Académique Documents
Professionnel Documents
Culture Documents
From the Department of Radiology, Howard University Hospital, 2041 Georgia Aye, NW, Washington, DC 20060.
From the AAPM/RSNA Physics Tutorial at the 1996 RSNA scientific assembly. Received August 21, 1997; revision re-
quested October 9 and received November 11; accepted November 13. Address reprint requests to the author.
RSNA, 1998
151
Incident beam, unattenuated
Ionization
Incident beam Attenuated beam
chamber
Figure 1. Diagrams
an unattenuated
show
x-ray beam
(top) and an x-ray beam pass-
Lj
ing through a foil (bottom)
into detectors. Foil
attenuation and the terms used to character- The thickness of a material can be expressed
ize it, (b) the factors that affect attenuation, in different units of measure, for example,
(c) exponential attenuation relationships, meters, kilograms per meter squared, and
(d) concepts involved in the attenuation of electrons pen meter squared.
monochromatic and polychromatic x-ray An attenuation coefficient is a measure of
beams, and (e) half-value layer (HVL) mea- the quantity of radiation attenuated by a
sunements and their significance. given thickness of an absorbing material. The
linear attenuation coefficient, symbolized by the
U DEFINITION OF ATTENUATION Greek letter .t, is the fractional change in x-
Attenuation is the reduction of the intensity of ray intensity per the thickness of the attenu-
an x-ray beam as it traverses matter. The re- ating material because of interactions in a
duction may be caused by absorption (in this given material:
process, energy is transferred from the pho-
tons to atoms of the target or irradiated ma- =N/NAx, (1)
terial) or by deflection of photons from the
beam (scatter). where N is the number of photons removed
In the example of a beam of x rays passing from the x-ray beam in thickness \x. In
through a foil and into an x-ray detector, Equation (1), for any given Ii, Ex must be
some of the photons will interact with the foil chosen so that the number of photons re-
and be absorbed completely from the beam moved from the beam is much smaller than
and some photons may be scattered (Fig 1). If the total number of photons. As the thickness
one measures the intensity of the beam (a) af- of the attenuating material increases, the
ten it has been attenuated by the foil and as it equation is no longer correct and the rela-
strikes the detector and then (b) without the tionship becomes nonlinear.
foil and as it strikes the detector, one obtains The linear attenuation coefficient is mea-
a quantitative measurement of the interaction sured in units of pen unit length, which is
of the x rays with the material contained in most commonly expressed in terms of centi-
the foil. meters or millimeters. Attenuation rate can
The intensity of an x-ray beam passing also be expressed in terms of the mass of the
through a layer of attenuating material de- material encountered by photons. The mass
pends on the thickness and type of material. attenuation coefficient is obtained by dividing
the linear attenuation coefficient by the den-
sity of the material through which the pho-
Table 2
Physical Properties of Selected Materials
tons pass and thus is represented by the sym- efficient is obtained by dividing the mass at-
bol pjp. Mass attenuation coefficient is the tenuation coefficient by the number of elec-
rate of photon interactions per unit area mass trons per gram.
and is independent of the physical state of
the material. The typical unit of the mass at- U FACTORS AFFECTING ATTENU-
tenuation coefficient is per gram per centime- ATION
ten squared (cm2/g), since the unit in which Several factors affect attenuation. Some are
thickness is measured is gram per centimeter related to the x-ray beam or radiation and
squared (the mass of a 1-cm2 area of mate- the others to properties of the matter through
na!). The coefficient is the inverse of the unit which the radiation is passing. The factors in-
in which thickness is measured. dude beam energy, the number of photons
Other attenuation coefficients are the elec- traversing the attenuating medium or ab-
tronic and atomic coefficients, in which the sorber, the density of the absorber, and the
thickness of the attenuating medium is ex- atomic number of the absorber. As noted, the
pressed as the number of electrons or atoms greater the thickness of the attenuating mate-
per unit area, respectively. The relationship rial, the greater is the attenuation. Similarly,
among the attenuation coefficients is shown as the atomic number or density of the mate-
in Table 1. nial increases, the attenuation produced by a
The atomic attenuation coefficient 1a is the given thickness increases. Thus, different ma-
fraction of an incident x-ray or gamma ray terials such as water, fat, bone, and air have
beam that is attenuated by a single atom (ie, different linear attenuation coefficients, as do
the probability that an absorber atom will in- the different physical states or densities of a
teract with one of the photons in the beam). material, such as water vapor, ice, and water
The atomic coefficient is obtained by divid- (Table 2; Figs 2, 3).
ing the mass attenuation coefficient by the
number of atoms per gram. The electronic co-
U,
ZA>ZB f E1>E2
a
C
a, 0.5 . 0.5 Energy E1
>
a, a,
0 0
0 5 10
0
Nal - . -.
Photoelectric
a) Water- - - -
8 10
C
0
Ca
C
a)
I Air
C,
0)
-
-
-
. -
-
.
Compton
Rayleigh
Pair
Ca
U)
E
U)
C)
Ca
E 0.1 .
Ca
0
0.01
0.01 0. 10
Energy (MeV) Energy (MeV)
4. 5.
Figures 4, 5. (4) Mass attenuation coefficients for selected materials as a function of photon energy.
Graph shows the variation of si/p for sodium iodide, lead, water, and air. (5) Mass attenuation coefficients
for photons in air. Graph displays the mass attenuation coefficient for air (with an effective atomic number
of about 7.6) for specific interactions with x rays and the total attenuation as a function of energy.
Photoelectric interactions are important ample, the attenuation curve for sodium io-
for a low-energy range (up to 50 keV) and dide will show an increase at 33 keV because
materials with large atomic numbers. Pair the K electron binding energy is 33 keV for
production interaction is important only for a iodine (Figs 4, 5). The attenuation curve for
very high energy range (5-100 MeV) and ma- lead will show an increase at 88 keV (Figs 4,
terials with large atomic numbers. Compton 5). The curves decrease more slowly in the
interaction is predominant in the intermedi- region in which the Compton effect is impor-
ate energy range (60 keV-2 MeV) for all ma- tant. Because the mass attenuation coeffi-
tenials, regardless of atomic number (1). The cients do not depend on density and the
relative probability of each type of interac- physical state of the absorber, numeric data
tion is proportional to the cross section for are often expressed in terms of these coeffi-
that process. Cross section is defined as the cients, rather than linear attenuation coeffi-
probability that a particular reaction will oc- cients.
cur. The total linear attenuation coefficient is The range of energies used in x-ray imag-
equal to the sum of the individual interac- ing is chosen to optimize the diagnostic x-ray
tions and their cross-sectional values: information and to minimize the radiation
absorbed by the patient. Both these factors
totaI = c + K, depend on the mass attenuation coefficients
of various materials and tissues.
where t = photoelectric, Compton and The importance of linear and mass attenu-
classical, and x pair production interac- ation coefficients can be demonstrated in sev-
tions. This equation with the appropriate eral clinical situations. Contrast agents that
subscripts applies to the mass, electronic, and contain iodine and barium are used because
atomic coefficients. of their large attenuation coefficients, which
In radiography performed with low ener- increase the visibility of anatomic structures
gies (<30 keV), photoelectric effect is most that contain the contrast agent. The increased
important in soft tissue and bone. As the x- attenuation is caused by the atomic number
ray energy is increased, Compton scattering and K absorption edge of the contrast agent
becomes the predominant interaction. If t is being greater than those of the surrounding
plotted versus photon energy for air, soft tis- tissue. In cases in which the penetration
sue, and lead, the curves fall rapidly with in- of x rays must be reduced, a shielding mate-
creases in energy because of the rapid de- rial with a large attenuation coefficient is
crease of the photoelectric effect. However, at
the K absorption edge, there will be an in-
crease in the attenuation coefficient. For ex-
_____
.
,- a,..
______
5
a,. .
p:-.
:5
____ a,.
,,
.; . -
Figure 6. Exponential attenuation .. ____
a, .
required. Shielding is achieved by using ma- number of photons in a beam weighted by its
tenials with a high atomic number, such as energy) is the quantity that is most often
lead. measured. Exponential reduction in the num-
ben of photons is demonstrated in Figure 6. If
U EXPONENTIAL ATTENUATION I/Jo 5 plotted as a function of x on linear
RELATIONSHIPS graph paper, an exponential curve will be ob-
Attenuation measurements of a monoener- tamed (Fig 7). The logarithm of the number
getic (monochromatic) beam of x or gamma of photons transmitted varies linearly with
rays depend on the number of photons inci- the thickness of the attenuating material;
dent on an absorber, the number of photons therefore, if the logarithm of I/I,, is plotted
transmitted through the absorber, and the against x, a straight line graph will result.
absorber thickness. The expression j.t = This plot is referred to as a semilogarithmic
Lx previously discussed must be trans- plot because one axis is logarithmic and the
formed into a more convenient form. If N other linear.
and Ex are very small, they are known as dif- Polychromatic beams contain a spectrum
ferentials and the differential equation is of photon energies. With an x-ray beam, the
solved by using calculus to give the follow- maximum photon energy is determined by
ing equations: the peak kilovoltage (kVp) used to generate
the beam. Because of the spectrum of photon
I = I et
0
(2) energies, the transmission of a polychromatic
and beam through an absorber does not strictly
N = N e, 0
(3) follow Equation (3). When a polychromatic
beam passes through an absorber, photons of
where I = beam intensity at an absorber low energy are attenuated more rapidly than
thickness of zero, x absorber thickness, I the higher energy photons; therefore, both
beam intensity transmitted through an ab- the number of transmitted photons and the
sonben of thickness of x, e base of the natu- quality of the beam change with increasing
nal logarithm system, .t attenuation coeffi- amounts of an absorber. A semilogarithmic
cient, N = number of transmitted photons, plot of the number of photons in a polychro-
and N0 = number of incident photons. matic beam as a function of the thickness
These equations may be used to calculate of the attenuating materials will not be a
attenuation by any thickness of material straight line but will be a curve (Fig 8). The
when the incident and transmitted photon in- initial slope of the curve is steep because the
tensity or photon number is measured. In di- low-energy photons are attenuated, but, as
agnostic radiology, photon intensity (ie, the the beam becomes more monochromatic, the
slope decreases. A comparison of the curves
for polychromatic and monochromatic radia-
tion is shown in Figure 9.
U, U,
C 800 C
a 2
0 0
800
100
E 400 E
#{149}1 U,
C C
a, a,
I- 200
Figure 7. Attenuation of
0 10
monochromatic radiation
0 4 8 12 16 20 0 4 8 12 16 20 plotted on a linear scale and
cm of water cm of water semilogarithmic scale.
C
0
U,
Ce
E
U,
C
a,
I-
C
a,
U
a,
0.
30 40 50 80 70 80 90 100 10
Energy(keV) 0 1 2 3 4 5
Attenuator
Figure 11. Diagrams illus-
Attenuator
trate the geometry for nar-
row-beam and broad-beam
Source I, Source
conditions. HVL measure-
----p Detector Detector ments should always be
made under narrow-beam
INScattered photons
geometry conditions to en-
arc sure that only primary (unat-
not detected
scattered into the tenuated) photons reach the
detector detector. (Redrawn from ref-
erence 3 and reprinted with
Narrow-Beam Geometry Broad-Beam Geometry permission.)
120
E 100 10
E
;8o
Ce
0 60
0.
C
a,
>.
a,
* 20
0 I
0 1 2 3 4 5
01234567
Absorber thickness (mm Al)
Filtration (mm Cu)
Figure 12. Results of a typical measurement series for HVL determination are shown
for a lower-energy beam (left) measured with aluminum and a higher-energy beam
(right) measured with copper. The graph on the right has several sequential HVLs mdi-
cated below the curve. For example, the first HVL is the thickness required to reduce the
original intensity of the beam from 68 R/min (1.75 x 102 C/kg/mm) to 34 R/min (8.77 x
10-s C/kg/mm), which graphically is determined as 0.35 mm copper. After the addition
of 1 mm copper, the beam is now reduced to 20 R/min (5.16 x 10 C/kg/mm). The HVL
of the beam including the 1 mm copper is the thickness required to reduce the beam to
10 R/min (2.58 x 10 C/kg/mm). The thickness is graphically determined as 1.3 mm
copper, indicating the greater penetrability of the beam with added filtration. Several
other HVLs indicated on the graph are determined in a similar fashion. (Right graph re-
drawn from reference 1 and reprinted with permission; left graph redrawn from refer-
ence 4 and reprinted with permission.)
75
E
I)
50
Broad a
11)
25 0 rge Filter
deor tector
near (B)field
0 10
0 1 2345 01234567
A complete attenuation curve is not essen- (ie, if the thickness of the absorber is 1 HVL),
tial for routine dosimetry; rather, thicknesses then:
of the attenuating material that reduce the
exposure rate to slightly more than haif and 1/10 - 0.5;
to slightly less than half are required. The therefore,
difference in apparent attenuation for broad 1/10 = 0.5 = e-
and narrow beams is seen in Figure 13. Un-
den broad-beam conditions, the beam will ap- If the natural logarithm (inverse function
pear to have greater penetrating power (ie, a of the exponential) is calculated for each side
greater HVL or half-value thickness) than if it of the equality,
were measured with narrow-beam geometry.
in [0.5] = in [e-]
U RELATIONSHIP OF HVL AND -0.693 = jiHVL
LINEAR ATTENUATION COEFFI-
CIENT HVL = 0.693/si (4)
For a monoenergetic beam of x-ray or gam-
ma ray photons, it was already determined l 0.693/I-IVL. (5)
in Equation (2) that I I0e. When x HVL =
a,
I = EEE =
Thus, if the first HVL reduces the beam to = = = = = = = = =
C 0.5 - - - - - - - - -
a,
half again to one-quarter. With a monoenen- U, 0.2
U,
getic beam, the first and second HVLs are a, 0.1
E 10 20 30 40 50 60
equal.
Energy (key)
With a polychromatic beam, photons of
Figure 15. Illustration shows how effective en-
low energy are attenuated more rapidly than
ergy can be determined with use of graphical in-
photons of higher energy. The second HVL
terpolation. HVL is measured in the same way as
(ie, the thickness required to reduce the pen- in Figure 14 (eg, in millimeters of aluminum), and
etration to one-quarter) is larger than the first the linear attenuation coefficient is calculated with
HVL. The ratio of the two HVLs first HVL/ - Equations (4) and (5). The correct energy is deter-
second HVL is called
- the homogeneity coef- mined from the graph at the intersection of the at-
ficient. It follows that the homogeneity coeffi- tenuation curve and the effective mass attenuation
coefficient value.
cient for a polychromatic beam is less than
one.
U,
3. HVL measurements and values are used to
C
a, indicate if these filtration criteria are met.
C
Figure 16 demonstrates the effect of added
a,
filtration on a polychromatic x-ray beam.
a,
U CONCLUSIONS
a, One of the technical principles on which radi-
a,
ography is based is the difference in attenua-
tion by different materials; thus, an under-
0 20 40 60 80 100 standing of attenuation probability, the units
Photon Energy (key) for describing it, and the factors affecting it is
Figure 16. Graph demonstrates the effect of essential. A practical way of expressing the
added filtration on the energy and intensity of a penetrating ability of x-ray beams from dif-
polychromatic x-ray beam. (Modified from refer- ferent x-ray tubes is by using the concept of
ence 2 and reprinted with permission.) HVL, which must be measured under nar-
row-beam geometry conditions.
This article meets the criteria for I .0 credit hour in category I of f/ic AMA Physician s Recognition Award.
To obtain credit, see the questionnaire on pp 145-150.