Vous êtes sur la page 1sur 4

I t is very important to biopharma-

ceutical manufacturing that processing


equipment be cleaned using validated
procedures to avoid cross
contamination from material remaining from
a previously manufactured product (13).
Potential residues remaining on the surfaces
determined trace levels of proteins,
nucleic acids, amino acids, sugars, and
detergents (9).
In this article we present a procedure for
the TOC analysis of swab samples for the
cleaning validation of fermentation
equipment used in the production of
of equipment following processing steps biopharmaceuticals. This procedure is based
(fermentation, for example) include, in in part on a method developed previously
addition to the product itself, a multitude of for the colorimetric determination of total
Mark A. Strege, Terry L. organic cellular components such as nucleic protein in swab samples (10).
Stinger, Brett T. Farrell, and acids, proteins, and carbohydrates.
Avinash L. Lagu Together with appropriate surface Experimental
sampling procedures, analytical techniques Materials. Potassium biphthalate and sodium
that can determine trace levels of hydroxide were purchased from Sigma (St.
Louis, MO) and Mallinckrodt (Paris, KY),
respectively. Texwipe Alpha Swab polyester

Total Organic Carbon Analysis of Swab swabs were acquired from Baxter Scientific
Products (McGaw Park, IL), and E. coli

Samples for the Cleaning Validation of cells were obtained from Eli Lilly and
Company (Indianapolis).

Bioprocess Fermentation Equipment It is important to note that the heads of


the polyester swabs used in this study were
thermally bonded to the handles, eliminating
the need for adhesives that can cause
Validated cleaning procedures are contaminants are required to support the contamination during extraction. The swabs
needed to ensure the absence of validation of equipment cleaning were also laundered by the manufacturer to
contaminants from bioprocessing procedures. Currently, the preferred method minimize inherent nonvolatile residues or
equipment, and these procedures of sampling is to directly swab the particulates that might otherwise have
must be supported by appropriate equipment surface (4). In this approach, the significantly decreased the sensitivity of the
swab head is wetted with an effective agent analysis (11).
analytical methodology. This
to facilitate transfer of contaminants from Equipment. The TOC analyzer used in this
article describes the development
the surface to the swab, a defined area of the study was a TOC-5000 (Shimadzu,
of a quantitative total organic
equipment surface is swabbed, and Columbia, MD) equipped with a 74-position
carbon (TOC) assay for residual contaminants acquired by the swab are
carbon-containing materials on extracted into solution before analysis.
stainless steel surfaces using Analytical methodology is then needed to
E. coli cells as a model substance. determine trace levels of contaminants in the
presence of both a solubilizing medium and
a swab.
Total organic carbon analysis (TOC) is a
method that has been used successfully for
monitoring water quality (57), including
the quality of water for injection in
pharmaceuticals (8). TOC analysis offers a
number of distinct practical advantages over
other commonly used residual testing
methods such as enzyme-linked
immunosorbent assays (ELISA) and protein
assays (Lowry, Bradford, and bicinchoninic
Corresponding author Mark A. Strege is an acid, for example) because of its high
associate senior analytical chemist, Terry L. sample throughput, lack of interfering Figure 1. Plot of TOC-measured carbon
Stinger is a technician, Brett T. Farrell is an substances, and inherent sensitivity. One concentration compared with prepared
associate senior analytical chemist, and Avinash L. study demonstrating the potential carbon concentration (in aqueous sucrose
Lagu is a senior research scientist at Lilly Research application of TOC for the analysis of rinse solutions) that demonstrates method
Laboratories, Lilly Corporate Center, Indianapolis, water samples for cleaning validation linearity from 0 to 30 g/mL carbon.
IN 46285, (317) 276-9116, fax (317) 276-5499.
autosampler. The TOC-5000 employs ANALYTICAL PERFORMANCE PARAMETERS
acidification of the sample with
hydrochloric acid followed by sparging with Accuracy. The accuracy of an analytical calculating the standard deviation of this
purified air to remove inorganic carbon from method is the closeness of test results response. The standard deviation
solution as carbon dioxide gas. The organic obtained by that method to the true value. multiplied by a factor of 10 provides an
compounds remaining in solution are then Accuracy may often be expressed as estimate of the limit of quantitation.
oxidized to carbon dioxide in a combustion percent recovery by the assay of known,
tube packed with a catalyst and incubated at added amounts of analyte. Accuracy is a Linearity. The linearity of an analytical
680 C. The carbon dioxide produced in this measure of the exactness of the analytical method is its ability to elicit test results
manner passes through a dehumidifier and method. that are directly, or by a well-defined
mathematical transformation, proportional
halogen scrubber before it reaches a cell
Limit of detection. The limit of detection to the concentration of analyte in samples
where it is quantified by a nondispersive is the lowest concentration of analyte in a within a given range. Linearity is usually
infrared detector tuned to the highly specific sample that can be detected, but not expressed in terms of the variance around
carbonyl absorption frequency in the necessarily quantitated, under the stated the slope of the regression line calculated
1,700-cm1 range. The signal response was experimental conditions. Investigators according to an established mathematical
stored in a microprocessor and compared may measure the magnitude of analytical relationship from test results obtained by
with a calibration curve generated by background response by analyzing a the analysis of samples with varying
analysis of solutions of 0, 26.6, and 53.2 number of blank samples and calculating concentrations of analyte.
ppm potassium biphthalate corresponding to the standard deviation of this response.
levels of 0, 12.5, and 25 ppm organic The standard deviation multiplied by a Precision. The precision of an analytical
carbon, respectively. Between three and five factor, usually 3, provides an estimate of method is the degree of agreement
the limit of detection. among individual test results when the
replicate measurements were obtained for
procedure is applied repeatedly to
each sample so that in all cases, coefficients Limit of quantitation. Limit of multiple samplings of a homogenous
of variation of less than 2% were obtained quantitation is a parameter of quantitative sample. The precision of an analytical
for the series of replicate injections. assays for low levels of compounds. It is method is usually expressed as the
Methods. Potassium biphthalate standards the lowest concentration of analyte in a standard deviation or relative standard
were made up and diluted in purified water sample that can be determined with deviation (coefficient of variation).
in glass volumetric flasks that had been acceptable precision and accuracy under
previously rinsed with 50% nitric acid and the stated experimental conditions.
copious amounts of purified water to remove Investigators may measure the magnitude Source: United States Pharmacopeia XXII,
any trace of organic contaminants. Purified of analytical background response by National Formulary XVII, pp. 17101712
water was obtained through the use of a analyzing a number of blank samples and (1990).
Milli-Q water system (Millipore, Bedford,
MA). All standards, blanks, spikes, and
controls were prepared at least in triplicate. the following protocol. The fabric head of a with the exception that clean swab heads not
Stainless steel pans were coated with swab was wetted in 1 N sodium hydroxide, exposed to the pan surface were used. For
E. coli cells in the following manner. Wet and excess liquid was removed by gently the determination of a limit of detection and
cell paste (1.43 mL), obtained after pressing the swab head against the walls of a limit of quantitation, a set of 10 swab
centrifugation of fermentation broth in a the reagent reservoir. 5 cm 5 cm sections blanks were prepared and analyzed.
TJ-6 centrifuge (Beckman Instruments, of the pan were swabbed in their entirety, Linearity was evaluated by analyzing four
Fullerton, CA) for 20 minutes at 1,000 g, after which the swab stem was cut levels of sucrose standards in purified water
was suspended to volume with purified approximately 1 cm above the swab head, corresponding to concentrations of 0, 10, 20,
water in a 100 mL volumetric flask to make dropping each swab head into a glass 5 mL and 25 g/mL carbon.
a 70-diluted cell suspension. A 3.825 mL TOC autosampler vial (Shimadzu) also Assay precision was determined by
aliquot of well-mixed 70-diluted cell containing a small stir bar. Before analysis, preparing a set of 25 swabs by spiking
suspension was then diluted to 4.0 L final 1.0 mL of 1 N sodium hydroxide, 3.0 mL of 100 L of 233 diluted cell suspension
volume with purified water, making a Milli-Q water, and a small stir bar were onto the swab heads and then analyzing five
73,200-diluted cell suspension. A 1.0 L added to tubes containing swab samples. spiked swabs per day for five days. To
volume of well-mixed 73,200-diluted cell The samples were then stirred on a stir plate validate recovery, two sets of three levels of
suspension was transferred to a at room temperature for one hour, after diluted cell paste (100 L volumes of 70-,
25.2 50.6-cm stainless steel pan and was which time 1.0 mL of 2 N hydrochloric acid 140-, and 280-diluted cell paste were
then lyophilized to dryness, depositing was added before sparging and injection by aliquoted four times each to sample tubes)
E. coli cells corresponding to approximately the instrument. were prepared, one set in the presence of
267 g of wet cell mass per 25 cm2 area. Swab blanks were prepared in a manner swab heads (spikes) and one set in the
Ten 25 cm2 sections of the stainless steel similar to that described above for the absence of swab heads (controls).
pan surface were sampled with swabs using preparation of stainless steel pan samples,
Table 1. Precision of the analysis of five Table 3. Stainless steel pan swab sample swab samples were then calculated.
replicates of swabs spiked with diluted recovery determined by the analysis of As displayed in Table 2, recovery
E. coli cell paste analyzed on each of E. coli cell paste deposited on a stainless ranged from 59.4% to 110.9% for
five days. steel pan. The pan swab signal was the set of 10 replicates,with a mean
compared with that generated by an recovery of 81.9%. A lower TOC
aqueous suspension of the cell paste and recovery (67%) was achieved by other
Precision is expressed as percent recovery.
Day No. (% RSD) researchers who used a swab wetted
with dilute phosphoric acid to sample
1 6.87
TOC Recovered %
a surface upon which whole cells had
2 2.05
Replicate (g/swab) Recovery been applied (12). A similar degree of
3 2.74
1 48.1 74.8 high variability of recovery was also
4 7.18
2 60.0 93.3 encountered in a previous study (10),
5 5.83 in which it was believed to reflect
3 57.4 89.3
Pooled five-day precision = 7.15% 4 55.0 85.5 limitations associated with precise
RSD. 5 59.1 91.9 sampling as well as the limitations in
6 46.1 71.7 the establishment of a consistent layer
7 47.9 74.5 of cells on the surface during drying.
Results and Discussion 8 44.1 68.6
We determined the TOC signals generated 9 38.2 59.4 General Applications
by the sucrose standards prepared in the 10 71.3 110.9 The method described in this article
absence of swab heads to increase linearly was developed as a quantitative
with carbon concentration, with a simple assay for trace amounts of carbon-
curve fit correlation coefficient of 0.999 containing components that may be
(Figure 1). Because the preparation of swab pooled five-day data set. present on the stainless steelsurfaces
samples required extraction and dilution to a The results of the evaluation of accuracy
5.0 mL volume before analysis, the linearity
determined for the concentrations of 025
g carbon/mL from the sucrose solutions
using diluted cell paste are presented in
Table 2. At the three levels of dilution (70,
140, and 280 in purified water), the
TOC is less labor
corresponded to levels of 0125 g/swab. In TOC signals from E. coli cell paste were
intensive and offers
comparison with a standard curve, the 10 determined to range from approximately 20 greater general
swab blanks generated signals equivalent to to 100 g/swab, corresponding to
a mean background of 8.0 g/swab. All 74.9103% recovery in the presence of swab application than other
sample determinations discussed in the heads.
remainder of this section were corrected for Table 3 shows the recoveries from each
assays that have
the swab blank background. A limit of of the 10 swab samples taken from the been used to support
detection and limit of quantitation were surface of the stainless steel pan, which had
established at 9.2 g/swab and 12.1 been coated with a layer of E. coli cells. As cleaning validation.
g/swab, respectively, for this analysis. described above, we calculated that
A determination of analysis precision approximately 267 g of wet cell mass was of bioprocess equipment. E. coli
indicated relative standard deviations (RSD) deposited per 25 cm2 area on the pan cells, which are present at levels
to range between 2.05% and 7.18% within a surface. A direct TOC determination of the of approximately 10131014 cells/L,
set of five replicate analyses of swabs spiked 73,200 diluted cell suspension revealed together with a multitude of soluble
with diluted cell paste, as assayed daily for that 267 g of wet cells corresponded to carbon-containing components during
five days (Table 1). Day-to-day precision 64.3 g of organic carbon, against which a typical fermentation, were chosen
was determined to equal 7.15% RSD for the number the percent recoveries of the assayed to represent a most challenging
case model for this study because
of their insoluble particulate nature.
Table 2. Method accuracy determined by the analysis of diluted E. coli cell paste both in In comparison with the fermentor cell
the presence (spikes) and absence (controls) of swab heads expressed as percent concentration mentioned above, the
recovery of spike signal compared with control signal. TOC method accuracy was validated
by the recovery of levels of carbon
Dilution of Mean Control Mean Spike Mean % Recovery corresponding to approximately
Cell Paste Signal (g) Signal (g) (5 Replicates) 109 cells/25 cm2 surface area. It is
70 97.8 100.6 102.9
important to note that the method will
determine the presence of cellular
140 50.4 44.7 88.7
material in swab samples irrelevant
280 25.9 19.4 74.9
of the condition of the cells (intact validation, with a specific focus on Organic Carbon and Total Organic Halide as
Broad-Spectrum Parameters for Drinking
compared with lysed, for example). recovery, in a manner similar to that Water Control, Proc. AWWA Water Qual.
TOC is less labor intensive and offers described in this article (13). Technol. Conf. 13 (1986), 339.
greater general application than other References (8) R. Matsuda et al., Total Organic Carbon as an
assays that have been used to support (1) The Pink Sheet, F-D-C Reports, 55 (22), T&G- Index for Specification of Water for Injection,
13 (1993). J. Assoc. Off. Anal. Chem. 70 (1987), 681686.
cleaning validation. For example, TOC (9) R. Baffi et al., A Total Organic Carbon
(2) H.L. Avallone, Drug Substance Manufacture
can be used to determine the presence and Control, Pharm. Eng. 9 (2), 3740, 57 Analysis Method for Validating Cleaning
of compounds such as amino acids, (1989). Between Products in Biopharmaceutical
nucleic acids, and sugars that are (3) S.W. Harder, The Validation of Cleaning Manufacturing, J. Parent. Sci. Tech. 45
Procedures, Pharm. Tech. 8 (5), 2934 (1991), 1319.
incapable of generating a response in (10) M.A. Strege et al., Total Protein Analysis of
(1984).
assay methods such as a total protein (4) FDA Mid-Atlantic Region Inspection Guide: Swab Samples for the Cleaning Validation of
determination. At Eli Lilly and Co., Cleaning Validation (Food and Drug Bioprocess Equipment, BioPharm 7 (9),
TOC has been routinely used for the Administration, Washington, DC, 1993), 3. 4042 (1994).
(5) Y. Egozy and J. P. Denoncourt, Trace Level (11) CleanTips Polyester Swabs product
analyses of swabs taken from process description, Texwipe Inc., 1995.
Analysis of High-Purity Water, II. Total
equipment exposed to a variety of Organic Carbon, Ultrapure Water 3 (6) (12) S. Lombado et al., Development of Surface
water-soluble drug substances (such as (1986). Swabbing Procedures for a Cleaning
(6) A. Henriksen, D.F. Brakke, and S.A. Norton, Validation Program in a Biopharmaceutical
antibiotics), excipients (such as lactose), Manufacturing Facility, Biotechnol. Bioeng.
and formulated cleaning agents (such Total Organic Carbon Concentrations in
Acidic Lakes in Southern Norway, Environ. 48 (1995), 513519.
as surfactants), and the methods used Sci. Technol. 22 (1988), 11031105. (13) B.T. Farrell, unpublished results, Eli Lilly and
for these assays have undergone general (7) C.D. Reach Jr. and J.T. OConnor, Total Co. (1995). BP

Reprinted from BIOPHARM INTERNATIONAL, April 1996 Printed in U.S.A.

Vous aimerez peut-être aussi