Académique Documents
Professionnel Documents
Culture Documents
for FRCR
Simplified and edited by
(Volume 2)
Dr. Khaled Elazhry
Dr. Khaled Elazhry 1
Chapter 5
DIGITAL RADIOGRAPHY
FOV
Pixel size =
No. of pixels in the matrix
The value on the pixels is stored in binary format, and the maximum value is related to
the bit depth of the pixel
8-bit pixel has a (28) 256 levels of gray, and 12-bit pixel has a (212) 4096 level of gray.
Computer memory expressed as bytes , with single byte carry up to 8 bits , so for pixels
with more than 8-bits use proportionally increased bytes.
Matrix size more than 512x512 will not improve resolution because of the limitation of
detector size and sapling frequency.
Compression ;
Compression of the images is required to reduce the storage requirements and
transmission time.
Lossless (reversible) compression: compression factor of 2 to 3 with images can be
restored to its origin value.
Lossy (irreversible) compression: compression factor up to 40 with some loss of the
details & images cannot be restored to its origin value.
Noise Reduction :
Low-pass spatial filtration: average after adding a proportion of the grayscale values of
the neighboring pixels, BUT blur small details and edges.
High-pass spatial filtration (Edge enhancement): adding a proportion of the difference
between the grayscale values of the neighboring pixels, exaggerate edge contrast BUT
increase noise, and may be create false structures.
Image Quality:
Spatial resolution
Spatial resolution of CR is limited to 5.5 lp/mm for small plates and about 3.5 lp/mm
for larger plates, (Film screen system is about 8 lp/mm and 12 lp/mm for detail
screens).
The limiting factors of reduced spatial resolution of CR:
1- Pixel size ( the main reason )
2- Pixel spacing (sampling interval).
3- Scattering of the laser light in the phosphors layer.
4- The size of the phosphors grains.
5- Phosphors layer thickening.
6- The diameter and orientation of the scanning laser beam.
Increasing the ability to detect fine details in CR by :
1- The partial volume effect.
2- Edge enhancement algorithms.
Contrast
Contrast of CR is more than film screen system and determined by the processing
techniques.
Image density is independent of the exposure factors
Noise
Arise from random flocculation in the count of the X-ray quanta absorbed.
Inversely proportional to square root of fractional x-ray absorption efficiency (n).
Inversely proportional to square root of photon fluence incident upon the image (N).
Analogue image:
Displayed on monitors.
Stored on magnetic tapes
Digital image:
Made of pixels
Higher value pixel displayed darker
Can be stored on computers and disks
##################
FLUOROSCOPY
The absorption edges of Cesium (at 36 keV) & Iodine (at 33 keV) are favorably
placed in relation to the effective energy of the X-ray spectrum "which typically
have a mean energy about 25 keV".
Photocathode
Photoemissive metal "Cesium-Antimony photoelectric screen".
It is coated directly over the phosphor screen so, the light from the CsI input
phosphor passes directly & strike the photocathode emit Photoelectrons in
numbers proportional to the brightness of the screen.
(MCQ: the max. fluoroscopic tube current is mainly limited by Patient Dose)
Accelerating Anode
Locatedintheneckoftheimagetube(Fig.123).
Hasapositivepotentialof25to35kVrelativetothephotocathode
Accelerateelectronsemittedfromthephotocathodetoatremendousvelocitytoward
theoutputscreen.
The output phosphor image is viewed either directly through a series of lenses and
mirrors( in old systems ), or indirectly through closed-circuit television
Gain:
The image is intensified ( brighter ), minified , inverted & reverted
For each X-ray photon absorbed by the input phosphor about 400 light photons are
emitted producing 400 photoelectrons which cause the output phosphor to emit
nearly 400000 light photons.
Brightness gain : Also called intensification factor. Equals the ratio
brightness of the output phosphor
brightness of the input phosphor
The brightness gain of an image intensifier comes from two completely unrelated
sources, called "minification gain" and "flux gain".
Minification Gain
Produced by a reduction in image size.
The quantity of the gain depends on the relative areas of the input and output
screens.
d1
Minification gain = ( )2
d0
where dl is the diameter of input screen, and d0 is the diameter of output screen
Magnification (Zooming)
Image quality increase as the size of the input field is decreased.
[
The central part of the input image then fills the whole of the output phosphor (zoomed )
Effect of Zooming (adv. & Disadv.)
Magnifiestheimage.
Improvesthesharpness"resolution"
But, makes the image less bright "less minification gain" exposure factors are
automatically increased to compensate for the decreased brightness increases
thepatient'sskindose.
BUTtheXraybeamisautomaticallyrecollimatedvolumeofirradiatedtissue
(sopatient'sskindoseincreasesbutthetotaldoseismoreorlessthesame)
6.2 TV System
Original TV image intensifier systems used a lens system to focus the image
from the output screen on the input screen of the camera.
The camera has an electron beam scan for that produce a voltage (video signal ) which is
proportional to the intensity on the light falling on the signal plate in its input screen
Digital Fluoroscopy: the signal from the video camera is converted to digital format
that is input into a computer before displayed.
Computer functions:
1- Image storage.
2- Image processing functions as noise reduction ,edge enhancement or black to
white reversal.
3- Geometrical conversions ( left to right or top to bottom )
4- Mapping to display look-up tables to maximize contrast.
CCD Camera :
Produce a digital signal not an analogue which is more effectively.
Its input screen is a thin layer of amorphous silicon divided into pixels ( generally 1024
x 1024) , each as a small capacitor to collect charge proportional to the intensity of the
light falling on it , then the charges measured very rapidly ( about 30 frames /s )
Has increased dynamic range ( 12-bit image depth )
The CCDs cameras are now the standard for the image intensifier based fluoroscopy.
Optics :
The traditional lens system between the output screen and the input screen on the
camera is now replaced by (Fibro- optic coupling) that provide efficient light collection
and improved geometrical integrity.
It does not act in changing the x-ray dose rate to the patient unnecessary
patient exposure.
Not improve quantum mottle and will increase electronic noise.
RULE Lower input dose rate , the greater noise in the image.
PULSED FLUOROSCOPY :
The X-ray generator supply the voltage to the tube is switched on & off at regular
intervals ( about 25 or 30 pulses /second )
At these rates the eye cannot dedifferentiates between continues and pulsed modes, then
may be reduced by successive halving of the rate ( 30 then 15 , 7.5 , 3.75 down to 1
pulse/second )
The input dose rate reduces with reducing the pulse rate , with a proportion depends on
the equipment.
Each successive image is retained on the display until the following image displayed
Movement with low pulse rate may result in unacceptable blurring or Lag
Grid-controlled tube : instead of switching the tube , using an additional circular
electrode build into the tube between the cathode and the anode to switch the beam itself
on & off
Applying negative 2 kV voltage to the grid electrode will repel the electrons back to the
cathode
It give greater control over the intensity of the X0ray beam in each pulse.
Noise:
The information in the image also suffers from statistical fluctuations in all the other
discrete 'events' occurring in the image chain:
Light photons emitted by the input phosphor
Photoelectrons emitted by the photocathode
Light photons emitted by the output phosphor
Electrons constituting the video signal, etc.
The noise produced in each of the above stages is relatively low since the number of
light photons and electrons involved is very large.
The Quantum Sink is It is the weakest link in the imaging chain
2
The part of the system where the number of photons or electrons/mm of the image
field is lowest the relative noise the largest and the SNR the lowest.
In TV system, it lies in the absorption of X-rays in the input phosphor.
In radiography, it lies in the absorption of X-rays in the intensifying screens.
No increase of the gain of the intensifier or camera will improve SNR
COMPUTED TOMOGRAPHY
Matrixoftissuevoxels,correspondingmemory
PrincipleofCTimaging locationsinthecomputerorimagepixels.
From steps (a), (b), & (d) contrast is more than conventional radiography
From step (c) the spatial resolution is less good.
N.B. contrast resolution = the ability to display low contrast structures.
The image is displayed as a matrix of pixels, each 0.5 x 0.5 mm.
The brightness or grey scale value of each pixel in the image = the average linear
attenuation coefficient of the contents of the corresponding voxel
CT numbers
The average linear attenuation coefficient t of each tissue within the voxel.
Compared with that w of water by the formula:
(t - w)
CT number ( CTn) = 1000
w
Windowing
Although the scanner can distinguish different CT numbers ( 1000 to 3000 )
about4000shadesofgray,BUTtheeyecannotdistinguishmorethan50.
Awindowischosenwhichjustembracesallthetissuesofinterest,andonlythese
aredisplayedasshadesofgreywithintherangeblacktomaximumwhiteonthe
monitor.
PixelswithCTnumbersoutsidethiswindowareundifferentiated,beingdisplayed
aseitherblackorwhite.
Windowlevelandwindowwidthcanbesetindependentlyatthecontrolpanel,
forexampletodifferentiatelungtissues.
Theyonlyaffectthedisplayedimage.
COLLIMATION:
Generally,scannershavetotaltubefiltrationofabout6mmaluminum.
Prepatientcollimatorshasafixedwidthgenerally50cm.effectivelylimittheamount
ofxraysthatreachthepatientbylimitingthebeamsizeandreducescatterradiation
Prepatientcollimatorforsingleslicescannersdefinestheslicethickness,whichfixed
thicknesswidthsettingbetween1and10mm.
Postpatient collimator , in single slice scanners only , its purpose to reduce scatter
radiation to reach the detectors if the slice thickness less than the width of the
detectorsandtogetaccuratethicknessforsmallestslicewidth.
PostpatientcollimatorinMSCTisnotnecessary orpossiblebecausethefullwidthof
thedetectorsisusedtoformtheimage,soshieldingwillpreventdirectradiation.
DETECTORS
Thedetectorsneedtohave:
1. Highdetectionefficiency;
2. Fastresponse(shortafterglow)tokeepupwithfastscanning;
3. Widedynamicrangeabletocopewithboththehighintensitybeamineither
side of the patient and the highly attenuated beam passing through the
patient(intensityratio5000:1);
4. LinearitysignalaccuratelyproportionaltotheXrayintensity;
5. Stabilityinfaceofvoltageandtemperaturefluctuations;
6. Smallsizetoallowclosepacking,givingbetterspatialresolution.
IONIZATIONCHAMBERSDETECTORS.
Features:
Highdetectionefficacy,about60%
Lesssensitivebutmoreeasilymatchedforsensitivity.
Theyareverystableunaffectedbyvoltagefluctuations.
Haveawidelinearresponsewithnolag.
Theyarenarrow(andcloselypacked).
NOTsuitableformultislicescanners,Sonolongermanufactured
Tocompensateforthelowersensitivity,ionizationchampersare:
Filled with a high atomic number gas (xenon , Z=54 , Kshell binding energy 35
KeV)ratherthanair&athighpressure(20atm,2MPa).
Maderelativelydeep&thick(e.g.6cm).
SOLIDSTATEDETECTORSSCINTILLATORS
Usescintillatewithembeddedsiliconphotodiodetodetectthelightoutput.
Scintillatemaybecadmiumtungstate,bismuthgerminateorrareearthceramic.
Features:
Detectionefficiencyisveryhigh,about98%.
Negligibleafterglow.
2
Verysmall(downto1x0.5mm )
A small gap between detector elements is necessary to reduce crossover
betweenadjacentdetectorswillreducingthehighefficiencyto80%.
Inprinciple,ifwehave512x512voxels&800(detectorsno.)x300(pulsesno.)
raysumsenoughdatafortheCTnumbersofallthevoxels.
TheCTnumberssocomputedarestoredinthecomputermemorylocations,each
ofwhichcorrespondstoavoxelandthereforetoapixel.
Backprojection:
Illustration(1)
representation.
(d) : Final representation , which is
not an accurate reproduction of the
original image but with approximate
relativevalues.
Theblurringintroducedbythebackprojectionprocessofthepencilbeamcan
becompensatedbyamathematicalprocesscarriedoutbythecomputercalled
'filtering'usingthedatanotjustfromthepencilbeamitselfbutalsofromthe
neighboringpencilbeamsbysubtractionoraddingmodifiesthebrightness
neartheedgeofeachbackprojectedbeam.
Data may be reconstructed from a reduced arc down to 180 , not full 360
rotation.AsinCTfluoroscopy.
There are different filters and back projection algorithms may be applied
dependsontheimagingtask.
Filteringalgorithms
'Bone'algorithmenhancefinedetailbutincreasenoise.
'Softtissue'filterimprovecontrastbysmoothingoutnoise,butimpairs
thespatialresolution.
Scannedprojectionradiographs:ScanogramScoutview
Scanned projection with a fixed projection angle with the collimator set at
narrowestslicewidth.
The image is different from the conventional radiograph because there is no
beamdivergencealongtheZaxis.
Computedtomographyfluoroscopy:CTfluoroscopy
DisplayofaCTimageinrealtime.
Donebycontinuousrotationofthegantrywithouttablemovement
Usingfastreconstructiontechniques,generallyfrom180datasets.
Usedforbiopsyneedleplacement.
Lesseffectivedose,BUThighpatientskindose,becausescanningisconfined
tonarrowregion.
SPIRAL(HELICAL)SCANNING
The couch moves continuously at a steady speed while the tube and detectors
makeanumberofrevolutionsaroundthepatient.
The tube receives its power supply through slipring technology, and the
detectorssendtheirsignalsbyradio.
Thedataareacquiredintheformofacontinuousribbonofcontiguousslices.
Advantages:
1. Itisfasterallowingagreaternumberofpatientsandalsotheuseofasmaller
volumeofcontrastmedium.
2. It overcomes the problem of slicetoslice misregistration, particularly in the
region of the diaphragm, caused by variations in inspiration between the
separatebreathholdsneededinsequentialscanning.
3. It reduces partial volume artefacts since the reconstructed slices can be
incrementedinsmallsteps.
4. Because of the volume acquisition of data, resolution in the axial direction is
good,andreformattingintootherplanesisimproved.
Disadvantages:
1. Nocoolingperiodsbetweensliceshighheatloadingofthetube.
2. EventhoughahighcapacitytubeandsensitivedetectorsareusedalowermA
must be employed patient dose but noise, particularly with thicker
patients.
3. Thereissomelossofspatialresolutionduetotheinterpolationprocess.
4. Ahighspeedcomputerwithalargedatastoreisnecessary.
Pitch
Def.: the distance (millimeters) moved by the table during one rotation of the
tubedividedbytheslicethickness(millimeters)
Increasingthepitch,byincreasingthetablespeed,islikestretchingthespring;
Speedsupscanning.
Reducesdose,
BUTresolutionmaybelost"greaterinterpolationneeded".
Decreasingthepitch,willbeoverlap
Aboveapitchof2therearegapsinthevolumebeingscanned,andartefactsmay
ariseandgivesunacceptableimages.
MultisliceReconstruction:
Principally , it is the same as single slice helical reconstruction , with the
interpolationofthemeasureddatabetweenadjacentdatasets.
ConeBeamEffect:itistheeffectofbeamdivergence,havealittlevaluein
single slice scanners, BUT in multislice scanners the reconstruction is
complicated because of this effect which increases with increased number of
slicesandwithincreasingoftotaldetectorlength.
Correctionofconebeameffect,isdonebespecialalgorithmstominimizethe
influenceofthiseffect.
Multiplanarimaging(sagittal,coronal,andobliquesections)
Threedimensionalsurfacerenderingshadedsurfacedisplay(SSD)
Maximumintensityprojection(MIP)&Minimumintensityprojection(MinIP)
Virtualendoscopy.
SpatialResolution:
ThemaximumhighcontrastspatialresolutionthatmaybeachievedbyCTscanis
about20lp\cm2lp/mm(inCTweusecm,notmmlikeinconventional&
fluoroscopy),generallyitisfrom6lp\cmto10lp\cm0.6to1lp\mm
HighcontrastspatialresolutionisdeterminedmainlybyPixelSize.
Byincreasingmatrixsizeorreducingfieldofviewdecreasingpixelsize.
(matrix512x512usedwith40cmFOV,haspixelsizeabout0.8mmleadingto
resolutionabout6lp\mm)
Highcontrastspatialresolutionislimitedby:
1) PixelSize.
2) Theusedalgorithm.
(certainalgorithmsdesignedspecially,likeinbonystructuresenhancement)
3) Thesamplingfrequencynumberofprojectionsampledineachrotation.
(withinlimits,increasingsamplingwillincreasespatialresolution)
4) ThewidthoftheprojectpathoftheXraybeam:
Whichisfurtheraffectedby:
Focalspotsize.
Thegeometryofthescanner.
Physicalsizeofthesensitiveareaofthedetector.
5) Spatialresolutioninthezaxisalsodependsonslicethickness.
(increasingthepitch,willreducespatialresolutioninzdirection)
Noise:
Types:
A)Quantumnoise:causedbyrandomvariationsinthenumberofXrayphotons
absorbed.
B)Electronicnoise:producedinmeasuringsystem,notsignificant.
C)Structuralnoise:thatisaffectedbythereconstructionalgorithm.
1. Increasingthenumberofphotonsabsorbedineachvoxel,byincreasingtheslice
thicknessorwidth.
Improving contrast resolution due to less noise , BUT reduce spatial resolution
andincreasepartialvolumeeffect.
2. IncreasingeitherthemAorthescantime.
Bothoftheseinvolveincreasingthepatientdose.
3. IncreasingofkV.
due to increasing of penetration and number of photons detected ( even if mA
reducedtorestorepatientdose).
4. IncreasingFOVorsmallmatrixsize.
leadingtolargerdetectorareaineachpixelwithmorephotonsperpixel.
IncreasingPITCHdoesntaffectnoiseinsingleslicescanners: becausedataare
interpolated between successive rotations for 360 scan and the same number of
detectedphotonsiseffectivelyusedinthereconstruction.
YET ,increasing pitchwill increase slice width , leading to reduce spatial resolution
andincreasepartialvolumeeffect.
IncreasingPITCHwillincreasenoiseinmultislicescanners:becauseofthecone
beam effect , the interpolation algorithms will produce a net effect with minimal
boardingofthescansliceinthezdirection,increasingpoise.
IncreasingWindowWidth,willNOTincreasenoiseorSNRbutwillmakeitmore
apparent.
Contrast:
ContrastresolutionistiedtotheSNR
InCT,softtissuecontrastisbetterthanfilmradiographyatleast10timesbecause:
o Notobscuredbyoverlyingbone.
o SmallerScatter.
o WindowingallowsquitesmalldifferencesofCTnumbers
ii. Highattenuationobjects
Neurosurgical clips, dental amalgam, Small areas of bone or contrast medium
etc.,giverisetostarartefactswhichmayobscuretheareaofinterest.
Theeffectisaccentuatedbymotion.
iii. Photonstarvation
Variantofstreakartefacts,asinCTscanofthepelvisofpatientwithbilateralhip
implants , the tissue between the hips are poorly represented due to high
attenuation.
iv. Beamhardening
HardeningofthebeamasitpenetratesthepatientresultsintheCTnumberof
thesamekindoftissuedecreasingalongtheray.
ThereconstructionprocessassumesahomogeneousXraybeam,withtheresult
thatCTnumbersarelowerinthecenterofthepatient(knownas'cupping').
Correctionby:
Usinga'BeamHardeningAlgorithm'bythecomputer
'Bow Tie" compensating copper filter is used. diminishing patient thickness
towardtheedgesofthefanbeam.producesarelativelyhomogeneousbeam.
v. RingartefactDetectormalfunction
vi. Partialvolumeorvolumeaveragingartefacts(discussedbefore..)
vii. Conebeamartefact.(discussedbefore..)
CTrepresentonlyabout4%ofionizingradiation,yetitisresponsibleforabout
40%oftotalpopulationmedicalexposureinUK.
Although the detector collimators are set to the nominal slice thickness, the
actualXraybeamsoverlap,astheirwidthismuchgreater,beingdeterminedby
thecollimationnearthetube.
TheCTdoseindex(CTDI):
Itisthemeasureofdosefromsinglerotation.
Usedforcalculatingspreadofdoseoutsideanominalslice.
D z dz
CTDI =
T
where: = the integral sign
D(z) = the dose at position z
dz = complete dose profile
T = nominal thickness of the slice.
CTDIw
CTDIvol
pitch
Maximum skin dose is approximately equal to CTDIw for head scan and 20%
higherforbodyscans.
DLP = CTDIvolxL
L = the total scan time = number of rotations x collimated length x pitch
Effective Dose ( E ) mSv , can be derived from DLP depending on the body
regionandonthescannerdesign.
E/DLP
Head 0.0023
Chest 0.018
Liver 0.015
Abdomen & Pelvis 0.017
The difference of E/DLP between body regions is due to the distribution of the
organs and tissue having the highest tissue weighting factor wT used for
calculatingeffectivedose.
Effectivedoseisdirectlyproportionaltothetubecurrentandtotalscanlength.
CTDIW DLP E
Head 60 700 1.5
Chest 14 400 7
Liver 16 350 5.5
Abdomen & Pelvis 16 500 8.5
Factorsinfluencingpatientdose:
1) kV:increasingkVincreasesthedose,butreducingmAmayreducethedosefor
thesameimagequalityandnoise.
2) mA&scamtime(mAs):increasingmAsincreasesthedose,andreducingthe
noise(noiseininverselyproportionaltosquarerootofmAs).
3) Reconstructedslicewidth:
In few single slice scanner slice width identified by post patient collimation ,
usingnarrowestslicewidth,significantlyincreasethedose.
In both single & multislice scanners , using narrower reconstructed slices
increasesthedose,duetoassociatedincreaseinmAs.
5) Pitch: pitch is inversely proportional to the dose , so increasing the pitch to 1.5 leads to 33%
decrease in the dose.
6) Numberofdetectorsrows(inmultislice):hasnoinfluenceondose,butsome
scannershasmoregeometricalefficiencymaycausedosereduction.
7) Reconstruction algorithm & FOV: has no influence on dose , but will influence
themAwhichwillaffectthedose.
##########################
GAMMA IMAGING
In isomeric transformation: ray is not emitted immediately, but after appreciable time from
emission of particle.
For example,
- - ,
99 Mo , 99mTc 99Tc
42 43 43
Stable nuclide
67 h 6 h 0.2 Ma
The daughter nucleus technetium-99 (99Tc) remains in the excited state for 6 hours.
It is said to be metastable, and is written technetium-99m (99mTc).
Technetium-99m decays to the ground state 99Tc, with the emission of a -ray of
energy 140 keV.
Isomeric transition here result in rays only "no ray emission"
99Tc & 99mTc are said to be ISOMERS:
Def. of isomers: nuclei having different energy states but otherwise
indistinguishable as regards A number, Z number and other properties.
Another isomeric transformation used in nuclear imaging occurs when rubidium-81 (81Rb)
decays to kryplon-81 (81Kr)
81
Rb 81mKr 81Kr
A proton change into a neutron and a positive electron the latter is ejected from the
nucleus with high energy "+ve particle".
p n + +
For example,
Carbon-11 (11C, Z= 6) Boron-11 (11B, Z= 5).
The nucleus capturing an extra-nuclear electron from the nearest (K) shell increase its
number of neutrons relative to the number of protons.
p+en
Same Mass and charge.
The atomic number decreased by 1 .
The daughter nuclide will emit characteristic X-rays when the hole so created in the K-
shell is filled by an electron from an outer shell gamma rays if still excited.
Iodine-123 (123I) decays by electron capture and emits 160 keV gamma and 28 keV X-
rays, but no -particles.
Gamma Rays:
GammarayshaveidenticalpropertiestoXrays.
Thegammaraysemittedduringradioactivedecayhaveatmostalinespectrumwhich
ischaracteristicofthenuclidewhichemitsthem.
Forexample,131Iemitsmostly360keVgammarays.
The rays emitted by nuclei photoelectrically absorbed in its Kshell emit both
photoelectronsandcharacteristicXrays,usuallyoffairlylowenergy.
Beta rays
HaveacontinuousspectrumofenergiesuptoamaximumEmaxwhichischaracteristic
oftheradionuclide.
TheiraverageenergyisaboutEmax/3.
When positive and negative beta particles travel through a material excitation &
ionizationofatoms"trackoftheparticleisdottedwithionpairs"tillthelossofthe
wholeinitialenergystop"endofitsrange".
The range of beta particles is greater with initial energy of particle &
densityofthematerial"inverselyproportional".
Themostenergeticbetarayshavearangeofabout2mmintissue.
Positron emitters
Positron(=positiveelectrons)isantimatterverybriefexistence.
FATE:Whena+vebetaparticlecomestotheendofitsrangeitcombineswithave
electronthechargesneutralizeeachother,andthemassesofthetwoelectronsare
whollyconvertedintotwophotonsofannihilationradiation(eachof511keV)&travel
inoppositedirections
AccordingtoEinstein'sformula
E=mc2
Positronemittersareusedinpositronemissiontomography(PET)imaging.
The activity can be regarded as having an effective halflife teff which is shorter
thaneitherthebiologicalorphysicalhalflives.
1/teff=1/tboil+1/tphys
Theeffectivehalflifedependson:
a) Theradiopharmaceutical. c) Personalvariations.
b) Theorganinvolved. d) Healthstateoftheorgan.
8.3 RADIOPHARMACEUTICALS
Desirable properties of a radionuclide for imaging are:
1. A physical half-life of a few hours, similar to the time from preparation to injection.
If the half-life is too short, much more activity must be prepared than is actually injected.
2. Decay to a stable daughter or one with a very long half-life (e.g. 200 000 yrs for 99Tc).
3. Emit rays (which produce the image) but no or particles or very low-energy photons.
Decay by isomeric transition or electron capture is preferred.
4. Emit gamma rays of energy 50-300 keV (ideally 150 keV)
high enough to exit the patient but low enough to be easily collimated and easily measured,
being largely absorbed in the collimator septa and the crystal.
5. Ideally emit monoenergetic gamma rays so that scatter can be eliminated by energy
discrimination with the pulse height analyzer.
6. Be easily and firmly attached to the pharmaceutical at room temperature & not affect its
metabolism.
7. Be readily available on the hospital site.
8. Have a high specific activity.
Fig. 5.10
Decay of activity of 99Mo and growth and
regrowth of activity of 99mTc in a generator which
is eluted daily.
Positron emitters are needed for PET , the commonest PET radionuclide is 18F (half-life
110 min) and often used in the form 18F 2-fluoro-2-deoxyglucose (2-FDG ) for brain & heart.
Other radionuclide are 11C , nitrogen-13,oxygen-15 and rubidium-82 82Rb , these have very
short half-life 20 min , 10 min , 2 min , 75s respectively
82Rb produced from strontium-82 generator and last about 1 month. Used for myocardial
perfusion imaging
Technetium-94m produced by cyclotron with half life 50 min , and used to label many
pharmaceuticals already available for 99mTc imaging.
Preparation of radiopharmaceuticals
By simple mixing at room temperature:
Radionuclide (e.g. 99mTc) + Pharmaceutical (e.g. MDP) + other necessary chemicals
Shielded syringes transfer the components between sterile vials.
Manipulations under sterile conditions in a 'workstation' glove box or laminar downflow
cabinet.
Room is under a positive pressure of filtered sterile air.
Impervious surfaces: continuous floors- gloss-painted walls- Formica-topped benches.
Entry is via an air lock and changing room.
The pharmacy must meet the conditions of both the Medicines Act and the Ionising Radiations
Regulations in the UK.
Instead of a large number of tiny detectors, behind the collimator there lies a single
largephosphorcrystal500mmindiameterand912mmthick.
ItismadeofSODIUMIODIDE(activatedwithatraceofthallium).
NaIhaveahighatomicnumber(Z=53)anddensityitabsorbsabout90%of 99mTc
gammarays,principallybythephotoelectricprocess;butonly30%of131Irays.
Itisfragile,hygroscopicandeasilydamagedbytemperaturechanges.
ToprotectitisencapsulatedinanaluminumcylinderwithoneflatPyrexface.
producedbyrayssufferedComptoninteractionsinthepatientorthe
crystal.
Subsidiary 'iodine escape peak'
At 30 keV below the photopeak.
Due to some of the K-characteristic rays from iodine escaping from the crystal.
Only pulses in the photopeak are of use in locating the source of the radioactivity
in the patient, A pulse height analyzer is used to reject those in the Compton tail
TheZpulsesenterapulseheightanalyzerrejectpulseswhichare(a)lowerthana
presetvalueor(b)higherthananotherpresetvalue.
Itallowsonlypulseswhichliewithinawindowof,say,10%ofthephotopeakenergy.
Anyhighenergybackgroundradiationisalsorejected.
Thepulsessoselectedarereferredtoas'counts'.
In the case of 99mTc the window might be set at 126-154 keV, centered on 140 keV.
Because a 140 keV photon will lose only 10 keV of energy even when scattered through
45 some scattered photons may 'pass through' the window & produce counts
degrade the image.
For 67Ga or 111In 2 or 3 windows must be used each selecting one ray energy.
TheX,Y,andZpulsesarenextapplied:
1. Directlytoamonitorforvisualinterpretation"inoldermachines",or
2. Newersystemsanalogtodigitalconverters(ADCs)intoacomputer,enables:
Dynamicandgatedstudies.
Imageprocessing.
The monitor
TheXandYpulsessteertheelectronbeaminthemonitortube.
If and only if the Zpulse has passed through the window of the pulse height analyzer a
pinpointoflightappearmomentarilyonthescreenattheXandYcoordinatescorresponding
totheheightsoftheXandYpulsesandthepositionsoftheradioactiveatomemittingeach
gammaray.
Thousandsofsuchdots,equallybright,makeuptheimage.
Twotypesofmonitorareused:
1. Longpersistencescreen:
Eachdotoflightpersistsforasufficientlylongtimeforavisualimagetobuildupofthe
radioactivityinthepatient.
Theimagequalityisnotgoodenoughfordiagnosisonlyhelpfulforpositioningthe
patientandmakingsurethattheactivityhasbeentakenup.
2. Shortpersistencescreen:ofhighquality;
Eachdotquicklydiesawaybutcanbecapturedonfilm,usingafilmcamera.
Typesoffilmcamera:
1. Polaroidfilmcamera:
Adv.: highspeedhighresolutionprocessedveryquickly.
Disadv: expensivelowcontrastlimiteddynamicrange.
2. Multiformatfilmcamera:
Employs singlecoated cut transparency film, formulated for Xray automatic
processor.
Anumberofimagescanberecordedinsequencesidebysideonasinglefilm.
Ifthefilmisexposedfortoolong,itwillbesaturated;
iffortooshorttime,theimagewillbeunderexposed,faint,andgrainy.
Typically, a total of 0.5 million counts is acquired for each image,
takingabout5min.
Its a gamma camera with a parallel hole collimator rotates slowly around the patient. Every
6, the camera halts for 20-30s and acquires a view of the patient; 60 views with about 3
million counts are acquired in an overall scanning time ( around 30 min ).
Acquisition time is halved or the sensitivity improved by using a dual or triple-headed camera.
Gamma ray attenuation that takes place in the patient would result in fewer counts from the
centre than from the edges, in reconstructing the image. An approximate method of attenuation
correction is to add the counts, pixel by pixel, in each pair of opposing views where the
combined counts are then more nearly the same.
A mathematical 'filtering' process uses the 60 sets of data (or two-dimensional projections) are
synthesized into a set of transverse images by methods similar to those used in X-ray CT -
filtered back-projection , YET an area of very high activity (e.g. the bladder or the injection
site) can cause streaking of the image and even hide adjacent areas of activity.
Iterative reconstruction, An alternative algorithm, whereby the image activity content is
adjusted in steps until the calculated activity in the image is as close as possible to the
measured activity (counts).
Advantages of iterative reconstruction:
- It is less sensitive to streaking artefacts.
- Can correct accurately for attenuation.
- Insensitivity to noise.
- The ability to reconstruct an image even when data acquisition is incomplete.
Positronemissiontomography(PET)
ThemostcommonpositronemitterusedinPETis18F.
When 18Femitsapositivebetaparticle(positron),thistravelsforabout2mmthrough
the patient before being annihilated by a negative electron. Their combined mass
(positron plus electron) is converted into two energetic photons, each of exactly 511
keV,emittedsimultaneouslyandinpracticallyoppositedirections.
PETimagingisbasedondetectingthesetwoannihilationphotonsincoincidenceand
identifyingtheirorigininthepatienttolocatetheradioactivesource.
A positron or PET camera comprises a ring or hexagon surrounding the patient and
composedofaverylargenumberofsolidscintillationdetectors(1000020000),often
of bismuth germanate (BGO). OR lutetium oxyorthosilicate (LSO) or gadolinium
oxyorthosilicate(GSO).
PETdetectorsaremadeinblockformat,coupledtophotomultipliertubes.
If,asinFigure8.17,theannihilationphotonsfromtheeventat(a)enterdetectorsA
and B, they produce simultaneous (coincident) pulses, which are then accepted and
combinedbytheelectronics.Anypulsesthatdonotcoincideintimeareignoredbythe
electronics, as are any single photons of background radiation. These two detectors
thereforemeasureonlythesumoftheactivitypresentalongalineAB,calledtheline
ofresponse(LOR)andsimilarlyforeachofthemanypairsofdetectorswhichisjust
theinformationrequiredfortomographicreconstruction,asinCTorSPECT.However,
asthedetectionmechanismisdifferent,soisthedataacquisition.
Figure8.17 Figure8.18
Figure8.17 Apositronemissiontomographyimagingsystemshowingsomelinesofresponse.
TheeventataisdetectedatAandBincoincidence.
Figure8.18 Differentcoincidenceevents:a,truecoincidence;b,randomcoincidence;andc,
scattercoincidence.
Therearethreetypesofcoincidencesthatcanoccurtrue,random,andscatter,
Toreducetherandomandscattereventsfromadjacent.rings,narrowleadortungsten
septacanbeusedbetweeneachringofdetectorstoactlikeaparallelantiscattergrid
inradiography.
InmodernPETscannerswithblockdetectorssetinaseriesofrings,dataacquisitionis
fastandcontiguous.TheaxialFOVisgivenbythewidthofthecompletesetofrings.As
in CT, the patient is moved along the axis of the. Scanner to obtain the next set of
sequentialtransverseslices.
Uniformity of field
Tested by a flood field phantom consists of a flat sealed dish, larger than the field of view
(FOV), filled with 99mTc or its longer lived analog cotalt-57 (57Co).
Should give a uniform image, with and without the collimator in place
A defective photomultiplier seen as a dark area in the image.
A cracked crystal a linear defect.
Slight differences in the performance of individual photomultipliers Non-
uniformity due to:
Isassessedbyinstructingthecomputertocomputeahistogramofthecounts
inindividualpixels.(Asufficientlylongexposureisnecessarytoreducenoise).
Itcalculatesthemean&standarddeviationtypicallytheuniformityis12%.
Modern gamma cameras can be instructed to compensate subsequent images
automaticallyforminornonuniformitiesofthefield.
The detector uniformity of PET scanner is checked using
longlivedsource(68Geor137Cs)mountedonthegantryand
rotating it around the field to expose all the detectors
uniformity, OR standard phantom with centrally located
positron source. The sinogram should be uniform, any
malfunction will appear as streak covering all the angles
witheachdetectorblock.
Spatial resolution:
Intrinsic resolution:
Refers to THE CAMERA (crystal, photomultipliers, and position logic circuits) in the absence of
the collimator and patient.
Total blurring due to Intrinsic blurring, Collimator & Scattering of rays in the
patient (according to the same formula as for radiographic blurrings).
Resolution worsens the farther the activity from the collimator. Consequently it is
worse for fat patients than for thin.
Testing the resolution:
The computer is instructed to plot the counts along a line of pixels at right angles to a thin
tube filled with 99mTc. The graph (Fig. 5.8) is called "line spread function".
[
Against the crystal face after removing the collimator (dotted curve).
The FWHM of this curve measures intrinsic resolution, typically 1 - 2 mm.
Linearity
Linearity (i.e. lack of distortion) can also be checked by imaging a line source.
Importance:
Better energy resolution better scatter rejection & better spatial resolution
Calculated from FWHM of the photopeak divided by the energy of the photon & expressed
as a percentage
Typically 12% of the peak energy
Flashoflightproducedbyagammaphotoninthecrystalhasadecaytimeof0.2s
About 95% of the light has been emitted in 1 s, at which point the pulse is cut off
electronically.Forthatperiodthecountingcircuitdoesnotrecognizeotherpulses,and
issaidtobe'dead'.
Duetothestochasticnatureofradioactivedecay,countsarriveatirregularintervals.
If 2 gamma photon enters the camera during the dead time (indicated by vertical
dashedlinesinthefigure)
thetwoflashesoflightoverlapandtheyaretreatedasone.
Ifthecombinedpulseistoolarge,itwillberejectedbythepulseheightanalyzer.
IftwoComptonscatteredphotons(60and80keV)enterthesystemwithin1sofeach
other recorded as a single photon (140 keV) pass through the pulse height
analyzer&produceaspuriousimageinafalselocationdeteriorationofthespatial
resolution
Lost counts:
Athighcountrates,asignificantproportionofthegammaphotonsaremissed
underestimatedcountrate
Typically there is 20% loss at 40000 cps and more at higher count rates as in cardiac
imaging.
In order to minimize patient dose, patients should drink a good deal of water and
empty the bladder frequently to reduce the dose to the gonads and pelvic bone
marrow.
Theactivityofeachadministrationofradiopharmaceuticaliskeptwithinthelimitsset
bytheAdministrationofRadioactiveSubstancesAdvisoryCommittee(ARSAC)ofthe
UKDepartmentofHealth,andchecked&recordedbeforeadministration.
Thephialisplacedinthe"well'ofalargereentrantionizationchamber,knowas:
Theradionuclideordosecalibrator
The ionization current produced by the gamma rays is:
Proportional to the activity of the sample.
Depends on the gamma energy and half-life of the radionuclide.
The radionuclide is entered on the control panel and the activity in MBq is displayed on a
digital read-out.
The accuracy of the radionuclide calibrator must be checked regularly using a reasonably
long-lived source, such as 57Co.
The calibrator can also be used with a lead sleeve to check for the higher-energy gamma rays
from molybdenum impurity.
Aqueous liquid waste well diluted with water disposed of via designated sinks or
sluices with drains running direct to the foul drain.
Solid waste (swabs, syringes, bottles, etc.) placed in designated sacks for disposal by
incineration or, if suitably diluted with ordinary waste, to waste disposal sites.
Old generators are kept in a secure shielded store until they are returned to the manufacturer.
Contaminated clothing and bedding is appropriately bagged and stored in a secure protected
area until sufficiently decayed for release to the laundry.
Records must be kept, for inspection, of all deliveries, stocks, administration, stored waste, and
disposals of waste.
###########################
Chapter9
ULTRASOUNDPHYSICS
Dr. Khaled Elazhry 70
9.1 PiezoelectricEffect:
Pulsedmode:
By applying a few hundred volts DC , leading to disc
expansion , producing a wave of compression in the
adjacent material that travels with a velocity ( ) ,
followed by a corresponding wave of decompression or
rarefaction(Figureb,c)
Continuousmode:
By applying an alternating voltage AC, the transducer
act as a piston producing successive compressions and
rarefactions.(Figured)
Wavelength (), distance between two successive
compressionsorrarefactions
Frequency(),numberofcompressionspassesatanygiven
point,equalsthefrequencyoftheAC,measuredinMHz
Remember,=
9.2 INTERFERENCE:
Naturalorresonantfrequency:
In continues mode , control of the
ultrasoundfrequencyisdonebycontrolling
thefrequencyoftheAC
Largest output , is achieved when the
wavelengthoftheproducedwavesequalto
twicethethickness(t)ofthepiezoelectric
disc , because of the full constructive
Interference or ( resonance ) of the wave
comingfromthe front faceofthe disc and
itsreflectionatthebackface.
Natural or resonant frequency of a transducer, it is the frequency which the
transduceristhemostefficientasatransmitterandalsowhichitismostsensitiveasa
receiver.
9.3 SingleTransducerProbe:
Backingblockfunctionis,scatteringandabsorbingofthebackwardswavesproduced
bythebackfaceofthedisc,leadingtodampingofthevibration,resultingin:
Decreasethepulselength.
improvingaxialresolution
LowQ
Increasesbandwidth,causingdecreasestransducersensitivity.
Bandwidth:
Itisuncommonforatransducertoemitasoundbeamwithasinglefrequency;rather,
thereisabroadrangeoffrequenciesemitted.
The bandwidth is the difference between the highest and the lowest frequency
emittedfromthetransducer.
Thebandwidth,isthefullwidthathalfmaximumintensity(FWHM).
Transducer bandwidth is inversely related to pulse length (duration). pulse length is
determined by the wavelength and number of cycles in the pulse. The number of
cycles, or waves, in the pulse is determined by the damping characteristics of the
transducer.
Highly damped transducers are characterized by short pulses and wide bandwidths
and poorly damped transducers are characterized by long pulses and narrow
bandwidths.Therefore,frequencybandwidthisdirectlyproportionaltodamping.
DAMPING BANDWIDTH
TheMechanicalCoefficient:Qfactor
Isaunitlessnumberthatrepresentstheabilityofthetransducertoemita"clean"or
"quality"ultrasoundfrequency.
ThehighertheQfactor,thenarrowerthetransducerbandwidthasatransmitterand
receiver.
Transducers with high Q is good for continuous mode as it produce pure note and
respondsonlytothatnote.
Transducers with low Qisgood forpulsed mode asit has short ring downtime and
producesshortpulsesandrespondstoarangeoffrequencies.
MH
Qfactor=
MH
BANDWIDTHQFACTOR
RESONSNTFREQUENCYQFACTOR
Diameterofthetransducer:
Ifthediameterequalsonewavelengthorless,soundwavewillspreadinalldirections.
Ifthediametermuchgreaterthanonewavelength,soundisprojectedforwardsasa
plane wave, due to reinforce in this direction and destructive interference in other
directions.
Nearandfarfields:
Near field: also called Fresnel zone is the proximal part of an ultrasound beam.
adjacenttothetransducersurfaceandhasaconvergingnarrowbeamshapedueto
theconstructiveanddestructiveinterferencepatternsofsoundwavelengthsemitted
fromthetransducercrystal.
Far field: also called Fraunhofer zone is the distal part of an ultrasound beam
characterized by a diverging shape and continuous loss of ultrasound intensity with
distancefromthetransducer.
Higherfrequencytransducers,increasesthelengthofthenearfiledanddecreasesthe
divergenceofthefarfield(forthesamediscdiameter).
Largerdiametertransducers,increasesthelengthofthenearfiledanddecreasesthe
divergenceofthefarfield(forthesamefrequency).
Focusing:
Isfocusingtheintensityofthesoundbeanataparticulardepth,producingstronger
echoesandimprovingthelateralresolution.
Internalfocusing:byusingacurvedpiezoelectricelement(thegreaterthecurvature,
theshorterthefocallength)
Externalfocusing:byusingplasticacousticlenscementedtothetransducerface,or
usingcurvedmirrors.
Theshorterthefocallength,theincreaseddivergenceofthe
farfieldandthenarrowerandshorterthefocaldepth
Electronicfocusing:(annulararray)
Isarchivedbysubdividingthepiezoelectriccirculardiscintomultipleconcentricrings,
withfirstenergizingtheoutermostringthesubsequentlytheinnerrings,withapoint
(p)atwhichthepulsestravellingatdifferentpathsarrivetogetherandreinforce.
Shortening of the focal length can be made by increasing the time delay between
energizingsuccessiverings.
AcousticImpedance:(Z)Kg/m2S(Rayl)
(Istheproductofthedensity()ofthematerialandthevelocity()ofthesoundinit)
Z=
Theproportionsofenergyreflectedandtransmitteddependontheacousticimpedance
ofthetwomaterialsmakingtheinterference.
The acoustic impedance depends on density and elasticity of the material and
independentoffrequency.
Z Density
Reflection(notrefraction),dependsintheacousticimpedancemismatch
Theacousticimpedancedifferenceisveryhighbetweenairorgas(lung)andbonewith
other body tissues, with total reflection, and no information can be collected from
organsplacedbehindthem.
Couplinggelbetweentransducerandbodyisimportanttopreventreflectionbyair.
The fraction of sound energy that is reflected ( R ) at the interference between two
materialsdependsontheangleofincidence
Whentheangleonincidenceis90ornearly,RdependsontheZ
R=(Z1Z2)2/(Z1+Z2)2
R Z
Matching:whenZ1=Z2,therewillbe100%transmissionwithnoreflection
Matchingplate:usedtoreducemismatchbetweentransducerandtissues,itisa
plastic compound of wavelength thickness having an acoustic impedance
intermediatebetweenthatofthetransducerandtheskin.
Small fractions of ultrasound are reflected between different soft tissues and
reflectionlessthan0.01%isunlikelytobedetected.
Specular(mirror)reflection:
if the beam strikes a large smooth interface, the same laws of reflection and
refraction applies to it like light ray, where the incidence angle equals the
reflectionangle
therefractionangleissubmittedtoSnellslaw:
Ithappenswhenthereflectionsurfaceisroughandhasundulationsequaltothe
wavelength,sothereflectedbeamwillspreadoutoveranangle
Theshorterthewavelengthandtherougherthesurface,thewiderthespread.
Scatter:
It happens when a sound encounters a structure that is much smaller than its
wavelength(asredbloodcell),itscatteredmoreorlessinalldirections.
Thiseffectallowsevensmallstructurestobevisualizedassomescatterwillreach
thetransducer.
Bloodflowimagingdependsonscatteringbyredbloodcells.
9.5 ATTENUATIONOFULTRASOUND:
Attenuationofthesoundbeamoccursdueto:
Energy is absorbed and converted to heat by friction and viscous forces in the
material.
Energy leaves the forwards travelling beam due to scattering and to partial
reflectionbythebydifferentinterfaces.
Thehigherthefrequency,thegreatertheattenuation.
Attenuationisexpressedindecibels(dB)
Decibellosspercentimeterisproportionaltothefrequency,e.g.at3.5MHzthelossis
about3.5dB/cm,soin15cmjourneyto&frothetotallossabout(3.5x30)=100dB
Halfvaluelayer:
Thethicknessoftissuethatreducesthesoundintensitytohalfitsoriginalvalue.
Water;thereisalittleabsorptionorscatterinwater(fullbladdercanaidpenetration).
Bone;attenuationismuchgreater(about35dB/cmat2.5MHz)
Air;attenuationisgreat(inlungsabout40dB/cmat1MHz)
Penetration:Thehigherthefrequency,thelesseffectivethepenetration
Penetration=40/
((Halfvaluelayerpenetrationisinverselyproportionedtofrequency))
9.6 AMODE:(AmplitudeMode)(Echoranging)
When the ultrasound beam encounters an anatomic boundary, the received sound
impulseappearsasaverticalreflectionofapoint.Onthedisplay,itlookslikespikesof
differentheights(theamplitude).
The intensity of the returning impulse determined the height of the vertical reflection
andthetimeittookfortheimpulsetomaketheroundtripwoulddeterminethespace
betweenverticals.
Thepulseisrepeatedtypicallyabout1000times/s(PRFis1KHz),withtransmitmode
about1sandlisteningmodeabout999s
Usedin:eyeexamination,breastcysts,brainmidlinedisplacement.
It is the signal processing control, which allows amplifying the returning signal from
deeperstructuresinthebody.
TheTGCcontrolcompensatesfortheattenuationofsoundwaves.
Amplificationisproportionaltothetimethathaselapsed,thusthedistancethathas
been travelled by the sound, so all echoes from identical interfaces are rendered the
sameindependentfromtheirdepth.
Theoperatorcanvarythethresholdandtheslopeoftheramp,andtheresultingTGC
curve.
9.7 BMODE:(BrightnessMode)
The ultrasound signal is used to produce various points whose brightness depends on
theamplitudeinsteadofthespikingverticalmovementsintheAmode.
The vertical position of each bright dot is determined by the time delay from pulse
transmission to return of the echo and the horizontal position by the location of the
receivingtransducerelement.
ThereturningsoundpulsesinBmodehavedifferentshadesofdarknessdependingon
theirintensities.Thevaryingshadesofgrayreflectvariationsinthetextureofinternal
organs.
TGCcontrolalsocanbeusedtocompensatefortheattenuationofsoundwaves.
9.8 REALTIMEIMAGING:
To generate a rapid series of individual 2D images that show motion, the ultrasound
beamissweptrepeatedly.
DifferentfromstaticBmodethattheframesneedto
bereproducedmorethan25timespersecond.
Scanners:
(A). Mechanical(sector)scanning:
Mechanicalscannersarecharacterizedbytheuseof
anelectricmotortomovetheultrasonictransducer
and when the transducer sweeps over the area of
interest, the returning echoes are displayed as a
seriesofsinglelineBscan.
There are two basic ways in which the transducer
canbemoved,rotationoroscillation.
Mechanicaltransducersuseacombinationofsingle
elementoscillation,multipleelementrotation,ora
single element and set of acoustic mirrors to
generatethesweepingbeamfor2Dmode.
Sectorscannersaremostusefulforcardiacultrasoundexaminationswherethebeam
isdirectedbetweentheribstoimagetheheart.
Elongated transducer , divided into a large number of separated narrow strips each
aboutawavelengthinwidth
Theyenergizedinoverlappinggroupsinsuccession(say16,27,38..)
Faultysingletransducerelementisnoticeableasadarkstreak,becauseelementsare
notenergizedsimultaneously
(C). Electronicsectorscanner:(steeredorphasedarray)
Similarto lineararray, with fewer elements and they all energized simultaneously at
smalltimedifference(phasing)
Sectralscanningisdonebychangingthetimedelaysequencebetweeneachpulse.
Steering , if they energized separately in rapid sequence, the pulse reinforce in one
directionandthebeamswingtothisdirection,ifenergizedinthereversesequence
thebeamswingtotheotherdirection.
Electronicfocusingisusedtoreinforcethepulsesinbothlinearandphasedarrays.
FocalDepth:
Inlinear&phasedarrays,focusinoneplaneonly(Azimuthatplane)paralleltothe
lengthofthearray.Inannularcirculararray,focusintwoplanes(twodimensions)
In linear & phased arrays, the greater the time delays between energizing the
successivepairsofelements,theshorterthefocallength.
Focusing in the perpendicular plane, done by shaping the transducer into curve or
usingacousticlens.
Multiplezonefocusing:threeorfourpulsesaresentinsuccession,withphasedelays
alteredtofocusindifferentdepths,andthetransducerisgatedtoreceivetheechoes
fromthedesiredfocalzoneonly.
Aspectsofrealtimeimaging:
Pulserepetitionfrequency(PRF):isthenumberofpulsespersecond.
Scanlinedensity:
Eachframemustbemadeupofasufficientlylargenumberofscanlines.
Patients tissues are sampled along a number of scan lines depending on the
numberofelementsinamultielementarray.
IncreasedbyincreasingPRF
FrameRate:(Framerepetitionfrequency)(FRF):
Dependsonthenumberoflinesperframe.
IncreasedbyincreasingPRFFramerate
Depthofview:
IncreasedbydecreasingPRF
NOTpossibletoachievehighframerateandhighscanlinedensitywithlarge
depthofview.
.
Depthofview
Depthofviewxframeratexlinesperframe=0.5xsoundvelocity
PRF=Frameratexlinesperframe=velocityofsoundintissue/2xdepthofview
Improveimagequality;mainlybeincreasingthesoundreflections.
Musthavelowtoxicityandreadilyeliminatedbythebody.
Generally,aremicrobubbles(lessthan4m)ornanoparticles(lessthan1m).
Normal static diffusion leads to total bubble destruction with few hours after the
examination.
Although, microbubbles are destroyed by high intensity ultrasound, this effect can
beusedtostudysubsequentrefilldynamicsallowerpowerforreperfusionimaging.
Types:
o Airfilledmicrospheresencapsulatedinathinshellofalbumin:forcardiac
imagingandDVTdiagnosis.
o Lowsolubility gas encapsulated in a lipid shell : used for all vascular
applicationsincludingtumorvasculature.
o Perfluorocarbonnanoparticles:stayinthebloodfoemanyhours,usedfor
metastasesimaging.
o Goldboundcolloidalmicrotubes:immunologicallytarget.
Harmonicimaging:
TypesofHarmonicImaging:
1. Tissue harmonic imaging (THI) (negative harmonic imaging): does not use
contrastagent.Usestheharmonicgeneratedwithinthebodytissue.
2. Harmonicimagingusingcontrastagent(positiveharmonicimaging):
Injects contrast agents into blood vessels to increase the generation of
harmonics,thentheharmonicsignalgeneratedinthebloodvesselisusedto
obtaintheimage.
A Harmonicbandfiltering:
The fundamental frequency is removed using a filter, leaving the tissue
generatedharmonicstobeprocessed.
The pulse is stretched to produce a narrow transmission band to eliminate
thehigherfrequencies
Decrease the axial resolution, with partial compensation by improved
detectionofsecondharmonic.
B PulseInversion:
Every other transmitted pulse will has its polarity reversed and the echoes
receivedfromeachpairoftransmittedpulsesaresummed.
Then,alltheoddharmonicsdisappearfromthesignalincludingtheinitial=ly
transmittedfrequency.
At the same time, the amplitude of every even harmonic is doubled, which
willenhancestheSNRforthesecondharmonic.
Preserve axial resolution, yet takes longer time with possible motion
artefacts.
BenefitsofTissueHarmonicImaging:
1. Reduced reverberation artefacts (the sound wave is reflected back into the body
from the transducer-skin interface), as low amplitude echoes dont produce
harmonics.
2. Distortion & scattering from fatty tissues are reduced, as these are at the
fundamentalfrequencysosuppressedalso,socontrastresolutionisimproved.
3. Lowcontrastlesionsandliquidfilledcavitiesarebettervisualized,asthereisa
reducedacousticnoise.
ThreeDimensionalImaging:
Threedimensionalimagesareobtainedfromasetoftwodimensionalscans.
The movement of the probe in controlled either electrically so that orientation and
positionarelikenedwithdatacollection,orwithmanuallyoperatormovementswith
orientationandpositionareconstantlyregistered.
Imagesaresetofparallelslicesoraseriesofwedgeshapes
Imagereconstructionandinterpretationarenoteasy
Veryusefulinvascularstudiesandfetalimaging.
If acquisition and display of volume images are in real time, this is called
(Fourdimensionalimaging).
Digital converter is used to improve the gray scale images, enable freeze frame, to
manipulatedataandarchiveit.
{Write}mode:afterdigitizingthesuccessivearrivingechoes,itenteredinthecomputer
memorylocationsalongthecorrespondingscanlinesofthevoxelsofthebody.
{Read}mode:afterpassingthroughdigitaltoanalogueconverter,thesignalisusedto
modulatethebrightnessofpixelsonthemonitor.
Imagescanbestoredintheusualways,inthecomputer memory,orasahardcopy
usingcameraorlaserimager.
DynamicRangingofsignals:
Is the range between the minimum low intensity and maximum high intensity
signalsthatasystemiscapableofdisplaying.
Typically7080dB,andafterTGCtypically4050dB.
A logarithmic amplifier is used to electronically compress the signal amplitudes,
because the grayscale monitor can only display about 25 dB, and the eye can
distinguishonlysome30graylevels.
Edgeenhancementisusedtoincreasethechangeinsignallevel.
Rejectcontrolisusedtoeliminatelowamplitudenoiseandscatter(signalfiltering)
9.10 RESOLUTION:
(A). Axial(longitudinalDepthRange)Resolution:
The ability to distinguish two objects parallel to the beam (objects above and
beloweachother)sothattheycanbeimagedastwodifferentinterfaces.
Axial resolution can be improved by higher frequency transducers, yet less
penetration.
Axialresolutioncanbeimprovedbyreducingthepulselength(duration).
Axial resolution is reduced by omitting the backing block which will increase Q
factor.
(B). Lateral(TransverseAzimuthalAngular)Resolution:
The ability to distinguish two objects perpendicular to the ultrasound beam at
thesamedepth,(objectsthataresidebyside)sothattheycanbeimagedastwo
differentinterfaces.
Lateral resolution can be improved by higher frequency transducers, yet less
penetration.
Lateralresolutionapproximatelyequalstothebeamwidthatthatdepth,socan
beimprovedbyusingsmallerbeamdiameter.
Theshorterthefocallengthoftransducer,thenarrowerandshorterthefocalregion.
In the focal Zone, both axial & lateral resolution are best, axial resolution may be
aboutonewavelengthandBUTlateralresolutionaboutthreetimesworsebecausethe
beamismaybenarrowerthanthegapbetweenobjects.
Speckle:interferencebetweenthewavesscatteredfrommanysmallstructureswithin
tissues,produceatexturedappearance.
Reverberation: strongly reflection interface near the surface, produce a series of
delayedechoesequallyspacedintimeandfalselyappearasdistantstructure.
Doublereflection:likethediaphragmactsasamirrorwithreflectingimagesfromthe
liverintothelung,alsomayhappenwiththepleuraorpacemakerwires.
Acousticshadowing:strongattenuationstructures(bowelgas,lung,bones,gallstones
andkidneystones),castshadowsintheregionbehindthem,MadeworsebyTGC
Acoustic enhancement : fluid filled structures ( cyst or full bladder ) , increase the
intensityofechoesbehindthem,producingnegativeshadow,MadeworsebyTGC
Refraction:likerefractionofabeamfallingobliquelyontwosurfacesofbone,displace
thebeamandtheimagesofthestructuresbehindthem.
Ringdown: caused by small gas bubble or air in the stomach, it resonates emitting
ultrasoundsignalscontinuouslyresultinginatrackthroughoutthescan.
Sidelobes:areformednearthetransducerduetismallvibrationsatthelateraledges
ofthepiezoelectric disc,they areoflowintensitycomparedtotheprimarybeam
butcancauseartifacts.
P
9.12 MMODE(TIMEMOTION):
Ultrasoundisusedforanalyzingmovingbody,commonlyincardiacandfetalcardiac
imaging.
VisualizationofthestructuresasafunctionofdepthandmotionontheYaxisandtime
onXaxis.
A single beam in an ultrasound scan is used to produce the onedimensional Mmode
picture, where movement of a structure such as a heart valve can be depicted in a
wavelikemanner.
9.13 DOPPLER:
DopplerEffectorshift:
the effect of motion of sound sources and the
frequencychangeofthesoundtotheobserver.
Proceeding:wavelengthfrequency.
Receding:wavelengthfrequency.
(Constantsoundvelocity).
The change in frequency (f f)is directly
proportional to the change of velocity of the
interface and the orginal transducer
frequency.
=2x
So,atrightangleofinsonation90,NODOPPLEREFFECT(COS90is0)
When=0(paralleltothesounddirection),MAXIMUMDOPPLEREFFECT(COS0is1)
ContinuousWaveDoppler:
ADopplermode,whichrecordsbloodflowvelocitiesalongthelengthofthebeamby
usingdifferentcrystalstosendandreceivethesignal(twoprobesinthetransducer).
Utilizing one half of the elements continuously sending sound waves of a single
frequencywhiletheotherhalfiscontinuouslyreceivingthereflectedsignals.
HighQisnecessary&Nobackingblockisused(omitted)
Originalf(equalsfofAC),issuppressedandtheDopplersignal(ff)isextracted
electrically.
Frequency analyzer is used to produce a spectrum of Doppler frequencies
correspondingtotherangeofvelocitiesdetected.
Ausefulaudiblesignalcanbeheardbyspeaker,thehigherthepitch,thegreaterthe
velocityandtheharsherthesound,thegreatertheturbulence.
MostimportantinDopplerechocardiography.
Advantagesofacontinuouswavetransducerare:
Highsensitivity
NoNyquistlimit(twicesamplingeveryperiod)
Noaliasing
Disadvantagesofacontinuouswavetransducerare
Nopossibletolocatemovingreflector
Nodepthprecision(distinguishbetweenflowintwooverlyingvessels)
PulsedDoppler(rangegating):(depthprecision)
A Doppler mode, which evaluates blood flow velocities in a range specific area
measuringthechangesinthereceivedfrequencyduetorelativemotion(flow):
UsingDuplexscanningofthevessels(Bmode+Doppler),Bmodehelpinchoosing
alineofsitefortheDopplerbeamandessentialfordirectionassessment
Asamplingvolumeisplacedoverthevesseltomeasuretheflow.
The normal short imaging pulses are interspersed with bursts of the Doppler
ultrasound, and a ranging gate is set to accept only those Doppler echoes which
cameonlyfromtheselectedsamplingvolume.
Thedepthofthetissuedependsonthetimewhichtheranginggateisopened,and
thethicknessdependsonthehowlongtheranginggatewasopened.
ThewidthofthesamplingvolumeisthewidthoftheDopplerbeam.
Theangleofinsonationisclearlydefinedbythesystemtoallowaccuratereadof
thefrequencyshifttobeconvertedintoflowvelocities.
Thediameterofthevesselisalsoestimatedtoallowvolumeflowratecalculation.
ItisagraphofDopplerfrequencyagainsttime.
Displaysthevariationsofbloodflowvelocityand
directionduringheartcycle.
Each pixel has a grey level representing the
numberofredbloodcellsinthesamplingvolume
thathavethevelocityatthatinstantoftime.
Pressuremeasurement:
Pressureinstenosedvesselorinthejetofbloodcomingfromdiseasedheartvalve
canbeestimatedbyDopplermeasurementofflow.
The velocity of the flow increased within any constriction , and the increase in
2
pressurecanbeestimatedbymodifiedBernoulliformulaP=4v
Aliasing:
With pulsed Doppler only , very high flow velocities will shown in the wrong
directionandunderestimated,likeawarproundtop&bottominthesonogram.
It is a consequence of the sampling requirement, that the wave form being
measuredmustbesampledtwiceeveryperiod,sothefastestflowcanbemeasured
withaccuracymustbePRF,otherwisealiasingoccursduetoundersampling.
The depth of the sampling volume gate, determines the PRF needed, and the PRF
determinesthemaximumvelocitycanbemeasuredwithoutaliasing.
Toreducealiasing:
Useoflowerfrequencyprobe(f).
Increasetheinsonationangle().
IncreasethePRF(PRF)
HighPulseRepetitionfrequencymode:(HighPRF)
DoublingorquadruplingPRF,willfurtherimprovethemeasurementofthemaximum
velocity,intheexpanseofsuperimposingvelocitydatafrommultiplegatedvolumes.
When PRF becomes very high, the pulses merge and we have reinvented continuous
waveDopplerwithnoaliasingbutnorangedata.
Color flow imaging is based on pulsed ultrasound Doppler technology. With this
technique multiple sample volumes among multiple planes are detected and a color
mapfordirectionandvelocityflowdataisdisplayed.
CommonmappingformatsareBART(BlueAway,RedTowards).Enhancedorvariance
flow maps show saturations and intensities that indicate higher velocities and
turbulenceoracceleration.Somemapsutilizeathirdcolor(greenoryellow)toindicate
acceleratingvelocitiesandturbulence.
ColorflowDopplerimagingisnotaspreciseasconventionalDopplerandisbestused
toscanalargerareaandthenuseotherDopplermodestoobtainmoreprecisedata.
Compared with normal Bmode imaging, the color Doppler pulse is longer, being a
compromisebetweenaccuratedepthinformationrequiresashortpulseandaccurate
velocityinformationrequireslongerpulse.
Aliasing is also a feature of colour scans, with exchange of red & blue direction and
mosaiccoloredappearance
Colourfactorsthatcanbevaried:
Framerate:shouldbefastenoughtofollowvelocitychanges.
Penetrationdepth:whichisinverselyproportionaltoPRF.
Fieldwidth:orsectorsize
Linedensity:scanlinesperframe,shouldbehighenoughforgoodspatialresolution.
Numberofpulsesinatrain:shouldbehigh,togiveaccuratevelocityinformation.
PowerDoppler:
PowerDopplerisshowingtheamplitudeoftheDopplersignalswithoutanyindication
ofvelocityordirection.
Itsmainadvantageistodifferentiatebetweenareasofflowornoflow.
Lessdependedontheinsonationangle.
MoresensitivetoweaksignalsthancolourDoppler.
Noaliasing.
9.14 QUALITYASSURANCE:
Usefulperiodictests,includethefollowing:
1) Resolution: tested by imaging a test rig composed of parallel wires mounted on a
frameandimmersedinaPerspexbathcontainsfluid.
2) Sensitivity & Dynamic range and accuracy of the A scan caliper: Tested using a
Perspexblockmachinedwithanumberofequallyspacedverticalrods.
3) GrayscaleperformanceandDopplerfunction:
o Testedwithcomplexequipment,basedongelatinortissuemimickingphantoms.
o Themaximumacceptabledistancemeasurementerrorshouldnotexceed2%.
4) Thepoweroutput:testedbyweighingthesoundpressurewithaforcebalanceorby
measuringtheheatingeffectusingacalorimeter.
Ultrasoundislowriskaswellaslowcostmethod.
Harmfulbioeffectshavebeenidentifiedathighexposurelevelsusuallywithultrasound
therapy.
Safetyguidelines:
Thetimeaveragedintensityshouldnotexceed100mW/cm2
Totalsoundenergy(intensityxdwelltime),shouldnotexceed50J/cm2.
Possiblerisks:
Localheating:duetofriction,viscousandmolecularrelaxationprocessesleadingto
chemicaldamage.
Acousticstreaming:ofcellularcontentsinthedirectionofthebeam,affectingcell
membranepermeability.
Cavitation: the high peak pressure changes cause microbubbles in the liquid ( or
near liquid ) medium to expand ,there might be a much rise in temperature with
consequentprofoundchemicaldamagetocells
Mechanicaldamage:tocellmembranescausedbyviolentaccelerationofparticles.
Thermalindex(TI):
Istheratioofthepoweremittedtothatrequiredtoincreasethetemperatureby
1Candgivesanindicationofthetemperatureriseintissues.
TI=
Indexvalueshouldntexceedvalue1,andbelow0.5isconsideredsafe.
TheexposuretimeshouldbereducedappropriatelyastheTIincreaseabovethe
value1.(e.g.about30minutesforfetalscanning).
Inapyrexialadult,atTIvalue2,examinationtimecanbesafefor10minutes.
Themechanicalindex(MI):
Isameasureofthemaximumamplitudeofthepressurepulse.
Calculated by, the peak rarefaction pressure divided by the square root of the
ultrasoundfrequency.
MI=
f
Potentialriskofcavitationwithuseofcontrastmediumatmorethan0.7value
MIshouldntexceed0.3valuefornondiagnosticandfetalscans,duetopossible
riskofminordamagetoneonatallungsorintestine
Chapter10
Magneticresonanceimaging
Dr. Khaled Elazhry 90
Nuclearmagneticresponse,ispropertyatomscontainsanoddnumber
ofprotons
HydrogenprovidesbestMRIsignals:
Hassingleproton,thuslargemagneticmoment
Abundantinthebody,inwaterandfat.
10.1 THESPINNINGPROTON:
Thestaticfieldalsocausesthespinningprotonstowobblein
theregularmannercalledprecession.
Thedirectionofthespinaxistiltsandspinrotatesaroundthe
directionofthemagneticfieldBwithfixedfrequency.
ThefrequencyofprecessioncalledLarmorfrequency.
The tilting of the spin axis splits its magnetic vector m into a
longitudinalcomponentmzthatpointintheZdirection,anda
transversecomponentmxythatrotatesintheXYplane.
All the mz vectors of all detectable protons, add up to a
combinednetlongitudinalmagnetismMz,butthismagnetism
cant be measured directly because it points in the same
directionasfieldB.
Themxyvectorsofthedetectableprotonsrotatesandprocess
independently and cancel each other with net transverse
magnetismMxy=0
Thefrequencyofprecession(Larmorfrequency)isproportionaltotheproductofthe
magneticfieldstrengthandthegyromagneticratioofthenucleus(constant).
for hydrogen nuclei in a field of 1T, the f =42.6 MHz , and according to quantum theory , this
frequency equals a quantum energy of 0.2 eV , and this is the energy difference between spin up
and spin down diploes.
Radiofrequencycoils:
Are set of coils surrounding the patient , that inject an RF pulse in a direction
perpendiculartoB,whichresultintwoeffects:
o Some or all spin up protons pick up energy and excited and turn spin down and this
affecttheMzvectorthatitmaybereducedordisappearorreversed.
o Theprotonsarepulledinsynchronism,andtheyprecessinphasewiththeirmxyvectors
adduptotransversemagneticfieldMxywhichrotatesintheXYplanewiththeLarmor
frequency.
The frequency of the RF generator, must be very accurately match the Larmor
frequency ( the photon energy of the radiowaves must be exactly the same as the
energydifferencebetweenspinupandspindownprotons0.2eV)
TypesofRFpulses:
180pulse:aRFpulseofcertaintotalenergywhichwillgiveeachandeverydipole
exactlytheenergy0.2eVrequiredtotipthemthrough180
90pulse: apulseofhalfofthatcertaintotalenergywilltiphalfofthedipolesso
equalnumbers pointspin upand spin down, and make them move into thesame
phaseandprecesstogether(phasecoherence)
Conclusion:180pulsetemporaryreverseMzvector
90 pulse reduce Mz to zero & instead produce Mxy with equal
magnetismtoMzbutperpendiculartoBconvertimmeasurableMzto
measurableMxy
StepsforproducingaMRIimage:
1) After a 90 pulse, The magnetic vector Mxy continue rotating for while in the
transverse plane XY, inducing in the RF coil a alternating RF voltage of few
microvolts.
2) RFamplifierisusedtoamplifythissignal.
3) Theamplitudeofthissignalissampled,digitizedandcomputeranalyzed.
4) Spatialencodingandsignalprocessingforthesignalfromeachindividualvoxel
inthescanmatrix,willproducethepixelgreyorcolourlevelintheMRimage.
Thesignalpeakandtheproportionalpixelbrightnessdependson:
Protonorspindensity(PD):thenumberofprotonspercubicmillimeteronthe
voxel.
Thegyromagneticratioofthenucleus.
ThestaticfieldstrengthB:becauseincreasedBstrengthwillincreasemorethe
initialdifferenceofspinupandspindownprotons.
Onlymobileprotonsgivesignal,thoseinlargemoleculesoreffectivelyimmobilizedin
bone,donotgivesignal.
Thegreaterpartfromsignalisduetobodywater(freeorbounded)
Airhasnohydrogen,willgivenosignalandappearblackintheimages.
FAThasgreaterPDthanothertissueswhichvaryintheirprotondensities.
FreeInductionDecay:((FID))
Spinlatticerelaxation:(T1recovery)
The excited protons give up their energy to the molecular lattice. One by one, the
dipolestipbackparalleltotheZaxis,andMzslowlyreappears(longitudinalrelaxation)
In the graph, curve a represent the Mz recover slowly and exponentially with time
constantT1,whichdependsonthetypeoftissue.
FactorsaffectingT1recovery:
Jostling ( pushing ) by large molecules , which are slow moving and near the
resonantfrequencyismosteffectiveatremovingenergyfromexciteddipoles.
e.g. fat and water bounded to proteins surface are large molecules , can absorb
energy easy and so have a relatively short T1.
Jostlingbysmall,lightweightmolecules,whicharerapidandrelativelyineffective
atremovingenergyfromexciteddipoles.
e.g. free water, urine, amniotic fluid, CSF and other solutions have a long T1, the
greater proportion of free water in tissues, the longer is T1
Theatomsinsolidsandrigidmacromolecules,arerelativelyfixedandtheyleast
effectiveatremovingenergy
e.g. Compact bone, teeth, calculi and metallic clips have a very long T1
Spinspinrelaxation:(T2decay)
Lossofphasecoherenceduetoenergytransferbetweennuclei,resultinaexponential
decayofthetransversemagneticvectorMxy
Asthedipolesareprogressivelydephasing,somerotatesfasterorslowerthanothers
withdecreasingnetstrengthofmagneticvectorMxy
In the graph, curve b represent the Mxy decays exponentially & rapidly with time
constantT2,andsodecreasingtheinducedMRsignal(curvec)
MostofthedephasingeffectisduetofieldBinhomogeneitiesfrommachineexternal
factors,yetanimportantfactoristissuerelatedcalledT2decay(transverserelaxation)
T2,isthetimeforMRsignaltofallto37%ofitsmaximumvalue.
After1timeT2,signalis37%
After2timesT2,signalis14%
After3timesT2,signalis5%
Theatomsinsolidsandrigidmacromolecules,thelocalvariationofmagneticfield
isgreatestandtheydephasequickly.TheyhaveaveryshortT2anddonotproduce
lastingsignal
Theatomsinlightermolecules,arerapidinthermalmotionwhichsmoothesout
thelocalfieldresultinginalongT2
TissueCharacteristics:
T2isalwaysshorterthanT1
T1isaffectmorethanT2byincreasestrengthofmagneticfield.
AbnormaltissuestendstohaveahigherPD,T1andT2thannormaltissuebecauseof
increasedwatercontentorincreasedvascularity
Typical relaxation times of tissues in a field 1T
Material T1 (ms) T2 (ms)
fat 250 80
liver 400 40
kidney 550 60
spleen 400 60
White matter 650 90
Grey matter 800 100
CSF 2000 150
water 3000 3000
Bone, teeth Very long Very short
10.3 SPINECHOSEQUENCE:
Freeinductiondecay(FID)signalisrarelymeasuredbecauseitdecaysveryrapidlywith
timeconstantT2*(T2star),whichismuchshorterthanT2becauseofStaticfieldB
inhomogeneitiesdueto:
Themagneticfieldproducedbygradientcoilsacrossthevoxel.
Unavoidableimperfectionsintheengineeringofthemagnet.
Theintroductionofthepatientwhichunavoidablydistortsthestaticfield.
3) Aftertime(t),the180pulseisappliedandtipsallthedipolesfromspinuptospin
down , this turns the individual mxy vectors through 180 in the Xdirection , so
laggersbecomeleadersandviceversa.
Mxy and the signal are still the same as mxy vectors continue to rotate in the XY
plane.Theynowrephase.ThefasteronescatchupwiththesloweronesandMxyand
MRsignalregrow.
4) After afurthertime (t),they areagaininphaseand MxyandtheMR signalare at
theirpeak.
Thereafter,theygrowoutofphaseagainandMxyandthesignaldecay.
180 pulse is called the rephasing or refocusing pulse, because is reverse and
eliminatesthedephasingeffectofthesystemicmagneticfieldinhomogeneities.
TheMRsignalreappearsasanechooftheinitialsignal,andcomposedessentiallyfrom
twosignalsbacktoback.
ThesignalismeasuredattimeTE=2t,itwillbereducedinamplitudebyT2decay.
(ThelongerTE,thesmallertheMRsignals.)
MRsignaliscontrolledbytwoparametersTR&TE:
TR(timetorepeat):thetimetorepeatthesecondpulse.
TE(timetoecho):equals2t.Thetimefortheechotoform.
ThestrengthoftheMRsignalfromeachvoxelandthebrightnessofthecorresponding
pixeldependson:
T1 contrast
1) How many protons there are in the
pixel (PD): the greater PD the larger
thesignal.
2) How far Mz has recovered from the
previous90pulsewhenthenext90
pulse tips it : the length of T1
comparedwithTR
Shorter T1 or longer TR, the larger
MRsignalandbrighterthepixeland
betterSNR.
3) How far Mxy has decayed when the T2 contrast
echo is formed : the length of T2
comparedtoTE
ThelongerT2orshorterTE,thelarger
MR signal and brighter the pixel and
betterSNR.
FairlyshortTR=maximumT1contrast
FairlylongTE=maximumT2contrast
FairlylongTR=maximumPDcontrast
WeightedImages:
The TE and TR are chosen so that pixel brightness depends on one of three
combinationsofPD,T1andT2
T1-weighted image:
ShortTR(300800ms)isusedtoproducemaximumT1contrast.
ShortTE(15ms)isusedtoreducetheeffectofT2,shortertimewontgivethe
systemtimetoapplythe180pulseandthegradientpulses.
TheresultedimageisprincipallyduetotheT1recoverypropertiesofthetissues,
withthetissueshasshorterT1morebright(fatisbright,waterandCSFaredark)
T2-weighted image:
LongTR(10002000ms)isusedtoreducetheeffectofT1.
LongTE(90140ms)isusedtoproducemaximumT2contrast..
The resulted image is principally due to the T2 decay properties of the tissues ,
withthetissueshaslongerT2morebright(waterandCSFmorebrightthanfat)
Air and cortical bone appear black in all images, because they have no hydrogen.
White & grey matter, grey matter is brighter and white matter in PD and T2 weighted
images. in T1 weight images, white matter is brighter but its shorter T1 is
counteracted by its lower PD.
Generally, tissues with long T1 also have long T2 and those with short T1 have a
short T2, this why images cannot be weighted for both T1 & T2.
Injudicious selection of TR & TE (not carefully selected), will lead to that tissues
with different relaxations times can produce equal signals showing no contrast and
indistinguishable.
10.4 SPATIALENCODING:
Includethreebasicprocesses:
Slicesection
Phaseencoding
Frequencyencoding
Sliceselection:
PhaseEncoding:
FOV=PixelsizexNumberofphaseencodingsteps
or
FOV=voxelsizexNumberofcomponentsintoofthesampledfrequency
Because phase angels repeat themselves every 360, there will be a several phase
cyclesacrossthewholeFOVsay,the10isrepeatedat370,730andsoon.
ImageTime:
Itisthetimeneededtoacquireanimage,itequals;
NexxphaseencodingstepsxTR
WhereNexisthenumberofsignalaveragesorexcitations
IncreasingtheNexwillreducethenoise,butwillincreaseimagetime.
Kspace:
Kspace is a temporary memory of the spatial frequency
informationintwoorthreedimensionsofanobject,storedas
greyscalevalues.
Kspace does not correspond to the image, but has axes
correspondingtofrequencyandphaseencodingdata.
Each data point in the Kspace matrix is a spatial frequency
componentofthesignal.
The central part of the Kspace contains data from shallow encoding gradients and
lowspatialfrequencies(lessdetailsbutstrongersignals),whiletheupperandlower
partsarefilledwithdatafromsteepergradientsandhighspatialfrequencies(better
detailsbutlowsignalsintensity).
Kspacehastobecompletelyfilledwithdatabeforethesignalscanbeprocessed.
o Fourieranalysisisdonebytwomethods:
Using the analysis along the phaseencoding axis (horizontal), and measure
thefrequencyinthisaxis(TRhereiscalledpseudotime).
Usingtheanalysisalongthefrequencyencodingaxis(vertical),andmeasure
thephasesignalinthisaxis(TRhereiscalledpseudofrequency).
o Onceallthesignalsanalysed,acompleximageisproducedwhichmayhaverealand
imaginaryparts.
o If any tissues move during the repetitions, they will be misregistered in the wrong
pixel of the phase encoding axis of KSpace, while any chemical artefact will be
reveledinthefrequencyencodingaxis).
SummaryofGradients:
- One gradient defines the slice
selection, appliedwhentheRF
pulseisturnedon.
- Secondorthogonalgradientis
used for phase encoding,
appliedbrieflybetweenthe90
and180pulses.
- Third orthogonal gradient is
used for frequency encoding
applied with echo formation,
during which the signal is
measured.
- GradientareusedtocontrolslicethicknessandtheFOV.
o Thesteeperthesliceselectiongradient,thethinnertheslice.
o Thesteeperthephaseandfrequencyencodinggradients,thesmallertheFOV
- Increasing the matrix size, makes no difference to scan time in the frequency
encodingdirectionasitisdonebysamplingasingleecho.Whileitwillincreasescan
timeinthephaseencodingdirectionasthespatialinformationinnotcompleteduntil
allthegradientsstepshavebeencompleted.
Multislicetechniques:
o MostoftheTRwastedifwehavetowait2sbeforerepeatingpulsesonagivenslice.
o This time can be used to deliver a succession of 90 and 180 pulses, each of a
differentfrequency,excitingaseriesofupto32separateslicesbeforerepeatingthe
firstslice.
o ShorterTEcomparedtoTR,willgivemoreslicescouldbeinterleavedinthisway(if
TR1000msandTE60ms,1000/60=16slicescouldinterleaved)
Multiechotechniques:
o Long TR can be used, after the 90
pulse a series of rapidly applied 180
rephasing pulses and multiple echoes
withincreasingTE.
o Theirpeakamplitudesdeceasewiththe
timeconstantT2
o The first echo may be produce a PD
weighted image, while following
successiveechoesproduceT2weighted
images.
o Dualechosequencecanbewrittenas90,180,180andproducetwoimagesperslice.
The first echo has short TE and long TR (PDweighted), second echo has longer TE
(T2weighted).
Fast(turbo)Spinecho:FSEorTSE
o TheMultiechosequence90,180,180,canbemodifiedbyphaseencodingeachof
the 4 16 echoes ( each echoes spaced by 20 ms ), with different phase encoding
gradient,withreducedfactorof416time(numberofTRintervals)
o FATisextremelyhigh(bright)signalinfastSET2weightedimages,becausetherapid
successionof180pulsesreducesthespinspininteractions.
o Muscle, is often darker than single SE sequence, due to increase the transfer of
magnetismwithresultinsaturation.
o Matrixsizecanbeincreased,toimprovespatialresolutionwhichwasaffectedbythe
reducedscantime.
o FSE is incompatible with respiratory compensation mechanism unless a powerful
gradientusedtoproduceimagefromsinglebreathhold.
InversionRecovery:(180,90,180)
o IsusedtoaccentuateT1weightedimages.
1) Initial180pulseisusedtotipthespinsantiparalleltotheZaxis(reverseMz).
2) Mzrecovers,passingthroughzeroandreversingdirectionaftertime(069xT1).
3) After a variable time (TI, time to inversion), a 90 pulse is applied, tilting the
availableMz
FatSuppressionshortTIinversionrecovery(STIR)(180,90)
Tipangle:
o OnewaytoreducescantimeistoreduceTRto200msoreven20ms,butthiswont
giveMztimetorecover.
o This is fixed by using RF of small strength (result in smaller tip angle or flip angle),
whichinvertsonlyasmallfractionofdipoles.
o 30 pulse, produce only half (50%) of the usual Mxy and the consequence MR signal,
butleavesabout87%oftheMz,whichrecoversquicklywithgoodsignalfollowingthe
nextRFpulse.
o 15pulse,produce(25%)oftheusualMxyandtheconsequenceMRsignal,butleaves
about97%oftheMz.
o ThesmallerinitialRF,thesmallerthetipangle,thesmallertheT1weighting.
Echoplanarimaging(EPI)(fastGRE)
o AnultrafastformofGRE,withsnapshotstakeninabout50mseach.
1) Following stranded SE (90, 180) sequence and the slice selection gradient, the
polarity of the frequency encoding gradient is continually reversed each time
inducingagradientecho.
2) The phase encoding gradient is also switched on and off briefly just before each
echo. Thus, encoding each of the echoes is done with a different phase encoding
gradient.
3) MultipleechoescanbecollectedbeforeMxyhasdecayedtoofar,togiveacomplete
imageinthe50msfollowingtheSEsequencepulses.
o PDandT2weightedimages,areobtainedbyusingshortorlongeffectiveTE
o T1 weighted image is possible if an inverting pulse is applied before the excitation
pulsetoproducesaturation.
o Superconductingmagnetisnecessaryandveryhighgradientsandveryfastswitchers.
o EPI, can be used for functional imaging, real time cardiac imaging and perfusion or
diffusionimaging.
Imaginginotherplanes
o Anydesiredimageplanecouldbeselectedwithoutmovingthepatient.
o Coronal plane, Ygradient used for slice selection & the other two for phase and
frequencyencoding.Insagittalplane,Xgradientisusedforsliceselection.
o Generally,phaseencodinggradientbestusedalongshorterdimensionofthepatient.
10.6 SPECIALIZEDIMAGINGTECHNIQUES:
Magneticresonanceangiography:
TheeffectontheMRimageoftheflowofblooddependsonmanyfactorsincluding;
Velocity&Flowprofile.
Directionrelativetotheslice(greatestifperpendicular)
Thepulsesequenceanditsparameters.
o RegardingbloodflowwithSEsequence:
Vesselscontainingslowflowingblood(e.g.veins),mayappearbright,because
bloodentersthesliceunexcitedduringthe90pulse,makinghimmoreaffected
and producestronger echo than other tissues in which their Mz has not yet fully
recoveredfromtheprevious90pulse.
Vessels containing fastflowing blood ( e.g. aorta), may appear dark or void ,
becausesomeofthebloodexcitedbythe90pulsealreadylefttheslicebeforethe
180,andsoproducenosignal.AndinconventionalT1imagesthebloodappear
asblackflow,soanybrightsignalsmayindicatestagnantfloworocclusion.
Turbulentflowproducesarapidlossofcoherence,thusreducingMxyandusually
appearsblack,asinturbulencedownstreamofstrictureinavessel.
PerfusionImaging:
o It is measurement of the rate at which blood is delivered to the capillary bed, thus
measuringmetabolicactivity.
o DoneusuallywithEPItechnique,usingaparamagneticcontrastagent,ineitherway:
BolusofcontrastagentArterialspinlabelling.
o Gadolinium as in ion Gd3+ , is highly suitable as contrast agent having 7 unpaired
electrons, but being very toxic is must be chelated with diethyl lenetriaminepenta
aceticacid(DTPA),butstillcontraindicatedinpatientwithrenaldysfunction.
o GdDTPA itself not visible in the MR images, but tumbling around the Larmor
frequency,itwillshortenbothT1&T2ofthehydrogennucleiintheirvicinity.butthe
effect on T1 is greater than T2, so called positive contrast agent and used mainly
withT1weightedimages
o T2*isalsosensitivetothechangesofthesusceptibilityeffectsofGdDTAPinperfused
vessels,butfastimagingtechniquesneededasthesechangeslastfewsecondsonly.
o Measurementofperfusion,intermsofmillilitersofbloodper100gtissueperminute.
o GdDTPA,iswatersolublesoitmayproduceincreasedcontrastbetweenpathological
andnormaltissue,andisdoesntcrossthenormalbloodbrainbarrier.
o Fatsuppressiontechniquesmaybeneededasfatmayappearsequallybrightaswater.
o GdDTPAconcentrationmustnotbesogreat,whichwillshortentherelaxationtimes
makingtheT2effect,cancelsouttheT1effect.
o Inarterialspinlabelling,perfusionimageobtainedbysubtractingthetwoimagesare
generallyobtained,
Oneinwhichthebloodwatermagnetizationisdifferent,obtainedbyspininversion
spinlabelledimage.
Secondisthecontrolimageinwhichtheyareinthesamestatewithnoinversion.
AnoffresonanceRFpulseisusuallyappliedbeforeacquisitionofthecontrolimage
to cancel the effect of magnetic transfer between free and bounded water which
willdecreasetheagentcontrast.
FunctionalImaging:(FunctionalMRI)
o Itdepends oncomparingofnormal brain rest images and activity(stimulus) images,
andthechangesbetweenthemareduetobloodoxygenationlevel.
o Oxyhaemoglobin is diamagnetic (weak compared to the main field), while
deoxyhaemoglobin is paramagnetic and produce magnetic filed inhomogeneities in
theneighboringtissuesandwillincreaseT2*
o Atrest,tissueshavealmostequalamountsofoxyanddeoxyhaemoglobin.
o Atactivity,therewillbeincreasedbloodflowwithincreaseddeoxyhaemoglobin,thus
increasedMRsignal.
o Veryrapidsequencesrequired(EPIorfastGRE),astheeffectlastforveryshorttime,
producing lowresolution volume images with the areas of subtracted images that
showincreasedsignalintensitycorrespondstoactivatedbrainareasbythisstimulus.
o The subtracted images can be overlaid on normal highresolution images to provide
thefunctionalmap.
Magneticresonancespectroscopicimaging:
o Magnetic resonance spectroscopy provides a frequency spectrum fingerprint of the
tissuesbasedonitsmolecularandchemicalcomposition.
o Mostclinicalspectroscopystudieshydrogen,butotherMRclinicalnucleiarepossible.
o Peakintensitiesandposition;indicatehowanatomisbondedtoamolecule.
o OnlynucleiwithoddnumbersofprotonsorneutronscangiveMRsignal.Anddifferent
nucleihavedifferentgyromagneticratios.
o Abundantelementsinthehumanbody:
1Hhasoneprotonandhighgyromagneticration.
16O,14Nand12Careabundantbuthasnonuclearmagnetism(evennumbers)
13Chasoddnumberofneutronsbutaccountsonlyfor1%ofcarbonatomsinbody.
o Becauseofchemicalshift,phosphorus
nuclei have different resonant
frequencies when bound in different
compounds (inorganic phosphorus,
adenosine triphosphate ATP, phospho
monoester, phospho diester and
phosphocreatine)
o Using a broadband RF, all these
compounds can be made to resonate
and theMR signalfromdefined tissue
can be analysed as a frequency
spectrum.
o Ahighmagneticfieldisneeded(2Tormore)andthefieldmustbeuniformtobetter
than1ppm,togivesufficientsignalstrengthandgoodspectralresolution
o Only phaseencoding gradients can be used in imaging, because spectroscopy
dependsonfrequency.
o Toreduceimagingtime,amatrixof1cmmustbeused.
Dixonmethodforchemicalshiftimaging
o The resonant frequency of protons is affected by its chemical environment and
measurementofthischemicalshiftcangiveinformationaboutmolecularstructure.
o The valence electrons in HO bond in water produce a slightly smaller magnetic field
thandothoseinHObondinlipids,andasresultthefrequencyoftheprotonisabout
3ppmgreaterinfatthaninwater.Andthisdifferenceandbeusedtoproduceseparate
imagesofwaterandfat.
o WaterplusfatimagesobtainedbysettingTEwhentheyareexactlyinphase.
WaterminusfatimagesobtainedbydelayingTE,tomakethemexactlyoutofphase.
o Subtractingtheseimageswillgivefatonlyimage.
Addingtheseimageswillgiveaugmentedwateronlyimage.
10.7 MAGNETICRESONANCEIMAGEQUALITY
Signaltonoiseratio(SNR):
o ItisrandomvariationintheMRsignaloccurringatallfrequenciesallthetime,dueto:
1) Patient : random thermal movements of hydrogen atoms in the tissues, induces
currentsofwiderangeofRFinthereceivercoilscalled"whitenoise"
2) Scanner:electronicnoisefromelectronsincurrentsflowingintheelectroniccircuits.
3) Environment:comefromRFinterferencefromoutsideorinsidetheroom.
(B). Reducingthenoise,by:
IncreasingNex(thenumberofexcitations).
Reducing the bandwidth of the receiver, so it can picks up less of the
spectrumofnoisefrequencies,althoughthisincreasesthechemicalshift
andmotionartefacts
Reducingcrosstalk,byhavinglargergapsorusingmultislicetechniques.
Reducingthevolumeoftissues,bygoodwellpositionedsurfacecoils.
Contrast:
o ItisthedifferenceintheSNRbetweenadjacenttissues.
o Contrastcanbeenhancedby:
1) Magnetization transfer contrast, using offresonant frequency RF pulses to
transfermagnetizationtofreeprotonstosuppressthesignalfromprotonsbound
tomacromolecules.
2) Fatsuppression,byusingSTIRsequence.
3) T2 weighting, specifically increase contrast between normal and abnormal
tissueswhicharebrighterasitcontainsmorewater.
4) Paramagneticcontrastmedium:
GdDTPA, shorten T1 of nearby hydrogen nuclei, thus enhancing the
inherentcontrast.
Manganese (as in ion Mn3+) and iron (as ion Fe2+) , can be also used as
positivecontrastagents.
5) Superparamagneticcontrastmedia:
AppeardarkinMRI"negativecontrastagentsorbulksusceptibility"
Asminuteparticles(30nm)oftheironoxideFe3O4withaninertcoating,
anddysprosium(DyDTPA).
TheyproducelocalmagneticfieldgradientslargeenoughtoshortenT2*
andT2,withareasofuptakeappearblack
6) Hyperpolarizedgas,newcontrastagentofhyperpolarized(bylaser)Xenon(129Xe)
that dissolves in blood and shows a large chemical shift, used with low fields
givinggoodSNR,suitableforlungMRIandlowfieldangiography.
Scantime:
itequals;NexxphaseencodingstepsxTR
improvedby:
ReducingNex:ButitwillreduceSNRandincreasemotionartefacts.
Reducingphasematrix:Butitwillreduceresolution
ReducingTR:ButitwillreduceSNR,numberofslices&increaseT1weighting.
10.8 ARTEFACTS
Aliasing:
Partoftheimageisshiftedbodilytotheoppositesidedirection.
ItisasignoftoosmallFOV.
Imagewarproundinthephaseencodingdirectionbecauseelectroniccircuits
suppressaliasinginthefrequencyencodingdirection
ReducedBy:
Doubling the FOV in the phaseencoding direction and doubling the
phaseencoding steps to keep the same pixel size and resolution and
HalvingNextokeepimagingtimeandSNR.
Using a surface coil more closely matches the FOV or increase FOV to
matchthecoil.
Motionartefacts:phasemismapping
Appearsasghostimages
Apparentonlyinthephaseencodingdirection,evenifthemotionthroughthe
sliceorinthefrequencyencodingdirection.
Cyclicalmotion,asinheartmotionorbreathing.Alsocausemotionartefacts.
But cardiac triggering is possible by making TR equal Rwave to Rwave
intervalECG,andrespiratorytriggeringisalsopossiblebutmoredifficultand
timeconsuming.
Pulsatileflowaspulsationsofabdominalaorta,alsocausemotionartefacts,
but using techniques as motion artefact suppression removes these
artefactsbymodifyingthefieldgradientsgradientmomentrephasing
ReducedBy:
Immobilizationofsedationofthepatient
Exchangephaseandfrequencyencodingdirections.
Chemicalshiftartefacts:
Duringfrequencyencoding,fatprotonsprecessslowerthanwaterprotonsin
the same slice because of their magnetic shielding. Through the difference in
resonancefrequencybetweenwaterandfat,protonsatthesamelocationare
misregistrated(dislocated)bytheFouriertransformation.
This chemical shift misregistration cause accentuation of any fatwater
interfacesalongthefrequencyencodingdirection.
Thisartifactcanbeseenasabrightordarkbandattheedgeoftheanatomy
as in kidney and perinephric fat, around optic nerve & at the margins of the
vertebrabodies.
ReducedBy:
Usingasteepergradient.
Widerreceiverbandwidth.
Superimpositionoftwochemicalshits
CentrallineZipperartefact:
When RF leaks from the transmitter to the receiver or RF interference from
outsideiftheshieldeddoorispartiallyopen
Producedacrossmiddleoftheimage,inthephaseencodingdirection.
ReducedBy:
LargerFOV
Propershielding
Truncationringingartefact:
Itisaparallelstriationsappearathighcontrastinterfaces(fat&muscleorCSF
&spinalcord)
Morelikelyinthephaseencodingdirection.Butalsoinfrequencydirection.
ReducedBy:
Increasephaseencodingsteps(Increasethematrix)
ReduceFOV
Typesofmainmagnet:
1) Apermanentmagnet:
Composedoftwoopposingflatfacedhighlymagnetizedpole pieces(ironwith
alloysofaluminumornickelorcobalt)fixedtoironframe.
Verylargeandheavy.
Expensive,butcheapertorun.
Nopowerrequired,butcantbeshutdown.
Lowstrength,verticalfields,onlyupto0.3Tesla.
Suitable for claustrophobic patients, children, obese adults & interventional
procedures.
2) Aresistiveelectromagnet:
ItisasetofDCcoilswithcopperoraluminumconductors.
Consumeabout55100kW
Heatproducedremovedbycoolingwater
Heatproducedlimitstheverticalorhorizontalmagneticfieldto0.5Tesla
Ithassignificantfringefields
Itcanbeswitchedonandoff,buttakesabout1530mintorampup.
Itisthecheapestandthesmallest.
3) Asuperconductingelectromagnet:
ItisaDCsolenoid,about1mindiameter,withconductorsmadeofaniobium
titaniumalloyinacoppermatrix.
Itissupercooledbyacryogenliquidhilumat4K(269C).
Atthistemperature,largeDCcurrentscanbeusedwithnegligibleresistanceto
producehorizontalfieldupto3Tesla.
Ithassignificantfringefields
Largeandveryexpensiveandheavy.
Ittakesseveralhoursforthecoiltocooldownandthecurrenttobuildup.
Theexpensiveliquidhilumiscontainedwithinafragilecryostatandreplenished
periodically.Witharefrigeratorsystemusedtoreduceheliumlosses.
Ifthemachineisshutdown,theelectromagneticenergy(about20kWh)stored
withinthesuperconductingcoilhastoberemovedcarefullytoavoidhilumleak.
If the temperature rises, the liquid gas boils off rapidly and must be vented
outsidethebuilding.
Themagneticfield:
Themainfieldmustbestableanduniformto5ppm(1ppmforspectroscopy).
Fringe fields (magnetic lines form closed crowed loops within a solenoid but
spread widely outside it),reducedbyadditionalshimcoilsorironshroud.
Optimum field strength is large enough to produce adequate signal, but not so
largetoexceedssafetyguidelines.
Coils:
Arrangedfromoutsidetoinside:
1) Theshimcoils:
CarryingDC,finetunedtomakethemainmagneticfieldmoreuniform.
2) Threesetsofgradientcoils:
Carrying DC, and varied to alter the
slopeofthemagneticfield
Typically20mT/m.
The currents must be switched off
rapidlyin1msorless,whichcausesthe
coilstoemitaloudrepeatedbangs.
The steeper the slice gradient, the
thinnertheslice.
The steeper the phaseencoding
gradient,thefinephasematrix
Steeper the frequencyencoding
gradient,neededforsmallFOV.
3) RF(transmitter/receiver)coils:
Tunedlikearadiototheresonantfrequency.
Producemagneticfieldatrightangelofthemainfield.
Types:
1) Standardbodycoil:permanentpartofthescanner,transmittheRFinallscan
typesandreceivesonlywhenimaginglargeparts(e.g.abdomen).
2) Headcoil:transmitandreceiveasapartofthehelmetusedinbrainscan.
3) Surfacecoils:receivers,asseparatecoilsappliedclosetotheimagedpart,they
allowsmallervoxelsandgivebetterresolutionastheypickupalargersignal
andlessnoise,thusimprovingSNR.
4) Phased array coils: multiple individual receiver coils, they receive signals
individuallythencombinedtogivebetterSNRwithlargeFOV.
5) Transmitphasedarraycoil:produceacurrentoneachelementswithspecial
amplifierneededtodefinethiscurrent.
o Magnets&coils.
o Pulsecontrollerunit:synchronize
the gradient and RF pulses for
theselectedparameters.
o Main computer: with an array
processor for Fourier transform
andalsoimageprocessor.
o The machine setting: With steel
grindersandreinforcedconcrete.
o Liftsandpowercablesshielding:
as it may cause RF interference
anddistorttheimagewithlinear
artefacts.
o Roomwallsanddoorsshielding:
with wire mesh to prevent
externalRFinterference.
10.10 QUALITYASSURANCE
Thehomogeneityofthemagneticfieldiscrucial,andbemeasured:
DirectlybyaspecialnuclearMRprobeatdifferentpositions.
Indirectlybyusingimagingspecialdevices.
Qualityassurancechecklist:
Signal:SNR,uniformityandghosting
Geometrical:distortion,slicewidthandpositionandresolution.Usingphantoms
Functional:T1&T2,imagingspeed,stabilityandflow
Hardware:Magnet,fieldcontoursandgradientcalibration.
10.11 HAZARDSANDSAFEPRACTICE
MedicinesandHealthcareProductsRegulatoryAgency(MHRA)guidelines:
Staticmagneticfield:
Normalmode:lessthan2.5T
Controlledmode:between2.5Tand4T(Panic button must be available with visual
contact and verbal if possible)
Researchorexperimentalmode:morethan4T
Pregnantwomen:notmorethan2.5T
Staff:notmore2Tforwholebody&5Tforlimbs&over24hournotmorethan0.2T
(200mT)&0.6Tin8hours.
Usingpulseoximeterisrecommended(ECGisnormallyaffectedabove0.3T).
Ethicscommitteeapprovalisrequiredabove4T.
MRIiscontraindicatedifthepatienthasimplantedpacemaker.
Acousticnoise:duetofastswitchingmagneticfields,withmachineslimitis
140dB,andhearingprotectionisrequiredtopreventirreversibledamageat
90db(earplugsreducenoiseby1030dB).
Radiofrequencyfields:
Microwaveheating:
Mayoccurathigherfrequenciesassociatedwithstrongstaticfields.
Compensatedbyvasodilation.
Cornea(nobloodsupply)andtestesmaybeatrisk.
Metallicimplantsheatingalsomaycauseproblems.
Skinandrectaltemperatureriseshouldnotexceed1C
Specificabsorptionratio(SAR):
ItistheRFenergydepositedpermassoftissue.
Expressedaswattsperkilogram(W/Kg)
1W/kgwholebodySARwillrisebodytemperature0.5C
Lessthan15minutesexposuremustnotexceed2.0w/KgSAR(1.0C)
SARisgreaterfor:
Largebodypatient,morethansmall
Highstaticfield,morethanlowerfields
180pulse,morethan90pulses.
InSEmorethanGRE
Highconductivitytissues(brain,blood,liverandCSF)morethan
lowconductivitytissues(fatandbonemarrow)
Otherhazards:
Mechanicalattractionofferromagneticobjectsduetofringefield:
o Whichcanextendtofewmetersmayattractoxygentanks,patientbeds,
firefightingapparatusmayhavethemallcausemajoraaccidents.
o Mayconvertscissorsandscalpelsintopotentiallylethalprojectiles
o Aneurysmclipsmaybedisplacedorrotated.
o Fringefieldalsaffectsomewatches,destroydataoncomputerdisksand
creditcards
o Controlled area around the magnet must be carefully supervised with
areasoutsidenotexceed0.5mT
Emergencies:
Cardiacarrest:
RemovethepatientfromthemagnettoMRcompatibletrolley
Keepairway
Cardiacmassage
Quicklytakentoresuscitationarea.
Fire:
Resistivemagnetsshouldbeswitchedoff
Nonferrous carbon dioxide fire extinguishers devices should be used from
distance1meterormore
Superconducting magnets should be quenched only if firemen need to enter
thecontrolledarea.
Quench:
Allscanroomsshouldcontainanoxygenmonitorandgasventingsystem.
In controlled quench in case of fire or trapped person in the magnet, fixed
opendoorshouldbemaintainedtopreventbuilduppressure.
Cleartheroomincaseofaccidentalquenchandreleasingofheliumgasinthe
roomthatmaycausesuffocationorfrostbite.
********************************