Does problem
Problem resolved No
recur?
Yes
Go to
Page 1 of 3 A
Yes
Check staging pathway from
WOC mixing chamber to flow
Does problem cell:
Problem resolved No Yes
recur?
Massage and reseat tubing in
staging solenoids
Does problem
Problem resolved No
recur?
Yes
Check channel
offsets
(CELL-DYN 3000/
3500/3700 only)
Are channel
offsets less
than 2.00?
Yes
Check 0/10
scattergram
Are lymph/neut
populations
No Dispatch Service
- correctly positioned?
- well defined?
Yes
Go to
B
Page 2 of 3
Verify WOC
calibration
Recalibrate if
indicated
Does problem
Problem resolved No
recur?
Yes
Dispatch Service
Page 3 of 3
Repeat
Problem resolved Yes background
OK?
No
Yes
Yes
Replace milipore
Does problem filter
Clean Aperture
recur?
Plates
No
Does problem
No Problem resolved
recur/
Yes Does problem
Problem resolved No
recur?
Yes
Does problem Go to
Problem resolved No Yes
recur? A
Page 1 of 2
Repeat
Clean vacuum accumulator
background Yes Problem resolved
Clean fan filters
OK?
No
Perform prepare
for ship
Repeat
Problem resolved Yes background No Cycle power
OK?
Yes Repeat
background
OK?
No
Dispatch Service
Page 2 of 2
Start
First time to
Has it worked
No be hooked-up
before?
to an LIS?
Yes
Yes
Yes
Is set-up
No Correct set-up
correct?
Does problem
Yes No Problem resolved
recur?
Yes
NOTE:
Perform serial loop Go to
Refer to Operator's Manual for
back test A
computer setup
Page 1 of 2
Pass? Failed?
Yes Yes
Dispatch Service
Cycle power and
reseat LIS cable
Does problem
No Problem resolved
recur?
Yes
Refer to LIS
vendor
Page 2 of 2
Problems with
Go to
Both patients or QC out Patients Only
A
of range or both?
Is it sample Go to
Rerun QC QC Only No
specific? A
Yes
Still out of
Issue resolved No Rerun
range?
Yes
Is issue
No Issue resolved
Prepare a new dilution recurring?
of QC and rerun
Yes
Yes
Is issue
No Issue resolved
Try new vials of recurring?
QC, prepare a new
dilution and rerun
Yes
Yes
No
Is issue Go to
Issue resolved No Yes
recurring? A Issue resolved
Page 1 of 4
Is background on
the stain <100? No Rerun background
Yes
Check maintenance
- extended autoclean - weekly
Is background
- clean metering syringe Yes
<100?
- clean sheath syringe
- drain and fill flow cell
No
No
Yes Controls within
range? Yes
Controls within
No
range? Yes
Yes
Follow procedure below:
1. For WOC gains: go to Main menu, Diagnostics, F12,
F1, Set point entry (0, 10, 90), and print screen
2. For reticulocyte gains: go to Reticulocyte mode,
Diagnostics, F12, F1, WOC (0, 10, 90)
3. Compare WOC gains and reticulocyte gains (0, 10,
90)
Do WOC and
Rerun QC Yes reticulocyte gains
match?
No
Go to Go to
B C
Page 2 of 4
No
Patient
Issue resolved Yes Run 3 -- 5 normal patient
samples OK? Controls within
samples in regular patient No
range? Yes
mode
No
Issue resolved
No
Issue resolved
Go to
D
Page 3 of 4
No
Controls within
Issue resolved Yes Rerun QC
range?
No Dispatch Service
References:
1. CELL-DYN 3500 Operator's Manual, June 1999: Chapter 14, 14-1 to 14-8
2. CELL-DYN 3700 Operator's Manual, November 2000: Chapter 14, 14-1 to 14-87
3. CELL-DYN 3200 Operator's Manual, October 2001: Chapter 14
Page 4 of 4
Step Action
1 Fill syringe with 25% bleach solution.
2 From the Main Menu, press
SPECIAL PROTOCOLS
DISABLE/ENABLE ANALYZER
5 Perform 2 autocleans with water to allow bleach to enter and clean flow cell.
6 Run 3 background counts.
7 Verify background counts are within specifications.
Step Action
1 Fill syringe with 25% bleach solution.
2 From the Main Menu, press
SPECIAL PROTOCOLS
DISABLE/ENABLE ANALYZER
5 Perform 2 autocleans with water to allow bleach to enter and clean flow cell.
6 Run 3 background counts.
7 Verify background counts are within specifications.