Vous êtes sur la page 1sur 18

ARTICLE 12

T-1210 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222


T-1220 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
T-1221 Vessel Stressing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
T-1222 Noise Reduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
T-1223 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
T-1224 Location of Acoustic Emission Sources . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222
T-1225 Procedure Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1230 Equipment and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1240 Application Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1241 Equipment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1242 System Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1243 Pre-Examination Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223
T-1244 Examination Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 224
T-1260 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
T-1280 Evaluation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
T-1281 Evaluation Criteria . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
T-1290 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
T-1291 Written Report. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225
T-1292 Record. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227

Figure
T-1244.3.2 An Example of Pressure Vessel Test Stressing Sequence. . . . . . . . . . . . . . . . . . . . . . . . 225

Table
T-1281 An Example of Evaluation Criteria for Zone Location . . . . . . . . . . . . . . . . . . . . . . . . . . 226

Mandatory Appendices
Appendix I Instrumentation Performance Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1210 Acoustic Emission Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1220 Signal Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1230 Couplant . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1240 Preamplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1250 Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1260 Power-Signal Cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1270 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1280 Main Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228
I-1290 Main Processor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
I-1291 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
I-1292 Peak Amplitude Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229

Appendix II Instrument Calibration and Cross-Referencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229


II-1210 Manufacturers Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229

220

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
II-1211 Annual Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
II-1220 Instrument Cross-Referencing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
II-1221 Sensor Characterization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
Appendix III Glossary of Terms for Acoustic Emission Examination of Metal
Pressure Vessels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
III-1210 Scope. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229
III-1220 General Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230
III-1230 Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230

Nonmandatory Appendices
Appendix A Sensor Placement Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231
Appendix B Supplemental Information for Conducting Acoustic Emission
Examinations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
B-10 Frequency Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
B-20 Combining More Than One Sensor in a Single Channel . . . . . . . . . . . . . . . . . . . . . . . . 237
B-30 Attenuative Welds. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237
B-40 Production Line Testing of Identical Vessels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237

221

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
ARTICLE 12
ACOUSTIC EMISSION EXAMINATION OF
METALLIC VESSELS DURING PRESSURE TESTING

T-1210 SCOPE shall be made for holding the pressure at designated


hold points.
This Article describes methods for conducting acous-
tic emission (AE) examination of metallic pressure
vessels during acceptance pressure testing when speci- T-1222 Noise Reduction
fied by a referencing Code Section. When AE examina-
tion in accordance with this Article is specified, the External noise sources such as rain, foreign objects
referencing Code Section shall be consulted for the contacting the vessel, and pressurizing equipment noise
following specific requirements: must be below the system examination threshold.
(a) personnel qualification /certification requirements;
(b) requirements /extent of examination and /or vol- T-1223 Sensors
ume(s) to be examined;
(c) acceptance /evaluation criteria; T-1223.1 Sensor Frequency. Selection of sensor
(d) standard report requirements; frequency shall be based on consideration of background
noise, acoustic attenuation, and vessel configuration.
(e) content of records and record retention.
Frequencies in the range of 100 kHz400 kHz have been
When this Article is specified by a referencing Code
shown to be effective. (See Nonmandatory Appendix B.)
Section, the AE method described in the Article shall
be used together with Article 1, General Requirements. T-1223.2 Sensor Mounting. The location and spacing
Definitions of terms used in this Article may be found of the sensors are referenced in T-1243. The sensors
in Mandatory Appendix III of this Article. shall be acoustically coupled using couplant specified
in the written procedure. Suitable adhesive systems are
those whose bonding and acoustic coupling effectiveness
T-1220 GENERAL have been demonstrated.
When examining austenitic stainless steels, titanium,
T-1220.1 The principal objectives of AE examination or nickel alloys, the need to restrict chloride /fluoride
are to locate and monitor emission sources caused by ion content, total chlorine /fluorine content, and sulfur
surface and internal discontinuities in the vessel wall, content in the couplant or other materials used on the
welds, and fabricated parts and components. vessel surface shall be considered and limits agreed
T-1220.2 All relevant indications caused by AE upon between contracting parties.
sources shall be evaluated by other methods of nonde- The sensor shall be held in place utilizing methods
structive examination. of attachment, as specified in the written procedure.
The signal cable and preamplifier must be supported.
T-1223.3 Surface Contact. Sensors shall be mounted
T-1221 Vessel Stressing directly on the vessel surface, or on integral waveguides.
Arrangements shall be made to stress the vessel
using internal pressure as specified by the referencing
T-1224 Location of Acoustic Emission Sources
Code Section. The rate of application of pressure shall
be specified in the examination procedure and the T-1224.1 Sources shall be located to the specified
pressurizing rate shall be sufficient to expedite the accuracy by multichannel source location, zone location,
examination with minimum extraneous noise. Provisions or both, as required by the referencing Code Section.

222

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
T-1224.1 ARTICLE 12 ACOUSTIC EMISSION EXAMINATION T-1243.2

All hits detected by the instrument shall be recorded ously monitored to an accuracy of 2% of the maximum
and used for evaluation. test pressure.
(1) Analog type indicating pressure gages used in
T-1224.2 Multichannel source location accuracy shall
testing shall be graduated over a range not less than
be within a maximum of 2 component wall thicknesses
112 times nor more than 4 times the test pressure.
or 5% of the sensor spacing distance, whichever is
(2) Digital type pressure gages may be used with-
greater.
out range restriction provided the combined error due
to calibration and readability does not exceed 1% of
01 T-1225 Procedure Requirements the test pressure.

Acoustic emission examination shall be performed


in accordance with a written procedure. Each procedure T-1240 APPLICATION REQUIREMENTS
shall include at least the following information, as
applicable: T-1241 Equipment
(a) material and configurations to be examined, in- (See T-1230 and Mandatory Appendix I.)
cluding dimensions and product form;
(b) background noise measurements; T-1242 System Calibration
(c) sensor type, frequency, and Manufacturer;
(d) method of sensor attachment (See Mandatory Appendix II.)
(e) couplant;
(f) acoustic emission instrument type and filter fre- T-1243 Pre-Examination Measurements
quency;
T-1243.1 On-Site System Calibration. Prior to each
(g) sensor locations;
vessel test or series of tests, the performance of each
(h) method for selection of sensor locations;
utilized channel of the AE instrument shall be checked
(i) description of system calibration(s); by inserting a simulated AE signal at each main ampli-
(j) data to be recorded and method of recording; fier input.
(k) post-examination vessel cleaning; A series of tests is that group of tests using the
(l) report requirements; and same examination system which is conducted at the
(m) qualification /certification of the examiner(s). same site within a period not exceeding 8 hr or the
test duration, whichever is greater.
This device shall input a sinusoidal burst-type signal
T-1230 EQUIPMENT AND SUPPLIES of measurable amplitude, duration, and carrier fre-
(a) The AE system consists of sensors, signal pro- quency. As a minimum, on-site system calibration shall
cessing, display, and recording equipment (see Appen- be able to verify system operation for threshold, counts,
dix I). MARSE, and peak amplitude. Calibration values shall
(b) Data measurement and recording instrumentation be within the range of values specified in Appendix I.
shall be capable of measuring the following parameters T-1243.2 Attenuation Characterization. An attenua-
from each AE hit on each channel: counts above system tion study is performed in order to determine sensor
examination threshold, peak amplitude, arrival time, spacing. This study is performed with the test fluid in
and Measured Area of the Rectified Signal Envelope the vessel using a simulated AE source. For production
(MARSE). Mixing or otherwise combining the acoustic line testing of identical vessels see Nonmandatory Ap-
emission signals of different sensors in a common pendix B.
preamplifier is not permitted except to overcome the The typical signal propagation losses shall be deter-
effects of local shielding. (See Nonmandatory Appendix mined according to the following procedure: select a
B.) The data acquisition system shall have sufficient representative region of the vessel away from manways,
channels to provide the sensor coverage defined in T- nozzles, etc., mount a sensor, and strike a line out
1243.4. Amplitude distribution, by channel, is required from the sensor at a distance of 10 ft (3 m) if possible.
for source characterization. The instrumentation shall Break 0.3 mm (2H) leads next to the sensor and then
be capable of recording the measured acoustic emission at a 2 ft (0.6 m) interval along this line. The breaks
data by hit and channel number. shall be done with the lead at an angle of approximately
(c) Time and pressure shall be measured and recorded 30 deg. to the surface and with a 0.1 in. (2.5 mm)
as part of the AE data. The pressure shall be continu- lead extension.

223

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
T-1243.3 2001 SECTION V T-1244.3.2

T-1243.3 Sensor Location. Sensor locations on the T-1244.2 Background Noise. Extraneous noise must
vessel shall be determined by the vessel configuration be identified, minimized, and recorded.
and the maximum sensor spacing (see T-1243.4). A
further consideration in locating sensors is the need to T-1244.2.1 Background Noise Check Prior to
detect structural flaws at critical sections, e.g., welds, Loading. Acoustic emission monitoring of the vessel
high stress areas, geometric discontinuities, nozzles, during intended examination conditions is required to
manways, repaired regions, support rings, and visible identify and determine the level of spurious signals
flaws. Additional consideration should be given to the following the completion of the system performance
possible attenuation effects of welds. See Nonmandatory check and prior to stressing the vessel. A recommended
Appendix B. Sensor location guidelines for zone location monitoring period is 15 min. If background noise is
for typical vessel types are given in Nonmandatory above the evaluation threshold, the source of the noise
Appendix A. shall be eliminated or the examination terminated.

T-1243.4 Sensor Spacing T-1244.2.2 Background Noise During Examina-


tion. In the AE examiners analysis of examination
T-1243.4.1 Sensor Spacing for Zone Location.
results, background noise shall be noted and its effects
Sensors shall be located such that a lead break at any
on test results evaluated. Sources of background noise
location in the examination area is detected by at least
include:
one sensor and have a measured amplitude not less
than as specified by the referencing Code Section. The (1) liquid splashing into a vessel;
maximum sensor spacing shall be no greater than 112 (2) a pressurizing rate that is too high;
times the threshold distance. The threshold distance is (3) pumps, motors, and other mechanical devices;
defined as the distance from a sensor at which a pencil- (4) electromagnetic interference; and
lead break on the vessel has a measured amplitude (5) environment (rain, wind, etc.).
value equal to the evaluation threshold. Leaks from the vessel such as valves, flanges, and
safety relief devices can mask AE signals from the
T-1243.4.2 Sensor Spacing for Multichannel structure. Leaks must be eliminated prior to continuing
Source Location Algorithms. Sensors shall be located the examination.
such that a lead break at any location in the examination
area is detected by at least the minimum number of T-1244.3 Vessel Pressurization
sensors required for the algorithms. T-1244.3.1 Rates of pressurization, pressurizing
medium, and safety release devices shall be as specified
T-1243.5 Systems Performance Check. A verifica-
by the referencing Code Section. The pressurization
tion of sensor coupling and circuit continuity shall be
should be done at a rate that will expedite the test
performed following sensor mounting and system
with a minimum of extraneous noise.
hookup and again immediately following the test. The
peak amplitude response of each sensor to a repeatable T-1244.3.2 Pressurization Sequence. The exami-
simulated acoustic emission source at a specific distance nation shall be done in accordance with the referencing
from each sensor should be taken prior to and after Code Section. Pressure increments shall generally be
the test. The measured peak amplitude should not vary to 50%, 65%, 85%, and 100% of maximum test pressure.
more than 4 dB from the average of all the sensors. Hold periods for each increment shall be 10 min and
Any channel failing this check should be investigated for the final hold period shall be at least 30 min. (See
and replaced or repaired as necessary. If during any Fig. T-1244.3.2.) Normally, the pressure test will cause
check it is determined that the testing equipment is local yielding in regions of high secondary stress. Such
not functioning properly, all of the product that has local yielding is accompanied by acoustic emission
been tested since the last valid system performance which does not necessarily indicate discontinuities. Be-
check shall be re-examined. cause of this, only large amplitude hits and hold period
data are considered during the first loading of vessels
without post-weld heat treatment (stress relief). If the
T-1244 Examination Procedure
first loading data indicates a possible discontinuity or
T-1244.1 General Guidelines. The vessel is sub- is inconclusive, the vessel shall be repressurized from
jected to programmed increasing stress levels to a 50% to 100% of the test pressure with intermediate
predetermined maximum while being monitored by load holds at 50%, 65%, and 85%. Hold periods for
sensors that detect acoustic emission caused by growing the second pressurization shall be the same as for the
structural discontinuities. original pressurization.

224

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
T-1244.3.3 ARTICLE 12 ACOUSTIC EMISSION EXAMINATION T-1291

FIG. T-1244.3.2 AN EXAMPLE OF PRESSURE VESSEL TEST STRESSING SEQUENCE

T-1244.3.3 Test Termination. Departure from a indications. These criteria are based on a specific set
linear count or MARSE vs. load relationship should of AE monitoring conditions. The criteria to be used
signal caution. If the AE count or MARSE rate increases shall be as specified in the referencing Code Section.
rapidly with load, the vessel shall be unloaded and
either the test terminated or the source of the emission
determined and the safety of continued testing evaluated. T-1290 DOCUMENTATION
A rapidly (exponentially) increasing count or MARSE
rate may indicate uncontrolled, continuing damage indic- T-1291 Written Report
ative of impending failure. The report shall include the following:
(a) complete identification of the vessel, including
material type, method of fabrication, Manufacturers
T-1260 CALIBRATION name, and certificate number;
(See Mandatory Appendix II.) (b) vessel sketch of Manufacturers drawing with
dimensions and sensor locations;
(c) test fluid employed;
T-1280 EVALUATION (d) test fluid temperature;
(e) test sequence load rate, hold times, and hold
T-1281 Evaluation Criteria levels;
The AE criteria shown in Table T-1281 are set forth (f) attenuation characterization and results;
as one basis for assessing the significance of AE (g) record of system performance verifications;

225

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
Table T-1281

Licensed by Information Handling Services


TABLE T-1281
AN EXAMPLE OF EVALUATION CRITERIA FOR ZONE LOCATION
Evaluation
Emission During Load Threshold,

COPYRIGHT American Society of Mechanical Engineers


Hold Count Rate Number of Hits Large Amplitude Hits MARSE or Amplitude Activity dB

(First Loading) Not more than EH hits Not applied Not applied Not more than EA hits MARSE or amplitudes Activity does not increase VTH
Pressure vessels beyond time TH above a specified do not increase with with increasing load.
without full amplitude increasing load.
postweld heat

226
treatment
Pressure vessels other Not more than EH hits Less than NT counts Not more than ET hits Not more than EA hits MARSE or amplitudes Activity does not increase VTH
2001 SECTION V

than those covered beyond time TH per sensor for a above a specified above a specified do not increase with with increasing load.
above specified load amplitude amplitude increasing load.
increase

GENERAL NOTES:
(a) EH , NT , ET , and EA are specified acceptance criteria values specified by the referencing Code Section.
(b) VTH is the specified evaluation threshold.
(c) TH is the specified hold time.
T-1291 ARTICLE 12 ACOUSTIC EMISSION EXAMINATION T-1292

(h) correlation of test data with the acceptance cri-


teria;
(i) a sketch or Manufacturers drawings showing the
location of any zone not meeting the evaluation criteria;
(j) any unusual effects or observations during or
prior to the test;
(k) date(s) of test(s);
(l) name(s) and qualifications of the test opera-
tor(s); and
(m) complete description of AE instrumentation in-
cluding Manufacturers name, model number, sensor
type, instrument settings, calibration data, etc.

T-1292 Record
(a) A permanent record AE data includes:
(1) AE hits above threshold vs time and /or pres-
sure for zones of interest;
(2) total counts vs time and /or pressure; and
(3) written reports.
(b) The AE data shall be maintained with the records
of the vessel.

227

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
ARTICLE 12
MANDATORY APPENDICES

APPENDIX I I-1240 PREAMPLIFIER


INSTRUMENTATION PERFORMANCE The preamplifier shall be mounted in the vicinity of
REQUIREMENTS the sensor, or in the sensor housing. If the preamplifier
I-1210 ACOUSTIC EMISSION SENSORS is of differential design, a minimum of 40 dB of
common-mode noise rejection shall be provided. Fre-
I-1210.1 General. Acoustic emission sensors in the quency response shall not vary more than 3 dB over
range of 100 kHz400 kHz shall be temperature-stable the operating frequency and temperature range of the
over the range of intended use, and shall not exhibit sensors.
sensitivity changes greater than 3 dB over this range
as guaranteed by the Manufacturer. Sensors shall be
shielded against radio frequency and electromagnetic
noise interference through proper shielding practice I-1250 FILTER
and /or differential (anticoincident) element design. Sen-
Filters shall be of the band pass or high pass type
sors shall have a frequency response with variations
and shall provide a minimum of 24 dB /octave signal
not exceeding 4 dB from the peak response.
attenuation. Filters shall be located in preamplifier.
I-1210.2 Sensor Characteristics. Sensors shall have Additional filters shall be incorporated into the proces-
a resonant response between 100 kHz400 kHz. Mini- sor. Filters shall insure that the principal processing
mum sensitivity shall be 80 dB referred to 1 frequency corresponds to the specified sensor frequency.
volt /microbar, determined by face-to-face ultrasonic
test.
I-1260 POWER-SIGNAL CABLE
NOTE: This method measures relative sensitivity of the sensor.
Acoustic emission sensors used in the same test should not vary in The cable providing power to the preamplifier and
peak sensitivity more than 3 dB from the average.
conducting the amplified signal to the main processor
shall be shielded against electromagnetic noise. Signal
loss shall be less than 1 dB per 100 ft (30.5 m) of
cable length. The recommended maximum cable length
I-1220 SIGNAL CABLE is 500 ft (152 m) to avoid excessive signal attenuation.
The signal cable from sensor to preamplifier shall
not exceed 6 ft (1.8 m) in length and shall be shielded
against electromagnetic interference. I-1270 POWER SUPPLY
A stable grounded electrical power supply, meeting
the specifications of the instrumentation, shall be used.

I-1230 COUPLANT
Couplant selection shall provide consistent coupling
I-1280 MAIN AMPLIFIER
efficiency during a test. Consideration should be given
to testing time and the surface temperature of the The gain in the main amplifier shall be linear within
vessel. The couplant and method of sensor attachment 3 dB over the temperature range of 40F to 125F
shall be specified in the written procedure. (4C to 52C).

228

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
I-1290 ARTICLE 12 MANDATORY APPENDICES III-1210

I-1290 MAIN PROCESSOR APPENDIX II


I-1291 General INSTRUMENT CALIBRATION AND
CROSS-REFERENCING
The main processor(s) shall have processing circuits
through which sensor data will be processed. It shall II-1210 MANUFACTURERS CALIBRATION
be capable of processing hits, counts, peak amplitudes, Acoustic emission system components will be pro-
and MARSE on each channel. vided from the Manufacturer with certification of per-
(a) Threshold. The AE instrument used for examina- formance specifications and tolerances.
tion shall have a threshold control accurate to within
1 dB over its useful range. II-1211 Annual Calibration
(b) Counts. The AE counter circuit used for examina-
tion shall detect counts over a set threshold within an The instrument shall have an annual comprehensive
calibration following the guidelines provided by the
accuracy of 5%.
Manufacturer using calibration instrumentation meeting
(c) Hits. The AE instrument used for examination the requirements of a recognized national standard.
shall be capable of measuring, recording, and displaying
a minimum of 20 hits /sec total for all channels for a
minimum period of 10 sec and continuously measuring,
II-1220 INSTRUMENT CROSS-
recording, and displaying a minimum of 10 hits /sec REFERENCING
total for all channels. The system shall display a warning
if there is greater than a 5 sec lag between recording The performance and threshold definitions vary for
and display during high data rates. different types of AE instrumentation. Parameters such
as counts, amplitude, energy, etc., vary from Manufac-
(d) Peak Amplitude. The AE circuit used for examina-
turer to Manufacturer and from model to model by the
tion shall measure the peak amplitude with an accuracy
same Manufacturer. This section of appendix describes
of 2 dB.
techniques for generating common baseline levels for
(e) Energy. The AE circuit used for examination the different types of instrumentation.
shall measure MARSE with an accuracy of 5%. The The procedures are intended for baseline instrument
usable dynamic range for energy shall be a minimum calibration at 60F to 80F (16C to 27C). For field
of 40 dB. use, small portable signal generators and calibration
(f) Parametric Voltage. If parametric voltage is mea- transducers can be carried with the equipment and
sured by the AE instrument, it should measure to an used for periodic checking of sensor, preamplifier, and
accuracy of 2% of full scale. channel sensitivity.

II-1221 Sensor Characterization


Threshold of acoustic emission detectability is an
amplitude value. All sensors shall be furnished with
documented performance data. Such data shall be trace-
able to NBS standards. A technique for measuring
threshold of detectability is described in Article XI,
I-1292 Peak Amplitude Detection Appendix II.
Comparative calibration must be established per the
requirements of Appendix II. Usable dynamic range
shall be a minimum of 60 dB with 1 dB resolution APPENDIX III GLOSSARY OF
over the frequency band width of 100 kHz to 400 TERMS FOR ACOUSTIC EMISSION
kHz, and the temperature range of 40F to 125F (4C EXAMINATION OF METAL
to 52C). Not more than 2 dB variation in peak detection PRESSURE VESSELS
accuracy shall be allowed over the stated temperature
range. Amplitude values shall be stated in dB, and III-1210 SCOPE
must be referenced to a fixed gain output of the system This Mandatory Appendix is used for the purpose
(sensor or preamplifier). of establishing standard terms and definitions of terms

229

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
III-1210 2001 SECTION V III-1230

relating to metal pressure vessel examination with acous- dB scale a relative logarithmic scale of signal
tic emission. amplitude defined by dB V p 20 log Vin /Vout. The
reference voltage is defined as 1 volt out of the sensor
and V is measured amplitude in volts.
electronic waveform generator a device which
can repeatably induce a transient signal into an acoustic
III-1220 GENERAL REQUIREMENTS emission processor for the purpose of checking, veri-
(a) The Standard Terminology for Nondestructive fying, and calibrating the instrument
Examinations (ASTM E 1316) has been adopted by measured area of the rectified signal envelope a
the Committee as SE-1316. measurement of the area under the envelope of the
(b) SE-1316 defines the terms that are used in con- rectified linear voltage time signal from the sensor
multi-channel source location a source location
junction with this Article.
technique which relies on stress waves from a single
(c) For general terms, such as Interpretation, Flaw,
source producing hits at more than one sensor. Position
Discontinuity, Evaluation, etc., refer to Article 1, Man-
of the source is determined by mathematical algorithms
datory Appendix I. using difference in time of arrival
(d) In addition to those terms listed in SE-1316, the simulated AE source a device which can repeatedly
terms listed in II-1230 are also applicable. induce a transient elastic stress wave into the structure
threshold of detectability a peak amplitude mea-
surement used for cross calibration of instrumentation
from different vendors
zone the area surrounding a sensor from which
III-1230 REQUIREMENTS
AE sources can be detected
The following Code terms are used in conjunction zone location a method of locating the approximate
with this Article: source of emission

230

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
ARTICLE 12
NONMANDATORY APPENDICES

Appendix A begins on the next page.

231

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
Case 1 2001 SECTION V

232

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
ARTICLE 12 NONMANDATORY APPENDICES Case 2

233

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
Case 3 2001 SECTION V

234

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
ARTICLE 12 NONMANDATORY APPENDICES Case 4

235

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
Case 5 2001 SECTION V

236

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services
B ARTICLE 12 NONMANDATORY APPENDICES B-40

APPENDIX B SUPPLEMENTAL example of this is the use of several sensors (with


INFORMATION FOR CONDUCTING preamplifiers around a manway or nozzle).
ACOUSTIC EMISSION
EXAMINATIONS
B-10 FREQUENCY SELECTION
The frequency band of 100 kHz200 kHz is the
lowest frequency band that should be considered for
general AE pressure vessel examination. Higher fre- B-30 ATTENUATIVE WELDS
quency bands may be considered if background noise
cannot be eliminated. If a higher frequency band is Some have been shown to be highly attenuative to
used the following items must be considered. non-surface waves. This situation predominantly affects
(a) Attenuation characteristics will change. multichannel source location algorithms. This situation
(b) Sensor spacings will decrease and more sensors can be identified by modifying the attenuation character-
will be required to adequately cover the evaluation area. ization procedure to produce a stress wave which does
not contain surface waves traveling across the weld.
(c) Instrumentation performance requirements de-
scribed in Appendix I must be adjusted to the higher
frequency band.
(d) Instrumentation calibration described in Appendix
II must be performed at the higher frequency band.
(e) Alternate evaluation /acceptance criteria must be
obtained from the referencing Code Section.
B-40 PRODUCTION LINE TESTING OF
IDENTICAL VESSELS
B-20 COMBINING MORE THAN ONE
For situations which involve repeated tests of identical
SENSOR IN A SINGLE CHANNEL
vessels where there is no change in the essential
Two or more sensors (with preamplifiers) may be variables such as material, thickness, product form and
plugged into a single channel to overcome the effects type, the requirement for attenuation characterization
of local shielding in a region of the vessel. One specific on each vessel is waived.

237

COPYRIGHT American Society of Mechanical Engineers


Licensed by Information Handling Services

Vous aimerez peut-être aussi