Académique Documents
Professionnel Documents
Culture Documents
Model FT2A
105397
Rev. C
Model FT2A
Notice
FOX THERMAL INSTRUMENTS
This publication must be read in its entirety before performing any operation. Failure
to understand and follow these instructions could result in serious personal injury
and/or damage to the equipment. Should this equipment require repair or adjustment
beyond the procedures given herein, contact the factory at:
Fox Thermal Instruments believes that the information provided herein is accurate;
however, be advised that the information contained herein is NOT a guarantee
for satisfactory results. Specifically, this information is neither a warranty nor
guarantee, expressed or implied, regarding performance, merchantability, fitness,
or any other matter with respect to the products; nor recommendation for the use
of the product/process information in conflict with any patent. Please note that Fox
Thermal Instruments, Inc. reserves the right to change and/or improve the product
design and specification without notice.
F O X I S I S O 9 0 0 1 C E R T I F I E D
2
Model FT2A
Table Of Contents
1. Introduction Page 4
TABLE OF CONTENTS
3. Wiring (Electrical) Page 14
a. Local Wiring Page 17
b. Input Power Page 17
c. Signal Wiring Page 19
d. Alarm Wiring Page 21
e. Remote Switch Page 23
f. Remote Wiring Page 24
5. Maintenance Page 58
a. Troubleshooting Page 60
7. Appendices Page 66
a. Specifications Page 66
b. Agency Approvals Page 69
c. FT2A with 2 Gas Curves Page 70
d. Dimensions Page 74
e. Mounting the FT2A Enclosure Page 82
f. Warranty Page 84
g. Returning your meter Page 86
9. Index Page 90
3
Model FT2A
Introduction
Scope
Welcome Thank you for purchasing the Model FT2A Thermal Gas Mass Flow Meter and
Temperature Transmitter from Fox Thermal Instruments. The Model FT2A is
one of the most technically advanced flow meters in the world. Extensive
engineering effort has been invested to deliver advanced features, accuracy
measurement performance and outstanding reliability.
4
Model FT2A
Introduction
INTRODUCTION
Calibration files include details on process conditions, calibration fluid, line size
and other information. All NIST-traceable equipment utilized for the calibration
procedure is identified, as is the calibration history of all reference equipment.
The first 4 to 20mA output is for flow rate. The second 4 to 20mA output
can be configured either for flow rate or process gas temperature. Both 4 to
20mA outputs can be scaled by the user. The digital output can be used for
frequency or alarm, is programmable to represent flow rate and can be scaled
for maximum flow/maximum frequency, units-per-pulse or pulse-per-units.
The maximum frequency is 100 Hz.
FT2A View interfaces to the USB port and is a free Fox PC-based
software program that displays flow meter readings and permits flow meter
configuration. The software is available for download on the Fox website.
5
Model FT2A
Introduction
Standard I/O
24VDC Input Power
4 to 20 mA Flow
100-240VAC Optional
4 to 20 mA Flow or Temperature
Frequency or Alarm
Contact Input
Display and
Configuration Panel
Standard Digital Communications
USB
MODEL FT2A (Free FT2A View Software)
Gas Mass Flowmeter & Temperature Transmitter
6
Model FT2A
Installation: General
INSTALLATION
2. Sensor installation depth
3. Sensor orientation in relation to sensor length and direction of flow
4. Proper tightening of compression fitting for mounting meter
General Precautions
The following general precautions should be observed:
1. Exercise care when handling the flow meter to avoid damaging the probe,
sensor or enclosure.
2. The enclosure covers must be closed except during installation.
3. Mounting FT2A in direct sunlight can cause the temperature inside the
enclosure to increase beyond design limits, resulting in failure of LCD
display and reduced component life. It is recommended that a sunshade
be installed to avoid direct sunlight.
4. Ensure the flow direction arrow points in the direction of flow.
5. Do not install the FT2A enclosure near an igniter, igniter-controller or
switching equipment.
6. Do not install an external power supply in a cabinet containing an
igniter controller or switching equipment.
7. Ensure that good engineering practices and applicable industry codes are
followed throughout the installation process.
8. For accurate flow measurement: review flow meter placement instructions
before installation to ensure a proper flow profile in the pipe.
7
Model FT2A
Installation: Insertion Type
Insertion Flow Meter Instructions for Insertion Flow Meter Lateral Placement
Lateral Placement Install the Model FT2A Insertion style flow meter so that it is far enough away
from bends in the pipe, obstructions, or changes in line sizes to ensure a
consistent flow profile. Fifteen diameters of straight pipe upstream and ten
downstream are recommended.
For example, a 2" pipe would require 30" upstream and 20" downstream, but a
INSTALLATION
Fig. 2.1: Upstream and Downstream Pipe IDs for Insertion Meters
Proper
Profile
FLOW
Irregular
Profile
8
Model FT2A
INSTALLATION
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
Compression Fitting
Half Coupling, (Supplied by Fox)
3/4" NPT Female
(Supplied by Probe Diameter, 1/2
Customer)
Customers Pipe
CL
.87" (22.098 mm)
9
Model FT2A
Installation: Insertion Type
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
Pipe
FLOW
i
Note: Some flow meters are shipped with the sensor elements that are offset (see
figure 2.4). Others are shipped with sensors that have equal length elements (see
figure 2.5). The sensor type supplied was selected at the factory to be the best suited
for your application. Follow the appropriate sensor orientation instructions.
+10
FLOW
-10
FLOW
10
Model FT2A
Installation: Insertion Type
INSTALLATION
+10
FLOW
-10
FLOW
11
Model FT2A
Installation: Inline Type
For example, a 2" pipe would require 16" upstream from the edge of the flow
body and 8" downstream from the other end of the flow body, whereas a 4"
pipe would require 32" upstream and 16" downstream.
The Model FT2A is welded, threaded or flanged to the customers pipe. Care
should be taken to ensure that the diameter of the mating pipe is the same
diameter as the Model FT2A flow body or errors in flow readings can occur.
The installation procedure should be a combination of the end users best
engineering practices, in compliance with local codes, and the manufacturers
recommendations.
See Figure 2.7 for a detailed look at upstream and downstream pipe diameters
for inline meters.
Fig. 2.7: Upstream and Downstream Pipe IDs for Inline Meters
Proper
Profile
FLOW
8X Pipe ID 4X Pipe ID
Irregular
Profile
12
Model FT2A
INSTALLATION
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
Flow Body
FLOW
13
Model FT2A
Wiring: General
Grounding Grounding
The enclosure must be properly grounded with a quality earth ground. 16 gauge,
stranded wire is recommended.
14
Model FT2A
Wiring: General
WIRING
Wiring
Fox FT2A Modbus / BACnet MS/TP Manual
Fox FT2A Profibus, DeviceNet, Ethernet Manual
Five wire shielded cable required, the recommended wire gauge is 18 AWG.
Run the shielded cable through a separate grounded steel conduit to the
bottom conduit opening of the electronics enclosure (no other cables or wires
in the conduit). If you are using your own cable, make sure that the cable
length does not exceed 100 feet and the wire resistance does not exceed one
ohm. Do not connect the cable shield at the electronics enclosure end.
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
15
Model FT2A
Wiring: General
TS7 1 2 3 1 2 3 TS1
AC POWER DC POWER
TS2 TS3
1 1 ETHERNET
WIRING
2 2
3 3
I/O WIRING 4 4
AND 5 5
OPTIONAL
MODBUS 6 6
COMMUNICATION 7 7
J1
8 8
9 OPTIONAL
10 COMMUNICATION
(DEVICENET
11
PROFIBUS)
TS8
1
2
REMOTE
3 USB
SENSOR
COMMUNICATION
4 WIRING
5 J4
FT2A Enclosure
i Note:
Serial numbers: If you have more than one meter, you must ensure
that the serial numbers of the probe/J-Box, remote electronics,
housing, and/or flow body match one another. These items have
been manufactured and calibrated to operate as a unit and cannot be
mismatched.
16
Model FT2A
The enclosure must be properly grounded with a quality earth ground. Sixteen
(16) gauge, stranded wire, is recommended for power and earth ground.
WIRING
24V Return
Earth Ground
TS7 1 2 3 1 2 3 TS1
(+) (-)
TS2
1
2
3
4
5
6
7
8
9
10
11
TS8
1
2
3
4
5
Caution:
Supply connection wiring must be rated for at least 90C.
17
Model FT2A
Wiring: Input Power
The enclosure must be properly grounded with a quality earth ground. Sixteen
(16) gauge, stranded wire, is recommended for power and earth ground.
Earth Ground
WIRING
AC (L)
AC (N)
TS7 1 2 3 1 2 3 TS1
(N) (L)
TS2
1
2
3
4
5
6
7
8
9
10
11
TS8
1
2
3
4
5
Caution:
Supply connection wiring must be rated for at least 90C.
18
Model FT2A
Wiring: Signal Wiring
TS7 1 2 3 1 2 3 TS1
FT2A Customer PLC or DCS
WIRING
+24VDC
TS2
1 4 to 20mA FLOW RATE
2 * (see important note below)
3
4
24VDC Return
5
6
+24VDC
7
8
4 to 20mA TEMPERATURE OR
FLOW RATE
9 * (see important note below)
10
11 24VDC Return
TS8
1
2
3
4
5
i Important Note
The load resistor on the Fox Flow Meter 4 to 20mA signal is typically
250 ohms and is located in or at the customers PLC or DCS. A 250
ohm resistor in the 4 to 20mA line will result in a 1 to 5VDC signal to the
PLC or DCS. Some PLC/DCS equipment has the load resistor built in to
the unit; please refer to the PLC/DCS technical manual. Do not exceed a
600 ohm load on the Fox Flow Meter 4 to 20mA signal.
19
Model FT2A
Fig. 3.6: 4 to 20mA Output Wiring for Loop Power Provided by FT2A
24V
1 2 3 1 2 3
+ -
TS2
1
+
2
3 4 to 20mA FLOW RATE
4 * (see important note below)
5 -
6 +
7 4 to 20mA TEMPERATURE
8 OR FLOW RATE
9 - * (see important note below)
10
11
TS8
1
2
3
4
5
i Important Note:
The load resistor on the Fox Flow Meter 4 to 20mA signal is typically
250 ohms and is located in or at the customers PLC or DCS. A 250
ohm resistor in the 4 to 20mA line will result in a 1 to 5VDC signal to the
PLC or DCS. Some PLC/DCS equipment has the load resistor built in to
the unit; please refer to the PLC/DCS technical manual. Do not exceed a
600 ohm load on the Fox Flow Meter 4 to 20mA signal.
20
Model FT2A
Wiring: Frequency/Alarm Wiring
WIRING
FT2A Customer PLC or DCS
TS7 1 2 3 1 2 3 TS1
TS2
1 +24VDC
2
3
2.4K to 10K OHM
4
5
Frequency or Alarm Output
6
7
8 24VDC Return
9
10
11
TS8
1
2
3
4
5
21
Model FT2A
Wiring: Frequency/Alarm Wiring
TS2
1
2
2.4K to 10K OHM
3
4
5
WIRING
TS8
1
2
3
4
5
i Note:
The FT2A Frequency/Alarm output is typically used to drive digital circuitry or solid-
state relays. The output of a solid state relay may, in turn, operate loads such as
electromechanical relays or alarm indicators.
22
Model FT2A
When the 2 gas curve option is ordered, the switch can be used to switch
between curves.
WIRING
TS7 1 2 3 1 2 3 TS1
TS2
1
FT2A Customer PLC
2 or DCS
3
4
5
6
7
8
EXTERNAL
9 SWITCH
10
11
TS8
1
2
3
4
5
23
Model FT2A
i Note: Remote wiring is only required when the Remote Electronics option is
provided.
Five wire 18 AWG shielded cable required. Run the cable through a separate
grounded steel conduit (no other cables or wires in the conduit). Make sure
that the cable length does not exceed 100 feet and the wire resistance does
not exceed one ohm.
WIRING
The extension cable connects the terminals of the remote probe enclosure to
connector TS8 of the electronics enclosure as shown in Figure 3.11 and Table
3.1 (p. 25).
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
i Installation wiring: Keep the wires inside the FT2A enclosure short.
Obtain the correct length for the FT2A remote wires using one of these
methods:
Trim the wires to extend 5 inches out of the enclosure after the
conduit and wires are routed to the FT2A.
Trim the wires to extend 6 inches from the end of the conduit before
it is attached to the FT2A
24
Model FT2A
TS7 1 2 3 1 2 3 TS1
TS2
1
2
3
4
5
6
WIRING
7
8
9
10 Shield
#4 BLK
11 #2
RED*
RED YEL RED
BLK*
RED
BRN* BRN RED
YEL #3 #1
WHT*
WHT
GRN*
WHT
Sensor Wires
Remote Enclosure
*Wire colors listed here represent the wire colors of cables supplied by
Fox. Colors may vary if customer is supplying their own cable.
25
Model FT2A
Operation: Start Up
i Note: Fox also offers FT2A View, a Windows-based PC program, that can
be used to program the meter.
Display Window
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
26
Model FT2A
Operation: Display Screens
OPERATION
to Display #1. Pressing the F3 and F4 keys at the same time enters the Reset
Total screen (see p. 46).
F1 F2
(Fixed screen)
Display #3 Alarm = None
Elp = 14.6 HR
F1 F2
27
Model FT2A
Operation: Engineering Displays
BrShtDnCnt=0 cnt
Bridge shutdown detection count Display 25
28
Model FT2A
Operation: Programming
VALUE = 0.91234
CHG OK
OPERATION
F1 F2 F3 F4
Press CHG (F1) key to change the value, OK (F4) to accept the value.
VALUE = 0.91234
UP DN NXT OK
F1 F2 F3 F4
Press the UP (F1) or DN (F2) key to select a new digit or character, the cursor
points to the selected digit. Press NXT (F3) to select the next digit and OK
(F4) to accept the entry.
F1 F2 F3 F4
Press NXT (F1) key repeatedly until the correct selection is made and OK (F4)
key to accept the entry.
SET PARAMETERS ?
No Yes
F1 F2 F3 F4
Press YES (F4) and the following screen will prompt user to enter password:
29
Model FT2A
Operation: Programming
F1 F2 F3 F4
Enter the correct password, then follow the instructions for changing a value
as specified on page p. 29. The default Level 1 password is 1234.
OPERATION
If the wrong password is entered, the message Wrong Password will display
and then return to the programming entry screen.
If the password is accepted, the base programming screen will be shown:
SET PARAMETERS
I/O FLO DSP EXIT
F1 F2 F3 F4
Press EXIT (F4) repeatedly until Normal Mode is seen briefly to exit the
programming mode.
From the base screen, press I/O (F1) and then in the next screen press 420
(F3).
SET I/O
COM I/O 420 EXIT
F1 F2 F3 F4
SET 4-20 mA
CH1 CH2 EXIT
F1 F2 F3 F4
30
Model FT2A
Operation: Programming
F1 F2 F3 F4
Enter the value for the 20mA and press OK (F4) key to accept the setting.
OPERATION
4 mA = 0 SCFM
CHG OK
F1 F2 F3 F4
mA Fault = 3.6 mA
NXT OK
F1 F2 F3 F4
This menu allows the user to select an alarm level on the 4 to 20mA when a
serious issue is detected that is preventing the calculation of a correct flow
value.
The options are:
Force the 4 to 20mA signal to 3.6mA
Force the 4 to 20mA signal to 21mA
Do not force 4 to 20mA signal (not used)
Press (F4) repeatedly until Normal Mode is seen briefly to exit the
programming mode.
i Note: When the flow rate exceeds the programmed value for the 20mA set
point, the analog output will stay at 20mA and an alarm code will be generated.
31
Model FT2A
Operation: Programming
F1 F2 F3 F4
Press NXT (F1) to select Flow or Temperature and then press OK (F4).
OPERATION
20 mA = 300 F
CHG OK
F1 F2 F3 F4
Enter the value for the 20mA and press OK (F4) key to accept setting.
4 mA = 10 F
CHG OK
F1 F2 F3 F4
i Note: When the flow rate exceeds the programmed value for the 20mA set
point, the analog output will stay at 20mA and an alarm code will be generated.
SET I/O
INP OUT EXIT
F1 F2 F3 F4
32
Model FT2A
Operation: Programming
F1 F2 F3 F4
Press NEXT (F1) to cycle through output options until you have the selection
for "OUT=Frequency" and press OK (F4).
OPERATION
The frequency output can be configured in one of three ways:
(1) specifying a maximum frequency to a defined maximum value of flow
rate, (2) specifying how many flow units total per pulse, U/P (i.e., 0.1 SCF per
pulse) or (3) specifying how many pulses per unit, P/U (i.e., 10 pulses per
SCF). All of these approaches are equivalent.
FREQUENCY OUTPUT
P/U U/P FEQ EXIT
F1 F2 F3 F4
Use P/U (F1) to enter pulse per unit, U/P (F2) for Unit per pulse or FEQ (F3) to
enter the flow and maximum frequency to scale the frequency output.
i Note: When data is entered with any of the three described methods, the other
values will be re-calculated according to the settings.
PLS/UNT = 1.2
CHG OK
F1 F2 F3 F4
Press CHG (F1) to change the setting and then OK (F4) to accept entry.
The value entered is in pulse per selected flow unit total (i.e., 10 pulses per
SCF).
33
Model FT2A
Operation: Programming
F1 F2 F3 F4
OPERATION
Press CHG (F1) to change the setting and then OK (F4) to accept entry.
The value entered is in unit per pulse (i.e. 0.01 flow unit total per pulse)
Max Flow and Entering data with flow and maximum frequency:
Frequency From the Frequency Output Menu on p. 33, press FEQ (F3) and the
following screen will show:
MaxFreq=98.5 Hz
CHG OK
F1 F2 F3 F4
MaxFlo=4999.8 SCFM
CHG OK
F1 F2 F3 F4
i Note: When the flow rate exceeds the maximum frequency set point, the output
will stay at that maximum frequency but the FT2A will issue an alarm code.
Caution: Equation to ensure pulse rate must not exceed 100 Hz.
34
Model FT2A
Operation: Programming
F1 F2 F3 F4
OUT = HiFloAlm
NXT OK
OPERATION
F1 F2 F3 F4
Then press NXT (F1) to select the correct alarm and press OK (F4).
Selections are:
HiFloAlm = High Flow Alarm
LoFloAlm = Low Flow Alarm
HiTempAlm = High Temperature Alarm
LoTempAlm = Low Temperature Alarm
Not used
Frequency
HiFloAlm=500 SCFM
CHG OK
F1 F2 F3 F4
Enter the value for the limit by pressing CHG (F1) and then OK (F4).
F1 F2 F3 F4
Press NXT (F1) repeatedly until the correct selection is shown and then press
OK (F4) to accept the setting.
35
Model FT2A
Operation: Programming
SET I/O
COM I/O 420 EXIT
F1 F2 F3 F4
Comm=Modbus
NXT OK
F1 F2 F3 F4
i
Note: Any selection other than None requires the communication option for
the selected communication type. If enabling a communication option, see the
associated Fox Communication Manual for specific programming information:
FT2A Modbus & BACnet MS/TP manual or FT2A Profibus, DeviceNet &
Ethernet Manual.
36
Model FT2A
Operation: Programming
Programming Using are programmable to show the information that you prefer and is discussed in
the Local Display this section.
Display #1 Display #2
DSP1L1 DSP2L1
DSP1L2 DSP2L2
F1 F2 F3 F4 F1 F2 F3 F4
OPERATION
Selections are:
DSP1L1 Display 1, Line 1
DSP1L2 Display 1, Line 2
DSP2L1 Display 2, Line 1
DSP2L2 Display 2, Line 2
DISPLAY/PASSWORD
DSP PSW EXIT
F1 F2 F3 F4
DSP1L1 = Total
NXT OK
F1 F2 F3 F4
When the selection is correct, press OK (F4) to accept. The display will then
go through the same process for all 4 lines of the 2 programmable displays
(DSP1L1, DSP1L2, DSP2L1 and DSP2L2).
37
Model FT2A
Operation: Programming
Programming Using After the last line of display 2 is accepted, the display will show the following
the Local Display menu:
ALTERNATE = Off
NXT OK
F1 F2 F3 F4
This menu allows you to alternate between menu display 1 and 2 every few
OPERATION
seconds.
Selections are: On or Off
Press EXIT (F4) repeatedly until Normal Mode is seen briefly to exit the
programming mode.
Password Password
There are two user level passwords, only Level 1 is programmable and gives
access to all the normal settings. The second password is used to allow
access to calibration factors and should normally never be changed unless
advised by the Fox service department, or to set a new password in the event
that the user forgets the Level 1 password.
From the base programming menu press DSP (F3) to select the display menu:
F1 F2 F3 F4
38
Model FT2A
Operation: Programming
F1 F2 F3 F4
OPERATION
Press OK (F4) to accept new data and exit programming by pressing EXIT (F4)
These values will be set at Fox, using the Application Data Sheet values. If
the customer changes the application, these values can be changed to match
the new application. Check with Fox customer service before changing the
application gas.
SET PARAMETERS
I/O FLO DSP EXIT
F1 F2 F3 F4
FLOW PARAMETERS 1
DGN UNT PRM EXIT
F1 F2 F3 F4
39
Model FT2A
Operation: Programming
F1 F2 F3 F4
Press NXT (F1) to change selection and OK (F4) to accept.
OPERATION
i Note: The totalizer (total flow measured) will roll over when reaching a certain
value. The maximum value is dependent on the flow units selected.
WARNING:
The FT2A re-calculates area, 4 and 20mA values, maximum flow for the
frequency output and zero flow cutoff when changing flow units except
for velocity units. When going to or from velocity units, the FT2A will not
recalculate these values and these values must be re-entered manually.
After pressing OK (F4) to accept the Flow unit the display will prompt for the
temperature unit setting:
40
Model FT2A
Operation: Programming
F1 F2 F3 F4
OPERATION
Deg F
After pressing OK (F4) to accept the temperature unit setting, the display will
prompt for temperature reference in selected unit.
TmpRef = 60 F
CHG OK
F1 F2 F3 F4
F1 F2 F3 F4
Selections are:
mmHG Millimeters of mercury
Psia Pounds per square inch atmosphere
bara Bar atmosphere
After the pressure unit selection is made, the display will show a menu to enter
the pressure reference:
PresRef= 14.7
CHG OK
F1 F2 F3 F4
41
Model FT2A
Operation: Programming
F1 F2 F3 F4
i
Note: The density entry is only used when KG/time or LBS/time is selected for
flow rate units.
SET PARAMETERS
I/O FLO DSP EXIT
F1 F2 F3 F4
The menu is accessed from the base programming menu by pressing FLO
(F2):
FLOW PARAMETER 1
DGN UNT PRM EXIT
F1 F2 F3 F4
FLOW PARAMETER 2
CAL SPC PRM EXIT
i F1 F2 F3 F4
Note: The CAL and SPC function keys will only appear and be accessible from
a Level 2 password.
42
Model FT2A
Operation: Programming
Programming Using Then press PRM (F3). The first parameter is Flow Cutoff:
the Local Display
Flow Cutoff CUTOFF = 2.0 SCFM
CHG OK
F1 F2 F3 F4
Enter the value for the percent low flow cutoff and then press OK (F4).
OPERATION
When the flow rate falls below the zero flow cutoff, the flow meter will display a
flow value of zero.
F1 F2 F3 F4
Enter the pipe area in square meters or square feet and then press OK (F4).
Use square meters for metric flow unit selections and square feet for English
flow unit selections.
Filter Value The Filter Value is also referred to as a dampening factor and is used to quiet
the readings. The filter value is an exponential filter that dampens the noise
and is used as follows:
Flow Value = (FA * new value) + (FB * average)
Where FA = filter value, FA + FB is equal to 1.0.
A lower filter value will increase dampening of the flow rate and smooth the
reading. A lower filter value will also slow the meters response. For example, if
we enter a filter of 0.8, the weight ratio for the new average is:
New average = (80% new sample) + (20% last average)
Filter range is 0.01 to 1.0, 0.01 being a high filter value and 1.0
= no filter.
FILTER = 0.8
CHG OK
F1 F2 F3 F4
43
Model FT2A
Operation: Programming
0.2 0.60
0.1 1.00
0.05 2.00
0.03 3.00
0.01 10.3
High Flow Rate Alarm To set the parameters for a High Flow Rate Alarm:
F1 F2 F3 F4
This is the upper flow limit alarm value that can be associated with a discrete
output. An alarm code is generated when the flow value exceeds this limit. If
no checking is needed, this value should be set to zero.
Low Flow Rate Alarm To set the parameters for a Low Flow Rate Alarm:
F1 F2 F3 F4
This is the lower flow limit alarm value that can be associated with a discrete
output. An alarm code is generated when the flow value is below this limit. If
no checking is needed, this value should be set to zero.
44
Model FT2A
Operation: Programming
F1 F2 F3 F4
This is the upper temperature limit alarm value that can be associated with
a discrete output. An alarm code is generated when the temperature value
OPERATION
exceeds this limit. If no checking is needed, this value should be set to zero.
Press OK (F4) to accept the value.
Low Temp Alarm To set the parameters for a Low Temperature Alarm:
LoTmpAlm = 50 C
CHG OK
F1 F2 F3 F4
This is the lower temperature limit alarm value that can be associated with a
discrete output. An alarm code is generated when the temperature value is
below this limit. If no checking is needed, this value should be set to zero.
i
Note: If the programming menu was entered with a Level 2 password, then
more menus will be shown concerning factory-set parameters that should not
be changed.
This menu allows changing the factory calibrated setting of the flow meter and
is accessible with a Level 2 password. Calibration parameter values are set for
temperature and pressure at 0 degree C and 760 mmHg.
This menu is entered from the base menu and pressing FLO, PRM and CAL.
45
Model FT2A
Operation: Programming
F1 F2 F3 F4
CAL TABLE
TB1 --- --- EXIT
OPERATION
F1 F2 F3 F4
Volt1 = 0.92367
CHG PRV NXT EXIT
F1 F2 F3 F4
Flo1 = 0
CHG PRV NXT EXIT
F1 F2 F3 F4
Use the CHG (F1) key to change the entry, PRV (F2) to move to the previous
entry, NXT (F3) to move to the next entry and EXIT (F4) to return.
Pressing the NXT (F3) key will show the data point voltage, then mass velocity,
and then go to the next data point. The number after Volt (i.e., Volt1) or Flo
(i.e., Flo1) indicated the data point number. The calibration table can hold
up to 20 data pair points. Each data point has a voltage and mass velocity
associated with it.
RESET TOTAL ?
NO YES
F1 F2 F3 F4
46
Model FT2A
Operation: Programming
Programming Using Press YES (F4) to reset total and NO (F1) to cancel.
the Local Display
Totalizer Rollover: The FT2A has an automatic roll-over function. The total
Totalizer Rollover flow count of the FT2A will roll over after the following values:
OPERATION
MSCFD: 999,999,999
MMSCFM: 9,999,999
MMSCFD: 999,999
FLOW PARAMETER 2
CAL SPC PRM EXIT
F1 F2 F3 F4
K fact=0%
CHG OK
F1 F2 F3 F4
RESTORE DATABASE ?
YES NO
F1 F2 F3 F4
Press YES (F1) ONLY if you want to restore your database to the initial factory
setting that the meter was shipped with. All current user-entered settings will
be overwritten.
The green LP1 LED will flash at a faster pace until the recall is performed. The
"RESET CRC" screen will follow "RESTORE DATABASE".
47
Model FT2A
Operation: Programming
RESET CRC?
YES NO
OPERATION
F1 F2 F3 F4
Press YES (F1) ONLY if you want to reset the CRC and generate a new CRC
value.
Simulation Simulation
This menu allows for the simulation of flow rate, temperature and flow input
voltage. It should only be used for testing and demonstration purposes.
Make sure to return all of these simulation values to zero, before returning
to the normal mode of operation.
i Note: Simulated values are only enabled when not set to zero.
The menu is accessible from the main programming menu by pressing FLO,
and DGN (F1):
FLOW PARAMETER 1
DGN UNT PRM EXIT
F1 F2 F3 F4
DIAGNOSTIC
SIM TST EXIT
F1 F2 F3 F4
48
Model FT2A
Operation: Programming
F1 F2 F3 F4
OPERATION
TmpSim = 0 C
CHG OK
F1 F2 F3 F4
CsvSim = 0 V
CHG OK
F1 F2 F3 F4
i Note: This value is used to simulate the Current Sense Voltage (CSV) and
should be set to zero for normal mode.
ENABLE SIM?
YES NO
F1 F2 F3 F4
Press YES (F1) to start the simulation mode, otherwise press NO (F4). Upon
pressing either key, the program will return to the FLOW PARAMETER 1 menu.
49
Model FT2A
Operation:
Operation Menu Tree
Section 1
Main Menu Fig. 4.4: FT2A Menu Tree - Main Menu
SET PARAMETERS
I/O FLO DSP EXIT
OPERATION
Display Menu, p. 54
Flow Parameters 1 Menu, p. 52
SET I/O
COM I/O 420 EXIT
Analog Outputs
Level 2
4 mACnt=367 20mACnt=3748
CHG OK CHG OK
Refer to appropriate Serial
Communication Manual Level 2
50
Model FT2A
Select 1 of 3
methods to scale
OPERATION
High Limit Alarm OUT= FREQUENCY
NXT OK the frequency
output
OUT= HiFloAlm
NXT OK
FREQUENCY OUTPUT
P/U U/P FEQ EXIT
HiFloAlm=500 SCFM
CHG OK
OUT= LoFloAlm
NXT OK MaxFlo=5678.5 SCFM
CHG OK
LoFloAlm=100 SCFM
CHG OK
OUT= HiTmpAlm
NXT OK
HiTmpAlm=300 F
CHG OK
OUT= LoTmpAlm
NXT OK
LoTmpAlm=10 F
CHG OK
51
Model FT2A
FLOW PARAMETER 1
Flow Parameter 2 Menu, p. 53
OPERATION
Simulate CSV?
YES NO
Note: simulation
value needs to be
CsvSim=0 V greater than zero
CHG OK to be taken
(i.e 0.0001 for a
value close to
zero)
ENABLE SIM?
YES NO
www.foxthermalinstruments.com
52
Model FT2A
FLOW PARAMETER 2
CAL SPC PRM EXIT
Level 2 Level 1
OPERATION
Calibration Table Parameters
Flo1 = 0.0
CHG PRV NXT EXIT LoFloAlm=0 SCFM Low Flow Alarm in selected
CHG OK units
velocity in Meter/
hour for point #1 HiTempAlm=0 F High Temperature Alarm in
CHG OK selected units
Volt20 = 1.2479
CHG PRV NXT EXIT Low Temperature Alarm in
LoTempAlm=0 F
CHG OK selected units
Flo20 = 7046.5
CHG PRV NXT EXIT
Cycle through
up to 20 settings
DISPLAY/PASSWORD
DSP PSW EXIT
Display 1
OPERATION
Line 1
FLo rate
Total
DSP1L1=FLo rate PASSWD=1234
Elps NXT OK CHG OK
Display 1 Temp
Line 2 Alarm
FLo rate
Total
DSP1L2=Temp
Elps NXT OK
Display 2 Temp
Line 1 Alarm
FLo rate
Total
DSP2L1=Alarm
Elps NXT OK
Display 2 Temp
Line 2 Alarm
When alternate
FLo rate "ON", flashes
Total
DSP2L2=Elps
Elps NXT OK between the 2
Temp displays
Alarm
On
Off
ALTERNATE=Off
NXT OK
Note:
All readings updated every second
Flo Rate = Flow rate of process gas
Total = Total flow of process gas
Elps = Elapsed time since reset of flow total
Temp = Temperature of process gas
Alarm = High/Low Flow Rate or Temperature Alarm
54
Model FT2A
Operation: Menu Tree
OPERATION
Velocity in selected units Vel=112345.7 FT/M
BrShtDnCnt=0 cnt
Bridge shutdown detection count Display 25
55
Model FT2A
Maintenance: Precautions
Precautions GB
WARNING! BEFORE ATTEMPTING ANY MAINTENANCE, TAKE THE NECESSARY SAFETY
PRECAUTIONS BEFORE REMOVING THE PROBE FROM THE DUCT (EXAMPLE: PURGE
LINES OF TOXIC AND/OR EXPLOSIVE GAS, DEPRESSURIZE, ETC...).
Access to Electronics
Accessing electronics is not normally required for maintenance purposes. If a loose
connection is suspected, unscrew the rear end-cap of the meter enclosure to access the
terminations
Acceso a la Electrnica
Normalmente no se requiere tener acceso a la electrnica para propsitos de mantenimiento.
Si se sospecha de una conexin suelta, desatornille la tapa posterior de la caja del medidor
para tener acceso a las terminales
56
Model FT2A
Maintenance: Precautions
Prcautions FR
AVERTISSEMENT ! AVANT TOUTE TENTATIVE DE MAINTENANCE, OBSERVER LES
CONSIGNES DE SECURITE NECESSAIRES AVANT DE RETIRER LA SONDE DE LA
CONDUITE (PAR EXEMPLE, PURGER LES LIGNES DES GAZ EXPLOSIFS/TOXIQUES
QUELLES POURRAIENT CONTENIR, DEPRESSURISER LE CONTENEUR, ETC.).
MAINTENANCE
AVERTISSEMENT ! RISQUE DEXPLOSION. NE PAS DEBRANCHER UN EQUIPEMENT
DANS UNE AMBIANCE COMBUSTIBLE OU INFLAMMABLE.
57
Model FT2A
Maintenance: General
Broken or Damaged Probe
If the sensor is broken or damaged, the probe and electronics must be returned to the factory.
A new sensor will be installed and calibrated. Refer to "Returning Your Meter" on p. 86.
Fox for a calibration check in our NIST traceable labs after two years of operation.
Fuse Replacement
Verify the fuse is defective by measuring it with an Ohm Meter (Two replacement fuses are
provided with each unit). The fuse F1 is located near the power terminal block and can be
removed by using tweezers or needle-nose pliers. Replacement fuse is Littelfuse part number
0454.750MR.
Warning:
Turn input power OFF before removing or installing a fuse. Use only recommended
fuse replacements.
It is the users responsibility to install the flow meter in an appropriately designed
system with adequate safety protections.
DO NOT remove the flow instrument from the flow body while the system is under flow
conditions.
This product may experience temperatures from as low as -40F (-40C) and as high
as 649F (343C). It is the users responsibility to take safety precautions regarding
operating temperature of the flow meter.
If large flow body has been purchased, user is advised to use a double sling to prevent
rotation of flow body during installation or other method to prevent damage of flow
instrument.
Sensor Wiring
i Note: Sensor terminations are performed at the factory except when the Remote Electronics
option is used or ordered.
Sensor Cleaning
The sensor is insensitive to small amounts of residue, but continued use in dirty environments
will necessitate periodic cleaning. To inspect the sensor, remove power from electronics and
remove the unit from the pipe or duct, exposing the sensor elements. If they are visibly dirty,
clean them with water or alcohol (ethanol) using an appropriate brush until they appear clean
again. Even though the sensor elements are rugged, avoid touching them with any solid object
and use a light touch while cleaning them.
58
Model FT2A
Troubleshooting: General
Troubleshooting Troubleshooting
Caution! The electronics, sensor and sensor interconnect wires supplied
by Fox are calibrated as a single precision mass flow meter. Interchanging
sensors or sensor wiring will affect the accuracy of the flow meter. If you
experience any problem with your Model FT2A Flow meter, call Fox Customer
TROUBLESHOOTING
Service Department, Technical Assistance at (831) 384-4300.
OPERATION
Display Error Display not operating properly. Check status of LP1 on the
display board. Is green LED
blinking once per second?
If LED is not blinking, cycle
power to reset meter. Call
Fox Tech Support.
Meter does not read up to full Calibration table may be Check the calibration table
scale corrupted for a corrupted location.
Enter the password 9111.
Start on p. 42 of the
FT2A manual and follow
the steps to get to Flow
Parameters 2 menu screen.
Select CAL (F1)
Select NXT (F1) to cycle
through calibration table
to verify entries match
calibration certificate.
Check for CRC error code
Velocity measurement seems 1.Probe not oriented properly 1.Orient probe per installation
low sections: Insertion (p. 8),
Inline (p. 12).
59
Model FT2A
Troubleshooting: General
Unit will not power-up a)No power input Check fuse (F1) located
b)Bad fuse next to TS1 on main board.
c)Bad Power supply Check for correct power
TROUBLESHOOTING
Meter does not initialize Electromechanical interference Check meter power cycles.
Press and release F1 and
F2 at the same time; the
display will enter Engineering
screens.
Press F1 to get to screen
#23; record power cycle
value.
Press F4 to return to normal
operation; monitor meter until
problem returns.
Return to screen #23 to
see if power cycles have
increased; microprocessor is
resetting due to EMI electrical
noise entering the meter.
Check Power input and
output cables grounding and
routing.
60
Model FT2A
Troubleshooting: General
Precautions
Problem Possible Cause Action
TROUBLESHOOTING
2.Sensor dirty 2.Clean sensor (Refer to
Maintenance section, p.
58)
61
Model FT2A
Troubleshooting: Installation Problems
The FT2A requires a separate power source for the main board and the two
4 to 20mA output signals. Two wires supply 24VDC power to the main
board. Two wires are used for each of the 4 to 20mA output signals. Refer
to Figure 3.4 and Figure 3.5 (p. 17). Also refer to "Wiring Precautions"
and "Helpful Hints" in Wiring section (p. 14) for further guidance.
62
Model FT2A
Troubleshooting: Installation Problems
TROUBLESHOOTING
4. Erratic flow reading (especially a flow reading spiking high).
This may be a symptom of moisture in the flow stream. Fox flow meters
are designed to work in relatively dry gas applications only. Contact Fox to
discuss resolutions to this problem.
6. Display and/or 4 to 20mA signal reading above zero flow when no flow is
occurring in the pipe.
If the reading is less than 5% of full scale, it is likely this is a normal
condition caused by convection flow created by the heated sensor. It does
not mean that the zero of the instrument is improperly set. The Fox sensor
is extremely sensitive to gas flow and can even read the small flow caused
by convection. If this is an unacceptable condition, please contact Fox
Customer Service for alternatives.
63
Model FT2A
64
Model FT2A
Troubleshooting: Alarm Codes
TROUBLESHOOTING
range limits under FLO Set 4 to 20mA.
33 4 to 20mA for Refer to p. 30 of this Manual. Use the Set
temperature is out of I/O section to verify range limits under FLO Set
range 4 to 20mA. Channel #2 can be set for flow or
temperature. Be sure to check the Alarm limits
(refer to p. 45).
34 Busy Meter is recalculating new parameters.
65
Model FT2A
Appendices: Specifications
Performance Specs
Flow Accuracy:
Inline meter: 1% of reading 0.2% of full scale.
" size: 6 (152 mm) of straight, unobstructed pipe upstream and downstream required.
All other sizes: 8 diameters of straight, unobstructed pipe upstream and 4 downstream required.
Insertion meter: 1% of reading 0.2% of full scale.
15 diameters of straight, unobstructed pipe upstream and 10 downstream required.
APPENDICES
Temperature Accuracy:
1.8 F ( 1.0 C) over the range -40 to 250 F (-40 to 121 C);
3.6 F ( 2.0 C), over the range 250 to 650 F (121 to 343 C);
Minimum velocity 60 SFPM.
Calibration:
Factory Calibration to NIST traceable standards
Operating Specs
Units of Measurement (field selectable):
SCFM, SCFH, NMPS, NM3/M, NM3/H, NM3/D, NLPS, NLPM, NLPH, MCFD, MSCFD, SCFD, MMSCFD,
MMSCFM, SMPS, SM3/D, SM3/H, SM3/M, LB/S, LB/M, LB/H, LB/D, KG/S, KG/M, KG/H, SLPM, SFPM,
MT/H
66
Model FT2A
Appendices: Specifications
Full Scale Flow Ranges for In-Line Flow meters:
Size SCFM NM3/hr
0.25" 0-20 0-32
0.5" 0-90 0-140
0.75" 0-180 0-280
1" 0-320 0-500
1.25" 0-580 0-910
APPENDICES
1.5" 0-840 0-1,320
2" 0-1,400 0-2,200
2.5" 0-2,000 0-3,150
3" 0-3,080 0-4,860
4" 0-5,300 0-8,360
6" 0-12,000 0-18,900
Note: Standard conditions of air at 70F and one atmosphere. Consult factory for other
gases and for flow ranges above those listed.
Relative Humidity:
90% RH maximum; non-condensing
Temperature:
ST sensor: -40 to 250F (-40 to 121C)
HT Sensor: -40 to 650F (-40 to 343C)
Enclosure: -40 to 158F (-40 to 70C) DC power*
-4 to 158F (-20 to 70C) AC power
*Note: Display dims below -4F (-20C), function returns once temperature rises again.
Remote sensor junction box ambient temperature: -40 to 212F (-40 to 100C)
67
Model FT2A
Appendices Specifications
Appendices:
Section 1
Note: Fluctuations of AC and DC power supply are not to exceed 10% of rating.
Outputs:
Two isolated 4 to 20mA outputs (output one is for flow rate and output two is programmable for flow
rate or temperature); fault indication per NAMUR NE43.
APPENDICES
Isolated pulse output 0 to 100Hz, 5 to 24 volts p/p for flow (the pulse output can be used as an
isolated solid state output for alarms); 10mA max.
Serial Communication:
USB connector for connecting to a laptop or computer is standard; free PC-based software tool - FT2A
View - provides complete configuration, remote process monitoring, and data logging functions.
Optional isolated communication outputs: RS485 Modbus, BACnet MS/TP, Profibus-DP, DeviceNet, or
Ethernet Modbus TCP.
Physical Specs
Sensor material:
316 stainless steel standard; Hastelloy C276 optional
Enclosure:
NEMA 4X, aluminum, dual conduit entries with NPT or optional M20 x 1.5mm.
Retractor Assemblies:
Packing gland assembly: 125 psig (8.6 barg ) max.
High pressure (crank) retractor: NPT 600 psig (41.4 barg), ANSI 150 flange & ANSI 300 flange, no
valve supplied.
68
Model FT2A
Appendices:
Appendices Agency Approvals
Section 1
Agency Approvals
CE: Approved
EMC Directive; 2004/108/EC
Emissions and Immunity Testing: EN61326-1:2008
Low Voltage Directive (LVD): 2006/95/EC
Product Safety Testing: EN 61010-1: 2010
Pressure Equipment Directive: 97/23/EC
Weld Testing: EN ISO 15614-1 and EN ISO 9606-1, ASME B31.3
APPENDICES
FM and FMc: Approved
Class I, II, III, Division 2, Groups A, B, C, D, E, F, G, T4A hazardous locations.
NEMA 4X Approved
Note: The EU Pressure Equipment Directive (PED) requires that the minimum ambient and fluid
temperature rating for carbon steel flow bodies not be below -29C.
69
Model FT2A
SWITCH TO CRV 1?
NO YES
F1 F2 F3 F4
PASSWORD:_
UP DN NXT OK
F1 F2 F3 F4
SWITCH TO CRV 1
F1 F2 F3 F4
www.foxthermalinstruments.com
70
Model FT2A
APPENDICES
INP=SWITCH CRV
NXT OK
F1 F2 F3 F4
i
Helpful Hint:
From normal display mode, press F4 to view the current gas curve
selection.
CHG OK
F1 F2 F3 F4
F1 F2 F3 F4
20mA = 500SCFM
CHG OK
F1 F2 F3 F4
F1 F2 F3 F4
F1 F2 F3 F4
APPENDICES
4 maCv2=0
CHG OK
F1 F2 F3 F4
6) Operation:
To avoid confusion, only one of two techniques is enabled. If the
contact input is assigned to switch gas curves, then the ability to switch
using the F2 and F3 function keys on the front panel is disabled.
Two totalizers (Total 1 and Total 2) and two elapsed time counters are
available on the display and through the USB serial communication. The
reset function will reset all totalizers and elapsed time counter to zero.
In the event of a power failure, the software will remember the last curve
in use. Upon powering up again, the FT2A unit will continue to use that
curve.
Switching between gas curves will require a password unless the
password is set to "0", which disables it.
The calibration certificates for flow meters with 2 gas curves will identify
which gas is Gas 1 and Gas 2.
When measuring in mass units, a density value must be entered for
each gas curve.
7.12
APPENDICES
8.44 (18.1)
(21.4)
MODEL TFH1000
Gas Mass Flowmeter & Temperature Transmitter
2X 3/4 NPT,
FEMALE
10.75.25
(27.3.6)
LL
0.87
(2.2)
www.foxthermalinstruments.com
74
Model FT2A
Appendices:
Appendices Dimensions
Section 1
Remote with Fig. 7.2 Remote Insertion Meter with Retractor Dimensions
Retractor
3.3
(8.4)
1.50
5.35 (3.8)
4.4
(11.2) (13.6)
2.0 1.4
(5.1) (3.5)
CONDUIT, 3/4, METAL
APPENDICES
7.12
5.9 2X 3/4 NPT, (18.1)
(15.0) FEMALE
MODEL TFH1000
Gas Mass Flowmeter & Temperature Transmitter
10.75.25
(27.3.6)
LL
0.87
(2.2)
Appendices: Dimensions
Local Insertion Meter Fig. 7.3 Insertion Meter Dimensions
3.9 3.2
(10.0) (8.1)
Input Signal
DIMENSIONS
Power Wiring
7.12
(18.1)
MODEL TFH1000
Gas Mass Flowmeter & Temperature Transmitter
HH
LL
FLOW CL
76
Model FT2A
FT
Appendices: Dimensions
Remote Fig 7.4: Insertion Remote Meter Dimensions
Insertion Meter
3.3
(8.4)
4.4 1.50
(11.2) 3.9 (3.8)
2.0
(5.1) (10.0)
CONDUIT, 3/4, METAL
DIMENSIONS
1.4
Input (3.5)
Power
5.9
(15.0)
7.12
2X 3/4 NPT, (18.1)
FEMALE
MODEL FT2
LL
REMOTE CABLE, 5 CONDUCTOR,
SHIELDED, 100ft (30.48m) MAX
FLOW CL
Table 7.4 Insertion Remote Meter with 316 stainless steel probe
77
Model FT2A
Appendices: Dimensions
Local Inline NPT Meter Fig. 7.5: Inline Meter with 316 Stainless Steel Flow Body and
NPT End Connections Dimensions
3.3
(8.4)
1.50
5.35 (3.8)
(13.6)
1.4
Signal
Input Wiring (3.5)
Power
DIMENSIONS
7.12
(18.1)
2x 3/4 inch NPT,
Female
MODEL FT2A
Gas Mass Flowmeter & Temperature Transmitter
CL
Table 7.5 Inline Meter with 316 stainless steel flow body and NPT End Connections
* Also available in A106 Grade B carbon steel pipe (20PC, 25PC, 30PC, and 40PC model codes)
78
Model FT2A
Appendices: Dimensions
Remote Inline NPT Fig. 7.6: Inline Remote Meter with 316 Stainless Steel Flow
Body and NPT End Connections Dimensions
3.3
(8.4)
1.50
5.35 (3.8)
(13.6)
4.4
DIMENSIONS
(11.2) 1.4
2.0 Input (3.5)
(5.1)
Power
(15.0)
HH
CL
Table 7.6 Inline Remote Meter with 316 stainless steel flow body and NPT End Connections
* Also available in A106 Grade B carbon steel pipe (20PC, 25PC, 30PC, and 40PC model codes)
79
Model FT2A
Appendices: Dimensions
Local Inline Fig. 7.7: Inline Meter with 316 Stainless Steel Flow Body and
Flange Meter 150# RF Flange End Connections Dimensions
3.3
(8.4)
1.50
(3.8)
3.9
(10.0)
1.4
DIMENSIONS
7.12
2X 3/4 NPT, (18.1)
FEMALE
MODEL FT2
Gas Mass Flowmeter & Temperature Transmitter
CL
L
2X Flange, Raised Face, ANSI B16.5, 316 SST
Table 7.7 Inline Meter with 316 stainless steel flow body and 150# RF Flange End Connections
* Also available in A106 Grade B carbon steel pipe with A105 flanges (20FC, 25FC, 30FC, and 40FC model codes)
80
Model FT2A
Appendices: Dimensions
Remote Inline Fig. 7.8: Inline Remote Meter with Stainless Steel Flow Body
Flange Meter and 150# RF Flange End Connections Dimensions
3.3
(8.4)
1.50
5.35 (3.8)
4.4 (13.6)
(11.2)
2.0 1.4
(5.1) Input (3.5)
DIMENSIONS
Power
CONDUIT, 3/4, METAL
7.12
2X 3/4 NPT, (18.1)
FEMALE
5.9
(15.0) MODEL FT2
Gas Mass Flowmeter & Temperature Transmitter
HH
CL
Table 7.8 Inline Remote Meter with stainless steel flow body and 150# RF Flange End Connections
* Also available in A106 Grade B carbon steel pipe with A105 flanges (20FC, 25FC, 30FC, and 40FC model codes)
81
Model FT2A
5.35
(13.6)
6.06
7.12 (15.4)
(18.1)
Figure 7.9 shows the rear view of the FT2A enclosure without sensor
or conduit. AC power enters on one side, DC power and signals enter
on either side.
www.foxthermalinstruments.com
82
Model FT2A
APPENDICES
Appendices: Warranty
Warranty Warranty
(a) FOX warrants that the products furnished under this Agreement will be
free from defects in material and workmanship for a period of one year from
the date of shipment. The customer shall provide notice of any defect to FOX,
within one week after the Customers discovery of such defect. The sole
obligation and liability of FOX, under this warranty shall be repair or replace, at
its option, without cost to the Customer, the defective product or part.
APPENDICES
(b) Upon request by FOX, the product or part claimed to be defective shall
immediately be returned at the Customers expense to FOX. Replaced or
repaired products or parts will be shipped to the Customer at the expense of
FOX. FOX shall have the right of final determination as to the existence and
cause of defect.
(c) There shall be no warranty or liability for any products or parts that have
been subject to misuse, accident, negligence, failure of electric power or
modifications by the Customer without the written approval of FOX. Final
determination of warranty eligibility shall be made by FOX. If a warranty claim
is considered invalid for any reason, the Customer will be charged for services
performed and expenses incurred by FOX, in handling and shipping the
returned unit.
(d) The liability of FOX shall be limited to replacing or repairing, at its option,
any defective parts which are returned. Labor and related expenses incurred to
install replacement parts are not covered by this warranty.
(e) As to replacement parts supplied or repairs made during the original
warranty period, the warranty period for the replacement or repaired part shall
terminate with the termination of the warranty period of the original product or
part.
(f) The use of these products is under exclusive control of the purchaser and
FOX specifically denies any responsibility for the calibration of units and/
or accuracy of work performed or the safety of the system in which FOX
products is used. EXTERNAL SAFETY DEVICES MUST BE USED WITH THIS
EQUIPMENT.
84
Model FT2A
Appendices: Warranty
APPENDICES
LIABILITIES, AND REMEDIES. EXCEPT AS THUS PROVIDED, FOX, DISCLAIMS
ALL WARRANTIES, EXPRESS OR IMPLIED, INCLUDING ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
85
Model FT2A
i
Note: Be sure to review all of the information on the Customer Information
Form before sending your meter to the Fox Customer Service Department. The
Fox Shipping/Receiving Department cannot accept meters that have not been
prepared appropriately.
www.foxthermalinstruments.com
86
Model FT2A
Disclaimer
Section 1
What to Expect What to expect while your meter is being serviced
During Servicing Depending on the type of service required when returning your Fox meter, there
are varying turnover times for servicing a meter. The average time needed
to service the meter is 7-10 days (not including shipping or peak production
times).
If you have already shipped your meter to Fox for servicing and would like to
check the status of your meter, please fill out our online Service Order Status
DISCLAIMER
form located at www.foxthermalinstruments.com and you will hear from a
Customer Service Rep within 1 business day of your requested update.
Definitions Aa
Glossary of Terms Glossary of Terms and Definitions
and Definitions
AC Alternating Current
AWG American Wire Gauge
Bara Bar absolute
CTC Contact
CAL Calibration
CHG Change
DEFINITIONS
COM Communication
CSV Current Sense Voltage
DC Direct Current
DN Down
DSP Display
ELP Elapsed time
Feq Frequency
Ft^2 Square Feet
I/O Input/Output
INP Input
IR Infrared (IR Buttons = optical switches)
LB Pound
LB/D Pound per Day
LB/H Pound per Hour
LB/M Pound per Minute
LB/S Pound per Second
LCD Liquid Crystal Display
KG Kilogram
KG/H Kilogram per Hour
KG/M Kilogram per Minute
KG/S Kilogram per Second
M^2 Square Meter
mmHG Pressure in millimeters of mercury
MMSCFD Million Standard Cubic Feet per Day
MXFLO Maximum Flow
NEMA National Electrical Manufactures Association
NIST National Institute of Standards and Technology
NLP Normal Liter
NLPH Normal Liter per Hour
NLPM Normal Liter per Minute
NM3 Normal cubic Meter
88
Model FT2A
Definitions Aa
Glossary of Terms NM3/H Normal cubic Meter per Hour
and Definitions NM3/M Normal cubic Meter per Minute
NPT National Pipe Thread
PDA Personal hand held computer
UNT Unit
U/P Unit per Pulse
PC Personal Computer
P/U Pulse per Unit
DEFINITIONS
PIP A^2 Pipe Area
PLC Programmable Logic Controller
PRM Parameters
PRS Pressure
PSIA Pounds per Square Inch Absolute
Pt Point
PSW Password
SIM Simulation
SCF Standard Cubic Feet
SCFM Standard Cubic Feet per Minute
SCFH Standard Cubic Feet per Hour
SCFD Standard Cubic Feet per Day
SPC Special Control
STP Standard Temperature and Pressure
TMP Temperature
TSI Internal Variable
TSV Internal Variable
420 4 to 20mA output
89
Model FT2A
Index
Index Introduction, p. 4
Level 2 password, p. 38
Access to Electronics, p. 14 Local Display, p. 6
Alarm Codes, p. 64 Local Input Wiring, p. 17
Alarm wiring, p. 21 Mass Flow, p. 4
Analog 4-20mA outputs, p. 30 Measurement Mode, p. 27
Breakage or Damage of Probe, p. 58 Menu Tree,
Calibration Parameters, p. 45 Digital Output, p. 51
Dimension Details, Display Menu, p. 54
Local Inline Flange, p. 80 Engineering Display, p. 55
Local Inline NPT, p. 78 Main Menu, p. 50
Local Insertion, p. 76 Parameter Menu 1, p. 52
INDEX
90
Model FT2A
Index
Programming (cont'd)
Serial Communication, p. 36
Simulation, p. 48
Unit-per-pulse, p. 34
Unit settings, p. 39
Using the Local Display, p. 29
Remote Wiring Installation, p. 24
Replacements Fuses, p. 58
Reset Total and Elapsed Time, p. 46
Restore Database, p. 47
Return Procedure, p. 86
Scope, p. 7
Sensor Cleaning, p. 58
INDEX
Sensor Orientation,
Inline, p. 13
Insertion, p. 10
Signal wiring, p. 19
Simulation Mode, p. 48
Theory of Operation, p. 4
To program the display, p. 37
Totalizer Rollover, p. 47
Troubleshooting, p. 63
Alarm Codes, p. 64
General, p. 60
Installation Problems, p. 63
Unequal length sensors, p. 10
USB Interface, p. 26
Warranty, p. 84
Wiring
Alarms, p. 21
Frequency/Alarm Outputs, p. 21
Grounding, p. 14
Input - Local, p. 17
Input - Remote, p. 24
Instructions, p. 14
Local meters, p. 17
Power Input, p. 17
Precautions, p. 14
Remote meters, p. 24
91
Wiring
Definition of Terms
Aa
Troubleshooting Tips
i Information