Vous êtes sur la page 1sur 36

General X-ray QA and QC

Guideline

THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS


General X-ray QA and QC Guideline

Faculty of Clinical Radiology

Name of document and version:


General X-ray QA and QC Guideline, Version 1

Approved by:
Faculty of Clinical Radiology

Date of approval:
1 November 2013

ABN 37 000 029 863


Copyright for this publication rests with The Royal Australian and New Zealand College of
Radiologists

The Royal Australian and New Zealand College of Radiologists


Level 9, 51 Druitt Street
Sydney NSW 2000 Australia

Email: ranzcr@ranzcr.edu.au
Website: www.ranzcr.edu.au
Telephone: +61 2 9268 9777
Facsimile: +61 2 9268 9799

Disclaimer: The information provided in this document is of a general nature only and is not intended as a
substitute for medical or legal advice. It is designed to support, not replace, the relationship that exists
between a patient and his/her doctor.
TABLE OF CONTENTS

1. Introduction 3
2. Daily / As Required Tests 5
3. Weekly Tests 7
4. Monthly Tests 9
5. Quarterly Tests 13
6. Six-Monthly Tests 28
7. Review of QC Results by Designated Person 30
8. Related policy documents 31
9. Appendices 31
10. References 31

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 2 of 33
1. INTRODUCTION

1.1 Preamble

Quality Assurance is a planned and systematic approach to monitoring and assessing the
care provided, or the service being delivered, that identifies opportunities for improvement
and maintains a mechanism through which action is taken to make and maintain these
improvements (Australian Council of Healthcare Standards).

The scope of activities in the quality management process for general radiology imaging
systems includes, but is not limited to:

i. Radiation protection and radiation monitoring (optimisation of radiation dose and


image quality, structured dose reports, establishment of diagnostic reference
levels);

ii. Equipment testing (routine Quality Control);

iii. Other recommended QA systems for general imaging (CR and DR systems), film
printing, screening/fluoroscopy, mobile imaging, and theatre imaging;

iv. Clinical image review and repeat / reject analysis

This document has been produced as a resource document for best practice in these areas
of quality management in Medical Imaging departments and assumes that all additional
state / legislative requirements such as equipment compliance testing, licencing,
registration, record keeping and regular service maintenance tests are conducted as part
of routine site accreditation requirements.

1.2 Purpose and Scope

(a) This guideline is intended to assist The Royal Australian and New Zealand College
of Radiologists (ABN 37 000 029 863) (the College), radiologists, radiographers,
equipment assessors in addressing quality control requirements relating to
radiography equipment. This document sets out standardised procedures and
instructions to complete quality control processes.

(b) This document will be reviewed at the end of 2015. Please refer to individual
manufacturers guidelines for any additional Quality Assurance requirements that
may be applicable for specific systems.

1.3 Overarching Management of QA and QC

It is recommended that at least one staff member at each facility is identified and
responsible for ensuring quality checks are undertaken at specified times. The designated
person(s) may include one or more of the following staff members: Chief Radiographer /
Assistant Chief Radiographer / Quality Manager / QA Radiographer / Section Senior
Radiographer / RSO. The designated quality manager should also review non-compliance
issues and provide strategies for corrective and preventative action in a timely manner and
formulate quality improvement strategies within the department.

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 3 of 33
The designated quality manager is also responsible for the organisation, dissemination and
document control of all incoming quality guidelines, policies and directives from accrediting
/ regulatory bodies (e.g. RANZCR & ISO/IEC / Radiation Protection Authority), Local
Health Service and State Departments of Health.

1.4 Reviewing QC Results and Taking Action

It is essential to review QC results immediately and take action if they are out of tolerance.
In some cases, it may be necessary to contact the service engineer. The urgency of
remedial action by the service engineer should be determined by personal judgment. For
example, a minor failure in x-ray to light-beam alignment could be fixed at the next service
visit but anything that may affect patient dose or image quality should be addressed
immediately. Prior to contacting the service engineer, you may wish to carry out the
following steps:

i. Repeat the test. If you are not confident in carrying out QC tests, consult an
experienced colleague ;

ii. Check that you have used correct and consistent settings e.g. kV, mAs, test object,
focus to detector distance;

iii. If artefacts are present, try to identify their location e.g. on the monitor, the
detector, the test object. Clean equipment and repeat the test;

iv. Contact your medical physicist / equipment assessor for advice (if available)

1.5 Acknowledgements

The Royal Australian and New Zealand College of Radiologists gratefully acknowledges
the extensive work undertaken by the General X-Ray working group consisting of Dr
Patrick Brennan, Ms Ingrid Egan, Dr Nick Pang, Mr John Robinson, Dr Andrew Scott and
representatives of the ACPSEM Dr Lee Collins, Dr Jenny Diffey and Dr Ian Smith, in
development of this guideline.

1.6 Definitions

In this Guideline:

College means The Royal Australian and New Zealand College of Radiologists.

Member means a member of the College.

1.7 College Mission

The Colleges Mission is:

As a fellowship based organisation, The Royal Australian and New Zealand College of
Radiologists sets, promotes and continuously improves the standards of training and
practice in radiology and radiation oncology, for the betterment of the people of Australia
and New Zealand.

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 4 of 33
2. DAILY / AS REQUIRED TESTS

2.1 Detector Calibration / Flat Field / Dark Noise Test (DR Only)

Frequency: As recommended by manufacturer (see below).

Tests such as Detector Calibration are not specifically included in the RANZCR guidelines,
due to vendor variability. For example, Carestream require a daily detector calibration,
known as the Daily Dark, whilst other vendors require an annual detector calibration which
is normally performed by the service engineer.

For this reason, no specific protocol or frequency is defined by RANZCR. Instead, the
advice of the Working Group is simply to follow the advice of the vendor.

These tests are generally pass / fail and no output is recorded. An example form for
recording that a daily detector calibration has been completed is included as Record Sheet
2.1.a.

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 5 of 33
2.1.a Record Sheet - Daily Detector Calibration

X-ray equipment ID: ________________ DR manufacturer: ________________ Test (e.g. Daily Dark): ___________________

Month: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Pass () / Fail ()

Initials

Month: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Pass () / Fail ()

Initials

Month: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Pass () / Fail ()

Initials

Month: 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31

Pass () / Fail ()

Initials
3. WEEKLY TESTS

3.1 Image Plate Erasure (CR Only)

Frequency: Weekly

CR image plates are sensitive to scattered and naturally occurring radiation sources and if
left unused for long periods of time will store energy absorbed from these sources. It is
recommended that all CR image plates be subjected to a primary erasure procedure on a
weekly basis, ideally on a Monday morning.

Note that this frequency is based on the assumption that unused cassettes are stored
under conditions whereby they are unlikely to receive any significant unintended exposure.
Should unexposed cassette storage be less than optimal, an increased frequency of plate
erasure will be required.

3.1.1 Procedure

Perform a primary erasure on all CR cassettes


Complete Record Sheet 3.1.a

3.1.2 Record

Date test was performed


Person performing test

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 7 of 33
3.1.a Record Sheet Image Plate Erasure (CR only)

Frequency: Weekly

Site / Department: ____________________________

Date Initials Date Initials Date Initials

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 8 of 33
4. MONTHLY TESTS

4.1 Monitor QC

Frequency: Monthly Test

The monitor in the x-ray room, attached to the operating console, is referred to as the
secondary or acquisition monitor. The high resolution monitors used for reporting the x-ray
examination images are known as the primary or reporting / diagnostic review monitors. It
is ONLY necessary to carry out this test on the primary (reporting) monitors.

4.1.1 Procedure

Display the appropriate version of the TG18-QC test pattern (see Appendix 1)
If window width (WW) and window level (WL) are displayed, these should be set to
4096 and 2048 respectively. Please note that these will not always be displayed, or be
adjustable. If the pattern has been installed correctly, these will be the default settings
Ensure viewing conditions are acceptable
Examine image paying particular attention to the following:
o Visibility and distortion of items used for the evaluation of the quality of the image
o Disturbing artefacts
Visually check for fingerprints and dust and clean gently using a lint-free cloth, or as per
manufacturers recommendations
If monitors are installed as a pair, visually check that the luminance of the left and right
monitors are matched i.e. white, dark and grey squares should appear to have the same
greyscale (please be aware that this test is subjective!)
Complete monthly checklist and Record Sheet 4.1.a (you will need one Record Sheet
for each monitor / pair of monitors)

4.1.2 Record

Date test was performed


Person performing test
Monitor identification
Test results

4.1.3 Limits

Borders are visible


Lines are straight
Squares appear square
No smearing at black-white transitions
Squares of different shades from black to white are distinct
The finest horizontal and vertical line pairs are visible in the high contrast pattern in all
four corners and in the centre
The 5% and 95% pixel value squares are clearly visible within the 0% and 100%
backgrounds
The pattern is centred in the active area
No disturbing artefacts are visible
At least 11 letters (i.e. QUALITY CONT) are visible in the phrase QUALITY CONTROL
for the dark, mid-grey and light renditions. Note that you should record how many letters
can be seen

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 9 of 33
4.1.a Record Sheet Monitor QC

Frequency: Monthly Test

Monitor ID (e.g. vendor, model, location): _______________________________________

Date:
Initials:
Left / right (if applicable) Left / Right / NA Left / Right / NA Left / Right / NA Left / Right / NA
Pass () / Fail ()

General Image Quality


No smearing
No artefacts
Ramps continuous
General distortion
Lines straight
Boxes square
Luminance
Greyscale patches
distinct
5% square visible in
0% background
95% square visible in
100% background
Finest high contrast resolution elements visible (in all 4 corners and at centre)
Horizontal line pairs
Vertical line pairs
Number of letters visible (at least 11 or QUALITY CONT)
Dark
Mid-grey
Light
Visual check for fingerprints / dust and cleaning (if required)
Completed
Monitor matching if applicable (subjectively by eye)
Dark areas matched
White areas matched
Comments

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 10 of 33
4.2 Printer QC (if applicable)

Frequency: Monthly Test, or prior to printing if used less frequently

Please note: Although many practices will not be printing films for primary reporting, it is
still important to make sure that printed images are of a relatively high standard if being
sent as the medical record.

It is acknowledged that sites will no longer own a densitometer and for this reason, a visual
check only is considered sufficient. It is expected that the Service Engineer will confirm that
the optical density range is acceptable during their routine service visit.

Note that a Printer Calibration is also required at installation, but this should be done in
conjunction with the Service Engineer. Given that there may be a large number of x-ray
equipment vendors at a single site, it is recommended that clinical images are sent to the
printer from PACS, rather than from the Acquisition Workstation.

4.2.1 Procedure

Print the TG18-QC test pattern (see Appendix 1) directly from the printer.
Check visibility and distortion of several items used for evaluating the quality of the
image
Check for disturbing artefacts
Clean the rollers (as required; based on frequency of use)
Complete monthly check list and Record Sheet 4.2.a (you will need one Record Sheet
for each printer)

4.2.2 Record

Date test was performed


Person performing test
Printer identification
Test results

4.2.3 Limits

Borders are visible


Lines are straight
Squares appear square
Squares of different shades from black to white are distinct
The finest horizontal and vertical line pairs are visible in the high contrast pattern in all
four corners and in the centre
The 5% and 95% grey squares are clearly visible within the 0% and 100% backgrounds
At least 11 letters (i.e. QUALITY CONT) are visible in the phrase QUALITY CONTROL
for each of the dark, mid-grey and light renditions. Note that you should record how
many letters can be seen
No disturbing artefacts are visible
Rollers clean

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 11 of 33
4.2.a Record Sheet Printer QC

Frequency: Monthly Test, or prior to printing if used less frequently

Printer ID (e.g. vendor, model, location): ____________________________________

Date:
Initials:
Pass () / Fail ()
General Image Quality
No smearing
No artefacts
Ramps continuous
General distortion
Lines straight
Boxes square
Luminance
Greyscale patches distinct
5% square visible in 0% background
95% square visible in 100% background
Finest high contrast resolution elements visible (in all 4 corners and at centre)
Horizontal line pairs
Vertical line pairs
Number of letters visible (at least 11 or QUALITY CONT)
Dark
Mid-grey
Light
Printer rollers
Cleaned (if required)
Comments

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 12 of 33
5. QUARTERLY TESTS

5.1 Mechanical Inspection

Frequency: Quarterly

The practice staff must perform an overall mechanical inspection of the digital x-ray system
and associated components. The inspection should be carried out quarterly to ensure that
there are no hazardous, inoperative, out of alignment or improperly operating items on the
system.

The Service Engineer may perform additional checks during their service visit, for example
CR Plate Sensitivity Matching. It is good practice to keep their service report alongside the
site QC records.

5.1.1 Procedure

Visual inspection of the system to ensure safe and optimum operation, using the checklist
provided on Record Sheet 5.1.a.

5.1.2 Record

Date inspection performed


Inspection results
X-ray equipment identification
Person performing test

5.1.3 Limits

Mechanical and safety functions operating correctly

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 13 of 33
5.1.a Record Sheet Mechanical Inspection

Frequency: Quarterly

X-ray Equipment ID: ____________________________________________________

Date: Initials:
Item Description Pass () Fail ()
Check that all cables are free from breaks, kinks or knots. Cables should not be
1
under other heavy equipment

2 Verify that the interlocks and brakes are working correctly

3 Ensure that the table, tube and bucky move smoothly

4 Ensure that control panel switches, indicator lights and meters are functioning

Ensure that the field light is functioning, has adequate intensity and collimator is
5
free from dust

6 Ensure that the current technique charts are displayed near the control panel

Ensure that there are no oil leaks around the x-ray tube and generator, and that
7
these are free from dust

8 Check that CR plates are clean and free from artefacts (CR only)

On the reporting workstation, display a recent clinical image and verify that the
9 time and date as well as the facility identification are correct in the image
annotation
Check that the x-ray tube and generator model and serial numbers are clearly
10 marked and readable (if labels are inaccessible, serial numbers must be
displayed elsewhere e.g. in a file or label drawer)
Visually inspect gonad shields and personal protective devices (e.g. lead aprons,
11 a
thyroid shields) for shielding integrity
Confirm that the operators view of the patient from the control window is not
12
obstructed by notices / charts
b Ensure that the radiation warning sign on the door is intact and that the room
13
warning lights are working
b
14 Ensure that cassette localization / auto-collimation and locks are working

b Check that centring and Source to Image Distance (SID) detents work correctly.
15
Verify the accuracy of the distance displayed on the collimator

a
Although a visual inspection will suffice on a quarterly basis, please be aware that some State
legislation (e.g. NSW EPA) requires that lead aprons are fluoroscopically screened on an
annual basis
b
Not applicable to mobile units

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 14 of 33
5.2 Alignment: X-ray to light field (CR and wireless DR detectors) / X-ray to
detector (integrated DR detectors)

Frequency: Quarterly

To ensure accurate collimation and full coverage of image receptor without unnecessary
exposure of patient. This test should be performed on all x-ray units.

5.2.1 Procedure

CR and wireless DR detectors

Ensure x-ray tube and table are level.


Place CR cassette / wireless DR detector on table top.
Set a source to detector distance of 100cm
Dim the room lights so that you can clearly see the edges of the light field
Collimate to an area greater than 10 x 10cm the exact area is not important as long as
it is within the CR cassette / wireless DR detector.
Use coins to mark the edges of the light field. It is good practice to place a coin within
and outside the light field (see Figure 1). Use an additional coin to mark the anode side.
Edges of light
field

Coin to mark
anode side

Figure 1: Position of coins for x-ray to light-beam alignment test

Make an exposure using 60kV and 5mAs


Open up the collimators fully and make an exposure using 60kV and 1mAs.
Readout the CR / DR image. Measure the difference between the x-ray field and the
edge of the coin (light field)
Repeat for broad and fine focus (if available).
Note that if the x-ray field is greater than the light field, this is a +ve difference
Complete Record Sheet 5.2.a

Integrated DR detectors

These are units where the digital detector is built in to the bucky. Markings on the top of the
bucky show the position of the detector.

Ensure x-ray tube and table are level


Set a source to detector distance of 100cm
Set auto-collimation
Dim the room lights so that you can clearly see the edges of the light field. Visually
check that the light field coincides with the markings on the bucky

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 15 of 33
Switch the room lights on. Align coins with the edges of the markings on the bucky. Use
an additional coin to mark the anode side.
Make an exposure using 60kV and 5mAs
Measure the difference between the edge of the image and the edge of the coin (light-
field)
Repeat for broad and fine focus (if available)
Note that if the imaged field is greater than the edge of the markings, this is a +ve
difference
The x-rays should extend to all edges of the detector (i.e. there should be no white
bands around the edge of the image)
Complete Record Sheet 5.2.b

5.2.2 Record

Date test was performed


Deviation of x-ray from light field / detector
X-ray equipment identification
Person performing test

5.2.3 Limits

X-ray to light-field must be within 1% of SID


X-ray to detector must agree to within 2% of SID. X-ray field should extend to all edges
of the detector

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 16 of 33
5.2.a Record Sheet Alignment: X-ray to light field (CR and wireless DR detectors)

Frequency: Quarterly

X-ray Equipment ID: _____________________

LIMITS
X-ray to light field must be within 1% of SID (i.e. 1 cm for SID of 100 cm)

Date

Initials

Broad Focus

Deviation between X-ray and light field (mm)


(positive indicates x-rays extend beyond light field)

Anode Edge

Cathode Edge

Top Edge

Bottom Edge

Fine Focus

Deviation between X-ray and light field (mm)


(positive indicates x-rays extend beyond light field)

Anode Edge

Cathode Edge

Top Edge

Bottom Edge

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 17 of 33
5.2.b Record Sheet Alignment: X-ray to detector (integrated DR detectors)

Frequency: Quarterly

X-ray Equipment ID: _____________________

LIMITS
X-ray to detector must agree to within 2% of SID
X-ray field should extend to all edges of the detector
Light field should coincide with markings on bucky

Date

Initials

Broad Focus
Deviation between x-ray imaged field and detector markings (mm)
(positive indicates that the imaged field is greater than the detector markings on the bucky)
Anode Edge

Cathode Edge

Top Edge

Bottom Edge

Confirm that the light field coincides with the markings on the bucky

Pass () Fail
()

Comments

Fine Focus
Deviation between x-ray imaged field and detector markings (mm)
(positive indicates that the imaged field is greater than the detector markings on the bucky)
Anode Edge

Cathode Edge

Top Edge

Bottom Edge

Confirm that the light field coincides with the markings on the bucky

Pass () Fail
()

Comments

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 18 of 33
5.3 Consistency of Exposure Index

Frequency: Quarterly

To ensure that exposure index remains approximately constant under given exposure
conditions.

Note that this test requires you to set a baseline. This is done by following the procedure
below and performing the test on 3 occasions (separated by e.g. an hour or a day). Doing it
3 times in a row is not recommended. If the results are consistent, set the average value as
your baseline. Make a note of the date you set the baseline. All future results are compared
to this value.

If results are not consistent, repeat the test checking that you have followed the correct
procedure using identical exposure conditions.

There are occasions where it may be necessary to set a new baseline value, for example,
following software upgrades or tube changes. Follow the steps described above and note
the date the new baseline was set and the reason for the change. You should not set a
new baseline simply because the test fails, unless an investigation of the reason for the
failure indicates that this is justified.

5.3.1 Procedure

CR

A designated CR test cassette (freshly erased) and CR reader must be assigned for this
test; this ensures consistency of measurement conditions. It is preferable to assign a
cassette which is in clinical use as this will enable any deterioration due to general wear
and tear to be detected, which may indicate that image plates are due for replacement. The
test should be performed on all x-ray units.

Note that in departments with multiple CR readers, it is considered good practice to repeat
the test for all CR readers using one designated x-ray unit to make the exposures (there is
no need to test all combinations of x-ray units and CR readers).

Place the CR test cassette on the table


Set focus to table distance of 110cm and centre x-ray beam on CR cassette
Ensure collimator position covers the full area of the CR plate
Attach 1mm copper sheet to tube collimator, ensuring beam fully intercepted
Make a manual exposure using 70kV, 4mAs
Process the image plate using a consistent image processing algorithm (e.g. L-spine)
Record the Exposure Index (note that this will be manufacturer-specific e.g. Fuji = S
number, Carestream = EI, Agfa = SAL or LgM)
Use Record Sheet 5.3a

DR

If using a wireless DR detector, place this on the table


Set focus to detector distance of 110cm and centre x-ray beam on detector
Ensure collimator position covers the full area of the detector
Attach 1mm copper sheet to tube collimator, ensuring beam fully intercepted
Use a consistent image processing algorithm (e.g. L-spine)
Make a manual exposure using 70kV, 4mAs

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 19 of 33
Record the Exposure Index (note that this will be manufacturer-specific e.g. Carestream
& Philips = EI, Canon = EXI, Shimadzu = S number)
Use Record Sheet 5.3a

5.3.2 Record

Date test was performed


Processing algorithm
Exposure Index
X-ray equipment identification
Person performing test

5.3.3 Limits

Exposure Index should be within 10% of the baseline value


If the change in Exposure Index over consecutive periods follows a trend (e.g. a
continual increase or decrease in value), discuss the results with the Service Engineer

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 20 of 33
5.3.a Record Sheet Consistency of Exposure Index

Frequency: Quarterly

CR Systems Only

CR Test Cassette ID: _____________________

CR Reader ID: __________________________

X-ray Equipment ID: _____________________

LIMITS
Variation of the EI from the baseline value < 10%

BASELINE DATA: Specify Exposure Indicator e.g. S number, EI ____________________

Date ____________ Baseline Lower Limit Upper Limit


EI

Date

Initials

Processing Algorithm

EI

EI within limits (P / F)

Comments

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 21 of 33
5.4 Image Uniformity and Artefact Evaluation

Frequency: Quarterly

To ensure that image is uniform and artefact-free. Additionally for DR, to ensure that flat-
field correction is working.

Note that as part of the process of image critique, every clinical image acquired should be
subject to a quick artefact check before being sent to PACS or printed

5.4.1 Procedure

Use the image acquired for 5.3 Consistency of Exposure Index. Visually inspect the image
for non-uniformity and artefacts, accepting that CR images will suffer from the anode-heel
effect (i.e. a decrease in greyscale intensity as you move from the cathode to the anode
side of the image). Use Record Sheet 5.4.a.

5.4.2 Record

Date test was performed


Comments on uniformity and artefacts
X-ray equipment identification
Person performing test

5.4.3 Limits

Uniform image
No evidence of artefacts

5.4.4 Action

If artefacts are seen it is important to determine whether they are due to the monitor, the
detector, x-ray beam non-uniformity or the copper. Take the following steps:

To eliminate the possibility of display artefacts, rotate or pan the image. If the artefact
moves with the image it is due to the imaging system; if it stays in the same place, it is
due to the display system (monitor).
To eliminate the possibility of artefacts (dust, scratches or thickness non-uniformity) on
the copper, rotate the copper and repeat the test. If the artefact stays in the same
place, it is most likely to be due to the detector; if it moves, it may be attributable to non-
uniformity on the copper.
Clean the equipment (if possible and following vendor guidelines) and repeat the test.

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 22 of 33
5.4.a Record Sheet Image Uniformity and Artefact Evaluation

Frequency: Quarterly

X-ray Equipment ID: _____________________

LIMITS
Image appears uniform
No artefacts present

Date

Initials

Image appears
uniform? (Y/N)
Comments (heel
effect etc?)

Artefacts present?
(Y/N)

Describe

Action taken (if


applicable)

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 23 of 33
5.5 AEC Consistency

It is important to ensure that the AEC operation remains stable over time. This should be
checked using table and vertical buckies and by choosing examinations which encompass
a range of AEC chambers.

Note that this test requires you to set baseline values for mAs and EI for each examination.
This is done by following the procedure below and performing the test on 3 occasions
(separated by e.g. an hour or a day). Doing it 3 times in a row is not recommended. If the
results are consistent, set the average value as your baseline. Make a note of the date you
set the baseline. All future results are compared to this value.

If results are not consistent, repeat the test checking that you have followed the correct
procedure using identical exposure conditions.

There are occasions where it may be necessary to set a new baseline value, for example,
following software upgrades or tube changes. Follow the steps described above and note
the date the new baseline was set and the reason for the change. You should not set a
new baseline simply because the test fails, unless an investigation of the reason for the
failure indicates that this is justified.

5.5.1 Procedure

Select up to 3 frequently performed clinical examinations which are carried out under
AEC. Try to include examinations which cover table and vertical buckies (if applicable)
and a range of AEC chambers (e.g. Table bucky: L-Spine & Abdomen / Pelvis; Vertical
bucky: Chest). Record these in the table on Record Sheet 4.5a
Place designated CR test cassette / wireless digital detector in bucky (if applicable)
For the table bucky, set the source to detector distance that would be used clinically
(e.g. 110cm) and centre the x-ray beam to the centre of the bucky and the detector
Ensure collimator position covers the full area of the AEC chambers
Attach 1mm copper sheet to tube collimator, ensuring beam fully intercepted
Select your examination (see table) and choose the clinically relevant kVp (e.g. 70kVp)
Make an exposure under AEC and record indicated post-exposure mAs value
Record the Exposure Index (note that this will be manufacturer-specific e.g. Fuji = S
number, Carestream = EI, Agfa = SAL or LgM)
Repeat for the second selected examination (see table)
Repeat for the vertical bucky (if applicable), setting the SID and kVp that would be used
clinically
Use Record Sheet 5.5.a

5.5.2 Record

Complete table of up to 3 frequently performed examinations


Date test was performed
Person performing test
X-ray system identification
Current density setting
post exposure mAs
Exposure Index

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 24 of 33
5.5.3 Limits

Post-exposure mAs within 20% of baseline values

No limits are set for Exposure Index, due to vendor variability, but if the change in
Exposure Index over consecutive periods follows a trend (e.g. a continual increase or
decrease in value), discuss the results with the Service Engineer

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 25 of 33
5.5.a Record Sheet AEC Consistency

Frequency: Quarterly

X-ray Equipment ID: ____________________

Choose up to 3 frequently performed clinical examinations which are carried out under
AEC and record the factors below.

Examination Bucky used kVp Chamber(s) Grid SID


(Y/N) (cm)

1 Table

2 Table

3 Vertical

LIMITS
mAs 20%

No limits are set for Exposure Index, due to vendor variability, but if the change in Exposure
Index over consecutive periods follows a trend (e.g. a continual increase or decrease in
value), discuss the results with the Service Engineer

BASELINE DATA
Date: Baseline Lower Limit Upper Limit
mAs for Exam 1
mAs for Exam 2
mAs for Exam 3
EI for Exam 1
EI for Exam 2 Not required
EI for Exam 3

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 26 of 33
Date : ____________________ Initials: _____________________________
mAs
Post within EI within
Density Exposure
Examination kV Exposure acceptable acceptable
setting Indicator
mAs range range (/)
(/)
1
2

Date : ____________________ Initials: _____________________________


mAs
Post within EI within
Density Exposure
Examination kV Exposure acceptable acceptable
setting Indicator
mAs range range (/)
(/)
1

Date : ____________________ Initials: _____________________________


mAs
Post within EI within
Density Exposure
Examination kV Exposure acceptable acceptable
setting Indicator
mAs range range (/)
(/)
1

Date : ____________________ Initials: _____________________________


mAs
Post within EI within
Density Exposure
Examination kV Exposure acceptable acceptable
setting Indicator
mAs range range (/)
(/)
1

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 27 of 33
6. SIX-MONTHLY TESTS

6.1 Image Quality

Frequency: Six Monthly

To ensure adequate image quality from each x-ray unit.

Procedure

Select 5 patients (mix of male and female) who have had a chest PA / AP x-ray for each
x-ray unit
Perform an assessment of image quality using the checklist on Record Sheet 6.1.a

Record

Date of assessment
X-ray equipment ID
Radiologist performing test
Complete checklist

Limits

Images must be deemed clinically acceptable. If not, the reasons must be investigated.

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 28 of 33
6.1.a Record Sheet Image Quality

Frequency: Six Monthly

Date: _________________________ X-ray Equipment ID: ____________________

Name of Radiologist _____________________________________________________

Chest AP / PA Film 1 Film 2 Film 3 Film 4 Film 5


Exposure Indicator
Image Assessment Criteria for Chest PA
1. Performed at deep inspiration (as
assessed by the position of ribs above the
diaphragm either 6 anteriorly or 10
posteriorly)
2. The spinous processes of the dorsal
vertebrae are equidistant from the inner
borders of the clavicles (no thorax rotation)

3. Medial border of the scapulae to be


projected outside the lung fields

4. Reproduction from the 7th cervical vertebra


to the bottom of both costophrenic sinuses
(reproduction of the whole thoracic cavity)

5. Visually sharp reproduction of the


peripheral vessels (performed with
suspended respiration)

6. Visually sharp reproduction of the trachea,


major bronchi, and borders of the cardiac
silhouette

7. Visually sharp reproduction of both


hemidiaphragms

8. Visualization of the lung vessels and


mediastinal outline through the cardiac
silhouette

9. The lower dorsal spine is no more than


faintly appreciable with visualization of the
intervertebral spaces

10. Quality overall acceptable for clinical


purpose

Poor Features invisible, detail invisible or not clear


Satisfactory Features just visible, detail just visible but not clearly defined
Good Features detected and fully reproduced, detail visible and clearly defined

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 29 of 33
7. REVIEW OF QC RESULTS BY DESIGNATED PERSON

The Designated Person may wish to use the form below to record the results of their review.

Site: _____________________________________________ X-ray Equipment ID: _____________________________________

Year

Jan Feb Mar Apr May Jun Jul Aug Sep Oct Nov Dec
Image Plate Erasure (Weekly)

Monitor QC (Monthly)

Printer QC (Monthly)

Mechanical inspection (Quarterly)

X-ray Alignment (Quarterly)

Consistency of Exposure Index (Quarterly)

Image Uniformity and Artefact Evaluation (Quarterly)

AEC Consistency (Quarterly)

Image Quality (Quarterly)

Comments

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 30 of 33
8. RELATED POLICY DOCUMENTS
Guideline for Quality Control Testing for Digital (CR and DR) Mammography V3.

9. APPENDICES
A. Monitor and Printer Quality Control: TG18 Test Pattern

10. REFERENCES

CRCPD (Conference of Radiation Control Program Directors) QC Recommendation Vol 3

EPA Radiation Guidelines

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 31 of 33
Appendix A

Monitor and Printer Quality Control: TG18-QC Test Pattern

Display devices (monitors and printers) form part of the imaging chain and it is imperative that QC
is carried out to ensure that:

1) they meet the required standards at baseline, and

2) performance does not deteriorate over time.

In order to do this, the American Association of Physicists in Medicine (AAPM) has developed a
1
series of test patterns known as the Task Group 18 (TG18) series . These are not physical test
objects, but exist as files which can be loaded on to systems and displayed on the monitor or
printed. It is important to know how to access them. Sometimes they are locked as a patient in
PACS, but more commonly, they may be accessed through vendor-specific software such as
Barco QA Web or Eizo RadiCS.

The TG18-QC test pattern is shown overleaf. This should be displayed at full resolution (one
display pixel for each pixel in the digital image). A visual inspection of the image should be
carried out including tests such as:

Are the squares from black (100%) to white (0%) distinct?


Can you see the 95% square inside the 100% square?
Can you see the 5% square inside the 0% square?
Can you see the contrast patches in the corners of each grayscale square?
Can you resolve the high contrast and low contrast line pair groups in all four corners and
the centre?
Can you see at least 11 letters in the phrase QUALITY CONTROL?
Are the grayscale ramps on the left and right of the image continuous (i.e. does the black
blend smoothly into the white)?

Further details are given in the test protocols (4.1 and 4.2).

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 32 of 33
Appendix A

Figure 2: TG18-QC test pattern showing resolution patterns, ramps, 5% and 95% squares and
corner patches

1. Assessment of Display Performance for Medical Imaging Systems. American Association


of Physicists in Medicine. www.aapm.org/pubs/reports/OR_03.pdf

General X-ray QA and QC Guideline V1


The Royal Australian and New Zealand College of Radiologists
November 2013
Page 33 of 33
THE ROYAL AUSTRALIAN AND NEW ZEALAND COLLEGE OF RADIOLOGISTS

Vous aimerez peut-être aussi