Vous êtes sur la page 1sur 1

Principles of Conductivity Measurement

Importance of conductivity in process Conductivity measurement The cell constant


technology Electrolytic conductivity is determined by There are countless types of conductivity
The Electrolytic conductivity is important for a resistance measurement. In the easiest sensors whose measurement values vary by
the characterization of liquids in laboratories case the measurement cell consists of two a great margin - depending on the electrode
and factories. Measurements of single samples metal electrodes. Between these alternating assembly.
or a continuous process facilitates monitoring voltage is applied. The more ions that are in To compensate for the geometry of the
the health of the system. the sample solution, the higher the current conductivity cell a cell constant is used:
Conductivity is a measure for the total ion between the electrodes. With help of Ohms
concentration in a sample. The more acids, law the conductivity is calculated. The unit
bases and salts present in a solution, the of the electrolytic conductivity is Siemens per Conductivity [S/cm] =
higher the conductivity. In water and waste centimeter (S/cm). Measurement [S] x Cell constant [1/cm]
water mainly ions of dissolved salts are present.
Therefore, conductivity is a likely indicator for 1 S/cm = 1000 mS/cm The cell constant is either known or it is
the salt contamination of water and waste determined by means of conductivity standards
1 mS/cm = 1000 S/cm
water. Additionally the determination of the (view page 10). The cell constant has to be put
salt content in food is very important. into the transmitter prior to measurement.

Conductivity ranges of aqueous solutions:

1000 mS/cm

100 mS/cm

10 mS/cm
Conductivity measurements help to determine
the purity of water
1 mS/cm

Why liquids are conductive


100 S/cm
The electrolytic conductivity of liquids is
caused by the decomposition of dissolved
acids, bases or salts into positive cations 10 S/cm
and negative anions. For example, common
salt (NaCl) decomposes into Sodium ions
(Na+) and Cloride ions (Cl-). The conductivity 1 S/cm
of ultra pure water is very low, because
approximately one molecule of water (H2O) in
0.1 S/cm
a billion decomposes into H+ and OH- ions. This
intrinsic conductivity of water also represents
the lower border of the conductivity scale. It is 0.01 S/cm
of greatest importance in the control of pure
water in power plants, electronics and the
pharmaceutical industry. Ultra Pure Drinking Waste Process Conc.
pure water water water water acids
water

24
THE MEASURE OF EXCELLENCE TM