Académique Documents
Professionnel Documents
Culture Documents
The analytical instruments that are in use in our power plant can be broadly
classified as stack monitoring instruments, gas analysers and steam and
water analysers . However, a few more portable instruments are used in
chemical laboratory.
Oxygen Analyser
Where:
P2 is the partial pressure of the oxygen in the measured gas on
one side of the cell
P1 is the partial pressure of the oxygen in the reference air on
the other side
T is the absolute temperature
C is the cell constant
K is the arithmatic constant
When the cell is at the operating temperature and there are unequal
concentrations across the cell, oxygen ions will travel from high partial
pressure oxygen side to the low partial pressure side of the cell. The result is
a logarithmic related output voltage. Because the magnitude of the output is
proportional to the logarithm of the inverse of the sample of the oxygen
partial pressure, the output signal increases as the oxygen concentration of
the sample gas decreases. This ensures a high output even for low oxygen
concentration.
1
Hydrogen Purity meter. (analyser)
The measurement principle used in the on-line hydrogen purity meter is the
thermal conductivity of gases. Gases have thermal conductivities, which are
different for different gases. The thermal conductivity of a gas mixture is
a function of individual conductivities and gas concentrations and is
approximately given by
Two sets of cell sealed in a glass are placed within a chamber; one flushed
with sample gas and the other with encapsulated reference gas. The sample
gas free from any measured components should have thermal conductivity
approximately same as reference gas. The four wires connected in a wheat-
stone bridge pattern is powered by a bridge current source, thus heating the
resistance wires and making them assume a temperature dependent on the
thermal conductivity of the reference or sample gas. The lesser the difference
between the thermal conductivity of the sample gas and that of the reference
gas, the lesser will be the effect of fluctuation in the ambient temperature on
the equilibrium of the bridge.
If the sample gas contains the measured component its thermal conductivity
changes thus changing the temperature and resistance in the wires flushed
by the sample gas; equilibrium of the bridge is disturbed. The resultant
2
bridge diagonal voltage is a measurement of the concentration of the
measured component in the sample gas.
where:
= Transmittance
I = Intensity of light into the medium
I = Intensity of light out of the medium
a = Attenuation coefficient
c = Concentration of pollutant
= Distance of light beam travels through the medium
O.D. =
=
3
By comparing a series of optical density measurements with gravity dust
measurements, it is possible to derive the value of K in real time application.
It should be noted here that in a double pass system a path length correction
factor is required for correcting the change in path lengths between the
measurement and exit diameters of the stack.
Equal intensity chopped infrared beams are passed through a measuring cell
(MC) and a reference cell (RC) simultaneously. The reference cell is filled with
a standard non-absorbing gas while the measuring cell contains the process
gas.
The beams from the MC and RC pass on to the detector D that consists of two
sealed absorption chambers A1 and A2 filled with the measuring gas and
separated by a thin metal diaphragm which forms a parallel plate capacitor
with an adjacent perforated fixed metal plate. These gas chambers will
therefore, selectively absorb radiation. Chamber A1 receives radiation which
has already been absorbed in parts in the MC and the gas in A1 will not
further absorb the same energy as would the gas in A2 so that the chamber
gases would be unequally heated to cause a movement of the metal
diaphragm to produce a change in the capacitance which is measured by a
bridge and amplifier-detector.
SWAS
SWAS is the commercial name adopted for the steam and water analytical
system. The instruments available under this system have a separate
sampling handling system and henceforth are not in-citu type. In the main
4
power plant these package is provided by ABB and for the DM plant lab.
these are mostly of YBL make.
The sensing probe for the dissolved oxygen is a membrane electrode, which
allows electro chemical reaction.
A typical cell has inner cylindrical anode made of lead and outer perforated
cathode usually of silver- the combination being immersed in an aqueous
solution of KHCO3, which is held in a position by a silicone rubber membrane.
This membrane is permeable to oxygen but not to water or other ions likely to
be present.
pH measurement
5
immersed. The inner tube also consists of glass wool placed at the bottom
where a small hole is made for bridging. The inner tube is placed in an outer
glass envelope containing sat. Kcl soln. The outer tube is provided with a
filling neck and a bridging hole for contact with the measuring solution.
The measuring electrode a glass electrode has special glass wall bulb at the
bottom, which separates a buffer solution contained in the bulb from the
measuring solution. The electrode wire is dipped in the buffer. Between this
buffer whose pH is known and constant and the unknown solution an emf is
generated which is proportional to the measuring PH.
Conductivity measurement
Silica analyser
The technique of analysis is partly chemical and partly optical with both
techniques automated in one single unit.
The chemical method used in the monitor utilizes the reaction between
silicate species in the sample, acid and molybdate reagents to form yellow
molybdosilicic acid complex. The acid conditions are chosen so that
specifically beta-molybdosilicic acid is produced which excludes interference
from other forms of the acid complex. To increase the sensitivity of the
method the yellow acid is reduced to the blue form, which is measured
colorimetrically in the optical system.
6
Sulphuric acid is added to the sample to reduce the pH to a value between
1.4 to 1.8 pH. .
The solution enters the first reaction coil in the temperature controlled block
(providing two minutes delay) where the yellow beta-molybdosilicic acid is
developed.
Sulphuric acid plus citric acid is added before the solution enters the second
reaction coil in the temperature controlled block (providing two minute delay)
to reduce the pH further to a value between 0.8 to 1.0 pH. This is the value
required to stop further development of silica and to provide conditions for
the next reaction (reduction) to take place. Citric acid is used to destroy any
phosphate complexes if present which would augment the colour developed
in the next stage.
The reduction solution (ascorbic acid) is added before the solution enters the
third reaction coil in the temperature controlled block (providing a one minute
delay) reducing the yellow complex to the blue form.
The fully developed solution passes to the measuring cuvette in the optical
system where the intensity of the colour is proportional to the original silica
concentration.
Sodium analyser
The sample passes through one half of a heat exchanger, which is used
during the calibration sequence to bring the standard solution close to the
temperature of the sample. This reduces the calibration time.
From the heat exchanger, the sample passes through a solenoid valve to the
constant head unit, which removes the effect of changes in the sample
pressure and flow-rate. The over flow unit also ensure self starting of the
system when the sample is lost and enables the monitor to function over a
wide sample flow.
The sample is then delivered to the T piece and a stainless steel entrainment
tube, where an alkaline vapour reagent (ammonia soln.) is added to the
7
sample to raise the pH value, before flowing past the sodium and reference
electrodes mounted in the flowcell. The sample leaves the flowcell and is
passed on to the drain.
where:
R = gas constant
F = Faraday constant
T = temperature in K
= Sodium ion concentration
K = a constant
A temperature sensor fitted in the flow cell detects the temperature of the
sample and is used for the automatic temperature compensation.
Hydrazine analyser
The hydrazine sensor and its overflow funnel are mounted on a separate
chamber whose height relative to standard solution container allows correct
rate of flow through the sensor. Even the flow can be adjusted by raising or
lowering the overflow tube in constant head unit.
and
8
9