ApplicationNote

Extraction of Mineral Oils from Water Solid Phase Extraction


for evaluation purposes only

Field Environmental

Sample Mineral Oils (C9-C28)

Matrix Water

Extraction Column BAKERBOND spe C18, 6 mL, 1000 mg

Sample Preparation Filter the sample if necessary.


Column Conditioning Wash the column with 2 4 mL of Tetrachloroethylene. Vacuum dry the column for 2 minutes. Add 2 4 mL
of methanol to the column and slowly aspirate through the column, followed by 2 4 mL of water. Do not
allow the column to run dry during and after conditioning the column.
Sample Addition/Wash With vacuum off, add 2 mL of the sample to the column. Attach adapter and 15 mL reservoir to the column
and add the remaining sample to it. Slowly aspirate through the column. Wash the column with 2 4 mL of
water. Vacuum dry the column for 5 minutes
Sample Elution Elute the mineral oil from the column with tetrachloroethylene as the eluting solvent.

Final Analytical Method GC-FID

Product List
Product Number Product Number Description
(Europe) (USA)
7020-07 7020-07 BAKERBOND spe C18, 6 mL, 1000 mg
4218 4218 Water, BAKER ANALYZED HPLC Reagent
8402 9093 Methanol, BAKER ANALYZED HPLC Reagent
7072 9465 Tetrachloroethylene, BAKER ANALYZED Reagent
7018-94 7018-00 BAKER spe-12G System
7122-00 7122-00 SPE Column Adapter
7119-01 7119-01 Reservoir, 15 mL

A P P L I C A T I O N N O T E | www.avantormaterials.com
Phillipsburg, NJ 9001: 2008 & 14001: 2004
Paris, KY 9001: 2008
Mexico City, Mexico 9001: 2008
Deventer, the Netherlands 9001: 2008 & 14001: 2004 & 13485: 2003
Selangor, Malaysia 9001: 2008

About Avantor Performance Materials


Avantor Performance Materials manufactures and markets high-performance chemistries and materials
around the world under several respected brand names, including the J.T.Baker, Macron Fine Chemicals,
Rankem, Diagnova, BeneSphera, and POCH brands.

Avantor products are used in a wide range of industries. Our biomedical and life science solutions are
used in academic, industry and quality control laboratories for research, pharmaceutical production and
medical lab testing, while our electronics solutions are used in the manufacturing of semiconductors and
flat panel displays. Based in Center Valley, Pennsylvania (USA), Avantor is owned by an affiliate of New
Mountain Capital, LLC.

For additional information please visit www.avantormaterials.com or follow www.twitter.com/avantor_news

Ordering Information and Assistance Corporate Headquarters Worldwide Locations


Customer Service and Technical Service AskAvantor Avantor Performance Materials, Inc. China Malaysia North America
toll free: +1-855-AVANTOR (+1-855-282-6867) Our Web site features ASK Avantor, 3477 Corporate Parkway India Mexico Taiwan
outside of u. s. tel: +1-610-573-2600 which includes live chat capabilities Suite #200 Korea The Netherlands
fax: +1-610-573-2610 with customer service representatives. Center Valley, PA 18034 For contact information at these locations, visit
e- mail: info@avantormaterials.com www.avantormaterials.com/askavantor USA www.avantormaterials.com/Support/Contact-
www.avantormaterials.com Us/Worldwide-Directory.aspx
Lit Number: U-PP-MineralOilsWater-10V1-E-MIN-OILS
2012 Avantor Performance Materials, Inc. All rights reserved.
Trademarks are owned by Avantor Performance Materials, Inc. or its affiliates unless otherwise noted.

Vous aimerez peut-être aussi